Data Quality Reports for Session: 142420 User: christoph5 Completed: 05/15/2012


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D060928.4NSA/SONDE/C1 - Increased precision for alt, dp, pres, rh, tdrynsasondewnpnC1.b1
D061011.1NSA/SONDE/C1 - Incorrect surface values of T and RHnsasondewnpnC1.b1
D061114.3NSA/SONDE/C1 - RH sensor failurensasondewnpnC1.b1
D070124.1NSA/SONDE/C1 - Incorrect surface RH valuensasondewnpnC1.b1
D070302.1NSA/SONDE/C1 - Temperature sensor broke at launchnsasondewnpnC1.b1


DQRID : D060928.4
Start DateStart TimeEnd DateEnd Time
09/29/2006050012/31/99992359
Subject:
NSA/SONDE/C1 - Increased precision for alt, dp, pres, rh, tdry
DataStreams:nsasondewnpnC1.b1
Description:
The output precision of the following variables was increased 20060929:

   alt	changed from 1m to 0.1m
   dp	changed from 0.1degC to 0.01degC
   pres changed from 0.1 hPa to 0.01 hPa
   RH	changed from 1%RH to 0.01%RH
   tdry changed from .1degC to 0.01degC

Note that the increase in precision does not necessarily imply an increase
in accuracy.  Users are cautioned to continue to use their best scientific
judgement when using these or any other data.
Measurements:nsasondewnpnC1.b1:
  • alt
  • Dry Bulb Temperature(tdry)
  • Dewpoint Temperature(dp)
  • Humidity, relative(rh)
  • Pressure, atmospheric, constant value of model level, RUC model output(pres)


Back To Table of Contents

DQRID : D061011.1
Start DateStart TimeEnd DateEnd Time
10/07/2006052410/07/20060525
Subject:
NSA/SONDE/C1 - Incorrect surface values of T and RH
DataStreams:nsasondewnpnC1.b1
Description:
The radiosonde was launched too quickly after being removed from the ground check set.  As 
a result, the system did not equilibrate with the local ground conditions and the 
automatically entered surface values are incorrect.
Measurements:nsasondewnpnC1.b1:
  • Dry Bulb Temperature(tdry)
  • Dewpoint Temperature(dp)
  • Humidity, relative(rh)


Back To Table of Contents

DQRID : D061114.3
Start DateStart TimeEnd DateEnd Time
10/28/2006053210/28/20060710
Subject:
NSA/SONDE/C1 - RH sensor failure
DataStreams:nsasondewnpnC1.b1
Description:
It appears that one of the two RH sensors on this radiosonde was affected by ice after 
emerging from a low cold cloud.  As a result the reported RH values oscillate by about 10 
percent until the alternate heating cycle stops.  I suspect the lower value is more close 
to correct, but can't be absolutely sure.
Measurements:nsasondewnpnC1.b1:
  • Dewpoint Temperature(dp)
  • Humidity, relative(rh)


Back To Table of Contents

DQRID : D070124.1
Start DateStart TimeEnd DateEnd Time
12/29/2006051712/29/20060518
Subject:
NSA/SONDE/C1 - Incorrect surface RH value
DataStreams:nsasondewnpnC1.b1
Description:
For an unknown reason the surface RH was reported a 0%.  Becuase these values are entered 
automatically by the system, this is NOT likely an operator error.  Examination of the 
raw data shows that the surface RH at launch time was approximately 81.2 %RH.
Measurements:nsasondewnpnC1.b1:
  • Dewpoint Temperature(dp)
  • Humidity, relative(rh)


Back To Table of Contents

DQRID : D070302.1
Start DateStart TimeEnd DateEnd Time
02/27/2007195502/27/20072131
Subject:
NSA/SONDE/C1 - Temperature sensor broke at launch
DataStreams:nsasondewnpnC1.b1
Description:
The temperature sensor on this sonde broke at launch.  As a result all the thermodynamic 
data are incorrect.  The wind data are OK, though there will be some uncertainty in the 
altitude values.  This sounding was done at a non-standard time and it isn't clear whether 
it was delayed for some reason or a RHUBC sounding.  In either case, there is no fix.
Measurements:nsasondewnpnC1.b1:
  • alt
  • Dry Bulb Temperature(tdry)
  • Dewpoint Temperature(dp)
  • Humidity, relative(rh)
  • asc


Back To Table of Contents