Data Quality Reports for Session: 141464 User: gaun Completed: 03/26/2012


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D110815.1PGH/MWR/M1 - Reprocess: Incorrect calibrationpghmwrlosM1.b1, pghmwrtipM1.a1


DQRID : D110815.1
Start DateStart TimeEnd DateEnd Time
06/20/2011053207/15/20110000
Subject:
PGH/MWR/M1 - Reprocess: Incorrect calibration
DataStreams:pghmwrlosM1.b1, pghmwrtipM1.a1
Description:
When the radiometer started to collect data it was not at its final location with 
obstructed field of view. The radiometer was moved to its final location on 7/13/2011. After the 
radiometer was moved it started to collect good data for calibration.
Measurements:pghmwrtipM1.a1:
  • tnd23
  • tnd_nom31
  • tnd23I
  • tnd31
  • tnd31I
  • tbsky31tip
  • tbsky23tip
  • tc31
  • tc23
  • tnd_nom23

pghmwrlosM1.b1:
  • tnd31
  • tc31
  • Radiation, longwave, brightness temperature, 23.8 GHz(tbsky23)
  • tnd23
  • tc23
  • tnd_nom23
  • tnd_nom31
  • 31.4 GHz sky brightness temperature(tbsky31)


Back To Table of Contents



END OF DATA