Data Quality Reports for Session: 142391 User: lijun1ljr Completed: 05/11/2012


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D090211.3HFE/AOS/M1 - No sub 1 micron datahfeaosM1.a1
D090211.4HFE/AOS/M1 - Humidifier failurehfeaosM1.a1
D090211.5HFE/AOS/M1 - CPC data missinghfeaosM1.a1
D090211.6HFE/AOS/M1 - CCN missinghfeaosccnM1.a1
D090211.7HFE/AOS/M1 - PSAP data missinghfeaosM1.a1
D101020.3HFE/AOS/M1 - Data processing problemhfeaosM1.a1, hfenoaaaosM1.b1, hfenoaaaosavgM1.b1
D111122.2HFE/AOS/M1 - CPC and CCN data questionablehfeaosccnM1.a1, hfenoaaaosM1.b1


DQRID : D090211.3
Start DateStart TimeEnd DateEnd Time
06/06/2008171707/24/20080000
Subject:
HFE/AOS/M1 - No sub 1 micron data
DataStreams:hfeaosM1.a1
Description:
The aerosol size switching impactors were stuck in the sub 10 micron mode. So all aerosol 
is sub 10 um.
Measurements:hfeaosM1.a1:
  • Absorption Coefficient Red, corrected for flow and sample area, 3 wavelength
    PSAP, 10 um size cut(Ba_R_Dry_1um_PSAP3W_1)
  • Aerosol backwards-hemispheric light scattering coefficient, Humidograph
    nephelometer blue channel, 1(Bbs_B_Wet_1um_Neph3W_2)
  • Aerosol backwards-hemispheric light scattering coefficient, Ref. Humidograph
    nephelometer blue chann(Bbs_B_Dry_1um_Neph3W_1)
  • Aerosol total light scattering coefficient, Humidograph nephelometer blue
    channel, 1 um particle dia(Bs_B_Wet_1um_Neph3W_2)
  • Absorption Coefficient Green, corrected for flow and sample area, 3 wavelength
    PSAP, 1 um size cut(Ba_G_Dry_1um_PSAP3W_1)
  • Aerosol total light scattering coefficient, Ref. Humidograph nephelometer red
    channel, 1 um particle(Bs_R_Dry_1um_Neph3W_1)
  • Aerosol total light scattering coefficient, Ref. Humidograph nephelometer blue
    channel, 1 um particl(Bs_B_Dry_1um_Neph3W_1)
  • Aerosol backwards-hemispheric light scattering coefficient, Ref. Humidograph
    nephelometer red channe(Bbs_R_Dry_1um_Neph3W_1)
  • Aerosol backwards-hemispheric light scattering coefficient, Humidograph
    nephelometer green channel,(Bbs_G_Wet_1um_Neph3W_2)
  • Aerosol backwards-hemispheric light scattering coefficient, Humidograph
    nephelometer red channel, 1(Bbs_R_Wet_1um_Neph3W_2)
  • Aerosol total light scattering coefficient, Humidograph nephelometer red
    channel, 1 um particle diam(Bs_R_Wet_1um_Neph3W_2)
  • Absorption Coefficient Blue, corrected for flow and sample area, 3 wavelength
    PSAP, 1 um size cut(Ba_B_Dry_1um_PSAP3W_1)
  • Aerosol light absorption coefficient, green channel, 1 um particle diameter(Ba_G_Dry_1um_PSAP1W_1)
  • Aerosol total light scattering coefficient, Ref. Humidograph nephelometer green
    channel, 1 um partic(Bs_G_Dry_1um_Neph3W_1)
  • Aerosol total light scattering coefficient, Humidograph nephelometer green
    channel, 1 um particle di(Bs_G_Wet_1um_Neph3W_2)
  • Aerosol backwards-hemispheric light scattering coefficient, Ref. Humidograph
    nephelometer green chan(Bbs_G_Dry_1um_Neph3W_1)


Back To Table of Contents

DQRID : D090211.4
Start DateStart TimeEnd DateEnd Time
05/19/2008000005/29/20080429
Subject:
HFE/AOS/M1 - Humidifier failure
DataStreams:hfeaosM1.a1
Description:
The humidifier either wasn't scanning or the wet nephelometer was flooded with water.
Measurements:hfeaosM1.a1:
  • Aerosol backwards-hemispheric light scattering coefficient, Humidograph
    nephelometer green channel,(Bbs_G_Wet_10um_Neph3W_2)
  • Aerosol total light scattering coefficient, Humidograph nephelometer green
    channel, 10 um particle d(Bs_G_Wet_10um_Neph3W_2)
  • Aerosol backwards-hemispheric light scattering coefficient, Humidograph
    nephelometer green channel,(Bbs_G_Wet_1um_Neph3W_2)
  • Aerosol backwards-hemispheric light scattering coefficient, Humidograph
    nephelometer blue channel, 1(Bbs_B_Wet_1um_Neph3W_2)
  • Aerosol backwards-hemispheric light scattering coefficient, Humidograph
    nephelometer red channel, 1(Bbs_R_Wet_1um_Neph3W_2)
  • Aerosol total light scattering coefficient, Humidograph nephelometer blue
    channel, 10 um particle di(Bs_B_Wet_10um_Neph3W_2)
  • Aerosol backwards-hemispheric light scattering coefficient, Humidograph
    nephelometer blue channel, 1(Bbs_B_Wet_10um_Neph3W_2)
  • Aerosol backwards-hemispheric light scattering coefficient, Humidograph
    nephelometer red channel, 10(Bbs_R_Wet_10um_Neph3W_2)
  • Aerosol total light scattering coefficient, Humidograph nephelometer blue
    channel, 1 um particle dia(Bs_B_Wet_1um_Neph3W_2)
  • Aerosol total light scattering coefficient, Humidograph nephelometer red
    channel, 1 um particle diam(Bs_R_Wet_1um_Neph3W_2)
  • Aerosol total light scattering coefficient, Humidograph nephelometer red
    channel, 10 um particle dia(Bs_R_Wet_10um_Neph3W_2)
  • Aerosol total light scattering coefficient, Humidograph nephelometer green
    channel, 1 um particle di(Bs_G_Wet_1um_Neph3W_2)


Back To Table of Contents

DQRID : D090211.5
Start DateStart TimeEnd DateEnd Time
05/19/2008000007/22/20080450
Subject:
HFE/AOS/M1 - CPC data missing
DataStreams:hfeaosM1.a1
Description:
The CPC wasn't installed until July 22nd when the mentor visited the site.
Measurements:hfeaosM1.a1:
  • Condensation Nuclei Concentration Number(N_CPC_1)


Back To Table of Contents

DQRID : D090211.6
Start DateStart TimeEnd DateEnd Time
07/08/2008032007/25/20080429
Subject:
HFE/AOS/M1 - CCN missing
DataStreams:hfeaosccnM1.a1
Description:
A power regulator on the CCN circuit board failed. The mentor replaced it and it was up 
and working on 7/25/2008.
Measurements:hfeaosccnM1.a1:
  • CCN_laser_current
  • AOS bin number of lowest channel of OPC include in summation of N_CCN(N_CCN_bin_number)
  • CCN_FirstStageMonitorVoltage
  • CCN_ss_calc
  • CCN_T_sample
  • CCN_T_TEC1
  • CCN_T_TEC3
  • CCN_T_TEC2
  • CCN_Q_sheath
  • CCN_T_nafion
  • AOS number concentration of CCN(N_CCN)
  • CCN_P_sample
  • CCN_dT_TEC3_TEC1_StdDev
  • CCN_T_inlet
  • Size of upper limit of each CCN bin(CCN_UpperSizeLimit)
  • CCN_ProportionalValveVoltage
  • CCN_temp_unstable_flag
  • CCN_Q_sample
  • CCN_Temperature_Gradient
  • AOS CCN number concentration by bin(N_CCN_dN)
  • CCN_T_OPC
  • CCN_ss_set


Back To Table of Contents

DQRID : D090211.7
Start DateStart TimeEnd DateEnd Time
08/23/2008015512/28/20082300
Subject:
HFE/AOS/M1 - PSAP data missing
DataStreams:hfeaosM1.a1
Description:
The PSAP was disconnected during this time period.
Measurements:hfeaosM1.a1:
  • Absorption Coefficient Red, corrected for flow and sample area, 3 wavelength
    PSAP, 10 um size cut(Ba_R_Dry_1um_PSAP3W_1)
  • Aerosol light absorption coefficient, green channel, 10 um particle diameter(Ba_G_Dry_10um_PSAP1W_1)
  • Absorption Coefficient Blue, corrected for flow and sample area, 3 wavelength
    PSAP, 10 um size cut(Ba_B_Dry_10um_PSAP3W_1)
  • Absorption Coefficient Red, corrected for flow and sample area, 3 wavelength
    PSAP, 10 um size cut(Ba_R_Dry_10um_PSAP3W_1)
  • Absorption Coefficient Green, corrected for flow and sample area, 3 wavelength
    PSAP, 10 um size cut(Ba_G_Dry_10um_PSAP3W_1)
  • Aerosol light absorption coefficient, green channel, 1 um particle diameter(Ba_G_Dry_1um_PSAP1W_1)
  • Absorption Coefficient Blue, corrected for flow and sample area, 3 wavelength
    PSAP, 1 um size cut(Ba_B_Dry_1um_PSAP3W_1)
  • Absorption Coefficient Green, corrected for flow and sample area, 3 wavelength
    PSAP, 1 um size cut(Ba_G_Dry_1um_PSAP3W_1)


Back To Table of Contents

DQRID : D101020.3
Start DateStart TimeEnd DateEnd Time
05/09/2008000012/28/20080000
Subject:
HFE/AOS/M1 - Data processing problem
DataStreams:hfeaosM1.a1, hfenoaaaosM1.b1, hfenoaaaosavgM1.b1
Description:
The AOS data was reprocessed for the AMF, SGP and NSA sites. There was an error in the 
processing of the corrections for the PSAP absorption signal such that a low pass filter was 
implemented for the scattering data used in these corrections.The difference between the 
corrected and previous data is < 0.05 Mm-1 (below the instrument detection limit) and 
should not have a significant effect on the previous data. Data will be reprocessed at a 
later time when more significant changes to the data processing are implemented.
Measurements:hfenoaaaosM1.b1:
  • Absorption Coefficient Blue, corrected for flow and sample area, 3 wavelength
    PSAP, 1 um size cut(Ba_B_Dry_1um_PSAP3W_1)
  • Absorption Coefficient Red, corrected for flow and sample area, 3 wavelength
    PSAP, 10 um size cut(Ba_R_Dry_1um_PSAP3W_1)
  • Absorption Coefficient Red, corrected for flow and sample area, 3 wavelength
    PSAP, 10 um size cut(Ba_R_Dry_10um_PSAP3W_1)
  • Aerosol light absorption coefficient, green channel, 10 um particle diameter(Ba_G_Dry_10um_PSAP1W_1)
  • Absorption Coefficient Green, corrected for flow and sample area, 3 wavelength
    PSAP, 1 um size cut(Ba_G_Dry_1um_PSAP3W_1)
  • Absorption Coefficient Green, corrected for flow and sample area, 3 wavelength
    PSAP, 10 um size cut(Ba_G_Dry_10um_PSAP3W_1)
  • Aerosol light absorption coefficient, green channel, 1 um particle diameter(Ba_G_Dry_1um_PSAP1W_1)
  • Absorption Coefficient Blue, corrected for flow and sample area, 3 wavelength
    PSAP, 10 um size cut(Ba_B_Dry_10um_PSAP3W_1)

hfeaosM1.a1:
  • Aerosol light absorption coefficient, green channel, 10 um particle diameter(Ba_G_Dry_10um_PSAP1W_1)
  • Absorption Coefficient Red, corrected for flow and sample area, 3 wavelength
    PSAP, 10 um size cut(Ba_R_Dry_1um_PSAP3W_1)
  • Absorption Coefficient Green, corrected for flow and sample area, 3 wavelength
    PSAP, 10 um size cut(Ba_G_Dry_10um_PSAP3W_1)
  • Absorption Coefficient Green, corrected for flow and sample area, 3 wavelength
    PSAP, 1 um size cut(Ba_G_Dry_1um_PSAP3W_1)
  • Absorption Coefficient Blue, corrected for flow and sample area, 3 wavelength
    PSAP, 10 um size cut(Ba_B_Dry_10um_PSAP3W_1)
  • Absorption Coefficient Red, corrected for flow and sample area, 3 wavelength
    PSAP, 10 um size cut(Ba_R_Dry_10um_PSAP3W_1)
  • Absorption Coefficient Blue, corrected for flow and sample area, 3 wavelength
    PSAP, 1 um size cut(Ba_B_Dry_1um_PSAP3W_1)
  • Aerosol light absorption coefficient, green channel, 1 um particle diameter(Ba_G_Dry_1um_PSAP1W_1)

hfenoaaaosavgM1.b1:
  • Aerosol light absorption coefficient, green channel, 1 um particle diameter(Ba_G_Dry_1um_PSAP1W_1)
  • Aerosol light absorption coefficient, green channel, 10 um particle diameter(Ba_G_Dry_10um_PSAP1W_1)


Back To Table of Contents

DQRID : D111122.2
Start DateStart TimeEnd DateEnd Time
07/25/2008000012/28/20082300
Subject:
HFE/AOS/M1 - CPC and CCN data questionable
DataStreams:hfeaosccnM1.a1, hfenoaaaosM1.b1
Description:
The CPC slowly clogged and the signal declined until the instrument was cleaned at which 
point the signal would increase. The data is still a good relative measure of the daily 
changes in the CN concentration, but the absolute value may be too low.

The mentor was unable to monitor the flow rate and operation in the CCN during operation. 
The relative changes in CCN with %ss are likely fine, but the absolute values may be too 
low or too high. Dust was prevalent at the site and likely contributed to the CCN signal 
such that large dust particles appeared as droplets.
Measurements:hfenoaaaosM1.b1:
  • AOS number concentration of CCN(N_CCN)
  • Condensation Nuclei Concentration Number(N_CPC_1)

hfeaosccnM1.a1:
  • CCN_Q_sheath
  • AOS number concentration of CCN(N_CCN)
  • CCN_Q_sample
  • AOS CCN number concentration by bin(N_CCN_dN)


Back To Table of Contents



END OF DATA