Data Quality Reports for Session: 141691 User: daniel1per Completed: 04/04/2012


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D050725.3SGP/MWR/B4 - Reprocess: Revised Retrieval Coefficientssgp5mwravgB4.c1, sgpmwrlosB4.a1, sgpmwrlosB4.b1, sgpmwrtipB4.a1, sgpqmemwrcolB4.c1
D951005.2SGP/5MWRAVG/B1/B4/B5 - Valid LWP > 1 mm excluded from 5 min avgssgp5mwravgB1.c1, sgp5mwravgB4.c1, sgp5mwravgB5.c1
D960404.12SGP/MWR/B1/B4/B5 - Reprocess: Error in MWR calibrationsgp5mwravgB1.c1, sgp5mwravgB4.c1, sgp5mwravgB5.c1
D960404.8SGP/MWR/B1/B4/B5 - Reprocess: MWR Tuning Functionssgp5mwravgB4.c1, sgp5mwravgB5.c1, sgp5mwravgB1.c1
D961117.12Loss of thermal stabilizationsgp5mwravgB4.c1
D961117.16Loss of thermal stabilizationsgp5mwravgB5.c1, sgp5mwravgB1.c1, sgp5mwravgB4.c1
D961117.17Loss of thermal stabilizationsgp5mwravgB4.c1
D961120.1SGP/MWR/B1/B4/B5/B6/C1 - Thermal Stabilization Adjustmentsgp1mwravgC1.c1, sgp5mwravgB1.c1, sgp5mwravgB4.c1, sgp5mwravgB5.c1, sgp5mwravgB6.c1,
sgp5mwravgC1.c1, sgpmwrlosB1.a0, sgpmwrlosB1.a1, sgpmwrlosB4.a0, sgpmwrlosB4.a1,
sgpmwrlosB5.a0, sgpmwrlosB5.a1, sgpmwrlosB6.a0, sgpmwrlosB6.a1, sgpmwrlosC1.a1, sgpmwrlosC1.b1
D991102.1SGP/MWR/B4 - power lines in field of viewsgpmwrlosB4.a1, sgpmwrlosB4.b1, sgpmwrtipB4.a1, sgp5mwravgB4.c1, sgpmwrlosB4.00,
sgpmwrtipB4.00


DQRID : D050725.3
Start DateStart TimeEnd DateEnd Time
04/15/2002230006/24/20052100
Subject:
SGP/MWR/B4 - Reprocess: Revised Retrieval Coefficients
DataStreams:sgp5mwravgB4.c1, sgpmwrlosB4.a1, sgpmwrlosB4.b1, sgpmwrtipB4.a1, sgpqmemwrcolB4.c1
Description:
IN THE BEGINNING (June 1992), the retrieval coefficients used to derive the precipitable 
water vapor (PWV) and liquid water path (LWP) from the MWR brightness temperatures were 
based on the Liebe and Layton (1987) water vapor and oxygen absorption model and the Grant 
(1957) liquid water absorption model.

Following the SHEBA experience, revised retrievals based on the more recent Rosenkranz 
(1998) water vapor and oxygen absorption models and the Liebe (1991) liquid waer absorption 
model were developed.  The Rosenkranz water vapor absorption model resulted a 2 percent 
increase in PWV relative to the earlier Liebe and Layton model.  The Liebe liquid water 
absorption model decreased the LWP by 10% relative to the Grant model.  However, the 
increased oxygen absorption caused a 0.02-0.03 mm (20-30 g/m2) reduction in LWP, which was 
particularly significant for low LWP conditions (i.e. thin clouds encountered at SHEBA).

Recently, it has been shown (Liljegren, Boukabara, Cady-Pereira, and Clough, TGARS v. 43, 
pp 1102-1108, 2005) that the half-width of the 22 GHz water vapor line from the HITRAN 
compilation, which is 5 percent smaller than the Liebe and Dillon (1969) half-width used in 
Rosenkranz (1998), provided a better fit to the microwave brightness temperature 
measurements at 5 frequencies in the range 22-30 GHz, and yielded more accurate retrievals. 
Accordingly, revised MWR retrieval coefficients have been developed using MONORTM, which 
utilizes the HITRAN compilation for its spectroscopic parameters.  These new retrievals 
provide 3 percent less PWV and 2.6 percent greater LWP than the previous retrievals based on 
Rosenkranz (1998).

Although the MWR data will be reprocessed to apply the new monortm-based retrievals, for 
most purposes it will be sufficient to correct the data using the following factors:

PWV_MONORTM = 0.9695 * PWV_ROSENKRANZ
LWP_MONORTM = 1.026  * LWP_ROSENKRANZ

The Rosenkranz-based retrieval coefficients became active at SGP.B4 20020415.2300.  The 
MONORTM-based retrieval coefficients became active at SGP.B4 20050624.2100.

Note: a reprocessing effort is already underway to apply the Rosenkranz-based retrieval 
coefficients to all MWR prior to April 2002.  An additional reprocessing task will be 
undertaken to apply the MONORTM retrieval to all MWR data when the first is completed. Read 
reprocessing comments in the netcdf file header carefully to ensure you are aware which 
retrieval is in play.
Measurements:sgp5mwravgB4.c1:
  • Total water vapor along LOS path(vap)
  • Total liquid water along LOS path(liq)

sgpmwrtipB4.a1:
  • vaptip
  • liqtip

sgpqmemwrcolB4.c1:
  • mean_vap_mwr
  • mean_liq_mwr

sgpmwrlosB4.b1:
  • Total water vapor along LOS path(vap)
  • Total liquid water along LOS path(liq)

sgpmwrlosB4.a1:
  • Total water vapor along LOS path(vap)
  • Total liquid water along LOS path(liq)


Back To Table of Contents

DQRID : D951005.2
Start DateStart TimeEnd DateEnd Time
03/01/1994000010/12/19952359
Subject:
SGP/5MWRAVG/B1/B4/B5 - Valid LWP > 1 mm excluded from 5 min avgs
DataStreams:sgp5mwravgB1.c1, sgp5mwravgB4.c1, sgp5mwravgB5.c1
Description:
Note: These data have not been and will not be reprocessed.  The MWRAVG VAP has been 
retired.

The limit of maximum valid liquid water path was set at 1 mm.  Although this limit was 
selected 'conservatively' so as to definitely flag precipitation-contaminated data in the 
20-second (sgpmwrlos) files, the effect has been to exclude valid liquid water paths 
greater than 1 mm from the 5-minute averages (sgp5mwravg).

The following actions are recommended:
1) the maximum limits for precipitable water vapor (PWV) and liquid water path (LWP) be 
removed and, 2) the averaging algorithm instead exclude data on the basis of the brightness 
temperature flags.  These flags are set below a minimum of 2.75 K (cosmic background) 
and above a maximum of 100 K (precipitation).
Measurements:sgp5mwravgB4.c1:
  • liq_sdev
  • Total water vapor along LOS path(vap)
  • Total liquid water along LOS path(liq)
  • vap_sdev

sgp5mwravgB1.c1:
  • Total water vapor along LOS path(vap)
  • Total liquid water along LOS path(liq)
  • vap_sdev
  • liq_sdev

sgp5mwravgB5.c1:
  • Total water vapor along LOS path(vap)
  • vap_sdev
  • Total liquid water along LOS path(liq)
  • liq_sdev


Back To Table of Contents

DQRID : D960404.12
Start DateStart TimeEnd DateEnd Time
03/01/1994000504/04/19962359
Subject:
SGP/MWR/B1/B4/B5 - Reprocess: Error in MWR calibration
DataStreams:sgp5mwravgB1.c1, sgp5mwravgB4.c1, sgp5mwravgB5.c1
Description:
The effect of this error is small.  At most, it results in a bias of
about -0.015 cm in precipitable water vapor and -0.015 mm in liquid
water path during clear sky conditions.  The error is largest when the
brightness temperatures are small (i.e.  clear skies and low PWV).

The error results from failing to correctly account for the effect of
the Teflon window covering the radiometer mirror.  Although the
contribution of the window is subtracted when the tip curve data are
reduced to determine the true zenith brightness temperature, it is not
added back in when the zenith brightness temperature is used to
calibrate the noise diode.  This would still not be a problem if the
contribution of the window where not subtracted (again) during zenith
line-of-sight (LOS) operations.  But it is and the net effect is to
subtract the contribution of the window twice.

The calibrations ('Noise Injection Temperatures') are off by a factor
of 1.00164 and 1.00217 for the 23.8 and 31.4 GHz frequencies,
respectively.

The magnitude of the error is equal to the emissivity of the window
multiplied by the difference between the brightness temperature and the
temperature of the window.  The latter is taken to be equal to the
temperature of the internal blackbody target (which is about 10 deg C
above ambient.)  The emissivity of the window is 0.00164 at 23.8 GHz
and 0.00217 at 31.4 GHz.  For a reference temperature of 292.6 K and
brightness temperatures of 32.3 and 20.8 K at 23.8 and 31.4 GHz
respectively, this amounts to errors of -0.43 and -0.59 K at the
respective frequencies.  The average PWV for this date (5 April 1995)
was 1.4 cm.

At higher levels of PWV and for cloudy conditions, the brightness
temperatures are higher and so the error is smaller.

I will adjust the calibrations of all SGP radiometers to account
for this problem by the end of tomorrow (4 April 1996).
Measurements:sgp5mwravgB4.c1:
  • liq_sdev
  • Total water vapor along LOS path(vap)
  • Total liquid water along LOS path(liq)
  • vap_sdev

sgp5mwravgB1.c1:
  • Total water vapor along LOS path(vap)
  • Total liquid water along LOS path(liq)
  • vap_sdev
  • liq_sdev

sgp5mwravgB5.c1:
  • Total water vapor along LOS path(vap)
  • vap_sdev
  • Total liquid water along LOS path(liq)
  • liq_sdev


Back To Table of Contents

DQRID : D960404.8
Start DateStart TimeEnd DateEnd Time
03/01/1994000504/04/19962359
Subject:
SGP/MWR/B1/B4/B5 - Reprocess: MWR Tuning Functions
DataStreams:sgp5mwravgB4.c1, sgp5mwravgB5.c1, sgp5mwravgB1.c1
Description:
The 'tuning functions' used to adjust the equivalent brightness
temperatures (TBs) measured by the ARM microwave radiometers (MWRs) are
now believed to be both incorrect and unnecessary.  They should no longer
be used and the data (going back to 1992) that incorporated them should
be reprocessed.  By eliminating these tuning functions the radiometer
retrievals would be independent of the soundings.

BACKGROUND

A recent comparison by Barry Lesht (ANL) of the precipitable water
vapor (PWV) retrieved from the MWR-measured brightness temperatures
against PWV derived by integrating along the trajectory of radiosonde
ascents has revealed that the MWR values are about 90% of those derived
from the soundings.  This is directly attributable to the slope of the
tuning function for the vapor-sensing channel (23.8 GHz) of 0.915 which
is applied to the measured brightness temperatures prior to retrieval
of PWV.

The rationale behind the use of the tuning functions is that the
radiation model (Liebe 87), on which the retrieval is based, is
imperfect whereas the radiosondes represent 'ground truth.'  Thus the
observed brightness temperatures must be adjusted to match those
calculated with the model using co-located soundings so that the
retrieval yields precipitable vapor amounts that agree with the
soundings.

Tuning functions were developed for the present ARM MWRs using
co-located soundings launched between October 1992 and December 1993.
These were adjusted slightly in January 1995 to account for the effects
of the 1-point calibration check performed prior to launch (see DQR
P950110.1):

     23.8 GHz:  TB_model = 0.789 + 0.915 TB_measured  (R2 = 0.998)
     31.4 GHz:  TB_model = 1.142 + 0.910 TB_measured  (R2 = 0.984)

However, repeating this exercise for soundings launched during 1994 and
1995 (excepting those that were mis-calibrated by the manufacturer; see
D960229.1) it now appears that the model-calculated
brightness temperatures are in much closer agreement with the measured
values and that the tuning functions account more for variations in the
radiosonde calibration than for any deficiencies in the radiation
model.  

Consequently, it appears that the present tuning functions are
incorrect and bias the retrieved PWV low by 10%.  In addition, given
the present agreement between measured and modeled brightness
temperatures, the tuning functions are also unnecessary.

METHODOLOGY

Brightness temperatures measured with microwave radiometer (MWR) serial
number 10, which was deployed at the central facility in December 1993,
have been compared against calculations using measurements from the
co-located Balloon-Borne Sounding System (BBSS).  The results are
summarized in two tables.  In each table, the calibration dates of the
sondes and MWR are listed as well as the time period and number of
samples included in each regression.  Each MWR sample is a 40-minute
average, centered on the time of the sonde launch, of the microwave
brightness temperature.  In order to include only clear sky conditions,
samples for which the standard deviation of the liquid-sensing (31.4
GHz) channel exceeded 0.3 K were eliminated.  To assure that the water
vapor was reasonably homogeneous horizontally, samples for which the
standard deviation in the vapor-sensing (23.8 GHz) channel exceeded 0.4
K (in 1995) or 0.5 K (in 1994) were eliminated.  The 1994 threshold is
larger in order to increase the number of samples and reduce the
standard error in the results.

The microwave radiometer measurements used in this comparison have been
reprocessed to account for calibration changes and other problems (see
P940813.1)

                            TB vs PWV

The first table is a comparison of microwave brightness temperature
(TB_mwr) regressed against the precipitable water vapor (PWV) computed
by integrating along the trajectory of the radiosonde ascent.  The
sondes launched during May - December 1994 are compared against two
sets of MWR data; the first uses the May 1994 calibration, and the
second uses the calibration of July 1994.  A comparison is also made of
TB_model vs PWV ('Liebe87') for reference.

The intercepts indicate the contribution due to molecular oxygen (i.e.
the tail of the 60 GHz line) which is affected by temperature and
pressure.  Note that the 'Liebe87' intercepts vary seasonally as the
temperature changes.  Note also that the effect of MWR calibration
changes is most evident in the intercept: offsets of 1-2 K are
observed.  Because the MWR calibration values represent the slope of
the radiometer equation (see Appendix), the magnitude of the offset is
largest at 0 K (i.e. the intercept) and declines to zero at ambient
temperature (~290 K).

The slope of the regression is essentially unaffected by the MWR
calibration.  Variations in the slope of the regression correlate with
sonde calibration date.  The sondes calibrated in May 1994 or later
appear to yield much closer agreement between the measured brightness
temperatures and those calculated with the Liebe 87 model than those
calibrated in January 1994 or earlier, with which the present tuning
functions were developed.

TABLE 1.  Microwave brightness temperature vs. precipitable water vapor

Relationship:  TB_mwr (K) = intercept (K) + slope (K/cm) * PW_sonde (cm)
Standard Error of the intercepts and slopes are given in parentheses.

Date of   Date of  Period          ------ 23.8 GHz -----  ----- 31.4 GHz -----
Sonde Cal MWR tip  Covered    N    intercept   slope      intercept  slope

1991-93    92-93  Oct92-Dec93 91   6.7        14.7        8.1        5.3

1992,93   Dec 93  Jan-Feb 94  85   6.9(0.19)  15.8(0.26)  8.8(0.13)  5.6(0.17)
1992,93   Liebe87 Jan-Feb 94  85   6.5(0.02)  13.8(0.03)  8.9(0.07)  4.5(0.09)

 Jun 93   Dec 93      Apr 94  16  10.6(1.11)  14.8(0.55) 10.1(0.51)  5.6(0.25)
 Jun 93   Liebe87     Apr 94  16   6.9(0.05)  13.6(0.02)  8.1(0.09)  5.0(0.05)

1992,93   May 94  May-Jun 94  48   7.0(1.03)  14.9(0.45)  7.8(0.41)  5.7(0.17)
1992,93   Jul 94  May-Jun 94  48   5.1(1.03)  14.9(0.44)  6.6(0.39)  5.7(0.17)
1992,93   Liebe87 May-Jun 94  48   7.1(0.11)  13.5(0.05)  8.4(0.16)  4.9(0.07)

 Jan 94   Dec 93  Feb-May 94  95   7.6(0.27)  14.3(0.14)  8.5(0.15)  5.5(0.08)
 Jan 94   Liebe87 Feb-May 94  95   6.9(0.05)  13.6(0.02)  8.1(0.09)  5.0(0.05)

 May 94   May 94  Jun-Aug 94  78  12.3(1.04)  13.0(0.34) 11.0(0.39)  4.8(0.13)
 May 94   Jul 94  Jun-Aug 94  78  10.3(1.04)  13.1(0.34)  9.8(0.39)  4.8(0.13)
 May 94   Liebe87 Jun-Aug 94  78   7.8(0.22)  13.3(0.07)  8.6(0.29)  4.9(0.10)

 Jun 94   May 94  Jul-Dec 94  57   8.3(0.37)  13.6(0.21)  8.8(0.19)  5.2(0.11)
 Jun 94   Jul 94  Jul-Dec 94  57   6.4(0.37)  13.6(0.21)  7.7(0.18)  5.2(0.10)
 Jun 94   Liebe87 Jul-Dec 94  57   6.9(0.08)  13.5(0.04)  8.3(0.10)  4.9(0.06)

 Aug 94   May 94  Sep-Dec 94  90   7.4(0.15)  13.5(0.09)  8.8(0.11)  5.1(0.07)
 Aug 94   Jul 94  Sep-Dec 94  90   5.5(0.14)  13.6(0.09)  7.8(0.12)  5.0(0.07)
 Aug 94   Liebe87 Sep-Dec 94  90   6.8(0.05)  13.6(0.03)  8.6(0.09)  4.9(0.06)
Measurements:sgp5mwravgB4.c1:
  • Total water vapor along LOS path(vap)
  • Total liquid water along LOS path(liq)

sgp5mwravgB1.c1:
  • Total water vapor along LOS path(vap)
  • Total liquid water along LOS path(liq)

sgp5mwravgB5.c1:
  • Total water vapor along LOS path(vap)
  • Total liquid water along LOS path(liq)


Back To Table of Contents

DQRID : D961117.12
Start DateStart TimeEnd DateEnd Time
07/02/1996192507/22/19960200
Subject:
Loss of thermal stabilization
DataStreams:sgp5mwravgB4.c1
Description:
DQR No:                               Platform: sgpmwrlos, sgp5mwravg, 
                                                sgpqmemwrlos

Subject: Loss of thermal stabilization

Date Submitted:
Submitted By:    JIM LILJEGREN        _X_  Instrument Mentor
                                      ___  EST Member
                                      ___  Science Team Member
                                      ___  Other _____________________________
 
For questions or problems, please contact the ARM Experiment Center at
509-375-6898 or via email at dqr@arm.gov.

Platform/Measurement:
      What level data: a0,a1,c1

    What location was the data collected at: SGP B1 (Hillsboro, KS)
                                             SGP B4 (Vici, OK)
                                             SGP B5 (Morris, OK)
                                             SGP B6 (Purcell, OK)
                                             SGP C1 (Lamont, OK) 

    Period of time in question  (see table below)

 Data should be labeled:
 ___  questionable                      ___  All data fields affected
 _X_  incorrect                         _X_  Only some data fields affected:
 ___  wrong calibration
 ___  others                                 "23tbsky","31tbsky","vap","liq"
 
 Discussion of Problem:

I pointed out in a previous DQR (P960405.1) that during August of 1994
and 1995 the microwave radiometers would lose thermal stabilization on
very hot, sunny days when the temperature in the radiometer enclosure
rose above the set point for thermal stabilization (~50 deg C).  Although
I had planned to travel to the SGP prior to August 1996 to adjust the
set points upward to prevent this problem from occuring this year, the
temperatures in July 1996 were hotter than in previous years and the
loss of stabilization problem occurred before I could make the needed
adjustment.

The purpose of this note is the identify the time periods for which
this problem occurred.  More specific information about the problem,
including how the problem is detected and its effect on the reported
values of integrated water vapor and integrated cloud liquid water are
provided in the earlier DQR.

It is useful to repeat here that when the thermal stabilization is
lost, the reported precipitable water vapor increases and the liquid
water path decreases (and becomes SIGNIFICANTLY NEGATIVE (-0.1 mm) on
clear sky days).  The RMS noise level in the data also increases
sharply.  THESE ANOMALOUS VALUES HAVE BEEN INCLUDED IN THE 5-MINUTE
AVERAGES.

LOCATION                BEGINNING DATE AND TIME    ENDING DATE AND TIME

C1 (Central Facility)  
                         1 July 96  18:25 GMT      2 July 96  02:00 GMT
                         2          16:45          3          02:30
                         3          17:40          4          02:15
                         4          15:50          4          17:10
                         5          17:00          6          03:30
                         6          16:00          7          03:00
                         7          20:20          8          01:30
                        18          20:00         18          23:45
                        19          18:30         20          01:20
                        20          19:45         21          02:25
                        21          17:00         22          02:20
                        22          19:30         23          23:00

B1 (Hillsboro, KS)
                         1 July 96  20:30          2 July 96  01:15
                         2          17:00          3          02:00
                        17          20:30         18          00:15
                        18          19:25         19          02:00
                        19          19:00         20          02:30
                        20          18:35         21          00:30
                        21          20:20         22          01:40
                        28          20:55         28          00:15

B4 (Vici, OK)            2 July 96  19:25          2 July 96  23:15
                         3          19:35          3          21:15
                         4          20:40          5          00:30
                         5          19:15          6          02:00
                         6          19:00          6          22:40
                         7          20:45          8          00:30
                        21          19:45         22          02:00

B5 (Morris, OK)
                         1 July 96  18:35          2 July 96  00:45
                         2          17:20          3          01:15
                         3          17:25          4          02:00
                         5          20:20          6          01:45
                         6          16:45          7          02:30
                         7          18:10          8          01:00
                        19          20:00         20          00:20
                        20          19:30         21          00:55
                        21          18:15         22          01:30
                        22          19:30         23          01:15
                        23          22:00         24          00:15

B6 (Purcell, OK)

                         1 July 96  20:15          2 July 96  00:10
                         2          18:40          3          00:05
                         3          20:40          4          00:35
                         4          20:15          5          00:10
                         5          19:45          6          01:15
                         6          19:10          7          01:40
                        19          21:30         19          23:00
                        20          20:15         21          00:05
                        21          21:45         22          01:00
                        22          21:45         23          00:00

Other observations/measurements impacted by this problem:

none

Suggested Corrections of the Problem: (e.g. change calibration factor and
recompute, flag data with this comment, etc.)

Flag with this comment.
Measurements:sgp5mwravgB4.c1:
  • liq_sdev
  • tbsky23_sdev
  • Total water vapor along LOS path(vap)
  • tbsky31_sdev
  • alt
  • Total liquid water along LOS path(liq)
  • vap_sdev
  • lon
  • 31tbsky
  • num_obs
  • base_time
  • lat
  • time_offset
  • water_flag_fraction
  • 23tbsky


Back To Table of Contents

DQRID : D961117.16
Start DateStart TimeEnd DateEnd Time
07/01/1996183507/24/19960015
07/01/1996203007/28/19960015
07/02/1996192507/22/19960200
Subject:
Loss of thermal stabilization
DataStreams:sgp5mwravgB5.c1, sgp5mwravgB1.c1, sgp5mwravgB4.c1
Description:
DQR No:                               Platform: sgpmwrlos, sgp5mwravg, 
                                                sgpqmemwrlos

Subject: Loss of thermal stabilization

Date Submitted:
Submitted By:    JIM LILJEGREN        _X_  Instrument Mentor
                                      ___  EST Member
                                      ___  Science Team Member
                                      ___  Other _____________________________
 
For questions or problems, please contact the ARM Experiment Center at
509-375-6898 or via email at dqr@arm.gov.



Platform/Measurement:
      What level data: a0,a1,c1

    What location was the data collected at: SGP B1 (Hillsboro, KS)
                                             SGP B4 (Vici, OK)
                                             SGP B5 (Morris, OK)
                                             SGP B6 (Purcell, OK)
                                             SGP C1 (Lamont, OK) 

    Period of time in question  (see table below)



Data should be labeled:
 ___  questionable                      ___  All data fields affected
 _X_  incorrect                         _X_  Only some data fields affected:
___  wrong calibration
___  others                                 "23tbsky","31tbsky","vap","liq"

 Discussion of Problem:

I pointed out in a previous DQR (D960405.1) that during August of 1994
and 1995 the microwave radiometers would lose thermal stabilization on
very hot, sunny days when the temperature in the radiometer enclosure
rose above the set point for thermal stabilization (~50 deg C).  Although
I had planned to travel to the SGP prior to August 1996 to adjust the
set points upward to prevent this problem from occuring this year, the
temperatures in July 1996 were hotter than in previous years and the
loss of stabilization problem occurred before I could make the needed
adjustment.

The purpose of this note is the identify the time periods for which
this problem occurred.  More specific information about the problem,
including how the problem is detected and its effect on the reported
values of integrated water vapor and integrated cloud liquid water are
provided in the earlier DQR.

It is useful to repeat here that when the thermal stabilization is
lost, the reported precipitable water vapor increases and the liquid
water path decreases (and becomes SIGNIFICANTLY NEGATIVE (-0.1 mm) on
clear sky days).  The RMS noise level in the data also increases
sharply.  THESE ANOMALOUS VALUES HAVE BEEN INCLUDED IN THE 5-MINUTE
AVERAGES.

LOCATION                BEGINNING DATE AND TIME    ENDING DATE AND TIME

C1 (Central Facility)  
                         1 July 96  18:25 GMT      2 July 96  02:00 GMT
                         2          16:45          3          02:30
                         3          17:40          4          02:15
                         4          15:50          4          17:10
                         5          17:00          6          03:30
                         6          16:00          7          03:00
                         7          20:20          8          01:30
                        18          20:00         18          23:45
                        19          18:30         20          01:20
                        20          19:45         21          02:25
                        21          17:00         22          02:20
                        22          19:30         23          23:00

B1 (Hillsboro, KS)
                         1 July 96  20:30          2 July 96  01:15
                         2          17:00          3          02:00
                        17          20:30         18          00:15
                        18          19:25         19          02:00
                        19          19:00         20          02:30
                        20          18:35         21          00:30
                        21          20:20         22          01:40
                        28          20:55         28          00:15

B4 (Vici, OK)            2 July 96  19:25          2 July 96  23:15
                         3          19:35          3          21:15
                         4          20:40          5          00:30
                         5          19:15          6          02:00
                         6          19:00          6          22:40
                         7          20:45          8          00:30
                        21          19:45         22          02:00

B5 (Morris, OK)
                         1 July 96  18:35          2 July 96  00:45
                         2          17:20          3          01:15
                         3          17:25          4          02:00
                         5          20:20          6          01:45
                         6          16:45          7          02:30
                         7          18:10          8          01:00
                        19          20:00         20          00:20
                        20          19:30         21          00:55
                        21          18:15         22          01:30
                        22          19:30         23          01:15
                        23          22:00         24          00:15

B6 (Purcell, OK)

                         1 July 96  20:15          2 July 96  00:10
                         2          18:40          3          00:05
                         3          20:40          4          00:35
                         4          20:15          5          00:10
                         5          19:45          6          01:15
                         6          19:10          7          01:40
                        19          21:30         19          23:00
                        20          20:15         21          00:05
                        21          21:45         22          01:00
                        22          21:45         23          00:00

Other observations/measurements impacted by this problem:

none

Suggested Corrections of the Problem: (e.g. change calibration factor and
recompute, flag data with this comment, etc.)

Flag with this comment.
Measurements:sgp5mwravgB1.c1:
  • time_offset
  • vap_sdev
  • liq_sdev
  • lon
  • Temperature, brightness, longwave(ir_temp)
  • base_time
  • tbsky31_sdev
  • Total water vapor along LOS path(vap)
  • tbsky23_sdev
  • 31tbsky
  • 23tbsky
  • alt
  • ir_temp_sdev
  • num_obs
  • Total liquid water along LOS path(liq)
  • lat
  • water_flag_fraction

sgp5mwravgB4.c1:
  • Temperature, brightness, longwave(ir_temp)
  • alt
  • 31tbsky
  • ir_temp_sdev
  • tbsky31_sdev
  • Total liquid water along LOS path(liq)
  • vap_sdev
  • lat
  • water_flag_fraction
  • 23tbsky
  • liq_sdev
  • tbsky23_sdev
  • Total water vapor along LOS path(vap)
  • num_obs
  • base_time
  • time_offset
  • lon

sgp5mwravgB5.c1:
  • Total liquid water along LOS path(liq)
  • Temperature, brightness, longwave(ir_temp)
  • liq_sdev
  • water_flag_fraction
  • lat
  • Total water vapor along LOS path(vap)
  • vap_sdev
  • tbsky23_sdev
  • 23tbsky
  • 31tbsky
  • lon
  • tbsky31_sdev
  • alt
  • num_obs
  • ir_temp_sdev
  • base_time
  • time_offset


Back To Table of Contents

DQRID : D961117.17
Start DateStart TimeEnd DateEnd Time
07/02/1996192507/22/19960200
Subject:
Loss of thermal stabilization
DataStreams:sgp5mwravgB4.c1
Description:
DQR No:                               Platform: sgpmwrlos, sgp5mwravg, 
                                                sgpqmemwrlos

Subject: Loss of thermal stabilization

Date Submitted:
Submitted By:    JIM LILJEGREN        _X_  Instrument Mentor
                                      ___  EST Member
                                      ___  Science Team Member
                                      ___  Other _____________________________
 
For questions or problems, please contact the ARM Experiment Center at
509-375-6898 or via email at dqr@arm.gov.

Platform/Measurement:
      What level data: a0,a1,c1

    What location was the data collected at: SGP B1 (Hillsboro, KS)
                                             SGP B4 (Vici, OK)
                                             SGP B5 (Morris, OK)
                                             SGP B6 (Purcell, OK)
                                             SGP C1 (Lamont, OK) 

    Period of time in question  (see table below)

 Data should be labeled:
 ___  questionable                      ___  All data fields affected
 _X_  incorrect                         _X_  Only some data fields affected:
 ___  wrong calibration
 ___  others                                 '23tbsky','31tbsky','vap','liq'
 
 Discussion of Problem:

I pointed out in a previous DQR (D960405.1) that during August of 1994
and 1995 the microwave radiometers would lose thermal stabilization on
very hot, sunny days when the temperature in the radiometer enclosure
rose above the set point for thermal stabilization (~50 deg C).  Although
I had planned to travel to the SGP prior to August 1996 to adjust the
set points upward to prevent this problem from occuring this year, the
temperatures in July 1996 were hotter than in previous years and the
loss of stabilization problem occurred before I could make the needed
adjustment.

The purpose of this note is the identify the time periods for which
this problem occurred.  More specific information about the problem,
including how the problem is detected and its effect on the reported
values of integrated water vapor and integrated cloud liquid water are
provided in the earlier DQR.

It is useful to repeat here that when the thermal stabilization is
lost, the reported precipitable water vapor increases and the liquid
water path decreases (and becomes SIGNIFICANTLY NEGATIVE (-0.1 mm) on
clear sky days).  The RMS noise level in the data also increases
sharply.  THESE ANOMALOUS VALUES HAVE BEEN INCLUDED IN THE 5-MINUTE
AVERAGES.

LOCATION                BEGINNING DATE AND TIME    ENDING DATE AND TIME

C1 (Central Facility)  
                         1 July 96  18:25 GMT      2 July 96  02:00 GMT
                         2          16:45          3          02:30
                         3          17:40          4          02:15
                         4          15:50          4          17:10
                         5          17:00          6          03:30
                         6          16:00          7          03:00
                         7          20:20          8          01:30
                        18          20:00         18          23:45
                        19          18:30         20          01:20
                        20          19:45         21          02:25
                        21          17:00         22          02:20
                        22          19:30         23          23:00

B1 (Hillsboro, KS)
                         1 July 96  20:30          2 July 96  01:15
                         2          17:00          3          02:00
                        17          20:30         18          00:15
                        18          19:25         19          02:00
                        19          19:00         20          02:30
                        20          18:35         21          00:30
                        21          20:20         22          01:40
                        28          20:55         28          00:15

B4 (Vici, OK)            2 July 96  19:25          2 July 96  23:15
                         3          19:35          3          21:15
                         4          20:40          5          00:30
                         5          19:15          6          02:00
                         6          19:00          6          22:40
                         7          20:45          8          00:30
                        21          19:45         22          02:00

B5 (Morris, OK)
                         1 July 96  18:35          2 July 96  00:45
                         2          17:20          3          01:15
                         3          17:25          4          02:00
                         5          20:20          6          01:45
                         6          16:45          7          02:30
                         7          18:10          8          01:00
                        19          20:00         20          00:20
                        20          19:30         21          00:55
                        21          18:15         22          01:30
                        22          19:30         23          01:15
                        23          22:00         24          00:15

B6 (Purcell, OK)

                         1 July 96  20:15          2 July 96  00:10
                         2          18:40          3          00:05
                         3          20:40          4          00:35
                         4          20:15          5          00:10
                         5          19:45          6          01:15
                         6          19:10          7          01:40
                        19          21:30         19          23:00
                        20          20:15         21          00:05
                        21          21:45         22          01:00
                        22          21:45         23          00:00

Other observations/measurements impacted by this problem:

none

Suggested Corrections of the Problem: (e.g. change calibration factor and
recompute, flag data with this comment, etc.)

Flag with this comment.
Measurements:sgp5mwravgB4.c1:
  • liq_sdev
  • Temperature, brightness, longwave(ir_temp)
  • tbsky23_sdev
  • Total water vapor along LOS path(vap)
  • tbsky31_sdev
  • alt
  • Total liquid water along LOS path(liq)
  • vap_sdev
  • lon
  • 31tbsky
  • num_obs
  • ir_temp_sdev
  • base_time
  • lat
  • time_offset
  • water_flag_fraction
  • 23tbsky


Back To Table of Contents

DQRID : D961120.1
Start DateStart TimeEnd DateEnd Time
08/01/1996000008/31/19962359
Subject:
SGP/MWR/B1/B4/B5/B6/C1 - Thermal Stabilization Adjustment
DataStreams:sgp1mwravgC1.c1, sgp5mwravgB1.c1, sgp5mwravgB4.c1, sgp5mwravgB5.c1, sgp5mwravgB6.c1,
sgp5mwravgC1.c1, sgpmwrlosB1.a0, sgpmwrlosB1.a1, sgpmwrlosB4.a0, sgpmwrlosB4.a1,
sgpmwrlosB5.a0, sgpmwrlosB5.a1, sgpmwrlosB6.a0, sgpmwrlosB6.a1, sgpmwrlosC1.a1, sgpmwrlosC1.b1
Description:
In order to correct a thermal stabilization problem identified earlier
I adjusted the thermal set point of the microwave radiometers at the
SGP upward from 48-50 deg C to 55 deg C in early August 1996 according
to the schedule given below.

B6    5 August 1996
C1    6 August 1996
B1    7 August 1996
B5    8 August 1996

Subsequent to making this adjustment the MWRs were put in TIP mode to
check on whether the change in set point temperature affected their
calibration.  Because clear sky conditions were quite intermittent, it
is difficult to determine whether the substantial variability in the
tip data were attributable to the change in thermal set point.  The
instrument calibration was not altered in August.

Tip data were again collected with these instruments in September prior
to the beginning and at the close of the Water Vapor IOP.  For example,
the calibration of the instrument at the central facility (S/N 10)
derived from the September data was essentially the same as that
derived from calibration data acquired in February 1996.  Although this
would lead one to believe that altering the thermal set point did not
affect the instrument calibration, it may be that some transient effect
was induced.

In comparing soundings launched from the central facility with the
microwave radiometer there, I noticed that those sondes calibrated in
June 1996 consistently reported lower integrated water vapor than the
radiometer in July and September (during the IOP) but were in better
agreement with the radiometer for the two weeks period immediately
after the set point was adjusted.  I suspect that adjusting the thermal
set point may have temporarily increased the radiometer gain
(kelvins/volt) thereby lowering the measured brightness temperature and
the retrieved integrated water vapor.

It is not clear why a temporary change in gain should occur or even
whether it did.  But users of the data should be aware that the data
from the microwave radiometers at the SGP may be anomalous during
August 1996.
Measurements:sgpmwrlosB1.a1:
  • 23tbsky
  • Total liquid water along LOS path(liq)
  • Total water vapor along LOS path(vap)
  • 31tbsky

sgpmwrlosB1.a0:
  • 23tbsky
  • Total water vapor along LOS path(vap)
  • 31tbsky
  • Total liquid water along LOS path(liq)

sgp1mwravgC1.c1:
  • 31tbsky
  • Total water vapor along LOS path(vap)
  • 23tbsky
  • Total liquid water along LOS path(liq)

sgpmwrlosC1.a1:
  • 31.4 GHz sky brightness temperature(tbsky31)
  • Total water vapor along LOS path(vap)
  • Total liquid water along LOS path(liq)
  • Radiation, longwave, brightness temperature, 23.8 GHz(tbsky23)

sgpmwrlosC1.b1:
  • Radiation, longwave, brightness temperature, 23.8 GHz(tbsky23)
  • Total liquid water along LOS path(liq)
  • Total water vapor along LOS path(vap)
  • 31.4 GHz sky brightness temperature(tbsky31)

sgp5mwravgB5.c1:
  • Total liquid water along LOS path(liq)
  • Total water vapor along LOS path(vap)
  • 23tbsky
  • 31tbsky

sgp5mwravgB1.c1:
  • Total water vapor along LOS path(vap)
  • 31tbsky
  • 23tbsky
  • Total liquid water along LOS path(liq)

sgp5mwravgB4.c1:
  • 31tbsky
  • Total liquid water along LOS path(liq)
  • 23tbsky
  • Total water vapor along LOS path(vap)

sgpmwrlosB6.a1:
  • 23tbsky
  • Total water vapor along LOS path(vap)
  • Total liquid water along LOS path(liq)
  • 31tbsky

sgp5mwravgC1.c1:
  • 23tbsky
  • Total water vapor along LOS path(vap)
  • 31tbsky
  • Total liquid water along LOS path(liq)

sgpmwrlosB6.a0:
  • Total water vapor along LOS path(vap)
  • Total liquid water along LOS path(liq)
  • 31tbsky
  • 23tbsky

sgpmwrlosB4.a0:
  • Total water vapor along LOS path(vap)
  • Total liquid water along LOS path(liq)
  • 31tbsky
  • 23tbsky

sgpmwrlosB5.a0:
  • Total liquid water along LOS path(liq)
  • 31tbsky
  • Total water vapor along LOS path(vap)
  • 23tbsky

sgpmwrlosB4.a1:
  • 31tbsky
  • Total liquid water along LOS path(liq)
  • Total water vapor along LOS path(vap)
  • 23tbsky

sgpmwrlosB5.a1:
  • Total water vapor along LOS path(vap)
  • 23tbsky
  • Total liquid water along LOS path(liq)
  • 31tbsky

sgp5mwravgB6.c1:
  • 23tbsky
  • Total liquid water along LOS path(liq)
  • Total water vapor along LOS path(vap)
  • 31tbsky


Back To Table of Contents

DQRID : D991102.1
Start DateStart TimeEnd DateEnd Time
03/05/1999000006/08/19991710
Subject:
SGP/MWR/B4 - power lines in field of view
DataStreams:sgpmwrlosB4.a1, sgpmwrlosB4.b1, sgpmwrtipB4.a1, sgp5mwravgB4.c1, sgpmwrlosB4.00,
sgpmwrtipB4.00
Description:
On 5 March 1999, the local power company installed transmission lines north 
of BF4 at an elevation angle of about 30 degrees from the MWR, in the field 
of view of tip scans. This affected the calibration of the instrument and 
the retrieved values of precipitable water vapor and liquid water path. 
On 8 June 1999 at 1710 GMT, the MWR was remotely rotated 90 degrees so the 
instrument would scan from east to west.
This reorientation has corrected the problem.
Measurements:sgpmwrtipB4.a1:
  • bbn31
  • tc31
  • tnd_nom31
  • bb23
  • tnd31I
  • bb31
  • tnd31
  • tnd23I
  • tipsky23
  • r31
  • tnd_nom23
  • tc23
  • tnd23
  • r23
  • tipsky31
  • bbn23

sgp5mwravgB4.c1:
  • liq_sdev
  • tbsky23_sdev
  • Total water vapor along LOS path(vap)
  • 31tbsky
  • tbsky31_sdev
  • Total liquid water along LOS path(liq)
  • vap_sdev
  • 23tbsky

sgpmwrtipB4.00:
  • Raw data stream - documentation not supported

sgpmwrlosB4.b1:
  • Total liquid water along LOS path(liq)
  • 31.4 GHz sky brightness temperature(tbsky31)
  • Total water vapor along LOS path(vap)
  • Radiation, longwave, brightness temperature, 23.8 GHz(tbsky23)

sgpmwrlosB4.00:
  • Raw data stream - documentation not supported

sgpmwrlosB4.a1:
  • Total water vapor along LOS path(vap)
  • Total liquid water along LOS path(liq)


Back To Table of Contents



END OF DATA