DQR ID | Subject | Data Streams Affected |
---|---|---|
D000210.1 | SGP/EBBR/E26 - Temperature from Left T/RH Probe Incorrect | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D000210.2 | SGP/EBBR - EBBR E19 Changeout, Data Missing/Incorrect | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
D000210.3 | SGP/EBBR - EBBR E22 Moved, Data Missing/Incorrect | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D000210.4 | SGP/EBBR - EBBR E22 Net Radiation Incorrect | sgp30ebbrE22.b1, sgp5ebbrE22.b1, sgp15ebbrE22.b1 |
D000211.1 | SGP/EBBR - EBBR E25 Data Missing/Incorrect During System Changeout | sgp15ebbrE25.b1, sgp30ebbrE25.b1, sgp5ebbrE25.b1 |
D000211.10 | SGP/EBBR - EBBR E20 Data Incorrect from SHF #5 | sgp30ebbrE20.b1, sgp15ebbrE20.b1 |
D000211.12 | SGP/EBBR - EBBR E20 Thermocouple Data Incorrect | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D000211.13 | SGP/EBBR/E26 - Left T/RH Probe Temperature Data Incorrect | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D000211.8 | SGP/EBBR - EBBR E25 Data Incorrect During J-panel Change | sgp15ebbrE25.b1, sgp30ebbrE25.b1, sgp5ebbrE25.b1 |
D000211.9 | SGP/EBBR - EBBR E25 Thermocouple Temperatures Incorrect | sgp15ebbrE25.b1, sgp30ebbrE25.b1, sgp5ebbrE25.b1 |
D000214.2 | SGP/EBBR - EBBR E7 Soil Moisture Jumping Around | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D000214.3 | SGP/EBBR - EBBR E7 Soil Moisture Sensors #2 and #4 Negative | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D000214.4 | SGP/EBBR - EBBR E7 Soil Moisture Sensor #5 Negative | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D000221.2 | SGP/EBBR - EBBR E7 Left Aspirator Not Working | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D000221.3 | SGP/EBBR/E26 - Right Aspirator Not Working | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D000221.4 | SGP/EBBR - EBBR E19 Wind Speed Incorrect | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
D000221.5 | SGP/EBBR - EBBR E15 Anemometer Cup Off, Wind Speed Incorrect | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D000221.6 | SGP/EBBR - EBBR E9 Anemometer Frozen, Wind Speed Incorrect | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
D000221.7 | SGP/EBBR/E26 - Anemometer Not Working, Wind Speed Incorrect | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D000221.8 | SGP/EBBR - EBBR E12 Anemometer Cups Frozen, Wind Speed Incorrect | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D000224.10 | SGP/EBBR - EBBR E7 Soil Moisture Spike | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D000224.11 | SGP/EBBR - EBBR E7 Soil Moisture Spike | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D000224.12 | SGP/EBBR - EBBR E7 Soil Moisture Spike | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D000224.13 | SGP/EBBR - EBBR E7 Soil Moisture Spike | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D000224.14 | SGP/EBBR - EBBR E7 Soil Moisture Spike | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D000224.15 | SGP/EBBR - EBBR E7 Soil Moisture Spike | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D000224.16 | SGP/EBBR - EBBR E7 Soil Moisture Spike | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D000224.17 | SGP/EBBR - EBBR E7 Soil Moisture Spike | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D000224.5 | SGP/EBBR - EBBR E7 Soil Moisture Spike | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D000224.6 | SGP/EBBR - EBBR E7 Soil Moisture Spike | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D000224.7 | SGP/EBBR - EBBR E7 Soil Moisture Spike | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D000224.8 | SGP/EBBR - EBBR E7 Soil Moisture Spike | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D000224.9 | SGP/EBBR - EBBR E7 Soil Moisture Spike | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D000225.1 | SGP/EBBR/26 - Low Battery Condition | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D000225.10 | SGP/EBBR - EBBR E12 Net Radiation Incorrect | sgp30ebbrE12.b1, sgp5ebbrE12.b1, sgp15ebbrE12.b1 |
D000225.2 | SGP/EBBR- EBBR E15 Air Temperatures Incorrect | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D000225.4 | SGP/EBBR - EBBR E18 Right Humidity Incorrect | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
D000225.5 | SGP/EBBR - EBBR E18 Data Incorrect After Lightning Strike | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
D000225.6 | SGP/EBBR - EBBR E18 Much Data Missing After Lightning Strike | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
D000225.7 | SGP/EBBR - EBBR E13 Left Aspirator Not Working | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D000225.9 | SGP/EBBR - EBBR E7 Soil Heat Flows Spiked | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D000228.1 | SGP/EBBR - EBBR E20 Right Air Temperature Incorrect | sgp30ebbrE20.b1, sgp5ebbrE20.b1, sgp15ebbrE20.b1 |
D000229.1 | SGP/EBBR - EBBR E7 Soil Moistures #2 and #4 Incorrect | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D000229.2 | SGP/EBBR - EBBR E8 Left Air Temperature (thermocouple) Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D000301.1 | SGP/EBBR - EBBR E12 Net Radiation Incorrect | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D000301.10 | SGP/EBBR - EBBR E8 Wind Direction Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D000301.11 | SGP/EBBR - EBBR E22 Soil Moistures Negative | sgp15ebbrE22.b1, sgp30ebbrE22.b1 |
D000301.2 | SGP/EBBR - EBBR E19 Net Radiation Incorrect | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
D000301.3 | SGP/EBBR - EBBR E13 Net Radiation Incorrect | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D000301.4 | SGP/EBBR - EBBR E18 Net Radiation Incorrect | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
D000301.5 | SGP/EBBR - EBBR E25 Net Radiation Incorrect | sgp30ebbrE25.b1, sgp5ebbrE25.b1, sgp15ebbrE25.b1 |
D000301.6 | SGP/EBBR - EBBR E2 Net Radiation Incorrect | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
D000301.7 | SGP/EBBR - EBBR E7 Net Radiation Incorrect | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D000301.8 | SGP/EBBR - EBBR E9 Net Radiation Incorrect | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
D000301.9 | SGP/EBBR/E26 - Net Radiation Incorrect | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D000302.2 | SGP/EBBR - EBBR E22 Soil Moistures Negative | sgp15ebbrE22.b1, sgp30ebbrE22.b1 |
D000303.1 | SGP/EBBR - EBBR E20 Left Air Temperature Incorrect | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D000307.10 | SGP/EBBR - EBBR E25 Soil Heat Flow #1 Incorrect | sgp15ebbrE25.b1, sgp30ebbrE25.b1 |
D000307.11 | SGP/EBBR - EBBR E25 Soil Heat Flow #1 Incorrect | sgp15ebbrE25.b1, sgp30ebbrE25.b1 |
D000307.12 | SGP/EBBR - EBBR E25 Soil Heat Flow #1 Incorrect | sgp15ebbrE25.b1, sgp30ebbrE25.b1 |
D000307.13 | SGP/EBBR - EBBR E25 Soil Heat Flow #1 Incorrect | sgp15ebbrE25.b1, sgp30ebbrE25.b1 |
D000307.14 | SGP/EBBR - EBBR E25 Soil Heat Flow #1 Incorrect | sgp15ebbrE25.b1, sgp30ebbrE25.b1 |
D000307.15 | SGP/EBBR - EBBR E25 Soil Heat Flow #1 Incorrect | sgp15ebbrE25.b1, sgp30ebbrE25.b1 |
D000307.16 | SGP/EBBR - EBBR E25 Soil Heat Flow #1 Incorrect | sgp15ebbrE25.b1, sgp30ebbrE25.b1 |
D000307.17 | SGP/EBBR - EBBR E25 Soil Heat Flow #1 Incorrect | sgp15ebbrE25.b1, sgp30ebbrE25.b1 |
D000307.18 | SGP/EBBR - EBBR E25 Soil Heat Flow #1 Incorrect | sgp15ebbrE25.b1, sgp30ebbrE25.b1 |
D000307.2 | SGP/EBBR - EBBR E25 Soil Heat Flow #1 Incorrect | sgp15ebbrE25.b1, sgp30ebbrE25.b1 |
D000307.5 | SGP/EBBR - EBBR E25 Soil Heat Flow #1 Incorrect | sgp30ebbrE25.b1, sgp15ebbrE25.b1 |
D000307.6 | SGP/EBBR - EBBR E25 Soil Heat Flow #1 Incorrect | sgp15ebbrE25.b1, sgp30ebbrE25.b1 |
D000307.7 | SGP/EBBR - EBBR E25 Soil Heat Flow #1 Incorrect | sgp15ebbrE25.b1, sgp30ebbrE25.b1 |
D000307.8 | SGP/EBBR - EBBR E25 Soil Heat Flow #1 Incorrect | sgp15ebbrE25.b1, sgp30ebbrE25.b1 |
D000307.9 | SGP/EBBR - EBBR E25 Soil Heat Flow #1 Incorrect | sgp15ebbrE25.b1, sgp30ebbrE25.b1 |
D000531.4 | SGP/EBBR/E13 - failure of sensors | sgp30ebbrE13.b1, sgp5ebbrE13.b1, sgp15ebbrE13.b1, sgp5ebbrE13.00 |
D000531.5 | SGP/EBBR/E13 - Failed Right Humidity Probe | sgp5ebbrE13.b1, sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.00 |
D000622.1 | SGP/EBBR/E26 - Reprocess Soil Moisture Off-Scale | sgp30ebbrE26.b1, sgp15ebbrE26.b1 |
D000623.1 | SGP/EBBR/E26 - EBBR E26 Left Aspirator Not Working | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D000623.10 | SGP/EBBR/E26 - EBBR E26 All Data Incorrect | sgp5ebbrE26.b1, sgp30ebbrE26.b1, sgp15ebbrE26.b1 |
D000623.11 | SGP/EBBR/E26 - All Data Incorrect | sgp5ebbrE26.b1, sgp30ebbrE26.b1, sgp15ebbrE26.b1 |
D000623.2 | SGP/EBBR/E26 - All Data Incorrect | sgp30ebbrE26.b1, sgp15ebbrE26.b1, sgp5ebbrE26.b1 |
D000623.3 | SGP/EBBR/E26 - Low Battery Voltage Causes Incorrect Data | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D000623.5 | SGP/EBBR/E26 - Low Battery Voltage Causes Incorrect Data | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D000623.6 | SGP/EBBR/E26 - Low Battery Voltage Causes Incorrect Data | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D000623.7 | SGP/EBBR/E26 - Low Battery Voltage Causes Incorrect Data | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D000623.8 | SGP/EBBR/E26 - All Data Incorrect | sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp15ebbrE26.b1 |
D000623.9 | SGP/EBBR/E26 - All Data Incorrect | sgp5ebbrE26.b1, sgp30ebbrE26.b1, sgp15ebbrE26.b1 |
D000627.10 | SGP/EBBR/E26 - Left Air Temperature Periodically Incorrect | sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp15ebbrE26.b1 |
D000627.11 | SGP/EBBR/E26 - Reference Temperature Incorrect | sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp15ebbrE26.b1 |
D000627.4 | SGP/EBBR/E26 - Electronics Problem | sgp30ebbrE26.b1, sgp15ebbrE26.b1, sgp5ebbrE26.b1 |
D000627.6 | SGP/EBBR/E26 - Soil Heat Flow #5 Incorrect | sgp30ebbrE26.b1, sgp15ebbrE26.b1 |
D000627.7 | SGP/EBBR/E26 - Soil Heat Flow #1 Incorrect | sgp15ebbrE26.b1, sgp30ebbrE26.b1 |
D000627.8 | SGP/EBBR/E26 - Soil Heat Flow #2 Large Negative | sgp30ebbrE26.b1, sgp15ebbrE26.b1 |
D000627.9 | SGP/EBBR/E26 - Left Air Temperature Incorrect | sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp15ebbrE26.b1 |
D000628.1 | SGP/EBBR/E13- Left humidity probe replaced | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1, sgp5ebbrE13.00 |
D010213.1 | SGP/EBBR/E12 - EBBR E12 Wind Speed Iced | sgp5ebbrE12.b1, sgp30ebbrE12.b1, sgp15ebbrE12.b1 |
D010213.10 | SGP/EBBR/E25 - EBBR E25 Wind Speed Iced | sgp30ebbrE25.b1, sgp15ebbrE25.b1, sgp5ebbrE25.b1 |
D010213.11 | SGP/EBBR/E25 - EBBR E25 Wind Speed Iced | sgp30ebbrE25.b1, sgp15ebbrE25.b1, sgp5ebbrE25.b1 |
D010213.12 | SGP/EBBR/E26 - Wind Speed Iced | sgp5ebbrE26.b1, sgp30ebbrE26.b1, sgp15ebbrE26.b1 |
D010213.13 | SGP/EBBR/E2 - EBBR E2 Wind Speed Iced | sgp5ebbrE2.b1, sgp30ebbrE2.b1, sgp15ebbrE2.b1 |
D010213.16 | SGP/EBBR/E7 - EBBR E7 Wind Speed Iced | sgp5ebbrE7.b1, sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010213.2 | SGP/EBBR/E12 - EBBR E12 Wind Speed Iced | sgp30ebbrE12.b1, sgp15ebbrE12.b1, sgp5ebbrE12.b1 |
D010213.3 | SGP/EBBR/E15 - EBBR E15 Wind Speed Iced | sgp30ebbrE15.b1, sgp15ebbrE15.b1, sgp5ebbrE15.b1 |
D010213.4 | SGP/EBBR/E18 - EBBR E18 Wind Speed Iced | sgp15ebbrE18.b1, sgp5ebbrE18.b1, sgp30ebbrE18.b1 |
D010213.5 | SGP/EBBR/E18 - EBBR E18 Wind Speed Iced | sgp5ebbrE18.b1, sgp30ebbrE18.b1, sgp15ebbrE18.b1 |
D010213.6 | SGP/EBBR/E19 - EBBR E19 Wind Speed Iced | sgp30ebbrE19.b1, sgp15ebbrE19.b1, sgp5ebbrE19.b1 |
D010213.7 | SGP/EBBR/E20 - EBBR E20 Wind Speed Iced | sgp30ebbrE20.b1, sgp15ebbrE20.b1, sgp5ebbrE20.b1 |
D010213.8 | SGP/EBBR/E22 - EBBR E22 Wind Speed Iced | sgp30ebbrE22.b1, sgp15ebbrE22.b1, sgp5ebbrE22.b1 |
D010213.9 | SGP/EBBR/E25 - EBBR E25 Wind Speed Iced | sgp30ebbrE25.b1, sgp15ebbrE25.b1, sgp5ebbrE25.b1 |
D010302.1 | SGP/EBBR/E2 - E2 EBBR Most Data Incorrect | sgp30ebbrE2.b1, sgp15ebbrE2.b1, sgp5ebbrE2.b1 |
D010302.13 | SGP/EBBR/E2 - Missing data | sgp5ebbrE2.b1, sgp15ebbrE2.b1, sgp30ebbrE2.b1 |
D010302.19 | SGP/EBBR/E2 - E2 EBBR Data Incorrect | sgp30ebbrE2.b1, sgp15ebbrE2.b1, sgp5ebbrE2.b1 |
D010302.2 | SGP/EBBR/E2 - E2 EBBR Ts4 and SM4 Bouncing Around | sgp30ebbrE2.b1, sgp15ebbrE2.b1 |
D010302.20 | SGP/EBBR/E2 - E2 EBBR Ts2 Incorrect | sgp30ebbrE2.b1, sgp15ebbrE2.b1 |
D010302.21 | SGP/EBBR/E2 - E2 EBBR Wind Direction Incorrect | sgp30ebbrE2.b1, sgp5ebbrE2.b1, sgp15ebbrE2.b1 |
D010302.22 | SGP/EBBR/E2 - E2 EBBR Data Incorrect | sgp30ebbrE2.b1, sgp15ebbrE2.b1, sgp5ebbrE2.b1 |
D010302.23 | SGP/EBBR/E2 - E2 EBBR Low Battery Condition | sgp30ebbrE2.b1, sgp15ebbrE2.b1, sgp5ebbrE2.b1 |
D010302.24 | SGP/EBBR/E2 - E2 EBBR Left Aspirator Fan Stopped | sgp30ebbrE2.b1, sgp5ebbrE2.b1, sgp15ebbrE2.b1 |
D010302.25 | SGP/EBBR/E2 - E2 EBBR AEM Stuck In One Position | sgp30ebbrE2.b1, sgp5ebbrE2.b1, sgp15ebbrE2.b1 |
D010302.26 | SGP/EBBR/E2 - E2 EBBR AEM Stuck In One Position | sgp30ebbrE2.b1, sgp5ebbrE2.b1, sgp15ebbrE2.b1 |
D010302.27 | SGP/EBBR/E2 - E2 EBBR AEM Stuck In One Position | sgp30ebbrE2.b1, sgp5ebbrE2.b1, sgp15ebbrE2.b1 |
D010302.28 | SGP/EBBR/E2 - E2 EBBR AEM Stuck In One Position | sgp30ebbrE2.b1, sgp5ebbrE2.b1, sgp15ebbrE2.b1 |
D010302.29 | SGP/EBBR/E2 - E2 EBBR AEM Stuck In One Position | sgp30ebbrE2.b1, sgp5ebbrE2.b1, sgp15ebbrE2.b1 |
D010302.30 | SGP/EBBR/E2 - E2 EBBR AEM Stuck In One Position | sgp30ebbrE2.b1, sgp5ebbrE2.b1, sgp15ebbrE2.b1 |
D010302.31 | SGP/EBBR/E2 - E2 EBBR AEM Stuck In One Position | sgp30ebbrE2.b1, sgp5ebbrE2.b1, sgp15ebbrE2.b1 |
D010302.32 | SGP/EBBR/E2 - E2 EBBR AEM Stuck In One Position | sgp30ebbrE2.b1, sgp5ebbrE2.b1, sgp15ebbrE2.b1 |
D010307.1 | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.10 | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.100 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.101 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.102 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.103 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.104 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.105 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.11 | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.12 | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.13 | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.14 | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.15 | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.16 | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.17 | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.18 | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.19 | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.20 | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.21 | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.22 | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.23 | SGP/EBBR/E8 - E8 EBBR Low Battery Condition | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.24 | SGP/EBBR/E8 - E8 EBBR Wind Speed Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.25 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.26 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.27 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.28 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.29 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.3 | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.30 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.31 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.32 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.33 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.34 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.35 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.36 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.37 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.38 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.39 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.4 | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.40 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.41 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.42 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.43 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.44 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.45 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.46 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.47 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.48 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.49 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.5 | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.50 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.51 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.52 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.53 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.54 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.55 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.56 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.57 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.58 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.59 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.6 | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.60 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.61 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.62 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.63 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.64 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.65 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.66 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.67 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.68 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.69 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.7 | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.70 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.71 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.73 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.74 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.75 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.76 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.77 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.78 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.79 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.8 | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.80 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.81 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.82 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.83 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.85 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.86 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.87 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.88 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.89 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.9 | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.90 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.91 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.92 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.93 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.94 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.95 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.96 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.97 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.98 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.99 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.100 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.101 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.102 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.103 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.105 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.106 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.107 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.108 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.109 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.110 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.111 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.112 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.113 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.114 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.115 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.116 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.117 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.118 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.119 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.12 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.120 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.121 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.122 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.123 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.124 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.125 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.126 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.127 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.128 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.129 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.13 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.130 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.131 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.132 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.133 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.134 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.135 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.136 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.137 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.138 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.139 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.14 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.140 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.141 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.142 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.143 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.144 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.145 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.146 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.148 | SGP/EBBR/E7 - E7 EBBR Datalogger Replacement/Upgrade - Data Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1, sgp5ebbrE7.b1 |
D010316.149 | SGP/EBBR/E7 - E7 EBBR Soil Temperature #1 Incorrect | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D010316.15 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.150 | SGP/EBBR/E7 - E7 EBBR Soil Temperature #1 Incorrect | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D010316.151 | SGP/EBBR/E7 - E7 EBBR Soil Temperature #1 Incorrect | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D010316.152 | SGP/EBBR/E7 - E7 EBBR Soil Temperature #1 Incorrect | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D010316.153 | SGP/EBBR/E7 - E7 EBBR Soil Temperature #1 Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010316.154 | SGP/EBBR/E7 - E7 EBBR Soil Temperature #1 Incorrect | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D010316.155 | SGP/EBBR/E7 - E7 EBBR Soil Temperature #1 Incorrect | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D010316.156 | SGP/EBBR/E7 - E7 EBBR Soil Temperature #1 Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010316.157 | SGP/EBBR/E7 - E7 EBBR Soil Temperature #1 Incorrect | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D010316.158 | SGP/EBBR/E7 - E7 EBBR Soil Temperature #1 Incorrect | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D010316.159 | SGP/EBBR/E7 - E7 EBBR Soil Temperature #1 Incorrect | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D010316.16 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.160 | SGP/EBBR/E7 - E7 EBBR Soil Temperature #1 Incorrect | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D010316.161 | SGP/EBBR/E7 - E7 EBBR Soil Temperature #1 Incorrect | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D010316.162 | SGP/EBBR/E7 - E7 EBBR Soil Temperature #1 Incorrect | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D010316.163 | SGP/EBBR/E7 - E7 EBBR Right Aspirator Replacement | sgp5ebbrE7.b1, sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010316.164 | SGP/EBBR/E7 - E7 EBBR Left Aspirator Fan Not Working | sgp15ebbrE7.b1, sgp5ebbrE7.b1, sgp30ebbrE7.b1 |
D010316.165 | SGP/EBBR/E7 - E7 EBBR Net Radiometer Dome Replacement | sgp5ebbrE7.b1, sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010316.166 | SGP/EBBR/E7 - E7 EBBR All Data Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1, sgp5ebbrE7.b1 |
D010316.167 | SGP/EBBR/E7 - E7 EBBR All Data Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1, sgp5ebbrE7.b1 |
D010316.168 | SGP/EBBR/E7 - E7 EBBR Soil Moisture Spikes | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010316.169 | SGP/EBBR/E7 - E7 EBBR Soil Moisture Spikes | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010316.17 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.170 | SGP/EBBR/E7 - E7 EBBR Soil Moisture Spikes | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010316.171 | SGP/EBBR/E7 - E7 EBBR Soil Moisture Spikes | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010316.172 | SGP/EBBR/E7 - E7 EBBR Soil Moisture Spikes | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010316.173 | SGP/EBBR/E7 - E7 EBBR Soil Moisture Spikes | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010316.174 | SGP/EBBR/E7 - E7 EBBR Soil Moisture Spikes | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010316.175 | SGP/EBBR/E7 - E7 EBBR Soil Temperature #4 Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010316.18 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.19 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.20 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.21 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.22 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.23 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.24 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.25 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.26 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.27 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.28 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.29 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.30 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.31 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.32 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.33 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.34 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.35 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.36 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.37 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.38 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.39 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.40 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.41 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.42 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.43 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.44 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.45 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.46 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.47 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.48 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.49 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.50 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.51 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.52 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.53 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.55 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.56 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.57 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.58 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.59 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.60 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.61 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.62 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.63 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.64 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.65 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.66 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.67 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.68 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.69 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.70 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.71 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.72 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.73 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.74 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.75 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.76 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.77 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.78 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.79 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.80 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.81 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.82 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.83 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.84 | SGP/EBBR/E8 - E8 EBBR Data Incorrect During System Replacement | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.85 | SGP/EBBR/E8 - E8 EBBR Data Not Based On 30 Minute Average | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.87 | SGP/EBBR/E8 - E8 EBBR Wind Direction Frozen | sgp30ebbrE8.b1, sgp15ebbrE8.b1, sgp5ebbrE8.b1 |
D010316.88 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.89 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.90 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.91 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.92 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.93 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.94 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.95 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.96 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.97 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.98 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.99 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010319.31 | SGP/EBBR/E7 - E7 EBBR AEM Frozen Stopped | sgp5ebbrE7.b1, sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D010319.32 | SGP/EBBR/E7 - E7 EBBR AEM Frozen Stopped | sgp5ebbrE7.b1, sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D010319.33 | SGP/EBBR/E7 - E7 EBBR AEM Frozen Stopped | sgp5ebbrE7.b1, sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D010319.34 | SGP/EBBR/E7 - E7 EBBR AEM Frozen Stopped | sgp5ebbrE7.b1, sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D010319.35 | SGP/EBBR/E7 - E7 EBBR AEM Frozen Stopped | sgp5ebbrE7.b1, sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D010319.36 | SGP/EBBR/E7 - E7 EBBR AEM Frozen Stopped | sgp5ebbrE7.b1, sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D010319.37 | SGP/EBBR/E7 - E7 EBBR AEM Frozen Stopped | sgp5ebbrE7.b1, sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D010319.40 | SGP/EBBR/E7 - E7 EBBR Soil Moisture # 2 Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010319.41 | SGP/EBBR/E7 - E7 EBBR Soil Moisture # 2 Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010319.42 | SGP/EBBR/E7 - E7 EBBR Soil Moisture # 2 Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010319.43 | SGP/EBBR/E7 - E7 EBBR Soil Moisture # 2 Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010319.44 | SGP/EBBR/E7 - E7 EBBR Soil Moisture # 2 Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010319.45 | SGP/EBBR/E7 - E7 EBBR Soil Moisture # 2 Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010319.46 | SGP/EBBR/E7 - E7 EBBR Soil Moistures 2, 4, and 5 Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010319.50 | SGP/EBBR E7 - E7 EBBR Soil Moisture #4 Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010319.51 | SGP/EBBR E7 - E7 EBBR Soil Moisture #4 Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010319.52 | SGP/EBBR E7 - E7 EBBR Soil Moisture #4 Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010319.53 | SGP/EBBR E7 - E7 EBBR Soil Moisture #4 Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010319.54 | SGP/EBBR/E7 - E7 EBBR soil Moisture #5 Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010319.55 | SGP/EBBR/E7 - E7 EBBR soil Moisture #5 Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010319.56 | SGP/EBBR/E7 - E7 EBBR soil Moisture #5 Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010319.57 | SGP/EBBR/E7 - E7 EBBR soil Moisture #5 Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010319.58 | SGP/EBBR/E7 - E7 EBBR soil Moisture #5 Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010319.59 | SGP/EBBR/E7 - E7 EBBR soil Moisture #5 Incorrect | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D010319.60 | SGP/EBBR/E7 - E7 EBBR soil Moisture #5 Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010319.61 | SGP/EBBR/E7 - E7 EBBR Soil Moisture #3 Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010319.62 | SGP/EBBR/E7 - E7 EBBR Soil Moisture #3 Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010319.63 | SGP/EBBR/E7 - E7 EBBR Soil Moisture #3 Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010319.64 | SGP/EBBR/E7 - E7 EBBR Soil Moisture #3 Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010319.65 | SGP/EBBR/E7 - E7 EBBR Soil Moisture #3 Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010319.66 | SGP/EBBR/E7 - E7 EBBR Soil Moisture #3 Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010319.67 | SGP/EBBR/E7 - E7 EBBR Soil Moisture #3 Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010319.68 | SGP/EBBR/E7 - E7 EBBR Soil Moisture #3 Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010319.69 | SGP/EBBR/E7 - E7 EBBR Soil Moisture #3 Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010319.70 | SGP/EBBR/E7 - E7 EBBR Soil Moisture #3 Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010319.71 | SGP/EBBR/E7 - E7 EBBR Soil Moisture #3 Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010319.72 | SGP/EBBR/E7 - E7 EBBR Soil Moisture #3 Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010319.73 | SGP/EBBR/E7 - E7 EBBR Soil Moisture #3 Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010319.74 | SGP/EBBR/E7 - E7 EBBR Soil Moisture #3 Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010319.75 | SGP/EBBR/E7 - E7 EBBR Soil Moisture #3 Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010319.76 | SGP/EBBR/E7 - E7 EBBR Soil Moisture #3 Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010319.77 | SGP/EBBR/E7 - E7 EBBR Soil Moisture #3 Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010319.78 | SGP/EBBR/E7 - E7 EBBR Soil Moisture #3 Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010319.79 | SGP/EBBR/E7 - E7 EBBR Soil Moisture #3 Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010319.80 | SGP/EBBR/E7 - E7 EBBR Soil Moisture #3 Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D010402.1 | SGP/EBBR/E9 - E9 EBBR Wind Direction Incorrect | sgp5ebbrE9.b1, sgp30ebbrE9.b1, sgp15ebbrE9.b1 |
D010402.13 | SGP/EBBR/E12 - E12 EBBR Wind Direction Incorrect | sgp30ebbrE12.b1, sgp5ebbrE12.b1, sgp15ebbrE12.b1 |
D010402.14 | SGP/EBBR/E12 - E12 EBBR Wind Speed Incorrect | sgp30ebbrE12.b1, sgp15ebbrE12.b1, sgp5ebbrE12.b1 |
D010402.16 | SGP/EBBR/E12 - E12 EBBR AEM Stuck - Data Incorrect | sgp30ebbrE12.b1, sgp15ebbrE12.b1, sgp5ebbrE12.b1 |
D010402.2 | SGP/EBBR/E9 - E9 EBBR Wind Speed Incorrect | sgp30ebbrE9.b1, sgp5ebbrE9.b1, sgp15ebbrE9.b1 |
D010402.3 | SGP/EBBR/E9 - E9 EBBR Soil Heat Flow 1 Incorrect | sgp30ebbrE9.b1, sgp15ebbrE9.b1 |
D010402.5 | SGP/EBBR/E9 - E9 EBBR Soil Heat Capacities Incorrect | sgp30ebbrE9.b1 |
D010403.10 | SGP/EBBR/E13 - E13 EBBR Flux Data Incorrect | sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
D010403.16 | SGP/EBBR/E13 - E13 EBBR All Data Incorrect | sgp5ebbrE13.b1, sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
D010403.17 | SGP/EBBR/E13 - E13 EBBR All Data Incorrect | sgp5ebbrE13.b1, sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
D010403.18 | SGP/EBBR/E13 - E13 EBBR All Data Incorrect | sgp5ebbrE13.b1, sgp30ebbrE13.b1, sgp15ebbrE13.b1 |
D010403.19 | SGP/EBBR/E13 - E13 EBBR All Data Incorrect | sgp5ebbrE13.b1, sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
D010403.2 | SGP/EBBR/E13 - E13 EBBR Heat Capacities Incorrect | sgp30ebbrE13.b1 |
D010403.20 | SGP/EBBR/E13 - E13 EBBR All Data Incorrect | sgp5ebbrE13.b1, sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
D010403.22 | SGP/EBBR/E13 - E13 EBBR All Data Incorrect | sgp5ebbrE13.b1, sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
D010403.24 | SGP/EBBR/E13 - E13 EBBR All Data Incorrect | sgp5ebbrE13.b1, sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
D010403.3 | SGP/EBBR/E13 - E13 EBBR Flux Data Incorrect | sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
D010403.35 | SGP/EBBR/E13 - E13 EBBR All Data Incorrect | sgp5ebbrE13.b1, sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
D010403.36 | SGP/EBBR/E13 - E13 EBBR All Data Incorrect | sgp5ebbrE13.b1, sgp30ebbrE13.b1, sgp15ebbrE13.b1 |
D010403.37 | SGP/EBBR/E13 - E13 EBBR All Data Incorrect | sgp5ebbrE13.b1, sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
D010403.38 | SGP/EBBR/E13 - E13 EBBR All Data Incorrect | sgp5ebbrE13.b1, sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
D010403.39 | SGP/EBBR/E13 - E13 EBBR All Data Incorrect | sgp5ebbrE13.b1, sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
D010403.4 | SGP/EBBR/E13 - E13 EBBR Flux Data Incorrect | sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
D010403.40 | SGP/EBBR/E13 - E13 EBBR All Data Incorrect | sgp5ebbrE13.b1, sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
D010403.41 | SGP/EBBR/E13 - E13 EBBR All Data Incorrect | sgp5ebbrE13.b1, sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
D010403.42 | SGP/EBBR/E13 - E13 EBBR All Data Incorrect | sgp5ebbrE13.b1, sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
D010403.43 | SGP/EBBR/E13 - E13 EBBR All Data Incorrect | sgp5ebbrE13.b1, sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
D010403.44 | SGP/EBBR/E13 - E13 EBBR All Data Incorrect | sgp5ebbrE13.b1, sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
D010403.45 | SGP/EBBR/E13 - E13 EBBR All Data Incorrect | sgp5ebbrE13.b1, sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
D010403.6 | SGP/EBBR/E13 - E13 EBBR Flux Data Incorrect | sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
D010403.8 | SGP/EBBR/E13 - E13 EBBR Flux Data Incorrect | sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
D010403.9 | SGP/EBBR/E13 - E13 EBBR Flux Data Incorrect | sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
D010412.1 | SGP/EBBR/E13 - E13 EBBR Soil Temperature #4 Suspect | sgp30ebbrE13.b1, sgp15ebbrE13.b1 |
D010413.1 | SGP/EBBR/E13 - E13 EBBR Wind Direction Iced | sgp5ebbrE13.b1, sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
D010413.2 | SGP/EBBR/E13 - E13 EBBR Soil Heat Flow #4 Incorrect | sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
D010413.3 | SGP/EBBR/E15 - E15 EBBR Reference Temperature Incorrect | sgp5ebbrE15.b1, sgp30ebbrE15.b1, sgp15ebbrE15.b1 |
D010413.4 | SGP/EBBR/E15 - E15 EBBR Data Incorrect | sgp30ebbrE15.b1, sgp15ebbrE15.b1, sgp5ebbrE15.b1, sgp5ebbrE15.00 |
D010413.5 | SGP/EBBR/E15 - E15 EBBR Data Missing | sgp30ebbrE15.b1, sgp15ebbrE15.b1, sgp5ebbrE15.b1, sgp5ebbrE15.00 |
D010413.6 | SGP/EBBR/E15 - E15 EBBR Data Incorrect | sgp30ebbrE15.b1, sgp15ebbrE15.b1, sgp5ebbrE15.b1, sgp5ebbrE15.00 |
D010413.7 | SGP/EBBR/E15 - E15 EBBR Data Missing | sgp30ebbrE15.b1, sgp15ebbrE15.b1, sgp5ebbrE15.b1, sgp5ebbrE15.00 |
D010413.8 | SGP/EBBR/E15 - E15 EBBR Some Soil Temperatures and Heat Flows Incorrect | sgp30ebbrE15.b1, sgp15ebbrE15.b1 |
D010413.9 | SGP/EBBR/E15 - E15 EBBR Some Soils and Flux Data Incorrect | sgp30ebbrE15.b1 |
D010416.1 | SGP/EBBR/E15 - E15 EBBR AEM Stuck in Middle | sgp5ebbrE15.b1, sgp30ebbrE15.b1, sgp15ebbrE15.b1 |
D010416.10 | SGP/EBBR/E18 - E18 EBBR Data Incorrect | sgp30ebbrE18.b1, sgp15ebbrE18.b1, sgp5ebbrE18.b1 |
D010416.11 | SGP/EBBR/E18 - E18 EBBR Right RH Incorrect | sgp30ebbrE18.b1, sgp5ebbrE18.b1, sgp15ebbrE18.b1 |
D010416.12 | SGP/EBBR/E18 - E18 EBBR Thum Right Incorrect | sgp30ebbrE18.b1, sgp15ebbrE18.b1, sgp5ebbrE18.b1 |
D010416.2 | SGP/EBBR/E15 - E15 EBBR AEM Replacement | sgp5ebbrE15.b1, sgp30ebbrE15.b1, sgp15ebbrE15.b1 |
D010416.3 | SGP/EBBR/E15 - E15 EBBR Battery Voltage Low | sgp15ebbrE15.b1, sgp5ebbrE15.b1, sgp30ebbrE15.b1 |
D010416.4 | SGP/EBBR/E15 - E15 EBBR Incorrect Net Radiation | sgp5ebbrE15.b1, sgp30ebbrE15.b1, sgp15ebbrE15.b1 |
D010416.8 | SGP/EBBR/E18 - E18 EBBR Data Incorrect | sgp30ebbrE18.b1, sgp15ebbrE18.b1, sgp5ebbrE18.b1 |
D010416.9 | SGP/EBBR/E18 - E18 EBBR Data Incorrect | sgp30ebbrE18.b1, sgp15ebbrE18.b1, sgp5ebbrE18.b1 |
D010427.1 | SGP EBBR E18 - E18 EBBR Wind Direction Frozen | sgp30ebbrE18.b1, sgp15ebbrE18.b1, sgp5ebbrE18.b1 |
D010427.11 | SGP EBBR E22 - E22 EBBR Data Incorrect During Logger Replacement | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1, sgp5ebbrE22.00 |
D010427.13 | SGP EBBR E22 - E22 EBBR Soil Moisture Lower Than Measurement Capability | sgp30ebbrE22.b1, sgp15ebbrE22.b1 |
D010427.14 | SGP EBBR E22 - E22 EBBR Soil Moisture #3 Lower Than Can Be Measured | sgp30ebbrE22.b1, sgp15ebbrE22.b1 |
D010427.15 | SGP EBBR E22 - E22 EBBR Net Radiation Incorrect | sgp15ebbrE22.b1, sgp5ebbrE22.b1, sgp30ebbrE22.b1 |
D010427.2 | SGP EBBR E18 - E18 EBBR Wind Direction Frozen | sgp30ebbrE18.b1, sgp15ebbrE18.b1, sgp5ebbrE18.b1 |
D010427.3 | SGP EBBR E18 - E18 EBBR Left Ventilator Fan Not Working | sgp30ebbrE18.b1, sgp5ebbrE18.b1, sgp15ebbrE18.b1 |
D010427.4 | SGP EBBR E19 - E19 EBBR Left RH Probe Temperature Incorrect | sgp5ebbrE19.b1, sgp30ebbrE19.b1, sgp15ebbrE19.b1 |
D010427.5 | SGP EBBR E19 - E19 EBBR Left Aspirator Housing Off | sgp5ebbrE19.b1, sgp30ebbrE19.b1, sgp15ebbrE19.b1 |
D010427.6 | SGP EBBR E19 - E19 EBBR Left Air Temperature Incorrect | sgp30ebbrE19.b1, sgp5ebbrE19.b1, sgp15ebbrE19.b1 |
D010427.7 | SGP EBBR E19 - E19 EBBR Data Incorrect During J-panel Replacement | sgp15ebbrE19.b1, sgp5ebbrE19.b1, sgp30ebbrE19.b1 |
D010427.8 | SGP EBBR E19 - E19 EBBR Wind Direction Incorrect | sgp30ebbrE19.b1, sgp5ebbrE19.b1, sgp15ebbrE19.b1 |
D010427.9 | SGP EBBR E19 - E19 EBBR Soil Heat Flow #1 Sign Reversed | sgp15ebbrE19.b1, sgp30ebbrE19.b1 |
D010517.1 | SGP EBBR E20 - E20 EBBR Soil Heat Flow #3 Incorrect | sgp15ebbrE20.b1, sgp30ebbrE20.b1 |
D010517.2 | SGP EBBR E20 - E20 EBBR Right Air T Incorrect | sgp5ebbrE20.b1, sgp30ebbrE20.b1, sgp15ebbrE20.b1 |
D010517.3 | SGP EBBR E20 - E20 EBBR Right Air T Incorrect | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D010605.11 | SGP EBBR E20 - E20 EBBR Soil Temperatures Incorrect | sgp30ebbrE20.b1, sgp15ebbrE20.b1 |
D010605.12 | SGP EBBR E20 - E20 EBBR Data Incorrect During STP #3 Replacement | sgp30ebbrE20.b1, sgp15ebbrE20.b1, sgp5ebbrE20.b1, sgp5ebbrE20.00 |
D010605.13 | SGP EBBR E20 - E20 EBBR Data Incorrect After Program Restart | sgp30ebbrE20.b1, sgp15ebbrE20.b1, sgp5ebbrE20.b1, sgp5ebbrE20.00 |
D010605.14 | SGP EBBR E20 - E20 EBBR Some Data Incorrect | sgp30ebbrE20.b1 |
D010605.16 | SGP EBBR E25 - E25 EBBR Soil Heat Flow #5 Incorrect | sgp30ebbrE25.b1, sgp15ebbrE25.b1 |
D010605.17 | SGP EBBR E25 - E25 EBBR Soil Heat Flow #3 Incorrect | sgp30ebbrE25.b1, sgp15ebbrE25.b1 |
D010605.18 | SGP EBBR E25 - E25 EBBR Left Aspirator Not Working | sgp30ebbrE25.b1, sgp15ebbrE25.b1, sgp5ebbrE25.b1 |
D010605.19 | SGP EBBR E25 - E25 EBBR Soil Heat Flow #1 Incorrect | sgp30ebbrE25.b1, sgp15ebbrE25.b1 |
D010605.20 | SGP EBBR E26 - Data Suspect, Low Battery Voltage | sgp30ebbrE26.b1, sgp15ebbrE26.b1, sgp5ebbrE26.b1, sgp5ebbrE26.00 |
D010605.22 | SGP EBBR E26 - All Data Incorrect | sgp30ebbrE26.b1, sgp15ebbrE26.b1, sgp5ebbrE26.b1, sgp5ebbrE26.00 |
D010605.23 | SGP EBBR E26 - Soil Heat Flow #1 Incorrect | sgp30ebbrE26.b1, sgp15ebbrE26.b1 |
D010605.24 | SGP EBBR E26 - Soil Heat Flow #2 Incorrect | sgp15ebbrE26.b1, sgp30ebbrE26.b1 |
D010605.25 | SGP EBBR E26 - Soil Heat Flow #3 Incorrect | sgp30ebbrE26.b1, sgp15ebbrE26.b1 |
D010605.5 | SGP EBBR E13 - E13 EBBR Soil T #4 Incorrect | sgp30ebbrE13.b1, sgp15ebbrE13.b1 |
D010605.7 | SGP EBBR E20 - E20 EBBR Data Incorrect | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1, sgp5ebbrE20.00 |
D010605.8 | SGP EBBR E20 - E20 EBBR Data Incorrect | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1, sgp5ebbrE20.00 |
D010605.9 | SGP EBBR E20 - E20 EBBR Net Radiation Data Incorrect | sgp30ebbrE20.b1, sgp5ebbrE20.b1, sgp15ebbrE20.b1 |
D010723.3 | SGP EBBR E26 - Soil Moisture #1 Incorrect | sgp30ebbrE26.b1, sgp15ebbrE26.b1 |
D010723.4 | SGP EBBR E26 - AEM Inbetween, Data Incorrect | sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp15ebbrE26.b1 |
D010723.5 | SGP EBBR E26 - Soil Heat Flow #5 Incorrect | sgp30ebbrE26.b1, sgp15ebbrE26.b1 |
D010723.7 | SGP EBBR E26 - Some Data Incorrect | sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp15ebbrE26.b1 |
D010723.8 | SGP EBBR E26 - Some Data Incorrect | sgp30ebbrE26.b1 |
D011102.3 | SGP EBBR E19 - E19 EBBR Soil Temperature 3 Intermittenmtly Incorrect | sgp30ebbrE19.b1, sgp15ebbrE19.b1 |
D011102.7 | SGP EBBR E19 - E19 EBBR TS3 Intermittent Incorrect | sgp30ebbrE19.b1, sgp15ebbrE19.b1 |
D011102.8 | SGP EBBR E19 - E19 EBBR Wind Speed Incorrect | sgp30ebbrE19.b1, sgp5ebbrE19.b1, sgp15ebbrE19.b1 |
D011127.21 | SGP EBBR E2 - E2 EBBR Data Incorrect During 6 Month Checks | sgp30ebbrE2.b1, sgp15ebbrE2.b1, sgp5ebbrE2.b1 |
D011127.23 | SGP EBBR E7 - E7 EBBR Data Incorrect During 6 Month Checks | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D011127.24 | SGP EBBR E8 - E8 EBBR Data Incorrect During 6 Month Checks | sgp30ebbrE8.b1, sgp15ebbrE8.b1, sgp5ebbrE8.b1 |
D011127.25 | SGP EBBR E9 - E9 EBBR Data Incorect During 6 Month Checks | sgp5ebbrE9.b1, sgp15ebbrE9.b1, sgp30ebbrE9.b1 |
D011127.26 | SGP EBBR E12 - E12 EBBR Data Incorrect During 6 Month Checks | sgp30ebbrE12.b1, sgp15ebbrE12.b1, sgp5ebbrE12.b1 |
D011127.27 | SGP EBBR E13 - E13 EBBR Data Incorrect During 6 Month Checks | sgp5ebbrE13.b1, sgp30ebbrE13.b1, sgp15ebbrE13.b1 |
D011127.28 | SGP EBBR E15 - E15 EBBR Data Incorrect During 6 Month Checks | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D011127.29 | SGP EBBR E18 - E18 EBBR Data Incorrect During 6 Month Checks | sgp5ebbrE18.b1, sgp30ebbrE18.b1, sgp15ebbrE18.b1 |
D011127.30 | SGP EBBR E19 - E19 EBBR Data Incorrect During 6 Month Checks | sgp30ebbrE19.b1, sgp15ebbrE19.b1, sgp5ebbrE19.b1 |
D011127.31 | SGP EBBR E20 - E20 EBBR Data Incorrect During 6 Month Checks | sgp30ebbrE20.b1, sgp15ebbrE20.b1, sgp5ebbrE20.b1 |
D011127.32 | SGP EBBR E22 - E22 EBBR Data Incorrect During 6 Month Checks | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D011127.33 | SGP EBBR E25 - E25 EBBR Data Incorrect During 6 Month Checks | sgp15ebbrE25.b1, sgp5ebbrE25.b1, sgp30ebbrE25.b1 |
D011127.34 | SGP EBBR E26 - Data Incorrect During 6 Month Checks | sgp30ebbrE26.b1, sgp15ebbrE26.b1, sgp5ebbrE26.b1 |
D011127.4 | SGP EBBR E20 - E20 EBBR Soil Temperature #3 Incorrect | sgp30ebbrE20.b1, sgp15ebbrE20.b1 |
D011129.2 | SGP EBBR E7 - E7 EBBR Data Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D011129.3 | SGP EBBR E7 - E7 EBBR Soil Moisture #3 and Other Data Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D011129.4 | SGP EBBR E7 - E7 EBBR AEM Position Caused Incorrect Data | sgp5ebbrE7.b1, sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D011129.5 | SGP EBBR E7 - E7 EBBR Soil Moisture #2 Spiked, Incorrect Data Resulted | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D011129.6 | SGP EBBR E7 - E7 EBBR AEM Position Caused Incorrect Data | sgp30ebbrE7.b1, sgp15ebbrE7.b1, sgp5ebbrE7.b1 |
D011129.7 | SGP EBBR E7 - E7 EBBR All Data Incorrect | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D011129.8 | SGP EBBR E7 - E7 EBBR Data Incorrect After Recalibration | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D011129.9 | SGP EBBR E7 - E7 EBBR Data Incorrect After Missing Data Period | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D011206.1 | SGP EBBR E26 - Low Battery Voltage | sgp5ebbrE26.b1, sgp30ebbrE26.b1, sgp15ebbrE26.b1 |
D011206.11 | SGP EBBR E25 - E25 EBBR Wind Direction Incorrect | sgp30ebbrE25.b1, sgp5ebbrE25.b1 |
D011206.12 | SGP EBBR E25 - E25 EBBR Wind Direction Incorrect | sgp30ebbrE25.b1, sgp5ebbrE25.b1 |
D011206.2 | SGP EBBR E26 - Battery Voltage Low | sgp30ebbrE26.b1, sgp15ebbrE26.b1, sgp5ebbrE26.b1 |
D011206.4 | SGP EBBR E22 - E22 EBBR 30 Minute Data Incorrect After Missing Data | sgp30ebbrE22.b1 |
D011206.6 | SGP EBBR E25 - E25 EBBR Soil Heat Flow #3 Incorrect | sgp30ebbrE25.b1, sgp15ebbrE25.b1 |
D011206.7 | SGP EBBR E25 - E25 EBBR Soil Heat Flow #1 Incorrect | sgp30ebbrE25.b1, sgp15ebbrE25.b1 |
D011206.8 | SGP EBBR E25 - E25 EBBR Soil Heat Flow #3 Incorrect | sgp30ebbrE25.b1, sgp15ebbrE25.b1 |
D011206.9 | SGP EBBR E25 - E25 EBBR Soil Heat Flow #1 Incorrect | sgp30ebbrE25.b1, sgp15ebbrE25.b1 |
D011207.11 | SGP EBBR E12 - E12 EBBR Data Incorrect From AEM In One Position | sgp5ebbrE12.b1, sgp30ebbrE12.b1, sgp15ebbrE12.b1 |
D011207.12 | SGP EBBR E15 - E15 EBBR Battery Voltage Low, Data Incorrect | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D011218.1 | SGP EBBR E15 - E15 EBBR Data Zeros Incorrect | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D011218.10 | SGP EBBR E20 - E20 EBBR Soil Heat Flow #2 Incorrect | sgp30ebbrE20.b1, sgp15ebbrE20.b1 |
D011218.11 | SGP EBBR E20 - E20 EBBR Battery Voltage Low, Incorrect Data | sgp15ebbrE20.b1, sgp30ebbrE20.b1 |
D011218.13 | SGP EBBR E20 - E20 EBBR Battery Voltage Low, Data Incorrect | sgp15ebbrE20.b1, sgp30ebbrE20.b1 |
D011218.14 | SGP/EBBR/E20 - Reprocess: Soil Temperature #3 Incorrect | sgp15ebbrE20.b1, sgp30ebbrE20.b1 |
D011218.5 | SGP EBBR E18 - E18 EBBR AEM Frozen, Data Incorrect | sgp30ebbrE18.b1, sgp5ebbrE18.b1, sgp15ebbrE18.b1 |
D011218.6 | SGP EBBR E18 - E18 EBBR Wind Direction Frozen, Data Incorrect | sgp5ebbrE18.b1, sgp15ebbrE18.b1, sgp30ebbrE18.b1 |
D011218.7 | SGP EBBR E18 - E18 EBBR Wind Speed Frozen, Data Incorrect | sgp15ebbrE18.b1, sgp5ebbrE18.b1, sgp30ebbrE18.b1 |
D011218.8 | SGP EBBR E18 - E18 EBBR Right RH Incorrect | sgp30ebbrE18.b1, sgp5ebbrE18.b1, sgp15ebbrE18.b1 |
D011218.9 | SGP EBBR E20 - E20 EBBR Soil Heat Flow #4 Incorrect | sgp30ebbrE20.b1, sgp15ebbrE20.b1 |
D011219.2 | SGP EBBR E8 - E8 EBBR Soil Heat Flow #1 Incorrect | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D011219.3 | SGP EBBR E8 - E8 EBBR Soil Heat Flow #1 Incorrect | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D011219.4 | SGP EBBR E8 - E8 EBBR AEM Frozen In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D011219.5 | SGP EBBR E8 - E8 EBBR Wind Speed Frozen | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D011220.1 | SGP EBBR E8 - E8 EBBR Battery Voltage Low, Data Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.00, sgp15ebbrE8.b1 |
D011220.2 | SGP EBBR E8 - E8 EBBR Battery Voltage Low, Data Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1, sgp5ebbrE8.00 |
D011221.2 | SGP EBBR E9 - E9 EBBR AEM Hung, Data Incorrect | sgp30ebbrE9.b1, sgp5ebbrE9.b1, sgp15ebbrE9.b1 |
D011221.3 | SGP EBBR E9 - E9 EBBR AEM Hung, Data Incorrect | sgp5ebbrE9.b1, sgp30ebbrE9.b1, sgp15ebbrE9.b1 |
D011221.4 | SGP EBBR E9 - E9 EBBR Wind Speed Frozen Stopped | sgp30ebbrE9.b1, sgp5ebbrE9.b1, sgp15ebbrE9.b1 |
D011221.5 | SGP EBBR E9 - E9 EBBR Wind Direction Frozen In One Direction | sgp5ebbrE9.b1, sgp30ebbrE9.b1, sgp15ebbrE9.b1 |
D011221.6 | SGP/EBBR/E9 - Reprocess: E9 EBBR Soil Temperature #5 Intermittently Incorrect | sgp30ebbrE9.b1, sgp15ebbrE9.b1 |
D011221.8 | SGP EBBR E9 - E9 EBBR Gradients Incorrect During Sensor Replacement | sgp5ebbrE9.b1, sgp30ebbrE9.b1, sgp15ebbrE9.b1 |
D020110.2 | SGP EBBR E9 - E9 EBBR Right Air Temperature Incorrect | sgp5ebbrE9.b1, sgp15ebbrE9.b1, sgp30ebbrE9.b1 |
D020115.1 | SGP/EEBR/E2 - AEM Stopped In One Position | sgp30ebbrE2.b1, sgp15ebbrE2.b1, sgp5ebbrE2.b1 |
D020115.2 | SGP/EEBR/E2 - AEM Stopped In One Position | sgp5ebbrE2.b1, sgp30ebbrE2.b1, sgp15ebbrE2.b1 |
D020115.3 | SGP/EEBR/E2 - Wind Direction Frozen | sgp30ebbrE2.b1, sgp5ebbrE2.b1, sgp15ebbrE2.b1 |
D020115.4 | SGP/EEBR/E2 - Wind Speed Sensor Frozen | sgp30ebbrE2.b1, sgp5ebbrE2.b1, sgp15ebbrE2.b1 |
D020115.5 | SGP/EEBR/E2 - Soil Temperature #2 Incorrect | sgp30ebbrE2.b1, sgp15ebbrE2.b1 |
D020117.14 | SGP/EBBR/E2 - Reprocess: E2 EBBR Soil Temperature #2 Incorrect | sgp30ebbrE2.b1, sgp15ebbrE2.b1 |
D020117.23 | SGP/EBBR/E2 - Reprocess: E2 EBBR Soil Temperature #2 Incorrect | sgp30ebbrE2.b1, sgp15ebbrE2.b1 |
D020117.4 | SGP/EBBR/E2 - Reprocess: E2 EBBR Soil Temperature #2 Incorrect | sgp30ebbrE2.b1, sgp15ebbrE2.b1 |
D020129.1 | SGP EBBR E15 - E15 EBBR Recalibration Changeout | sgp30ebbrE15.b1, sgp15ebbrE15.b1, sgp5ebbrE15.b1, sgp5ebbrE15.00 |
D020129.10 | SGP/EBBR/E13 - Reprocess: E13 EBBR Soil heat Flows #1 and #2 Incorrect | sgp30ebbrE13.b1, sgp15ebbrE13.b1 |
D020129.11 | SGP EBBR E13 - E13 EBBR Data Incorrect | sgp30ebbrE13.b1, sgp15ebbrE13.b1, sgp5ebbrE13.b1 |
D020129.12 | SGP EBBR E13 - E13 EBBR Left T/RH Probe Temperature Incorrect | sgp30ebbrE13.b1, sgp15ebbrE13.b1 |
D020129.13 | SGP EBBR E13 - E13 EBBR Left T/RH Probe Temperature Incorrect | sgp30ebbrE13.b1, sgp15ebbrE13.b1 |
D020129.2 | SGP EBBR E13 - E13 EBBR AEM in Mid Position | sgp30ebbrE13.b1, sgp15ebbrE13.b1, sgp5ebbrE13.b1 |
D020129.3 | SGP EBBR E13 - E13 EBBR AEM Home 30 Offset | sgp30ebbrE13.b1, sgp15ebbrE13.b1, sgp5ebbrE13.b1 |
D020129.4 | SGP EBBR E13 - E13 EBBR AEM Home 30 Offset or AEM in Mid Position | sgp30ebbrE13.b1, sgp15ebbrE13.b1, sgp5ebbrE13.b1 |
D020129.5 | SGP EBBR E13 - E13 EBBR Left Air Temperature Incorrect | sgp30ebbrE13.b1, sgp15ebbrE13.b1, sgp5ebbrE13.b1 |
D020129.7 | SGP EBBR E13 - E13 EBBR Left Air Temperature Incorrect | sgp30ebbrE13.b1, sgp15ebbrE13.b1, sgp5ebbrE13.b1 |
D020129.8 | SGP/EBBR/E13 - Reprocess: E13 EBBR Soil Temperature #4 Incorrect | sgp30ebbrE13.b1, sgp15ebbrE13.b1 |
D020129.9 | SGP/EBBR/E13 - Reprocess: E13 EBBR Soil Temperature #4 Incorrect | sgp30ebbrE13.b1, sgp15ebbrE13.b1 |
D020131.13 | SGP/EBBR/E13 - Left T/RH Probe Temperature Incorrect | sgp30ebbrE13.b1, sgp15ebbrE13.b1, sgp5ebbrE13.b1 |
D020214.1 | SGP/EBBR/E13 - Left Humidity Probe Temperature Incorrect | sgp30ebbrE13.b1, sgp15ebbrE13.b1, sgp5ebbrE13.b1 |
D020214.2 | SGP/EBBR/E13 - Data incorrect during maintenance | sgp30ebbrE13.b1, sgp15ebbrE13.b1, sgp5ebbrE13.b1 |
D020214.5 | SGP/EBBR/E13 - Data Incorrect During 6 Month Checks | sgp30ebbrE13.b1, sgp15ebbrE13.b1, sgp5ebbrE13.b1 |
D020214.7 | SGP/EBBR/E15 - Data Incorrect During 6 Month Checks | sgp30ebbrE15.b1, sgp15ebbrE15.b1, sgp5ebbrE15.b1 |
D020626.28 | SGP/EBBR/E2 - EBBR E2 Wind Speed Sensors Frozen | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
D020628.1 | SGP/EBBR/E12 - 3-month Check | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D020628.11 | SGP/EBBR/E7 - T/RH Check | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D020628.13 | SGP/EBBR/E8 - T/RH Check | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D020628.14 | SGP/EBBR/E9 - T/RH Check | sgp30ebbrE9.b1, sgp5ebbrE9.b1, sgp15ebbrE9.b1 |
D020628.15 | SGP/EBBR/E12 - T/RH Check | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D020628.16 | SGP/EBBR/E13 - T/RH Check | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D020628.17 | SGP/EBBR/E15 - T/RH Check | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D020628.18 | SGP/EBBR/E20 - T/RH Check | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D020628.19 | SGP/EBBR/E22 - T/RH Check | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D020628.2 | SGP/EBBR/E13 - 3-month Check | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D020628.3 | SGP/EBBR/E15 - 3-month Check | sgp5ebbrE15.b1, sgp30ebbrE15.b1, sgp15ebbrE15.b1 |
D020628.4 | SGP/EBBR/E20 - 3-month Check | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D020628.5 | SGP/EBBR/E22 - 3-month Check | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D020628.6 | SGP/EBBR/E26 - 3-month Check | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D020628.8 | SGP/EBBR/E7 - 3-month Check | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D020628.9 | SGP/EBBR/E8 - 3-month Check | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D020716.1 | SGP/EBBR/E9 - Annual calibration changeout | sgp30ebbrE9.b1, sgp15ebbrE9.b1 |
D020716.10 | SGP/EBBR/E9 - 6 Month Checks | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
D020716.11 | SGP/EBBR/E12 - 6 Month Checks | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D020716.12 | SGP/EBBR/E13 - 6 Month Checks | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D020716.13 | SGP/EBBR/E15 - 6 Month Checks | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D020716.14 | SGP/EBBR/E18 - 6 Month Checks | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
D020716.15 | SGP/EBBR/E19 - 6 Month Checks | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
D020716.16 | SGP/EBBR/E20 - 6 Month Checks | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D020716.17 | SGP/EBBR/E22 - 6 Month Checks | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D020716.18 | SGP/EBBR/E26 - 6 Month Checks | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D020716.19 | SGP/EBBR/E2 - AEM Hung | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
D020716.2 | SGP/EBBR/E9 - Annual calibration changeout | sgp30ebbrE9.b1, sgp5ebbrE9.b1, sgp15ebbrE9.b1 |
D020716.20 | SGP/EBBR/E2 - Wind Direction Incorrect | sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
D020716.21 | SGP/EBBR/E2 - Reprocess: Soil Temperature #2 Incorrect | sgp30ebbrE2.b1 |
D020716.22 | SGP/EBBR/E2 - Data Incorrect | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.00, sgp5ebbrE2.b1 |
D020716.26 | SGP/EBBR/E7 - Measurements Zero | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.00, sgp5ebbrE7.b1 |
D020716.27 | SGP/EBBR/E7 - AEM Hung | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D020716.28 | SGP/EBBR/E7 - Wind Speed Incorrect | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D020716.29 | SGP/EBBR/E8 - Low Battery Voltage Causes Incorrect Measurements | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.00, sgp5ebbrE8.b1 |
D020716.3 | SGP/EBBR/E22 - Annual Calibration Changeout | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D020716.30 | SGP/EBBR/E8 - AEM Hung | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D020716.5 | SGP/EBBR/E2 - 6 Month Checks | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
D020716.6 | SGP/EBBR/E2 - 6 Month Checks | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
D020716.8 | SGP/EBBR/E7 - 6 Month Checks | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D020716.9 | SGP/EBBR/E8 - 6 Month Checks | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D020717.10 | SGP/EBBR/E12 - AEM Hung | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D020717.11 | SGP/EBBR/E12 - Measurements Incorrect | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.00, sgp5ebbrE12.b1 |
D020717.12 | SGP/EBBR/E12 - Resistance Measurements Incorrect | sgp30ebbrE12.b1, sgp5ebbrE12.b1, sgp15ebbrE12.b1 |
D020717.13 | SGP/EBBR/E13 - Reprocess: Soil Heat Flow #1 Incorrect | sgp30ebbrE13.b1 |
D020717.14 | SGP/EBBR/E13 - Reprocess: Soil Heat Flow #2 Incorrect | sgp30ebbrE13.b1 |
D020717.15 | SGP/EBBR/E13 - Wrong CR10 Program Installed | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.00, sgp5ebbrE13.b1 |
D020717.16 | SGP/EBBR/E13 - Wind Direction Incorrect | sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D020717.17 | SGP/EBBR/E13 - AEM Hung up | sgp30ebbrE13.b1, sgp5ebbrE13.b1, sgp15ebbrE13.b1 |
D020717.2 | SGP/EBBR/E9 - Reprocess: Soil Temperature #5 Incorrect | sgp30ebbrE9.b1 |
D020717.20 | SGP/EBBR/E18 - Version 6 Produced Noisy Fluxes | sgp30ebbrE18.b1 |
D020717.21 | SGP/EBBR/E18 - AEM Hung | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
D020717.22 | SGP/EBBR/E18 - Soil Heat Flows Incorrect After Calibration Changeout | sgp30ebbrE18.b1 |
D020717.23 | SGP/EBBR/E18 - Reprocess: Soil Heat Flow #1 Incorrect | sgp30ebbrE18.b1 |
D020717.24 | SGP/EBBR/E18 - Left Air Temperature Incorrect | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
D020717.25 | SGP/EBBR/E18 - Reference Temperature Incorrect | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
D020717.26 | SGP/EBBR/E19 - Reprocess: Soil Heat Flow #3 Incorrect | sgp30ebbrE19.b1 |
D020717.27 | SGP/EBBR/E19 - Reprocess: Soil Temperature #3 Incorrect | sgp30ebbrE19.b1 |
D020717.29 | SGP/EBBR/E19 - Most Measurements Incorrect | sgp15ebbrE19.b1, sgp5ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.00 |
D020717.3 | SGP/EBBR/E9 - Measurements All Zeros | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.00, sgp5ebbrE9.b1 |
D020717.30 | SGP/EBBR/E19 - All Measurements Zero | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.00, sgp5ebbrE19.b1 |
D020717.31 | SGP/EBBR/E19 - Pressure Incorrect | sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
D020717.32 | SGP/EBBR/E19 - Measurements Incorrect After CR10 Replacement | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.00, sgp5ebbrE19.b1 |
D020717.33 | SGP/EBBR/E19 - Low Battery Voltage | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.00, sgp5ebbrE19.b1 |
D020717.34 | SGP/EBBR/E22 - All Measurements Zero | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.00, sgp5ebbrE22.b1 |
D020717.35 | SGP/EBBR/E22 - AEM Hung or High Home Signals | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D020717.36 | SGP/EBBR/E25 - Reprocess: Soil Heat Flow #5 Negative Offset | sgp30ebbrE25.b1 |
D020717.37 | SGP/EBBR/E26 - Measurements Zero | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.00, sgp5ebbrE26.b1 |
D020717.38 | SGP/EBBR/E26 - Measurements Incorrect | sgp30ebbrE26.b1 |
D020717.39 | SGP/EBBR/E26 - Multiplexer 2 Failure From High Temperature | sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp15ebbrE26.b1 |
D020717.4 | SGP/EBBR/E9 - Humidity Probe Measurements Incorrect | sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
D020717.40 | SGP/EBBR/E20 - AEM Hung | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D020717.42 | SGP/EBBR/E20 - Battery Voltage Low | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.00, sgp5ebbrE20.b1 |
D020717.43 | SGP/EBBR/E20 - Pressure Almost Constant | sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D020717.45 | SGP/EBBR/E20 - Reprocess: Soil Heat Flow #3 Incorrect During Sensor Changeout | sgp30ebbrE20.b1 |
D020717.46 | SGP/EBBR/E20 - Reprocess: Soil Heat Flow #2 Has a Negative Offset | sgp30ebbrE20.b1 |
D020717.47 | SGP/EBBR/E20 - Reprocess: Soil Heat Flow #4 Incorrect | sgp30ebbrE20.b1 |
D020717.48 | SGP/EBBR/E20 - Measurements Zero | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.00, sgp5ebbrE20.b1 |
D020717.49 | SGP/EBBR/E20 - Reprocess: Soil Temperature #3 Incorrect | sgp30ebbrE20.b1 |
D020717.5 | SGP/EBBR/E9 - Right Humidity Suspect | sgp5ebbrE9.b1, sgp30ebbrE9.b1 |
D020717.50 | SGP/EBBR/E20 - Measurements Incorrect During Sensor Changeout | sgp30ebbrE20.b1 |
D020717.51 | SGP/EBBR/E20 - Measurements Incorrect | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.00, sgp5ebbrE20.b1 |
D020717.6 | SGP/EBBR/E12 - Reprocess: Soil Temperature #3 Incorrect | sgp30ebbrE12.b1 |
D020717.7 | SGP/EBBR/E12 - Wind Direction Incorrect | sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D020717.8 | SGP/EBBR/E12 - Reprocess: Soil Moisture 4 and 5 Incorrect | sgp30ebbrE12.b1 |
D020717.9 | SGP/EBBR/E12 - Reprocess: Soil Moisture 2 and 5 Incorrect | sgp30ebbrE12.b1 |
D020723.1 | SGP/EBBR/E13 - Thermocouple out of range, data incorrect | sgp30ebbrE13.b1, sgp5ebbrE13.b1, sgp15ebbrE13.b1 |
D020724.48 | SGP/EBBR/E26 - Annual Calibration Changeout | sgp5ebbrE26.b1, sgp15ebbrE26.b1, sgp30ebbrE26.b1 |
D020726.1 | SGP/EBBR/E20 - Reprocess: Soil Heat Flow #2 and #4 incorrect | sgp30ebbrE20.b1 |
D020726.2 | SGP/EBBR/E13 - Reprocess: Soil Heat Flow #1 and #2 Incorrect | sgp30ebbrE13.b1 |
D021113.1 | SGP/EBBR - Incorrect Units for Soil Heat Capacity (cs1, cs2, cs3, cs4, cs5) in 30EBBR Header | sgp30ebbrE12.b1, sgp30ebbrE13.b1, sgp30ebbrE15.b1, sgp30ebbrE18.b1, sgp30ebbrE19.b1, sgp30ebbrE2.b1, sgp30ebbrE20.b1, sgp30ebbrE22.b1, sgp30ebbrE26.b1, sgp30ebbrE4.b1, sgp30ebbrE7.b1, sgp30ebbrE8.b1, sgp30ebbrE9.b1 |
D021113.4 | SGP/EBBR - Incorrect Units for Soil Heat Capacity (cs1, cs2, cs3, cs4, cs5) in 30EBBR Header | sgp30ebbrE12.b1, sgp30ebbrE13.b1, sgp30ebbrE15.b1, sgp30ebbrE20.b1, sgp30ebbrE22.b1, sgp30ebbrE26.b1, sgp30ebbrE4.b1, sgp30ebbrE7.b1, sgp30ebbrE8.b1, sgp30ebbrE9.b1 |
D030110.23 | SGP/EBBR/E18 - Data Incorrect during 6 month checks | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
D030110.24 | SGP/EBBR/E19 - Data Incorrect during 6 month checks | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
D030110.25 | SGP/EBBR/E20 - Data Incorrect during 6 month checks | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D030110.26 | SGP/EBBR/E22 - Data Incorrect during 6 month checks | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D030110.27 | SGP/EBBR/E26 - Data Incorrect during 6 month checks | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D030110.28 | SGP/EBBR/E2 - Data Incorrect during 6 month checks | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
D030110.30 | SGP/EBBR/E7 - Data Incorrect during 6 month checks | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D030110.31 | SGP/EBBR/E8 - Data Incorrect during 6 month checks | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D030110.32 | SGP/EBBR/E9 - Data Incorrect during 6 month checks | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
D030110.33 | SGP/EBBR/E12 - Data Incorrect during 6 month checks | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D030110.34 | SGP/EBBR/E13 - Data Incorrect during 6 month checks | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D030110.35 | SGP/EBBR/E15 - Data Incorrect during 6 month checks | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D030113.1 | SGP/EBBR/E7 - Wind speed incorrect | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D030113.10 | SGP/EBBR/E8 - AEM Frozen | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D030113.11 | SGP/EBBR/E8 - Wind Speed Frozen | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D030113.12 | SGP/EBBR/E8 - Water in broken Net Rad Dome | sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D030113.13 | SGP/EBBR/E2 - REPROCESS: Soil Heat Flow #2 Incorrect | sgp30ebbrE2.b1 |
D030113.3 | SGP/EBBR/E7 - CES data incorrect | sgp30ebbrE7.b1 |
D030113.4 | SGP/EBBR/E7 - REPROCESS: Soil Heat Flow #2 Incorrect | sgp30ebbrE7.b1 |
D030113.5 | SGP/EBBR/E7 - Change In Energy Storage Calculations Incorrect | sgp30ebbrE7.b1 |
D030113.6 | SGP/EBBR/E7 - REPROCESS: Soil Heat Flow #5 too low | sgp30ebbrE7.b1 |
D030113.7 | SGP/EBBR/E7 - AEM Problems | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D030113.8 | SGP/EBBR/E7 - T/RH Ventilator Fans Stopped | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D030113.9 | SGP/EBBR/E8 - Battery Voltage Low | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D030114.11 | SGP/EBBR/E12 - Resistance-based Measurements Incorrect | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D030114.12 | SGP/EBBR/E12 - Wind Direction Incorrect | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D030114.13 | SGP/EBBR/E12 - Data zero following Logger Restart | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D030114.14 | SGP/EBBR/E12 - Battery Voltage Low | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D030114.19 | SGP/EBBR/E12 - AEM Stuck in One Position | sgp30ebbrE12.b1 |
D030114.2 | SGP/EBBR/E7 - Data Reported as Zero following Logger Replacement | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D030114.5 | SGP/EBBR/E9 - REPROCESS: Soil Temperature #5 Incorrect | sgp30ebbrE9.b1 |
D030114.6 | SGP/EBBR/E9 - REPROCESS: Soil Heat Flow #4 Incorrect | sgp30ebbrE9.b1 |
D030114.8 | SGP/EBBR/E9 - Net Rad data bad during dome replacement | sgp30ebbrE9.b1, sgp5ebbrE9.b1, sgp15ebbrE9.b1 |
D030114.9 | SGP/EBBR/E9 - AEM Malfunction | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
D030115.10 | SGP/EBBR/E18 - Reprocess: Soil Heat Flow #1 High | sgp30ebbrE18.b1 |
D030115.11 | SGP/EBBR/E18 - Reprocess: Soil Heat Flow #4 High | sgp30ebbrE18.b1 |
D030115.12 | SGP/EBBR/E18 - Data Mostly Zero Following Logger Restart | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
D030115.13 | SGP/EBBR/E18 - Data Incorrect After Logger Restart | sgp30ebbrE18.b1 |
D030115.14 | SGP/EBBR/E18 - Water in Net Radiometer Dome | sgp30ebbrE18.b1 |
D030115.2 | SGP/EBBR/E12 - REPROCESS: Net Rad Calibration Coefficients | sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D030115.3 | SGP/EBBR/E12 - Data Mostly Zero Following Logger Restart | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D030115.4 | SGP/EBBR/E13 - Wind Speed Stalling at Low Speeds | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D030115.5 | SGP/EBBR/E13 - REPROCESS: Soil Moisture #1 Offset1 | sgp30ebbrE13.b1 |
D030115.6 | SGP/EBBR/E13 - AEM Jung between Home Positions | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D030115.7 | SGP/EBBR/E15 - REPROCESS: Soil Moisture #1 Offset | sgp30ebbrE15.b1 |
D030115.8 | SGP/EBBR/E15 - Data Mostly Zero Following Logger Restart | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D030115.9 | SGP/EBBR/E15 - Pressure Low | sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D030116.10 | SGP/EBBR/E19 - REPROCESS: Soil Heat Flow #3 Incorrect Calibration | sgp30ebbrE19.b1 |
D030116.11 | SGP/EBBR/E19 - Water in Net Radiometer Dome | sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
D030116.6 | SGP/EBBR/E19 - Incorrect data after power interruption | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
D030116.7 | SGP/EBBR/E19 - Most Measurements Zero following Restart | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
D030116.8 | SGP/EBBR/E19 - REPROCESS: Soil Temperature #3 Incorrect | sgp30ebbrE19.b1 |
D030116.9 | SGP/EBBR/E19 - REPROCESS: Soil Heat Flow #1 Incorrect | sgp30ebbrE19.b1 |
D030117.1 | SGP/EBBR/E19 - Pressure Low | sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
D030117.10 | SGP/EBBR/E20 - Data Zero following logger restarts | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D030117.11 | SGP/EBBR/E20 - REPROCESS: Soil Temperature #2 Incorrect | sgp30ebbrE20.b1 |
D030117.12 | SGP/EBBR/E20 - Change of Energy Storage Incorrect | sgp30ebbrE20.b1 |
D030117.13 | SGP/EBBR/E20 - Battery voltage low | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D030117.14 | SGP/EBBR/E20 - Wind speed and direction frozen | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE2.b1 |
D030117.16 | SGP/EBBR/E22 - AEM Hung between home positions | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D030117.17 | SGP/EBBR/E22 - Water inside Broken Net Rad Dome | sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D030117.18 | SGP/EBBR/E20 - AEM Hung between Home Positions | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D030117.19 | SGP/EBBR/E20 - REPROCESS: Heat Flow #3 and #5 Incorrect | sgp30ebbrE20.b1 |
D030117.2 | SGP/EBBR/E20 - Data incorrect during instrument maintenance | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D030117.20 | SGP/EBBR/E20 - REPROCESS: Soil Heat Flow #5 Incorrect | sgp30ebbrE20.b1 |
D030117.22 | SGP/EBBR/E20 - REPROCESS: Soil Temperature #1 Incorrect | sgp30ebbrE20.b1 |
D030117.24 | SGP/EBBR/E18 - REPROCESS: Soil Heat Flow #1 and #4 High | sgp30ebbrE18.b1 |
D030117.25 | SGP/EBBR/E20 - REPROCESS: Soil Temperature #3 Incorrect | sgp30ebbrE20.b1 |
D030117.26 | SGP/EBBR/E26 - Measurements zero following logger restart | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D030117.27 | SGP/EBBR/E26 - Multiplexer Problem | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D030117.28 | SGP/EBBR/E26 - Water in broken Net Radiometer Dome | sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D030117.29 | SGP/EBBR/E26 - Battery Voltage Low | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D030117.3 | SGP/EBBR/E20 - Resistance-based Measurements Incorrect | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D030117.4 | SGP/EBBR/E19 - Soil Temperatures #2 and #3 Intermittently Bad | sgp30ebbrE19.b1 |
D030117.6 | SGP/EBBR/E19 - Measurements zero following logger restart | sgp5ebbrE19.b1, sgp15ebbrE19.b1, sgp30ebbrE19.b1 |
D030117.7 | SGP/EBBR/E19 - Change of Energy Storage Incorrect | sgp30ebbrE19.b1 |
D030117.8 | SGP/EBBR/E20 - REPROCESS: Net Rad Coefficients Incorrect | sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D030121.1 | SGP/EBBR/E20 - REPROCESS: Soil Temperature #1 and Soil Heat Flow #5 Incorrect | sgp30ebbrE20.b1 |
D030128.2 | SGP/EBBR/E26 - AEM Hung | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D030128.3 | SGP/EBBR/E26 - Data zeros following logger restart | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D030128.4 | SGP/EBBR/E26 - Left Air Temperature Incorrect | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D030514.8 | SGP/EBBR/E26 - E26 EBBR Soil Heat Flow #1 Incorrect | sgp15ebbrE26.b1, sgp30ebbrE26.b1 |
D030530.3 | SGP/EBBR/E27 - E27 EBBR Incorrect Units for Soil Heat Capacity (cs1, cs2, cs3, cs4, cs5) in 30EBBR | sgp30ebbrE27.b1 |
D030620.1 | SGP/EBBR/E7 - Soil Heat Flow #5 Incorrect | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D030709.1 | SGP/EBBR/E25 - Incorrect Units for Soil Heat Capacity (cs1, cs2, cs3, cs4, cs5) in 30EBBR Header | sgp30ebbrE25.b1 |
D030709.4 | SGP/EBBR/E25 - Incorrect Units for Soil Heat Capacity (cs1, cs2, cs3, cs4, cs5) in 30EBBR Header | sgp30baebbrE25.c1, sgp30ebbrE25.b1 |
D030711.10 | SGP/EBBR/E20 - RHtop Zero | sgp30ebbrE20.b1 |
D030711.11 | SGP/EBBR/E19 - Soil Temperature #2 Incorrect | sgp30ebbrE19.b1, sgp15ebbrE19.b1 |
D030711.12 | SGP/EBBR/E19 - All Data Zeros | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
D030711.13 | SGP/EBBR/E19 - Several Soil Measurements Incorrect | sgp15ebbrE19.b1, sgp30ebbrE19.b1 |
D030711.14 | SGP/EBBR/E19 -Several Soil Measurements Were Incorrect | sgp15ebbrE19.b1, sgp30ebbrE19.b1 |
D030711.16 | SGP/EBBR/E19 - AEM Hung in home30 position | sgp30ebbrE19.b1 |
D030711.17 | SGP/EBBR/E19 - Soil Heat Flow #3 Incorrect | sgp15ebbrE19.b1, sgp30ebbrE19.b1 |
D030711.19 | SGP/EBBR/E19 - Soil Heat Flow#1 Incorrect | sgp15ebbrE19.b1, sgp30ebbrE19.b1 |
D030711.2 | SGP/EBBR/E27 - Soil Heat Flow #2 Incorrect | sgp15ebbrE27.b1, sgp30ebbrE27.b1 |
D030711.20 | SGP/EBBR/E19 - Many Measurements Incorrect | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
D030711.21 | SGP/EBBR/E12 - AEM Hung Inbetween Home Positions | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D030711.22 | SGP/EBBR/E18 - AEM Hung | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
D030711.3 | SGP/EBBR/E27 - Soil Heat Flow #4 Incorrect | sgp15ebbrE27.b1, sgp30ebbrE27.b1 |
D030711.4 | SGP/EBBR/E27 - Reference T Sensor Failure | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
D030711.5 | SGP/EBBR/E27 - Wind Direction Hung | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
D030711.6 | SGP/EBBR/E26 -Right Air Temperature Incorrect | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D030711.7 | SGP/EBBR/E22 - home30 too large | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D030711.8 | SGP/EBBR/E20 - Soil Heat Flow#5 Incorrect | sgp15ebbrE20.b1, sgp30ebbrE20.b1 |
D030711.9 | SGP/EBBR/E20 - Soil Storage Incorrect After Startup | sgp30ebbrE20.b1 |
D030712.1 | SGP/EBBR/E9 - AEM Hung Inbetween Home Positions | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
D030712.11 | SGP/EBBR/E2 - 6 Month Checks | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
D030712.13 | SGP/EBBR/E7 - 6 Month Checks | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D030712.14 | SGP/EBBR/E8 - 6 Month Checks | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D030712.15 | SGP/EBBR/E9 - 6 Month Checks | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
D030712.16 | SGP/EBBR/E12 - 6 Month Checks | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D030712.17 | SGP/EBBR/E13 - 6 Month Checks | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D030712.18 | SGP/EBBR/E15 - 6 Month Checks | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D030712.19 | SGP/EBBR/E18 - 6 Month Checks | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
D030712.2 | SGP/EBBR/E9 - Soil Heat Flow #1 Incorrect | sgp30ebbrE9.b1, sgp15ebbrE9.b1 |
D030712.20 | SGP/EBBR/E19 - 6 Month Checks | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
D030712.21 | SGP/EBBR/E20 - 6 Month Checks | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D030712.22 | SGP/EBBR/E22 - 6 Month Checks | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D030712.23 | SGP/EBBR/E26 - 6 Month Checks | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D030712.24 | SGP/EBBR/E27 - 6 Month Checks | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
D030712.3 | SGP/EBBR/E9 - Battery Voltage Low | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
D030712.4 | SGP/EBBR/E8 - AEM Hung Inbetween Home Positions | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D030712.5 | SGP/EBBR/E8 - Battery Voltage Low | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D030712.6 | SGP/EBBR/E2 - Soil Heat Flow #2 Incorrect | sgp30ebbrE2.b1, sgp15ebbrE2.b1 |
D030712.7 | SGP/EBBR/E2 - Data All Zeros | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
D030712.8 | SGP/EBBR/E2 - Bottom RH Incorrect | sgp5ebbrE2.b1, sgp30ebbrE2.b1 |
D030712.9 | SGP/EBBR/E7 - AEM Hung Inbetween Home Position | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D030724.2 | SGP/EBBR/E15 - Wind Direction 10 Degrees Low | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D030725.1 | SGP/EBBR/E2 - Soil Heat Flow #1 Incorrect | sgp30ebbrE2.b1, sgp15ebbrE2.b1 |
D030728.1 | SGP/EBBR/E2 - Soil Heat Flow #3 Spike | sgp15ebbrE2.b1, sgp30ebbrE2.b1 |
D030728.2 | SGP/EBBR/E15 - Top Humidity reported as zero in 30 minute data. | sgp30ebbrE15.b1 |
D030728.4 | SGP/EBBR/E26 - Soil Heat Flow #5 Large Positive | sgp15ebbrE26.b1, sgp30ebbrE26.b1 |
D030728.5 | SGP/EBBR/E27 - Soil Heat Flow #1 Large | sgp15ebbrE27.b1, sgp30ebbrE27.b1 |
D030728.6 | SGP/EBBR/E27 - Incorrect Data From Logger Restart | sgp30ebbrE27.b1 |
D030729.1 | SGP/EBBR/E8 - Pressure Sensor Measurements Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgpebbrE8.01 |
D030729.2 | SGP/EBBR/E9 - AEM Hung in home 30 position. | sgp30ebbrE9.b1, sgp5ebbrE9.b1, sgp15ebbrE9.b1 |
D030729.3 | SGP/EBBR/E7 - AEM Malfunction, Some Data Incorrect | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D030729.4 | SGP/EBBR/E22 - AEM Hung Inbetween Home Positions | sgp30ebbrE22.b1, sgp5ebbrE22.b1, sgp15ebbrE22.b1 |
D030916.2 | SGP/EBBR/E22 - AEM Home Signals Too High | sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D030917.15 | SGP/EBBR/E2 - Water in Net Radiometer, Surface Fluxes Incorrect | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
D030917.2 | SGP/EBBR/E7 - Reprocess: Soil Heat Flow #5 Offscale, Surface Fluxes Incorrect | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
D030919.1 | SGP/EBBR/E7 - Data all zero | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D030919.10 | SGP/EBBR/E19 - Water in Net Radiometer; Surface Fluxes are Incorrect | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
D030919.11 | SGP/EBBR/E26 - Soil Storage Incorrect; Surface Fluxes Incorrect | sgp30ebbrE26.b1 |
D030919.12 | SGP/EBBR/E27 - Water in Net Radiometer; Surface Fluxes are Incorrect | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
D030919.13 | SGP/EBBR/E27 - 15 and 30 Minute Data All Zeros | sgp15ebbrE27.b1, sgp30ebbrE27.b1 |
D030919.2 | SGP/EBBR/E7 - Energy Storage Incorrect, Fluxes Incorrect | sgp30ebbrE7.b1 |
D030919.3 | SGP/EBBR/E7 - Water in Net Radiometer Domes; Surface Fluxes are Incorrect | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D030919.4 | SGP/EBBR/E2 - 15 and 30 Minute Data are Zeros | sgp15ebbrE2.b1, sgp30ebbrE2.b1 |
D030919.5 | SGP/EBBR/E2 - Soil Heat #1 Offscale | sgp15ebbrE2.b1, sgp30ebbrE2.b1 |
D030919.6 | SGP/EBBR/E8 - Water in Net Radiometer, Surface Fluxes Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D030919.7 | SGP/EBBR/E9 - Water in Net Radiometer; Surface Fluxes are Incorrect | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
D030919.8 | SGP/EBBR/E13 - Water In Net Radiometer; Surface Fluxes are Incorrect | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D030919.9 | SGP/EBBR/E18 - Water in Net Radiometer; Surface Fluxes are Incorrect | sgp30ebbrE18.b1, sgp5ebbrE18.b1, sgp15ebbrE18.b1 |
D030922.6 | SGP/EBBR/E19 - Tair Right Offscale | sgp5ebbrE19.b1, sgp15ebbrE19.b1, sgp30ebbrE19.b1 |
D030922.7 | SGP/EBBR/E27 - Right Air Temperature Spiked Offscale | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
D031021.3 | SGP/EBBR/E27 - Missing data during system update | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
D031029.1 | SGP/EBBR/E20 - Fuse Blown in AEM | sgp30ebbrE20.b1, sgp15ebbrE20.b1, sgp5ebbrE20.b1 |
D031029.2 | SGP/EBBR/E20 - Reprocess: Soil Temperature Probe #1 Measurements Spiked | sgp15ebbrE20.b1, sgp30ebbrE20.b1 |
D031031.1 | SGP/EBBR/E7 - Reference Temperature Failure | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D031031.2 | SGP/EBBR/E15 - Reference Temperature Failed | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D031031.3 | SGP/EBBR/E9 - Measurements All Zero | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
D031031.4 | SGP/EBBR/E12 - Water in Net Radiometer, Surface Fluxes Incorrect | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D031031.5 | SGP/EBBR/E13 - Water in Net Radiometer, Surface Fluxes Incorrect | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D031031.6 | SGP/EBBR/E15 - Water In Net Radiometer, Surface Fluxes Incorrect | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D031031.7 | SGP/EBBR/E20 - Water in Net Radiometer, Surface Fluxes Incorrect | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D031031.8 | SGP/EBBR/E22 - Water in Net Radiometer, Surface Fluxes Incorrect | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D031031.9 | SGP/EBBR/E15 - Soil Temperature #5 Incorrect, Surface Fluxes Incorrect | sgp15ebbrE15.b1, sgp30ebbrE15.b1 |
D031103.1 | SGP/EBBR/E26 - AEM Home Signals Incorrect, Surface Fluxes Incorrect | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D031106.1 | SGP/EBBR/E15 - Soil Temperature #5 Incorrect, Surface Fluxes Incorrect | sgp15ebbrE15.b1, sgp30ebbrE15.b1 |
D031110.7 | SGP/EBBR/E15 - AEM Hung Between Home Positions, Surface Fluxes Incorrect | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D031110.8 | SGP/EBBR/E19 - Battery Voltage Too Low, Most Measurements Incorrect | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
D031117.2 | SGP/EBBR/E19 - AEM Stuck In One Position, Surface Fluxes Incorrect | sgp30ebbrE19.b1, sgp5ebbrE19.b1, sgp15ebbrE19.b1 |
D031117.3 | SGP/EBBR/E27 - Right Air Temperature Spiked Offscale | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
D031124.11 | SGP/EBBR/E19 - Data Missing During Recalibration Changeout | sgp15ebbrE19.b1, sgp30baebbrE19.c1, sgp30ebbrE19.b1, sgp5ebbrE19.b1, sgpebbrE19.00, sgpebbrE19.01 |
D031124.12 | SGP/EBBR/E12 - Pressure Sensor Measurement Scales With Battery Voltage | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D031204.10 | SGP/EBBR/E13 - 6 Month Checks | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D031204.11 | SGP/EBBR/E15 - 6 Month Checks | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D031204.16 | SGP/EBBR/E18 - 6 Month Checks | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
D031204.17 | SGP/EBBR/E19 - 6 Month Checks | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
D031204.18 | SGP/EBBR/E19 - 6 Month Checks | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
D031204.19 | SGP/EBBR/E20 - 6 Month Checks | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D031204.2 | SGP/EBBR/E19 - Data All Zeros | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
D031204.20 | SGP/EBBR/E22 - 6 Month Checks | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D031204.21 | SGP/EBBR/E26 - 6 Month Checks | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D031204.22 | SGP/EBBR/E27 - 6 Month Checks | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
D031204.4 | SGP/EBBR/E2 - 6 Month Checks | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
D031204.6 | SGP/EBBR/E7 - 6 Month Checks | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D031204.7 | SGP/EBBR/E8 - 6 Month Checks | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D031204.8 | SGP/EBBR/E9 - 6 Month Checks | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
D031204.9 | SGP/EBBR/E12 - 6 Month Checks | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D031208.5 | SGP/EBBR/E20 - Wind Direction Incorrect | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D031210.1 | SGP/EBBR/E27 - Reprocess: Soil Heat Flow #5 Offscale, Surface Fluxes Incorrect | sgp15ebbrE27.b1, sgp30ebbrE27.b1 |
D031211.3 | SGP/EBBR/E26 - Spark Gap Bad, Pressure Incorrect | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D031211.5 | SGP/EBBR/E20 - Pressure Sensor Data Incorrect | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D031211.6 | SGP/EBBR/E7 - Pressure Sensor Data Incorrect | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D031217.4 | SGP/EBBR/E12 - Pressure Sensor Measurement Scales With Battery Voltage | sgp30ebbrE12.b1, sgp5ebbrE12.b1, sgp15ebbrE12.b1 |
D031222.10 | SGP/EBBR/E2 - All Data Zero After Logger Restart | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
D031222.11 | SGP/EBBR/E2 - Net Radiometer Iced Over During Storm | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
D031222.12 | SGP/EBBR/E8 - Net Radiometer Iced Over During Storm | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D031222.13 | SGP/EBBR/E9 - Net Radiometer Iced Over During Storm | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
D031222.14 | SGP/EBBR/E9 - Wind Speed Sensor Frozen By Ice Storm | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
D031222.15 | SGP/EBBR/E12 - Wind Speed Sensor Frozen By Ice Storm | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D031222.16 | SGP/EBBR/E13 - Wind Speed Sensor Frozen By Ice Storm | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D031222.17 | SGP/EBBR/E13 - Net Radiometer Iced Over During Storm | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D031222.18 | SGP/EBBR/E15 - Net Radiometer Iced Over During Storm | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D031222.19 | SGP/EBBR/E15 - Wind Speed Sensor Frozen By Ice Storm | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D031222.20 | SGP/EBBR/E20 - Wind Speed Sensor Frozen By Ice Storm | sgp30ebbrE20.b1, sgp5ebbrE20.b1, sgp15ebbrE20.b1 |
D031222.21 | SGP/EBBR/E22 - Wind Speed Sensor Frozen By Ice Storm | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D031222.22 | SGP/EBBR/E22 - Wind Direction Sensor Frozen By Ice Storm | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D031222.24 | SGP/EBBR/E26 - Wind Speed Sensor Frozen By Ice Storm | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D031222.25 | SGP/EBBR/E26 - Wind Direction Sensor Frozen By Ice Storm | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D031222.26 | SGP/EBBR/E27 - Soil Temperature #3 Spike in Data | sgp15ebbrE27.b1, sgp30ebbrE27.b1 |
D031222.9 | SGP/EBBR/E7 - Wind Speed Sensor Frozen By Ice Storm | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D040106.1 | SGP/EBBR/E19 - Wind Speed Sensor Frozen Stopped by Ice Storm | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
D040112.1 | SGP/EBBR/E15 - Wind Direction About 130 Degrees Low | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D040206.1 | SGP/EBBR/E18 - AEM Sitting At Home 15 Position | sgp30baebbrE18.c1, sgp30ebbrE18.b1, sgp5ebbrE18.b1, sgp15ebbrE18.b1 |
D040206.3 | SGP/EBBR/E7 - Pressure Sensor Data Incorret | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D040206.4 | SGP/EBBR/E19 - Low Battery Voltage Caused Incorrect Data | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
D040227.1 | SGP/EBBR/E15 - Net Radiometer not level | sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D040303.2 | SGP/EBBR/E22 - Wind Speed Sensor Bearings Worn | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D040304.10 | SGP/EBBR/E12 - Wind Speed Frozen Stopped By Ice | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D040304.11 | SGP/EBBR/E12 - Wind Direction Frozen Stopped By Ice | sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D040304.12 | SGP/EBBR/E13 - Wind Speed Frozen Stopped By Ice | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D040304.13 | SGP/EBBR/E13 - Wind Direction Frozen Stopped By Ice | sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D040304.14 | SGP/EBBR/E15 - Wind Speed Frozen Stopped By Ice | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D040304.17 | SGP/EBBR/E18 - Wind Speed Frozen Stopped By Ice | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
D040304.18 | SGP/EBBR/E18 - Wind Direction Frozen Stopped By Ice | sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
D040304.19 | SGP/EBBR/E20 - Wind Speed Frozen Stopped By Ice | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D040304.20 | SGP/EBBR/E20 - Wind Direction Frozen Stopped By Ice | sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D040304.21 | SGP/EBBR/E22 - Wind Speed Frozen Stopped By Ice | sgp30ebbrE22.b1, sgp5ebbrE22.b1, sgp15ebbrE22.b1 |
D040304.22 | SGP/EBBR/E22 - Wind Direction Frozen Stopped By Ice | sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D040304.25 | SGP/EBBR/E27 - Wind Speed Frozen Stopped By Ice | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
D040304.26 | SGP/EBBR/E19 - Wind Speed Frozen Stopped By Ice | sgp30ebbrE19.b1, sgp5ebbrE19.b1, sgp15ebbrE19.b1 |
D040304.28 | SGP/EBBR/E26 - Wind Speed Frozen Stopped By Ice | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D040304.29 | SGP/EBBR/E26 - Wind Direction Frozen Stopped By Ice | sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D040304.3 | SGP/EBBR/E7 - Wind Speed Frozen Stopped By Ice | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D040304.4 | SGP/EBBR/E7 - Tref Problem Causes All Temps. to be Incorrect | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D040304.5 | SGP/EBBR/E2 - Wind Speed Frozen Stopped By Ice | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
D040304.6 | SGP/EBBR/E8 - Wind Speed Frozen Stopped By Ice | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D040304.7 | SGP/EBBR/E8 - Wind Direction Frozen Stopped By Ice | sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D040304.8 | SGP/EBBR/E9 - Wind Speed Frozen Stopped By Ice | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
D040304.9 | SGP/EBBR/E9 - Wind Direction Frozen Stopped By Ice | sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
D040318.13 | SGP/EBBR/E27 - Soil Moisture #5 Insensitive | sgp30ebbrE27.b1 |
D040318.17 | SGP/EBBR/E8 - Wind Speed Sensor Frozen Stopped By Icing | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D040331.2 | SGP/EBBR/E8 - AEM Replacement | sgp15ebbrE8.b1, sgp30baebbrE8.c1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D040407.2 | SGP/EBBR/E19 - Reprocess: Incorrect Net Rad Calibration Coefficient | sgp30baebbrE19.c1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
D040505.1 | SGP/EBBR/E2 - Soil Heat Flow #1 Spikes Down | sgp30ebbrE2.b1 |
D040505.3 | SGP/EBBR/E18 - Pressure Low | sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
D040507.1 | SGP/EBBR/E15 - Wind Speed Incorrect | sgp15ebbrE15.b1, sgp30baebbrE15.c1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D040507.2 | SGP/EBBR/E20 - Left Air Temperature Incorrect | sgp15ebbrE20.b1, sgp30baebbrE20.c1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D040507.3 | SGP/EBBR/E27 - Right Air Temperature Incorrect | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
D040525.3 | SGP/EBBR/E19 - Metadata errors | sgp15ebbrE19.b1, sgp30baebbrE19.c1, sgp30ebbrE19.b1 |
D040525.4 | SGP/EBBR/E19 - Metadata errors | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
D040608.6 | SGP/EBBR/E13 - Automatic Exchange Mechanism Failure | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D040608.7 | SGP/EBBR/E19 - AEM Hung Inbetween Home Positions | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
D040615.3 | SGP/EBBR/E12 - Incorrect Times | sgp15ebbrE12.b1, sgp30baebbrE12.c1, sgp30ebbrE12.b1, sgp5ebbrE12.b1, sgpebbrE12.00 |
D040706.2 | SGP/EBBR/E22 - Net Radiation Incorrect | sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D040706.3 | SGP/EBBR/E27 - Loose Tair Left Connections | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
D040722.1 | SGP/EBBR/E20 - Wind Direction Stuck at 100 Degrees | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D040723.2 | SGP/EBBR/E9 - net radiometer top dome broken, water in bottom dome | sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
D040723.3 | SGP/EBBR/E22 - Water In Net Radiometer | sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D040723.4 | SGP/EBBR/E18 - Soil Temperature and Moisture Probes in Set #1 Spiked Outputs | sgp30ebbrE18.b1 |
D040813.2 | SGP/EBBR/E20 - AEM Failure | sgp5ebbrE20.b1, sgp15ebbrE20.b1, sgp30ebbrE20.b1 |
D040813.5 | SGP/EBBR/E26 - Soil Temperature #4 and Soil Heat Flow #4 Incorrect | sgp30ebbrE26.b1 |
D040813.6 | SGP/EBBR/E18 - Soil Temperature #2 Incorrect | sgp30ebbrE18.b1 |
D040813.8 | SGP/EBBR/E7 - Pressure Too High | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D040915.2 | SGP/EBBR/E13 - Net Radiation Incorrect, Fluxes Incorrect | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D040915.3 | SGP/EBBR/E2 - Net Radiometer Dome Broken, Reducing Net Radiation Measurement, Fluxes Incorrect | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
D040915.5 | SGP/EBBR/E8 - Net Radiation Incorrect, Fluxes Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D040915.6 | SGP/EBBR/E13 - AEM Failure, Fluxes Incorrect | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D040915.7 | SGP/EBBR/E20 - Soil Heat Flow #3 and Soil Temperature #3 Incorrect | sgp15ebbrE20.b1, sgp30ebbrE20.b1 |
D041004.4 | SGP/EBBR/E15 - Soil Moisture Coefficients Incorrect | sgp30ebbrE15.b1 |
D041005.3 | SGP/EBBR/E7 - Data Missing | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D041006.3 | SGP/EBBR/E20 - All Soil Measurements Incorrect | sgp15ebbrE20.b1, sgp30ebbrE20.b1 |
D041014.2 | SGP/EBBR/E2 - Soil Heat Flow Sensor #2 Bad | sgp15ebbrE2.b1, sgp30ebbrE2.b1 |
D041222.1 | SGP/EBBR/E26 - Spark Gap Bad, Pressure Incorrect | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D041222.10 | SGP/EBBR/E9 - 6 Month Checks | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
D041222.11 | SGP/EBBR/E12 - 6 Month Checks | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D041222.12 | SGP/EBBR/E13 - 6 Month Checks | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D041222.13 | SGP/EBBR/E15 - 6 Month Checks | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D041222.14 | SGP/EBBR/E18 - 6 Month Checks | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
D041222.15 | SGP/EBBR/E19 - 6 Month Checks | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
D041222.16 | SGP/EBBR/E20 - 6 Month Checks | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D041222.17 | SGP/EBBR/E22 - 6 Month Checks | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D041222.18 | SGP/EBBR/E26 - 6 Month Checks | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D041222.19 | SGP/EBBR/E27 - 6 Month Checks | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
D041222.2 | SGP/EBBR/E12 - AEM home_15 Incorrect, Flux Measurements not affected | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D041222.3 | SGP/EBBR/E20 - Soil Moisture #4 and Soil Temperature #4 Incorrect | sgp15ebbrE20.b1, sgp30ebbrE20.b1 |
D041222.4 | SGP/EBBR/E26 - Low Battery Voltage - Many Measurements Incorrect | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D041222.5 | SGP/EBBR/E26 - Most Measurements Incorrect | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D041222.6 | SGP/EBBR/E2 - 6 Month Checks | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
D041222.8 | SGP/EBBR/E7 - 6 Month Checks | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D041222.9 | SGP/EBBR/E8 - 6 Month Checks | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D041223.1 | SGP/EBBR/E20 - Soil Heat Flow #4 Incorrect | sgp15ebbrE20.b1, sgp30ebbrE20.b1 |
D050113.2 | SGP/EBBR/E9 - Wind Speed Threshhold Low | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
D050206.10 | SGP/EBBR/E13 - Wind Speed Frozen Stopped By Ice Storm | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D050206.11 | SGP/EBBR/E15 - Wind Speed Frozen Stopped By Ice Storm | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D050206.12 | SGP/EBBR/E18 - Wind Speed Frozen Stopped By Ice Storm | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
D050206.13 | SGP/EBBR/E19 - Wind Speed Frozen Stopped By Ice Storm | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
D050206.14 | SGP/EBBR/E20 - Wind Speed Frozen Stopped By Ice Storm | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D050206.15 | SGP/EBBR/E22 - Wind Speed Frozen Stopped By Ice Storm | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D050206.16 | SGP/EBBR/E26 - Wind Speed Frozen Stopped By Ice Storm | sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp15ebbrE26.b1 |
D050206.17 | SGP/EBBR/E27 - Wind Speed Frozen Stopped By Ice Storm | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
D050206.4 | SGP/EBBR/E2 - Wind Speed Frozen Stopped By Ice Storm | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
D050206.6 | SGP/EBBR/E7 - Wind Speed Frozen Stopped By Ice Storm | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D050206.7 | SGP/EBBR/E8 - Wind Speed Frozen Stopped By Ice Storm | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D050206.8 | SGP/EBBR/E9 - Wind Speed Frozen Stopped By Ice Storm | sgp30ebbrE9.b1, sgp5ebbrE9.b1, sgp15ebbrE9.b1 |
D050206.9 | SGP/EBBR/E12 - Wind Speed Frozen Stopped By Ice Storm | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D050207.1 | SGP/EBBR/E2 - Wind Direction Frozen Stopped By Ice Storm | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
D050207.10 | SGP/EBBR/E20 - Wind Direction Frozen Stopped By Ice Storm | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D050207.11 | SGP/EBBR/E22 - Wind Direction Frozen Stopped By Ice Storm | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D050207.12 | SGP/EBBR/E26 - Wind Direction Frozen Stopped By Ice Storm | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D050207.13 | SGP/EBBR/E27 - Wind Direction Frozen Stopped By Ice Storm | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
D050207.3 | SGP/EBBR/E8 - Wind Direction Frozen Stopped By Ice Storm | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D050207.4 | SGP/EBBR/E9 - Wind Direction Frozen Stopped By Ice Storm | sgp30ebbrE9.b1, sgp5ebbrE9.b1, sgp15ebbrE9.b1 |
D050207.5 | SGP/EBBR/E12 - Wind Direction Frozen Stopped By Ice Storm | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D050207.6 | SGP/EBBR/E13 - Wind Direction Frozen Stopped By Ice Storm | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D050207.7 | SGP/EBBR/E15 - Wind Direction Frozen Stopped By Ice Storm | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D050207.8 | SGP/EBBR/E18 - Wind Direction Frozen Stopped By Ice Storm | sgp30ebbrE18.b1, sgp5ebbrE18.b1, sgp15ebbrE18.b1 |
D050207.9 | SGP/EBBR/E19 - Wind Direction Frozen Stopped By Ice Storm | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
D050211.1 | SGP/EBBR/E26 - Aging Pressure Sensor Output Changes With Supply Voltage | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D050211.2 | SGP/EBBR/E27 - Soil Moisture #5 Incorrect | sgp30ebbrE27.b1 |
D050215.11 | SGP/EBBR/E7 - Wind Speed Frozen Stopped By Ice | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D050215.12 | SGP/EBBR/E8 - Wind Speed Frozen Stopped By Ice Storm | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D050215.13 | SGP/EBBR/E8 - Wind Direction Frozen Stopped By Ice | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D050215.14 | SGP/EBBR/E9 - Wind Direction Frozen Stopped By Ice | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
D050215.15 | SGP/EBBR/E9 - Wind Speed Frozen Stopped By Ice | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
D050215.16 | SGP/EBBR/E12 - Wind Speed Frozen Stopped By Ice | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D050215.17 | SGP/EBBR/E12 - Wind Direction Frozen Stopped By Ice | sgp30ebbrE12.b1, sgp5ebbrE12.b1, sgp15ebbrE12.b1 |
D050215.18 | SGP/EBBR/E15 - Wind Speed Frozen Stopped By Ice | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D050215.19 | SGP/EBBR/E15 - Wind Direction Frozen Stopped By Ice | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D050215.20 | SGP/EBBR/E19 - Wind Speed Constant | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
D050215.21 | SGP/EBBR/E22 - Wind Speed Frozen Stopped By Ice | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D050215.22 | SGP/EBBR/E22 - Wind Direction Frozen Stopped By Ice | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D050215.7 | SGP/EBBR/E2 - Wind Speed Frozen Stopped By Ice | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
D050215.8 | SGP/EBBR/E2 - Wind Direction Frozen Stopped By Ice | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
D050304.1 | SGP/EBBR/E12 - Wind Speed and Direction Frozen by Ice Storm | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D050323.2 | SGP/EBBR/E22 - Net Radiation Too Low - Top Dome Broken | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D050324.1 | SGP/EBBR/E15 - Soil Heat Flows Undermeasured, and Latent and Sensible Heat Fluxes Overstimated | sgp30ebbrE15.b1 |
D050324.2 | SGP/EBBR/E19 - Reprocess: Negative Net Rad calibrations | sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
D050324.3 | SGP/EBBR/E26 - Incorrect soil type coefficients used | sgp30ebbrE26.b1 |
D050330.2 | SGP/EBBR/E7 - Wrong Datalogger Program | sgp30ebbrE7.b1 |
D050404.1 | SGP/EBBR/E20 - Wrong Datalogger Program Installed | sgp30ebbrE20.b1 |
D050404.4 | SGP/EBBR/E20 - Temp. of thum_left Incorrect At Times | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D050413.3 | SGP/EBBR/E8 - Battery Voltage Problem | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D050413.6 | SGP/EBBR/E27 - SM2 Coefficients Incorect | sgp15ebbrE27.b1, sgp30ebbrE27.b1 |
D050510.3 | SGP/EBBR/E13 - Net Radiometer Tilted Half Degree West | sgp30ebbrE13.b1, sgp5ebbrE13.b1, sgp15ebbrE13.b1 |
D050601.2 | SGP/EBBR/E9 - Soil moisture too high | sgp15ebbrE9.b1, sgp30ebbrE9.b1 |
D050622.1 | SGP/EBBR/E13 - Sensible and Latent Heat Fluxes Incorrect (AEM failure) | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D050624.2 | SGP/EBBR/E15 - Net Radiation Too Large - Water in Bottom | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D050624.3 | SGP/EBBR/E19 - Net Radiation Incorrect - Cable Chewed by Rodents | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
D050624.7 | SGP/EBBR/E13 - Improved EBBR CR10 Program | sgp30ebbrE13.b1 |
D050719.10 | SGP/EBBR/E15 - Improved EBBR CR10 Program | sgp30ebbrE15.b1 |
D050719.11 | SGP/EBBR/E18 - Improved EBBR CR10 Program | sgp30ebbrE18.b1 |
D050719.12 | SGP/EBBR/E19 - Improved EBBR CR10 Program | sgp30ebbrE19.b1 |
D050719.13 | SGP/EBBR/E20 - Improved EBBR CR10 Program | sgp30ebbrE20.b1 |
D050719.14 | SGP/EBBR/E22 - Improved EBBR CR10 Program | sgp30ebbrE22.b1 |
D050719.15 | SGP/EBBR/E26 - Improved EBBR CR10 Program | sgp30ebbrE26.b1 |
D050719.16 | SGP/EBBR/E27 - Improved EBBR CR10 Program | sgp30ebbrE27.b1 |
D050719.4 | SGP/EBBR/E2 - Improved EBBR CR10 Program | sgp30ebbrE2.b1 |
D050719.6 | SGP/EBBR/E7 - Improved EBBR CR10 Program | sgp30ebbrE7.b1 |
D050719.7 | SGP/EBBR/E8 - Improved EBBR CR10 Program | sgp30ebbrE8.b1 |
D050719.8 | SGP/EBBR/E9 - Improved EBBR CR10 Program | sgp30ebbrE9.b1 |
D050719.9 | SGP/EBBR/E12 - Improved EBBR CR10 Program | sgp30ebbrE12.b1 |
D050922.2 | SGP/EBBR/E20 - Soil Temperature and Soil Moisture #4 Measurements Incorrect | sgp15ebbrE20.b1, sgp30ebbrE20.b1 |
D051024.1 | SGP/EBBR/E7 - Data Missing | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D051104.1 | SGP/EBBR/E9 - Net Radiation Incorrect: Sensible and Latent Heat Fluxes Incorrect | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
D051104.2 | SGP/EBBR/E13 - Fluxes Incorrect | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D051104.3 | SGP/EBBR/E13 - Loose connections | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D051106.1 | SGP/EBBR/E2 - Net Radiation Incorrect; Latent and Sensible Heat Fluxes Incorrect | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
D051106.10 | SGP/EBBR/E18 - Net Radiation Incorrect; Sensible and Latent Heat Fluxes Incorrect | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
D051106.13 | SGP/EBBR/E22 - Net Radiation Incorrect; Sensible and Latent Heat Fluxes Incorrect | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D051106.15 | SGP/EBBR/E27 - Net Radiation Incorrect; Sensible and Latent Heat Fluxes Incorrect | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
D051106.16 | SGP/EBBR/E27 - Net Radiation Incorrect; Sensible and Latent Heat Fluxes Incorredt | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
D051106.18 | SGP/EBBR/E2 - Average Soil Heat Flux Incorrect | sgp30ebbrE2.b1 |
D051106.20 | SGP/EBBR/E7 - Average Soil Heat Flux Incorrect | sgp30ebbrE7.b1 |
D051106.21 | SGP/EBBR/E8 - Average Soil Heat Flux Incorrect | sgp30ebbrE8.b1 |
D051106.22 | SGP/EBBR/E9 - Average Soil Heat Flux Incorrect | sgp30ebbrE9.b1 |
D051106.23 | SGP/EBBR/E12 - Average Soil Heat Flux Incorrect | sgp30ebbrE12.b1 |
D051106.24 | SGP/EBBR/E13 - Average Soil Heat Flux Incorrect | sgp30ebbrE13.b1 |
D051106.25 | SGP/EBBR/E15 - Average Soil Heat Flux Incorrect | sgp30ebbrE15.b1 |
D051106.26 | SGP/EBBR/E18 - Average Soil Heat Flux Incorrect | sgp30ebbrE18.b1 |
D051106.27 | SGP/EBBR/E19 - Average Soil Heat Flux Incorrect | sgp30ebbrE19.b1 |
D051106.28 | SGP/EBBR/E20 - Average Soil Heat Flux Incorrect | sgp30ebbrE20.b1 |
D051106.29 | SGP/EBBR/E22 - Average Soil Heat Flux Incorrect | sgp30ebbrE22.b1 |
D051106.3 | SGP/EBBR/E7 - Net Radiation Incorrect; Sensible and Latent Heat Fluxes Incorrect | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D051106.30 | SGP/EBBR/E26 - Average Soil Heat Flux Incorrect | sgp30ebbrE26.b1 |
D051106.31 | SGP/EBBR/E27 - Average Soil Heat Flux Incorrect | sgp30ebbrE27.b1 |
D051106.4 | SGP/EBBR/E8 - Soil Heat Flow #2 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D051106.5 | SGP/EBBR/E8 - Net Radiation Incorrect; Sensible and Latent Heat Fluxes Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D051106.9 | SGP/EBBR/E15 - Soil Set #5 Temperature, Moisture, and Heat Flow Incorrect | sgp15ebbrE15.b1, sgp30ebbrE15.b1 |
D051111.10 | SGP/EBBR/E12 - metadata corrections | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D051111.11 | SGP/EBBR/E13 - metadata corrections | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D051111.12 | SGP/EBBR/E15 - metadata corrections | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D051111.13 | SGP/EBBR/E18 - metadata corrections | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
D051112.1 | SGP/EBBR/E19 - metadata corrections | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
D051112.10 | SGP/EBBR/E27 - metadata corrections | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
D051112.2 | SGP/EBBR/E2 - metadata corrections | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
D051112.4 | SGP/EBBR/E7 - metadata corrections | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D051112.5 | SGP/EBBR/E8 - metadata corrections | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D051112.6 | SGP/EBBR/E9 - metadata corrections | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
D051112.7 | SGP/EBBR/E20 - metadata corrections | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D051112.8 | SGP/EBBR/E22 - metadata corrections | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D051112.9 | SGP/EBBR/E26 - metadata corrections | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D051115.10 | SGP/EBBR/E25 - metadata errors | sgp15ebbrE25.b1, sgp30ebbrE25.b1 |
D051214.2 | SGP/EBBR/E19 - Net Radiation, Sensible and Latent Heat Fluxes Incorrect | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
D051214.3 | SGP/EBBR/E19 - Incorrect Sensible and Latent Heat Fluxes | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
D051215.2 | SGP/EBBR/E9 - Net Radiation, Sensible and Latent Heat Fluxes Incorrect | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
D051215.3 | SGP/EBBR/E15 - Net Radiation, Sensible and Latent Heat Fluxes Incorrect. | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D051222.1 | SGP/EBBR/E26 - Data Suspect During Low Supply Voltage | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D060110.2 | SGP/EBBR/E2 - Sensible and Latent Heat Fluxes Incorrect | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
D060111.10 | SGP/EBBR/E15 - Incorrect Sensible and Latent Heat Fluxes | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D060111.11 | SGP/EBBR/E15 - Reprocess: Sensible and Latent Heat Fluxes, Soil Set #2 Sometimes Incorrect | sgp30ebbrE15.b1 |
D060111.13 | SGP/EBBR/E20 - Data Incorrect During J-Panel Replacement | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D060111.3 | SGP/EBBR/E7 - Barometric Pressure Incorrect | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D060111.4 | SGP/EBBR/E8 - AEM Not Exchanging | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D060111.5 | SGP/EBBR/E12 - Sensible and Latent Heat Fluxes Incorrect at Times | sgp15ebbrE12.b1, sgp30ebbrE12.b1 |
D060111.8 | SGP/EBBR/E12 - Multiplexer failure | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D060111.9 | SGP/EBBR/E13 - Wind Speed Sensor Frozen | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D060112.1 | SGP/EBBR/E22 - Sensible and Latent Heat Fluxes Incorrect | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D060425.1 | SGP/EBBR/E27 - Sensible and Latent Heat Fluxes Incorrect | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
D060428.4 | SGP/EBBR/E22 - Fluxes Incorrect When Soil Heat Flow #3 Out of Range | sgp30ebbrE22.b1 |
D060501.11 | SGP/EBBR/E12 - Fluxes and Gradient Measurements Incorrect | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D060501.12 | SGP/EBBR/E13 - Missing Data | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D060501.13 | SGP/EBBR/E13 - Flux and Gradient Measurements Incorrect | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D060501.14 | SGP/EBBR/E13 - Net Radiation and Fluxes Suspect | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D060501.15 | SGP/EBBR/E15 - Wind Speed Sensor Frozen Stopped | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D060501.17 | SGP/EBBR/E18 - Wind Speed Zero | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
D060501.2 | SGP/EBBR/E2 - Wind Speed Sensor Frozen Stopped | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
D060501.22 | SGP/EBBR/E20 - Suspect Measurements | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D060501.25 | SGP/EBBR/E22 - Wind Speed Sensor Frozen Stopped | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D060501.26 | SGP/EBBR/E26 - Measurements Incorrect | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D060501.27 | SGP/EBBR/E27 - Flux and Gradient Measurements Incorrect | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
D060501.4 | SGP/EBBR/E7 - Fluxes and Gradient Measurements Incorrect | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D060501.6 | SGP/EBBR/E8 - Flux and Gradient Measurements Suspect | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D060501.8 | SGP/EBBR/E8 - Wind Speed Sensor Frozen Stopped | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D060501.9 | SGP/EBBR/E8 - Fluxes Incorrect; Left Humidity Near Zero | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D060508.2 | SGP/EBBR/E8 - Sensible and Latent Heat Flux Suspect | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D060508.3 | SGP/EBBR/E27 - Instrument problem, Loose connections | sgp15ebbrE27.b1, sgp30ebbrE27.b1 |
D060621.7 | SGP/EBBR/E9 - Wind Speed Incorrect | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
D060622.1 | SGP/EBBR/E13 - Left Air Temperature Incorrect - Fluxes Affected | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D060622.2 | SGP/EBBR/E13 - Fluxes Incorrect, AEM Problems | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D060622.6 | SGP/EBBR/E20 - Net Radiation, Fluxes Too Large | sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D060712.1 | SGP/EBBR/E27 - Field fire burned AEM | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
D060809.1 | SGP/EBBR/E27 - Sensible, Latent, and Average Soil Heat Fluxes Sometimes Incorrect | sgp15ebbrE27.b1, sgp30ebbrE27.b1 |
D060809.2 | SGP/EBBR/E27 - Low Battery Condition | sgp30ebbrE27.b1, sgp5ebbrE27.b1, sgp15ebbrE27.b1 |
D060818.10 | SGP/EBBR/E13 - Sensible and Latent Heat Fluxes Incorrect When Pressure Offscale | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D060818.11 | SGP/EBBR/E27 - Most Measurements Suspect | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
D060818.13 | SGP/EBBR/E27 - Battery Voltage Problem - Measurements Incorrect | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
D060818.6 | SGP/EBBR/E18 - Sensible and Latent Heat Fluxes and Net Radiation Incorrect | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
D060818.8 | SGP/EBBR/E8 - Sensible and Latent Heat Fluxes and Net Radiation Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D060921.3 | SGP/EBBR/E12 - Net Radiation, and Sensible and Latent Heat Fluxes Incorrect | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D060921.4 | SGP/EBBR/E12 - Soil Temperature #1 Incorrect | sgp15ebbrE12.b1, sgp30ebbrE12.b1 |
D060921.5 | SGP/EBBR/E13 - Soil Heat Flow #3 Large Negative | sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
D060921.6 | SGP/EBBR/E13 - Pressure and Sensible and Latent Heat Fluxes Incorrect | sgp30ebbrE13.b1, sgp5ebbrE13.b1, sgp15ebbrE13.b1 |
D060921.7 | SGP/EBBR/E13 - Most Measurements Incorrect When Battery Voltage Too low | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D060921.8 | SGP/EBBR/E22 - Net Radiation and Sensible and Latent Heat Fluxes Incorrect | sgp30ebbrE22.b1, sgp5ebbrE22.b1, sgp15ebbrE22.b1 |
D061103.3 | SGP/EBBR/E18 - Sensible and Latent Heat Fluxes Incorrect | sgp30ebbrE18.b1 |
D061106.5 | SGP/EBBR/E22 - Soil Heat Flow #4 Incorrect - Sensible and Latent Heat Fluxes Sometimes Affected | sgp15ebbrE22.b1, sgp30ebbrE22.b1 |
D061114.6 | SGP/EBBR/E13 - Battery failure | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D061114.9 | SGP/EBBR/E19 - Sensible and Latent Heat Fluxes Incorrect | sgp30ebbrE19.b1 |
D061220.1 | SGP/EBBR/E2 - Wind Direction Dependency of Surface Fluxes and General Data Quality Information | sgp30ebbrE2.b1 |
D061220.10 | SGP/EBBR/E18 - Wind Direction Dependency of Surface Fluxes and General Data Quality Information | sgp30ebbrE18.b1 |
D061220.11 | SGP/EBBR/E19 - Wind Direction Dependency of Surface Fluxes and General Data Quality Information | sgp30ebbrE19.b1 |
D061220.12 | SGP/EBBR/E20 - Wind Direction Dependency of Surface Fluxes and General Data Quality Information | sgp30ebbrE20.b1 |
D061220.13 | SGP/EBBR/E22 - Wind Direction Dependency of Surface Fluxes and General Data Quality Information | sgp30ebbrE22.b1 |
D061220.14 | SGP/EBBR/E25 - Wind Direction Dependency of Surface Fluxes and General Data Quality Information | sgp30ebbrE25.b1 |
D061220.15 | SGP/EBBR/E26 - Wind Direction Dependency of Surface Fluxes and General Data Quality Information | sgp30ebbrE26.b1 |
D061220.16 | SGP/EBBR/E27 - Wind Direction Dependency of Surface Fluxes and General Data Quality Information | sgp30ebbrE27.b1 |
D061220.29 | SGP/EBBR/E25 - Wind Direction Dependency of Surface Fluxes and General Data Quality Information | sgp30ebbrE25.b1 |
D061220.3 | SGP/EBBR/E7 - Wind Direction Dependency of Surface Fluxes and General Data Quality Information | sgp30ebbrE7.b1 |
D061220.4 | SGP/EBBR/E8 - Wind Direction Dependency of Surface Fluxes and General Data Quality Information | sgp30ebbrE8.b1 |
D061220.5 | SGP/EBBR/E9 - Wind Direction Dependency of Surface Fluxes and General Data Quality Information | sgp30ebbrE9.b1 |
D061220.6 | SGP/EBBR/E12 - Wind Direction Dependency of Surface Fluxes and General Data Quality Information | sgp30ebbrE12.b1 |
D061220.8 | SGP/EBBR/E13 - Wind Direction Dependency of Surface Fluxes and General Data Quality Information | sgp30ebbrE13.b1 |
D061220.9 | SGP/EBBR/E15 - Wind Direction Dependency of Surface Fluxes and General Data Quality Information | sgp30ebbrE15.b1 |
D070208.2 | SGP/EBBR/E2 - Missing data | sgp15ebbrE2.b1, sgp30baebbrE2.c1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
D070208.3 | SGP/EBBR/E2 - Missing data | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
D070215.2 | SGP/EBBR/E2 - Windspeed sensor failure | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
D070215.3 | SGP/EBBR/E7 - Windspeed threshhold too high | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D070223.2 | SGP/EBBR/E8 - Failed Temp/RH sensor | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D070426.1 | SGP/EBBR/E22 - Wind Direction Incorrect | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D070716.2 | SGP/EBBR/E20 - Reprocess: Wind Direction 20 Degrees Low | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D070809.1 | SGP/EBBR/E27 - Multiplexer #2 failure | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
D070815.6 | SGP/EBBR/E22 - H and LE Incorrect | sgp15ebbrE22.b1, sgp30ebbrE22.b1 |
D070917.1 | SGP/EBBR/E13 - Radiometer dome broken | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D071001.1 | SGP/EBBR/E26 - Sensible and Latent Heat Fluxes Affected by bad temp sensor | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D071115.1 | SGP/EBBR/E20 - Reprocess: Wind Direction 65 Degrees Low | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D080102.2 | SGP/EBBR/E9 - Net Radiation Incorrect | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
D080114.3 | SGP/EBBR/E27 - Measurements offscale | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
D080212.1 | SGP/EBBR/E7 - Wind Speed Frozen Stopped | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D080318.1 | SGP/EBBR/E13 - Wind direction failure | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D080410.2 | SGP/EBBR/E20 - Wind Direction 50 Degrees High | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D080424.1 | SGP/EBBR/E18 - Soil Heat Flow #4 Offscale Low; Sensible and Latent Heat Flux Sometimes Incorrect | sgp15ebbrE18.b1, sgp30ebbrE18.b1 |
D080520.1 | SGP/EBBR/E8 - Wind Direction Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D080523.2 | SGP/EBBR/E2 - Net Radiation and Fluxes Incorrect | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
D080527.2 | SGP/EBBR/E19 - Sensible and latent heat flux and net radiation incorrect | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
D080616.3 | SGP/EBBR/E2 - AEM Ventilator Dangling - Sensible and Latent Heat Fluxes Incorrect | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
D080616.4 | SGP/EBBR/E15 - Water in dome | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D080616.5 | SGP/EBBR/E15 - Soil Heat Flow #4 Incorrect | sgp15ebbrE15.b1, sgp30ebbrE15.b1 |
D080702.2 | SGP/EBBR/E18 - SHF #4 Measurements Too Low; Sensible and Latent Heat Fluxes Suspect | sgp15ebbrE18.b1, sgp30ebbrE18.b1 |
D080902.1 | SGP/EBBR/E20 - Wind Direction Incorrect | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D090312.2 | SGP/EBBR/E7 - AEM voltage too high | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D090313.11 | SGP/EBBR/E9 - Wind Speed Incorrect | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
D090313.12 | SGP/EBBR/E12 - Wind Speed Incorrect | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D090313.13 | SGP/EBBR/E13 - Wind Speed Incorrect | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D090313.14 | SGP/EBBR/E15 - Wind Speed Incorrect | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D090313.15 | SGP/EBBR/E8 - Wind Speed Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D090313.16 | SGP/EBBR/E20 - Wind Speed Incorrect | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D090313.2 | SGP/EBBR/E15 - Soil Moisture and Soil Temp #1 Incorrect | sgp15ebbrE15.b1, sgp30ebbrE15.b1 |
D090313.3 | SGP/EBBR/E18 - Wind Speed Incorrect | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
D090313.4 | SGP/EBBR/E19 - Wind Speed Incorrect | sgp30ebbrE19.b1, sgp5ebbrE19.b1, sgp15ebbrE19.b1 |
D090313.6 | SGP/EBBR/E22 - Wind Speed Incorrect | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D090313.8 | SGP/EBBR/E26 - Wind Speed Incorrect | sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp15ebbrE26.b1 |
D090313.9 | SGP/EBBR/E27 - Wind Speed Incorrect | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
D090414.1 | SGP/EBBR/E20 - Wind Direction Incorrect | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D090428.1 | SGP/EBBR/E7 - Fluxes intermittently incorrect | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D090617.3 | SGP/EBBR/E7 - Wind direction incorrect | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D090817.1 | SGP/EBBR/E8 - Insufficient voltage | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D091001.2 | SGP/EBBR/E2 - Low Supply Voltage | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
D091015.2 | SGP/EBBR/E26 - Low supply voltage | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D100202.2 | SGP/EBBR/E7 - Wind speed frozen stopped | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D100202.3 | SGP/EBBR/E19 - Wind direction sensor frozen near north | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
D100202.4 | SGP/EBBR/E20 - Wind speed sensor frozen stopped | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D100203.1 | SGP/EBBR/E19 - Wind speed sensor frozen stopped | sgp5ebbrE19.b1, sgp15ebbrE19.b1, sgp30ebbrE19.b1 |
D100223.1 | SGP/EBBR/E12 - Wind Speed Sensor Frozen Stopped by Icing | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D100510.1 | SGP/EBBR/E15 - Sensible and Latent Heat Fluxes Low (Net Radiation too low) | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D100521.2 | SGP/EBBR/E9 - Sensible and latent heat flux incorrect | sgp15ebbrE9.b1, sgp30ebbrE9.b1 |
D100720.2 | SGP/EBBR/E12 - Intermittent data loss | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D100720.3 | SGP/EBBR/E12 - Sensible and Latent Heat Fluxes and Other Measurements Incorrect | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D100727.4 | SGP/EBBR/E7 - Incorrect Sensible and Latent Heat Fluxes | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D100727.5 | SGP/EBBR/E13 - Incorrect Sensible and Latent Heat Fluxes | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D100803.1 | SGP/EBBR/E20 - Incorrect Sensible and Latent Heat Fluxes | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D100816.1 | SGP/EBBR/E20 - Incorrect Sensible and Latent Heat Fluxes - a result of low battery voltage. | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D101004.1 | SGP/EBBR/E12 - Net radiation too high | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D101004.2 | SGP/EBBR/E13 - Net Radiation Too Large | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D101007.1 | SGP/EBBR/E15 - Sensible and Latent Heat Flux Suspect - Soil Heat Flow 5 Problem | sgp15ebbrE15.b1, sgp30ebbrE15.b1 |
D101007.3 | SGP/EBBR/E7 - Sensible and latent heat fluxes incorrect - soil heat flow 1 problem | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D101127.1 | SGP/EBBR/E9 - T/RH Probe Temperature Problem - Sensible and Latent Heat Fluxes Affected | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
D110105.3 | SGP/EBBR/E8 - RH Maximum too high | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D110105.4 | SGP/EBBR/E9 - Sensible and Latent Heat Fluxes Incorrect | sgp5ebbrE9.b1, sgp15ebbrE9.b1, sgp30ebbrE9.b1 |
D110721.1 | SGP/EBBR/E13 - Soil temperature #3 offscale | sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
D110830.1 | SGP/EBBR/E12 - Loose connection (SHF #4) | sgp15ebbrE12.b1, sgp30ebbrE12.b1 |
D110830.2 | SGP/EBBR/E9 - Sensible and Latent Heat Fluxes Incorrect | sgp30ebbrE9.b1 |
D110830.3 | SGP/EBBR/E7 - Loose connection (ST #3) | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D111004.2 | SGP/EBBR/E19 - Nighttime data missing | sgp30ebbrE19.b1, sgp5ebbrE19.b1, sgp15ebbrE19.b1 |
D111011.2 | SGP/EBBR/E7 - Sensible and Latent Heat Flux Incorrect | sgp30ebbrE7.b1 |
D111026.2 | SGP/EBBR/E20 - Sensible and latent heat fluxes overestimated | sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D120128.2 | SGP/EBBR/E15 - Soil heat flow #1 failure | sgp15ebbrE15.b1, sgp30ebbrE15.b1 |
D120305.1 | SGP/EBBR/E15 - Soil heat flow #5 failure | sgp15ebbrE15.b1, sgp30ebbrE15.b1 |
D120307.1 | SGP/EBBR/E12 - Soil heat flux #4 intermittently incorrect | sgp30ebbrE12.b1 |
D930503.1 | EBBR - soil heat flux plate giving opposite sign | sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D931108.3 | SGP/EBBR/E13 - pressure calibration | sgp30ebbrE13.b1 |
D931108.4 | SGP/EBBR/E13 - solid moisture correction | sgp30ebbrE13.b1 |
D931108.6 | SGP/EBBR/E9 - pressure calibration | sgp30ebbrE9.b1 |
D931115.3 | SGP/EBBR/E8 - Drop in atmospheric pressure | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D940105.1 | Incorrect readings for pressure sensor | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D940105.10 | SGP/EBBR/E22 - Air Temperature Sensor malfunction | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D940105.2 | SGP/EBBR/E20 - Blown fuse in AEM at E20 | sgp30ebbrE20.b1, sgp5ebbrE20.b1, sgp15ebbrE20.b1 |
D940105.7 | SGP/EBBR/E22 - Incorrect averaging interval at E22 | sgp30ebbrE22.b1 |
D940105.8 | SGP/EBBR/E20 - Reprocess: Questionable data in soil heat fluxes at E20 | sgp30ebbrE20.b1, sgp15ebbrE20.b1 |
D940106.1 | SGP/EBBR/E8 - Invalid data due to AEM fuse blow out at E8 | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D940106.2 | SGP/EBBR/E12 - Invalid data due to blown AEM fuse at E12 | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D940106.5 | SGP/EBBR/E13 - Invalid data during disconnection of AEM at E13 | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D940211.1 | SGP/EBBR/E26 - Invalid data during Preventative Maintenance | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D940211.2 | SGP/EBBR/E8 - Invalid wind data frozen sensor | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D940301.2 | SGP/EBBR/E26 - Winds, SM2 Invalid data 10/10/93 - 2/2/94 | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D940301.3 | SGP/EBBR/E8 - Invalid wind speed frozen sensor | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D940301.4 | Invalid wind speed frozen sensor | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D940301.6 | SGP/EBBR/E9 - Invalid wind speed frozen sensor | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
D940329.1 | SGP/EBBR/E26 - Wind Direction Sensor Malfunction | sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp15ebbrE26.b1 |
D940329.2 | SGP/EBBR/E20 - Pressure Sensor Malfunction | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D940329.3 | SGP/E9/EBBR - AEM Malfunction | sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
D940503.1 | SGP/EBBR/E9 - AEM failure | sgp30ebbrE9.b1 |
D940511.2 | SGP/EBBR/E13 - SM4 Replacement | sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
D940606.1 | SGP/EBBR/E26 - Wind Speed Malfunction | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D940606.2 | SGP/EBBR/E26 - Wind Direction Sensor Failure | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D940616.3 | SGP/EBBR/E12 - Wind Direction | sgp30ebbrE12.b1, sgp5ebbrE12.b1, sgp15ebbrE12.b1 |
D940616.4 | SGP/EBBR/E20 - Soil Moisture Probe | sgp15ebbrE20.b1, sgp30ebbrE20.b1 |
D940711.2 | SGP/EBBR/E9 - Reprocess: Soil Heat Flow Sensor Failure | sgp30ebbrE9.b1, sgp15ebbrE9.b1 |
D940711.3 | SGP/EBBR/E12 - Aspirator Failure | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D940711.4 | SGP/EBBR/E7 - Reprocess: Soil Temperature Probe Failure | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D940808.2 | SGP/EBBR/E12 - Stopped Ventilation Fan | sgp30ebbrE12.b1, sgp5ebbrE12.b1, sgp15ebbrE12.b1 |
D940818.1 | SGP/EBBR/E26 - Soil Moisture Sensor Malfunction | sgp15ebbrE26.b1, sgp30ebbrE26.b1 |
D940826.1 | SGP/EBBR/E9 - Malfunctioning Wind Dir Sensor, Incorrect SM2 Results | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
D940826.2 | SGP/EBBR/E9 - Reprocess: Soil Heat Flux Out of Range | sgp15ebbrE9.b1, sgp30ebbrE9.b1 |
D940830.10 | SGP/EBBR/E8 - Battery Voltage Drop | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D940830.7 | SGP/EBBR/E9 - Hail Damage to Net Radiometer and Wind Speed Sensor | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
D940830.8 | Stopped Ventilation Fan | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D940909.1 | SGP/EBBR/E12 - Miscellaneous Problems Affecting Data | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D940912.1 | SGP/EBBR/E8 - Data Corruption and Loss | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D940921.1 | SGP/EBBR/E20 - Fan Stopped, T and RH Biases | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D940926.1 | Intermittent Stopped or Slowed Aspirator Fans | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D941021.1 | SGP/EBBR/E20 - Soil Moisture Probe Failure | sgp15ebbrE20.b1, sgp30ebbrE20.b1 |
D941021.10 | SGP/EBBR/E22 - Net Radiometer Hemisphere Changes | sgp5ebbrE22.b1, sgp30ebbrE22.b1, sgp15ebbrE22.b1 |
D941021.11 | SGP/EBBR/E26 - Net Radiometer Hemisphere Changes | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D941021.12 | SGP/EBBR/E12 - Net Radiometer Hemisphere Changes | sgp5ebbrE12.b1, sgp30ebbrE12.b1, sgp15ebbrE12.b1 |
D941021.13 | SGP/EBBR/E15 - Net Radiometer Hemisphere Changes | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D941021.2 | SGP/EBBR/E20 - Aspirator Fan Stoppage | sgp15ebbrE20.b1, sgp30ebbrE20.b1 |
D941021.5 | SGP/EBBR/E13 - Soil Temperature Aberations | sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
D941021.6 | SGP/EBBR/E12 - Low home_15 Values | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D941021.7 | SGP/EBBR/E20 - Net Radiometer Hemisphere Changes | sgp30ebbrE20.b1, sgp15ebbrE20.b1, sgp5ebbrE20.b1 |
D941021.8 | SGP/EBBR/E15 - Net Radiometer Hemisphere Changes | sgp30ebbrE15.b1, sgp15ebbrE15.b1, sgp5ebbrE15.b1 |
D941021.9 | SGP/EBBR/E20 - Net Radiometer Hemisphere Changes | sgp30ebbrE20.b1, sgp15ebbrE20.b1, sgp5ebbrE20.b1 |
D941114.2 | SGP/EBBR/E8 - Water in Net Radiometer | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D941114.3 | SGP/EBBR/E7 - Net Radiometer not level | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D941208.1 | SGP/EBBR/E13 - Net Radiation - Hemisphere Replacement | sgp30ebbrE13.b1, sgp5ebbrE13.b1, sgp15ebbrE13.b1 |
D941208.2 | SGP/EBBR/E7 - Radiation Shield Problem | sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D941212.1 | SGP/EBBR/E26 - Wind Direction Sensor Affect on Soil Moisture 2 | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D941212.3 | SGP/EBBR/E8 - Frozen Wind Instruments | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D941220.3 | SGP/EBBR/E7 - Water in Net Radiometer | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D941221.1 | SGP/EBBR/E26 - T/RH Check | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D950126.1 | SGP/EBBR/E13 - Incorrect Soil Temperature #4 | sgp30ebbrE13.b1, sgp15ebbrE13.b1 |
D950126.2 | SGP/EBBR/E9 - Reprocess: Soil Heat Flow Sensor 5 Incorrect Data | sgp30ebbrE9.b1, sgp15ebbrE9.b1 |
D950126.3 | SGP/EBBR/E8 - Frozen Wind Speed Instrument | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D950126.4 | SGP/EBBR/E15 - Frozen Wind Speed Instrument | sgp30ebbrE15.b1, sgp5ebbrE15.b1, sgp15ebbrE15.b1 |
D950308.2 | SGP/EBBR/E9 - Radiation Shield Problem | sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
D950308.3 | SGP/EBBR/E26 - Frozen Wind Speed Sensor | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D950626.1 | SGP/EBBR/E13 - Frozen Wind Speed Sensor | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D950626.10 | SGP/EBBR/E26 - Stopped Ventilation Fan | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D950626.11 | SGP/EBBR/E9 - Stopped Ventilation Fan | sgp30ebbrE9.b1, sgp5ebbrE9.b1, sgp15ebbrE9.b1 |
D950626.12 | SGP/EBBR/E9 - Water in Net Radiometer | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
D950626.14 | SGP/EBBR/E26 - Reprocess: Soil Heat Flow Sensor 1 Replacement | sgp15ebbrE26.b1, sgp30ebbrE26.b1 |
D950626.15 | SGP/EBBR/E12 - Wind Direction Malfunction | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D950626.16 | SGP/EBBR/E12 - Wind Direction Malfunction | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D950626.19 | SGP/EBBR/E26 - Right Air Temperature Malfunction | sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp15ebbrE26.b1 |
D950626.2 | SGP/EBBR/E13 - Frozen Wind Speed Sensor | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D950626.20 | SGP/EBBR/E15 - Battery Voltage Drop, Storage Module Change | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D950626.3 | SGP/EBBR/E15 - Frozen Wind Speed Sensor | sgp30ebbrE15.b1, sgp5ebbrE15.b1, sgp15ebbrE15.b1 |
D950626.4 | SGP/EBBR/E20 - Frozen Wind Speed Sensor | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D950626.5 | SGP/EBBR/E26 - Frozen Wind Speed Sensor | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D950626.6 | SGP/EBBR/E9 - Frozen Wind Speed Sensor | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
D950626.7 | SGP/EBBR/E13 - Reprocess: Soil Temperature Malfunction | sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
D950626.8 | SBP/EBBR/E9 - Reprocess: Soil Heat Flow Sensor 5 Incorrect Data | sgp15ebbrE9.b1, sgp30ebbrE9.b1 |
D950626.9 | SGP/EBBR/E26 - Frozen Wind Speed Sensor | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D950719.2 | SGP/EBBR/E12 - Wind Direction Malfunction | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D950803.2 | SGP/EBBR/E20 - Reprocess: Soil Temperature Malfunction | sgp15ebbrE20.b1, sgp30ebbrE20.b1 |
D950815.2 | SGP/EBBR/E13 - Low Battery Voltage and AEM Left on Manual | sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D950815.3 | SGP/EBBR/E20 - Annual Calibration Changeout | sgp30ebbrE20.b1 |
D950816.2 | SGP/EBBR/E12 - AEM Failure | sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D950816.3 | SGP/EBBR/E15 - AEM Home 15 Values High | sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D950816.4 | SGP/EBBR/E20 - AEM Home 30 Values Low | sgp5ebbrE20.b1, sgp30ebbrE20.b1 |
D950828.1 | SGP/EBBR/E13 - Battery Voltage Drop | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D950828.2 | AEM Home 30 Values Low | sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D950908.1 | SGP/EBBR/E15 - AEM Not Exchanging | sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D950908.2 | SGP/EBBR/E26 - AEM not exchanging | sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D950908.3 | SGP/EBBR/E12 - Wind Direction Malfunction | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D950921.2 | SGP/EBBR/E15 - AEM Replaced | sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D950921.3 | SGP/EBBR/E12 - AEM replacement | sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D950921.4 | SGP/EBBR/E26 - AEM replacement | sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D950922.2 | SGP/EBBR/E8 - Home 30 Values Correct | sgp30ebbrE8.b1 |
D951016.1 | SGP/EBBR/E26 - Battery Voltage Drop | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D951016.2 | SGP/EBBR/E26 - Net Radiometer Dome Punctured | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D951020.1 | SGP/EBBR/E9 - 3-month Check | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
D951113.1 | SGP/EBBR/E13 - Wind Direction Set Pin Slippage | sgp30ebbrE13.b1, sgp5ebbrE13.b1, sgp15ebbrE13.b1 |
D951113.4 | SGP/EBBR/E26 - Right Tair Offscale | sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp15ebbrE26.b1 |
D951113.6 | SGP/EBBR/E13 - Battery voltage drop | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D951116.1 | SGP/EBBR/E13 - AEM Exchange Stopped in Snow and Freezing Rain | sgp30ebbrE13.b1 |
D951116.2 | SGP/EBBR/E20 - Soil Heat Flow Zero Under Unchanging Conditions | sgp30ebbrE20.b1 |
D951116.3 | SGP/EBBR/E26 - Wind Speed Cups Frozen | sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp15ebbrE26.b1 |
D951116.4 | SGP/EBBR/E15 - AEM Exchange Stopped in Snow and Freezing Rain | sgp30ebbrE15.b1 |
D951116.5 | SGP/EBBR/E9 - AEM Exchange Stopped in Snow and Freezing Rain | sgp30ebbrE9.b1 |
D951116.6 | SGP/EBBR/E20 - AEM Exchange Stopped in Snow and Freezing Rain | sgp30ebbrE20.b1 |
D951127.1 | SGP/EBBR/E9 - Annual calibration changeout | sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
D951127.3 | SGP/EBBR/E12 - AEM Failure/Right Radiation Shield OFF | sgp30ebbrE12.b1 |
D951220.5 | SGP/EBBR/E15 - Annual calibration changeout | sgp5ebbrE15.b1, sgp30ebbrE15.b1 |
D951220.6 | SGP/EBBR/E13 - battery voltage drop | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D951221.1 | Annual calibration changeout | sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D960104.2 | SGP/EBBR/E26 - Wind Speed Cups Frozen | sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp15ebbrE26.b1 |
D960104.3 | SGP/EBBR/E22 - Wind Speed Cups Frozen | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D960104.4 | SGP/EBBR/E15 - Wind Speed Cups Frozen | sgp30ebbrE15.b1, sgp5ebbrE15.b1, sgp15ebbrE15.b1 |
D960104.5 | SGP/EBBR/E7 - Wind Speed Cups Frozen | sgp30ebbrE7.b1, sgp5ebbrE7.b1, sgp15ebbrE7.b1 |
D960104.8 | SGP/EBBR/E26 - Right Tair Offscale | sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp15ebbrE26.b1 |
D960119.1 | SGP/EBBR/E15 - Reprocess: Soil Temperature Probe Malfunction | sgp30ebbrE15.b1, sgp15ebbrE15.b1 |
D960119.2 | SGP/EBBR/E15 - Battery Voltage Drop | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D960119.3 | SGP/EBBR/E26 - Right tair Offscale | sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp15ebbrE26.b1 |
D960119.5 | SGP/EBBR/E7 - Soil Moisture Probes Offscale | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
D960320.2 | SGP/EBBR/E7 - Wind Speed Cups Frozen | sgp30ebbrE7.b1, sgp5ebbrE7.b1, sgp15ebbrE7.b1 |
D960320.3 | SGP/EBBR/E12 - Wind Speed Cups Frozen | sgp30ebbrE12.b1, sgp5ebbrE12.b1, sgp15ebbrE12.b1 |
D960321.1 | SGP/EBBR/E9 - Wind Speed Cups Frozen | sgp30ebbrE9.b1, sgp5ebbrE9.b1, sgp15ebbrE9.b1 |
D960408.2 | SGP/EBBR/E22 - Reprocess: Soil Heat Flow Sensor 4 Incorrect Data | sgp15ebbrE22.b1, sgp30ebbrE22.b1 |
D960409.1 | SGP/EBBR/E26 - Right Air Temperature Malfunction | sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp15ebbrE26.b1 |
D960501.4 | SGP/EBBR/E26 - Right Air Temp Malfunction | sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp15ebbrE26.b1 |
D960502.1 | SGP/EBBR/E20 - Temperature/RH Probes Malfunction and Replacement | sgp30ebbrE20.b1, sgp5ebbrE20.b1, sgp15ebbrE20.b1, sgp5ebbrE20.00 |
D960508.10 | SGP/EBBR/E26 - 3-month check | sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp15ebbrE26.b1 |
D960508.2 | SGP/EBBR/E7 - 3-month check | sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D960508.3 | SGP/EBBR/E8 - 3-month check | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D960508.4 | SGP/EBBR/E9 - 3-month check | sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
D960508.5 | SGP/EBBR/E12 - 3-month check | sgp30ebbrE12.b1, sgp5ebbrE12.b1, sgp15ebbrE12.b1 |
D960508.6 | SGP/EBBR/E13 - 3-month check | sgp30ebbrE13.b1, sgp5ebbrE13.b1, sgp15ebbrE13.b1 |
D960508.7 | SGP/EBBR/E15 - 3-month check | sgp30ebbrE15.b1, sgp5ebbrE15.b1, sgp15ebbrE15.b1 |
D960508.8 | SGP/EBBR/E20 - 3-month check | sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D960508.9 | SGP/EBBR/E22 - 3-month check | sgp30ebbrE22.b1, sgp5ebbrE22.b1, sgp15ebbrE22.b1 |
D960606.1 | SGP/EBBR/E13 - Annual Calibration changeout | sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D961122.1 | SGP/EBBR/E8 - Annual Calibration Changeout | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D961122.10 | SGP/EBBR/E9 - Air Temperature Thermocouples Incorrect | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
D961122.11 | SGP/EBBR/E13 - Wind Direction Incorrect | sgp30ebbrE13.b1, sgp5ebbrE13.b1, sgp15ebbrE13.b1 |
D961122.12 | SGP/EBBR/E26 - Wind Direction Sensor Failure | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D961122.13 | SGP/EBBR/E26 - Right Air Temperature Malfunction | sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp15ebbrE26.b1 |
D961122.14 | SGP/EBBR/E8 - Incorrect Right Humidity Probe Temperature | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D961122.15 | SGP/EBBR/E8 - Reprocess: Soil Heat Flow Sensor 2 Incorrect Data | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D961122.17 | SGP/EBBR/E20 - Incorrect Data from Left Humidity Probe | sgp30ebbrE20.b1, sgp15ebbrE20.b1 |
D961122.2 | SGP/EBBR/E22 - Annual Calibration Changeout | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D961122.20 | SGP/EBBR/E7 - EBBR 3-Month Checks - 10 EFs | sgp30ebbrE7.b1, sgp5ebbrE7.b1, sgp15ebbrE7.b1 |
D961122.21 | SGP/EBBR/E8 - EBBR 3 Month Checks - 10 EFs | sgp15ebbrE8.b1, sgp5ebbrE8.b1, sgp30ebbrE8.b1 |
D961122.22 | SGP/EBBR/E9 - EBBR 3 Month Checks - 10 EFs | sgp15ebbrE9.b1, sgp5ebbrE9.b1, sgp30ebbrE9.b1 |
D961122.23 | SGP/EBBR/E12 - EBBR 3 Month Checks - 10 EFs | sgp30ebbrE12.b1, sgp15ebbrE12.b1, sgp5ebbrE12.b1 |
D961122.24 | SGP/EBBR/E13 - EBBR 3 Month Checks - 10 EFs | sgp15ebbrE13.b1, sgp5ebbrE13.b1, sgp30ebbrE13.b1 |
D961122.25 | SGP/EBBR/E15 - EBBR 3 Month Checks - 10 EFs | sgp30ebbrE15.b1, sgp15ebbrE15.b1, sgp5ebbrE15.b1 |
D961122.26 | SGP/EBBR/E20 - EBBR 3 Month Checks - 10 EFs | sgp30ebbrE20.b1, sgp15ebbrE20.b1, sgp5ebbrE20.b1 |
D961122.27 | SGP/EBBR/E22 - EBBR 3 Month Checks - 10 EFs | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D961122.28 | SGP/EBBR/E26 - EBBR 3 Month Checks - 10 EFs | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D961122.3 | SGP/EBBR/E12 - Annual Calibration Changeout | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D961122.4 | SGP/EBBR/E26 - Annual Calibration Changeout | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D961122.5 | SGP/EBBR/E26 - Wind Direction Sensor Failure | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D961122.6 | SGP/EBBR/E8 - Incorrect Soil Moisture #1 | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D961122.7 | SGP/EBBR/E8 - Incorrect Soil Moisture #2, Soil Temperature #2 | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D961122.8 | SGP/EBBR/E12 - Left Air Temperature Thermocouple Incorrect | sgp30ebbrE12.b1, sgp5ebbrE12.b1, sgp15ebbrE12.b1 |
D961122.9 | SGP/EBBR/E12 - Battery Voltage Low - Fluxes Affected | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D970330.1 | SGP/EBBR/E13 - Frozen Wind Speed Sensor | sgp30ebbrE13.b1, sgp5ebbrE13.b1, sgp15ebbrE13.b1 |
D970330.2 | SGP/EBBR/E15 - Wind Speed Cups frozen | sgp30ebbrE15.b1, sgp5ebbrE15.b1, sgp15ebbrE15.b1 |
D970330.4 | SGP/EBBR/E7- Wind Speed Cups Frozen | sgp30ebbrE7.b1, sgp5ebbrE7.b1, sgp15ebbrE7.b1 |
D970330.5 | SGP/EBBR/E20 - Wind Speed Cups Frozen | sgp30ebbrE20.b1, sgp5ebbrE20.b1, sgp15ebbrE20.b1 |
D970330.6 | SGP/EBBR/E26 - Wind Direction Sensor Failure | sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp15ebbrE26.b1 |
D970401.1 | Controlled burn at E12 on 3/20/97 and affected data | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1, sgpsirosE12.a1 |
D970616.1 | SGP/EBBR/E8 - Incorrect Soil Temperature #s 1 and 2 | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D970623.2 | SGP/EBBR/E9 - Temperature Portion of T/RH Probe | sgp15ebbrE9.b1, sgp30ebbrE9.b1 |
D970627.2 | SGP/EBBR/E26 - Reprocess: E26 Soil Heat Flow Sensor 1 Replacement | sgp15ebbrE26.b1, sgp30ebbrE26.b1 |
D970703.1 | SGP/EBBR/E8 - Incorrect Soil Temperature #s 1 and 2 | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D970703.3 | SGP/EBBR/E15 - Connections Affecting Home Signals | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D970707.1 | SGP/EBBR/E22 - Reprocess: E22 Soil Heat Flow Sensor # 5 Incorrect Data | sgp15ebbrE22.b1, sgp30ebbrE22.b1 |
D970716.2 | SGP/EBBR/E8 - AEM Inoperative | sgp30ebbrE8.b1 |
D970729.1 | SGP/EBBR/E20 - Reprocess: E20 Soil Probes #3 Cables Cut | sgp15ebbrE20.b1, sgp30ebbrE20.b1 |
D970730.1 | SGP/EBBR/E22 - Right Aspirator Unit Stolen | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D970730.2 | SGP/EBBR/E12 - Wind Direction Sensor Malfunctions | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D970730.3 | SGP/EBBR/E13 - CR10 Malfunction | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D971017.1 | SGP/EBBR/E8 - Reprocess: E8 EBBR Program Compile Error, Effect On Net Radiation | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D971017.2 | E8 EBBR Lightning Strike | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D971017.3 | SGP/EBBR/E19 - Reprocess: E19 EBBR Soil heat Flow Probe #5 Upsidedown | sgp15ebbrE19.b1, sgp30ebbrE19.b1 |
D971017.7 | SGP/EBBR/E22 - Reprocess: E22 EBBR Soil Heat Flow Probe #5 Incorrect Data | sgp15ebbrE22.b1, sgp30ebbrE22.b1 |
D971017.8 | SGP/EBBR/E8 - Reprocess: E8 EBBR Lightning Strike / sensor Replacements | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D971021.1 | SGP/EBBR/E7 - EBBR Biannual Checks | sgp30ebbrE7.b1, sgp5ebbrE7.b1, sgp15ebbrE7.b1 |
D971021.10 | SGP/EBBR/E2 - EBBR Biannual Checks | sgp15ebbrE2.b1, sgp5ebbrE2.b1, sgp30ebbrE2.b1 |
D971021.2 | SGP/EBBR/E8 - EBBR Biannual Checks | sgp15ebbrE8.b1, sgp5ebbrE8.b1, sgp30ebbrE8.b1 |
D971021.3 | SGP/EBBR/E9 - EBBR Biannual Checks | sgp15ebbrE9.b1, sgp5ebbrE9.b1, sgp30ebbrE9.b1 |
D971021.4 | SGP/EBBR/E12 - EBBR Biannual Checks | sgp30ebbrE12.b1, sgp15ebbrE12.b1, sgp5ebbrE12.b1 |
D971021.5 | SGP/EBBR/E13 - EBBR Biannual Checks | sgp15ebbrE13.b1, sgp5ebbrE13.b1, sgp30ebbrE13.b1 |
D971021.6 | SGP/EBBR/E15 - EBBR Biannual Checks | sgp30ebbrE15.b1, sgp5ebbrE15.b1, sgp15ebbrE15.b1 |
D971021.7 | SGP/EBBR/E20 - EBBR Biannual Checks | sgp30ebbrE20.b1, sgp15ebbrE20.b1, sgp5ebbrE20.b1 |
D971021.8 | SGP/EBBR/E22 - EBBR Biannual Checks | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D971021.9 | SGP/EBBR/E26 - EBBR Biannual Checks | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D971027.6 | SGP/EBBR/E18 - Left RH Probe Temperature Incorrect | sgp15ebbrE18.b1, sgp30ebbrE18.b1 |
D971027.7 | SGP/EBBR/E2 - EBBR E2 Half Hour Incorrect | sgp30ebbrE2.b1 |
D971027.8 | SGP/EBBR/E9 - Wind Direction Sensor Malfunction | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
D971027.9 | SGP/EBBR/E26 - Bad Dates | sgp30ebbrE26.b1 |
D971111.1 | SGP/EBBR/E8 - Reprocess: SHF2 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D980110.1 | SGP/EBBR/E19 - Battery Voltage Drop | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
D980110.2 | SGP/EBBR/E19 - Noisy Sensible and Latent Heat Fluxes | sgp30ebbrE19.b1, sgp5ebbrE19.b1, sgp15ebbrE19.b1 |
D980117.1 | SGP/EBBR/E22 - Reprocess: E22 Soil Moisture Values Off End of Calibration at High Soil Moistures | sgp15ebbrE22.b1, sgp30ebbrE22.b1 |
D980117.10 | SGP/EBBR/E2 - Reprocess: E2 Soil Heat Flow #4 Incorrect - Replaced | sgp15ebbrE2.b1, sgp30ebbrE2.b1 |
D980117.11 | SGP/EBBR/E2 - EBBR Frozen Wind Speed | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
D980117.13 | SGP/EBBR/E9 - EBBR Frozen Wind Speed | sgp30ebbrE9.b1, sgp5ebbrE9.b1, sgp15ebbrE9.b1 |
D980117.14 | SGP/EBBR/E15 - EBBR Frozen Wind Speed | sgp5ebbrE15.b1, sgp30ebbrE15.b1, sgp15ebbrE15.b1 |
D980117.15 | SGP/EBBR/E7 - EBBR Frozen Wind Speed | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D980117.16 | SGP/EBBR/E8- EBBR Frozen Wind Speed | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D980117.17 | SGP/EBBR/E12- EBBR Frozen Wind Speed | sgp5ebbrE12.b1, sgp15ebbrE12.b1, sgp30ebbrE12.b1 |
D980117.18 | SGP/EBBR/E19- EBBR Frozen Wind Speed | sgp5ebbrE19.b1, sgp30ebbrE19.b1, sgp15ebbrE19.b1 |
D980117.19 | SGP/EBBR/E13- EBBR Frozen Wind Speed | sgp5ebbrE13.b1, sgp30ebbrE13.b1, sgp15ebbrE13.b1 |
D980117.20 | SGP/EBBR/E20- EBBR Frozen Wind Speed | sgp5ebbrE20.b1, sgp30ebbrE20.b1, sgp15ebbrE20.b1 |
D980117.21 | SGP/EBBR/E25- EBBR Frozen Wind Speed | sgp15ebbrE25.b1, sgp30ebbrE25.b1, sgp5ebbrE25.b1 |
D980117.22 | SGP/EBBR/E26- EBBR Frozen Wind Speed | sgp5ebbrE26.b1, sgp15ebbrE26.b1, sgp30ebbrE26.b1 |
D980117.23 | SGP/EBBR/E2 - EBBR AEMs Not Exchanging (Iced Or Fuse Blown) | sgp15ebbrE2.b1, sgp5ebbrE2.b1, sgp30ebbrE2.b1 |
D980117.24 | SGP/EBBR/E15 - EBBR AEMs Not Exchanging (Iced Or Fuse Blown) | sgp30ebbrE15.b1, sgp5ebbrE15.b1, sgp15ebbrE15.b1 |
D980117.25 | SGP/EBBR/E9 - EBBR AEMs Not Exchanging (Iced Or Fuse Blown) | sgp15ebbrE9.b1, sgp5ebbrE9.b1, sgp30ebbrE9.b1 |
D980117.26 | SGP/EBBR/E12 - EBBR AEMs Not Exchanging (Iced Or Fuse Blown) | sgp30ebbrE12.b1, sgp15ebbrE12.b1, sgp5ebbrE12.b1 |
D980117.27 | SGP/EBBR/E13 - EBBR AEMs Not Exchanging (Iced Or Fuse Blown) | sgp15ebbrE13.b1, sgp5ebbrE13.b1, sgp30ebbrE13.b1 |
D980117.28 | SGP/EBBR/E20 - EBBR AEMs Not Exchanging (Iced Or Fuse Blown) | sgp30ebbrE20.b1, sgp5ebbrE20.b1, sgp15ebbrE20.b1 |
D980117.29 | SGP/EBBR/E15 - EBBR Net Radiometer Domes Pierced, Water Inside - Data Affected | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D980117.3 | SGP/EBBR/E25 - Bowen Ratio Usually Negative - Fluxes Incorrect | sgp15ebbrE25.b1, sgp30ebbrE25.b1, sgp5ebbrE25.b1 |
D980117.31 | SGP/EBBR/E9 - EBBR Net Radiometer Domes Pierced, Water Inside - Data Affected | sgp30ebbrE9.b1, sgp5ebbrE9.b1, sgp15ebbrE9.b1 |
D980117.32 | SGP/EBBR/E12 - EBBR Net Radiometer Domes Pierced, Water Inside - Data Affected | sgp5ebbrE12.b1, sgp30ebbrE12.b1, sgp15ebbrE12.b1 |
D980117.33 | SGP/EBBR/E26 - EBBR Net Radiometer Domes Pierced, Water Inside - Data Affected | sgp5ebbrE26.b1, sgp30ebbrE26.b1, sgp15ebbrE26.b1 |
D980117.5 | SGP/EBBR/E26 - EBBR AEMs Not Exchanging (Iced Or Fuse Blown) | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D980117.6 | SGP/EBBR/E12 - AEM Failure and Replacement | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D980117.7 | SGP/EBBR/E8 - Battery Low - Data Affected | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D980117.8 | SGP/EBBR/E2 - EBBR Net Radiometer Domes Pierced, Water Inside - Data Affected | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
D980117.9 | SGP/EBBR/E8 - Reprocess: E8 Soil Heat Flow #2 Failure - Replaced | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D980410.1 | SGP/EBBR/E7 - Right T/RH Probe Temperature Incorrect | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1, sgp5ebbrE7.00 |
D980410.10 | SGP/EBBR/E20 - Wind Speed Sensor Frozen | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1, sgp5ebbrE20.00 |
D980410.2 | SGP/EBBR/E2 - EBBR E2 Wind Direction Sensors Frozen | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
D980410.3 | SGP/EBBR/E7 - Wind Speed Sensor Frozen | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D980410.4 | SPG/EBBR/E8 - Wind Speed Sensor and AEM Frozen | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D980410.5 | SGP/EBBR/E9 - Wind Speed Sensor Frozen | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1, sgp5ebbrE9.00 |
D980410.8 | SGP/EBBR/E13 - Wind Speed Sensor Frozen | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D980410.9 | SGP/EBBR/E15 - Wind Speed Sensor Frozen | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1, sgp5ebbrE15.00 |
D980413.5 | SGP/EBBR/E22 - Low Battery - Resulting In Incorrect Data | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D980413.6 | SGP/EBBR/E2 - EBBR E2 Incorrect Data | sgp30ebbrE2.b1, sgp5ebbrE2.b1, sgp15ebbrE2.b1 |
D980416.1 | SGP/EBBR/E2 - EBBR E2 3 Month Check | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
D980416.10 | SGP/EBBR/E9 - 3-Month Check | sgp30ebbrE9.b1, sgp5ebbrE9.b1, sgp15ebbrE9.b1 |
D980416.11 | SGP/EBBR/E12 - 3-Month Check | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D980416.12 | SGP/EBBR/E13 - 3-Month Check | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D980416.13 | SGP/EBBR/E15 - 3-Month Check | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D980416.14 | SGP/EBBR/E18 - 3 Month Check | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
D980416.15 | SGP/EBBR/E19 - 3-Month Check | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
D980416.16 | SGP/EBBR/E20 - 3 Month Check | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D980416.17 | SGP/EBBR/E22 - 3-Month Check | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D980416.18 | SGP/EBBR/E26 - 3-Month Check | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D980416.36 | SGP/EBBR/E8 - Wind Speed Frozen | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D980416.37 | SGP/EBBR/E8 - Wind Speed Frozen | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D980416.38 | SGP/EBBR/E8 - Wind Speed Frozen | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D980416.8 | SGP/EBBR/E7 - 3-Month Check | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D980416.9 | SGP/EBBR/E8 - 3-Month Check | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D980507.5 | SGP/EBBR/E26 - Character In EBBR E26, 30 Minute Data | sgp30ebbrE26.b1 |
D980513.4 | SGP/EBBR/E15 - Pressure Diurnal Fluctuations | sgp30ebbrE15.b1 |
D980513.5 | SGP/EBBR/E15 - Pressure Too High | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D980513.6 | SGP/EBBR/E15 - Left T/RH Probe Temperature Too Large | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D980514.1 | SGP/EBBR/E25 - Right Humidity Incorrect | sgp15ebbrE25.b1, sgp30ebbrE25.b1, sgp5ebbrE25.b1 |
D980526.2 | EBBR E8 Net Radiation Data Quality | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D980528.13 | SGP/EBBR/E12 - Wind Direction Sensor Failure | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D980529.10 | SGP/EBBR/E13 - Six Month Checks | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D980529.11 | SGP/EBBR/E15 - Six Month Checks | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D980529.12 | SGP/EBBR/E18 - Six Month Checks | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
D980529.13 | SGP/EBBR/E19 - Six-Month Checks | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
D980529.14 | SGP/EBBR/E20 - Six Month Checks | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D980529.15 | SGP/EBBR/E22 - Six Month Checks | sgp30ebbrE22.b1, sgp5ebbrE22.b1, sgp15ebbrE22.b1 |
D980529.16 | SGP/EBBR/E25 - Six-Month Checks | sgp15ebbrE25.b1, sgp30ebbrE25.b1, sgp5ebbrE25.b1 |
D980529.17 | SGP/EBBR/E26 - Six Month Checks | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D980529.4 | SGP/EBBR/E2 - EBBR E2 Six Month Checks | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
D980529.6 | SGP/EBBR/E7 - Six-Month Checks | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D980529.7 | SGP/EBBR/E8 - Six Month Checks | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D980529.8 | SGP/EBBR/E9 - Six Month Checks | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
D980529.9 | SGP/EBBR/E12 - Six Month Checks | sgp30ebbrE12.b1, sgp5ebbrE12.b1, sgp15ebbrE12.b1 |
D980603.1 | EBBR E8 Adjustment of Negative Net Radiation Values | sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D980603.2 | EBBR E8 Adjustment of Negative Net Radiation Values | sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D980603.3 | SGP/EBBR/E8 - Wind Direction Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D980623.2 | SGP/EBBR/E18 - Reprocess: E18 EBBR Net Radiometer Calibration | sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
D980731.2 | SGP/EBBR/E20 - Wind Speed Incorrect | sgp5ebbrE20.b1, sgp15ebbrE20.b1, sgp30ebbrE20.b1 |
D980731.3 | SGP/EBBR/E25 - Data Incorrect | sgp15ebbrE25.b1, sgp30ebbrE25.b1, sgp5ebbrE25.b1 |
D980731.4 | SGP/EBBR/E25 - Data Incorrect | sgp5ebbrE25.b1, sgp15ebbrE25.b1, sgp30ebbrE25.b1 |
D981112.4 | SGP/EBBR/E2 - EBBR E2 Net Radiation Incorrect - Water in Bottom Dome | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
D981112.5 | SGP/EBBR/E13 - Net Radiation Incorrect - Water in Bottom Dome | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D981113.1 | SGP/EBBR/E20 - Incorrect Data - Cattle Vandalism | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D981113.2 | SGP/EBBR/E7 - Data Incorrect Because of Recompile | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D981113.3 | SGP/EBBR/E15 - Aspirator loosened from AEM mount | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D990127.10 | SGP/EBBR/E18 - Water In Net Radiometer, Flux Data Incorrect | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
D990127.11 | SGP/EBBR/E25 - All Data Incorrect (zeros) | sgp30ebbrE25.b1, sgp15ebbrE25.b1, sgp5ebbrE25.b1, sgp5ebbrE25.00 |
D990127.12 | SGP/EBBR/E9 - Wind Direction Data Incorrect | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1, sgp5ebbrE9.00 |
D990127.13 | SGP/EBBR/E8 - Wind Direction and Speed Data Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D990127.15 | SGP/EBBR/E25 - Wind Direction and Speed Data Incorrect | sgp15ebbrE25.b1, sgp30ebbrE25.b1, sgp5ebbrE25.b1, sgp5ebbrE25.00 |
D990127.2 | SGP/EBBR/E22 - Battery Voltage Low, Data Incorrect | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1, sgp5ebbrE22.00 |
D990127.21 | SGP/EBBR/E12 - Wind Direction Data Incorrect (revised) | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D990127.22 | SGP/EBBR/E7 - Wind Direction Data Incorrect (revised) | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1, sgp5ebbrE7.00 |
D990127.23 | SGP/EBBR/E18 - Wind Direction Data Incorrect (revised) | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1, sgp5ebbrE18.00 |
D990127.24 | SGP/EBBR/E19 - Wind Direction Data Incorrect (revised) | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1, sgp5ebbrE19.00 |
D990127.25 | SGP/EBBR/E20 - Wind Direction Data Incorrect (revised) | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D990127.26 | SGP/EBBR/E15 - Wind Direction Data Incorrect (revised) | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D990127.28 | SGP/EBBR/E25 - Wind Direction Data Incorrect | sgp15ebbrE25.b1, sgp30ebbrE25.b1, sgp5ebbrE25.b1 |
D990127.29 | SGP/EBBR/E20 - Wind Direction Data Incorrect | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D990127.30 | SGP/EBBR/E22 - Wind Direction Data Incorrect | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1, sgp5ebbrE22.00 |
D990127.31 | SGP/EBBR/E26 - Wind Direction Data Incorrect | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D990127.32 | SGP/EBBR/E13 - Wind Speed Data Incorrect | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D990127.33 | SGP/EBBR - EBBR E13 Wind Speed Data Incorrect | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1, sgp5ebbrE13.00 |
D990127.34 | SGP/EBBR - EBBR E13 Wind Speed Data Incorrect | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1, sgp5ebbrE13.00 |
D990127.35 | SGP/EBBR/E19 - Wind Speed Data Incorrect | sgp30ebbrE19.b1, sgp5ebbrE19.b1, sgp5ebbrE19.00, sgp15ebbrE19.b1 |
D990127.36 | SGP/EBBR/E19 - Wind Speed Data Incorrect | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1, sgp5ebbrE19.00 |
D990127.37 | SGP/EBBR - EBBR E18 Wind Speed Data Incorrect | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1, sgp5ebbrE18.00 |
D990127.38 | SGP/EBBR/E18 - Wind Speed Data Incorrect | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1, sgp5ebbrE18.00 |
D990127.39 | SGP/EBBR/E18 - Wind Speed Data Incorrect | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1, sgp5ebbrE18.00 |
D990127.40 | SGP/EBBR/E20 - Wind Speed Data Incorrect | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D990127.41 | SGP/EBBR/E20 - Wind Speed Data Incorrect | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D990127.42 | SGP/EBBR/E26 - Wind Speed Data Incorrect | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D990127.43 | SGP/EBBR - EBBR E26 Wind Speed Data Incorrect | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp5ebbrE26.00 |
D990127.44 | SGP/EBBR/E15 - Wind Speed Data Incorrect | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1, sgp5ebbrE15.00 |
D990127.45 | SGP/EBBR - EBBR E15 Wind Speed Data Incorrect | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1, sgp5ebbrE15.00 |
D990127.46 | SGP/EBBR - EBBR E15 Wind Speed Data Incorrect | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1, sgp5ebbrE15.00 |
D990127.7 | SGP/EBBR/E26 - Battery Voltage Low, Data Incorrect | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp5ebbrE26.00 |
D990127.8 | SGP/EBBR/E9 - EBBR E9 Net Radiometer Dome Broken, Flux Data Incorrect | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1, sgp5ebbrE9.00 |
D990128.1 | SGP/EBBR/E12 - Wind Speed Data Incorrect | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D990128.10 | SGP/EBBR/E25 - Wind Speed Data Incorrect | sgp15ebbrE25.b1, sgp30ebbrE25.b1, sgp5ebbrE25.b1 |
D990128.11 | SGP/EBBR/E25 - Wind Speed Data Incorrect | sgp15ebbrE25.b1, sgp30ebbrE25.b1, sgp5ebbrE25.b1 |
D990128.2 | SGP/EBBR - EBBR E12 Wind Speed Data Incorrect | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1, sgp5ebbrE12.00 |
D990128.3 | SGP/EBBR/E7 - EBBR E7 Wind Speed Data Incorrect | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D990128.4 | SGP/EBBR/E7 - Wind Speed Data Incorrect | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
D990128.5 | SGP/EBBR/E2 - EBBR E2 Wind Speed Data Incorrect | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1, sgp5ebbrE2.00 |
D990128.6 | SGP/EBBR/E9 - Wind Speed Data Incorrect | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1, sgp5ebbrE9.00 |
D990128.7 | SGP/EBBR/E9 - Wind Speed Data Incorrect | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1, sgp5ebbrE9.00 |
D990128.8 | SGP/EBBR/E8 - Wind Speed Data Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D990129.1 | SGP/EBBR - EBBR E12 Wind Speed Data Incorrect | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1, sgp5ebbrE12.00 |
D990129.2 | SGP/EBBR/E12 - Data Incorrect | sgp30ebbrE12.b1, sgp5ebbrE12.b1, sgp5ebbrE12.00, sgp15ebbrE12.b1 |
D990129.3 | SGP/EBBR/E12 - Temperature, RH, and Vapor Pressure Data Incorrect | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
D990129.4 | SGP/EBBR/E12 - All Temperature Measurements Incorrect | sgp30ebbrE12.b1, sgp5ebbrE12.b1, sgp15ebbrE12.b1 |
D990129.5 | SGP/EBBR/E26 - Battery Voltage Low | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp5ebbrE26.00 |
D990129.8 | SGP/EBBR/E20 - Data Incorrect | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1, sgp5ebbrE20.00 |
D990202.1 | SGP/EBBR/E2 - Wind Speed Iced, Data Incorrect | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1, sgp5ebbrE2.00 |
D990202.3 | SGP/EBBR - EBBR E8 Wind Speed Iced, Data Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp5ebbrE8.00 |
D990208.1 | SGP/EBBR/E8 - Wind Direction Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D990208.2 | SGP/EBBR/E8 - Wind Speed Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp5ebbrE8.00 |
D990218.2 | SGP/EBBR/E20 - Pressure Low | sgp5ebbrE20.b1, sgp15ebbrE20.b1, sgp30ebbrE20.b1 |
D990311.1 | SGP/EBBR/E15 - Temperatures Incorrect | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D990316.1 | SGP/EBBR/E22 - Right Aspirator Fan Replaced | sgp15ebbrE22.b1, sgp30ebbrE22.b1 |
D990409.6 | SGP/EBBR/E26 - Calibration Changeout | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp5ebbrE26.00 |
D990409.8 | SGP/EBBR/E25 - CR10 Program Revision (Vs. 7) | sgp15ebbrE25.b1, sgp30ebbrE25.b1, sgp5ebbrE25.b1, sgp5ebbrE25.00 |
D990409.9 | SGP/EBBR/E19 - Net Radiation Incorrect | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
D990415.1 | SGP/EBBR/E13 - Wind Speed Frozen | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
D990415.2 | SGP/EBBR/E8 - Wind Speed Frozen | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D990415.3 | SGP/EBBR/E20 - Air Temperature Incorrect | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D990421.2 | SGP/EBBR/E26 - Incorrect Data | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgpebbrE26.00, sgp5ebbrE26.00 |
D990421.4 | SGP/EBBR/E19 - Incorrect Tref, Tair Data, Bowen Ratio, latent and sensible heat fluxes | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
D990622.2 | SGP/EBBR/E18 - Pressure Incorrect | sgp30ebbrE18.b1, sgp5ebbrE18.b1, sgp5ebbrE18.00 |
D990622.3 | SGP/EBBR/E19 - Temperature of Left Humidity Probe Incorrect | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
D990622.4 | SGP/EBBR/E12 - RH top Data Zero | sgp30ebbrE12.b1 |
D990622.5 | SGP/EBBR/E9 - Wind Speed Incorrect | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1, sgp5ebbrE9.00 |
D990622.6 | SGP/EBBR/E26 - Battery Wire Problem Causes Incorrect Data | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp5ebbrE26.00 |
D990622.7 | SGP/EBBR/E26 - Battery Wire Problem Causes Questionable Data | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp5ebbrE26.00 |
D990623.1 | SGP/EBBR/E9 - Right Aspirator Not Working | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
D990623.3 | SGP/EBBR/E25 - Data Incorrect During PM | sgp15ebbrE25.b1, sgp30ebbrE25.b1, sgp5ebbrE25.b1, sgp5ebbrE25.00 |
D990623.4 | SGP/EBBR/E25 - Data Incorrect During PM | sgp15ebbrE25.b1, sgp30ebbrE25.b1, sgp5ebbrE25.b1 |
D990623.5 | SGP/EBBR/E25 - Data Incorrect During PM | sgp15ebbrE25.b1, sgp30ebbrE25.b1, sgp5ebbrE25.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/11/1999 | 0000 | 06/29/1999 | 2145 |
Subject: | SGP/EBBR/E26 - Temperature from Left T/RH Probe Incorrect |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | The temperature from the left T/RH probe went offscale frequently. This temperature is used in the calculation of the vapor pressure. Therefore, top and bottom Thum measurements, left Thum, top and bottom vapor pressure measurements, vapor pressure gradients, and the resulting latent and sensible heat fluxes and bowen ratio were frequently incorrect for the period stated. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/20/1999 | 1430 | 07/20/1999 | 1830 |
Subject: | SGP/EBBR - EBBR E19 Changeout, Data Missing/Incorrect |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | During changeout of the EBBR system at E19, data was missing for the time period, except for 1815-1830 GMT, when the data was incorrect. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/29/1999 | 1415 | 11/02/1999 | 2030 |
Subject: | SGP/EBBR - EBBR E22 Moved, Data Missing/Incorrect |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | During, and as a consequence of, the relocation of E22 EBBR, some data is missing or incorrect during the period above. The data was sneakernetted on 3 Nov 99, so some of the data may have been recovered. Some of the data during the period is incorrect as a consequence of damage done by cattle during the period; inspection of the data is required to determine incorrect data. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/29/1999 | 1430 | 11/03/1999 | 1930 |
Subject: | SGP/EBBR - EBBR E22 Net Radiation Incorrect |
DataStreams: | sgp30ebbrE22.b1, sgp5ebbrE22.b1, sgp15ebbrE22.b1 |
Description: | The upper dome of the net radiometer was broken, probably by migrating birds, and water had drained into the bottom dome. Since much of the data is missing during the period, it is not clear exactly when the top dome was broken during the period. All net radiation data and sensible and latent heat fluxes data are incorrect for this period. The affected data fields are: 5 minute: q 15 minute: mv_q 30 minute: q, e, h |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/25/1999 | 1430 | 10/25/1999 | 1845 |
Subject: | SGP/EBBR - EBBR E25 Data Missing/Incorrect During System Changeout |
DataStreams: | sgp15ebbrE25.b1, sgp30ebbrE25.b1, sgp5ebbrE25.b1 |
Description: | All data was missing or incorrect during the EBBR system changeout. |
Measurements: | sgp30ebbrE25.b1:
sgp5ebbrE25.b1:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/07/1999 | 1800 | 04/21/1999 | 1730 |
Subject: | SGP/EBBR - EBBR E20 Data Incorrect from SHF #5 |
DataStreams: | sgp30ebbrE20.b1, sgp15ebbrE20.b1 |
Description: | something, since the problem began during J-panel changeout the 4/7/99 PM. At the 7/21/99 PM, the reference temperature was repaired and reseated. At that time the SHF #5 data became correct again; the two events are probably related only in that other connections, including that of the SHF #5, were probably reseated at the same time. The report of the PM activity is not detailed enough to determine the exact causes and fixes. The incorrect SHF #5 data can be identified by it being an order of magnitude larger than the values from the other soil heat flow sensors. Affected data are: 5 minute: none 15 minute: mv_hft5 30 minute: shf5, c_shf5, g5, ave_shf, e, h |
Measurements: | sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/18/1999 | 1715 | 06/18/1999 | 1745 |
Subject: | SGP/EBBR - EBBR E20 Thermocouple Data Incorrect |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | Left and right thermocouple air temperature data are incorrect during the replacement of the right thermocouple. Affected data are: 5 minute: tair_bot, tair_top 15 minute: tai_r, tair_l 30 minute: tair_bot, tair_top, e, h, bowen |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/11/1999 | 0000 | 06/29/1999 | 2145 |
Subject: | SGP/EBBR/E26 - Left T/RH Probe Temperature Data Incorrect |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | Left T/RH probe temperature is frequently offscale during the period. Examine the data (compare with the the T/RH right temperature) to determine which times are incorrect. Replacement of the probe on 29 June 1999 solved the problem. Affected data are: 5 minute: tair_bot, tair_top 15 minute: tair_l 30 minute: tair_bot, tair_top, e, h, bowen |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/16/1999 | 1915 | 06/16/1999 | 2000 |
Subject: | SGP/EBBR - EBBR E25 Data Incorrect During J-panel Change |
DataStreams: | sgp15ebbrE25.b1, sgp30ebbrE25.b1, sgp5ebbrE25.b1 |
Description: | Some data was incorrect during the replacement of the J-panel #4. Affected data are: 5 minute: tair_top, tair_bot, thum_top, thum_bot, vp_top, vp_bot, tref 15 minute: rr_thum_r, rr_thum_l, rr_ts1, rr_ts2, rr_ts3, rr_ts4, rr_ts5, r_sm1, r_sm2, r_sm3, r_sm4, r_sm5, tair_r, tair_l, r_tref 30 minute: bowen, e, h, tair_top, tair_bot, thum_top, thum_bot, vp_top, vp_bot, sm1, sm2, sm3, sm4, sm5, shf1, shf2, shf3, shf4, shf5, c_shf1, c_shf2, c_shf3, c_shf4, c_shf5, ts1, ts2, ts3, ts4, ts5, cs1, cs2, cs3, cs4, cs5, ces1, ces2, ces3, ces4, ces5, g1, g2, g3, g4, g5, ave_shf, tref |
Measurements: | sgp30ebbrE25.b1:
sgp5ebbrE25.b1:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/16/1999 | 2015 | 06/17/1999 | 0500 |
Subject: | SGP/EBBR - EBBR E25 Thermocouple Temperatures Incorrect |
DataStreams: | sgp15ebbrE25.b1, sgp30ebbrE25.b1, sgp5ebbrE25.b1 |
Description: | Thermocouple temperature data was incorrect during the period, for unknown reasons. Affected data are: 5 minute: tair_top, tair_bot 15 minute: tair_r, tair_l 30 minute: bowen, e, h, tair_top, tair_bot |
Measurements: | sgp30ebbrE25.b1:
sgp5ebbrE25.b1:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/23/1999 | 0900 | 12/23/1999 | 1700 |
Subject: | SGP/EBBR - EBBR E7 Soil Moisture Jumping Around |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | All soil moisture sensor outputs jumped around during the period, causing the soil moisture values and the quantities calculated from them to be incorrect. The data values affected are shown below. 15 minute: r_sm1, r_sm2, r_sm3, r_sm4, r_sm5 30 minute: sm1, sm2, sm3, sm4, sm5, c_shf1, c_shf2, c_shf3, c_shf4, c_shf5, g1, g2, g3, g4, g5, ave_shf, e, h |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/21/1999 | 0230 | 10/30/1999 | 0330 |
Subject: | SGP/EBBR - EBBR E7 Soil Moisture Sensors #2 and #4 Negative |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | Soil moisture sensors #2 and #4 produced negative values for soil moisture frequently during the period, causing the quantities calculated from them to be incorrect. The version of the CR10X program at E7 does not have max/min limits on the outputs. The data needs to be examined to determine times of incorrect soil moisture values. The data values affected are shown below. 15 minute: r_sm1, r_sm2, r_sm3, r_sm4, r_sm5 30 minute: sm1, sm2, sm3, sm4, sm5, c_shf1, c_shf2, c_shf3, c_shf4, c_shf5, g1, g2, g3, g4, g5, ave_shf, e, h |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/04/1999 | 0530 | 09/10/1999 | 1330 |
Subject: | SGP/EBBR - EBBR E7 Soil Moisture Sensor #5 Negative |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | Soil moisture sensor #5 produced small negative values for soil moisture infrequently during the period. Because the vales were small negative, no calculated data values are adversely affected. The data needs to be examined to determine times of incorrect soil moisture values. The data values affected are shown below. 15 minute: r_sm5 30 minute: sm5 |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/26/2000 | 2315 | 02/01/2000 | 2050 |
Subject: | SGP/EBBR - EBBR E7 Left Aspirator Not Working |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | The left aspirator fan did not work during the period. The following values were therefore incorrect: 5 minute: tair_top, tair_bot, thum_top. thum_bot, hum_top, hum_bot, vp_top, vp_bot 15 minute: rr_thum_l, mv_hum_l, tair_l 30 minute: tair_top, tair_bot, thum_top. thum_bot, hum_top, hum_bot, vp_top, vp_bot, e, h, bowen |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/08/2000 | 0230 | 01/27/2000 | 1700 |
Subject: | SGP/EBBR/E26 - Right Aspirator Not Working |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | The right aspirator fan did not work during the period. The following values were therefore incorrect: 5 minute: tair_top, tair_bot, thum_top. thum_bot, hum_top, hum_bot, vp_top, vp_bot 15 minute: rr_thum_l, mv_hum_l, tair_l 30 minute: tair_top, tair_bot, thum_top. thum_bot, hum_top, hum_bot, vp_top, vp_bot, e, h, bowen |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/31/2000 | 2245 | 02/02/2000 | 1640 |
Subject: | SGP/EBBR - EBBR E19 Wind Speed Incorrect |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | The anemometer cups were frozen stalled during the period. The following values were therefore incorrect: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/04/1999 | 2215 | 08/10/1999 | 1500 |
Subject: | SGP/EBBR - EBBR E15 Anemometer Cup Off, Wind Speed Incorrect |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | An anemometer cup fell off. The following values were therefore incorrect: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/18/1999 | 0800 | 12/18/1999 | 1930 |
Subject: | SGP/EBBR - EBBR E9 Anemometer Frozen, Wind Speed Incorrect |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Description: | The anemometer cups were frozen stalled. The following values were therefore incorrect: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/25/1999 | 0635 | 10/07/1999 | 1620 |
Subject: | SGP/EBBR/E26 - Anemometer Not Working, Wind Speed Incorrect |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | The anemometer was not working during this period. The following values were therefore incorrect: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/21/1999 | 0000 | 12/21/1999 | 1320 |
Subject: | SGP/EBBR - EBBR E12 Anemometer Cups Frozen, Wind Speed Incorrect |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | The anemometer cups were frozen stalled during this period. The following values were therefore incorrect: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/10/2000 | 0200 | 01/10/2000 | 0200 |
Subject: | SGP/EBBR - EBBR E7 Soil Moisture Spike |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | All soil moisture sensor outputs are spiking up for a single half hour for this day, causing the soil moisture values and the quantities calculated from them to be incorrect. The spiked half hours are normally near sunrise. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/13/2000 | 1300 | 01/13/2000 | 1300 |
Subject: | SGP/EBBR - EBBR E7 Soil Moisture Spike |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | All soil moisture sensor outputs are spiking up for a single half hour for this day, causing the soil moisture values and the quantities calculated from them to be incorrect. The spiked half hours are normally near sunrise. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/14/2000 | 1330 | 01/14/2000 | 1330 |
Subject: | SGP/EBBR - EBBR E7 Soil Moisture Spike |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | All soil moisture sensor outputs are spiking up for a single half hour for this day, causing the soil moisture values and the quantities calculated from them to be incorrect. The spiked half hours are normally near sunrise. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/01/2000 | 1330 | 02/01/2000 | 1330 |
Subject: | SGP/EBBR - EBBR E7 Soil Moisture Spike |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | All soil moisture sensor outputs are spiking up for a single half hour for this day, causing the soil moisture values and the quantities calculated from them to be incorrect. The spiked half hours are normally near sunrise. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/02/2000 | 1330 | 02/02/2000 | 1330 |
Subject: | SGP/EBBR - EBBR E7 Soil Moisture Spike |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | All soil moisture sensor outputs are spiking up for a single half hour for this day, causing the soil moisture values and the quantities calculated from them to be incorrect. The spiked half hours are normally near sunrise. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/04/2000 | 1400 | 02/04/2000 | 1400 |
Subject: | SGP/EBBR - EBBR E7 Soil Moisture Spike |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | All soil moisture sensor outputs are spiking up for a single half hour for this day, causing the soil moisture values and the quantities calculated from them to be incorrect. The spiked half hours are normally near sunrise. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/05/2000 | 1330 | 02/05/2000 | 1330 |
Subject: | SGP/EBBR - EBBR E7 Soil Moisture Spike |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | All soil moisture sensor outputs are spiking up for a single half hour for this day, causing the soil moisture values and the quantities calculated from them to be incorrect. The spiked half hours are normally near sunrise. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/06/2000 | 1130 | 02/06/2000 | 1130 |
Subject: | SGP/EBBR - EBBR E7 Soil Moisture Spike |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | All soil moisture sensor outputs are spiking up for a single half hour for this day, causing the soil moisture values and the quantities calculated from them to be incorrect. The spiked half hours are normally near sunrise. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/24/1999 | 0100 | 12/24/1999 | 0100 |
Subject: | SGP/EBBR - EBBR E7 Soil Moisture Spike |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | All soil moisture sensor outputs are spiking up for a single half hour for this day causing the soil moisture values and the quantities calculated from them to be incorrect. The spiked half hours are normally near sunrise. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/22/1999 | 1330 | 12/22/1999 | 1330 |
Subject: | SGP/EBBR - EBBR E7 Soil Moisture Spike |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | All soil moisture sensor outputs are spiking up for a single half hour this day causing the soil moisture values and the quantities calculated from them to be incorrect. The spiked half hours are normally near sunrise. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/25/1999 | 1300 | 12/25/1999 | 1300 |
Subject: | SGP/EBBR - EBBR E7 Soil Moisture Spike |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | All soil moisture sensor outputs are spiking up for a single half hour for this day, causing the soil moisture values and the quantities calculated from them to be incorrect. The spiked half hours are normally near sunrise. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/01/2000 | 0230 | 01/01/2000 | 0230 |
Subject: | SGP/EBBR - EBBR E7 Soil Moisture Spike |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | All soil moisture sensor outputs are spiking up for a single half hour for this day, causing the soil moisture values and the quantities calculated from them to be incorrect. The spiked half hours are normally near sunrise. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/05/2000 | 1330 | 01/05/2000 | 1330 |
Subject: | SGP/EBBR - EBBR E7 Soil Moisture Spike |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | All soil moisture sensor outputs are spiking up for a single half hour for this day, causing the soil moisture values and the quantities calculated from them to be incorrect. The spiked half hours are normally near sunrise. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/27/1999 | 0615 | 08/12/1999 | 1600 |
Subject: | SGP/EBBR/26 - Low Battery Condition |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | When battery voltage was less than 10.4 V DC during this period, all data should be considered to be incorrect. The cause of the problem was a broken trace on J-panel #3. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/17/1999 | 1700 | 12/27/1999 | 1845 |
Subject: | SGP/EBBR - EBBR E12 Net Radiation Incorrect |
DataStreams: | sgp30ebbrE12.b1, sgp5ebbrE12.b1, sgp15ebbrE12.b1 |
Description: | Condensation inside the domes (possibly water in the bottom dome also) during the period made daytime net radiation values as much as 50% too large and nighttime net radiations as much as 100% too small. The affected data are: 5 minute: q 15 minute: mv_q 30 minute: q, h, e |
Measurements: | sgp30ebbrE12.b1:
sgp15ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/27/1999 | 1530 | 07/27/1999 | 1530 |
Subject: | SGP/EBBR- EBBR E15 Air Temperatures Incorrect |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | When the reference temperature sensor was being replaced, the left and right thermocouple air temperatures and humidity probe temperatures were incorrect. Affected data values were: 5 Minute: tref, tair_top, tair_bot, thum_top, thum_bot, vp_top, vp_bot 15 minute: rr_tref, rr_thum_l, rr_thum_r, tair_r, tair_l 30 minute: tref, tair_top, tair_bot, thum_top, thum_bot, vp_top, vp_bot, bowen, e, h |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/03/1999 | 0545 | 01/11/2000 | 2135 |
Subject: | SGP/EBBR - EBBR E18 Right Humidity Incorrect |
DataStreams: | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
Description: | After the 3 December 99 lightning strike, the humidity portion of the right T/RH probe drifted downwards at times, causing vapor pressure gradients to be incorrectly measured. Therefore, sensible and latent heat fluxes were also incorrect. Since much of the data from the period is missing (because of latent effects of the lightning strike), little data is actually affected. The affected data values were: 5 minute: hum_top, hum_bot, vp_top, vp_bot 15 minute: mv_hum_r 30 minute: hum_top, hum_bot, vp_top, vp_bot, bowen, h, e |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/03/1999 | 0535 | 12/03/1999 | 0600 |
Subject: | SGP/EBBR - EBBR E18 Data Incorrect After Lightning Strike |
DataStreams: | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
Description: | Immediately after the 3 December 99 lightning strike, all data from this EBBR system was incorrect. The system crashed at the end of this period. |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/03/1999 | 0535 | 02/14/2000 | 0915 |
Subject: | SGP/EBBR - EBBR E18 Much Data Missing After Lightning Strike |
DataStreams: | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
Description: | On 3 December 99 this EBBR system was struck by lightning. Much of the data from this EBBR system for the stated period is missing. |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/13/2000 | 2115 | 01/13/2000 | 2130 |
Subject: | SGP/EBBR - EBBR E13 Left Aspirator Not Working |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | A loss of the 12V DC power to the left aspirator caused it to stop operating. The affected measurements are: 5 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot 15 minute: rr_thum_r, rr_thum_l, mv_hum_r, mv_hum_l, tair_r, tair_l 30 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, bowen, h, e |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/05/2000 | 1330 | 01/05/2000 | 1330 |
Subject: | SGP/EBBR - EBBR E7 Soil Heat Flows Spiked |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | All soil heat flows were spiked during the period. The affected data are: 15 minute: mv_hft1, mv_hft2, mv_hft3, mv_hft4, mv_hft5 30 minute: shf1, shf2, shf3, shf4, shf5, c_shf1, c_shf2, c_shf3, c_shf4, c_shf5, g1, g2, g3, g4, g5, ave_shf, h, e |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/30/1999 | 0700 | 02/09/2000 | 1715 |
Subject: | SGP/EBBR - EBBR E20 Right Air Temperature Incorrect |
DataStreams: | sgp30ebbrE20.b1, sgp5ebbrE20.b1, sgp15ebbrE20.b1 |
Description: | The right air temperature (thermocouple) output was frequently offscale. Examination of the 15 minute data will clearly show which times are affected during the period. The affected data are: 5 minute: tair_top, tair_bot 15 minute: tair_r 30 minute: tair_top, tair_bot, bowen, h, e |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/17/1999 | 1500 | 10/30/1999 | 0330 |
Subject: | SGP/EBBR - EBBR E7 Soil Moistures #2 and #4 Incorrect |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | Soil moisture #2 and #4 were negative, often large negative, during very dry soil conditions, causing several other quantities to also be incorrect. The affected data are: 15 minute: r_sm2, r_sm4 30 minute: sm2, sm4, c_shf2, c_shf4, cs2, cs4, ces2, ces4, g2, g4, ave_shf, h, e |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/13/1999 | 1600 | 02/01/2000 | 1900 |
Subject: | SGP/EBBR - EBBR E8 Left Air Temperature (thermocouple) Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The left air temperature (thermocouple) frequently went offscale, for an unknown reason, thereby affecting many quantities. The affected data are: 5 minute: tair_bot, tair_top 15 minute: tair_l 30 minute: tair_bot, tair_top, bowen, h, e |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/09/1999 | 0700 | 12/14/1999 | 1500 |
Subject: | SGP/EBBR - EBBR E12 Net Radiation Incorrect |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | The net radiometer top dome was broken and water had entered. Net radiation values are therefore incorrect during the period. The affected data are: 5 minute: q 15 minute: mv_q 30 minute: q, h, e |
Measurements: | sgp30ebbrE12.b1:
sgp15ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/22/1999 | 1300 | 09/28/1999 | 1830 |
Subject: | SGP/EBBR - EBBR E8 Wind Direction Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | A malfunctioning wind direction sensor caused the wind direction measurement to be incorrect for the period. The affected data are: 5 minute: res_ws, wind_d 15 minute: mv_wind_d 30 minute: res_ws, wind_d |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/20/1999 | 0300 | 08/04/1999 | 1730 |
Subject: | SGP/EBBR - EBBR E22 Soil Moistures Negative |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1 |
Description: | Because of the lack of rain, soil moistures were negative during much of the period. All five soil moisture sensors had negative outputs by early on Aug 1. The negative soil moistures affected sensible and latent heat fluxes slightly. The affected data are: 15 minute: r_sm1, r_sm2, r_sm3, r_sm4, r_sm5 30 minute: sm1, sm2, sm3, sm4, sm5, c_shf1, c_shf2, c_shf3, c_shf4, c_shf5, cs1, cs2, cs3, cs4, cs5, ces1, ces2, ces3, ces4, ces5, g1, g2, g3, g4, g5, ave_shf, e, h |
Measurements: | sgp30ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/09/1999 | 0330 | 12/16/1999 | 1630 |
Subject: | SGP/EBBR - EBBR E19 Net Radiation Incorrect |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | The net radiometer top dome was broken and water had entered. Net radiation values are therefore incorrect during the period. The affected data are: 5 minute: q 15 minute: mv_q 30 minute: q, h, e |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/20/1999 | 1930 | 09/23/1999 | 1930 |
Subject: | SGP/EBBR - EBBR E13 Net Radiation Incorrect |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | The net radiometer top dome was broken and water had entered. Net radiation values are therefore incorrect during the period. The affected data are: 5 minute: q 15 minute: mv_q 30 minute: q, h, e |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/21/1999 | 1330 | 09/22/1999 | 0300 |
Subject: | SGP/EBBR - EBBR E18 Net Radiation Incorrect |
DataStreams: | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
Description: | The net radiometer top dome was broken and water had entered. Net radiation values are therefore incorrect during the period. The affected data are: 5 minute: q 15 minute: mv_q 30 minute: q, h, e |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/07/1999 | 1900 | 09/08/1999 | 2200 |
Subject: | SGP/EBBR - EBBR E25 Net Radiation Incorrect |
DataStreams: | sgp30ebbrE25.b1, sgp5ebbrE25.b1, sgp15ebbrE25.b1 |
Description: | The net radiometer top dome was broken and water had entered. Net radiation values are therefore incorrect during the period. The affected data are: 5 minute: q 15 minute: mv_q 30 minute: q, h, e |
Measurements: | sgp30ebbrE25.b1:
sgp5ebbrE25.b1:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/27/1999 | 0600 | 09/30/1999 | 1500 |
Subject: | SGP/EBBR - EBBR E2 Net Radiation Incorrect |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | The net radiometer top dome was broken and water had entered. Net radiation values are therefore incorrect during the period. The affected data are: 5 minute: q 15 minute: mv_q 30 minute: q, h, e |
Measurements: | sgp30ebbrE2.b1:
sgp15ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/27/1999 | 0930 | 09/28/1999 | 1700 |
Subject: | SGP/EBBR - EBBR E7 Net Radiation Incorrect |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | The net radiometer top dome was broken and water had entered. Net radiation values are therefore incorrect during the period. The affected data are: 5 minute: q 15 minute: mv_q 30 minute: q, h, e |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/28/1999 | 0230 | 09/30/1999 | 1500 |
Subject: | SGP/EBBR - EBBR E9 Net Radiation Incorrect |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Description: | The net radiometer top dome was broken and water had entered. Net radiation values are therefore incorrect during the period. The affected data are: 5 minute: q 15 minute: mv_q 30 minute: q, h, e |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/25/1999 | 0700 | 10/07/1999 | 1700 |
Subject: | SGP/EBBR/E26 - Net Radiation Incorrect |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | The net radiometer top dome was broken and water had entered. Net radiation values are therefore incorrect during the period. The affected data are: 5 minute: q 15 minute: mv_q 30 minute: q, h, e |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/15/1999 | 0800 | 10/30/1999 | 0300 |
Subject: | SGP/EBBR - EBBR E22 Soil Moistures Negative |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1 |
Description: | Because of the lack of rain, all soil moistures were negative during much of the period. The negative soil moistures caused soil heat storage and thus average soil heat flux to be smaller than they should have been. The sensible and latent heat fluxes were therefore reduced slightly (approximately 0-5%). The affected data are: 15 minute: r_sm1, r_sm2, r_sm3, r_sm4, r_sm5 30 minute: sm1, sm2, sm3, sm4, sm5, c_shf1, c_shf2, c_shf3, c_shf4, c_shf5, cs1, cs2, cs3, cs4, cs5, ces1, ces2, ces3, ces4, ces5, g1, g2, g3, g4, g5, ave_shf, e, h |
Measurements: | sgp30ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/23/2000 | 1935 | 02/23/2000 | 2000 |
Subject: | SGP/EBBR - EBBR E20 Left Air Temperature Incorrect |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | Left air temperature (thermocouple) was incorrect during part of the period, thereby affecting top and bottom air temperature. The affected data and times are: 5 minute: tair_top (1935 GMT) 15 minute: tair_l (1945 GMT) 30 minute: tair_top, bowen, e, h (2000 GMT) |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/04/1999 | 1800 | 08/04/1999 | 1800 |
Subject: | SGP/EBBR - EBBR E25 Soil Heat Flow #1 Incorrect |
DataStreams: | sgp15ebbrE25.b1, sgp30ebbrE25.b1 |
Description: | Soil heat flux #1 was large negative or offscale numerous times. |
Measurements: | sgp30ebbrE25.b1:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/05/1999 | 0100 | 08/05/1999 | 0200 |
Subject: | SGP/EBBR - EBBR E25 Soil Heat Flow #1 Incorrect |
DataStreams: | sgp15ebbrE25.b1, sgp30ebbrE25.b1 |
Description: | Soil heat flux #1 was large negative or offscale numerous times. |
Measurements: | sgp30ebbrE25.b1:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/06/1999 | 1830 | 08/06/1999 | 1830 |
Subject: | SGP/EBBR - EBBR E25 Soil Heat Flow #1 Incorrect |
DataStreams: | sgp15ebbrE25.b1, sgp30ebbrE25.b1 |
Description: | Soil heat flux #1 was large negative or offscale numerous times. |
Measurements: | sgp30ebbrE25.b1:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/07/1999 | 0130 | 08/07/1999 | 0130 |
Subject: | SGP/EBBR - EBBR E25 Soil Heat Flow #1 Incorrect |
DataStreams: | sgp15ebbrE25.b1, sgp30ebbrE25.b1 |
Description: | Soil heat flux #1 was large negative or offscale numerous times. |
Measurements: | sgp30ebbrE25.b1:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/17/1999 | 0130 | 08/17/1999 | 0130 |
Subject: | SGP/EBBR - EBBR E25 Soil Heat Flow #1 Incorrect |
DataStreams: | sgp15ebbrE25.b1, sgp30ebbrE25.b1 |
Description: | Soil heat flux #1 was large negative or offscale numerous times. |
Measurements: | sgp30ebbrE25.b1:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/18/1999 | 0200 | 08/18/1999 | 0400 |
Subject: | SGP/EBBR - EBBR E25 Soil Heat Flow #1 Incorrect |
DataStreams: | sgp15ebbrE25.b1, sgp30ebbrE25.b1 |
Description: | Soil heat flux #1 was large negative or offscale numerous times. |
Measurements: | sgp30ebbrE25.b1:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/18/1999 | 0930 | 08/18/1999 | 1030 |
Subject: | SGP/EBBR - EBBR E25 Soil Heat Flow #1 Incorrect |
DataStreams: | sgp15ebbrE25.b1, sgp30ebbrE25.b1 |
Description: | Soil heat flux #1 was large negative or offscale numerous times. |
Measurements: | sgp30ebbrE25.b1:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/15/1999 | 2000 | 10/15/1999 | 2330 |
Subject: | SGP/EBBR - EBBR E25 Soil Heat Flow #1 Incorrect |
DataStreams: | sgp15ebbrE25.b1, sgp30ebbrE25.b1 |
Description: | Soil heat flux #1 was large negative or offscale numerous times. |
Measurements: | sgp30ebbrE25.b1:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/26/1999 | 0000 | 10/26/1999 | 0030 |
Subject: | SGP/EBBR - EBBR E25 Soil Heat Flow #1 Incorrect |
DataStreams: | sgp15ebbrE25.b1, sgp30ebbrE25.b1 |
Description: | Soil heat flux #1 was large negative or offscale numerous times. |
Measurements: | sgp30ebbrE25.b1:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/26/1999 | 0100 | 07/26/1999 | 0530 |
Subject: | SGP/EBBR - EBBR E25 Soil Heat Flow #1 Incorrect |
DataStreams: | sgp15ebbrE25.b1, sgp30ebbrE25.b1 |
Description: | Soil heat flux #1 was large negative or offscale numerous times. |
Measurements: | sgp30ebbrE25.b1:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/03/1999 | 0100 | 08/03/1999 | 0430 |
Subject: | SGP/EBBR - EBBR E25 Soil Heat Flow #1 Incorrect |
DataStreams: | sgp30ebbrE25.b1, sgp15ebbrE25.b1 |
Description: | Soil heat flux #1 was large negative or offscale numerous times. |
Measurements: | sgp30ebbrE25.b1:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/02/1999 | 1730 | 08/16/1999 | 1800 |
Subject: | SGP/EBBR - EBBR E25 Soil Heat Flow #1 Incorrect |
DataStreams: | sgp15ebbrE25.b1, sgp30ebbrE25.b1 |
Description: | Soil heat flux #1 was large negative or offscale numerous times. |
Measurements: | sgp30ebbrE25.b1:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/02/1999 | 0100 | 08/02/1999 | 0230 |
Subject: | SGP/EBBR - EBBR E25 Soil Heat Flow #1 Incorrect |
DataStreams: | sgp15ebbrE25.b1, sgp30ebbrE25.b1 |
Description: | Soil heat flux #1 was large negative or offscale numerous times. |
Measurements: | sgp30ebbrE25.b1:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/03/1999 | 1800 | 08/03/1999 | 1800 |
Subject: | SGP/EBBR - EBBR E25 Soil Heat Flow #1 Incorrect |
DataStreams: | sgp15ebbrE25.b1, sgp30ebbrE25.b1 |
Description: | Soil heat flux #1 was large negative or offscale numerous times. |
Measurements: | sgp30ebbrE25.b1:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/04/1999 | 0100 | 08/04/1999 | 0200 |
Subject: | SGP/EBBR - EBBR E25 Soil Heat Flow #1 Incorrect |
DataStreams: | sgp15ebbrE25.b1, sgp30ebbrE25.b1 |
Description: | Soil heat flux #1 was large negative or offscale numerous times. |
Measurements: | sgp30ebbrE25.b1:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/17/1995 | 2200 | 07/24/1995 | 2300 |
Subject: | SGP/EBBR/E13 - failure of sensors |
DataStreams: | sgp30ebbrE13.b1, sgp5ebbrE13.b1, sgp15ebbrE13.b1, sgp5ebbrE13.00 |
Description: | A lightning strike to the central cluster area shortly after 0800 GMT on July 17 caused failure of several EBBR sensors: both humidity probes, the barometer, soil heat flux #3, soil moisture # 1 and the wind direction. These failed sensors caused sensible, latent, and average soil heat flux to be incorrect until site operations performed maintenance on July 24th. Site Operations found a loose connection from soil moisture probe #1, removed barometer SN 6578 and replaced it with SN 6507, removed soil heat flux sensor #3 with SN 923022 and replaced it with SN 933199, repositioned the wind direction sensor - which was rotated approximately ninety degrees clockwise from its proper position, and removed the right humidity probe SN 460412 and replaced it with SN 460428. |
Measurements: | sgp5ebbrE13.00:
sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/17/1995 | 2200 | 09/22/1995 | 2108 |
Subject: | SGP/EBBR/E13 - Failed Right Humidity Probe |
DataStreams: | sgp5ebbrE13.b1, sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.00 |
Description: | A lightning strike to the central cluster area on 17 July 1995 caused failure of several EBBR sensors including the right humidity probe. On July 24th, site operations removed the right humidity probe SN 460412 and replaced it with SN 460428 In less than a day the right humidity probe output soared to over 113% and has hovered around 112% since. I consider all right probe values to be incorrect due to this behavior. Without a properly functioning right humidity probe, all values of top and bottom relative humidity, top and bottom vapor pressure, Bowen ratio, and latent and sensible heat fluxes are incorrect. A working spare relative humidity probe was not available at that time. The probe was replaced with a working, calibrated probe on 22 September 1995 at about 2108 GMT. I have the examined the data since the day of the replacement and find that the new probe is working well. All E13 EBBR measurements are presently correct. The old probe S/N was 460428, the new S/N is 460413. |
Measurements: | sgp5ebbrE13.00:
sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/17/1996 | 0000 | 04/01/1999 | 1410 |
Subject: | SGP/EBBR/E26 - Reprocess Soil Moisture Off-Scale |
DataStreams: | sgp30ebbrE26.b1, sgp15ebbrE26.b1 |
Description: | During numerous periods within the days listed above, individual soil moisture probes indicate small positive or else negative values when soil moisture is high (greater than 35%). This situation has occured since the present EBBR unit was installed on 17 July 1996. The cause of the problem is inadequate range in the calibration curve equation. Cement soil is somewhat unique amongst the EBBR sites, as the soil is very sandy and porous; it can hold a larger amount of water per volume than the other soils. This high a soil moisture cannot be obtained with the calibration soil used by REBS. Therefore, the soil moisture equations need to be manipulated to allow larger soil moistures to be calculated. Much of the data from January 1998 is corrupted by this problem, making changes in the equations prudent. The soil moisture data must be inspected by the data user to see when soil moisture values for one or more probes is inconsistent with the others. Recalculation of ave_shf, e, and h must be made using only the soil moisture sets with correct soil moisture values. Affected data values are (# for probe number): 15 minute: rr_sm# 30 minute: sm#, c_shf#, cs#, ces#, g#, ave_shf, e, h Sensible, latent, and soil heat fluxes can be recalculated using the soil probe sets that provide correct data. For example, if soil moisture 2 is incorrect, ave_shf = (g1 + g3 + g4 + g5)/4 e = -(q + ave_shf)/(1 +bowen) h = -(e + ave_shf +q), If soil moisture 2 and 4 are incorrect, ave_shf = (g1 + g3 + g5)/3 e = -(q + ave_shf)/(1 +bowen) h = -(e + ave_shf +q), If soil moisture 2, 4, and 5 are incorrect, ave_shf = (g1 + g3)/2 e = -(q + ave_shf)/(1 +bowen) h = -(e + ave_shf +q), etc. On 5 May 1998 at 1815 GMT a calibration coefficient was changed in the CR10 program as a result of a BCR by the mentor. This coefficient has helped reduce the least severe problems but has not eliminated problems from large negative soil moistures measured even once during the integration period. However, now the incorrect average soil heat flow, latent heat flux, and sensible heat flux are so large, negative or positive, that they cannot be mistaken as being correct. The EBBR unit was replaced with a recalibrated system on 1 April 1999. The previous unit experienced soil moisture problems the entire time (almost 3 years) that the system was installed because of the inadequacy of the soil moisture sensor calibrations. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/28/2000 | 2045 | 05/04/2000 | 1500 |
Subject: | SGP/EBBR/E26 - EBBR E26 Left Aspirator Not Working |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | The data is suspect because of inadequate aspiration for the sensors in the left aspirated radiation shield. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/22/2000 | 1330 | 02/22/2000 | 1500 |
Subject: | SGP/EBBR/E26 - EBBR E26 All Data Incorrect |
DataStreams: | sgp5ebbrE26.b1, sgp30ebbrE26.b1, sgp15ebbrE26.b1 |
Description: | All data was incorrect for an unknown reason. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/22/2000 | 2300 | 02/23/2000 | 0030 |
Subject: | SGP/EBBR/E26 - All Data Incorrect |
DataStreams: | sgp5ebbrE26.b1, sgp30ebbrE26.b1, sgp15ebbrE26.b1 |
Description: | All data wqas incorrect for an unknown reason. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/06/2000 | 1630 | 04/06/2000 | 1700 |
Subject: | SGP/EBBR/E26 - All Data Incorrect |
DataStreams: | sgp30ebbrE26.b1, sgp15ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | Data was incorrect because of Preventative Maintenance activities. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/13/2000 | 1215 | 04/13/2000 | 1415 |
Subject: | SGP/EBBR/E26 - Low Battery Voltage Causes Incorrect Data |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | Battery voltage below about 10.5 volts DC caused a number of the EBBR sensors to not measure correctly. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/13/2000 | 1745 | 04/13/2000 | 1800 |
Subject: | SGP/EBBR/E26 - Low Battery Voltage Causes Incorrect Data |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | Battery voltage less than 10.5 volts DC causes some sensors to measure incorrectly. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/14/2000 | 0615 | 04/14/2000 | 1530 |
Subject: | SGP/EBBR/E26 - Low Battery Voltage Causes Incorrect Data |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | Battery voltage less than 10.5 volts DC causes some sensors to measure incorrectly. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/16/2000 | 0845 | 04/16/2000 | 1500 |
Subject: | SGP/EBBR/E26 - Low Battery Voltage Causes Incorrect Data |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | Battery voltage less than 10.5 volts DC causes some sensors to make incorrect measurements. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/24/2000 | 0000 | 03/24/2000 | 0030 |
Subject: | SGP/EBBR/E26 - All Data Incorrect |
DataStreams: | sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp15ebbrE26.b1 |
Description: | All data was incorrect for an unknown reason. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/15/2000 | 1030 | 03/15/2000 | 1430 |
Subject: | SGP/EBBR/E26 - All Data Incorrect |
DataStreams: | sgp5ebbrE26.b1, sgp30ebbrE26.b1, sgp15ebbrE26.b1 |
Description: | All data was incorrect for an unknown reason. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/26/2000 | 1530 | 04/20/2000 | 1450 |
Subject: | SGP/EBBR/E26 - Left Air Temperature Periodically Incorrect |
DataStreams: | sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp15ebbrE26.b1 |
Description: | Left air temperature periodically was incorrect, causing the indicated measurements to be incorrect. The user will need to inspect the data to see when the left air temperature was incorrect (compare left with right air temperature or with the humidity probe temperature). |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/07/2000 | 0900 | 04/20/2000 | 1450 |
Subject: | SGP/EBBR/E26 - Reference Temperature Incorrect |
DataStreams: | sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp15ebbrE26.b1 |
Description: | The reference temperature sensor did not work during the indicated period, causing the indicated measurements to be incorrect. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/24/2000 | 0100 | 04/06/2000 | 1730 |
Subject: | SGP/EBBR/E26 - Electronics Problem |
DataStreams: | sgp30ebbrE26.b1, sgp15ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | Incorrect home signal indicates an electronic problem that affected the indicated measurements; this may have been a multiplexer #2 problem as all sensors affected are connected to that multiplexer and no sensors on multiplexer #1 were affected. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/06/2000 | 1800 | 06/15/2000 | 1600 |
Subject: | SGP/EBBR/E26 - Soil Heat Flow #5 Incorrect |
DataStreams: | sgp30ebbrE26.b1, sgp15ebbrE26.b1 |
Description: | Soil Heat Flow #5 was incorrect, causing the indicated measurements to be incorrect. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/16/1999 | 0030 | 03/07/2000 | 1455 |
Subject: | SGP/EBBR/E26 - Soil Heat Flow #1 Incorrect |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1 |
Description: | Soil Heat Flow #1 was incorrect, causing the indicated measurements to be incorrect. A broken sensor wire was at fault. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/03/2000 | 0600 | 03/07/2000 | 1455 |
Subject: | SGP/EBBR/E26 - Soil Heat Flow #2 Large Negative |
DataStreams: | sgp30ebbrE26.b1, sgp15ebbrE26.b1 |
Description: | Soil Heat Flow #2 was large negative and incorrect for an unknown reason. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/25/2000 | 2145 | 02/26/2000 | 1130 |
Subject: | SGP/EBBR/E26 - Left Air Temperature Incorrect |
DataStreams: | sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp15ebbrE26.b1 |
Description: | Incorrect left air temperature caused the indicated measurements to be incorrect. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/17/1995 | 2200 | 07/25/1995 | 1930 |
Subject: | SGP/EBBR/E13- Left humidity probe replaced |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1, sgp5ebbrE13.00 |
Description: | A lightning strike to the central cluster area on 17 July 1995 caused the left humidity probe to fail. Some of the period of incorrect data has been discussed in D0000531.4. The left humidity probe was replaced on 25 July 1995 at 1925 GMT (old SN 460413, new SN 460435). |
Measurements: | sgp5ebbrE13.00:
sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/11/2000 | 0130 | 12/15/2000 | 1630 |
Subject: | SGP/EBBR/E12 - EBBR E12 Wind Speed Iced |
DataStreams: | sgp5ebbrE12.b1, sgp30ebbrE12.b1, sgp15ebbrE12.b1 |
Description: | The anemometer was frozen stopped by freezing rain; the indicated data is incorrect. |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/13/2000 | 0915 | 12/14/2000 | 1715 |
Subject: | SGP/EBBR/E25 - EBBR E25 Wind Speed Iced |
DataStreams: | sgp30ebbrE25.b1, sgp15ebbrE25.b1, sgp5ebbrE25.b1 |
Description: | The anemometer was frozen stopped by freezing rain. Indicated data was incorrect. |
Measurements: | sgp30ebbrE25.b1:
sgp5ebbrE25.b1:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/25/2000 | 1645 | 12/28/2000 | 2345 |
Subject: | SGP/EBBR/E25 - EBBR E25 Wind Speed Iced |
DataStreams: | sgp30ebbrE25.b1, sgp15ebbrE25.b1, sgp5ebbrE25.b1 |
Description: | The anemometer was frozen stopped by freezing rain. Indicated data was incorrect. The end date and time is a guess, based on conditions at surrounding sites (data was missing for a few days starting Dec. 27). |
Measurements: | sgp30ebbrE25.b1:
sgp5ebbrE25.b1:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/25/2000 | 0600 | 12/28/2000 | 1145 |
Subject: | SGP/EBBR/E26 - Wind Speed Iced |
DataStreams: | sgp5ebbrE26.b1, sgp30ebbrE26.b1, sgp15ebbrE26.b1 |
Description: | The anemometer was frozen stopped by freezing rain. Indicated data are incorrect. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/11/2000 | 0330 | 12/13/2000 | 0000 |
Subject: | SGP/EBBR/E2 - EBBR E2 Wind Speed Iced |
DataStreams: | sgp5ebbrE2.b1, sgp30ebbrE2.b1, sgp15ebbrE2.b1 |
Description: | The anemometer was frozen stopped by freezing rain. Indicated data was incorrect. |
Measurements: | sgp15ebbrE2.b1:
sgp30ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/10/2000 | 2215 | 12/15/2000 | 1715 |
Subject: | SGP/EBBR/E7 - EBBR E7 Wind Speed Iced |
DataStreams: | sgp5ebbrE7.b1, sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | The anemometer was frozen stopped by freezing rain. Indicated data is incorrect. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/18/2000 | 0730 | 12/18/2000 | 1115 |
Subject: | SGP/EBBR/E12 - EBBR E12 Wind Speed Iced |
DataStreams: | sgp30ebbrE12.b1, sgp15ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | The anemometer was frozen stopped by freezing rain; the indicated data is incorrect. |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/10/2000 | 2130 | 12/12/2000 | 1715 |
Subject: | SGP/EBBR/E15 - EBBR E15 Wind Speed Iced |
DataStreams: | sgp30ebbrE15.b1, sgp15ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | The anemometer was frozen stopped by freezing rain. Indicated data is incorrect. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/11/2000 | 0930 | 12/15/2000 | 0245 |
Subject: | SGP/EBBR/E18 - EBBR E18 Wind Speed Iced |
DataStreams: | sgp15ebbrE18.b1, sgp5ebbrE18.b1, sgp30ebbrE18.b1 |
Description: | The anemometer was frozen stopped by freezing rain. Indicated data is incorrect. |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/25/2000 | 2015 | 01/03/2001 | 1915 |
Subject: | SGP/EBBR/E18 - EBBR E18 Wind Speed Iced |
DataStreams: | sgp5ebbrE18.b1, sgp30ebbrE18.b1, sgp15ebbrE18.b1 |
Description: | The anemometer was frozen stopped by freezing rain. Indicated data is incorrect. |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/25/2000 | 1245 | 12/28/2000 | 1815 |
Subject: | SGP/EBBR/E19 - EBBR E19 Wind Speed Iced |
DataStreams: | sgp30ebbrE19.b1, sgp15ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | The anemometer was frozen stopped by freezing rain. Indicated data is incorrect. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/25/2000 | 1545 | 12/28/2000 | 2145 |
Subject: | SGP/EBBR/E20 - EBBR E20 Wind Speed Iced |
DataStreams: | sgp30ebbrE20.b1, sgp15ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | The anemometer was frozen stopped by freezing rain. Indicated data is incorrect. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/25/2000 | 1315 | 12/27/2000 | 2015 |
Subject: | SGP/EBBR/E22 - EBBR E22 Wind Speed Iced |
DataStreams: | sgp30ebbrE22.b1, sgp15ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | The anemometer was frozen stopped by freezing rain. Indicated data is incorrect. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/11/2000 | 0945 | 12/12/2000 | 1414 |
Subject: | SGP/EBBR/E25 - EBBR E25 Wind Speed Iced |
DataStreams: | sgp30ebbrE25.b1, sgp15ebbrE25.b1, sgp5ebbrE25.b1 |
Description: | The anemometer was frozen stopped by freezing rain. Indicated data was incorrect. |
Measurements: | sgp30ebbrE25.b1:
sgp5ebbrE25.b1:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/30/2000 | 1600 | 03/30/2000 | 1605 |
Subject: | SGP/EBBR/E2 - E2 EBBR Most Data Incorrect |
DataStreams: | sgp30ebbrE2.b1, sgp15ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | After logger replacement, most data was incorrect for the next timestamp. |
Measurements: | sgp15ebbrE2.b1:
sgp30ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/25/1998 | 1915 | 12/27/1998 | 1810 |
03/09/1998 | 1720 | 03/18/1998 | 1620 |
03/19/1998 | 1725 | 04/03/1998 | 2350 |
Subject: | SGP/EBBR/E2 - Missing data |
DataStreams: | sgp5ebbrE2.b1, sgp15ebbrE2.b1, sgp30ebbrE2.b1 |
Description: | Data are missing and unrecoverable. |
Measurements: | sgp30ebbrE2.b1:
sgp15ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/25/2000 | 1440 | 05/25/2000 | 1505 |
Subject: | SGP/EBBR/E2 - E2 EBBR Data Incorrect |
DataStreams: | sgp30ebbrE2.b1, sgp15ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | The data was incorrect for a period after logger replacement. |
Measurements: | sgp15ebbrE2.b1:
sgp30ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/30/2000 | 1635 | 04/27/2000 | 1605 |
Subject: | SGP/EBBR/E2 - E2 EBBR Ts4 and SM4 Bouncing Around |
DataStreams: | sgp30ebbrE2.b1, sgp15ebbrE2.b1 |
Description: | Ts4 and SM4 values bounced around in an unnatural way. The indicated data during the period indicated was incorrect. |
Measurements: | sgp15ebbrE2.b1:
sgp30ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/15/2000 | 0230 | 07/15/2000 | 1405 |
Subject: | SGP/EBBR/E2 - E2 EBBR Ts2 Incorrect |
DataStreams: | sgp30ebbrE2.b1, sgp15ebbrE2.b1 |
Description: | Ts2 data and quantities derived from it were incorrect for the indicated period. |
Measurements: | sgp15ebbrE2.b1:
sgp30ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/28/2000 | 1300 | 08/31/2000 | 1505 |
Subject: | SGP/EBBR/E2 - E2 EBBR Wind Direction Incorrect |
DataStreams: | sgp30ebbrE2.b1, sgp5ebbrE2.b1, sgp15ebbrE2.b1 |
Description: | The wind direction values were 135 degrees (an indication of an electronics failure in the sensor), 0 degrees, or 359 degrees during the indicated period, and were incorrect. |
Measurements: | sgp30ebbrE2.b1:
sgp15ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/18/2001 | 1700 | 01/18/2001 | 1705 |
Subject: | SGP/EBBR/E2 - E2 EBBR Data Incorrect |
DataStreams: | sgp30ebbrE2.b1, sgp15ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | Most of the data values were incorrect during the indicated time period, for an unknown reason. |
Measurements: | sgp15ebbrE2.b1:
sgp30ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/26/2000 | 1815 | 01/04/2001 | 1405 |
Subject: | SGP/EBBR/E2 - E2 EBBR Low Battery Condition |
DataStreams: | sgp30ebbrE2.b1, sgp15ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | All data should be considered to be incorrect when battery voltage is 10.4 volts or less. |
Measurements: | sgp15ebbrE2.b1:
sgp30ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/25/2000 | 2245 | 09/28/2000 | 1450 |
Subject: | SGP/EBBR/E2 - E2 EBBR Left Aspirator Fan Stopped |
DataStreams: | sgp30ebbrE2.b1, sgp5ebbrE2.b1, sgp15ebbrE2.b1 |
Description: | The indicated data was incorrect while the left aspirator fan was not operating. |
Measurements: | sgp30ebbrE2.b1:
sgp15ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/29/2000 | 0100 | 12/01/2000 | 0105 |
Subject: | SGP/EBBR/E2 - E2 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE2.b1, sgp5ebbrE2.b1, sgp15ebbrE2.b1 |
Description: | The indicated data was incorrect while the AEM was was not operating properly. |
Measurements: | sgp30ebbrE2.b1:
sgp15ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/02/2000 | 2330 | 12/03/2000 | 0635 |
Subject: | SGP/EBBR/E2 - E2 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE2.b1, sgp5ebbrE2.b1, sgp15ebbrE2.b1 |
Description: | The indicated data was incorrect while the AEM was was not operating properly. |
Measurements: | sgp30ebbrE2.b1:
sgp15ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/03/2000 | 1300 | 12/04/2000 | 1005 |
Subject: | SGP/EBBR/E2 - E2 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE2.b1, sgp5ebbrE2.b1, sgp15ebbrE2.b1 |
Description: | The indicated data was incorrect while the AEM was was not operating properly. |
Measurements: | sgp30ebbrE2.b1:
sgp15ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/05/2000 | 0500 | 12/07/2000 | 0005 |
Subject: | SGP/EBBR/E2 - E2 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE2.b1, sgp5ebbrE2.b1, sgp15ebbrE2.b1 |
Description: | The indicated data was incorrect while the AEM was was not operating properly. |
Measurements: | sgp30ebbrE2.b1:
sgp15ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/07/2000 | 0130 | 12/07/2000 | 0135 |
Subject: | SGP/EBBR/E2 - E2 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE2.b1, sgp5ebbrE2.b1, sgp15ebbrE2.b1 |
Description: | The indicated data was incorrect while the AEM was was not operating properly. |
Measurements: | sgp30ebbrE2.b1:
sgp15ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/07/2000 | 0930 | 12/07/2000 | 1535 |
Subject: | SGP/EBBR/E2 - E2 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE2.b1, sgp5ebbrE2.b1, sgp15ebbrE2.b1 |
Description: | The indicated data was incorrect while the AEM was was not operating properly. |
Measurements: | sgp30ebbrE2.b1:
sgp15ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/10/2000 | 1000 | 12/11/2000 | 1635 |
Subject: | SGP/EBBR/E2 - E2 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE2.b1, sgp5ebbrE2.b1, sgp15ebbrE2.b1 |
Description: | The indicated data was incorrect while the AEM was was not operating properly. |
Measurements: | sgp30ebbrE2.b1:
sgp15ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/12/2000 | 0600 | 12/31/2000 | 2250 |
Subject: | SGP/EBBR/E2 - E2 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE2.b1, sgp5ebbrE2.b1, sgp15ebbrE2.b1 |
Description: | The indicated data was incorrect while the AEM was was not operating properly. |
Measurements: | sgp30ebbrE2.b1:
sgp15ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/03/2000 | 1030 | 02/15/2000 | 1735 |
Subject: | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | When the AEM (automatic exchange mechanism) is stuck in one postion, half hour latent and sensible heat fluxes and the Bowen ratio are incorrect. Some of the raw EBBR measurements are also incorrect. I do not normally submit a DQR for this situation, as the qc checks for AEM home signals normally indicate the problem. However, the frequency of the problem at this site seems to warrant DQRs. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/20/2000 | 1530 | 03/28/2000 | 1805 |
Subject: | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The raw temperature and humidity data is incorrect and thus Bowen ratio, and half hour sensible and latent heat flux are incorrect. Normally I do not submit DQRs for this condition, but the frequency of the problem was high for this site. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/26/2000 | 2300 | 06/26/2000 | 2335 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/27/2000 | 0000 | 06/27/2000 | 0005 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/27/2000 | 1500 | 06/27/2000 | 1505 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/28/2000 | 1100 | 06/29/2000 | 1635 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/29/2000 | 1830 | 06/30/2000 | 2135 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/30/2000 | 2230 | 07/01/2000 | 1205 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/28/2000 | 2000 | 04/09/2000 | 1435 |
Subject: | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The raw temperature and humidity data is incorrect and thus Bowen ratio, and half hour sensible and latent heat flux are incorrect. Normally I do not submit DQRs for this condition, but the frequency of the problem was high for this site. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/09/2000 | 1600 | 04/11/2000 | 1735 |
Subject: | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The raw temperature and humidity data is incorrect and thus Bowen ratio, and half hour sensible and latent heat flux are incorrect. Normally I do not submit DQRs for this condition, but the frequency of the problem was high for this site. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/11/2000 | 1800 | 04/25/2000 | 1735 |
Subject: | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The raw temperature and humidity data is incorrect and thus Bowen ratio, and half hour sensible and latent heat flux are incorrect. Normally I do not submit DQRs for this condition, but the frequency of the problem was high for this site. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/25/2000 | 2330 | 05/13/2000 | 0235 |
Subject: | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The raw temperature and humidity data is incorrect and thus Bowen ratio, and half hour sensible and latent heat flux are incorrect. Normally I do not submit DQRs for this condition, but the frequency of the problem was high for this site. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/13/2000 | 0500 | 05/14/2000 | 0205 |
Subject: | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The raw temperature and humidity data is incorrect and thus Bowen ratio, and half hour sensible and latent heat flux are incorrect. Normally I do not submit DQRs for this condition, but the frequency of the problem was high for this site. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/14/2000 | 0330 | 05/14/2000 | 1035 |
Subject: | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The raw temperature and humidity data is incorrect and thus Bowen ratio, and half hour sensible and latent heat flux are incorrect. Normally I do not submit DQRs for this condition, but the frequency of the problem was high for this site. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/14/2000 | 1200 | 06/06/2000 | 1805 |
Subject: | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The raw temperature and humidity data is incorrect and thus Bowen ratio, and half hour sensible and latent heat flux are incorrect for much of this period. Please look at the home signal qc to determine whether the data is indeed incorrect. Normally I do not submit DQRs for this condition, but the frequency of the problem was high for this site. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/14/2000 | 0930 | 06/20/2000 | 1635 |
Subject: | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The raw temperature and humidity data is incorrect and thus Bowen ratio, and half hour sensible and latent heat flux are incorrect for this period. Normally I do not submit DQRs for this condition, but the frequency of the problem was high for this site. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/22/2000 | 0900 | 06/23/2000 | 1705 |
Subject: | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The raw temperature and humidity data is incorrect and thus Bowen ratio, and half hour sensible and latent heat flux are incorrect for this period. Normally I do not submit DQRs for this condition, but the frequency of the problem was high for this site. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/25/2000 | 0330 | 06/25/2000 | 1935 |
Subject: | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The raw temperature and humidity data is incorrect and thus Bowen ratio, and half hour sensible and latent heat flux are incorrect for this period. Normally I do not submit DQRs for this condition, but the frequency of the problem was high for this site. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/26/2000 | 1330 | 06/26/2000 | 1335 |
Subject: | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The raw temperature and humidity data is incorrect and thus Bowen ratio, and half hour sensible and latent heat flux are incorrect for this period. Normally I do not submit DQRs for this condition, but the frequency of the problem was high for this site. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/26/2000 | 1530 | 07/31/2000 | 1630 |
Subject: | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The raw temperature and humidity data is incorrect and thus Bowen ratio, and half hour sensible and latent heat flux are incorrect for this period. Normally I do not submit DQRs for this condition, but the frequency of the problem was high for this site. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/29/2000 | 0700 | 12/05/2000 | 1805 |
Subject: | SGP/EBBR/E8 - E8 EBBR Low Battery Condition |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | All indicated data fields should be considered to be incorrect when the battery voltage is less than 10.4 volts. Check the battery voltage to determine times of incorrect data. The battery was replaced on 5 Dec. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/01/2000 | 0800 | 11/20/2000 | 1750 |
Subject: | SGP/EBBR/E8 - E8 EBBR Wind Speed Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The indicated data fields are incorrect for the stated period because of failure of the anemometer electronics. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/27/2000 | 0030 | 03/27/2000 | 0105 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/07/2000 | 1600 | 04/08/2000 | 1705 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/10/2000 | 0530 | 04/10/2000 | 1405 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/10/2000 | 1800 | 04/10/2000 | 1805 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/10/2000 | 2030 | 04/10/2000 | 2035 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/17/2000 | 0130 | 02/21/2000 | 0035 |
Subject: | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The raw temperature and humidity data is incorrect and thus Bowen ratio, and half hour sensible and latent heat flux are incorrect. Normally I do not submit DQRs for this condition, but the frequency of the problem was high for this site. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/10/2000 | 2200 | 04/10/2000 | 2235 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/10/2000 | 2330 | 04/11/2000 | 1805 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/13/2000 | 0130 | 04/13/2000 | 0135 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/16/2000 | 0000 | 04/16/2000 | 0005 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/16/2000 | 0100 | 04/16/2000 | 0235 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/16/2000 | 0430 | 04/16/2000 | 0635 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/16/2000 | 0830 | 04/16/2000 | 2105 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/17/2000 | 1400 | 04/17/2000 | 1535 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/20/2000 | 0230 | 04/21/2000 | 0035 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/22/2000 | 0100 | 04/22/2000 | 0435 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/21/2000 | 1000 | 02/23/2000 | 1905 |
Subject: | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The raw temperature and humidity data is incorrect and thus Bowen ratio, and half hour sensible and latent heat flux are incorrect. Normally I do not submit DQRs for this condition, but the frequency of the problem was high for this site. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/22/2000 | 0600 | 04/22/2000 | 1335 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/23/2000 | 0030 | 04/24/2000 | 0005 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/25/2000 | 0630 | 04/25/2000 | 0635 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/25/2000 | 1400 | 04/25/2000 | 1905 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/25/2000 | 2000 | 04/25/2000 | 2305 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/26/2000 | 0430 | 04/26/2000 | 0505 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/09/2000 | 1530 | 05/09/2000 | 1535 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/09/2000 | 1700 | 05/09/2000 | 1735 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/10/2000 | 0230 | 05/10/2000 | 0435 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/12/2000 | 1530 | 05/12/2000 | 2305 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/24/2000 | 1530 | 02/25/2000 | 0405 |
Subject: | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The raw temperature and humidity data is incorrect and thus Bowen ratio, and half hour sensible and latent heat flux are incorrect. Normally I do not submit DQRs for this condition, but the frequency of the problem was high for this site. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/13/2000 | 0300 | 05/13/2000 | 0435 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/13/2000 | 1600 | 05/13/2000 | 1805 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/14/2000 | 0230 | 05/14/2000 | 0305 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/14/2000 | 1100 | 05/14/2000 | 1305 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/14/2000 | 1530 | 05/14/2000 | 1535 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/18/2000 | 1430 | 05/18/2000 | 1435 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/19/2000 | 1700 | 05/19/2000 | 1735 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/23/2000 | 1800 | 05/23/2000 | 1805 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/07/2000 | 0430 | 06/07/2000 | 0505 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/07/2000 | 1230 | 06/07/2000 | 1235 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/26/2000 | 0200 | 02/29/2000 | 1835 |
Subject: | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The raw temperature and humidity data is incorrect and thus Bowen ratio, and half hour sensible and latent heat flux are incorrect. Normally I do not submit DQRs for this condition, but the frequency of the problem was high for this site. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/08/2000 | 1030 | 06/08/2000 | 1105 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/09/2000 | 0000 | 06/09/2000 | 0005 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/09/2000 | 0100 | 06/09/2000 | 0105 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/10/2000 | 0000 | 06/10/2000 | 0205 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/10/2000 | 0330 | 06/10/2000 | 0335 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/10/2000 | 0500 | 06/10/2000 | 0605 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/10/2000 | 1230 | 06/10/2000 | 1635 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/10/2000 | 1730 | 06/10/2000 | 1805 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/10/2000 | 1900 | 06/10/2000 | 1905 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/12/2000 | 1500 | 06/12/2000 | 1505 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/29/2000 | 2000 | 03/19/2000 | 1735 |
Subject: | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The raw temperature and humidity data is incorrect and thus Bowen ratio, and half hour sensible and latent heat flux are incorrect. Normally I do not submit DQRs for this condition, but the frequency of the problem was high for this site. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/13/2000 | 1000 | 06/13/2000 | 1235 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/13/2000 | 1330 | 06/13/2000 | 1535 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/13/2000 | 1700 | 06/13/2000 | 1705 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/16/2000 | 0900 | 06/16/2000 | 0905 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/16/2000 | 1030 | 06/16/2000 | 1135 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/16/2000 | 1230 | 06/16/2000 | 1235 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/16/2000 | 1400 | 06/16/2000 | 1435 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/16/2000 | 1500 | 06/16/2000 | 1505 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/16/2000 | 1530 | 06/16/2000 | 1535 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/19/2000 | 2330 | 03/20/2000 | 0105 |
Subject: | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The raw temperature and humidity data is incorrect and thus Bowen ratio, and half hour sensible and latent heat flux are incorrect. Normally I do not submit DQRs for this condition, but the frequency of the problem was high for this site. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/16/2000 | 1630 | 06/16/2000 | 1735 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/16/2000 | 2200 | 06/16/2000 | 2205 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/17/2000 | 2030 | 06/17/2000 | 2035 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/21/2000 | 0100 | 06/21/2000 | 0105 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/21/2000 | 0430 | 06/21/2000 | 0505 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/21/2000 | 0730 | 06/21/2000 | 0805 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/21/2000 | 0900 | 06/21/2000 | 0905 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/21/2000 | 2230 | 06/21/2000 | 2305 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/24/2000 | 0000 | 06/24/2000 | 0005 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/20/2000 | 0230 | 03/20/2000 | 1105 |
Subject: | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The raw temperature and humidity data is incorrect and thus Bowen ratio, and half hour sensible and latent heat flux are incorrect. Normally I do not submit DQRs for this condition, but the frequency of the problem was high for this site. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/24/2000 | 0100 | 06/24/2000 | 0105 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/24/2000 | 0300 | 06/24/2000 | 0335 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/25/2000 | 1000 | 06/25/2000 | 1105 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/25/2000 | 1730 | 06/25/2000 | 1735 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/25/2000 | 1900 | 06/25/2000 | 1935 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/25/2000 | 2030 | 06/25/2000 | 2035 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/25/2000 | 2200 | 06/25/2000 | 2205 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/25/2000 | 2300 | 06/25/2000 | 2335 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/26/2000 | 0000 | 06/26/2000 | 0105 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/26/2000 | 1200 | 06/26/2000 | 2205 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/25/2000 | 0500 | 08/25/2000 | 0501 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/25/2000 | 1800 | 08/25/2000 | 1801 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/25/2000 | 1930 | 08/25/2000 | 1931 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/25/2000 | 2100 | 08/25/2000 | 2101 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/20/2000 | 2230 | 08/20/2000 | 2231 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/11/2000 | 0230 | 09/11/2000 | 0231 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/11/2000 | 1030 | 09/11/2000 | 1131 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/11/2000 | 1700 | 09/11/2000 | 1831 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/12/2000 | 0600 | 09/12/2000 | 1031 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/12/2000 | 0330 | 09/12/2000 | 0431 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/12/2000 | 1730 | 09/12/2000 | 1901 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/13/2000 | 0200 | 09/13/2000 | 0631 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/13/2000 | 1730 | 09/13/2000 | 1901 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/14/2000 | 0130 | 09/14/2000 | 0431 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/14/2000 | 1800 | 09/14/2000 | 2201 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/15/2000 | 0005 | 09/15/2000 | 0201 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/15/2000 | 1700 | 09/16/2000 | 0301 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/16/2000 | 1800 | 09/17/2000 | 0231 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/17/2000 | 1900 | 09/18/2000 | 0301 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/01/2000 | 1400 | 07/01/2000 | 1835 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/18/2000 | 1430 | 09/18/2000 | 1531 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/18/2000 | 1800 | 09/19/2000 | 0401 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/19/2000 | 1100 | 09/19/2000 | 1131 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/19/2000 | 1400 | 09/20/2000 | 0201 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/20/2000 | 1230 | 09/20/2000 | 2131 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/20/2000 | 2230 | 09/21/2000 | 0431 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/22/2000 | 1730 | 09/23/2000 | 0301 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/25/2000 | 2330 | 09/26/2000 | 0031 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/27/2000 | 1800 | 09/27/2000 | 1801 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/27/2000 | 2030 | 09/28/2000 | 0031 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/01/2000 | 1930 | 07/01/2000 | 2105 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/28/2000 | 1730 | 09/29/2000 | 0131 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/29/2000 | 0230 | 09/29/2000 | 0401 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/29/2000 | 1730 | 09/30/2000 | 0431 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/30/2000 | 1800 | 10/01/2000 | 0131 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/01/2000 | 1830 | 10/02/2000 | 0501 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/02/2000 | 1900 | 10/03/2000 | 0231 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/03/2000 | 1900 | 10/04/2000 | 0501 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/09/2000 | 2030 | 10/09/2000 | 2031 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/14/2000 | 2000 | 10/14/2000 | 2301 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/16/2000 | 1730 | 10/16/2000 | 2001 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/01/2000 | 2200 | 07/01/2000 | 2335 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/16/2000 | 2330 | 10/17/2000 | 0201 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/19/2000 | 0300 | 10/19/2000 | 0701 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/21/2000 | 0345 | 10/22/2000 | 1601 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/23/2000 | 1445 | 10/25/2000 | 1501 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/26/2000 | 2015 | 10/26/2000 | 2201 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/08/2000 | 2330 | 12/09/2000 | 0201 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/10/2000 | 0900 | 12/10/2000 | 2001 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/14/2000 | 2050 | 03/14/2000 | 2101 |
Subject: | SGP/EBBR/E7 - E7 EBBR Datalogger Replacement/Upgrade - Data Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | The datalogger was replaced with one that had been upgraded with a new operating system. Data just after the replacement was incorrect. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/07/1999 | 1430 | 12/07/1999 | 1431 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Temperature #1 Incorrect |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | Soil Temperature probe #1 was incorrect. A poor electronic connection was suspected. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/02/2000 | 0400 | 07/02/2000 | 0405 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/09/1999 | 1330 | 12/09/1999 | 1331 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Temperature #1 Incorrect |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | Soil Temperature probe #1 was incorrect. A poor electronic connection was suspected. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/12/1999 | 1330 | 12/12/1999 | 1331 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Temperature #1 Incorrect |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | Soil Temperature probe #1 was incorrect. A poor electronic connection was suspected. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/12/1999 | 1900 | 12/12/1999 | 1901 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Temperature #1 Incorrect |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | Soil Temperature probe #1 was incorrect. A poor electronic connection was suspected. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/04/1999 | 1400 | 08/04/1999 | 1431 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Temperature #1 Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil Temperature probe #1 was incorrect. A poor electronic connection was suspected. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/04/1999 | 1530 | 08/04/1999 | 1631 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Temperature #1 Incorrect |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | Soil Temperature probe #1 was incorrect. A poor electronic connection was suspected. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/04/1999 | 1800 | 08/04/1999 | 1801 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Temperature #1 Incorrect |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | Soil Temperature probe #1 was incorrect. A poor electronic connection was suspected. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/04/1999 | 2200 | 08/04/1999 | 2201 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Temperature #1 Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil Temperature probe #1 was incorrect. A poor electronic connection was suspected. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/04/1999 | 2330 | 08/05/1999 | 0031 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Temperature #1 Incorrect |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | Soil Temperature probe #1 was incorrect. A poor electronic connection was suspected. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/05/1999 | 0430 | 08/05/1999 | 0431 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Temperature #1 Incorrect |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | Soil Temperature probe #1 was incorrect. A poor electronic connection was suspected. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/05/1999 | 0830 | 08/05/1999 | 0831 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Temperature #1 Incorrect |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | Soil Temperature probe #1 was incorrect. A poor electronic connection was suspected. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/02/2000 | 0530 | 07/02/2000 | 0805 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/07/1999 | 0730 | 08/07/1999 | 0731 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Temperature #1 Incorrect |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | Soil Temperature probe #1 was incorrect. A poor electronic connection was suspected. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/01/2000 | 1600 | 03/01/2000 | 1631 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Temperature #1 Incorrect |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | Soil Temperature probe #1 was incorrect while it was being replaced. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/09/2000 | 1830 | 06/06/2000 | 1701 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Temperature #1 Incorrect |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | Soil Temperature probe #1 was incorrect. A poor electronic connection was suspected. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/01/2000 | 1530 | 03/01/2000 | 1701 |
Subject: | SGP/EBBR/E7 - E7 EBBR Right Aspirator Replacement |
DataStreams: | sgp5ebbrE7.b1, sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | A number of the basic measurements and measurements calculated from them were incorrect while the right AEM aspirator was replaced. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/04/1999 | 1245 | 08/15/1999 | 2001 |
Subject: | SGP/EBBR/E7 - E7 EBBR Left Aspirator Fan Not Working |
DataStreams: | sgp15ebbrE7.b1, sgp5ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | While the left aspirator fan was not working, the indicated measurements were incorrect. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/07/2000 | 1930 | 11/07/2000 | 2001 |
Subject: | SGP/EBBR/E7 - E7 EBBR Net Radiometer Dome Replacement |
DataStreams: | sgp5ebbrE7.b1, sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | The net radiation, sensible heat flux, and latent heat flux were incorrect while the net radiometer domes were being replaced. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/26/2000 | 2100 | 09/26/2000 | 2101 |
Subject: | SGP/EBBR/E7 - E7 EBBR All Data Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | All measurements were incorrect for an unknown reason. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/07/2000 | 2000 | 11/07/2000 | 2001 |
Subject: | SGP/EBBR/E7 - E7 EBBR All Data Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | All measurements were incorrect for an unknown reason. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/01/2000 | 1330 | 03/01/2000 | 1331 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Moisture Spikes |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | All soil moisture sensors spiked during the indicated period, causing incorrect data. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/10/2000 | 0100 | 03/10/2000 | 0101 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Moisture Spikes |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | All soil moisture sensors spiked during the indicated period, causing incorrect data. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/02/2000 | 1000 | 07/02/2000 | 1135 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/01/2000 | 1330 | 02/01/2000 | 1331 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Moisture Spikes |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | All soil moisture sensors spiked during the indicated period, causing incorrect data. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/04/2000 | 1400 | 02/04/2000 | 1401 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Moisture Spikes |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | All soil moisture sensors spiked during the indicated period, causing incorrect data. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/05/2000 | 1330 | 02/05/2000 | 1331 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Moisture Spikes |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | All soil moisture sensors spiked during the indicated period, causing incorrect data. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/06/2000 | 1130 | 02/06/2000 | 1131 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Moisture Spikes |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | All soil moisture sensors spiked during the indicated period, causing incorrect data. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/09/2000 | 1300 | 02/09/2000 | 1301 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Moisture Spikes |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | All soil moisture sensors spiked during the indicated period, causing incorrect data. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/13/2000 | 1930 | 09/26/2000 | 2031 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Temperature #4 Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil Temperature probe #4 failed, resulting in incorrect data. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/02/2000 | 1300 | 07/02/2000 | 1335 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/02/2000 | 1500 | 07/02/2000 | 1535 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/02/2000 | 1700 | 07/04/2000 | 0035 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/04/2000 | 0430 | 07/04/2000 | 0835 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/04/2000 | 0930 | 07/04/2000 | 1335 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/04/2000 | 1500 | 07/04/2000 | 1505 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/04/2000 | 1600 | 07/04/2000 | 1605 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/05/2000 | 2100 | 07/05/2000 | 2105 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/05/2000 | 2200 | 07/05/2000 | 2205 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/05/2000 | 2300 | 07/06/2000 | 0135 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/06/2000 | 0600 | 07/06/2000 | 0605 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/06/2000 | 1030 | 07/06/2000 | 1105 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/06/2000 | 1400 | 07/06/2000 | 2005 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/06/2000 | 2100 | 07/06/2000 | 2105 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/06/2000 | 2300 | 07/07/2000 | 0205 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/07/2000 | 1600 | 07/07/2000 | 1805 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/07/2000 | 1900 | 07/07/2000 | 1905 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/07/2000 | 2200 | 07/07/2000 | 2235 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/14/2000 | 0700 | 07/14/2000 | 0805 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/14/2000 | 1200 | 07/14/2000 | 1305 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/14/2000 | 2230 | 07/14/2000 | 2235 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/15/2000 | 0100 | 07/15/2000 | 0105 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/15/2000 | 0300 | 07/15/2000 | 1305 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/15/2000 | 1400 | 07/15/2000 | 1905 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/15/2000 | 2000 | 07/16/2000 | 0135 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/16/2000 | 0330 | 07/16/2000 | 0435 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/16/2000 | 0630 | 07/16/2000 | 0705 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/16/2000 | 0830 | 07/16/2000 | 0835 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/16/2000 | 1000 | 07/16/2000 | 1205 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/16/2000 | 1330 | 07/16/2000 | 1335 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/16/2000 | 1530 | 07/16/2000 | 1535 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/16/2000 | 1630 | 07/16/2000 | 1635 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/16/2000 | 1900 | 07/16/2000 | 1905 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/16/2000 | 2000 | 07/16/2000 | 2005 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/16/2000 | 2100 | 07/16/2000 | 2205 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/17/2000 | 0430 | 07/17/2000 | 0435 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/18/2000 | 0100 | 07/18/2000 | 0105 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/18/2000 | 0200 | 07/18/2000 | 0205 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/18/2000 | 0330 | 07/18/2000 | 0435 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/18/2000 | 0530 | 07/18/2000 | 0605 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/18/2000 | 0700 | 07/18/2000 | 0705 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/18/2000 | 1930 | 07/18/2000 | 1935 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/19/2000 | 2330 | 07/20/2000 | 0935 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/20/2000 | 1900 | 07/20/2000 | 1905 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/20/2000 | 2300 | 07/20/2000 | 2305 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/21/2000 | 0000 | 07/21/2000 | 0335 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/21/2000 | 0430 | 07/21/2000 | 0505 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/21/2000 | 0600 | 07/21/2000 | 0605 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/21/2000 | 0830 | 07/21/2000 | 0905 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/21/2000 | 1000 | 07/21/2000 | 1005 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/21/2000 | 1100 | 07/21/2000 | 1535 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/21/2000 | 1700 | 07/21/2000 | 1705 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/21/2000 | 1800 | 07/21/2000 | 1935 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/22/2000 | 0630 | 07/22/2000 | 0705 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/22/2000 | 1100 | 07/22/2000 | 1305 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/22/2000 | 2300 | 07/22/2000 | 2305 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/23/2000 | 0100 | 07/23/2000 | 0105 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/23/2000 | 0200 | 07/23/2000 | 0705 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/23/2000 | 0800 | 07/23/2000 | 0805 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/23/2000 | 0900 | 07/23/2000 | 2005 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/23/2000 | 2330 | 07/24/2000 | 0705 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/24/2000 | 1300 | 07/24/2000 | 1405 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/28/2000 | 0930 | 07/28/2000 | 0935 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/28/2000 | 1100 | 07/28/2000 | 1105 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/29/2000 | 0130 | 07/29/2000 | 0131 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/31/2000 | 1630 | 07/31/2000 | 1656 |
Subject: | SGP/EBBR/E8 - E8 EBBR Data Incorrect During System Replacement |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The data was incorrect while preparations were made to replace it with a recalibrated unit. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/31/2000 | 2030 | 07/31/2000 | 2031 |
Subject: | SGP/EBBR/E8 - E8 EBBR Data Not Based On 30 Minute Average |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The 30 minute value was based on only 15 minutes of data. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/10/2000 | 2330 | 12/11/2000 | 0731 |
Subject: | SGP/EBBR/E8 - E8 EBBR Wind Direction Frozen |
DataStreams: | sgp30ebbrE8.b1, sgp15ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The wind direction sensor was frozen into reading one direction during the indicated time period. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/05/2000 | 0130 | 08/05/2000 | 0131 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/12/2000 | 0400 | 08/12/2000 | 0501 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/13/2000 | 0230 | 08/13/2000 | 0231 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/16/2000 | 0530 | 08/16/2000 | 0731 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/16/2000 | 1730 | 08/16/2000 | 1801 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/16/2000 | 2100 | 08/16/2000 | 2231 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/17/2000 | 0530 | 08/17/2000 | 0531 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/17/2000 | 0700 | 08/17/2000 | 0801 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/17/2000 | 1430 | 08/17/2000 | 1431 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/17/2000 | 1800 | 08/17/2000 | 1901 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/24/2000 | 1830 | 08/24/2000 | 1931 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/25/2000 | 0100 | 08/25/2000 | 0331 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/12/2000 | 0300 | 12/12/2000 | 1531 |
Subject: | SGP/EBBR/E7 - E7 EBBR AEM Frozen Stopped |
DataStreams: | sgp5ebbrE7.b1, sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | Icing caused the Automatic Exchange Mechanism to stop. The indicated measurements are therefore incorrect. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/13/2000 | 0645 | 12/15/2000 | 1931 |
Subject: | SGP/EBBR/E7 - E7 EBBR AEM Frozen Stopped |
DataStreams: | sgp5ebbrE7.b1, sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | Icing caused the Automatic Exchange Mechanism to stop. The indicated measurements are therefore incorrect. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/17/2000 | 0000 | 12/17/2000 | 1401 |
Subject: | SGP/EBBR/E7 - E7 EBBR AEM Frozen Stopped |
DataStreams: | sgp5ebbrE7.b1, sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | Icing caused the Automatic Exchange Mechanism to stop. The indicated measurements are therefore incorrect. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/22/2000 | 0200 | 12/22/2000 | 1001 |
Subject: | SGP/EBBR/E7 - E7 EBBR AEM Frozen Stopped |
DataStreams: | sgp5ebbrE7.b1, sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | Icing caused the Automatic Exchange Mechanism to stop. The indicated measurements are therefore incorrect. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/25/2000 | 2330 | 12/27/2000 | 1931 |
Subject: | SGP/EBBR/E7 - E7 EBBR AEM Frozen Stopped |
DataStreams: | sgp5ebbrE7.b1, sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | Icing caused the Automatic Exchange Mechanism to stop. The indicated measurements are therefore incorrect. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/27/2000 | 2200 | 12/28/2000 | 1831 |
Subject: | SGP/EBBR/E7 - E7 EBBR AEM Frozen Stopped |
DataStreams: | sgp5ebbrE7.b1, sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | Icing caused the Automatic Exchange Mechanism to stop. The indicated measurements are therefore incorrect. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/30/2000 | 0800 | 12/30/2000 | 1031 |
Subject: | SGP/EBBR/E7 - E7 EBBR AEM Frozen Stopped |
DataStreams: | sgp5ebbrE7.b1, sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | Icing caused the Automatic Exchange Mechanism to stop. The indicated measurements are therefore incorrect. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/29/2000 | 0830 | 04/29/2000 | 1431 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Moisture # 2 Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil Moisture probe #2 was not capable of measuring very low soil moistures in the dry conditions present. The small positive, negative, and offscale measurements are incorrect. Several other measurements calculated from soil moisture are incorrect also, as a result. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/30/2000 | 0130 | 05/01/2000 | 0001 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Moisture # 2 Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil Moisture probe #2 was not capable of measuring very low soil moistures in the dry conditions present. The small positive, negative, and offscale measurements are incorrect. Several other measurements calculated from soil moisture are incorrect also, as a result. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/20/2000 | 0930 | 07/20/2000 | 1101 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Moisture # 2 Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil Moisture probe #2 was not capable of measuring very low soil moistures in the dry conditions present. The small positive, negative, and offscale measurements are incorrect. Several other measurements calculated from soil moisture are incorrect also, as a result. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/10/2000 | 0730 | 08/10/2000 | 1601 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Moisture # 2 Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil Moisture probe #2 was not capable of measuring very low soil moistures in the dry conditions present. The small positive, negative, and offscale measurements are incorrect. Several other measurements calculated from soil moisture are incorrect also, as a result. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/11/2000 | 0230 | 09/23/2000 | 0501 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Moisture # 2 Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil Moisture probe #2 was not capable of measuring very low soil moistures in the dry conditions present. The small positive, negative, and offscale measurements are incorrect. Several other measurements calculated from soil moisture are incorrect also, as a result. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/12/2000 | 0530 | 12/13/2000 | 2301 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Moisture # 2 Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil Moisture probe #2 was not capable of measuring very low soil moistures in the dry conditions present. The small positive, negative, and offscale measurements are incorrect. Several other measurements calculated from soil moisture are incorrect also, as a result. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/01/2000 | 1530 | 05/01/2000 | 1531 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Moistures 2, 4, and 5 Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil Moisture Probes 2, 4, and 5 were not capable of measuring very low soil moistures in the dry conditions present. The offscale measurements are incorrect. Several other measurements calculated from soil moisture are incorrect also as a result. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/13/2000 | 0400 | 07/20/2000 | 1201 |
Subject: | SGP/EBBR E7 - E7 EBBR Soil Moisture #4 Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil moisture probe #4 was not capable of measuring very low soil moistures in the dry conditions present. The small positive, negative, and offscale measurements are incorrect. Several other measurements calculated from soil moisture are incorrect also as a result. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/30/2000 | 2335 | 04/30/2000 | 2356 |
Subject: | SGP/EBBR E7 - E7 EBBR Soil Moisture #4 Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil moisture probe #4 was not capable of measuring very low soil moistures in the dry conditions present. The small positive, negative, and offscale measurements are incorrect. Several other measurements calculated from soil moisture are incorrect also as a result. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/30/2000 | 1800 | 04/30/2000 | 1826 |
Subject: | SGP/EBBR E7 - E7 EBBR Soil Moisture #4 Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil moisture probe #4 was not capable of measuring very low soil moistures in the dry conditions present. The small positive, negative, and offscale measurements are incorrect. Several other measurements calculated from soil moisture are incorrect also as a result. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/06/2000 | 0030 | 09/23/2000 | 0501 |
Subject: | SGP/EBBR E7 - E7 EBBR Soil Moisture #4 Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil moisture probe #4 was not capable of measuring very low soil moistures in the dry conditions present. The small positive, negative, and offscale measurements are incorrect. Several other measurements calculated from soil moisture are incorrect also as a result. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/16/2000 | 1100 | 08/16/2000 | 1431 |
Subject: | SGP/EBBR/E7 - E7 EBBR soil Moisture #5 Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil moisture probe #5 was not capable of measuring very low soil moistures in the dry conditions present. The small positive, negative, and offscale measurements are incorrect. Several other measurements calculated from soil moisture are also incorrect as a result. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/17/2000 | 0530 | 08/17/2000 | 1601 |
Subject: | SGP/EBBR/E7 - E7 EBBR soil Moisture #5 Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil moisture probe #5 was not capable of measuring very low soil moistures in the dry conditions present. The small positive, negative, and offscale measurements are incorrect. Several other measurements calculated from soil moisture are also incorrect as a result. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/18/2000 | 0230 | 08/18/2000 | 2001 |
Subject: | SGP/EBBR/E7 - E7 EBBR soil Moisture #5 Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil moisture probe #5 was not capable of measuring very low soil moistures in the dry conditions present. The small positive, negative, and offscale measurements are incorrect. Several other measurements calculated from soil moisture are also incorrect as a result. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/18/2000 | 2200 | 09/23/2000 | 0931 |
Subject: | SGP/EBBR/E7 - E7 EBBR soil Moisture #5 Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil moisture probe #5 was not capable of measuring very low soil moistures in the dry conditions present. The small positive, negative, and offscale measurements are incorrect. Several other measurements calculated from soil moisture are also incorrect as a result. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/08/2000 | 0730 | 10/08/2000 | 1531 |
Subject: | SGP/EBBR/E7 - E7 EBBR soil Moisture #5 Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil moisture probe #5 was not capable of measuring very low soil moistures in the dry conditions present. The small positive, negative, and offscale measurements are incorrect. Several other measurements calculated from soil moisture are also incorrect as a result. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/09/2000 | 0700 | 10/09/2000 | 1531 |
Subject: | SGP/EBBR/E7 - E7 EBBR soil Moisture #5 Incorrect |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | Soil moisture probe #5 was not capable of measuring very low soil moistures in the dry conditions present. The small positive, negative, and offscale measurements are incorrect. Several other measurements calculated from soil moisture are also incorrect as a result. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/10/2000 | 0900 | 10/10/2000 | 1501 |
Subject: | SGP/EBBR/E7 - E7 EBBR soil Moisture #5 Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil moisture probe #5 was not capable of measuring very low soil moistures in the dry conditions present. The small positive, negative, and offscale measurements are incorrect. Several other measurements calculated from soil moisture are also incorrect as a result. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/28/2000 | 1030 | 08/28/2000 | 1501 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Moisture #3 Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil moisture probe #3 was not capable of measuring very low soil moistures in the dry conditions that were present. The small positive, negative, and offscale measurements are incorrect. Several other measurements calculated from soil moisture are incorrect also as a result. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/29/2000 | 0830 | 08/29/2000 | 1531 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Moisture #3 Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil moisture probe #3 was not capable of measuring very low soil moistures in the dry conditions that were present. The small positive, negative, and offscale measurements are incorrect. Several other measurements calculated from soil moisture are incorrect also as a result. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/30/2000 | 0530 | 08/30/2000 | 1631 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Moisture #3 Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil moisture probe #3 was not capable of measuring very low soil moistures in the dry conditions that were present. The small positive, negative, and offscale measurements are incorrect. Several other measurements calculated from soil moisture are incorrect also as a result. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/31/2000 | 0500 | 08/31/2000 | 1701 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Moisture #3 Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil moisture probe #3 was not capable of measuring very low soil moistures in the dry conditions that were present. The small positive, negative, and offscale measurements are incorrect. Several other measurements calculated from soil moisture are incorrect also as a result. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/01/2000 | 0330 | 09/01/2000 | 1731 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Moisture #3 Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil moisture probe #3 was not capable of measuring very low soil moistures in the dry conditions that were present. The small positive, negative, and offscale measurements are incorrect. Several other measurements calculated from soil moisture are incorrect also as a result. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/02/2000 | 0330 | 09/02/2000 | 1731 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Moisture #3 Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil moisture probe #3 was not capable of measuring very low soil moistures in the dry conditions that were present. The small positive, negative, and offscale measurements are incorrect. Several other measurements calculated from soil moisture are incorrect also as a result. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/03/2000 | 0330 | 09/03/2000 | 1801 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Moisture #3 Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil moisture probe #3 was not capable of measuring very low soil moistures in the dry conditions that were present. The small positive, negative, and offscale measurements are incorrect. Several other measurements calculated from soil moisture are incorrect also as a result. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/04/2000 | 0130 | 09/04/2000 | 1901 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Moisture #3 Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil moisture probe #3 was not capable of measuring very low soil moistures in the dry conditions that were present. The small positive, negative, and offscale measurements are incorrect. Several other measurements calculated from soil moisture are incorrect also as a result. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/05/2000 | 0430 | 09/05/2000 | 1531 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Moisture #3 Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil moisture probe #3 was not capable of measuring very low soil moistures in the dry conditions that were present. The small positive, negative, and offscale measurements are incorrect. Several other measurements calculated from soil moisture are incorrect also as a result. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/06/2000 | 1030 | 09/06/2000 | 1501 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Moisture #3 Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil moisture probe #3 was not capable of measuring very low soil moistures in the dry conditions that were present. The small positive, negative, and offscale measurements are incorrect. Several other measurements calculated from soil moisture are incorrect also as a result. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/10/2000 | 1100 | 09/10/2000 | 1531 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Moisture #3 Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil moisture probe #3 was not capable of measuring very low soil moistures in the dry conditions that were present. The small positive, negative, and offscale measurements are incorrect. Several other measurements calculated from soil moisture are incorrect also as a result. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/11/2000 | 0900 | 09/11/2000 | 1601 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Moisture #3 Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil moisture probe #3 was not capable of measuring very low soil moistures in the dry conditions that were present. The small positive, negative, and offscale measurements are incorrect. Several other measurements calculated from soil moisture are incorrect also as a result. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/12/2000 | 0630 | 09/12/2000 | 1831 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Moisture #3 Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil moisture probe #3 was not capable of measuring very low soil moistures in the dry conditions that were present. The small positive, negative, and offscale measurements are incorrect. Several other measurements calculated from soil moisture are incorrect also as a result. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/13/2000 | 0200 | 09/13/2000 | 1701 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Moisture #3 Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil moisture probe #3 was not capable of measuring very low soil moistures in the dry conditions that were present. The small positive, negative, and offscale measurements are incorrect. Several other measurements calculated from soil moisture are incorrect also as a result. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/14/2000 | 1130 | 09/14/2000 | 1401 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Moisture #3 Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil moisture probe #3 was not capable of measuring very low soil moistures in the dry conditions that were present. The small positive, negative, and offscale measurements are incorrect. Several other measurements calculated from soil moisture are incorrect also as a result. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/15/2000 | 0800 | 09/15/2000 | 1631 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Moisture #3 Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil moisture probe #3 was not capable of measuring very low soil moistures in the dry conditions that were present. The small positive, negative, and offscale measurements are incorrect. Several other measurements calculated from soil moisture are incorrect also as a result. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/16/2000 | 0430 | 09/16/2000 | 1826 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Moisture #3 Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil moisture probe #3 was not capable of measuring very low soil moistures in the dry conditions that were present. The small positive, negative, and offscale measurements are incorrect. Several other measurements calculated from soil moisture are incorrect also as a result. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/17/2000 | 0230 | 09/17/2000 | 1831 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Moisture #3 Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil moisture probe #3 was not capable of measuring very low soil moistures in the dry conditions that were present. The small positive, negative, and offscale measurements are incorrect. Several other measurements calculated from soil moisture are incorrect also as a result. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/18/2000 | 0100 | 09/18/2000 | 2031 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Moisture #3 Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil moisture probe #3 was not capable of measuring very low soil moistures in the dry conditions that were present. The small positive, negative, and offscale measurements are incorrect. Several other measurements calculated from soil moisture are incorrect also as a result. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/18/2000 | 2230 | 09/23/2000 | 0431 |
Subject: | SGP/EBBR/E7 - E7 EBBR Soil Moisture #3 Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil moisture probe #3 was not capable of measuring very low soil moistures in the dry conditions that were present. The small positive, negative, and offscale measurements are incorrect. Several other measurements calculated from soil moisture are incorrect also as a result. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/11/2000 | 0000 | 12/11/2000 | 1331 |
Subject: | SGP/EBBR/E9 - E9 EBBR Wind Direction Incorrect |
DataStreams: | sgp5ebbrE9.b1, sgp30ebbrE9.b1, sgp15ebbrE9.b1 |
Description: | The wind direction vane was frozen in one direction by icing during the indicated period. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/18/2000 | 0645 | 12/18/2000 | 1101 |
Subject: | SGP/EBBR/E12 - E12 EBBR Wind Direction Incorrect |
DataStreams: | sgp30ebbrE12.b1, sgp5ebbrE12.b1, sgp15ebbrE12.b1 |
Description: | The wind direction vane was frozen stopped in one direction by icing. |
Measurements: | sgp30ebbrE12.b1:
sgp15ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/18/2000 | 0645 | 12/18/2000 | 1101 |
Subject: | SGP/EBBR/E12 - E12 EBBR Wind Speed Incorrect |
DataStreams: | sgp30ebbrE12.b1, sgp15ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | The wind speed sensor cups were frozen stopped by icing. |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/01/2000 | 0000 | 02/08/2000 | 1531 |
Subject: | SGP/EBBR/E12 - E12 EBBR AEM Stuck - Data Incorrect |
DataStreams: | sgp30ebbrE12.b1, sgp15ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | The AEM became stuck between levels, causing all gradient measurements to be incorrect, and thus the latent and sensible heat fluxes and Bowen ratio. |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/10/2000 | 2015 | 12/15/2000 | 1631 |
Subject: | SGP/EBBR/E9 - E9 EBBR Wind Speed Incorrect |
DataStreams: | sgp30ebbrE9.b1, sgp5ebbrE9.b1, sgp15ebbrE9.b1 |
Description: | The wind speed sensor cups were frozen in place by icing. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/18/2000 | 1530 | 02/13/2001 | 1546 |
Subject: | SGP/EBBR/E9 - E9 EBBR Soil Heat Flow 1 Incorrect |
DataStreams: | sgp30ebbrE9.b1, sgp15ebbrE9.b1 |
Description: | Soil Heat flow sensor #1 produced measurements that were large negative during the indicated period. This was the result of sensor failure. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/13/2000 | 1600 | 03/13/2000 | 1601 |
Subject: | SGP/EBBR/E9 - E9 EBBR Soil Heat Capacities Incorrect |
DataStreams: | sgp30ebbrE9.b1 |
Description: | Soil Heat Capacities were out-of-range during this period, as reported by the datalogger. The cause of this is unknown. Actual values used in the computations of fluxes were okay, as the fluxes were not affected. |
Measurements: | sgp30ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/04/2001 | 1730 | 01/04/2001 | 1731 |
Subject: | SGP/EBBR/E13 - E13 EBBR Flux Data Incorrect |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
Description: | The Automatic Exchange Mechanism was stuck inbetween the top and bottom positions. Incorrect Bowen ratio, and latent and sensible heat fluxes resulted. |
Measurements: | sgp15ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/16/2000 | 2130 | 03/19/2000 | 1701 |
Subject: | SGP/EBBR/E13 - E13 EBBR All Data Incorrect |
DataStreams: | sgp5ebbrE13.b1, sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
Description: | Low battery voltage caused all data to be incorrect. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/06/2000 | 0130 | 11/06/2000 | 0431 |
Subject: | SGP/EBBR/E13 - E13 EBBR All Data Incorrect |
DataStreams: | sgp5ebbrE13.b1, sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
Description: | Low battery voltage caused all data to be incorrect. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/06/2000 | 0120 | 11/06/2000 | 2255 |
Subject: | SGP/EBBR/E13 - E13 EBBR All Data Incorrect |
DataStreams: | sgp5ebbrE13.b1, sgp30ebbrE13.b1, sgp15ebbrE13.b1 |
Description: | All data was incorrect for an unknown reason. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/12/2000 | 0945 | 11/12/2000 | 1116 |
Subject: | SGP/EBBR/E13 - E13 EBBR All Data Incorrect |
DataStreams: | sgp5ebbrE13.b1, sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
Description: | Battery voltage was too low. All data was incorrect. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/24/2000 | 2230 | 02/24/2000 | 2231 |
Subject: | SGP/EBBR/E13 - E13 EBBR Heat Capacities Incorrect |
DataStreams: | sgp30ebbrE13.b1 |
Description: | The soil heat capacities were incorrect, a result of the missing data from the previous half hour. |
Measurements: | sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/13/2000 | 0015 | 11/14/2000 | 0031 |
Subject: | SGP/EBBR/E13 - E13 EBBR All Data Incorrect |
DataStreams: | sgp5ebbrE13.b1, sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
Description: | Battery voltage was too low. All data was incorrect. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/28/2000 | 1230 | 02/28/2000 | 1231 |
Subject: | SGP/EBBR/E13 - E13 EBBR All Data Incorrect |
DataStreams: | sgp5ebbrE13.b1, sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
Description: | Battery voltage was too low. All data was incorrect. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/29/2000 | 1300 | 02/29/2000 | 1301 |
Subject: | SGP/EBBR/E13 - E13 EBBR All Data Incorrect |
DataStreams: | sgp5ebbrE13.b1, sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
Description: | Battery voltage was too low. All data were incorrect. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/01/2000 | 0000 | 02/23/2000 | 1431 |
Subject: | SGP/EBBR/E13 - E13 EBBR Flux Data Incorrect |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
Description: | The Automatic Exchange Mechanism was stuck inbetween the top and bottom positions. Incorrect Bowen ratio, and latent and sensible heat fluxes resulted. |
Measurements: | sgp15ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/12/2000 | 1300 | 03/12/2000 | 1301 |
Subject: | SGP/EBBR/E13 - E13 EBBR All Data Incorrect |
DataStreams: | sgp5ebbrE13.b1, sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
Description: | Battery voltage was too low. All data was incorrect. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/11/2000 | 1300 | 03/11/2000 | 1301 |
Subject: | SGP/EBBR/E13 - E13 EBBR All Data Incorrect |
DataStreams: | sgp5ebbrE13.b1, sgp30ebbrE13.b1, sgp15ebbrE13.b1 |
Description: | Battery voltage was too low. All data was incorrect. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/08/2000 | 1300 | 03/08/2000 | 1301 |
Subject: | SGP/EBBR/E13 - E13 EBBR All Data Incorrect |
DataStreams: | sgp5ebbrE13.b1, sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
Description: | Battery voltage was too low. All data was incorrect. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/05/2000 | 1300 | 03/05/2000 | 1301 |
Subject: | SGP/EBBR/E13 - E13 EBBR All Data Incorrect |
DataStreams: | sgp5ebbrE13.b1, sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
Description: | Battery voltage was too low. All data was incorrect. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/01/2000 | 1300 | 03/01/2000 | 1301 |
Subject: | SGP/EBBR/E13 - E13 EBBR All Data Incorrect |
DataStreams: | sgp5ebbrE13.b1, sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
Description: | Battery voltage was too low. All data was incorrect. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/03/2000 | 0000 | 03/04/2000 | 1501 |
Subject: | SGP/EBBR/E13 - E13 EBBR Flux Data Incorrect |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
Description: | The Automatic Exchange Mechanism was stuck inbetween the top and bottom positions. Incorrect Bowen ratio, and latent and sensible heat fluxes resulted. |
Measurements: | sgp15ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/13/2000 | 1300 | 03/13/2000 | 1301 |
Subject: | SGP/EBBR/E13 - E13 EBBR All Data Incorrect |
DataStreams: | sgp5ebbrE13.b1, sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
Description: | Battery voltage was too low. All data was incorrect. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/15/2000 | 1300 | 03/15/2000 | 1301 |
Subject: | SGP/EBBR/E13 - E13 EBBR All Data Incorrect |
DataStreams: | sgp5ebbrE13.b1, sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
Description: | Battery voltage was too low. All data was incorrect. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/22/2000 | 1300 | 03/22/2000 | 1301 |
Subject: | SGP/EBBR/E13 - E13 EBBR All Data Incorrect |
DataStreams: | sgp5ebbrE13.b1, sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
Description: | Battery voltage was too low. All data was incorrect. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/23/2000 | 1300 | 03/23/2000 | 1301 |
Subject: | SGP/EBBR/E13 - E13 EBBR All Data Incorrect |
DataStreams: | sgp5ebbrE13.b1, sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
Description: | Battery voltage was too low. All data was incorrect. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/26/2000 | 1300 | 03/26/2000 | 1301 |
Subject: | SGP/EBBR/E13 - E13 EBBR All Data Incorrect |
DataStreams: | sgp5ebbrE13.b1, sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
Description: | Battery voltage was too low. All data was incorrect. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/27/2000 | 1300 | 03/27/2000 | 1301 |
Subject: | SGP/EBBR/E13 - E13 EBBR All Data Incorrect |
DataStreams: | sgp5ebbrE13.b1, sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
Description: | Battery voltage was too low. All data was incorrect. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/17/2000 | 1730 | 03/23/2000 | 1901 |
Subject: | SGP/EBBR/E13 - E13 EBBR Flux Data Incorrect |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
Description: | The Automatic Exchange Mechanism was stuck inbetween the top and bottom positions. Incorrect Bowen ratio, and latent and sensible heat fluxes resulted. |
Measurements: | sgp15ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/14/2000 | 0200 | 12/14/2000 | 2301 |
Subject: | SGP/EBBR/E13 - E13 EBBR Flux Data Incorrect |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
Description: | The Automatic Exchange Mechanism was stuck inbetween the top and bottom positions. Incorrect Bowen ratio, and latent and sensible heat fluxes resulted. |
Measurements: | sgp15ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/31/2000 | 2300 | 12/31/2000 | 2331 |
Subject: | SGP/EBBR/E13 - E13 EBBR Flux Data Incorrect |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
Description: | The Automatic Exchange Mechanism was stuck inbetween the top and bottom positions. Incorrect Bowen ratio, and latent and sensible heat fluxes resulted. |
Measurements: | sgp15ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/01/2000 | 0000 | 12/31/2000 | 0000 |
Subject: | SGP/EBBR/E13 - E13 EBBR Soil Temperature #4 Suspect |
DataStreams: | sgp30ebbrE13.b1, sgp15ebbrE13.b1 |
Description: | Soil Temperature #4 produced noisy measurements, resulting in quantities calculated from it to be of suspect quality. This condition normally occurred early in the GMT day. Sensible (h) and latent (e) heat fluxes were not normally greatly affected. Average surface heat flow was normally slightly underestimated. |
Measurements: | sgp15ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/11/2000 | 0000 | 12/14/2000 | 1931 |
Subject: | SGP/EBBR/E13 - E13 EBBR Wind Direction Iced |
DataStreams: | sgp5ebbrE13.b1, sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
Description: | The wind direction sensor was iced stopped for the indicated period. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/13/2000 | 1930 | 01/12/2001 | 2001 |
Subject: | SGP/EBBR/E13 - E13 EBBR Soil Heat Flow #4 Incorrect |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
Description: | Soil Heat Flow #4 was large positive or negative, but always incorrect during the indicated period. This resulted in overestimates or underestimates of laetnt and sensible heat fluxes and average soil heat flow. |
Measurements: | sgp15ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/01/2000 | 1730 | 02/05/2000 | 0131 |
02/05/2000 | 1500 | 02/06/2000 | 1231 |
02/06/2000 | 1430 | 02/08/2000 | 0701 |
02/08/2000 | 1500 | 02/10/2000 | 1931 |
02/19/2000 | 1700 | 02/20/2000 | 0201 |
02/20/2000 | 1530 | 02/21/2000 | 1701 |
02/23/2000 | 0300 | 02/23/2000 | 2001 |
02/24/2000 | 0200 | 02/24/2000 | 1501 |
02/25/2000 | 0030 | 02/27/2000 | 0501 |
02/29/2000 | 1630 | 03/01/2000 | 0931 |
03/01/2000 | 1430 | 03/04/2000 | 0801 |
03/04/2000 | 1430 | 03/07/2000 | 1601 |
03/11/2000 | 1530 | 03/12/2000 | 0331 |
03/12/2000 | 1430 | 03/16/2000 | 1831 |
03/17/2000 | 1800 | 03/21/2000 | 1531 |
02/27/2000 | 1430 | 02/29/2000 | 0231 |
Subject: | SGP/EBBR/E15 - E15 EBBR Reference Temperature Incorrect |
DataStreams: | sgp5ebbrE15.b1, sgp30ebbrE15.b1, sgp15ebbrE15.b1 |
Description: | The reference temperature was incorrect during the indicated periods. This temperature is used to adjust air temperature measurements; therefore air temperature and resulting sensible and latent heat fluxes and Bowen ratio are incorrect. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/17/2000 | 1700 | 03/17/2000 | 1701 |
Subject: | SGP/EBBR/E15 - E15 EBBR Data Incorrect |
DataStreams: | sgp30ebbrE15.b1, sgp15ebbrE15.b1, sgp5ebbrE15.b1, sgp5ebbrE15.00 |
Description: | All of the measurements (except, battery voltage, time_offset, lat, lon, etc.) were incorrect during the installation of a new reference temperature sensor. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp5ebbrE15.00:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/17/2000 | 1600 | 03/17/2000 | 1631 |
Subject: | SGP/EBBR/E15 - E15 EBBR Data Missing |
DataStreams: | sgp30ebbrE15.b1, sgp15ebbrE15.b1, sgp5ebbrE15.b1, sgp5ebbrE15.00 |
Description: | The data for this period was missing, during replacement of the reference temperature sensor. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp5ebbrE15.00:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/04/2000 | 1530 | 04/04/2000 | 1531 |
Subject: | SGP/EBBR/E15 - E15 EBBR Data Incorrect |
DataStreams: | sgp30ebbrE15.b1, sgp15ebbrE15.b1, sgp5ebbrE15.b1, sgp5ebbrE15.00 |
Description: | All of the measurements (except, battery voltage, time_offset, lat, lon, etc.) were incorrect during a logger upgrade. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp5ebbrE15.00:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/04/2000 | 1500 | 04/04/2000 | 1501 |
Subject: | SGP/EBBR/E15 - E15 EBBR Data Missing |
DataStreams: | sgp30ebbrE15.b1, sgp15ebbrE15.b1, sgp5ebbrE15.b1, sgp5ebbrE15.00 |
Description: | The data for this period was missing, during an upgrade of the data logger. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp5ebbrE15.00:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/17/2000 | 0130 | 03/17/2000 | 0331 |
Subject: | SGP/EBBR/E15 - E15 EBBR Some Soil Temperatures and Heat Flows Incorrect |
DataStreams: | sgp30ebbrE15.b1, sgp15ebbrE15.b1 |
Description: | For an unknown reason, some of the soil temperatures and soil heat flows were incorrect, leading to many of the calculated quantities being incorrect. |
Measurements: | sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/04/2000 | 1830 | 04/04/2000 | 1831 |
Subject: | SGP/EBBR/E15 - E15 EBBR Some Soils and Flux Data Incorrect |
DataStreams: | sgp30ebbrE15.b1 |
Description: | The soil heat capacity measurements were incorrect for an unknown reason, resulting in incorrect average soil heat flow and latent and sensible heat fluxes. |
Measurements: | sgp30ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/21/2000 | 0130 | 09/22/2000 | 1331 |
09/22/2000 | 2300 | 10/03/2000 | 1531 |
12/31/2000 | 1930 | 01/09/2001 | 1701 |
Subject: | SGP/EBBR/E15 - E15 EBBR AEM Stuck in Middle |
DataStreams: | sgp5ebbrE15.b1, sgp30ebbrE15.b1, sgp15ebbrE15.b1 |
Description: | When the AEM does not complete switching and becomes hung at a mid position, the indicated data measurements are incorrect. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/19/2000 | 2115 | 09/19/2000 | 2131 |
Subject: | SGP/EBBR/E18 - E18 EBBR Data Incorrect |
DataStreams: | sgp30ebbrE18.b1, sgp15ebbrE18.b1, sgp5ebbrE18.b1 |
Description: | The EBBR data was incorrect during maintenance. |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/01/2000 | 0000 | 04/04/2000 | 2131 |
Subject: | SGP/EBBR/E18 - E18 EBBR Right RH Incorrect |
DataStreams: | sgp30ebbrE18.b1, sgp5ebbrE18.b1, sgp15ebbrE18.b1 |
Description: | Right RH being incorrect caused a number of calculated measurements to be incorrect also, including Bowen ratio, and latent and sensible heat fluxes. |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/21/2000 | 2300 | 04/04/2000 | 2131 |
Subject: | SGP/EBBR/E18 - E18 EBBR Thum Right Incorrect |
DataStreams: | sgp30ebbrE18.b1, sgp15ebbrE18.b1, sgp5ebbrE18.b1 |
Description: | The temperature portion of the right T/RH probe was not working. Calculated vapor pressure, Bowen ratio, and sensible and latent heat fluxes were also incorrect. |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/03/2000 | 1545 | 10/03/2000 | 1815 |
Subject: | SGP/EBBR/E15 - E15 EBBR AEM Replacement |
DataStreams: | sgp5ebbrE15.b1, sgp30ebbrE15.b1, sgp15ebbrE15.b1 |
Description: | Large home signals occured while the AEM was being replaced; the indicated data measurements are incorrect. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/05/2000 | 1700 | 10/05/2000 | 1901 |
10/04/2000 | 1045 | 10/04/2000 | 1316 |
10/05/2000 | 0700 | 10/05/2000 | 1601 |
Subject: | SGP/EBBR/E15 - E15 EBBR Battery Voltage Low |
DataStreams: | sgp15ebbrE15.b1, sgp5ebbrE15.b1, sgp30ebbrE15.b1 |
Description: | When EBBR battery voltage is 10.4 volts or less, the measurements are considered to be incorrect. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/15/2000 | 0730 | 10/17/2000 | 1516 |
Subject: | SGP/EBBR/E15 - E15 EBBR Incorrect Net Radiation |
DataStreams: | sgp5ebbrE15.b1, sgp30ebbrE15.b1, sgp15ebbrE15.b1 |
Description: | Incorrect net radiation resulted in incorrect sensible and latent heat fluxes. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/18/2000 | 2100 | 04/18/2000 | 2101 |
Subject: | SGP/EBBR/E18 - E18 EBBR Data Incorrect |
DataStreams: | sgp30ebbrE18.b1, sgp15ebbrE18.b1, sgp5ebbrE18.b1 |
Description: | The EBBR data was incorrect while the datalogger was being upgraded/replaced. |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/28/2000 | 1730 | 07/28/2000 | 1731 |
Subject: | SGP/EBBR/E18 - E18 EBBR Data Incorrect |
DataStreams: | sgp30ebbrE18.b1, sgp15ebbrE18.b1, sgp5ebbrE18.b1 |
Description: | The EBBR data was incorrect just after power was restored. |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/11/2000 | 1000 | 12/13/2000 | 1701 |
Subject: | SGP EBBR E18 - E18 EBBR Wind Direction Frozen |
DataStreams: | sgp30ebbrE18.b1, sgp15ebbrE18.b1, sgp5ebbrE18.b1 |
Description: | The wind direction sensor was frozen stopped. The indicated measurements are incorrect as a result. |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/05/2000 | 1900 | 04/05/2000 | 1901 |
04/05/2000 | 2000 | 04/05/2000 | 2001 |
Subject: | SGP EBBR E22 - E22 EBBR Data Incorrect During Logger Replacement |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1, sgp5ebbrE22.00 |
Description: | The data was incorrect during replacement of the logger. |
Measurements: | sgp5ebbrE22.00:
sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/21/2000 | 1900 | 04/27/2000 | 1031 |
05/10/2000 | 2000 | 06/11/2000 | 0531 |
07/08/2000 | 1800 | 10/16/2000 | 0231 |
Subject: | SGP EBBR E22 - E22 EBBR Soil Moisture Lower Than Measurement Capability |
DataStreams: | sgp30ebbrE22.b1, sgp15ebbrE22.b1 |
Description: | The indicated measurements were incorrect because soil moisture dropped below (sometimes even negative) the ability of the probes to measure it. |
Measurements: | sgp30ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/16/2000 | 0235 | 10/22/2000 | 0316 |
Subject: | SGP EBBR E22 - E22 EBBR Soil Moisture #3 Lower Than Can Be Measured |
DataStreams: | sgp30ebbrE22.b1, sgp15ebbrE22.b1 |
Description: | The soil moisture was too low to be measured correctly by probe #3. |
Measurements: | sgp30ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/16/2000 | 0130 | 10/18/2000 | 1841 |
Subject: | SGP EBBR E22 - E22 EBBR Net Radiation Incorrect |
DataStreams: | sgp15ebbrE22.b1, sgp5ebbrE22.b1, sgp30ebbrE22.b1 |
Description: | Water inside the domes of the net radiometer caused measurements from it to be incorrect. Calculated latent and sensible heat fluxes were therefore incorrect also. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/26/2000 | 1330 | 12/29/2000 | 1901 |
Subject: | SGP EBBR E18 - E18 EBBR Wind Direction Frozen |
DataStreams: | sgp30ebbrE18.b1, sgp15ebbrE18.b1, sgp5ebbrE18.b1 |
Description: | The wind direction sensor was frozen stopped. The indicated measurements are incorrect as a result. |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/12/2000 | 0015 | 09/19/2000 | 2101 |
Subject: | SGP EBBR E18 - E18 EBBR Left Ventilator Fan Not Working |
DataStreams: | sgp30ebbrE18.b1, sgp5ebbrE18.b1, sgp15ebbrE18.b1 |
Description: | The left ventilator fan was not working. Some of the indicated measurements may be incorrect (under some wind speed and direction conditions, cloud cover, and time of day, some of the data may be correct). The data is therefore described as suspect for the indicated period. |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/02/1999 | 0030 | 12/02/1999 | 1631 |
Subject: | SGP EBBR E19 - E19 EBBR Left RH Probe Temperature Incorrect |
DataStreams: | sgp5ebbrE19.b1, sgp30ebbrE19.b1, sgp15ebbrE19.b1 |
Description: | The Left RH probe temperature was drifting. The indicated measurements are incorrect. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/02/2000 | 1500 | 12/08/2000 | 1946 |
Subject: | SGP EBBR E19 - E19 EBBR Left Aspirator Housing Off |
DataStreams: | sgp5ebbrE19.b1, sgp30ebbrE19.b1, sgp15ebbrE19.b1 |
Description: | The left aspirator housing fell off and thus the indicated measurements are incorrect. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/07/1999 | 0000 | 02/24/2000 | 1601 |
Subject: | SGP EBBR E19 - E19 EBBR Left Air Temperature Incorrect |
DataStreams: | sgp30ebbrE19.b1, sgp5ebbrE19.b1, sgp15ebbrE19.b1 |
Description: | A bad capacitor on the J-panel caused left air temperature, and thus other measurements to be incorrect. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/09/2000 | 1630 | 03/09/2000 | 1701 |
Subject: | SGP EBBR E19 - E19 EBBR Data Incorrect During J-panel Replacement |
DataStreams: | sgp15ebbrE19.b1, sgp5ebbrE19.b1, sgp30ebbrE19.b1 |
Description: | The indicated data was incorreect during a J-panel replacement. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/12/2000 | 2030 | 12/13/2000 | 0731 |
Subject: | SGP EBBR E19 - E19 EBBR Wind Direction Incorrect |
DataStreams: | sgp30ebbrE19.b1, sgp5ebbrE19.b1, sgp15ebbrE19.b1 |
Description: | The wind direction sensor was frozen stopped during the indicated period. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/24/1999 | 0445 | 01/25/2001 | 1531 |
Subject: | SGP EBBR E19 - E19 EBBR Soil Heat Flow #1 Sign Reversed |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1 |
Description: | The sign was reversed (heat flow plate #1 was installed upside down), causing incorrect measurements, as indicated. The average soil heat flow can be recalculated by using an average of soil sensor sets 2-5. The recalculated average soil heat flow can then be used to recalculate latent and sensible heat fluxes. |
Measurements: | sgp15ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/06/2000 | 2000 | 03/06/2000 | 1901 |
03/06/2000 | 1930 | 03/06/2000 | 2031 |
Subject: | SGP EBBR E20 - E20 EBBR Soil Heat Flow #3 Incorrect |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1 |
Description: | Soil Heat Flow #3 was incorrect during the indicated periods. Average soil heat flux, latent heat flux, and sensible heat flux were therefore also incorrect. |
Measurements: | sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/22/2000 | 0000 | 05/22/2000 | 0000 |
Subject: | SGP EBBR E20 - E20 EBBR Right Air T Incorrect |
DataStreams: | sgp5ebbrE20.b1, sgp30ebbrE20.b1, sgp15ebbrE20.b1 |
Description: | The right air temperature probe was incorrect while connections for it were checked. Bowen ratio, sensible heat flux, and latent heat flux were incorrect as a result. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/02/2000 | 2345 | 02/21/2001 | 1946 |
Subject: | SGP EBBR E20 - E20 EBBR Right Air T Incorrect |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | The right air temperature probe was incorrect frequently. Inspection of the right air temperature data will quickly show where it went offscale. Bowen ratio, sensible heat flux, and latent heat flux were incorrect as a result. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/01/2000 | 1930 | 11/01/2000 | 1931 |
Subject: | SGP EBBR E20 - E20 EBBR Soil Temperatures Incorrect |
DataStreams: | sgp30ebbrE20.b1, sgp15ebbrE20.b1 |
Description: | All soil temperatures were incorrect for the indicated period. Latent and sensible heat fluxes were therefore incorrect also. |
Measurements: | sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/15/2000 | 2031 | 11/15/2000 | 2101 |
Subject: | SGP EBBR E20 - E20 EBBR Data Incorrect During STP #3 Replacement |
DataStreams: | sgp30ebbrE20.b1, sgp15ebbrE20.b1, sgp5ebbrE20.b1, sgp5ebbrE20.00 |
Description: | All data was incorrect during replacement of soil temperature probe #3. |
Measurements: | sgp5ebbrE20.b1:
sgp5ebbrE20.00:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/29/2000 | 1831 | 11/29/2000 | 1901 |
Subject: | SGP EBBR E20 - E20 EBBR Data Incorrect After Program Restart |
DataStreams: | sgp30ebbrE20.b1, sgp15ebbrE20.b1, sgp5ebbrE20.b1, sgp5ebbrE20.00 |
Description: | All data was incorrect after a datalogger program restart. |
Measurements: | sgp5ebbrE20.b1:
sgp5ebbrE20.00:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/29/2000 | 1930 | 11/29/2000 | 1931 |
Subject: | SGP EBBR E20 - E20 EBBR Some Data Incorrect |
DataStreams: | sgp30ebbrE20.b1 |
Description: | Quantities requiring a starting value were incorrect; sensible and latent heat fluxes were therefore incorrect also. |
Measurements: | sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/07/2000 | 1830 | 02/07/2000 | 1831 |
02/08/2000 | 1830 | 02/08/2000 | 1831 |
02/12/2000 | 2000 | 02/12/2000 | 2001 |
03/06/2000 | 1730 | 03/06/2000 | 1731 |
02/14/2000 | 1900 | 02/14/2000 | 2001 |
02/14/2000 | 2100 | 02/14/2000 | 2101 |
02/15/2000 | 1830 | 02/15/2000 | 1901 |
02/19/2000 | 1830 | 02/20/2000 | 1831 |
02/20/2000 | 1930 | 02/20/2000 | 2201 |
02/27/2000 | 2200 | 02/27/2000 | 2201 |
03/04/2000 | 1930 | 03/04/2000 | 2131 |
03/05/2000 | 1930 | 03/05/2000 | 2101 |
02/27/2000 | 1930 | 02/27/2000 | 1931 |
Subject: | SGP EBBR E25 - E25 EBBR Soil Heat Flow #5 Incorrect |
DataStreams: | sgp30ebbrE25.b1, sgp15ebbrE25.b1 |
Description: | Soil Heat Flow probe #5 gave incorrect measurements; the measurements were either too low or low offscale. Average soil heat flow, sensible heat flux, and latent heat flux were therefore also incorrect, but can be calculated from the remaining correct soil heat flows. |
Measurements: | sgp30ebbrE25.b1:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/03/2000 | 0200 | 11/08/2000 | 1301 |
11/08/2000 | 1330 | 11/08/2000 | 2301 |
11/09/2000 | 1600 | 11/10/2000 | 0201 |
11/11/2000 | 1500 | 11/13/2000 | 0301 |
11/13/2000 | 1630 | 11/14/2000 | 0331 |
11/14/2000 | 1530 | 11/15/2000 | 0201 |
11/15/2000 | 1430 | 11/16/2000 | 1331 |
11/16/2000 | 1430 | 11/17/2000 | 0430 |
11/18/2000 | 1400 | 11/19/2000 | 0430 |
11/19/2000 | 1500 | 11/20/2000 | 0301 |
11/20/2000 | 1500 | 11/21/2000 | 0001 |
11/21/2000 | 1500 | 02/07/2001 | 1731 |
11/10/2000 | 1530 | 11/11/2000 | 0301 |
11/17/2000 | 1530 | 11/18/2000 | 0101 |
Subject: | SGP EBBR E25 - E25 EBBR Soil Heat Flow #3 Incorrect |
DataStreams: | sgp30ebbrE25.b1, sgp15ebbrE25.b1 |
Description: | Soil Heat Flow #3 was frequently too small or large negative. Average soil heat flow, sensible heat flux, and latent heat flux were also incorrect during these times; these can be re-calculated from the remaining correct soil heat flows. |
Measurements: | sgp30ebbrE25.b1:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/23/2000 | 0245 | 11/29/2000 | 1631 |
12/03/2000 | 1930 | 12/13/2000 | 1746 |
Subject: | SGP EBBR E25 - E25 EBBR Left Aspirator Not Working |
DataStreams: | sgp30ebbrE25.b1, sgp15ebbrE25.b1, sgp5ebbrE25.b1 |
Description: | The left aspirator unit of the AEM was not working. Therefore the indicated measurements are incorrect. |
Measurements: | sgp30ebbrE25.b1:
sgp5ebbrE25.b1:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/01/2000 | 0000 | 12/21/2000 | 1701 |
Subject: | SGP EBBR E25 - E25 EBBR Soil Heat Flow #1 Incorrect |
DataStreams: | sgp30ebbrE25.b1, sgp15ebbrE25.b1 |
Description: | Soil Heat Flow #1 was always positive and therefore incorrect. Average soil heat flow, sensible heat flux, and latent heat flux were therefore incorrect also. |
Measurements: | sgp30ebbrE25.b1:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/03/2000 | 0600 | 03/07/2000 | 0901 |
03/24/2000 | 0100 | 04/06/2000 | 1701 |
06/19/2000 | 0830 | 06/19/2000 | 1431 |
06/19/2000 | 1630 | 06/19/2000 | 1731 |
06/19/2000 | 1900 | 06/19/2000 | 1901 |
06/20/2000 | 1430 | 06/20/2000 | 2201 |
06/30/2000 | 0945 | 06/30/2000 | 1346 |
07/02/2000 | 0930 | 07/02/2000 | 1601 |
07/03/2000 | 1230 | 07/03/2000 | 1316 |
Subject: | SGP EBBR E26 - Data Suspect, Low Battery Voltage |
DataStreams: | sgp30ebbrE26.b1, sgp15ebbrE26.b1, sgp5ebbrE26.b1, sgp5ebbrE26.00 |
Description: | The data measurements are suspect because of low battery voltage. Some measurements may be correct, but most are probably incorrect. Large home signals indicate very low battery voltage is present. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
sgp5ebbrE26.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/06/2000 | 2230 | 09/06/2000 | 2231 |
09/10/2000 | 1430 | 09/10/2000 | 1431 |
10/16/2000 | 1500 | 10/16/2000 | 1501 |
10/23/2000 | 1900 | 10/23/2000 | 1901 |
10/26/2000 | 1530 | 10/26/2000 | 1601 |
10/27/2000 | 1430 | 10/27/2000 | 1501 |
01/03/2001 | 1530 | 01/03/2001 | 1631 |
01/02/2001 | 1600 | 01/02/2001 | 2201 |
Subject: | SGP EBBR E26 - All Data Incorrect |
DataStreams: | sgp30ebbrE26.b1, sgp15ebbrE26.b1, sgp5ebbrE26.b1, sgp5ebbrE26.00 |
Description: | All data was incorrect during these periods for unknown reasons. However, this tends to happen during preventative maintenance visits if sensors have to be changed out or tested and the datalogger program is reloaded, or if the battery voltage is low. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
sgp5ebbrE26.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/16/1999 | 1700 | 03/03/2000 | 0931 |
Subject: | SGP EBBR E26 - Soil Heat Flow #1 Incorrect |
DataStreams: | sgp30ebbrE26.b1, sgp15ebbrE26.b1 |
Description: | Soil Heat Flow #1 was incorrect because of a bad electronic connection. Average surface heat flow, sensible heat flux, and latent heat flux are incorrect as a result; these can be re-calculated by using the remaining correct soil heat flows. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/18/2000 | 0330 | 02/19/2000 | 1131 |
03/04/2000 | 1500 | 03/07/2000 | 0901 |
Subject: | SGP EBBR E26 - Soil Heat Flow #2 Incorrect |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1 |
Description: | Soil Heat Flow #2 was incorrect. Average soil heat flow, sensible heat flux, and latent heat flux were therefore also incorrect; these can be re-calculated from the remaining correct soil heat flows. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/05/2000 | 1900 | 11/16/2000 | 1601 |
11/30/2000 | 1600 | 12/03/2000 | 0731 |
12/14/2000 | 1600 | 12/16/2000 | 0301 |
06/15/2000 | 1500 | 06/19/2000 | 2201 |
Subject: | SGP EBBR E26 - Soil Heat Flow #3 Incorrect |
DataStreams: | sgp30ebbrE26.b1, sgp15ebbrE26.b1 |
Description: | Soil heat Flow #3 was incorrect (large negative). Therefore average soil heat flow, sensible heat flux, and latent heat flux were also incorrect; these can be re-calculated using the remaining correct soil heat flows. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/29/1999 | 0000 | 03/08/2001 | 1931 |
Subject: | SGP EBBR E13 - E13 EBBR Soil T #4 Incorrect |
DataStreams: | sgp30ebbrE13.b1, sgp15ebbrE13.b1 |
Description: | Soil Temperature #4 floated around, causing average soil heat flow to sometimes be too small (resulting too small a sensible heat flux and too large a latent heat flux) or too large (resulting in too large a sensible heat flux and too small a latent heat flux). Investigation of g4 will indicate when these conditions occurred and the likely effect on sensible and latent heat fluxes. |
Measurements: | sgp15ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/17/2000 | 2001 | 05/17/2000 | 2031 |
Subject: | SGP EBBR E20 - E20 EBBR Data Incorrect |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1, sgp5ebbrE20.00 |
Description: | Data was incorrect right after a datalogger upgrade. |
Measurements: | sgp5ebbrE20.b1:
sgp5ebbrE20.00:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/20/2000 | 1930 | 08/02/2000 | 0331 |
Subject: | SGP EBBR E20 - E20 EBBR Data Incorrect |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1, sgp5ebbrE20.00 |
Description: | The data was incorrect because battery voltage was low. |
Measurements: | sgp5ebbrE20.b1:
sgp5ebbrE20.00:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/04/2000 | 1930 | 10/04/2000 | 1931 |
Subject: | SGP EBBR E20 - E20 EBBR Net Radiation Data Incorrect |
DataStreams: | sgp30ebbrE20.b1, sgp5ebbrE20.b1, sgp15ebbrE20.b1 |
Description: | The net radiation data was incorrect while the top dome was being replaced. Sensible and latent heat fluxes are therefore also incorrect. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/06/2000 | 1730 | 02/08/2001 | 1546 |
Subject: | SGP EBBR E26 - Soil Moisture #1 Incorrect |
DataStreams: | sgp30ebbrE26.b1, sgp15ebbrE26.b1 |
Description: | Soil Moisture #1 was a constant 33.7 and was incorrect. Latent and Sensible Heat fluxes were normally minimally affected. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/23/2000 | 0345 | 08/23/2000 | 0401 |
Subject: | SGP EBBR E26 - AEM Inbetween, Data Incorrect |
DataStreams: | sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp15ebbrE26.b1 |
Description: | While the AEM was inbetween the correct measurement positions the measurements from the two levels were incorrect. The sensible and latent heat fluxes and Bowen ratio were therefore incorrect also. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/15/2000 | 1530 | 02/22/2001 | 1535 |
Subject: | SGP EBBR E26 - Soil Heat Flow #5 Incorrect |
DataStreams: | sgp30ebbrE26.b1, sgp15ebbrE26.b1 |
Description: | Soil Heat Flow #5 was large negative during this period. Therefore, calculated average soil heat flow, latent heat flux, and sensible heat flux were incorrect. These measurements can be recalculated from the use of measurements from only soil heat flow sensors #1-4. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/21/2000 | 2130 | 10/21/2000 | 2131 |
Subject: | SGP EBBR E26 - Some Data Incorrect |
DataStreams: | sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp15ebbrE26.b1 |
Description: | The reference temperature sensor malfunctioned, causing all associated measurements to be incorrect. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/10/2000 | 1500 | 09/10/2000 | 1501 |
11/02/2000 | 1600 | 11/02/2000 | 1601 |
01/03/2001 | 1700 | 01/03/2001 | 1701 |
Subject: | SGP EBBR E26 - Some Data Incorrect |
DataStreams: | sgp30ebbrE26.b1 |
Description: | Since soil temperature was not available for the previous half hour, soil energy storage could not be calculated properly for the indicated period. The individual soil heat storage from the five soil sets, average soil heat flow, sensible heat flux, and latent heat flux are therefore incorrect. |
Measurements: | sgp30ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/07/2001 | 0000 | 05/07/2001 | 2356 |
05/21/2001 | 0000 | 05/21/2001 | 2356 |
Subject: | SGP EBBR E19 - E19 EBBR Soil Temperature 3 Intermittenmtly Incorrect |
DataStreams: | sgp30ebbrE19.b1, sgp15ebbrE19.b1 |
Description: | Soil Temperature # 3 was incorrect at times during these periods. Compare G3 with G1, G2, G4, and G5 to determine when TS3 was incorrect. |
Measurements: | sgp15ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/01/2001 | 0000 | 04/01/2001 | 1631 |
Subject: | SGP EBBR E19 - E19 EBBR TS3 Intermittent Incorrect |
DataStreams: | sgp30ebbrE19.b1, sgp15ebbrE19.b1 |
Description: | Soil Temperature probe #3 was intermittently incorrect. Examine the data and compare G3 with G1-2 and G4-5 to see when TS3 is incorrect. |
Measurements: | sgp15ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/01/2001 | 0300 | 03/01/2001 | 1651 |
Subject: | SGP EBBR E19 - E19 EBBR Wind Speed Incorrect |
DataStreams: | sgp30ebbrE19.b1, sgp5ebbrE19.b1, sgp15ebbrE19.b1 |
Description: | The wind speed sensor was iced stopped for a while on this day. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/26/2001 | 1445 | 04/26/2001 | 1536 |
11/18/2001 | 1535 | 11/18/2001 | 1631 |
Subject: | SGP EBBR E2 - E2 EBBR Data Incorrect During 6 Month Checks |
DataStreams: | sgp30ebbrE2.b1, sgp15ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | The indicated data was incorrect during the 6 month checks. |
Measurements: | sgp30ebbrE2.b1:
sgp15ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/08/2001 | 1800 | 05/08/2001 | 1901 |
11/25/2001 | 1435 | 11/25/2001 | 1601 |
Subject: | SGP EBBR E7 - E7 EBBR Data Incorrect During 6 Month Checks |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | The data was incorrect during the 6 month checks. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/06/2001 | 1815 | 11/06/2001 | 1931 |
04/24/2001 | 1700 | 04/24/2001 | 1746 |
Subject: | SGP EBBR E8 - E8 EBBR Data Incorrect During 6 Month Checks |
DataStreams: | sgp30ebbrE8.b1, sgp15ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The data was incorrect during the 6 month checks. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/22/2001 | 1500 | 05/22/2001 | 1536 |
04/12/2001 | 1800 | 04/12/2001 | 1846 |
10/23/2001 | 1510 | 10/23/2001 | 1701 |
Subject: | SGP EBBR E9 - E9 EBBR Data Incorect During 6 Month Checks |
DataStreams: | sgp5ebbrE9.b1, sgp15ebbrE9.b1, sgp30ebbrE9.b1 |
Description: | The data was incorrect during the 6 month checks. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/03/2001 | 1625 | 04/03/2001 | 1716 |
10/18/2001 | 1930 | 10/18/2001 | 2116 |
Subject: | SGP EBBR E12 - E12 EBBR Data Incorrect During 6 Month Checks |
DataStreams: | sgp30ebbrE12.b1, sgp15ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | The data was incorrect during the 6 month checks. |
Measurements: | sgp30ebbrE12.b1:
sgp15ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/20/2001 | 1720 | 04/20/2001 | 1749 |
10/04/2001 | 1835 | 10/04/2001 | 1934 |
Subject: | SGP EBBR E13 - E13 EBBR Data Incorrect During 6 Month Checks |
DataStreams: | sgp5ebbrE13.b1, sgp30ebbrE13.b1, sgp15ebbrE13.b1 |
Description: | The data was incorrect during the 6 month checks. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/17/2001 | 1850 | 04/17/2001 | 1945 |
10/02/2001 | 1604 | 10/02/2001 | 1742 |
Subject: | SGP EBBR E15 - E15 EBBR Data Incorrect During 6 Month Checks |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | The data was incorrect during the 6 month checks. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/03/2001 | 2255 | 04/03/2001 | 2341 |
10/18/2001 | 1421 | 10/18/2001 | 1601 |
Subject: | SGP EBBR E18 - E18 EBBR Data Incorrect During 6 Month Checks |
DataStreams: | sgp5ebbrE18.b1, sgp30ebbrE18.b1, sgp15ebbrE18.b1 |
Description: | The data was incorrect during the 6 month checks. |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/19/2001 | 1700 | 04/19/2001 | 2001 |
10/03/2001 | 1413 | 10/03/2001 | 1506 |
Subject: | SGP EBBR E19 - E19 EBBR Data Incorrect During 6 Month Checks |
DataStreams: | sgp30ebbrE19.b1, sgp15ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | The data was incorrect during the 6 month checks. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/17/2001 | 1500 | 10/17/2001 | 1601 |
04/04/2001 | 1700 | 04/04/2001 | 1736 |
Subject: | SGP EBBR E20 - E20 EBBR Data Incorrect During 6 Month Checks |
DataStreams: | sgp30ebbrE20.b1, sgp15ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | The data was incorrect during the 6 month checks. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/18/2001 | 1715 | 04/18/2001 | 1821 |
10/10/2001 | 1800 | 10/10/2001 | 1857 |
Subject: | SGP EBBR E22 - E22 EBBR Data Incorrect During 6 Month Checks |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | The data was incorrect during the 6 month checks. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/04/2001 | 1530 | 04/04/2001 | 1631 |
10/17/2001 | 1705 | 10/17/2001 | 1811 |
Subject: | SGP EBBR E25 - E25 EBBR Data Incorrect During 6 Month Checks |
DataStreams: | sgp15ebbrE25.b1, sgp5ebbrE25.b1, sgp30ebbrE25.b1 |
Description: | The data was incorrect during the 6 month checks. |
Measurements: | sgp30ebbrE25.b1:
sgp5ebbrE25.b1:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/05/2001 | 1415 | 04/05/2001 | 1546 |
10/16/2001 | 1750 | 10/16/2001 | 1921 |
Subject: | SGP EBBR E26 - Data Incorrect During 6 Month Checks |
DataStreams: | sgp30ebbrE26.b1, sgp15ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | The data was incorrect during the 6 month checks. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/01/2000 | 2300 | 11/15/2000 | 2031 |
Subject: | SGP EBBR E20 - E20 EBBR Soil Temperature #3 Incorrect |
DataStreams: | sgp30ebbrE20.b1, sgp15ebbrE20.b1 |
Description: | The indicated measurements were incorrect while soil temperature probe #3 was not working. |
Measurements: | sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/28/2001 | 1330 | 02/28/2001 | 1331 |
Subject: | SGP EBBR E7 - E7 EBBR Data Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil moisture measurements were spiked by an unknown cause. This caused the indicated measurements and calculated quantities to be incorrect. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/09/2001 | 1230 | 03/09/2001 | 1231 |
Subject: | SGP EBBR E7 - E7 EBBR Soil Moisture #3 and Other Data Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil moisture #3 spiked and caused other measurements and calculated quantities to also be incorrect. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/27/2001 | 1745 | 03/27/2001 | 1936 |
Subject: | SGP EBBR E7 - E7 EBBR AEM Position Caused Incorrect Data |
DataStreams: | sgp5ebbrE7.b1, sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | The AEM stuck in one position, cuasing incorrect measurements and calculated quantities. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/29/2001 | 1500 | 03/29/2001 | 1501 |
Subject: | SGP EBBR E7 - E7 EBBR Soil Moisture #2 Spiked, Incorrect Data Resulted |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil moisture #2 spiked and caused other measurements and calculated quantities to also be incorrect. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/10/2001 | 1945 | 04/10/2001 | 2031 |
Subject: | SGP EBBR E7 - E7 EBBR AEM Position Caused Incorrect Data |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | The AEM stuck in one position, causing the indicated measurements and calculated quantities to be incorrect. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/07/2001 | 1900 | 05/07/2001 | 1901 |
Subject: | SGP EBBR E7 - E7 EBBR All Data Incorrect |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | The data was incorrect immediately after the re-calibrated system was installed and put into service. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/07/2001 | 1930 | 05/07/2001 | 1931 |
Subject: | SGP EBBR E7 - E7 EBBR Data Incorrect After Recalibration |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | Soil temperature measurements and the half hour average were incorrect the half hour after starting the recalibrated system, causing other measurements and calculated quantities to be incorrect. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/28/2001 | 1900 | 08/28/2001 | 1901 |
Subject: | SGP EBBR E7 - E7 EBBR Data Incorrect After Missing Data Period |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil temperature measurements and the half hour average were incorrect the half hour after a missing data period, causing other measurements and calculated quantities to be incorrect. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/14/2000 | 1130 | 03/31/2001 | 2356 |
Subject: | SGP EBBR E26 - Low Battery Voltage |
DataStreams: | sgp5ebbrE26.b1, sgp30ebbrE26.b1, sgp15ebbrE26.b1 |
Description: | During the indicated the period the battery voltage often dropped to 10.4 volts or below. At voltages less than 10.5 volts none of the selected measurements can be considered to be correct. The battery charging capacity was insufficient to maintain voltage at night. If the battery voltage is less than 10.5 volts (see battery voltage in the 15 minute data files), do not use the data. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/30/2001 | 0845 | 05/30/2001 | 1801 |
Subject: | SGP EBBR E25 - E25 EBBR Wind Direction Incorrect |
DataStreams: | sgp30ebbrE25.b1, sgp5ebbrE25.b1 |
Description: | The wind direction was incorrect at a constant 135 degrees during the indicated period. |
Measurements: | sgp30ebbrE25.b1:
sgp5ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/30/2001 | 0605 | 07/11/2001 | 1631 |
Subject: | SGP EBBR E25 - E25 EBBR Wind Direction Incorrect |
DataStreams: | sgp30ebbrE25.b1, sgp5ebbrE25.b1 |
Description: | The wind direction was incorrect at close to 0 degrees during the indicated period. The sensor was replaced to resolve the problem. |
Measurements: | sgp30ebbrE25.b1:
sgp5ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/01/2001 | 0000 | 08/23/2001 | 1530 |
Subject: | SGP EBBR E26 - Battery Voltage Low |
DataStreams: | sgp30ebbrE26.b1, sgp15ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | The battery voltage often dropped below 10.5 volts. All data for voltages below 10.5 volts should be considered to be incorrect. See the 15 minute data files, which contain the battery voltage, to determine when the data was incorrect. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/07/2001 | 2000 | 03/07/2001 | 2001 |
Subject: | SGP EBBR E22 - E22 EBBR 30 Minute Data Incorrect After Missing Data |
DataStreams: | sgp30ebbrE22.b1 |
Description: | Complete soil temperature information is not available when the data is missing for the previous half hour. Since soil temperature difference with time cannot be computed properly, the selected calculated quantities are incorrect. |
Measurements: | sgp30ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/15/2000 | 1630 | 02/07/2001 | 1731 |
Subject: | SGP EBBR E25 - E25 EBBR Soil Heat Flow #3 Incorrect |
DataStreams: | sgp30ebbrE25.b1, sgp15ebbrE25.b1 |
Description: | The selected measurements and calculated quantities were incorrect because soil heat flow sensor #3 indicated large negative soil heat flow. |
Measurements: | sgp30ebbrE25.b1:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/01/2000 | 0000 | 02/21/2001 | 1701 |
Subject: | SGP EBBR E25 - E25 EBBR Soil Heat Flow #1 Incorrect |
DataStreams: | sgp30ebbrE25.b1, sgp15ebbrE25.b1 |
Description: | The #1 soil heat flow sensor developed a +60 W m-2 bias and thus indicated higher soil heat flows than the other sensors. This results in slightly higher sensible and latent heat fluxes than is correct. The sensor was replaced on 2/21/01. |
Measurements: | sgp30ebbrE25.b1:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/18/2001 | 1645 | 04/18/2001 | 1701 |
Subject: | SGP EBBR E25 - E25 EBBR Soil Heat Flow #3 Incorrect |
DataStreams: | sgp30ebbrE25.b1, sgp15ebbrE25.b1 |
Description: | Soil heat flow #3 was large negative during the indicated period. |
Measurements: | sgp30ebbrE25.b1:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/13/2001 | 0030 | 05/16/2001 | 1931 |
Subject: | SGP EBBR E25 - E25 EBBR Soil Heat Flow #1 Incorrect |
DataStreams: | sgp30ebbrE25.b1, sgp15ebbrE25.b1 |
Description: | Soil heat flow #1 was large negative or large positive during the indicated period. |
Measurements: | sgp30ebbrE25.b1:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/16/2001 | 1900 | 04/17/2001 | 1631 |
04/17/2001 | 2330 | 04/18/2001 | 1631 |
04/18/2001 | 2100 | 04/20/2001 | 1301 |
04/20/2001 | 2130 | 04/21/2001 | 0931 |
04/21/2001 | 2030 | 04/26/2001 | 1501 |
04/27/2001 | 0030 | 05/01/2001 | 1431 |
04/06/2001 | 2030 | 04/11/2001 | 1731 |
Subject: | SGP EBBR E12 - E12 EBBR Data Incorrect From AEM In One Position |
DataStreams: | sgp5ebbrE12.b1, sgp30ebbrE12.b1, sgp15ebbrE12.b1 |
Description: | The AEM kept stopping in one position. Offset biases are not removed from the gradient data in this situation and the data indicated is therefore incorrect. |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/27/2001 | 1735 | 06/26/2001 | 1501 |
Subject: | SGP EBBR E15 - E15 EBBR Battery Voltage Low, Data Incorrect |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | When the battery voltage is 10.4 volts or below, all selected data should be considered to be incorrect. See the battery voltage in the 15 minute data to determine times of incorrect data. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/11/2001 | 1930 | 10/11/2001 | 1931 |
Subject: | SGP EBBR E15 - E15 EBBR Data Zeros Incorrect |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | The data was all zeros after a logger restart for 5 , 15, and 30 minute periods. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/21/2001 | 2000 | 03/07/2001 | 1901 |
Subject: | SGP EBBR E20 - E20 EBBR Soil Heat Flow #2 Incorrect |
DataStreams: | sgp30ebbrE20.b1, sgp15ebbrE20.b1 |
Description: | Broken leads for SHF #2 caused measurements from it to be incorrect during the indicated period. |
Measurements: | sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/30/2001 | 0615 | 03/30/2001 | 1431 |
03/31/2001 | 1115 | 03/31/2001 | 1401 |
Subject: | SGP EBBR E20 - E20 EBBR Battery Voltage Low, Incorrect Data |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1 |
Description: | Some measurements were incorrect because the battery voltage was low. |
Measurements: | sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/02/2001 | 1245 | 04/02/2001 | 1701 |
04/03/2001 | 0215 | 04/03/2001 | 1501 |
04/06/2001 | 1115 | 04/06/2001 | 1501 |
04/07/2001 | 0345 | 04/07/2001 | 1401 |
04/23/2001 | 1045 | 04/23/2001 | 1331 |
Subject: | SGP EBBR E20 - E20 EBBR Battery Voltage Low, Data Incorrect |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1 |
Description: | The indicated data was incorrect because the battery voltage was low. |
Measurements: | sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/15/2001 | 1500 | 10/17/2001 | 0201 |
Subject: | SGP/EBBR/E20 - Reprocess: Soil Temperature #3 Incorrect |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1 |
Description: | During the indicated period, TS #3 was large negative or positive or offscale. Average soil heat flow, sensible heat flux, and latent heat flux were incorrect when TS #3 was incorrect. Inspect the 15 and 30 minute TS #3 data to determine the times of incorrect data. |
Measurements: | sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/15/2001 | 0800 | 02/16/2001 | 2031 |
Subject: | SGP EBBR E18 - E18 EBBR AEM Frozen, Data Incorrect |
DataStreams: | sgp30ebbrE18.b1, sgp5ebbrE18.b1, sgp15ebbrE18.b1 |
Description: | The AEM was frozen in one position. The indicated data is incorrect. |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/15/2001 | 0625 | 02/16/2001 | 2031 |
02/16/2001 | 2355 | 02/17/2001 | 1650 |
Subject: | SGP EBBR E18 - E18 EBBR Wind Direction Frozen, Data Incorrect |
DataStreams: | sgp5ebbrE18.b1, sgp15ebbrE18.b1, sgp30ebbrE18.b1 |
Description: | The wind direction sensor was frozen stopped. The indicated data is incorrect. |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/15/2001 | 0620 | 02/17/2001 | 2006 |
Subject: | SGP EBBR E18 - E18 EBBR Wind Speed Frozen, Data Incorrect |
DataStreams: | sgp15ebbrE18.b1, sgp5ebbrE18.b1, sgp30ebbrE18.b1 |
Description: | The wind speed sensor was frozen stopped. The indicated data are incorrect. |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/10/2001 | 1810 | 06/12/2001 | 2240 |
Subject: | SGP EBBR E18 - E18 EBBR Right RH Incorrect |
DataStreams: | sgp30ebbrE18.b1, sgp5ebbrE18.b1, sgp15ebbrE18.b1 |
Description: | The RH portion of the T/RH probe produced incorrect data. Sensible and latent heat fluxes and the Bowen ratio were also incorrect. |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/01/2001 | 0000 | 02/22/2001 | 0531 |
03/01/2001 | 1230 | 03/01/2001 | 1231 |
03/07/2001 | 1805 | 03/07/2001 | 1901 |
Subject: | SGP EBBR E20 - E20 EBBR Soil Heat Flow #4 Incorrect |
DataStreams: | sgp30ebbrE20.b1, sgp15ebbrE20.b1 |
Description: | The SHF #4 produced large negative values during much of the period because of a broken lead wire. Inspect the data to determine the times of incorrect SHF #4 values. |
Measurements: | sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/10/2000 | 2015 | 03/31/2001 | 2356 |
Subject: | SGP EBBR E8 - E8 EBBR Soil Heat Flow #1 Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | Soil Heat Flow #1 was frequently large negative or large positive and was incorrect. The indicated measurements were also incorrect as a result. Correct sensible and latent heat fluxes can be determined by using the other four sets of soil measurements to determine the energy balance. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/01/2001 | 0000 | 12/14/2001 | 2356 |
Subject: | SGP EBBR E8 - E8 EBBR Soil Heat Flow #1 Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | Soil heat flow #1 was usually large negative or positive, and therefore incorrect. Sensible and latent heat fluxes can be recalculated by using the four good sets of soil measurements for the energy balance. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/08/2001 | 2145 | 02/13/2001 | 1830 |
Subject: | SGP EBBR E8 - E8 EBBR AEM Frozen In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The AEM was frozen in one position by an ice storm for 5 days. The indicated measurements are incorrect. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/08/2001 | 1615 | 02/11/2001 | 1745 |
Subject: | SGP EBBR E8 - E8 EBBR Wind Speed Frozen |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The wind speed sensor was frozen stopped by an ice storm. The indicated measurements are incorrect. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/13/2001 | 0600 | 03/31/2001 | 2356 |
Subject: | SGP EBBR E8 - E8 EBBR Battery Voltage Low, Data Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.00, sgp15ebbrE8.b1 |
Description: | The battery voltage was frequently less than 10.5 volts. When the battery voltage is 10.4 V or less, all of the data is suspect. Examine the battery voltage measurement in the 15 minute data to determine whether the measurements can be used. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
sgp5ebbrE8.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/01/2001 | 0000 | 12/14/2001 | 2356 |
Subject: | SGP EBBR E8 - E8 EBBR Battery Voltage Low, Data Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1, sgp5ebbrE8.00 |
Description: | The battery voltage was frequently less than 10.5 volts. When the voltage is 10.4 V or less, all of the measurements are suspect and should not be used (they are likely incorrect). |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
sgp5ebbrE8.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/01/2001 | 0000 | 01/03/2001 | 1901 |
02/22/2001 | 2031 | 02/22/2001 | 2301 |
02/23/2001 | 0545 | 02/25/2001 | 1201 |
02/28/2001 | 2015 | 03/11/2001 | 0831 |
03/20/2001 | 0145 | 03/29/2001 | 1601 |
02/09/2001 | 0615 | 02/11/2001 | 1701 |
01/20/2001 | 0915 | 02/09/2001 | 0001 |
Subject: | SGP EBBR E9 - E9 EBBR AEM Hung, Data Incorrect |
DataStreams: | sgp30ebbrE9.b1, sgp5ebbrE9.b1, sgp15ebbrE9.b1 |
Description: | A worn AEM belt caused the AEM to hang frequently. Gradient measurements, sensible heat flux, latent heat flux, and Bowen ratio are incorrect during the indicated periods. The AEM hung on 2/9/01 because of icing from an ice storm. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/08/2001 | 1315 | 04/09/2001 | 0931 |
04/23/2001 | 1915 | 04/24/2001 | 0031 |
04/10/2001 | 1545 | 04/10/2001 | 1831 |
04/12/2001 | 1745 | 04/12/2001 | 1901 |
04/24/2001 | 1245 | 04/25/2001 | 0131 |
04/25/2001 | 1215 | 04/26/2001 | 0501 |
04/29/2001 | 0000 | 04/29/2001 | 0201 |
04/29/2001 | 0745 | 04/29/2001 | 0801 |
04/29/2001 | 1000 | 05/08/2001 | 1501 |
04/03/2001 | 1945 | 04/03/2001 | 2331 |
Subject: | SGP EBBR E9 - E9 EBBR AEM Hung, Data Incorrect |
DataStreams: | sgp5ebbrE9.b1, sgp30ebbrE9.b1, sgp15ebbrE9.b1 |
Description: | The AEM belt was worn and the AEM hung frequently. Gradients, sensible heat flux, latent heat flux, and Bowen ratio were incorrect during the indicated period. The AEM was replaced during the recalibration changeout on 5/11/01. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/28/2001 | 0530 | 01/29/2001 | 2031 |
02/09/2001 | 0430 | 02/11/2001 | 1531 |
02/14/2001 | 1735 | 02/16/2001 | 1831 |
Subject: | SGP EBBR E9 - E9 EBBR Wind Speed Frozen Stopped |
DataStreams: | sgp30ebbrE9.b1, sgp5ebbrE9.b1, sgp15ebbrE9.b1 |
Description: | An ice storm froze the wind speed sensor to a stop. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/28/2001 | 0800 | 01/29/2001 | 1701 |
02/14/2001 | 1735 | 02/16/2001 | 1831 |
02/09/2001 | 0430 | 02/10/2001 | 1801 |
Subject: | SGP EBBR E9 - E9 EBBR Wind Direction Frozen In One Direction |
DataStreams: | sgp5ebbrE9.b1, sgp30ebbrE9.b1, sgp15ebbrE9.b1 |
Description: | An ice storm froze the wind direction sensor in one postion. Th indicated measurements are incorrect. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/15/2001 | 2315 | 12/14/2001 | 2356 |
Subject: | SGP/EBBR/E9 - Reprocess: E9 EBBR Soil Temperature #5 Intermittently Incorrect |
DataStreams: | sgp30ebbrE9.b1, sgp15ebbrE9.b1 |
Description: | Soil moisture probe #5 in the recalibrated unit installed on 5/11/01 has not been working all of the time. At times the measurement has been very large and incorrect. The indicated measurements are incorrect at those times. The data must be inspected to determine the times that have incorrect data. The sensible and latent heat fluxes can be recalculated by using soil sets 1, 2, 3, and 4 only in a recalculation of the energy budget. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/19/2001 | 1500 | 07/19/2001 | 1631 |
Subject: | SGP EBBR E9 - E9 EBBR Gradients Incorrect During Sensor Replacement |
DataStreams: | sgp5ebbrE9.b1, sgp30ebbrE9.b1, sgp15ebbrE9.b1 |
Description: | During the replacement of the right air temperature probe, all gradient measurements were incorrect. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/15/2001 | 0830 | 07/15/2001 | 2216 |
07/16/2001 | 0530 | 07/16/2001 | 1946 |
07/14/2001 | 0230 | 07/14/2001 | 2016 |
07/17/2001 | 1300 | 07/17/2001 | 1316 |
07/18/2001 | 1230 | 07/18/2001 | 1431 |
07/19/2001 | 0715 | 07/19/2001 | 1616 |
Subject: | SGP EBBR E9 - E9 EBBR Right Air Temperature Incorrect |
DataStreams: | sgp5ebbrE9.b1, sgp15ebbrE9.b1, sgp30ebbrE9.b1 |
Description: | The right air temperature probe measurements were incorrect during these periods. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/01/2001 | 0230 | 02/02/2001 | 1601 |
02/09/2001 | 0330 | 02/11/2001 | 2001 |
02/15/2001 | 0830 | 02/15/2001 | 1531 |
02/16/2001 | 1431 | 02/16/2001 | 1901 |
02/19/2001 | 0900 | 03/24/2001 | 2301 |
02/16/2001 | 2100 | 02/18/2001 | 1831 |
Subject: | SGP/EEBR/E2 - AEM Stopped In One Position |
DataStreams: | sgp30ebbrE2.b1, sgp15ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | All gradient related measurements, including sensible and latent heat flux and bowen ratio, are incorrect. The AEM kept stopping in the home30 position (home signal about 23). |
Measurements: | sgp30ebbrE2.b1:
sgp15ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/26/2001 | 1505 | 06/07/2001 | 1501 |
06/20/2001 | 1646 | 07/04/2001 | 1101 |
07/05/2001 | 1800 | 07/30/2001 | 0201 |
07/30/2001 | 0345 | 09/08/2001 | 1701 |
09/13/2001 | 0530 | 09/13/2001 | 1531 |
09/13/2001 | 1900 | 09/27/2001 | 1416 |
10/02/2001 | 0945 | 01/12/2002 | 0001 |
Subject: | SGP/EEBR/E2 - AEM Stopped In One Position |
DataStreams: | sgp5ebbrE2.b1, sgp30ebbrE2.b1, sgp15ebbrE2.b1 |
Description: | The AEM kept stopping in one of three positions, none of which allow sensible heat flux, latent heat flux, or Bown ratio to be calculated correctly. These and the gradient measurements are incorrect for the period indicated. The AEM stopped in either the home15 position (home signal about 53), the home 30 position (home signal about 23), or inbetween these two (home signal 1 to 2). |
Measurements: | sgp30ebbrE2.b1:
sgp15ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/08/2001 | 1745 | 02/10/2001 | 1630 |
Subject: | SGP/EEBR/E2 - Wind Direction Frozen |
DataStreams: | sgp30ebbrE2.b1, sgp5ebbrE2.b1, sgp15ebbrE2.b1 |
Description: | The wind direction sensor froze stopped for the indicated period of time. The indicated measurements are incorrect. |
Measurements: | sgp30ebbrE2.b1:
sgp15ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/08/2001 | 1745 | 02/11/2001 | 1731 |
Subject: | SGP/EEBR/E2 - Wind Speed Sensor Frozen |
DataStreams: | sgp30ebbrE2.b1, sgp5ebbrE2.b1, sgp15ebbrE2.b1 |
Description: | The wind speed sensor was frozen stopped. The indicated measurements were incorrect. |
Measurements: | sgp15ebbrE2.b1:
sgp30ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/24/2001 | 0300 | 03/24/2001 | 0431 |
03/24/2001 | 0700 | 03/24/2001 | 1501 |
03/28/2001 | 0745 | 03/28/2001 | 1301 |
03/25/2001 | 0530 | 03/25/2001 | 1431 |
03/26/2001 | 0430 | 03/26/2001 | 1531 |
03/30/2001 | 0645 | 03/30/2001 | 1431 |
03/15/2001 | 1400 | 03/15/2001 | 1601 |
Subject: | SGP/EEBR/E2 - Soil Temperature #2 Incorrect |
DataStreams: | sgp30ebbrE2.b1, sgp15ebbrE2.b1 |
Description: | Soil temperature #2 was intermittently incorrect. The cause is not yet known. The indicated other measurements were also incorrect because soil temperature #2 was used in the calculations of the quantities. Soil moisture (30 minute only) was affected because version 6 of the EBBR datalogger program used a multiplication of soil temperature and soil moisture as part of the adjustment of soil moisture for soil temperature. This was discovered to be a poor technique soon after it was implemented, but E2 will continue to use version 6 until a replacement EBBR unit is available. Changeout dates of the EBBR units can be found on the EBBR Instrument web page. |
Measurements: | sgp15ebbrE2.b1:
sgp30ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/01/2001 | 0115 | 05/01/2001 | 0231 |
05/03/2001 | 1100 | 05/03/2001 | 1901 |
05/04/2001 | 1600 | 05/04/2001 | 1631 |
05/05/2001 | 1730 | 05/06/2001 | 0401 |
05/06/2001 | 2015 | 05/07/2001 | 0831 |
05/12/2001 | 1815 | 05/13/2001 | 0130 |
05/19/2001 | 1330 | 05/19/2001 | 2301 |
05/20/2001 | 0115 | 05/20/2001 | 0231 |
05/21/2001 | 0330 | 05/21/2001 | 0801 |
05/21/2001 | 1500 | 05/21/2001 | 1931 |
05/22/2001 | 0130 | 05/22/2001 | 1531 |
05/22/2001 | 1700 | 05/22/2001 | 1701 |
05/24/2001 | 0330 | 05/24/2001 | 1431 |
05/27/2001 | 0930 | 05/27/2001 | 1431 |
06/15/2001 | 0900 | 06/15/2001 | 1131 |
06/19/2001 | 1215 | 06/19/2001 | 1331 |
06/20/2001 | 0445 | 06/20/2001 | 0701 |
06/21/2001 | 1100 | 06/21/2001 | 1401 |
05/01/2001 | 0415 | 05/02/2001 | 1301 |
05/07/2001 | 1045 | 05/09/2001 | 0301 |
05/21/2001 | 0945 | 05/21/2001 | 1331 |
05/22/2001 | 1830 | 05/23/2001 | 2001 |
06/02/2001 | 1145 | 06/02/2001 | 1231 |
06/21/2001 | 1500 | 06/21/2001 | 1531 |
Subject: | SGP/EBBR/E2 - Reprocess: E2 EBBR Soil Temperature #2 Incorrect |
DataStreams: | sgp30ebbrE2.b1, sgp15ebbrE2.b1 |
Description: | Soil Temperature #2 was intermittently incorrect. The cause is unknown. The indicated measurements that are calculated using soil temperature are also incorrect. Soil moisture #2 (30 minute only) was also affected because version 6 of the EBBR datalogger program uses a multiplication of soil temperature and soil moisture as part of the adjustment of soil moisture for soil temperature. This was discovered to be a poor technique soon after it was implemented, but E2 will continue to use version 6 until a replacement EBBR unit is available. When soil temperature #2 is large negative, soil moisture #2 is often twice as large as it should be. EBBR unit changeout dates are listed on the EBBR instrument web page. |
Measurements: | sgp15ebbrE2.b1:
sgp30ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/22/2001 | 1100 | 06/22/2001 | 1331 |
06/26/2001 | 0845 | 06/27/2001 | 0331 |
06/27/2001 | 0545 | 06/30/2001 | 1701 |
07/04/2001 | 1130 | 07/05/2001 | 0231 |
07/12/2001 | 0300 | 07/12/2001 | 0401 |
07/12/2001 | 1115 | 07/12/2001 | 1531 |
07/14/2001 | 0230 | 07/14/2001 | 1501 |
07/18/2001 | 0230 | 07/18/2001 | 0301 |
07/19/2001 | 0200 | 07/19/2001 | 0301 |
07/19/2001 | 0545 | 07/19/2001 | 0801 |
07/19/2001 | 0900 | 07/19/2001 | 0931 |
07/19/2001 | 1045 | 07/19/2001 | 1431 |
07/21/2001 | 1730 | 07/21/2001 | 1801 |
08/03/2001 | 1500 | 08/04/2001 | 0101 |
08/04/2001 | 1715 | 08/04/2001 | 1831 |
09/28/2001 | 1930 | 09/29/2001 | 0001 |
10/16/2001 | 1915 | 10/16/2001 | 2001 |
10/16/2001 | 2315 | 10/17/2001 | 0001 |
10/20/2001 | 1545 | 10/20/2001 | 1631 |
11/01/2001 | 0115 | 11/01/2001 | 0131 |
11/02/2001 | 1530 | 11/02/2001 | 1731 |
11/05/2001 | 1600 | 11/05/2001 | 1901 |
11/11/2001 | 1015 | 11/11/2001 | 1201 |
11/11/2001 | 1430 | 11/11/2001 | 1531 |
12/06/2001 | 1715 | 12/06/2001 | 2331 |
12/17/2001 | 1345 | 12/17/2001 | 1531 |
12/22/2001 | 0115 | 12/22/2001 | 0401 |
12/22/2001 | 0800 | 12/22/2001 | 1001 |
12/22/2001 | 1100 | 12/22/2001 | 1331 |
12/27/2001 | 1800 | 12/27/2001 | 1901 |
12/28/2001 | 1830 | 12/28/2001 | 1931 |
12/28/2001 | 2345 | 12/29/2001 | 0031 |
01/05/2002 | 1730 | 01/05/2002 | 2301 |
01/08/2002 | 1615 | 01/08/2002 | 2101 |
01/11/2002 | 0115 | 01/11/2002 | 0201 |
01/11/2002 | 1530 | 01/11/2002 | 1831 |
07/12/2001 | 0500 | 07/12/2001 | 0531 |
07/18/2001 | 0815 | 07/18/2001 | 0901 |
07/20/2001 | 1445 | 07/20/2001 | 1931 |
09/28/2001 | 1830 | 09/28/2001 | 1831 |
11/10/2001 | 1730 | 11/10/2001 | 1901 |
12/27/2001 | 2030 | 12/28/2001 | 0031 |
01/07/2002 | 0930 | 01/07/2002 | 1031 |
06/22/2001 | 1445 | 06/22/2001 | 1501 |
10/31/2001 | 1415 | 10/31/2001 | 1731 |
12/21/2001 | 2000 | 12/21/2001 | 2031 |
Subject: | SGP/EBBR/E2 - Reprocess: E2 EBBR Soil Temperature #2 Incorrect |
DataStreams: | sgp30ebbrE2.b1, sgp15ebbrE2.b1 |
Description: | Soil Temperature #2 was intermittently incorrect. The cause is unknown. The indicated measurements that are calculated using soil temperature are also incorrect. Soil moisture #2 (30 minute only) was also affected because version 6 of the EBBR datalogger program uses a multiplication of soil temperature and soil moisture as part of the adjustment of soil moisture for soil temperature. This was discovered to be a poor technique soon after it was implemented, but E2 will continue to use version 6 until a replacement EBBR unit is available. When soil temperature #2 is large negative, soil moisture #2 is often twice as large as it should be. EBBR unit changeout dates are listed on the EBBR instrument web page. |
Measurements: | sgp15ebbrE2.b1:
sgp30ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/12/2001 | 0130 | 04/12/2001 | 1331 |
04/13/2001 | 0530 | 04/13/2001 | 1631 |
04/14/2001 | 0745 | 04/14/2001 | 1501 |
04/15/2001 | 0915 | 04/15/2001 | 1231 |
04/16/2001 | 0245 | 04/16/2001 | 0301 |
04/16/2001 | 0515 | 04/16/2001 | 2301 |
04/17/2001 | 0145 | 04/17/2001 | 0931 |
04/19/2001 | 0615 | 04/19/2001 | 1601 |
04/23/2001 | 0515 | 04/24/2001 | 0001 |
04/25/2001 | 1030 | 04/25/2001 | 1331 |
04/29/2001 | 1500 | 04/29/2001 | 1601 |
04/30/2001 | 0130 | 04/30/2001 | 0401 |
04/30/2001 | 1745 | 04/30/2001 | 2101 |
04/01/2001 | 1145 | 04/01/2001 | 1301 |
04/15/2001 | 1445 | 04/15/2001 | 1831 |
04/17/2001 | 1545 | 04/18/2001 | 1601 |
04/25/2001 | 0215 | 04/25/2001 | 0331 |
04/30/2001 | 0615 | 04/30/2001 | 1601 |
Subject: | SGP/EBBR/E2 - Reprocess: E2 EBBR Soil Temperature #2 Incorrect |
DataStreams: | sgp30ebbrE2.b1, sgp15ebbrE2.b1 |
Description: | Soil Temperature #2 was intermittently incorrect. The cause is unknown. The indicated measurements that are calculated using soil temperature are also incorrect. Soil moisture #2 (30 minute only) was also affected because version 6 of the EBBR datalogger program uses a multiplication of soil temperature and soil moisture as part of the adjustment of soil moisture for soil temperature. This was discovered to be a poor technique soon after it was implemented, but E2 will continue to use version 6 until a replacement EBBR unit is available. When soil temperature #2 is large negative, soil moisture #2 is often twice as large as it should be. EBBR unit changeout dates are listed on the EBBR instrument web page. |
Measurements: | sgp15ebbrE2.b1:
sgp30ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/11/2001 | 1508 | 10/11/2001 | 1931 |
Subject: | SGP EBBR E15 - E15 EBBR Recalibration Changeout |
DataStreams: | sgp30ebbrE15.b1, sgp15ebbrE15.b1, sgp5ebbrE15.b1, sgp5ebbrE15.00 |
Description: | During the indicated period the EBBR unit was being changed out with a recalibrated unit. All data is incorrect or missing. Unit #14 was removed; #6 was installed. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp5ebbrE15.00:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/05/2001 | 1700 | 01/18/2002 | 0001 |
Subject: | SGP/EBBR/E13 - Reprocess: E13 EBBR Soil heat Flows #1 and #2 Incorrect |
DataStreams: | sgp30ebbrE13.b1, sgp15ebbrE13.b1 |
Description: | Soil Heat Flow probes #1 and #2 were incorrect during the indicated period (and for some time afterward, to be covered in another DQR). Outputs from the probes were typically too large negative or too large positive. |
Measurements: | sgp15ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/05/2001 | 1715 | 10/05/2001 | 1731 |
Subject: | SGP EBBR E13 - E13 EBBR Data Incorrect |
DataStreams: | sgp30ebbrE13.b1, sgp15ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | The 5 and 15 minute data was zero and thus the thirty minute data was also incorrect. This occurs after the restart of an EBBR system, in this case after a recalibration unit changeout. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/08/2001 | 2000 | 10/12/2001 | 1731 |
Subject: | SGP EBBR E13 - E13 EBBR Left T/RH Probe Temperature Incorrect |
DataStreams: | sgp30ebbrE13.b1, sgp15ebbrE13.b1 |
Description: | The left T/RH Probe Temperature was often incorrect, causing the other indicated measurements to be incorrect also. |
Measurements: | sgp15ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/12/2001 | 1735 | 10/18/2001 | 1601 |
01/06/2002 | 1930 | 01/06/2002 | 2101 |
01/07/2002 | 2330 | 01/07/2002 | 2331 |
01/09/2002 | 0330 | 01/18/2002 | 0001 |
01/05/2002 | 1930 | 01/05/2002 | 1931 |
11/22/2001 | 0130 | 11/25/2001 | 1531 |
01/08/2002 | 1730 | 01/08/2002 | 1801 |
Subject: | SGP EBBR E13 - E13 EBBR Left T/RH Probe Temperature Incorrect |
DataStreams: | sgp30ebbrE13.b1, sgp15ebbrE13.b1 |
Description: | The left T/RH Probe Temperature was often incorrect, causing the other indicated measurements to be incorrect also. When the home30 signal had an offset, the top RH probe temperature was incorrect but not the bottom RH probe temperature. |
Measurements: | sgp15ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/25/2001 | 1545 | 12/28/2001 | 1931 |
10/09/2001 | 0245 | 10/09/2001 | 2145 |
Subject: | SGP EBBR E13 - E13 EBBR AEM in Mid Position |
DataStreams: | sgp30ebbrE13.b1, sgp15ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | The AEM stopped before reaching the end of it's travel, causing the home signals to be between 0 and 3 during much of this period (therefore the data is colored yellow). All gradient data, sensible heat flux, latent heat flux, and the Bowen ratio are incorrect during this condition. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/28/2001 | 1930 | 01/18/2002 | 0001 |
10/12/2001 | 1315 | 11/25/2001 | 1541 |
Subject: | SGP EBBR E13 - E13 EBBR AEM Home 30 Offset |
DataStreams: | sgp30ebbrE13.b1, sgp15ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | The AEM home30 signal is offset by approximately 25; it appears that the electronics of the AEM is adding the home15 and home30 voltages together when the AEM is in the home30 position. All gradient data, sensible heat flux, latent heat flux, and the Bowen ratio are incorrect during this condition. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/09/2001 | 2245 | 10/12/2001 | 1301 |
Subject: | SGP EBBR E13 - E13 EBBR AEM Home 30 Offset or AEM in Mid Position |
DataStreams: | sgp30ebbrE13.b1, sgp15ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | The AEM stopped before reaching the end of it's travel, causing the home signals to be between 0 and 3 during part of this period. The AEM home30 signal is offset by approximately 25 during part of this period; it appears that the electronics of the AEM is adding the home15 and home30 voltages together when the AEM is in the home30 position. All gradient data, sensible heat flux, latent heat flux, and the Bowen ratio are incorrect during these conditions. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/01/2001 | 0000 | 10/05/2001 | 1401 |
Subject: | SGP EBBR E13 - E13 EBBR Left Air Temperature Incorrect |
DataStreams: | sgp30ebbrE13.b1, sgp15ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | This condition occurred most of the time during the indicated period (hence color yellow). The left air temperature was incorrect only when the AEM was in the home 30 position. This only affected the bottom air temperature calculation. However, it made the latent heat flux, sensible heat flux, and Bowen ratio incorrect. The sensor was improperly inserted into the aspirator shield and was pinched between the shield and the T/RH probe. When the AEM was in the home30 position, the aspirator hose position may have caused the sensor to short to ground. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/25/2001 | 0530 | 03/01/2001 | 2356 |
Subject: | SGP EBBR E13 - E13 EBBR Left Air Temperature Incorrect |
DataStreams: | sgp30ebbrE13.b1, sgp15ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | This condition occurred most of the time during the indicated period (hence color yellow). The left air temperature was incorrect only when the AEM was in the home 30 position. This only affected the bottom air temperature calculation. However, it made the latent heat flux, sensible heat flux, and Bowen ratio incorrect. The sensor was improperly inserted into the aspirator shield and was pinched between the shield and the T/RH probe. When the AEM was in the home30 position, the aspirator hose position may have caused the sensor to short to ground. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/08/2001 | 1935 | 03/31/2001 | 2356 |
Subject: | SGP/EBBR/E13 - Reprocess: E13 EBBR Soil Temperature #4 Incorrect |
DataStreams: | sgp30ebbrE13.b1, sgp15ebbrE13.b1 |
Description: | Soil Temperature #4 was often incorrect, typically too large negative, during this period, causing other measurements calculated with it to be incorrect also. This was caused by a bad multiplexor channel, which was discovered by the vendor during recalibration of the unit later. |
Measurements: | sgp15ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/01/2001 | 0000 | 10/05/2001 | 1401 |
Subject: | SGP/EBBR/E13 - Reprocess: E13 EBBR Soil Temperature #4 Incorrect |
DataStreams: | sgp30ebbrE13.b1, sgp15ebbrE13.b1 |
Description: | Soil Temperature #4 was often incorrect, typically too large negative, during this period, causing other measurements calculated with it to be incorrect also. This was caused by a bad multiplexor channel, which was discovered by the vendor during recalibration of the unit later. |
Measurements: | sgp15ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/25/2001 | 1545 | 12/28/2001 | 1931 |
Subject: | SGP/EBBR/E13 - Left T/RH Probe Temperature Incorrect |
DataStreams: | sgp30ebbrE13.b1, sgp15ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | During much of the indicated period the Left T/RH Temperature was incorrect. Because the home signal was between 0 and 3, the temperature was incorrect when the AEM was at the bottom position, but was correct when the AEM was at the top position. This was caused by a short created when the sensor was mispositioned in the aspirator housing and enough strain on the cable for it to bring it into contact with the housing in the bottom position. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/18/2002 | 0005 | 01/25/2002 | 2201 |
Subject: | SGP/EBBR/E13 - Left Humidity Probe Temperature Incorrect |
DataStreams: | sgp30ebbrE13.b1, sgp15ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | The left humidity probe temperature was incorrect much of the time, as was the top humidity temperature and vapor pressure. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/25/2002 | 2200 | 01/25/2002 | 2201 |
Subject: | SGP/EBBR/E13 - Data incorrect during maintenance |
DataStreams: | sgp30ebbrE13.b1, sgp15ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | The 5, 15, and 30 minute measurements were zero or offscale at 2200 GMT only. Electronic connections were being checked at that time. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/25/2002 | 2010 | 01/25/2002 | 2101 |
Subject: | SGP/EBBR/E13 - Data Incorrect During 6 Month Checks |
DataStreams: | sgp30ebbrE13.b1, sgp15ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | The measurements were incorrect during the 6 month checks. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/13/2001 | 1700 | 11/13/2001 | 1801 |
Subject: | SGP/EBBR/E15 - Data Incorrect During 6 Month Checks |
DataStreams: | sgp30ebbrE15.b1, sgp15ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | The data was incorrect during the 6 month checks. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/14/1998 | 0700 | 01/14/1998 | 1845 |
Subject: | SGP/EBBR/E2 - EBBR E2 Wind Speed Sensors Frozen |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | Wind speed sensors were frozen stalled during this period. Wind speed was frozen from 0700 through 1845 GMT. Data values that were affected were: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp15ebbrE2.b1:
sgp30ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/14/1995 | 1400 | 10/14/1995 | 1700 |
Subject: | SGP/EBBR/E12 - 3-month Check |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during this period, at least, do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/15/1994 | 1605 | 11/15/1994 | 1800 |
Subject: | SGP/EBBR/E7 - T/RH Check |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | During the month of November 1994, site operations personnel conducted field comparisons of the temperature and relative humidity sensors in the EBBRs and the HMI31 portable T/RH meter (with the removable probe). As is reflected in the home signal output, the automatic exchange mechanism was disabled in order to conduct the checks. Measurements were taken every 5 minutes, for a half hour each in the aspirated radiation shields (the HMI31 was inserted into the EBBR aspirated radiation shield to provide the same aspiration as the EBBR sensors received). Measurements of aspirator flow were also recorded. This information was provided to the mentor. The mentor then compared the field checks with the data recorded by the EBBRs. The periods of data and data fields that are incorrect because of the checks are listed below. Times are those at which the data are reported. Data Fields 5 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot 15 minute: rr_thum_r, rr_thum_l, mv_hum_r, mv_hum_l, tair_r, tair_l 30 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, bowen, e, h 5 minute: 1605, 1610, 1615, 1620, 1625, 1630, 1635, 1640, 1645, 1650, 1655, 1700, 1705, 1710, 1715, 1720, 1725, 1730, 1735, 1740, 1745, 1750 15 minute: 1615, 1630, 1645, 1700, 1715, 1730, 1745, 1800 30 minute: 1630, 1700, 1730, 1800 (EDITOR'S NOTE: The following analysis refers to all EBBRs installed in Nov 1994) Results of the T/RH comparisons: The combined accuracy of the HMI31 and EBBR temperature sensors (thermocouples or PRTDs) is +/- 0.4 degrees C. The combined accuracy of the HMI31 and EBBR RH sensors is +/- 4% for RH less than 80%, and +/- 6% for RH greater than 80%. The combined accuracy for each pair of EBBR sensors of the same type is essentially the same as the quantities stated above. All of the pairs of EBBR sensors show differences that fall within the stated combined accuracies. Although this sounds wonderful, a few of the differences are close to the combined accuracy, which, because of the small differences in temperature being measured (normally smaller than the magnitude of the combined accuracy) could lead to some incorrect estimation of sensible and latent heat flux. However, unless there would be substantial differences of calibration slope of two sensors in a pair, the switching of height of the sensors every 15 minutes by the automatic exchange mechanism will remove the offset between the sensors. The comparisons between the EBBR sensors and the HMI31 T/RH meter showed much larger differences. The range of differences were: -2.4 to 0.0 degrees for the thermocouples, -2.1 to 0.1 degrees for the RH probe PRTDs, and -6.6 to 6.7 percent for the RH sensors. Six Thermocouples indicated differences from the HMI31 greater than the combined accuracy. Seven PRTDs indicated differences from the HMI31 greater than the combined accuracy. Four RH sensors indicated differences from the HMI31 greater than the combined accuracy. The temperature differences (for both thermocouples and PRTDs) were nearly all of one sign; the HMI31 indicated the greater temperature, which may suggest a bias in the HMI31 temperature measurement, even though a dozen differences of +/- 0.2 degrees were measured. Four differences of zero and only one positive difference (out of 40) were measured. Thirteen of the 20 RH differences were positive (HMI31 indicating the lower RH), with the differences near the level of the combined accuracies being approximately 50% positive and 50% negative. It does not appear that the HMI31 has a RH bias. Plots of EBBR sensor/HMI31 differences as a function of time from the installation of the individual sensors (a range of 10 to 30 months in the field) shows no obvious pattern of sensor drift with time. This is encouraging. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/02/1994 | 1640 | 11/02/1994 | 1900 |
Subject: | SGP/EBBR/E8 - T/RH Check |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | During the month of November 1994, site operations personnel conducted field comparisons of the temperature and relative humidity sensors in the EBBRs and the HMI31 portable T/RH meter (with the removable probe). As is reflected in the home signal output, the automatic exchange mechanism was disabled in order to conduct the checks. Measurements were taken every 5 minutes, for a half hour each in the aspirated radiation shields (the HMI31 was inserted into the EBBR aspirated radiation shield to provide the same aspiration as the EBBR sensors received). Measurements of aspirator flow were also recorded. This information was provided to the mentor. The mentor then compared the field checks with the data recorded by the EBBRs. The periods of data and data fields that are incorrect because of the checks are listed below. Times are those at which the data are reported. Data Fields 5 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot 15 minute: rr_thum_r, rr_thum_l, mv_hum_r, mv_hum_l, tair_r, tair_l 30 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, bowen, e, h 5 minute: 1640, 1645, 1650, 1655, 1700, 1705, 1710, 1715, 1720, 1725, 1730, 1735, 1740, 1745, 1750, 1755, 1800, 1805, 1810, 1815, 1820, 1825, 1830, 1835 15 minute: 1645, 1700, 1715, 1730, 1745, 1800, 1815, 1830, 1845 30 minute: 1700, 1730, 1800, 1830, 1900 (EDITOR'S NOTE: The following analysis referrs to all EBBRs installed in Nov 1994) Results of the T/RH comparisons: The combined accuracy of the HMI31 and EBBR temperature sensors (thermocouples or PRTDs) is +/- 0.4 degrees C. The combined accuracy of the HMI31 and EBBR RH sensors is +/- 4% for RH less than 80%, and +/- 6% for RH greater than 80%. The combined accuracy for each pair of EBBR sensors of the same type is essentially the same as the quantities stated above. All of the pairs of EBBR sensors show differences that fall within the stated combined accuracies. Although this sounds wonderful, a few of the differences are close to the combined accuracy, which, because of the small differences in temperature being measured (normally smaller than the magnitude of the combined accuracy) could lead to some incorrect estimation of sensible and latent heat flux. However, unless there would be substantial differences of calibration slope of two sensors in a pair, the switching of height of the sensors every 15 minutes by the automatic exchange mechanism will remove the offset between the sensors. The comparisons between the EBBR sensors and the HMI31 T/RH meter showed much larger differences. The range of differences were: -2.4 to 0.0 degrees for the thermocouples, -2.1 to 0.1 degrees for the RH probe PRTDs, and -6.6 to 6.7 percent for the RH sensors. Six Thermocouples indicated differences from the HMI31 greater than the combined accuracy. Seven PRTDs indicated differences from the HMI31 greater than the combined accuracy. Four RH sensors indicated differences from the HMI31 greater than the combined accuracy. The temperature differences (for both thermocouples and PRTDs) were nearly all of one sign; the HMI31 indicated the greater temperature, which may suggest a bias in the HMI31 temperature measurement, even though a dozen differences of +/- 0.2 degrees were measured. Four differences of zero and only one positive difference (out of 40) were measured. Thirteen of the 20 RH differences were positive (HMI31 indicating the lower RH), with the differences near the level of the combined accuracies being approximately 50% positive and 50% negative. It does not appear that the HMI31 has a RH bias. Plots of EBBR sensor/HMI31 differences as a function of time from the installation of the individual sensors (a range of 10 to 30 months in the field) shows no obvious pattern of sensor drift with time. This is encouraging. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/14/1994 | 1820 | 11/14/1994 | 2000 |
Subject: | SGP/EBBR/E9 - T/RH Check |
DataStreams: | sgp30ebbrE9.b1, sgp5ebbrE9.b1, sgp15ebbrE9.b1 |
Description: | During the month of November 1994, site operations personnel conducted field comparisons of the temperature and relative humidity sensors in the EBBRs and the HMI31 portable T/RH meter (with the removable probe). As is reflected in the home signal output, the automatic exchange mechanism was disabled in order to conduct the checks. Measurements were taken every 5 minutes, for a half hour each in the aspirated radiation shields (the HMI31 was inserted into the EBBR aspirated radiation shield to provide the same aspiration as the EBBR sensors received). Measurements of aspirator flow were also recorded. This information was provided to the mentor. The mentor then compared the field checks with the data recorded by the EBBRs. The periods of data and data fields that are incorrect because of the checks are listed below. Times are those at which the data are reported. Data Fields 5 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot 15 minute: rr_thum_r, rr_thum_l, mv_hum_r, mv_hum_l, tair_r, tair_l 30 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, bowen, e, h 5 minute: 1820, 1825, 1830, 1835, 1840, 1845, 1850, 1855, 1900, 1905, 1910, 1915, 1920, 1925, 1930, 1935, 1940, 1945 15 minute: 1830, 1845, 1900, 1915, 1930, 1945 30 minute: 1830, 1900, 1930, 2000 (EDITOR'S NOTE: The following analysis refers to all EBBRs installed in Nov 1994) Results of the T/RH comparisons: The combined accuracy of the HMI31 and EBBR temperature sensors (thermocouples or PRTDs) is +/- 0.4 degrees C. The combined accuracy of the HMI31 and EBBR RH sensors is +/- 4% for RH less than 80%, and +/- 6% for RH greater than 80%. The combined accuracy for each pair of EBBR sensors of the same type is essentially the same as the quantities stated above. All of the pairs of EBBR sensors show differences that fall within the stated combined accuracies. Although this sounds wonderful, a few of the differences are close to the combined accuracy, which, because of the small differences in temperature being measured (normally smaller than the magnitude of the combined accuracy) could lead to some incorrect estimation of sensible and latent heat flux. However, unless there would be substantial differences of calibration slope of two sensors in a pair, the switching of height of the sensors every 15 minutes by the automatic exchange mechanism will remove the offset between the sensors. The comparisons between the EBBR sensors and the HMI31 T/RH meter showed much larger differences. The range of differences were: -2.4 to 0.0 degrees for the thermocouples, -2.1 to 0.1 degrees for the RH probe PRTDs, and -6.6 to 6.7 percent for the RH sensors. Six Thermocouples indicated differences from the HMI31 greater than the combined accuracy. Seven PRTDs indicated differences from the HMI31 greater than the combined accuracy. Four RH sensors indicated differences from the HMI31 greater than the combined accuracy. The temperature differences (for both thermocouples and PRTDs) were nearly all of one sign; the HMI31 indicated the greater temperature, which may suggest a bias in the HMI31 temperature measurement, even though a dozen differences of +/- 0.2 degrees were measured. Four differences of zero and only one positive difference (out of 40) were measured. Thirteen of the 20 RH differences were positive (HMI31 indicating the lower RH), with the differences near the level of the combined accuracies being approximately 50% positive and 50% negative. It does not appear that the HMI31 has a RH bias. Plots of EBBR sensor/HMI31 differences as a function of time from the installation of the individual sensors (a range of 10 to 30 months in the field) shows no obvious pattern of sensor drift with time. This is encouraging. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/23/1994 | 1515 | 11/23/1994 | 1630 |
Subject: | SGP/EBBR/E12 - T/RH Check |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | During the month of November 1994, site operations personnel conducted field comparisons of the temperature and relative humidity sensors in the EBBRs and the HMI31 portable T/RH meter (with the removable probe). As is reflected in the home signal output, the automatic exchange mechanism was disabled in order to conduct the checks. Measurements were taken every 5 minutes, for a half hour each in the aspirated radiation shields (the HMI31 was inserted into the EBBR aspirated radiation shield to provide the same aspiration as the EBBR sensors received). Measurements of aspirator flow were also recorded. This information was provided to the mentor. The mentor then compared the field checks with the data recorded by the EBBRs. The periods of data and data fields that are incorrect because of the checks are listed below. Times are those at which the data are reported. Data Fields 5 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot 15 minute: rr_thum_r, rr_thum_l, mv_hum_r, mv_hum_l, tair_r, tair_l 30 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, bowen, e, h 5 minute: 1515, 1520, 1525, 1530, 1535, 1540, 1545, 1550, 1555, 1600, 1605, 1610 15 minute: 1515, 1530, 1545, 1600, 1615 30 minute: 1530, 1600, 1630 (EDITOR'S NOTE: The following analysis refers to all EBBRs installed in Nov 1994) Results of the T/RH comparisons: The combined accuracy of the HMI31 and EBBR temperature sensors (thermocouples or PRTDs) is +/- 0.4 degrees C. The combined accuracy of the HMI31 and EBBR RH sensors is +/- 4% for RH less than 80%, and +/- 6% for RH greater than 80%. The combined accuracy for each pair of EBBR sensors of the same type is essentially the same as the quantities stated above. All of the pairs of EBBR sensors show differences that fall within the stated combined accuracies. Although this sounds wonderful, a few of the differences are close to the combined accuracy, which, because of the small differences in temperature being measured (normally smaller than the magnitude of the combined accuracy) could lead to some incorrect estimation of sensible and latent heat flux. However, unless there would be substantial differences of calibration slope of two sensors in a pair, the switching of height of the sensors every 15 minutes by the automatic exchange mechanism will remove the offset between the sensors. The comparisons between the EBBR sensors and the HMI31 T/RH meter showed much larger differences. The range of differences were: -2.4 to 0.0 degrees for the thermocouples, -2.1 to 0.1 degrees for the RH probe PRTDs, and -6.6 to 6.7 percent for the RH sensors. Six Thermocouples indicated differences from the HMI31 greater than the combined accuracy. Seven PRTDs indicated differences from the HMI31 greater than the combined accuracy. Four RH sensors indicated differences from the HMI31 greater than the combined accuracy. The temperature differences (for both thermocouples and PRTDs) were nearly all of one sign; the HMI31 indicated the greater temperature, which may suggest a bias in the HMI31 temperature measurement, even though a dozen differences of +/- 0.2 degrees were measured. Four differences of zero and only one positive difference (out of 40) were measured. Thirteen of the 20 RH differences were positive (HMI31 indicating the lower RH), with the differences near the level of the combined accuracies being approximately 50% positive and 50% negative. It does not appear that the HMI31 has a RH bias. Plots of EBBR sensor/HMI31 differences as a function of time from the installation of the individual sensors (a range of 10 to 30 months in the field) shows no obvious pattern of sensor drift with time. This is encouraging. |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/04/1994 | 2005 | 11/04/1994 | 2100 |
Subject: | SGP/EBBR/E13 - T/RH Check |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | During the month of November 1994, site operations personnel conducted field comparisons of the temperature and relative humidity sensors in the EBBRs and the HMI31 portable T/RH meter (with the removable probe). As is reflected in the home signal output, the automatic exchange mechanism was disabled in order to conduct the checks. Measurements were taken every 5 minutes, for a half hour each in the aspirated radiation shields (the HMI31 was inserted into the EBBR aspirated radiation shield to provide the same aspiration as the EBBR sensors received). Measurements of aspirator flow were also recorded. This information was provided to the mentor. The mentor then compared the field checks with the data recorded by the EBBRs. The periods of data and data fields that are incorrect because of the checks are listed below. Times are those at which the data are reported. Data Fields 5 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot 15 minute: rr_thum_r, rr_thum_l, mv_hum_r, mv_hum_l, tair_r, tair_l 30 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, bowen, e, h 5 minute: 2005, 2010, 2015, 2020, 2025, 2030, 2035, 2040, 2045, 2050, 2055, 2100, 2105 15 minute: 2015, 2045, 2100, 2115 30 minute: 2030, 2100 (EDITOR'S NOTE: The following analysis refers to all EBBRs installed in Nov 1994) Results of the T/RH comparisons: The combined accuracy of the HMI31 and EBBR temperature sensors (thermocouples or PRTDs) is +/- 0.4 degrees C. The combined accuracy of the HMI31 and EBBR RH sensors is +/- 4% for RH less than 80%, and +/- 6% for RH greater than 80%. The combined accuracy for each pair of EBBR sensors of the same type is essentially the same as the quantities stated above. All of the pairs of EBBR sensors show differences that fall within the stated combined accuracies. Although this sounds wonderful, a few of the differences are close to the combined accuracy, which, because of the small differences in temperature being measured (normally smaller than the magnitude of the combined accuracy) could lead to some incorrect estimation of sensible and latent heat flux. However, unless there would be substantial differences of calibration slope of two sensors in a pair, the switching of height of the sensors every 15 minutes by the automatic exchange mechanism will remove the offset between the sensors. The comparisons between the EBBR sensors and the HMI31 T/RH meter showed much larger differences. The range of differences were: -2.4 to 0.0 degrees for the thermocouples, -2.1 to 0.1 degrees for the RH probe PRTDs, and -6.6 to 6.7 percent for the RH sensors. Six Thermocouples indicated differences from the HMI31 greater than the combined accuracy. Seven PRTDs indicated differences from the HMI31 greater than the combined accuracy. Four RH sensors indicated differences from the HMI31 greater than the combined accuracy. The temperature differences (for both thermocouples and PRTDs) were nearly all of one sign; the HMI31 indicated the greater temperature, which may suggest a bias in the HMI31 temperature measurement, even though a dozen differences of +/- 0.2 degrees were measured. Four differences of zero and only one positive difference (out of 40) were measured. Thirteen of the 20 RH differences were positive (HMI31 indicating the lower RH), with the differences near the level of the combined accuracies being approximately 50% positive and 50% negative. It does not appear that the HMI31 has a RH bias. Plots of EBBR sensor/HMI31 differences as a function of time from the installation of the individual sensors (a range of 10 to 30 months in the field) shows no obvious pattern of sensor drift with time. This is encouraging. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/07/1994 | 1650 | 11/07/1994 | 1830 |
Subject: | SGP/EBBR/E15 - T/RH Check |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | During the month of November 1994, site operations personnel conducted field comparisons of the temperature and relative humidity sensors in the EBBRs and the HMI31 portable T/RH meter (with the removable probe). As is reflected in the home signal output, the automatic exchange mechanism was disabled in order to conduct the checks. Measurements were taken every 5 minutes, for a half hour each in the aspirated radiation shields (the HMI31 was inserted into the EBBR aspirated radiation shield to provide the same aspiration as the EBBR sensors received). Measurements of aspirator flow were also recorded. This information was provided to the mentor. The mentor then compared the field checks with the data recorded by the EBBRs. The periods of data and data fields that are incorrect because of the checks are listed below. Times are those at which the data are reported. Data Fields 5 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot 15 minute: rr_thum_r, rr_thum_l, mv_hum_r, mv_hum_l, tair_r, tair_l 30 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, bowen, e, h 5 minute: 1650, 1655, 1700, 1705, 1710, 1715, 1720, 1725, 1730, 1735, 1740, 1745, 1750, 1755, 1800, 1805 15 minute: 1700, 1715, 1730, 1745, 1800, 1815 30 minute: 1700, 1730, 1800, 1830 (EDITOR'S NOTE: The following analysis refers to all EBBRs installed in Nov 1994) Results of the T/RH comparisons: The combined accuracy of the HMI31 and EBBR temperature sensors (thermocouples or PRTDs) is +/- 0.4 degrees C. The combined accuracy of the HMI31 and EBBR RH sensors is +/- 4% for RH less than 80%, and +/- 6% for RH greater than 80%. The combined accuracy for each pair of EBBR sensors of the same type is essentially the same as the quantities stated above. All of the pairs of EBBR sensors show differences that fall within the stated combined accuracies. Although this sounds wonderful, a few of the differences are close to the combined accuracy, which, because of the small differences in temperature being measured (normally smaller than the magnitude of the combined accuracy) could lead to some incorrect estimation of sensible and latent heat flux. However, unless there would be substantial differences of calibration slope of two sensors in a pair, the switching of height of the sensors every 15 minutes by the automatic exchange mechanism will remove the offset between the sensors. The comparisons between the EBBR sensors and the HMI31 T/RH meter showed much larger differences. The range of differences were: -2.4 to 0.0 degrees for the thermocouples, -2.1 to 0.1 degrees for the RH probe PRTDs, and -6.6 to 6.7 percent for the RH sensors. Six Thermocouples indicated differences from the HMI31 greater than the combined accuracy. Seven PRTDs indicated differences from the HMI31 greater than the combined accuracy. Four RH sensors indicated differences from the HMI31 greater than the combined accuracy. The temperature differences (for both thermocouples and PRTDs) were nearly all of one sign; the HMI31 indicated the greater temperature, which may suggest a bias in the HMI31 temperature measurement, even though a dozen differences of +/- 0.2 degrees were measured. Four differences of zero and only one positive difference (out of 40) were measured. Thirteen of the 20 RH differences were positive (HMI31 indicating the lower RH), with the differences near the level of the combined accuracies being approximately 50% positive and 50% negative. It does not appear that the HMI31 has a RH bias. Plots of EBBR sensor/HMI31 differences as a function of time from the installation of the individual sensors (a range of 10 to 30 months in the field) shows no obvious pattern of sensor drift with time. This is encouraging. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/10/1994 | 1520 | 11/10/1994 | 1630 |
Subject: | SGP/EBBR/E20 - T/RH Check |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | During the month of November 1994, site operations personnel conducted field comparisons of the temperature and relative humidity sensors in the EBBRs and the HMI31 portable T/RH meter (with the removable probe). As is reflected in the home signal output, the automatic exchange mechanism was disabled in order to conduct the checks. Measurements were taken every 5 minutes, for a half hour each in the aspirated radiation shields (the HMI31 was inserted into the EBBR aspirated radiation shield to provide the same aspiration as the EBBR sensors received). Measurements of aspirator flow were also recorded. This information was provided to the mentor. The mentor then compared the field checks with the data recorded by the EBBRs. The periods of data and data fields that are incorrect because of the checks are listed below. Times are those at which the data are reported. Data Fields 5 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot 15 minute: rr_thum_r, rr_thum_l, mv_hum_r, mv_hum_l, tair_r, tair_l 30 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, bowen, e, h 5 minute: 1520, 1525, 1530, 1535, 1540, 1545, 1550, 1555, 1600, 1605 15 minute: 1530, 1545, 1600, 1615 30 minute: 1530, 1600, 1630 (EDITOR'S NOTE: The following analysis refers to all EBBRs installed in Nov 1994) Results of the T/RH comparisons: The combined accuracy of the HMI31 and EBBR temperature sensors (thermocouples or PRTDs) is +/- 0.4 degrees C. The combined accuracy of the HMI31 and EBBR RH sensors is +/- 4% for RH less than 80%, and +/- 6% for RH greater than 80%. The combined accuracy for each pair of EBBR sensors of the same type is essentially the same as the quantities stated above. All of the pairs of EBBR sensors show differences that fall within the stated combined accuracies. Although this sounds wonderful, a few of the differences are close to the combined accuracy, which, because of the small differences in temperature being measured (normally smaller than the magnitude of the combined accuracy) could lead to some incorrect estimation of sensible and latent heat flux. However, unless there would be substantial differences of calibration slope of two sensors in a pair, the switching of height of the sensors every 15 minutes by the automatic exchange mechanism will remove the offset between the sensors. The comparisons between the EBBR sensors and the HMI31 T/RH meter showed much larger differences. The range of differences were: -2.4 to 0.0 degrees for the thermocouples, -2.1 to 0.1 degrees for the RH probe PRTDs, and -6.6 to 6.7 percent for the RH sensors. Six Thermocouples indicated differences from the HMI31 greater than the combined accuracy. Seven PRTDs indicated differences from the HMI31 greater than the combined accuracy. Four RH sensors indicated differences from the HMI31 greater than the combined accuracy. The temperature differences (for both thermocouples and PRTDs) were nearly all of one sign; the HMI31 indicated the greater temperature, which may suggest a bias in the HMI31 temperature measurement, even though a dozen differences of +/- 0.2 degrees were measured. Four differences of zero and only one positive difference (out of 40) were measured. Thirteen of the 20 RH differences were positive (HMI31 indicating the lower RH), with the differences near the level of the combined accuracies being approximately 50% positive and 50% negative. It does not appear that the HMI31 has a RH bias. Plots of EBBR sensor/HMI31 differences as a function of time from the installation of the individual sensors (a range of 10 to 30 months in the field) shows no obvious pattern of sensor drift with time. This is encouraging. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/08/1994 | 1650 | 11/08/1994 | 1800 |
Subject: | SGP/EBBR/E22 - T/RH Check |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | During the month of November 1994, site operations personnel conducted field comparisons of the temperature and relative humidity sensors in the EBBRs and the HMI31 portable T/RH meter (with the removable probe). As is reflected in the home signal output, the automatic exchange mechanism was disabled in order to conduct the checks. Measurements were taken every 5 minutes, for a half hour each in the aspirated radiation shields (the HMI31 was inserted into the EBBR aspirated radiation shield to provide the same aspiration as the EBBR sensors received). Measurements of aspirator flow were also recorded. This information was provided to the mentor. The mentor then compared the field checks with the data recorded by the EBBRs. The periods of data and data fields that are incorrect because of the checks are listed below. Times are those at which the data are reported. Data Fields 5 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot 15 minute: rr_thum_r, rr_thum_l, mv_hum_r, mv_hum_l, tair_r, tair_l 30 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, bowen, e, h 5 minute: 1650, 1655, 1700, 1705, 1710, 1715, 1720, 1725, 1730, 1735 15 minute: 1700, 1715, 1730, 1745 30 minute: 1700, 1730, 1800 (EDITOR'S NOTE: The following analysis refers to all EBBRs installed in Nov 1994) Results of the T/RH comparisons: The combined accuracy of the HMI31 and EBBR temperature sensors (thermocouples or PRTDs) is +/- 0.4 degrees C. The combined accuracy of the HMI31 and EBBR RH sensors is +/- 4% for RH less than 80%, and +/- 6% for RH greater than 80%. The combined accuracy for each pair of EBBR sensors of the same type is essentially the same as the quantities stated above. All of the pairs of EBBR sensors show differences that fall within the stated combined accuracies. Although this sounds wonderful, a few of the differences are close to the combined accuracy, which, because of the small differences in temperature being measured (normally smaller than the magnitude of the combined accuracy) could lead to some incorrect estimation of sensible and latent heat flux. However, unless there would be substantial differences of calibration slope of two sensors in a pair, the switching of height of the sensors every 15 minutes by the automatic exchange mechanism will remove the offset between the sensors. The comparisons between the EBBR sensors and the HMI31 T/RH meter showed much larger differences. The range of differences were: -2.4 to 0.0 degrees for the thermocouples, -2.1 to 0.1 degrees for the RH probe PRTDs, and -6.6 to 6.7 percent for the RH sensors. Six Thermocouples indicated differences from the HMI31 greater than the combined accuracy. Seven PRTDs indicated differences from the HMI31 greater than the combined accuracy. Four RH sensors indicated differences from the HMI31 greater than the combined accuracy. The temperature differences (for both thermocouples and PRTDs) were nearly all of one sign; the HMI31 indicated the greater temperature, which may suggest a bias in the HMI31 temperature measurement, even though a dozen differences of +/- 0.2 degrees were measured. Four differences of zero and only one positive difference (out of 40) were measured. Thirteen of the 20 RH differences were positive (HMI31 indicating the lower RH), with the differences near the level of the combined accuracies being approximately 50% positive and 50% negative. It does not appear that the HMI31 has a RH bias. Plots of EBBR sensor/HMI31 differences as a function of time from the installation of the individual sensors (a range of 10 to 30 months in the field) shows no obvious pattern of sensor drift with time. This is encouraging. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/06/1995 | 0000 | 10/06/1995 | 2359 |
Subject: | SGP/EBBR/E13 - 3-month Check |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during this period, at least, do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. Time of the check is not known, so the entire day has been flagged as suspect. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/10/1995 | 1800 | 10/10/1995 | 2100 |
Subject: | SGP/EBBR/E15 - 3-month Check |
DataStreams: | sgp5ebbrE15.b1, sgp30ebbrE15.b1, sgp15ebbrE15.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during this period, at least, do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/13/1995 | 1300 | 10/13/1995 | 1600 |
Subject: | SGP/EBBR/E20 - 3-month Check |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during this period, at least, do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/11/1995 | 1630 | 10/11/1995 | 1930 |
Subject: | SGP/EBBR/E22 - 3-month Check |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during this period, at least, do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/13/1995 | 1400 | 10/13/1995 | 1700 |
Subject: | SGP/EBBR/E26 - 3-month Check |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during this period, at least, do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/04/1995 | 1500 | 10/04/1995 | 1800 |
Subject: | SGP/EBBR/E7 - 3-month Check |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during this period, at least, do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/02/1995 | 1930 | 10/02/1995 | 2230 |
Subject: | SGP/EBBR/E8 - 3-month Check |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during this period, at least, do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/15/1995 | 2100 | 11/25/1995 | 0000 |
Subject: | SGP/EBBR/E9 - Annual calibration changeout |
DataStreams: | sgp30ebbrE9.b1, sgp15ebbrE9.b1 |
Description: | On November 15, 1995 the changeout for annual calibration occurred at E9, Ashton, KS. The soil moisture sensors were shipped wet from the vendor and did not stabilize until 25 November 1995. The following data should be considered incorrect until 25 November 1995: 15 minute: rr_sm1, rr_sm2, rr_sm3, rr_sm4, rr_sm5 30 minute: h, e, ave_shf, shf1, shf2, shf3, shf4, shf5, sm1, sm2, sm3, sm4, sm5, cs1, cs2, cs3, cs4, cs5, ces1, ces2, ces3, ces4, g1, g2, g3, g4, g5. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/09/2002 | 1600 | 05/09/2002 | 1631 |
Subject: | SGP/EBBR/E9 - 6 Month Checks |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Description: | The indicated measurements were incorrect during the 6 month checks. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/30/2002 | 1540 | 04/30/2002 | 1611 |
Subject: | SGP/EBBR/E12 - 6 Month Checks |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | The indicated measurements were incorrect during the 6 month checks. |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/16/2002 | 1855 | 05/16/2002 | 1926 |
Subject: | SGP/EBBR/E13 - 6 Month Checks |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | The indicated measurements were incorrect during the 6 month checks. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/14/2002 | 1505 | 05/14/2002 | 1546 |
Subject: | SGP/EBBR/E15 - 6 Month Checks |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | The indicated measurements were incorrect during the 6 month checks. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/30/2002 | 2000 | 04/30/2002 | 2031 |
Subject: | SGP/EBBR/E18 - 6 Month Checks |
DataStreams: | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
Description: | The indicated measurements were incorrect during the 6 month checks. |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/16/2002 | 1430 | 05/16/2002 | 1501 |
Subject: | SGP/EBBR/E19 - 6 Month Checks |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | The indicated measurements were incorrect during the 6 month checks. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/01/2002 | 1910 | 05/01/2002 | 1946 |
Subject: | SGP/EBBR/E20 - 6 Month Checks |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | The indicated measurements were incorrect during the 6 month checks. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/15/2002 | 1810 | 05/15/2002 | 1846 |
Subject: | SGP/EBBR/E22 - 6 Month Checks |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | The indicated measurements were incorrect during the 6 month checks. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/02/2002 | 1500 | 05/02/2002 | 1531 |
Subject: | SGP/EBBR/E26 - 6 Month Checks |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | The indicated measurements were incorrect during the 6 month checks. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/12/2002 | 0000 | 01/17/2002 | 1631 |
Subject: | SGP/EBBR/E2 - AEM Hung |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | The automatic exchange mechanism hung, thereby making a number of measurements incorrect. |
Measurements: | sgp30ebbrE2.b1:
sgp15ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/15/1995 | 2100 | 11/16/1995 | 1700 |
Subject: | SGP/EBBR/E9 - Annual calibration changeout |
DataStreams: | sgp30ebbrE9.b1, sgp5ebbrE9.b1, sgp15ebbrE9.b1 |
Description: | On November 15, 1995 the changeout for annual calibration occurred at E9, Ashton, KS. After installation of the new SEBS unit, RH output was zero. The next day site operations personnel found that the RH probe cable connections were loose. RH returned to normal after the connections were tightened at 1700 GMT on 16 November 1995. During the period of time that the connections were not made, the following data values are incorrect: 5 minute: hum_top, hum_bot, vp_top, vp_bot 15 minute: mv_hum_r, mv_hum_l 30 minute: hum_top, hum_bot, vp_top, vp_bot. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/01/2002 | 0330 | 03/01/2002 | 1701 |
Subject: | SGP/EBBR/E2 - Wind Direction Incorrect |
DataStreams: | sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | The wind direction was always 65 to 90 degrees and sigma direction was near zero; the data is incorrect. |
Measurements: | sgp30ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/15/2002 | 0645 | 01/15/2002 | 1701 |
02/14/2002 | 1515 | 02/14/2002 | 1531 |
Subject: | SGP/EBBR/E2 - Reprocess: Soil Temperature #2 Incorrect |
DataStreams: | sgp30ebbrE2.b1 |
Description: | The indicated measurements were incorrect because soil temperature #2 was incorrect. Average soil heat flow, sensible heat and latent heat can be recalculated using the remaining good sets of soil sensors as follows: ave_shf = (g1 + g3 + g4 + g5)/4 e = -(q + ave_shf)/(1 + bowen) h = -(q + e + ave_shf) |
Measurements: | sgp30ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/15/2002 | 1900 | 01/15/2002 | 1901 |
Subject: | SGP/EBBR/E2 - Data Incorrect |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.00, sgp5ebbrE2.b1 |
Description: | The indicated measurements were zeros (30 minute) or incorrect (5 and 15 minute) the half hour after the recalibration unit was installed. |
Measurements: | sgp5ebbrE2.00:
sgp15ebbrE2.b1:
sgp30ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/12/2002 | 2100 | 03/12/2002 | 2101 |
Subject: | SGP/EBBR/E7 - Measurements Zero |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.00, sgp5ebbrE7.b1 |
Description: | The indicated measurements were zero after interruption of the CR10 program. |
Measurements: | sgp5ebbrE7.00:
sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/14/2002 | 2200 | 04/23/2002 | 1531 |
05/20/2002 | 1200 | 06/06/2002 | 1501 |
Subject: | SGP/EBBR/E7 - AEM Hung |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | The indicated measurements were incorrect because the automatic exchange mechanism hung in one position. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/20/2002 | 1200 | 05/21/2002 | 1601 |
Subject: | SGP/EBBR/E7 - Wind Speed Incorrect |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | The wind speed and resultant wind speed are incorrect for the indicated times because a cup had broken off of the cup anemometer. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/15/2001 | 0000 | 01/15/2002 | 2001 |
Subject: | SGP/EBBR/E8 - Low Battery Voltage Causes Incorrect Measurements |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.00, sgp5ebbrE8.b1 |
Description: | All measurements should be considered to be suspect when the battery voltage (see the 15 minute data) is less than 10.5 volts. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
sgp5ebbrE8.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/07/1996 | 1901 | 11/07/1996 | 1930 |
11/06/1996 | 1300 | 11/06/1996 | 2300 |
Subject: | SGP/EBBR/E22 - Annual Calibration Changeout |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | On 6 November 1996 the EBBR changeout for annual calibration occurred at E22, Cordell, OK. SEBS unit 7 was removed and unit 3 was installed. Between 1300 and 2300 GMT no data or incorrect data was collected; most of the data is missing. Data for half hour 1930 GMT on 7 November is incorrect; 3 month calibration checks were being performed. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/15/2002 | 2015 | 01/28/2002 | 1801 |
02/26/2002 | 0230 | 02/26/2002 | 1831 |
02/27/2002 | 0315 | 03/26/2002 | 1731 |
04/06/2002 | 0000 | 04/09/2002 | 1731 |
01/30/2002 | 0445 | 02/12/2002 | 1801 |
Subject: | SGP/EBBR/E8 - AEM Hung |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The indicated data is incorrect while the automatic exchange mechanism was hung in one position. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/28/2002 | 1615 | 02/28/2002 | 1646 |
Subject: | SGP/EBBR/E2 - 6 Month Checks |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | The indicated measurements were incorrect during the 6 month checks. |
Measurements: | sgp30ebbrE2.b1:
sgp15ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/23/2002 | 1500 | 05/23/2002 | 1531 |
Subject: | SGP/EBBR/E2 - 6 Month Checks |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | The indicated measurements were incorrect during the 6 month checks. |
Measurements: | sgp30ebbrE2.b1:
sgp15ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/07/2002 | 1645 | 05/07/2002 | 1716 |
Subject: | SGP/EBBR/E7 - 6 Month Checks |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | The indicated measurements were incorrect during the 6 month checks. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/21/2002 | 1730 | 05/21/2002 | 1801 |
Subject: | SGP/EBBR/E8 - 6 Month Checks |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The indicated measurements were incorrect during the 6 month checks. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/30/2002 | 0430 | 02/06/2002 | 1646 |
Subject: | SGP/EBBR/E12 - AEM Hung |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | The automatic exchange mechanism did not exchange properly, causing other measurements to be incorrect. |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/12/2002 | 1430 | 06/12/2002 | 1501 |
Subject: | SGP/EBBR/E12 - Measurements Incorrect |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.00, sgp5ebbrE12.b1 |
Description: | All measurements were incorrect or offscale for an unknown reason. |
Measurements: | sgp30ebbrE12.b1:
sgp15ebbrE12.b1:
sgp5ebbrE12.00:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/28/2002 | 1800 | 07/07/2002 | 2356 |
Subject: | SGP/EBBR/E12 - Resistance Measurements Incorrect |
DataStreams: | sgp30ebbrE12.b1, sgp5ebbrE12.b1, sgp15ebbrE12.b1 |
Description: | All sensors that depend on resistance measurements are incorrect. Average soil heat flow, sensible heat flux, and latent heat flux are incorrect as a result. This condition normally occurs when the reference temperature sensor is disconnected or bad. |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/25/2002 | 2202 | 03/03/2002 | 1129 |
Subject: | SGP/EBBR/E13 - Reprocess: Soil Heat Flow #1 Incorrect |
DataStreams: | sgp30ebbrE13.b1 |
Description: | Soil Heat Flow #1 was incorrect during the indicated period. Average soil heat flow, sensible heat flux, and latent heat flux were affected; these can be recalculated using the remaining good sets of soil sensors as follows: ave_shf = (g2 + g3 + g4 + g5)/4 e = -(q + ave_shf)/(1 + bowen) h = -(q + e + ave_shf) |
Measurements: | sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/07/2002 | 2032 | 05/20/2002 | 2101 |
Subject: | SGP/EBBR/E13 - Reprocess: Soil Heat Flow #2 Incorrect |
DataStreams: | sgp30ebbrE13.b1 |
Description: | Soil Heat Flow #2 was incorrect during the indicated period. Average soil heat flow, sensible heat flux, and latent heat flux were affected; these can be recalculated using the remaining good sets of soil sensors as follows: ave_shf = (g1 + g3 + g4 + g5)/4 e = -(q + ave_shf)/(1 + bowen) h = -(q + e + ave_shf) |
Measurements: | sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/05/2001 | 1715 | 02/22/2002 | 2131 |
Subject: | SGP/EBBR/E13 - Wrong CR10 Program Installed |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.00, sgp5ebbrE13.b1 |
Description: | An old version 4 program was installed on 5 Oct 01 instead of the correct version 8 program. All of the indicated measurements are suspect. |
Measurements: | sgp5ebbrE13.00:
sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/24/2002 | 0600 | 06/14/2002 | 1431 |
Subject: | SGP/EBBR/E13 - Wind Direction Incorrect |
DataStreams: | sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | The wind direction and sigma_wd was near zero for the indicated times. |
Measurements: | sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/08/2002 | 2300 | 05/09/2002 | 0101 |
05/11/2002 | 1530 | 05/11/2002 | 1601 |
01/18/2002 | 0015 | 03/03/2002 | 0801 |
03/03/2002 | 0815 | 03/08/2002 | 1401 |
03/08/2002 | 1415 | 03/09/2002 | 0931 |
03/09/2002 | 0945 | 03/09/2002 | 2001 |
03/12/2002 | 0115 | 04/12/2002 | 1901 |
04/16/2002 | 1430 | 04/16/2002 | 2001 |
04/16/2002 | 2230 | 04/17/2002 | 0031 |
04/17/2002 | 1530 | 04/17/2002 | 1701 |
04/18/2002 | 1230 | 04/18/2002 | 1231 |
05/06/2002 | 0600 | 05/06/2002 | 1931 |
05/07/2002 | 0000 | 05/07/2002 | 0001 |
05/07/2002 | 0300 | 05/07/2002 | 0301 |
05/07/2002 | 0400 | 05/07/2002 | 0401 |
05/11/2002 | 1900 | 05/11/2002 | 1901 |
03/12/2002 | 0000 | 03/12/2002 | 0101 |
04/17/2002 | 2230 | 04/17/2002 | 2231 |
05/07/2002 | 1700 | 05/07/2002 | 1731 |
Subject: | SGP/EBBR/E13 - AEM Hung up |
DataStreams: | sgp30ebbrE13.b1, sgp5ebbrE13.b1, sgp15ebbrE13.b1 |
Description: | The automatic exchange mechanism hung numerous times, but with various different measurements of home signals. In all cases, temperature and humidity gradients were incorrect; thus the Bowen ratio, sensible heat flux, and latent heat flux are incorrect for the indicated periods. Even when the home 15 and 30 signals were different (but too high) the measurements were incorrect because the program keys off of the home signals, which must be in the proper ranges. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/15/2001 | 0000 | 06/30/2002 | 2356 |
Subject: | SGP/EBBR/E9 - Reprocess: Soil Temperature #5 Incorrect |
DataStreams: | sgp30ebbrE9.b1 |
Description: | Soil temperature # 5 was incorrect most of the time, leading to incorrect surface heat flow, latent heat flux and sensible heat flux. The heat fluxes can be recalculated by using soil sets 1-4 as follows: ave_shf = (g1 + g2 + g3 + g4)/4 e = -(q + ave_shf)/(1 + bowen) h = -(q + e + ave_shf) |
Measurements: | sgp30ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/01/2001 | 0000 | 03/29/2002 | 1631 |
Subject: | SGP/EBBR/E18 - Version 6 Produced Noisy Fluxes |
DataStreams: | sgp30ebbrE18.b1 |
Description: | Version 6 of the CR10 program produced noisy latent and sensible heat fluxes. That ended on March 29 after a recalibration unit was installed with version 8. |
Measurements: | sgp30ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/06/2002 | 0030 | 04/17/2002 | 1401 |
05/01/2002 | 0300 | 05/28/2002 | 2031 |
11/14/2001 | 1945 | 11/14/2001 | 2001 |
11/25/2001 | 0215 | 11/25/2001 | 1901 |
11/28/2001 | 1145 | 11/28/2001 | 1501 |
11/29/2001 | 0745 | 11/29/2001 | 0931 |
11/29/2001 | 1415 | 11/30/2001 | 0431 |
12/02/2001 | 0145 | 12/02/2001 | 0201 |
12/03/2001 | 0045 | 12/03/2001 | 0101 |
12/24/2001 | 2015 | 12/24/2001 | 2031 |
12/25/2001 | 1915 | 12/25/2001 | 2031 |
01/18/2002 | 0115 | 01/18/2002 | 0201 |
01/24/2002 | 2115 | 01/24/2002 | 2131 |
01/28/2002 | 1345 | 01/28/2002 | 1401 |
02/13/2002 | 2215 | 02/14/2002 | 0801 |
02/25/2002 | 2345 | 02/26/2002 | 0001 |
03/14/2002 | 1515 | 03/29/2002 | 1831 |
04/18/2002 | 2230 | 04/30/2002 | 2031 |
01/22/2002 | 1615 | 01/22/2002 | 1631 |
Subject: | SGP/EBBR/E18 - AEM Hung |
DataStreams: | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
Description: | The automatic exchange mechanism did not work properly for a number of time periods. The Bowen ratio, latent heat flux, and sensible heat flux were incorrect also during this condition. |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/29/2002 | 1815 | 03/29/2002 | 2131 |
Subject: | SGP/EBBR/E18 - Soil Heat Flows Incorrect After Calibration Changeout |
DataStreams: | sgp30ebbrE18.b1 |
Description: | The soil heat flows were incorrect while the EBBR changeout was being done. |
Measurements: | sgp30ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/20/2002 | 0600 | 05/20/2002 | 2001 |
05/21/2002 | 0430 | 05/21/2002 | 1731 |
05/22/2002 | 1130 | 05/22/2002 | 1631 |
Subject: | SGP/EBBR/E18 - Reprocess: Soil Heat Flow #1 Incorrect |
DataStreams: | sgp30ebbrE18.b1 |
Description: | Soil Heat Flow 1 was incorrect during the indicated periods. Average soil heat flow, sensible heat flux, and latent heat flux were affected; these can be recalculated using the remaining good sets of soil sensors as follows: ave_shf = (g2 + g3 + g4 + g5)/4 e = -(q + ave_shf)/(1 + bowen) h = -(q + e + ave_shf) |
Measurements: | sgp30ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/09/2002 | 1900 | 05/14/2002 | 1846 |
Subject: | SGP/EBBR/E18 - Left Air Temperature Incorrect |
DataStreams: | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
Description: | The left air temperature was incorrect much of the time, causing the Bowen ratio, sensible heat flux, and latent heat flux to be incorrect also. |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/17/2002 | 1430 | 04/17/2002 | 1431 |
Subject: | SGP/EBBR/E18 - Reference Temperature Incorrect |
DataStreams: | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
Description: | The reference temperature was incorrect during the checks of the left air temperature sensor, leading to all gradient measurements, Bowen ratio, sensible heat flux, latent heat flux, soil temperature, and soil moisture measurements being incorrect. |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/08/2002 | 1630 | 02/21/2002 | 1601 |
Subject: | SGP/EBBR/E19 - Reprocess: Soil Heat Flow #3 Incorrect |
DataStreams: | sgp30ebbrE19.b1 |
Description: | Soil heat flow 3 was incorrect, with values too large positive and too small negative. Since the calibration in the program is the one supplied by the vendor, it appears that the calibration may be wrong. Average soil heat flow, sensible heat flux and latent heat flux can be recalculated using the remaining good sets of soil sensors as follows: ave_shf = (g1 + g2 + g4 + g5)/4 e = -(q + ave_shf)/(1 + bowen) h = -(q + e + ave_shf) |
Measurements: | sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/01/2001 | 0000 | 06/30/2002 | 2356 |
Subject: | SGP/EBBR/E19 - Reprocess: Soil Temperature #3 Incorrect |
DataStreams: | sgp30ebbrE19.b1 |
Description: | Soil Temperature 3 was incorrect for the period indicated, causing average soil flow, latent heat flux, and sensible heat flux to be incorrect also. Average soil heat flow, sensible heat flux, and latent heat flux can be recalculated using the remaining good sets of soil sensors as follows: ave_shf = (g1 + g2 + g4 + g5)/4 e = -(q + ave_shf)/(1 + bowen) h = -(q + e + ave_shf) |
Measurements: | sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/08/2002 | 0430 | 04/08/2002 | 0431 |
05/19/2002 | 0400 | 05/19/2002 | 0401 |
Subject: | SGP/EBBR/E19 - Most Measurements Incorrect |
DataStreams: | sgp15ebbrE19.b1, sgp5ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.00 |
Description: | Most measurements were incorrect for the indicated periods. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp5ebbrE19.00:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/28/2002 | 1600 | 03/28/2002 | 1601 |
06/04/2002 | 1500 | 06/04/2002 | 1501 |
Subject: | SGP/EBBR/E9 - Measurements All Zeros |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.00, sgp5ebbrE9.b1 |
Description: | The measurements were all zeros at the indicated times. This occurs after a CR10 program restart, which sometimes happens after some servicing of the EBBR is performed. |
Measurements: | sgp30ebbrE9.b1:
sgp5ebbrE9.00:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/27/2002 | 1500 | 06/27/2002 | 1501 |
04/18/2002 | 1730 | 04/18/2002 | 1731 |
05/05/2002 | 1400 | 05/05/2002 | 1401 |
Subject: | SGP/EBBR/E19 - All Measurements Zero |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.00, sgp5ebbrE19.b1 |
Description: | Measurements are often zero the first period after a CR10 restart. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp5ebbrE19.00:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/25/2002 | 2100 | 05/25/2002 | 2101 |
05/25/2002 | 2200 | 05/25/2002 | 2231 |
05/29/2002 | 1730 | 05/29/2002 | 1731 |
05/29/2002 | 2000 | 05/29/2002 | 2001 |
05/29/2002 | 2130 | 05/29/2002 | 2131 |
05/30/2002 | 1500 | 05/30/2002 | 1501 |
06/15/2002 | 1900 | 06/15/2002 | 1901 |
05/29/2002 | 1830 | 05/29/2002 | 1831 |
04/24/2002 | 1600 | 04/24/2002 | 2201 |
Subject: | SGP/EBBR/E19 - Pressure Incorrect |
DataStreams: | sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | The pressure measurement was incorrect. This is used to determine the Bowen ratio, thereby affecting the sensible and latent heat fluxes also. |
Measurements: | sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/16/2002 | 2000 | 05/17/2002 | 0531 |
Subject: | SGP/EBBR/E19 - Measurements Incorrect After CR10 Replacement |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.00, sgp5ebbrE19.b1 |
Description: | Replacement of the CR10 datalogger caused most measurements to be incorrect, reason unknown. A different logger was installed at the end of the specified period and this resolved the problem. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp5ebbrE19.00:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/27/2002 | 0000 | 06/02/2002 | 1031 |
Subject: | SGP/EBBR/E19 - Low Battery Voltage |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.00, sgp5ebbrE19.b1 |
Description: | A battery voltage of less than 10.5 causes many of the sensors to produce incorrect measurements. The 15 minute data contains the battery voltage and should be used to determine whether the measurements are usable, based on battery voltage. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp5ebbrE19.00:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/07/2002 | 1845 | 02/07/2002 | 1901 |
Subject: | SGP/EBBR/E22 - All Measurements Zero |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.00, sgp5ebbrE22.b1 |
Description: | The measurements are often all zero the first period after a CR10 restart, as occurred here. |
Measurements: | sgp5ebbrE22.00:
sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/12/2002 | 0515 | 01/15/2002 | 0031 |
02/08/2002 | 0845 | 02/08/2002 | 2031 |
01/09/2002 | 1545 | 01/12/2002 | 0001 |
02/08/2002 | 2145 | 02/14/2002 | 2001 |
02/14/2002 | 2015 | 02/16/2002 | 1531 |
02/16/2002 | 1945 | 02/17/2002 | 0001 |
02/17/2002 | 0415 | 02/20/2002 | 0531 |
02/20/2002 | 0815 | 02/20/2002 | 1831 |
02/28/2002 | 1030 | 03/06/2002 | 1901 |
04/21/2002 | 2000 | 05/29/2002 | 1831 |
12/17/2001 | 1230 | 01/09/2002 | 1531 |
Subject: | SGP/EBBR/E22 - AEM Hung or High Home Signals |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | The automatic exchange mechanism hung some times and continued to switch at other times, although with high home signal values. If home signal values are not in the proper ranges, the gradient calculation is not performed properly by the CR10 program. Affected flux measurements cannot be recalculated. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/15/2002 | 2215 | 04/08/2002 | 1431 |
Subject: | SGP/EBBR/E25 - Reprocess: Soil Heat Flow #5 Negative Offset |
DataStreams: | sgp30ebbrE25.b1 |
Description: | Soil heat flow 5 was either too large negative or negative when it should have been positive; the cause of the negative offset is unknown, but is often the result of a failed J-panel over-voltage protection device. Average soil heat flow, sensible heat flux, and latent heat flux were affected; these can be recalculated using the remaining good sets of soil sensors as follows: ave_shf = (g1 + g2 + g3 + g4)/4 e = -(q + ave_shf)/(1 + bowen) h = -(q + e + ave_shf) |
Measurements: | sgp30ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/30/2002 | 1500 | 05/30/2002 | 1501 |
Subject: | SGP/EBBR/E26 - Measurements Zero |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.00, sgp5ebbrE26.b1 |
Description: | The measurements are often zero after a CR10 datalogger restart, as occurred here. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
sgp5ebbrE26.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/30/2002 | 1530 | 05/30/2002 | 1531 |
Subject: | SGP/EBBR/E26 - Measurements Incorrect |
DataStreams: | sgp30ebbrE26.b1 |
Description: | Change of energy storage (CES) measurements (calculated from the change of soil temperature with time) are sometimes incorrect after a CR10 program restart because the beginning measurement of the averaging period is zero. Average soil heat flow, sensible heat flux, and latent heat flux were affected and cannot be recalculated. |
Measurements: | sgp30ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/01/2002 | 2115 | 06/30/2002 | 2356 |
Subject: | SGP/EBBR/E26 - Multiplexer 2 Failure From High Temperature |
DataStreams: | sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp15ebbrE26.b1 |
Description: | Multiplexer 2 is failing during times of control box temperatures above 29 degrees C. The failure begins in the last channel and progresses to lower channels as the temperature increases. It has not yet been determined whether the failure is in the relays or a solid state component of the multiplexer. This condition is probably due to the aging of the device and the millions of cycles that this 10 year old device has made. More failures of this sort may occur in the future. A cure is being sought aside from replacement of the multiplexers to prevent the problem. The indicated measurements need to be inspected to determine which ones are usable. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/28/2002 | 1615 | 03/28/2002 | 1801 |
Subject: | SGP/EBBR/E9 - Humidity Probe Measurements Incorrect |
DataStreams: | sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Description: | The temperature and humidity measurements from both T/RH probes were incorrect during changeout of the right probe. |
Measurements: | sgp30ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/28/2001 | 2045 | 01/23/2002 | 1801 |
Subject: | SGP/EBBR/E20 - AEM Hung |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | The automatic exchange mechanism frequently hung in one position. The home signals and automated qc in the netcdf data files indicate when this occurred. Bowen ratio, sensible heat flux, and latent heat flux are incorrect for this condition. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/28/2001 | 0045 | 11/28/2001 | 2031 |
02/06/2002 | 2130 | 02/07/2002 | 0331 |
Subject: | SGP/EBBR/E20 - Battery Voltage Low |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.00, sgp5ebbrE20.b1 |
Description: | When the battery voltage is below 10.5 volts, most of the measurements are suspect. |
Measurements: | sgp5ebbrE20.b1:
sgp5ebbrE20.00:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/14/2001 | 1605 | 11/28/2001 | 2031 |
Subject: | SGP/EBBR/E20 - Pressure Almost Constant |
DataStreams: | sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | During the indicated period the pressure went to an almost constant 102 to 105 kPa. This occurred after the replacement of another sensor. The incorrect pressure values affect the sensible and latent heat flux calculations. |
Measurements: | sgp5ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/12/2001 | 1915 | 12/12/2001 | 2031 |
Subject: | SGP/EBBR/E20 - Reprocess: Soil Heat Flow #3 Incorrect During Sensor Changeout |
DataStreams: | sgp30ebbrE20.b1 |
Description: | Soil heat flow #3 measurements were incorrect during it's replacement. Average soil heat flow, sensible heat flux, and latent heat flux were affected; these can be recalculated using the remaining good sets of soil sensors as follows: ave_shf = (g1 + g5)/2 e = -(q + ave_shf)/(1 + bowen) h = -(q + e + ave_shf) Note: during the time period referenced here, soil heat flow #4 and #2 were also incorrect. This quality information is documented in D020717.46, D020717.47 and D020726.1. |
Measurements: | sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/12/2002 | 0315 | 01/13/2002 | 2131 |
01/17/2002 | 0915 | 01/21/2002 | 0901 |
02/20/2002 | 1645 | 02/20/2002 | 1831 |
12/04/2001 | 0000 | 12/12/2001 | 1914 |
02/09/2002 | 0115 | 02/12/2002 | 0731 |
Subject: | SGP/EBBR/E20 - Reprocess: Soil Heat Flow #2 Has a Negative Offset |
DataStreams: | sgp30ebbrE20.b1 |
Description: | Soil heat flow #2 measurements had a negative offset. Negative values were too small and negative values occurred when positive values should have. This can occur when there is a poor wire connection or a J-panel over-voltage protection device has failed. Average soil heat flow, sensible heat flux, and latent heat flux were affected; these can be recalculated using the remaining good sets of soil sensors as follows: ave_shf = (g1 + g3 + g4 + g5)/4 e = -(q + ave_shf)/(1 + bowen) h = -(q + e + ave_shf) |
Measurements: | sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/24/2001 | 0532 | 12/27/2001 | 1901 |
Subject: | SGP/EBBR/E20 - Reprocess: Soil Heat Flow #4 Incorrect |
DataStreams: | sgp30ebbrE20.b1 |
Description: | Soil heat flow #4 was incorrect because it's wires were loose from the connection terminals. Average soil heat flow, sensible heat flux, and latent heat flux were affected; these can be recalculated using the remaining good sets of soil sensors as follows: ave_shf = (g1 + g2 + g3 + g5)/4 e = -(q + ave_shf)/(1 + bowen) h = -(q + e + ave_shf) |
Measurements: | sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/15/2002 | 1700 | 05/15/2002 | 1701 |
11/14/2001 | 1600 | 11/14/2001 | 1601 |
05/29/2002 | 1830 | 05/29/2002 | 1901 |
11/27/2001 | 1800 | 11/27/2001 | 1801 |
Subject: | SGP/EBBR/E20 - Measurements Zero |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.00, sgp5ebbrE20.b1 |
Description: | Most measurements were zero during the indicated periods. This often happens for the first reported data time after a CR10 datalogger restart. |
Measurements: | sgp5ebbrE20.b1:
sgp5ebbrE20.00:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/22/2002 | 2015 | 02/22/2002 | 2101 |
02/23/2002 | 1745 | 02/23/2002 | 1801 |
02/24/2002 | 0045 | 02/25/2002 | 0201 |
03/05/2002 | 2015 | 03/06/2002 | 0201 |
03/06/2002 | 1915 | 03/07/2002 | 1901 |
Subject: | SGP/EBBR/E20 - Reprocess: Soil Temperature #3 Incorrect |
DataStreams: | sgp30ebbrE20.b1 |
Description: | Soil temperature 3 was incorrect during the indicated periods. Average soil heat flow, sensible heat flux, and latent heat flux were affected; these can be recalculated using the remaining good sets of soil sensors as follows: ave_shf = (g1 + g2 + g4 + g5)/4 e = -(q + ave_shf)/(1 + bowen) h = -(q + e + ave_shf) |
Measurements: | sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/30/2002 | 1015 | 07/16/2002 | 1630 |
Subject: | SGP/EBBR/E9 - Right Humidity Suspect |
DataStreams: | sgp5ebbrE9.b1, sgp30ebbrE9.b1 |
Description: | The Right Humidity measurement was normally incorrect, although it is often difficult to determine it by looking at the measurement. The measurements float around, sometimes being too large and sometimes being too small, through most of the day, particularly for the hour before sunset through the hour after sunrise. Great care and comparison with surrounding EBBR sites would be needed to be able to use any of the latent and sensible heat flux measurements from E9 for this period. Another DQPR has been issued to fix this problem, which has eluded repair efforts so far. |
Measurements: | sgp30ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/20/2002 | 1745 | 02/20/2002 | 1901 |
Subject: | SGP/EBBR/E20 - Measurements Incorrect During Sensor Changeout |
DataStreams: | sgp30ebbrE20.b1 |
Description: | The replacement of soil heat flow 2 caused the loss of initial soil temperature values used in the calculation of CES, G, C, ave_shf, bowen, e, and h. |
Measurements: | sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/27/2001 | 1845 | 12/27/2001 | 1901 |
05/15/2002 | 1715 | 05/15/2002 | 1731 |
Subject: | SGP/EBBR/E20 - Measurements Incorrect |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.00, sgp5ebbrE20.b1 |
Description: | Many measurements were incorrect because of the lack of an initial 15 minute value to use for gradient or soil change of energy storage calculations. |
Measurements: | sgp5ebbrE20.b1:
sgp5ebbrE20.00:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/01/2001 | 0000 | 06/20/2002 | 1231 |
Subject: | SGP/EBBR/E12 - Reprocess: Soil Temperature #3 Incorrect |
DataStreams: | sgp30ebbrE12.b1 |
Description: | Soil Temperature #3 was flaky primarily during nighttime hours and sometimes during early daylight hours. Average soil heat flow, sensible heat flux, and latent heat flux are also affected; these can be recalculated using the remaining good sets of soil sensors as follows: The heat fluxes can be recalculated by using soil sets 1, 2, 4, and 5 as follows: ave_shf = (g1 + g2 + g4 + g5)/4 e = -(q + ave_shf)/(1 + bowen) h = -(q + e + ave_shf) |
Measurements: | sgp30ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/25/2002 | 2230 | 06/30/2002 | 2356 |
Subject: | SGP/EBBR/E12 - Wind Direction Incorrect |
DataStreams: | sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | The wind direction measurement was zero, in the 50s, or always less than 180 degrees. A DQPR has been issued to resolve the problem. |
Measurements: | sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/30/2002 | 1200 | 01/30/2002 | 1201 |
Subject: | SGP/EBBR/E12 - Reprocess: Soil Moisture 4 and 5 Incorrect |
DataStreams: | sgp30ebbrE12.b1 |
Description: | Soil moisture measurements from sets 4 and 5 were incorrect, leading to many other incorrect measurements calculated using them. Average soil heat flow, sensible heat flux, and latent heat flux are affected; these can be recalculated using the remaining good sets of soil sensors as follows: ave_shf = (g1 + g2 + g3)/3 e = -(q + ave_shf)/(1 + bowen) h = -(q + e + ave_shf) Note: D020717.6 reports Soil temperature sensor #3 was also bad during this time, but for the very specific records affected by the soil moisture #4 and #5 reporting incorrectly, soil temp #3 was okay. Therefore the equations given are correct. |
Measurements: | sgp30ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/30/2002 | 1430 | 01/30/2002 | 1431 |
Subject: | SGP/EBBR/E12 - Reprocess: Soil Moisture 2 and 5 Incorrect |
DataStreams: | sgp30ebbrE12.b1 |
Description: | Soil moisture measurements from sets 2 and 5 were incorrect, leading to many other incorrect measurements calculated using them. Average soil heat flow, sensible heat flux, and latent heat flux are affected; these can be recalculated using the remaining good sets of soil sensors as follows: ave_shf = (g1 + g3 + g4)/3 e = -(q + ave_shf)/(1 + bowen) h = -(q + e + ave_shf) Note: D020717.6 reports Soil temperature sensor #3 was also bad during this time, but for the very specific records affected by the soil moisture #2 and #5 reporting incorrectly, soil temp #3 was okay. Therefore the equations given are correct. |
Measurements: | sgp30ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/22/1995 | 2030 | 12/22/1995 | 2130 |
Subject: | SGP/EBBR/E13 - Thermocouple out of range, data incorrect |
DataStreams: | sgp30ebbrE13.b1, sgp5ebbrE13.b1, sgp15ebbrE13.b1 |
Description: | The left thermocouple temperature for 22 December was out of range between 2030 and 2130 GMT, yielding incorrect values as listed below: 5 minute: tair_top, tair_bot 15 minute: tair_l 30 minute: tair_top, tair_bot, bowen, h, e |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/17/1996 | 1400 | 07/17/1996 | 1430 |
Subject: | SGP/EBBR/E26 - Annual Calibration Changeout |
DataStreams: | sgp5ebbrE26.b1, sgp15ebbrE26.b1, sgp30ebbrE26.b1 |
Description: | On 16 November 1996 the EBBR changeout for annual calibration occurred at E26, Cement, OK. SEBS unit 2 was removed and unit 8 was installed. Data for half hours 1400 and 1430 GMT on 17 November is incorrect; 3 month calibration checks were being performed. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/12/2001 | 2032 | 12/24/2001 | 0531 |
Subject: | SGP/EBBR/E20 - Reprocess: Soil Heat Flow #2 and #4 incorrect |
DataStreams: | sgp30ebbrE20.b1 |
Description: | Soil heat flow #2 measurements had a negative offset. Negative values were too small and negative values occurred when positive values should have. This can occur when there is a poor wire connection or a J-panel over-voltage protection device has failed. Soil heat flow #4 was incorrect because it's wires were loose from the connection terminals. Average soil heat flow, sensible heat flux, and latent heat flux were affected; these can be recalculated using the remaining good sets of soil sensors as follows: ave_shf = (g1 + g3 + g5)/3 e = -(q + ave_shf)/(1 + bowen) h = -(q + e + ave_shf) |
Measurements: | sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/03/2002 | 1130 | 03/07/2002 | 2031 |
01/18/2002 | 0015 | 01/25/2002 | 2201 |
Subject: | SGP/EBBR/E13 - Reprocess: Soil Heat Flow #1 and #2 Incorrect |
DataStreams: | sgp30ebbrE13.b1 |
Description: | Soil Heat Flow #1 and #2 were incorrect during the indicated period. Average soil heat flow, sensible heat flux, and latent heat flux were affected; these can be recalculated using the remaining good sets of soil sensors as follows: ave_shf = (g3 + g4 + g5)/3 e = -(q + ave_shf)/(1 + bowen) h = -(q + e + ave_shf) |
Measurements: | sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/01/2001 | 0000 | 06/18/2003 | 2359 |
Subject: | SGP/EBBR - Incorrect Units for Soil Heat Capacity (cs1, cs2, cs3, cs4, cs5) in 30EBBR |
DataStreams: | sgp30ebbrE12.b1, sgp30ebbrE13.b1, sgp30ebbrE15.b1, sgp30ebbrE18.b1, sgp30ebbrE19.b1, sgp30ebbrE2.b1, sgp30ebbrE20.b1, sgp30ebbrE22.b1, sgp30ebbrE26.b1, sgp30ebbrE4.b1, sgp30ebbrE7.b1, sgp30ebbrE8.b1, sgp30ebbrE9.b1 |
Description: | The units for Soil Heat Capacity (cs1, cs2, cs3, cs4, cs5) in the EBBR 30 minute netcdf header have been incorrect from the beginning of processing the data into netcdf files. The correct units are MJ/m^3/C. This does not affect the data quality. |
Measurements: | sgp30ebbrE2.b1:
sgp30ebbrE26.b1:
sgp30ebbrE9.b1:
sgp30ebbrE22.b1:
sgp30ebbrE13.b1:
sgp30ebbrE7.b1:
sgp30ebbrE18.b1:
sgp30ebbrE12.b1:
sgp30ebbrE4.b1:
sgp30ebbrE15.b1:
sgp30ebbrE19.b1:
sgp30ebbrE20.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/01/1993 | 1200 | 06/18/2003 | 1159 |
Subject: | SGP/EBBR - Incorrect Units for Soil Heat Capacity (cs1, cs2, cs3, cs4, cs5) in 30EBBR |
DataStreams: | sgp30ebbrE12.b1, sgp30ebbrE13.b1, sgp30ebbrE15.b1, sgp30ebbrE20.b1, sgp30ebbrE22.b1, sgp30ebbrE26.b1, sgp30ebbrE4.b1, sgp30ebbrE7.b1, sgp30ebbrE8.b1, sgp30ebbrE9.b1 |
Description: | The units for Soil Heat Capacity (cs1, cs2, cs3, cs4, cs5) in the EBBR 30 minute netcdf header have been incorrect from the beginning of processing the data into netcdf files. The correct units are MJ/m^3/C. This does not affect the data quality. (NOTE: The actual begin date of some of these data streams is as early as 7/4/93. This DQR applies to all 30m EBBR data.) |
Measurements: | sgp30ebbrE26.b1:
sgp30ebbrE9.b1:
sgp30ebbrE7.b1:
sgp30ebbrE4.b1:
sgp30ebbrE12.b1:
sgp30ebbrE15.b1:
sgp30ebbrE22.b1:
sgp30ebbrE13.b1:
sgp30ebbrE8.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/25/2002 | 2215 | 11/25/2002 | 2245 |
Subject: | SGP/EBBR/E18 - Data Incorrect during 6 month checks |
DataStreams: | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
Description: | Some of the measurements are incorrect during the 6 month checks. |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/14/2002 | 1600 | 11/14/2002 | 1630 |
Subject: | SGP/EBBR/E19 - Data Incorrect during 6 month checks |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | Some of the measurements are incorrect during the 6 month checks. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/26/2002 | 1900 | 11/26/2002 | 1930 |
Subject: | SGP/EBBR/E20 - Data Incorrect during 6 month checks |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | Some of the measurements are incorrect during the 6 month checks. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/13/2002 | 1900 | 11/13/2002 | 1930 |
Subject: | SGP/EBBR/E22 - Data Incorrect during 6 month checks |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | Some of the measurements are incorrect during the 6 month checks. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/27/2002 | 1605 | 11/27/2002 | 1635 |
Subject: | SGP/EBBR/E26 - Data Incorrect during 6 month checks |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | Some of the measurements are incorrect during the 6 month checks. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/07/2002 | 1600 | 11/07/2002 | 1630 |
Subject: | SGP/EBBR/E2 - Data Incorrect during 6 month checks |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | Some of the measurements are incorrect during the 6 month checks. |
Measurements: | sgp30ebbrE2.b1:
sgp15ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/19/2002 | 2000 | 11/19/2002 | 2030 |
Subject: | SGP/EBBR/E7 - Data Incorrect during 6 month checks |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | Some of the measurements are incorrect during the 6 month checks. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/05/2002 | 1930 | 11/05/2002 | 2000 |
Subject: | SGP/EBBR/E8 - Data Incorrect during 6 month checks |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | Some of the measurements are incorrect during the 6 month checks. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/19/2002 | 1635 | 11/19/2002 | 1735 |
Subject: | SGP/EBBR/E9 - Data Incorrect during 6 month checks |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Description: | Some of the measurements are incorrect during the 6 month checks. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/20/2002 | 1600 | 08/20/2002 | 1630 |
11/25/2002 | 1815 | 11/25/2002 | 1845 |
Subject: | SGP/EBBR/E12 - Data Incorrect during 6 month checks |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | Some of the measurements are incorrect during the 6 month checks. |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/14/2002 | 2015 | 11/14/2002 | 2045 |
Subject: | SGP/EBBR/E13 - Data Incorrect during 6 month checks |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | Some of the measurements are incorrect during the 6 month checks. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/12/2002 | 1710 | 11/12/2002 | 1740 |
Subject: | SGP/EBBR/E15 - Data Incorrect during 6 month checks |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | Some of the measurements are incorrect during the 6 month checks. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/11/2002 | 2100 | 07/18/2002 | 1700 |
Subject: | SGP/EBBR/E7 - Wind speed incorrect |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | A wind speed sensor electronics problem limited the measurement to the sensor offset. This data was incorrect. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/24/2002 | 1230 | 11/05/2002 | 2000 |
Subject: | SGP/EBBR/E8 - AEM Frozen |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | When the AEM (Automatic Exchange Mechanism) is hung inbetween a home position, the gradients of temperature and relative humidity, and thus sensible and latent heat fluxes, are incorrect. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/24/2002 | 0230 | 10/25/2002 | 1300 |
Subject: | SGP/EBBR/E8 - Wind Speed Frozen |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The wind speed sensor was frozen stopped by ice. The wind speed and residual wind speed measurements are incorrect. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/12/2002 | 0000 | 08/12/2002 | 1730 |
Subject: | SGP/EBBR/E8 - Water in broken Net Rad Dome |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The net radiometer had water in the bottom, resulting in incorrect measurements of net radation. Calculated Sensible (h) and Latent Heat (e) Flux are therefore incorrect also. |
Measurements: | sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/01/2002 | 1800 | 08/01/2002 | 1930 |
08/13/2002 | 0430 | 09/04/2002 | 1730 |
09/12/2002 | 1900 | 10/01/2002 | 1830 |
10/03/2002 | 1930 | 10/23/2002 | 0630 |
10/26/2002 | 1730 | 10/30/2002 | 2330 |
11/03/2002 | 1800 | 01/01/2003 | 0000 |
09/09/2002 | 1430 | 09/09/2002 | 1459 |
Subject: | SGP/EBBR/E2 - REPROCESS: Soil Heat Flow #2 Incorrect |
DataStreams: | sgp30ebbrE2.b1 |
Description: | Soil Heat Flow #2 was incorrect during these periods. See below for a work-around. |
Measurements: | sgp30ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/30/2002 | 1900 | 07/30/2002 | 1901 |
Subject: | SGP/EBBR/E7 - CES data incorrect |
DataStreams: | sgp30ebbrE7.b1 |
Description: | After a missing half hour, the soil temperature at the beginning of the half hour cannot be used to determine the soil energy storage term. Therefore the change of energy calculation blows up. The indicated measurements are therefore incorrect. |
Measurements: | sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/05/2002 | 2100 | 11/05/2002 | 2129 |
Subject: | SGP/EBBR/E7 - REPROCESS: Soil Heat Flow #2 Incorrect |
DataStreams: | sgp30ebbrE7.b1 |
Description: | Soil Heat Flow #2 measurements were incorrect while it was being replaced. |
Measurements: | sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/05/2002 | 2130 | 11/05/2002 | 2159 |
Subject: | SGP/EBBR/E7 - Change In Energy Storage Calculations Incorrect |
DataStreams: | sgp30ebbrE7.b1 |
Description: | Because soil temperature measurements were not available from the beginning of the half hour, the change in soil energy storage could not be calculated and thus is incorrect. |
Measurements: | sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/22/2002 | 1630 | 12/03/2002 | 1830 |
Subject: | SGP/EBBR/E7 - REPROCESS: Soil Heat Flow #5 too low |
DataStreams: | sgp30ebbrE7.b1 |
Description: | Soil Heat Flow #5 was sometimes 2X the other soil heat flow measurements when heat flows were negative. This does not usually greatly affect h and e, therefore the data color has been set to yellow. |
Measurements: | sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/10/2002 | 0800 | 07/16/2002 | 1900 |
07/22/2002 | 1630 | 07/30/2002 | 1630 |
08/24/2002 | 0230 | 08/27/2002 | 1800 |
Subject: | SGP/EBBR/E7 - AEM Problems |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | The AEM (Automatic Exchange Mechanism) hung in a non-home position for a few lengthy periods, during which the indicated measurements are incorrect. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/01/2002 | 1700 | 10/08/2002 | 1645 |
Subject: | SGP/EBBR/E7 - T/RH Ventilator Fans Stopped |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | The temperature and relative humidity gradients are suspect because the ventilator fans were not operating during this time period. The data needs to be carefully inspected to determine when measurements, if any, are correct. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/28/2002 | 0945 | 10/28/2002 | 1530 |
10/28/2002 | 1630 | 10/28/2002 | 1700 |
10/28/2002 | 2130 | 11/04/2002 | 0200 |
Subject: | SGP/EBBR/E8 - Battery Voltage Low |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | When Battery Voltage drops below 10.5 volts, most measurements are suspect and some are often incorrect. The user will need to inspect the data carefully to determine the usability of the data. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/08/2002 | 0000 | 07/09/2002 | 1330 |
Subject: | SGP/EBBR/E12 - Resistance-based Measurements Incorrect |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | When the Reference Temperature in the EBBR enclosure is not reported correctly, it affects all resistance-based measurements that are adjusted with the temperature of the reference junction. |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/01/2002 | 0000 | 07/12/2002 | 1800 |
Subject: | SGP/EBBR/E12 - Wind Direction Incorrect |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | The wind direction and residual wind speed measurements are incorrect. |
Measurements: | sgp30ebbrE12.b1:
sgp15ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/23/2002 | 1700 | 07/23/2002 | 1729 |
08/16/2002 | 1600 | 08/16/2002 | 1629 |
09/03/2002 | 1700 | 09/03/2002 | 1729 |
Subject: | SGP/EBBR/E12 - Data zero following Logger Restart |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | Most of the measurements were zero, which normally happens after a datalogger restart. The restart on 08/16/02 occurred after an EBBR unit changeout. |
Measurements: | sgp30ebbrE12.b1:
sgp15ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/09/2002 | 0000 | 10/29/2002 | 1830 |
Subject: | SGP/EBBR/E12 - Battery Voltage Low |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | Normally, when Battery Voltage drops below 10.5 volts, most measurements are suspect and some are often incorrect. In this particular case, some of the data is suspect when the battery voltage is below 11.0 volts. Soil Heat Flow #3 is normally affected first, with other measurements following at lower voltages. |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/31/2002 | 0700 | 11/12/2002 | 1730 |
11/25/2002 | 1830 | 11/25/2002 | 1900 |
Subject: | SGP/EBBR/E12 - AEM Stuck in One Position |
DataStreams: | sgp30ebbrE12.b1 |
Description: | When the AEM (Automatic Exchange Mechanism) is hung at a home position, the gradients of temperature and relative humidity, and thus sensible and latent heat fluxes, are incorrect. |
Measurements: | sgp30ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/30/2002 | 1800 | 07/30/2002 | 1829 |
Subject: | SGP/EBBR/E7 - Data Reported as Zero following Logger Replacement |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | The data was reported as zero after a data logger replacement. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/01/2002 | 0000 | 07/16/2002 | 1530 |
Subject: | SGP/EBBR/E9 - REPROCESS: Soil Temperature #5 Incorrect |
DataStreams: | sgp30ebbrE9.b1 |
Description: | Soil Temperature #5 was intermittently incorrect. The cause is unknown. The indicated measurements that are calculated using soil temperature are also incorrect. |
Measurements: | sgp30ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/16/2002 | 1500 | 07/16/2002 | 1529 |
Subject: | SGP/EBBR/E9 - REPROCESS: Soil Heat Flow #4 Incorrect |
DataStreams: | sgp30ebbrE9.b1 |
Description: | Soil Heat Flow #4 was incorrect during these periods as a consequence of the repair of Soil Temperature #5. See below for a work-around. |
Measurements: | sgp30ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/24/2002 | 1600 | 09/24/2002 | 1630 |
Subject: | SGP/EBBR/E9 - Net Rad data bad during dome replacement |
DataStreams: | sgp30ebbrE9.b1, sgp5ebbrE9.b1, sgp15ebbrE9.b1 |
Description: | The net radiation measurement was incorrect during replacement of the top dome. Measurements calculated using net radiation are also incorrect. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/16/2002 | 1330 | 11/16/2002 | 1430 |
11/17/2002 | 0700 | 11/19/2002 | 1730 |
Subject: | SGP/EBBR/E9 - AEM Malfunction |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Description: | When the AEM (Automatic Exchange Mechanism) is hung inbetween a home position, the gradients of temperature and relative humidity, and thus sensible and latent heat fluxes, are incorrect. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/01/2002 | 2100 | 11/12/2002 | 2130 |
07/08/2002 | 1800 | 07/19/2002 | 1430 |
08/20/2002 | 2100 | 08/26/2002 | 0300 |
Subject: | SGP/EBBR/E18 - Reprocess: Soil Heat Flow #1 High |
DataStreams: | sgp30ebbrE18.b1 |
Description: | Soil Heat Flow #1 measurements were 2X to offscale high during these periods. See below for a work-around. |
Measurements: | sgp30ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/20/2002 | 2000 | 07/26/2002 | 2030 |
07/27/2002 | 0300 | 07/30/2002 | 2000 |
08/03/2002 | 0130 | 08/06/2002 | 2030 |
08/07/2002 | 0200 | 08/09/2002 | 1900 |
08/12/2002 | 0130 | 08/17/2002 | 2030 |
08/17/2002 | 2130 | 08/19/2002 | 1830 |
Subject: | SGP/EBBR/E18 - Reprocess: Soil Heat Flow #4 High |
DataStreams: | sgp30ebbrE18.b1 |
Description: | Soil Heat Flow #4 measurements were 2X or offscale high during this period when compared with the other soil heat flow measurements. See below for a work-around. |
Measurements: | sgp30ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/12/2002 | 2200 | 11/12/2002 | 2229 |
Subject: | SGP/EBBR/E18 - Data Mostly Zero Following Logger Restart |
DataStreams: | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
Description: | Most of the measurements were zero, which normally happens after a datalogger restart. The restart on 11/12/02 occurred after replacement of soil heat flow #1. |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/12/2002 | 2230 | 11/12/2002 | 2259 |
Subject: | SGP/EBBR/E18 - Data Incorrect After Logger Restart |
DataStreams: | sgp30ebbrE18.b1 |
Description: | Because soil temperature measurements were not available from the beginning of the half hour, the change in soil energy storage could not be calculated and thus is incorrect. There is no fix to this problem. Average Soil Heat Flow, Sensible Heat Flux (h), and Latent Heat Flux (e) can not be calculated using the available, remaining measurements. |
Measurements: | sgp30ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/09/2002 | 1400 | 10/15/2002 | 2030 |
Subject: | SGP/EBBR/E18 - Water in Net Radiometer Dome |
DataStreams: | sgp30ebbrE18.b1 |
Description: | The net radiation measurement is incorrect. Measurements calculated using net radiation are also incorrect. No work-around is possible for this situation. |
Measurements: | sgp30ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/16/2002 | 1720 | 12/10/2002 | 1720 |
Subject: | SGP/EBBR/E12 - REPROCESS: Net Rad Calibration Coefficients |
DataStreams: | sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | The Net Radiometer coefficients were incorrect in the CR10 program when the replacement EBBR unit was installed on 08/16/02. The data can be easily corrected; see the Suggestion area below. |
Measurements: | sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/10/2002 | 1700 | 12/10/2002 | 1729 |
Subject: | SGP/EBBR/E12 - Data Mostly Zero Following Logger Restart |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | Most of the measurements were zero, which normally happens after a datalogger restart. The restart on 12/10/02 occurred after a CR10 program change of net radiometer coefficients and subsequent restart of the CR10. |
Measurements: | sgp30ebbrE12.b1:
sgp15ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/23/2002 | 0000 | 10/31/2002 | 2130 |
Subject: | SGP/EBBR/E13 - Wind Speed Stalling at Low Speeds |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | Perhaps because of excessive bearing wear, the wind speed sensor had a high starting threshhold with wind speeds below about 2 m/s. Wind speed measurements at all speeds were probably too low. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/05/2001 | 1750 | 02/22/2002 | 2100 |
Subject: | SGP/EBBR/E13 - REPROCESS: Soil Moisture #1 Offset1 |
DataStreams: | sgp30ebbrE13.b1 |
Description: | The wrong offset (1.4656) was used in the calibration coefficients for Soil Moisture Probe #1. The correct offset is 14.656. Sensible and Latent Heat Flux measurements are not significantly affected. |
Measurements: | sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/18/2002 | 1210 | 09/19/2002 | 0245 |
Subject: | SGP/EBBR/E13 - AEM Jung between Home Positions |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | When the AEM (Automatic Exchange Mechanism) is hung inbetween home positions, the gradients of temperature and relative humidity, and thus sensible and latent heat fluxes, are incorrect. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/11/2001 | 1920 | 07/09/2002 | 1522 |
Subject: | SGP/EBBR/E15 - REPROCESS: Soil Moisture #1 Offset |
DataStreams: | sgp30ebbrE15.b1 |
Description: | The wrong offset (9.5264) was used in the calibration coefficients for Soil Moisture Probe #1. The correct offset is 0.95264. Sensible and Latent Heat Flux measurements are not significantly affected. |
Measurements: | sgp30ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/22/2002 | 1630 | 07/22/2002 | 1659 |
Subject: | SGP/EBBR/E15 - Data Mostly Zero Following Logger Restart |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | Most of the measurements were zero, which normally happens after a datalogger restart. The restart on 07/22/02 occurred after a pressure sensor replacement. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/10/2002 | 1300 | 12/23/2002 | 1600 |
Subject: | SGP/EBBR/E15 - Pressure Low |
DataStreams: | sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | The pressure measurements were too low during this period, as much as 40 mb too low. This occurred after a restart of the EBBR CR10 datalogger, for unknown reasons. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/08/2002 | 1600 | 04/18/2002 | 1700 |
Subject: | SGP/EBBR/E19 - REPROCESS: Soil Heat Flow #3 Incorrect Calibration |
DataStreams: | sgp30ebbrE19.b1 |
Description: | Soil Heat Flow #3 was incorrect during these periods. To use the measurements, multiply Soil Heat Flow #3 by 0.726. See below for a work-around. |
Measurements: | sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/25/2002 | 2100 | 09/05/2002 | 1430 |
Subject: | SGP/EBBR/E19 - Water in Net Radiometer Dome |
DataStreams: | sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | The net radiation measurement is incorrect. Measurements calculated using net radiation are also incorrect. No work-around is possible for this situation. |
Measurements: | sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/11/2002 | 1600 | 07/11/2002 | 1629 |
Subject: | SGP/EBBR/E19 - Incorrect data after power interruption |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | Most measurements are suspect after a power outage, caused by replacement of the solar panel regulator. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/11/2002 | 1500 | 07/11/2002 | 1529 |
Subject: | SGP/EBBR/E19 - Most Measurements Zero following Restart |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | Most of the measurements were zero, which normally happens after a datalogger restart. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/24/2002 | 1430 | 07/24/2002 | 1500 |
Subject: | SGP/EBBR/E19 - REPROCESS: Soil Temperature #3 Incorrect |
DataStreams: | sgp30ebbrE19.b1 |
Description: | Soil Temperature #3 was incorrect while electronic connections were being checked. The indicated measurements that are calculated using Soil Temperature #3 are also incorrect. See a work-around below. |
Measurements: | sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/25/2002 | 0930 | 07/25/2002 | 1300 |
07/26/2002 | 0900 | 07/26/2002 | 0900 |
07/26/2002 | 1130 | 07/26/2002 | 1130 |
07/27/2002 | 0630 | 07/27/2002 | 0830 |
07/28/2002 | 0300 | 07/28/2002 | 0630 |
07/28/2002 | 0930 | 07/28/2002 | 1230 |
07/29/2002 | 0800 | 07/29/2002 | 0830 |
08/08/2002 | 1600 | 08/08/2002 | 1600 |
08/09/2002 | 0230 | 08/09/2002 | 0300 |
08/09/2002 | 0730 | 08/09/2002 | 0730 |
08/09/2002 | 0930 | 08/09/2002 | 1100 |
08/09/2002 | 1400 | 08/09/2002 | 1600 |
08/10/2002 | 0130 | 08/10/2002 | 0230 |
08/10/2002 | 1200 | 08/10/2002 | 1330 |
08/13/2002 | 0600 | 08/13/2002 | 0700 |
08/13/2002 | 2030 | 08/13/2002 | 2030 |
08/13/2002 | 2130 | 08/14/2002 | 0330 |
08/14/2002 | 1300 | 08/14/2002 | 1330 |
08/21/2002 | 1630 | 08/22/2002 | 1430 |
Subject: | SGP/EBBR/E19 - REPROCESS: Soil Heat Flow #1 Incorrect |
DataStreams: | sgp30ebbrE19.b1 |
Description: | Soil Heat Flow #1 was incorrect during these periods. See below for a work-around. |
Measurements: | sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/17/2002 | 1430 | 10/31/2002 | 1600 |
Subject: | SGP/EBBR/E19 - Pressure Low |
DataStreams: | sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | The barometric pressure sensor read about 10 mb low during this period because a lead from the sensor was loose or disconnected. The error does not affect any other measurements. |
Measurements: | sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/13/2002 | 1830 | 11/13/2002 | 1859 |
12/11/2002 | 1900 | 12/11/2002 | 1929 |
Subject: | SGP/EBBR/E20 - Data Zero following logger restarts |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | Most of the measurements were zero, which normally happens after a datalogger restart. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/09/2002 | 1705 | 09/18/2002 | 1700 |
Subject: | SGP/EBBR/E20 - REPROCESS: Soil Temperature #2 Incorrect |
DataStreams: | sgp30ebbrE20.b1 |
Description: | Soil Temperature #2 was incorrect, apparently from am loose wire after replacement of the EBBR unit. The indicated measurements that are calculated using soil temperature are also incorrect. See the work-around below. |
Measurements: | sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/13/2002 | 1900 | 11/13/2002 | 1929 |
12/11/2002 | 2000 | 12/11/2002 | 2029 |
Subject: | SGP/EBBR/E20 - Change of Energy Storage Incorrect |
DataStreams: | sgp30ebbrE20.b1 |
Description: | Because soil temperature measurements were not available from the beginning of the half hour, the change in soil energy storage could not be calculated and thus is incorrect. There is no fix to this problem. Average Soil Heat Flow, Sensible Heat Flux (h), and Latent Heat Flux (e) can not be calculated using the available, remaining measurements. |
Measurements: | sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/23/2002 | 0215 | 10/30/2002 | 2000 |
Subject: | SGP/EBBR/E20 - Battery voltage low |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | When Battery Voltage drops below 10.5 volts, most measurements are suspect and some are often incorrect. The user will need to inspect the data carefully to determine the useability of the data. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/04/2002 | 1200 | 12/05/2002 | 1630 |
Subject: | SGP/EBBR/E20 - Wind speed and direction frozen |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE2.b1 |
Description: | Wind Speed and Dierction were frozen in place by an ice storm. |
Measurements: | sgp15ebbrE20.b1:
sgp5ebbrE2.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/04/2002 | 1715 | 09/04/2002 | 1730 |
10/13/2002 | 1300 | 10/15/2002 | 1930 |
10/15/2002 | 2145 | 10/15/2002 | 1630 |
10/13/2002 | 1230 | 10/13/2002 | 1234 |
Subject: | SGP/EBBR/E22 - AEM Hung between home positions |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | When the AEM (Automatic Exchange Mechanism) is hung inbetween home positions, the gradients of temperature and relative humidity, and thus sensible and latent heat fluxes, are incorrect. There is no work-around to this problem. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/29/2002 | 0830 | 08/07/2002 | 1800 |
Subject: | SGP/EBBR/E22 - Water inside Broken Net Rad Dome |
DataStreams: | sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | The net radiation measurement is incorrect. Measurements calculated using net radiation are also incorrect. No work-around is possible for this situation. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/07/2002 | 1630 | 08/07/2002 | 1730 |
Subject: | SGP/EBBR/E20 - AEM Hung between Home Positions |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | When the AEM (Automatic Exchange Mechanism) is hung inbetween home positions, the gradients of temperature and relative humidity, and thus sensible and latent heat fluxes, are incorrect. There is no work-around to this problem. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/30/2002 | 2030 | 01/08/2003 | 1730 |
Subject: | SGP/EBBR/E20 - REPROCESS: Heat Flow #3 and #5 Incorrect |
DataStreams: | sgp30ebbrE20.b1 |
Description: | Soil Heat Flow #3 was incorrect because part of the lead wire was cut; apparent vandalism. Soil Heat Flow #5 had been incorrect because of other causes, but was replaced at the same time as #3. |
Measurements: | sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/09/2002 | 1705 | 08/09/2002 | 1730 |
Subject: | SGP/EBBR/E20 - Data incorrect during instrument maintenance |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | Most of the data was incorrect just after the EBBR unit was replaced. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/03/2002 | 1200 | 12/04/2002 | 0400 |
12/06/2002 | 2000 | 12/07/2002 | 0230 |
12/08/2002 | 0230 | 12/09/2002 | 0500 |
12/09/2002 | 0930 | 12/20/2002 | 1830 |
12/20/2002 | 2100 | 12/21/2002 | 1630 |
12/21/2002 | 1730 | 12/22/2002 | 0100 |
12/22/2002 | 0130 | 12/30/2002 | 2000 |
Subject: | SGP/EBBR/E20 - REPROCESS: Soil Heat Flow #5 Incorrect |
DataStreams: | sgp30ebbrE20.b1 |
Description: | Soil Heat Flow #5 was incorrect during these periods. See below for a work-around. |
Measurements: | sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/19/2002 | 1700 | 11/19/2002 | 2200 |
11/18/2002 | 1900 | 11/18/2002 | 2200 |
11/20/2002 | 1430 | 11/20/2002 | 2200 |
11/21/2002 | 2000 | 11/21/2002 | 2030 |
11/22/2002 | 1630 | 11/22/2002 | 2000 |
11/23/2002 | 1430 | 11/23/2002 | 2230 |
11/24/2002 | 1700 | 11/24/2002 | 2200 |
11/27/2002 | 2030 | 11/27/2002 | 2300 |
11/28/2002 | 1700 | 11/28/2002 | 2300 |
11/29/2002 | 1330 | 11/29/2002 | 2130 |
12/01/2002 | 1930 | 12/01/2002 | 2030 |
12/02/2002 | 1430 | 12/02/2002 | 2230 |
Subject: | SGP/EBBR/E20 - REPROCESS: Soil Temperature #1 Incorrect |
DataStreams: | sgp30ebbrE20.b1 |
Description: | Soil Temperature #1 was intermittently incorrect. The cause is unknown. The indicated measurements that are calculated using soil temperature are also incorrect. |
Measurements: | sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/19/2002 | 1430 | 07/20/2002 | 2000 |
07/26/2002 | 2030 | 07/27/2002 | 0300 |
07/30/2002 | 2000 | 08/03/2002 | 0130 |
08/06/2002 | 2030 | 08/07/2002 | 0200 |
08/09/2002 | 1900 | 08/12/2002 | 0130 |
08/17/2002 | 2030 | 08/17/2002 | 2130 |
08/19/2002 | 1830 | 08/20/2002 | 2100 |
Subject: | SGP/EBBR/E18 - REPROCESS: Soil Heat Flow #1 and #4 High |
DataStreams: | sgp30ebbrE18.b1 |
Description: | Soil Heat Flow #1 and #4 measurements were 2X to offscale high during these periods. See suggestions for a work-around. |
Measurements: | sgp30ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/30/2002 | 2030 | 01/08/2003 | 1730 |
Subject: | SGP/EBBR/E20 - REPROCESS: Soil Temperature #3 Incorrect |
DataStreams: | sgp30ebbrE20.b1 |
Description: | Soil Temperature #3 was incorrect. The cause is apparently vandalism; the probe leads were cut. The indicated measurements that are calculated using soil temperature are also incorrect. DQR D030117.19 already provides the work-around for calculating correct sensible (h) and latent heat (e) fluxes, as soil heat flow #3 and soil heat flow #5 were also incorrect during this period. That work-around is repeated below. |
Measurements: | sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/11/2002 | 1600 | 07/11/2002 | 1629 |
Subject: | SGP/EBBR/E26 - Measurements zero following logger restart |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | Most of the measurements were zero, which normally happens after a datalogger restart. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/01/2002 | 0000 | 07/11/2002 | 1300 |
Subject: | SGP/EBBR/E26 - Multiplexer Problem |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | One of the multiplexers exhibited dropouts of the channels, beginning with the last channel at temperatures above about 30 C and moving backward through the channels as the temperature rose. High temperatures affected the switching speed of the multiplexer. This is the first time that this behavior has been documented in a Campbell Scientific multiplexer. The multiplexer had been in continuous field use for 10 years and had experienced probably more switching cycles than what is normally expected of the device in that period. The multiplexer was replaced to solve the problem. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/27/2002 | 0830 | 09/05/2002 | 1700 |
Subject: | SGP/EBBR/E26 - Water in broken Net Radiometer Dome |
DataStreams: | sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | The net radiation measurement is incorrect. Measurements calculated using net radiation are also incorrect. No work-around is possible for this situation. |
Measurements: | sgp30ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/01/2002 | 0845 | 11/06/2002 | 0130 |
Subject: | SGP/EBBR/E26 - Battery Voltage Low |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | When Battery Voltage drops below 10.5 volts, most measurements are suspect and some are often incorrect. The user will need to inspect the data carefully to determine the useability of the data. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/20/2002 | 1300 | 10/20/2002 | 1500 |
Subject: | SGP/EBBR/E20 - Resistance-based Measurements Incorrect |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | When the Reference Temperature in the EBBR enclosure is not reported correctly, it affects all resistance-based measurements that are adjusted with the temperature of the reference junction. No work-around is possible for this situation. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/14/2002 | 1430 | 01/09/2003 | 1600 |
Subject: | SGP/EBBR/E19 - Soil Temperatures #2 and #3 Intermittently Bad |
DataStreams: | sgp30ebbrE19.b1 |
Description: | Soil Temperatures #2 and #3 were intermittently incorrect. #3 appeared to be okay after the J-panel was replaced. #2 continues to be flaky, suggesting that the sensor itself has a problem. A DQPR has been submitted to replace #2. The indicated measurements that are calculated using soil temperature are also suspect. However, the intermittent problems normally occurred during night times and were fairly small in magnitude, thereby having little effect on the sensible and latent heat fluxes. |
Measurements: | sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/09/2003 | 1700 | 01/09/2003 | 1729 |
Subject: | SGP/EBBR/E19 - Measurements zero following logger restart |
DataStreams: | sgp5ebbrE19.b1, sgp15ebbrE19.b1, sgp30ebbrE19.b1 |
Description: | Most of the measurements were zero, which normally happens after a datalogger restart. The restart on 01/09/03 occurred after J-panel replacements. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/09/2003 | 1730 | 01/09/2003 | 1759 |
Subject: | SGP/EBBR/E19 - Change of Energy Storage Incorrect |
DataStreams: | sgp30ebbrE19.b1 |
Description: | Because soil temperature measurements were not available from the beginning of the half hour, the change in soil energy storage could not be calculated and thus is incorrect. There is no fix to this problem. Average Soil Heat Flow, Sensible Heat Flux (h), and Latent Heat Flux (e) can not be calculated using the available, remaining measurements. |
Measurements: | sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/09/2002 | 1730 | 12/11/2002 | 1930 |
Subject: | SGP/EBBR/E20 - REPROCESS: Net Rad Coefficients Incorrect |
DataStreams: | sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | The Net Radiometer coefficients were incorrect in the CR10 program when the replacement EBBR unit was installed on 08/09/02. The data can be easily corrected; see the Suggestion area below. |
Measurements: | sgp5ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/20/2002 | 1830 | 12/20/2002 | 2100 |
12/21/2002 | 1630 | 12/21/2002 | 1730 |
12/22/2002 | 0100 | 12/22/2002 | 0130 |
Subject: | SGP/EBBR/E20 - REPROCESS: Soil Temperature #1 and Soil Heat Flow #5 Incorrect |
DataStreams: | sgp30ebbrE20.b1 |
Description: | Soil Temperature #1 and Soil Heat Flow #5 were incorrect during these periods. The indicated calculated measurements are also incorrect. |
Measurements: | sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/04/2002 | 1015 | 01/10/2003 | 1600 |
Subject: | SGP/EBBR/E26 - AEM Hung |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | The AEM hung inbetween home positions during the period. The Bowen ratio, gradient measurements, sensible heat flux, and latent heat flux are incorrect. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/10/2003 | 1900 | 01/10/2003 | 1904 |
Subject: | SGP/EBBR/E26 - Data zeros following logger restart |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | Most measurements were zero after restart of the CR10 datalogger, after a calibration changeout of the EBBR unit. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/14/2003 | 1725 | 01/23/2003 | 1715 |
Subject: | SGP/EBBR/E26 - Left Air Temperature Incorrect |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | The left air temperature was incorrect during the period. Poor electronic connections may have caused this. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/19/2003 | 1900 | 05/16/2003 | 1500 |
Subject: | SGP/EBBR/E26 - E26 EBBR Soil Heat Flow #1 Incorrect |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1 |
Description: | Soil Heat Flow #1 was incorrect much of the time. The sensor was replaced. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/07/2003 | 1030 | 06/18/2003 | 2359 |
Subject: | SGP/EBBR/E27 - E27 EBBR Incorrect Units for Soil Heat Capacity (cs1, cs2, cs3, cs4, cs5) |
DataStreams: | sgp30ebbrE27.b1 |
Description: | The units for Soil Heat Capacity (cs1, cs2, cs3, cs4, cs5) in the EBBR 30 minute netcdf header have been incorrect from the beginning of processing the data into netcdf files. The correct units are MJ/m^3/C. This does not affect the data quality. |
Measurements: | sgp30ebbrE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/11/2003 | 1730 | 07/15/2003 | 2000 |
Subject: | SGP/EBBR/E7 - Soil Heat Flow #5 Incorrect |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | Soil Heat Flow Plate #5 data was intermittently incorrect at first (values too small negative), then was almost constantly so during the last half of the period. Comparison of the individual soil heat flows will quickly show the times that soil heat flow #5 was incorrect. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/01/2001 | 0000 | 04/08/2002 | 2359 |
Subject: | SGP/EBBR/E25 - Incorrect Units for Soil Heat Capacity (cs1, cs2, cs3, cs4, cs5) in 30EBBR |
DataStreams: | sgp30ebbrE25.b1 |
Description: | The units for Soil Heat Capacity (cs1, cs2, cs3, cs4, cs5) in the EBBR 30 minute netcdf header have been incorrect from the beginning of processing the data into netcdf files. The correct units are MJ/m^3/C. This does not affect the data quality. |
Measurements: | sgp30ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/22/1997 | 0000 | 04/08/2002 | 2359 |
Subject: | SGP/EBBR/E25 - Incorrect Units for Soil Heat Capacity (cs1, cs2, cs3, cs4, cs5) in 30EBBR |
DataStreams: | sgp30baebbrE25.c1, sgp30ebbrE25.b1 |
Description: | The units for Soil Heat Capacity (cs1, cs2, cs3, cs4, cs5) in the EBBR 30 minute netcdf header have been incorrect from the beginning of processing the data into netcdf files. The correct units are MJ/m^3/C. This does not affect the data quality. (NOTE: The actual begin date of some of these data streams is as early as 5/22/97. This DQR applies to all 30m EBBR data.) |
Measurements: | sgp30baebbrE25.c1:
sgp30ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/01/2003 | 1000 | 02/01/2003 | 1000 |
Subject: | SGP/EBBR/E20 - RHtop Zero |
DataStreams: | sgp30ebbrE20.b1 |
Description: | The RHtop value in the 30 minute file is zero. 5 and 15 minute files show RHtop correctly. This behavior has been seen infrequently and can probably be attributed to a CR10 communication error. No other data, including the fluxes, are affected. |
Measurements: | sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/09/2003 | 1630 | 01/09/2003 | 1700 |
Subject: | SGP/EBBR/E19 - Soil Temperature #2 Incorrect |
DataStreams: | sgp30ebbrE19.b1, sgp15ebbrE19.b1 |
Description: | Soil Temperature #2 was incorrect during J-panel replacements. |
Measurements: | sgp15ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/09/2003 | 1700 | 01/09/2003 | 1700 |
02/06/2003 | 1730 | 02/06/2003 | 1730 |
04/03/2003 | 1830 | 04/03/2003 | 1830 |
Subject: | SGP/EBBR/E19 - All Data Zeros |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | All data (5, 15, and 30 minute) was zero after replacement of sensors and restart of the CR10. This is a common occurrence. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/09/2003 | 1730 | 01/09/2003 | 1730 |
Subject: | SGP/EBBR/E19 - Several Soil Measurements Incorrect |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1 |
Description: | During replacement of a J-panel, several soil measurements were incorrect. |
Measurements: | sgp15ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/26/2003 | 1700 | 01/26/2003 | 1700 |
01/26/2003 | 1800 | 01/26/2003 | 1800 |
Subject: | SGP/EBBR/E19 -Several Soil Measurements Were Incorrect |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1 |
Description: | Several soil measurements were incorrect during replacement of soil temperature probe #2. |
Measurements: | sgp15ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/28/2003 | 1330 | 03/02/2003 | 0530 |
03/02/2003 | 1730 | 03/03/2003 | 0030 |
Subject: | SGP/EBBR/E19 - AEM Hung in home30 position |
DataStreams: | sgp30ebbrE19.b1 |
Description: | The automatic exchange mechanism (AEM) hung in the home30 position, resulting in sensible and latent heat fluxes being incorrect. |
Measurements: | sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/09/2003 | 1730 | 04/03/2003 | 1900 |
Subject: | SGP/EBBR/E19 - Soil Heat Flow #3 Incorrect |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1 |
Description: | Soil Heat Flow #3 was a flat -24 W/m2 for most of this period. |
Measurements: | sgp15ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/03/2003 | 1900 | 04/03/2003 | 1900 |
Subject: | SGP/EBBR/E19 - Soil Heat Flow#1 Incorrect |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1 |
Description: | Soil Heat Flow #1 was incorrect during replacement of some other sensors. |
Measurements: | sgp15ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/07/2003 | 1030 | 07/09/2003 | 1700 |
Subject: | SGP/EBBR/E27 - Soil Heat Flow #2 Incorrect |
DataStreams: | sgp15ebbrE27.b1, sgp30ebbrE27.b1 |
Description: | Soil Heat Flow sensor #2 was intermittently incorrect at first and then constantly so later in the period. |
Measurements: | sgp15ebbrE27.b1:
sgp30ebbrE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/15/2003 | 1600 | 05/15/2003 | 1700 |
Subject: | SGP/EBBR/E19 - Many Measurements Incorrect |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | Many measurements were incorrect during replacement of a multiplexer. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/01/2003 | 0000 | 01/10/2003 | 1630 |
02/06/2003 | 1730 | 02/18/2003 | 1715 |
02/23/2003 | 1530 | 03/04/2003 | 1645 |
04/15/2003 | 1645 | 04/15/2003 | 1700 |
Subject: | SGP/EBBR/E12 - AEM Hung Inbetween Home Positions |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | The AEM was hung inbetween home positions, causing many measurements to be incorrect. |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/01/2003 | 2145 | 04/01/2003 | 2200 |
Subject: | SGP/EBBR/E18 - AEM Hung |
DataStreams: | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
Description: | The AEM was hung inbetween home positions, causing many measurements to be incorrect. |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/15/2003 | 0100 | 06/23/2003 | 1900 |
06/27/2003 | 1030 | 06/27/2003 | 1300 |
06/30/2003 | 0030 | 06/30/2003 | 0730 |
06/30/2003 | 1000 | 06/30/2003 | 2000 |
07/05/2003 | 0400 | 07/10/2003 | 0500 |
07/10/2003 | 1430 | 07/20/2003 | 0300 |
07/20/2003 | 1730 | 07/21/2003 | 0230 |
07/21/2003 | 1930 | 07/22/2003 | 0230 |
07/22/2003 | 1730 | 07/23/2003 | 0600 |
07/23/2003 | 1630 | 07/23/2003 | 1900 |
Subject: | SGP/EBBR/E27 - Soil Heat Flow #4 Incorrect |
DataStreams: | sgp15ebbrE27.b1, sgp30ebbrE27.b1 |
Description: | Soil Heat flow sensor #4 outputs were often very small negative numbers. See below for a work-around. |
Measurements: | sgp15ebbrE27.b1:
sgp30ebbrE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/07/2003 | 1030 | 05/15/2003 | 1600 |
05/18/2003 | 1800 | 05/28/2003 | 1700 |
Subject: | SGP/EBBR/E27 - Reference T Sensor Failure |
DataStreams: | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
Description: | The reference T sensor was not working much of the time, affecting all other temperature sensors. Sensible and latent heat fluxes were incorrect when this occurred. |
Measurements: | sgp5ebbrE27.b1:
sgp15ebbrE27.b1:
sgp30ebbrE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/07/2003 | 1030 | 06/12/2003 | 1800 |
Subject: | SGP/EBBR/E27 - Wind Direction Hung |
DataStreams: | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
Description: | The Wind Direction frequently hung at values in the 70s or 80s. No other data was affected. Fluxes were not affected. |
Measurements: | sgp5ebbrE27.b1:
sgp15ebbrE27.b1:
sgp30ebbrE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/24/2003 | 1635 | 03/06/2003 | 1520 |
Subject: | SGP/EBBR/E26 -Right Air Temperature Incorrect |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | One of the Right Air Temperature connections was loose, producing incorrect data. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/01/2003 | 0000 | 06/25/2003 | 1700 |
Subject: | SGP/EBBR/E22 - home30 too large |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | The home30 measurements sometimes exceeded 34.9999, resulting in the qc flag being set. However, this does not create any data quality issues. The qc flag is set just to provide an alert that the automatic exchange mechanism may have a problem; in this case data was not affected. An electronics repair solved the problem. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/01/2003 | 0000 | 01/08/2003 | 1730 |
Subject: | SGP/EBBR/E20 - Soil Heat Flow#5 Incorrect |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1 |
Description: | Soil Heat Flow #5 was producing incorrect values. Average soil heat flow, sensible heat flux, and latent heat flux were affected. |
Measurements: | sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/08/2003 | 2100 | 01/08/2003 | 2100 |
Subject: | SGP/EBBR/E20 - Soil Storage Incorrect After Startup |
DataStreams: | sgp30ebbrE20.b1 |
Description: | For the half hour data after CR10 datalogger startup, stored soil temperature values do not exist from the beginning of the half hour. Therefore soil heat storage can not be calculated correctly. This affects numerous measurements that are calculated using soil heat storage. |
Measurements: | sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/23/2003 | 2000 | 03/13/2003 | 1745 |
Subject: | SGP/EBBR/E9 - AEM Hung Inbetween Home Positions |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Description: | The AEM was hung inbetween home positions, causing many measurements to be incorrect. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/08/2003 | 1545 | 05/08/2003 | 1615 |
Subject: | SGP/EBBR/E2 - 6 Month Checks |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp30ebbrE2.b1:
sgp15ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/22/2003 | 1650 | 04/22/2003 | 1720 |
Subject: | SGP/EBBR/E7 - 6 Month Checks |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/06/2003 | 1750 | 05/06/2003 | 1825 |
Subject: | SGP/EBBR/E8 - 6 Month Checks |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/24/2003 | 1800 | 04/24/2003 | 1830 |
Subject: | SGP/EBBR/E9 - 6 Month Checks |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/15/2003 | 1630 | 04/15/2003 | 1700 |
Subject: | SGP/EBBR/E12 - 6 Month Checks |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/01/2003 | 1905 | 05/01/2003 | 1935 |
Subject: | SGP/EBBR/E13 - 6 Month Checks |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/29/2003 | 1540 | 04/29/2003 | 1655 |
Subject: | SGP/EBBR/E15 - 6 Month Checks |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/15/2003 | 2120 | 04/15/2003 | 2215 |
Subject: | SGP/EBBR/E18 - 6 Month Checks |
DataStreams: | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/09/2003 | 1800 | 05/09/2003 | 1800 |
Subject: | SGP/EBBR/E9 - Soil Heat Flow #1 Incorrect |
DataStreams: | sgp30ebbrE9.b1, sgp15ebbrE9.b1 |
Description: | Soil Heat Flow #1 was incorrect. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/01/2003 | 1440 | 05/01/2003 | 1515 |
Subject: | SGP/EBBR/E19 - 6 Month Checks |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/16/2003 | 1950 | 04/16/2003 | 2045 |
Subject: | SGP/EBBR/E20 - 6 Month Checks |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/30/2003 | 1720 | 04/30/2003 | 1800 |
Subject: | SGP/EBBR/E22 - 6 Month Checks |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/17/2003 | 1445 | 04/17/2003 | 1535 |
Subject: | SGP/EBBR/E26 - 6 Month Checks |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/15/2003 | 1600 | 05/15/2003 | 1645 |
Subject: | SGP/EBBR/E27 - 6 Month Checks |
DataStreams: | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp5ebbrE27.b1:
sgp15ebbrE27.b1:
sgp30ebbrE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/14/2003 | 1100 | 05/30/2003 | 1700 |
Subject: | SGP/EBBR/E9 - Battery Voltage Low |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Description: | Many of the measurements are incorrect much of the time during this period, when the battery voltage was 10.9 volts or less. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/16/2003 | 1245 | 01/28/2003 | 1915 |
02/06/2003 | 1400 | 02/11/2003 | 1930 |
02/23/2003 | 0945 | 03/11/2003 | 1800 |
03/19/2003 | 1900 | 03/25/2003 | 1800 |
Subject: | SGP/EBBR/E8 - AEM Hung Inbetween Home Positions |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The AEM was hung inbetween home positions, causing many measurements to be incorrect. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/16/2003 | 0000 | 03/22/2003 | 1445 |
Subject: | SGP/EBBR/E8 - Battery Voltage Low |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | Many of the measurements are incorrect much of the time during this period, when the battery voltage was 10.5 volts or less. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/01/2003 | 0000 | 04/24/2003 | 1625 |
Subject: | SGP/EBBR/E2 - Soil Heat Flow #2 Incorrect |
DataStreams: | sgp30ebbrE2.b1, sgp15ebbrE2.b1 |
Description: | Soil Heat Flow #2 producing incorrect values much of the time. Average soil heat flow, sensible heat flux, and latent heat flux were affected. |
Measurements: | sgp15ebbrE2.b1:
sgp30ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/24/2003 | 1630 | 04/24/2003 | 1630 |
Subject: | SGP/EBBR/E2 - Data All Zeros |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | All data (5, 15, and 30 minute) were zero after replacement of a sensor and restart of the CR10. This is a common occurrence. |
Measurements: | sgp15ebbrE2.b1:
sgp30ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/16/2003 | 2200 | 04/16/2003 | 2200 |
Subject: | SGP/EBBR/E2 - Bottom RH Incorrect |
DataStreams: | sgp5ebbrE2.b1, sgp30ebbrE2.b1 |
Description: | The bottom RH was zero (and incorrect) for this period. |
Measurements: | sgp30ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/24/2003 | 0900 | 03/11/2003 | 1800 |
Subject: | SGP/EBBR/E7 - AEM Hung Inbetween Home Position |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | The AEM was hung inbetween home positions, causing many measurements to be incorrect. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/11/2001 | 1930 | 08/05/2003 | 1630 |
Subject: | SGP/EBBR/E15 - Wind Direction 10 Degrees Low |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | The wind direction at E15 EBBR was 10 degrees low for the period (since installation of the EBBR unit on 10/11/01). The wind direction sensor was apparently aligned to magnetic north instead of true north at installation. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/20/2003 | 0300 | 07/31/2003 | 1600 |
Subject: | SGP/EBBR/E2 - Soil Heat Flow #1 Incorrect |
DataStreams: | sgp30ebbrE2.b1, sgp15ebbrE2.b1 |
Description: | Soil Heat Flow #1 was very large most of the period. See below for a work-around. |
Measurements: | sgp15ebbrE2.b1:
sgp30ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/10/2003 | 0030 | 07/10/2003 | 0030 |
Subject: | SGP/EBBR/E2 - Soil Heat Flow #3 Spike |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1 |
Description: | Soil heat flow #3 spiked to a value much greater than the other soil heat flows. However, this did not affect sensible and latent heat fluxes outside the error of the EBBR system. |
Measurements: | sgp30ebbrE2.b1:
sgp15ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/19/2003 | 1900 | 06/19/2003 | 1900 |
Subject: | SGP/EBBR/E15 - Top Humidity reported as zero in 30 minute data. |
DataStreams: | sgp30ebbrE15.b1 |
Description: | The top relative humidity was reported as zero in the 30 minute data. This is an apparent misreporting by the CR10 datalogger or is caused by a communications problem in accessing the data from the datalogger. It is very infrequently seen; it's actual cause is unknown. No other data was affected. |
Measurements: | sgp30ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/12/2003 | 0230 | 06/12/2003 | 0300 |
06/13/2003 | 1200 | 06/13/2003 | 1230 |
Subject: | SGP/EBBR/E26 - Soil Heat Flow #5 Large Positive |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1 |
Description: | Soil Heat Flow #5 was large positive for these two periods, for an unknown reason. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/25/2003 | 1630 | 06/25/2003 | 1630 |
Subject: | SGP/EBBR/E27 - Soil Heat Flow #1 Large |
DataStreams: | sgp15ebbrE27.b1, sgp30ebbrE27.b1 |
Description: | Soil Heat Flow #1 was large for this period, for an unknown reason. |
Measurements: | sgp15ebbrE27.b1:
sgp30ebbrE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/09/2003 | 1700 | 07/09/2003 | 1700 |
Subject: | SGP/EBBR/E27 - Incorrect Data From Logger Restart |
DataStreams: | sgp30ebbrE27.b1 |
Description: | Some 30 minute was zero or incorrect after restart of the CR10. This is a common occurrence when sensor replacements are made or power outages occur. |
Measurements: | sgp30ebbrE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/12/2003 | 1800 | 09/02/2003 | 1800 |
Subject: | SGP/EBBR/E8 - Pressure Sensor Measurements Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgpebbrE8.01 |
Description: | The EBBR barometer did not agree with the SMOS at all times. However this problem only rarely affected the sensible and latent heat flux values. Replacement of the barometer on 29 july 2003, 12 aug 2003, and 26 aug 2003 caused some obvious disruption in data flow. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
sgpebbrE8.01:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/24/2003 | 1230 | 08/12/2003 | 1615 |
Subject: | SGP/EBBR/E9 - AEM Hung in home 30 position. |
DataStreams: | sgp30ebbrE9.b1, sgp5ebbrE9.b1, sgp15ebbrE9.b1 |
Description: | The AEM was hung in the home 30 position. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/27/2003 | 1330 | 10/07/2003 | 2000 |
Subject: | SGP/EBBR/E7 - AEM Malfunction, Some Data Incorrect |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | After a while, the teflon slides on the AEMs warp from heat and use. This tends to make the AEM drag, jam, and get hung. This happened at E7 EBBR. Replacement of the AEM resolved the problem. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/16/2003 | 0330 | 07/23/2003 | 1830 |
07/24/2003 | 1500 | 08/06/2003 | 1730 |
Subject: | SGP/EBBR/E22 - AEM Hung Inbetween Home Positions |
DataStreams: | sgp30ebbrE22.b1, sgp5ebbrE22.b1, sgp15ebbrE22.b1 |
Description: | The AEM was hung inbetween home positions. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/16/2003 | 1745 | 09/17/2003 | 1900 |
Subject: | SGP/EBBR/E22 - AEM Home Signals Too High |
DataStreams: | sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | The AEM home signals were too large at times during the indicated period; it has been determined by the vendor that this is usually caused by electrical noise in the vicinity. When the 30 minute data shows a home 30 signal of 35 or more, the bowen ratio, and sensible and latent heat flux measurements are incorrect Removing the jumper between 4L and AG on the CR10 and adding a jumper to Multiplexer 2, Set 6, between L1 and L2 fixed the problem. The vendor intends to make this modification to all EBBR systems as they return for re-calibration. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/29/2003 | 0300 | 09/25/2003 | 1600 |
Subject: | SGP/EBBR/E2 - Water in Net Radiometer, Surface Fluxes Incorrect |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | Water got inside the bird-pecked hole in the net radiometer during a rain. The net radiation and therefore latent and sensible heat fluxes are incorrect. |
Measurements: | sgp30ebbrE2.b1:
sgp15ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/13/2003 | 0000 | 11/18/2003 | 2000 |
Subject: | SGP/EBBR/E7 - Reprocess: Soil Heat Flow #5 Offscale, Surface Fluxes Incorrect |
DataStreams: | sgp30ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Soil Heat Flow #5 went offscale at times. Latent and Sensible Heat Fluxes were incorrect when this occurred. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/26/2003 | 1800 | 08/26/2003 | 1800 |
Subject: | SGP/EBBR/E7 - Data all zero |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | Data values were all zero after restarting the CR10 datalogger. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/30/2003 | 1400 | 09/04/2003 | 1445 |
Subject: | SGP/EBBR/E19 - Water in Net Radiometer; Surface Fluxes are Incorrect |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | The net radiometer top dome was punctured by migrating birds. This alone does not cause incorrect net radiation measurements. However, when it subsequently rains, water collects in the bottom dome, causing the net radiation measurements to be too large during the day and too small at night. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/21/2003 | 1600 | 08/21/2003 | 1600 |
Subject: | SGP/EBBR/E26 - Soil Storage Incorrect; Surface Fluxes Incorrect |
DataStreams: | sgp30ebbrE26.b1 |
Description: | The half hour after restart of the CR10 datalogger, the soil temperatures for the beginning of the period are missing. Therefore the soil energy storage and surface flux values are also incorrect. |
Measurements: | sgp30ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/30/2003 | 0630 | 09/03/2003 | 1700 |
Subject: | SGP/EBBR/E27 - Water in Net Radiometer; Surface Fluxes are Incorrect |
DataStreams: | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
Description: | The net radiometer top dome was punctured by migrating birds. This alone does not cause incorrect net radiation measurements. However, when it subsequently rains, water collects in the bottom dome, causing the net radiation measurements to be too large during the day and too small at night. |
Measurements: | sgp5ebbrE27.b1:
sgp15ebbrE27.b1:
sgp30ebbrE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/20/2003 | 1600 | 08/20/2003 | 1600 |
Subject: | SGP/EBBR/E27 - 15 and 30 Minute Data All Zeros |
DataStreams: | sgp15ebbrE27.b1, sgp30ebbrE27.b1 |
Description: | 15 and 30 minute measurements were all zero after restarting the CR10 datalogger. |
Measurements: | sgp15ebbrE27.b1:
sgp30ebbrE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/26/2003 | 1830 | 08/26/2003 | 1830 |
Subject: | SGP/EBBR/E7 - Energy Storage Incorrect, Fluxes Incorrect |
DataStreams: | sgp30ebbrE7.b1 |
Description: | The half hour after restart of the CR10 datalogger, the soil temperatures for the beginning of the period are missing. Therefore the soil energy storage and surface flux values are also incorrect. |
Measurements: | sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/29/2003 | 0230 | 09/05/2003 | 1730 |
Subject: | SGP/EBBR/E7 - Water in Net Radiometer Domes; Surface Fluxes are Incorrect |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | The net radiometer top dome was punctured by migrating birds. This alone does not cause incorrect net radiation measurements. However, when it subsequently rains, water collects in the bottom dome, causing the net radiation measurements to be too large during the day and too small at night. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/31/2003 | 1630 | 07/31/2003 | 1630 |
Subject: | SGP/EBBR/E2 - 15 and 30 Minute Data are Zeros |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1 |
Description: | 15 and 30 minute measurements were zero after restarting the CR10 datalogger. |
Measurements: | sgp30ebbrE2.b1:
sgp15ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/03/2003 | 1900 | 09/03/2003 | 1930 |
09/06/2003 | 1900 | 09/06/2003 | 1900 |
09/07/2003 | 1700 | 09/07/2003 | 1730 |
09/07/2003 | 1930 | 09/07/2003 | 1930 |
09/12/2003 | 1900 | 09/12/2003 | 2030 |
Subject: | SGP/EBBR/E2 - Soil Heat #1 Offscale |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1 |
Description: | Soil Heat Flow #1 was intermittently offscale. The reason is not clear. |
Measurements: | sgp15ebbrE2.b1:
sgp30ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/27/2003 | 0230 | 09/09/2003 | 1730 |
Subject: | SGP/EBBR/E8 - Water in Net Radiometer, Surface Fluxes Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The net radiometer top dome was punctured by migrating birds. This alone does not cause incorrect net radiation measurements. However, when it subsequently rains, water collects in the bottom dome, causing the net radiation measurements to be too large during the day and too small at night. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/02/2003 | 0600 | 08/04/2003 | 1800 |
Subject: | SGP/EBBR/E9 - Water in Net Radiometer; Surface Fluxes are Incorrect |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Description: | The net radiometer top dome was punctured by migrating birds. This alone does not cause incorrect net radiation measurements. However, when it subsequently rains, water collects in the bottom dome, causing the net radiation measurements to be too large during the day and too small at night. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/23/2003 | 1830 | 05/01/2003 | 2000 |
Subject: | SGP/EBBR/E13 - Water In Net Radiometer; Surface Fluxes are Incorrect |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | The net radiometer top dome was punctured by migrating birds. This alone does not cause incorrect net radiation measurements. However, when it subsequently rains, water collects in the bottom dome, causing the net radiation measurements to be too large during the day and too small at night. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/30/2003 | 1500 | 09/02/2003 | 2100 |
Subject: | SGP/EBBR/E18 - Water in Net Radiometer; Surface Fluxes are Incorrect |
DataStreams: | sgp30ebbrE18.b1, sgp5ebbrE18.b1, sgp15ebbrE18.b1 |
Description: | The net radiometer top dome was punctured by migrating birds. This alone does not cause incorrect net radiation measurements. However, when it subsequently rains, water collects in the bottom dome, causing the net radiation measurements to be too large during the day and too small at night. |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/17/2003 | 1830 | 09/17/2003 | 2000 |
09/17/2003 | 2230 | 09/18/2003 | 0100 |
Subject: | SGP/EBBR/E19 - Tair Right Offscale |
DataStreams: | sgp5ebbrE19.b1, sgp15ebbrE19.b1, sgp30ebbrE19.b1 |
Description: | The right air temperature data were going offscale. This affected the calculation of sensible and latent heat fluxes. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/15/2003 | 1030 | 11/17/2003 | 0630 |
Subject: | SGP/EBBR/E27 - Right Air Temperature Spiked Offscale |
DataStreams: | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
Description: | The right air temperature was going offscale at times. This problem seemed to end without Site Ops. intervention. |
Measurements: | sgp5ebbrE27.b1:
sgp15ebbrE27.b1:
sgp30ebbrE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/29/2003 | 1700 | 10/29/2003 | 1930 |
Subject: | SGP/EBBR/E27 - Missing data during system update |
DataStreams: | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
Description: | Some soil parameters were slightly incorrect since EBBR.E27 was installed in May 2003. While the coefficients were being updated, data collection was suspended resulting in missing data for the specified period. No significant effects (1% or less) on the surface heat flux data would have occurred from using the slightly incorrect coefficients. Therefore, flagging of the surface flux data from the time of installation of E27 is unwarranted. |
Measurements: | sgp5ebbrE27.b1:
sgp15ebbrE27.b1:
sgp30ebbrE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/15/2003 | 2100 | 10/29/2003 | 2300 |
Subject: | SGP/EBBR/E20 - Fuse Blown in AEM |
DataStreams: | sgp30ebbrE20.b1, sgp15ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | When the EBBR system restarted on 15 Oct. 2003, the AEM malfunctioned. Surface fluxes data during this period are incorrect. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/27/2003 | 1800 | 12/10/2003 | 2230 |
Subject: | SGP/EBBR/E20 - Reprocess: Soil Temperature Probe #1 Measurements Spiked |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1 |
Description: | The soil temperature #1 measurement spiked quite high at times. This affected a number of measurements that are dependent on the soil temperature measurements. |
Measurements: | sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/19/2003 | 1730 | 10/19/2003 | 1930 |
10/20/2003 | 1330 | 10/20/2003 | 1630 |
10/21/2003 | 1630 | 10/21/2003 | 1830 |
10/22/2003 | 0000 | 10/22/2003 | 1830 |
10/23/2003 | 0000 | 10/23/2003 | 1630 |
Subject: | SGP/EBBR/E7 - Reference Temperature Failure |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | The reference T sensor was not working, affecting other temperature sensors. Sensible and latent heat fluxes were incorrect when this occurred. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/11/2003 | 2230 | 10/12/2003 | 2330 |
Subject: | SGP/EBBR/E15 - Reference Temperature Failed |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | The reference temperature failed, making air temperatures, e, h, and Bowen ratio incorrect. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/23/2003 | 1600 | 09/23/2003 | 1600 |
Subject: | SGP/EBBR/E9 - Measurements All Zero |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Description: | Data values were all zero after restarting the CR10 datalogger. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/14/2003 | 0130 | 10/14/2003 | 1800 |
Subject: | SGP/EBBR/E12 - Water in Net Radiometer, Surface Fluxes Incorrect |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | The net radiometer top dome was punctured by migrating birds. This alone does not cause incorrect net radiation measurements. However, when it subsequently rains, water collects in the bottom dome, causing the net radiation measurements to be too large during the day and too small at night. |
Measurements: | sgp30ebbrE12.b1:
sgp15ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/29/2003 | 1800 | 10/02/2003 | 1845 |
Subject: | SGP/EBBR/E13 - Water in Net Radiometer, Surface Fluxes Incorrect |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | The net radiometer top dome was punctured by migrating birds. This alone does not cause incorrect net radiation measurements. However, when it subsequently rains, water collects in the bottom dome, causing the net radiation measurements to be too large during the day and too small at night. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/29/2003 | 1630 | 09/30/2003 | 1600 |
Subject: | SGP/EBBR/E15 - Water In Net Radiometer, Surface Fluxes Incorrect |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | The net radiometer top dome was punctured by migrating birds. This alone does not cause incorrect net radiation measurements. However, when it subsequently rains, water collects in the bottom dome, causing the net radiation measurements to be too large during the day and too small at night. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/30/2003 | 0000 | 10/01/2003 | 2100 |
Subject: | SGP/EBBR/E20 - Water in Net Radiometer, Surface Fluxes Incorrect |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | The net radiometer top dome was punctured by migrating birds. This alone does not cause incorrect net radiation measurements. However, when it subsequently rains, water collects in the bottom dome, causing the net radiation measurements to be too large during the day and too small at night. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/29/2003 | 1800 | 10/01/2003 | 1630 |
Subject: | SGP/EBBR/E22 - Water in Net Radiometer, Surface Fluxes Incorrect |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | The net radiometer top dome was punctured by migrating birds. This alone does not cause incorrect net radiation measurements. However, when it subsequently rains, water collects in the bottom dome, causing the net radiation measurements to be too large during the day and too small at night. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/11/2003 | 2230 | 10/12/2003 | 2359 |
Subject: | SGP/EBBR/E15 - Soil Temperature #5 Incorrect, Surface Fluxes Incorrect |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1 |
Description: | Soil Temperature #5 was incorrect. This affected many measurements, including sensible and latent heat fluxes. |
Measurements: | sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/30/2003 | 1545 | 11/13/2003 | 1545 |
Subject: | SGP/EBBR/E26 - AEM Home Signals Incorrect, Surface Fluxes Incorrect |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | During a preventative maintenance visit to perform the 6-month calibrations, the home15 and home30 outputs began reading the same as one another. A fuse replacement and lubrication of the AEM sprockets on 11/13 corrected the problem. During the interval specified, the home15, home30, and sensible and latent heat fluxes are incorrect. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/13/2003 | 0000 | 10/14/2003 | 1830 |
Subject: | SGP/EBBR/E15 - Soil Temperature #5 Incorrect, Surface Fluxes Incorrect |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1 |
Description: | Soil Temperature #5 was incorrect. This affected many measurements, including sensible and latent heat fluxes. |
Measurements: | sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/05/2003 | 1400 | 11/11/2003 | 1800 |
Subject: | SGP/EBBR/E15 - AEM Hung Between Home Positions, Surface Fluxes Incorrect |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | The AEM was hung inbetween home positions. Therefore the sensible and latent heat fluxes are also incorrect. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/03/2003 | 2000 | 11/13/2003 | 1615 |
Subject: | SGP/EBBR/E19 - Battery Voltage Too Low, Most Measurements Incorrect |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | Battery voltage dropped below levels required to sustain the production of valid data. The AC charger unit had failed. This prevented charging of the battery during the night and prevented the solar panel regulator from properly charging the battery during the day. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/14/2003 | 2115 | 11/26/2003 | 1600 |
Subject: | SGP/EBBR/E19 - AEM Stuck In One Position, Surface Fluxes Incorrect |
DataStreams: | sgp30ebbrE19.b1, sgp5ebbrE19.b1, sgp15ebbrE19.b1 |
Description: | The AEM home 15 and 30 outputs were both 40 mv because the AEM was not exchanging. Sensible and Latent Heat Fluxes are therefore incorrect. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/06/2003 | 1200 | 10/11/2003 | 0730 |
Subject: | SGP/EBBR/E27 - Right Air Temperature Spiked Offscale |
DataStreams: | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
Description: | The right air temperature was going offscale at times. |
Measurements: | sgp5ebbrE27.b1:
sgp15ebbrE27.b1:
sgp30ebbrE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/10/2003 | 1505 | 10/10/2003 | 1755 |
Subject: | SGP/EBBR/E19 - Data Missing During Recalibration Changeout |
DataStreams: | sgp15ebbrE19.b1, sgp30baebbrE19.c1, sgp30ebbrE19.b1, sgp5ebbrE19.b1, sgpebbrE19.00, sgpebbrE19.01 |
Description: | Data was missing during the recalibration changeout. |
Measurements: | sgpebbrE19.01:
sgpebbrE19.00:
sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
sgp30baebbrE19.c1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/13/2003 | 0000 | 11/26/2003 | 2300 |
Subject: | SGP/EBBR/E12 - Pressure Sensor Measurement Scales With Battery Voltage |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | The barometer output was scaling with supply voltage (i.e. it followed the charging of the battery by the solar panel). NOTE: The barometer was replaced later than the end date of this DQR. Why the problem ended on 26 Nov 2003 is unclear; there was no site ops intervention on that date. The problem did not affect the sensible and latent heat flux calculations beyond the expected system error. |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/13/2003 | 1830 | 10/13/2003 | 1900 |
Subject: | SGP/EBBR/E13 - 6 Month Checks |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/14/2003 | 1700 | 10/14/2003 | 1730 |
Subject: | SGP/EBBR/E15 - 6 Month Checks |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/28/2003 | 2225 | 10/28/2003 | 2315 |
Subject: | SGP/EBBR/E18 - 6 Month Checks |
DataStreams: | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/10/2003 | 1530 | 10/10/2003 | 1600 |
Subject: | SGP/EBBR/E19 - 6 Month Checks |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/16/2003 | 1500 | 10/16/2003 | 1530 |
Subject: | SGP/EBBR/E19 - 6 Month Checks |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/29/2003 | 2200 | 10/29/2003 | 2245 |
Subject: | SGP/EBBR/E20 - 6 Month Checks |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/10/2003 | 1800 | 10/10/2003 | 1800 |
Subject: | SGP/EBBR/E19 - Data All Zeros |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | The time period after restart of the datalogger produced all zero data. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/15/2003 | 1900 | 10/15/2003 | 1930 |
Subject: | SGP/EBBR/E22 - 6 Month Checks |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/30/2003 | 1535 | 10/30/2003 | 1615 |
Subject: | SGP/EBBR/E26 - 6 Month Checks |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/29/2003 | 1830 | 10/29/2003 | 1945 |
Subject: | SGP/EBBR/E27 - 6 Month Checks |
DataStreams: | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp5ebbrE27.b1:
sgp15ebbrE27.b1:
sgp30ebbrE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/09/2003 | 1700 | 10/09/2003 | 1730 |
Subject: | SGP/EBBR/E2 - 6 Month Checks |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp30ebbrE2.b1:
sgp15ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/04/2003 | 2200 | 11/04/2003 | 2230 |
Subject: | SGP/EBBR/E7 - 6 Month Checks |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/07/2003 | 1700 | 10/07/2003 | 1735 |
Subject: | SGP/EBBR/E8 - 6 Month Checks |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/04/2003 | 1730 | 11/04/2003 | 1800 |
Subject: | SGP/EBBR/E9 - 6 Month Checks |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/28/2003 | 1750 | 10/28/2003 | 1845 |
Subject: | SGP/EBBR/E12 - 6 Month Checks |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/03/2003 | 0530 | 12/23/2003 | 2140 |
Subject: | SGP/EBBR/E20 - Wind Direction Incorrect |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | Wind direction is incorrect, reading at or about 0 or 15 to 25 degrees. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/12/2003 | 1700 | 10/12/2003 | 2359 |
Subject: | SGP/EBBR/E27 - Reprocess: Soil Heat Flow #5 Offscale, Surface Fluxes Incorrect |
DataStreams: | sgp15ebbrE27.b1, sgp30ebbrE27.b1 |
Description: | Soil Heat Flow #5 went offscale at times. Latent and Sensible Heat Fluxes were incorrect when this occurred. |
Measurements: | sgp15ebbrE27.b1:
sgp30ebbrE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/13/2003 | 0000 | 09/29/2004 | 1845 |
Subject: | SGP/EBBR/E26 - Spark Gap Bad, Pressure Incorrect |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | In a trip to E26, I found that a bad spark gap through which the barometer was powered was causing pressure measurements to be incorrect. The pressure readings oscillated with the voltage of the EBBR battery. During daytime the voltage and thus the pressure would increase and vice versa during dark hours. Apparently only the positive side of the power supply was properly connected through the spark gap and thus the voltage received by the pressure sensor was lower than required to power it properly. Under this condition the perssure measurement varies with voltage level. The pressure measurements therefore had a diurnal signature and usually were incorrect. The decreased pressure during most times would cause latent heat flux to be slightly overestimated and sensible heat flux to be slightly underestimated. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/28/2003 | 1430 | 01/21/2004 | 2130 |
Subject: | SGP/EBBR/E20 - Pressure Sensor Data Incorrect |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | The E20 pressure sensor suffered from spikes in pressure, which were related to increases in voltage during the day as the solar panels charged the battery. These spikes were not of a magnitude that would affect the 30 minute surface fluxes. Replacement of the pressure sensor solved the problem. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/07/2003 | 1500 | 10/12/2003 | 2355 |
Subject: | SGP/EBBR/E7 - Pressure Sensor Data Incorrect |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | The E7 pressure sensor suffered from spikes in pressure, which were related to increases in voltage during the day as the solar panels charged the battery. These spikes were not of a magnitude that would affect the 30 minute surface fluxes. Replacement of the pressure sensor solved the problem. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/03/2003 | 1500 | 10/12/2003 | 2355 |
Subject: | SGP/EBBR/E12 - Pressure Sensor Measurement Scales With Battery Voltage |
DataStreams: | sgp30ebbrE12.b1, sgp5ebbrE12.b1, sgp15ebbrE12.b1 |
Description: | The barometer output scaled with supply voltage (it followed the charging of the battery by the solar panel). The problem did not affect the sensible and latent heat flux calculations beyond the expected system error. |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/09/2003 | 1800 | 10/09/2003 | 1800 |
Subject: | SGP/EBBR/E2 - All Data Zero After Logger Restart |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | All data was zero for the 5, 15, and 30 minute datastreams. |
Measurements: | sgp15ebbrE2.b1:
sgp30ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/14/2003 | 1500 | 12/14/2003 | 2200 |
Subject: | SGP/EBBR/E2 - Net Radiometer Iced Over During Storm |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | The net radiometer was covered with ice from a storm. Net radiation values are incorrect, and are sometimes even negative during the daytime. Sensible and latent heat fluxes are therefore incorrect. |
Measurements: | sgp30ebbrE2.b1:
sgp15ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/13/2003 | 1400 | 12/14/2003 | 2200 |
Subject: | SGP/EBBR/E8 - Net Radiometer Iced Over During Storm |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The net radiometer was covered with ice from a storm. Net radiation values are incorrect, and are sometimes even negative during the daytime. Sensible and latent heat fluxes are therefore incorrect. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/14/2003 | 0100 | 12/14/2003 | 1930 |
Subject: | SGP/EBBR/E9 - Net Radiometer Iced Over During Storm |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Description: | The net radiometer was covered with ice from a storm. Net radiation values are incorrect, and are sometimes even negative during the daytime. Sensible and latent heat fluxes are therefore incorrect. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/09/2003 | 2020 | 12/10/2003 | 0645 |
Subject: | SGP/EBBR/E9 - Wind Speed Sensor Frozen By Ice Storm |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Description: | The wind speed sensor was frozen stopped by ice from a storm. The wind speed and residual wind speed measurements are incorrect. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/09/2003 | 2235 | 12/10/2003 | 1705 |
Subject: | SGP/EBBR/E12 - Wind Speed Sensor Frozen By Ice Storm |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | The wind speed sensor was frozen stopped by ice from a storm. The wind speed and residual wind speed measurements are incorrect. |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/09/2003 | 2145 | 12/10/2003 | 0115 |
Subject: | SGP/EBBR/E13 - Wind Speed Sensor Frozen By Ice Storm |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | The wind speed sensor was frozen stopped by ice from a storm. The wind speed and residual wind speed measurements are incorrect. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/14/2003 | 1730 | 12/14/2003 | 1830 |
Subject: | SGP/EBBR/E13 - Net Radiometer Iced Over During Storm |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | The net radiometer was covered with ice from a storm. Net radiation values are incorrect, and are sometimes even negative during the daytime. Sensible and latent heat fluxes are therefore incorrect. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/13/2003 | 1500 | 12/15/2003 | 0000 |
Subject: | SGP/EBBR/E15 - Net Radiometer Iced Over During Storm |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | The net radiometer was covered with ice from a storm. Net radiation values are incorrect, and are sometimes even negative during the daytime. Sensible and latent heat fluxes are therefore incorrect. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/12/2003 | 1955 | 12/13/2003 | 1940 |
Subject: | SGP/EBBR/E15 - Wind Speed Sensor Frozen By Ice Storm |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | The wind speed sensor was frozen stopped by ice from a storm. The wind speed and residual wind speed measurements are incorrect. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/13/2003 | 1100 | 12/13/2003 | 1230 |
Subject: | SGP/EBBR/E20 - Wind Speed Sensor Frozen By Ice Storm |
DataStreams: | sgp30ebbrE20.b1, sgp5ebbrE20.b1, sgp15ebbrE20.b1 |
Description: | The wind speed sensor was frozen stopped by ice from a storm. The wind speed and residual wind speed measurements are incorrect. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/12/2003 | 1810 | 12/13/2003 | 2100 |
Subject: | SGP/EBBR/E22 - Wind Speed Sensor Frozen By Ice Storm |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | The wind speed sensor was frozen stopped by ice from a storm. The wind speed and residual wind speed measurements are incorrect. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/13/2003 | 2345 | 12/14/2003 | 1520 |
Subject: | SGP/EBBR/E22 - Wind Direction Sensor Frozen By Ice Storm |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | The wind direction sensor was frozen stopped by ice from a storm. The wind direction, sigma_wd, and residual wind speed measurements are incorrect. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/12/2003 | 1935 | 12/14/2003 | 1615 |
Subject: | SGP/EBBR/E26 - Wind Speed Sensor Frozen By Ice Storm |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | The wind speed sensor was frozen stopped by ice from a storm. The wind speed and residual wind speed measurements are incorrect. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/12/2003 | 2140 | 12/14/2003 | 1530 |
Subject: | SGP/EBBR/E26 - Wind Direction Sensor Frozen By Ice Storm |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | The wind direction sensor was frozen stopped by ice from a storm. The wind direction, sigma_wd, and residual wind speed measurements are incorrect. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/06/2003 | 0900 | 12/06/2003 | 0900 |
Subject: | SGP/EBBR/E27 - Soil Temperature #3 Spike in Data |
DataStreams: | sgp15ebbrE27.b1, sgp30ebbrE27.b1 |
Description: | The soil temperature #3 spiked to zero and was therefore incorrect. |
Measurements: | sgp15ebbrE27.b1:
sgp30ebbrE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/09/2003 | 2220 | 12/10/2003 | 1200 |
Subject: | SGP/EBBR/E7 - Wind Speed Sensor Frozen By Ice Storm |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | The wind speed sensor was frozen stopped by ice. The wind speed and residual wind speed measurements are incorrect. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/12/2003 | 1825 | 12/12/2003 | 2000 |
Subject: | SGP/EBBR/E19 - Wind Speed Sensor Frozen Stopped by Ice Storm |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | An ice storm froze the wind speed sensor to a stopped condition. It is not known how long this condition existed, as data is missing for more than 10 days after the end of the stated period. The wind speed and residual wind speed measurements are incorrect. This sensor failure does not affect the sensible and latent heat fluxes. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/25/2003 | 2150 | 01/20/2004 | 2030 |
Subject: | SGP/EBBR/E15 - Wind Direction About 130 Degrees Low |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | The EBBR wind direction sensor developed an approximately -130 degree offset. The pin holding the vane to the shaft had fallen out. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/26/2004 | 1915 | 04/13/2004 | 2045 |
Subject: | SGP/EBBR/E18 - AEM Sitting At Home 15 Position |
DataStreams: | sgp30baebbrE18.c1, sgp30ebbrE18.b1, sgp5ebbrE18.b1, sgp15ebbrE18.b1 |
Description: | After replacement of the AEM on 1/26/04, it exchanged briefly, then sat at the home 15 position until another replacement was available. |
Measurements: | sgp15ebbrE18.b1:
sgp30baebbrE18.c1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/13/2003 | 0000 | 01/28/2004 | 2030 |
Subject: | SGP/EBBR/E7 - Pressure Sensor Data Incorret |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | The E7 pressure sensor suffered from spikes in pressure, which were related to increases in voltage during the day as the solar panels charged the battery. These spikes were not of a magnitude that would affect the 30 minute surface fluxes. Replacement of the pressure sensor solved the problem. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/08/2004 | 1600 | 01/22/2004 | 1530 |
Subject: | SGP/EBBR/E19 - Low Battery Voltage Caused Incorrect Data |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | Battery voltage less than 10.5 V caused much data to be questionable or incorrect. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/26/2004 | 0000 | 03/02/2004 | 1705 |
Subject: | SGP/EBBR/E15 - Net Radiometer not level |
DataStreams: | sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | The EBBR net radiometer at E15 was suspected to be off of level because the daily peak net radiation was recorded later than it was recorded by the SIRS. Site Ops found the EBBR net radiometer at E15 was significantly off of level (1/2 bubble tipped to the north). Adjustment of the level improved the comparison of peak time with the co-located SIRS. There was no significant effect on the sensible and latent heat fluxes. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/13/2003 | 0000 | 03/17/2004 | 1835 |
Subject: | SGP/EBBR/E22 - Wind Speed Sensor Bearings Worn |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | Wind speed was too low and/or often at or near zero due to worn bearings in the wind speed sensor. Sensor was replaced 3/17/04. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/02/2004 | 0945 | 02/02/2004 | 1820 |
02/09/2004 | 1215 | 02/09/2004 | 1430 |
Subject: | SGP/EBBR/E12 - Wind Speed Frozen Stopped By Ice |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | An ice storm froze the wind speed sensor to a stopped condition. The wind speed and residual wind speed measurements are incorrect. This sensor failure does not affect the sensible and latent heat fluxes. |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/02/2004 | 0810 | 02/02/2004 | 1930 |
02/09/2004 | 1235 | 02/09/2004 | 1405 |
Subject: | SGP/EBBR/E12 - Wind Direction Frozen Stopped By Ice |
DataStreams: | sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | The wind direction sensor was frozen stopped by ice from a storm. The wind direction, sigma_wd, and residual wind speed measurements are incorrect. |
Measurements: | sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/09/2004 | 1125 | 02/09/2004 | 1500 |
Subject: | SGP/EBBR/E13 - Wind Speed Frozen Stopped By Ice |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | An ice storm froze the wind speed sensor to a stopped condition. The wind speed and residual wind speed measurements are incorrect. This sensor failure does not affect the sensible and latent heat fluxes. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/02/2004 | 0040 | 02/03/2004 | 1810 |
02/09/2004 | 1125 | 02/09/2004 | 1500 |
Subject: | SGP/EBBR/E13 - Wind Direction Frozen Stopped By Ice |
DataStreams: | sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | The wind direction sensor was frozen stopped by ice from a storm. The wind direction, sigma_wd, and residual wind speed measurements are incorrect. |
Measurements: | sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/01/2004 | 1810 | 02/02/2004 | 2040 |
02/04/2004 | 2240 | 02/05/2004 | 1840 |
02/09/2004 | 1020 | 02/09/2004 | 1430 |
Subject: | SGP/EBBR/E15 - Wind Speed Frozen Stopped By Ice |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | An ice storm froze the wind speed sensor to a stopped condition. The wind speed and residual wind speed measurements are incorrect. This sensor failure does not affect the sensible and latent heat fluxes. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/28/2004 | 0140 | 01/28/2004 | 0305 |
02/05/2004 | 0145 | 02/05/2004 | 0520 |
Subject: | SGP/EBBR/E18 - Wind Speed Frozen Stopped By Ice |
DataStreams: | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
Description: | An ice storm froze the wind speed sensor to a stopped condition. The wind speed and residual wind speed measurements are incorrect. This sensor failure does not affect the sensible and latent heat fluxes. |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/28/2004 | 0140 | 01/28/2004 | 0305 |
Subject: | SGP/EBBR/E18 - Wind Direction Frozen Stopped By Ice |
DataStreams: | sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
Description: | The wind direction sensor was frozen stopped by ice from a storm. The wind direction, sigma_wd, and residual wind speed measurements are incorrect. |
Measurements: | sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/02/2004 | 0505 | 02/02/2004 | 1845 |
Subject: | SGP/EBBR/E20 - Wind Speed Frozen Stopped By Ice |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | An ice storm froze the wind speed sensor to a stopped condition. The wind speed and residual wind speed measurements are incorrect. This sensor failure does not affect the sensible and latent heat fluxes. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/02/2004 | 0535 | 02/02/2004 | 1755 |
Subject: | SGP/EBBR/E20 - Wind Direction Frozen Stopped By Ice |
DataStreams: | sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | The wind direction sensor was frozen stopped by ice from a storm. The wind direction, sigma_wd, and residual wind speed measurements are incorrect. |
Measurements: | sgp5ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/24/2004 | 2345 | 01/25/2004 | 0135 |
02/01/2004 | 1920 | 02/02/2004 | 1740 |
02/03/2004 | 0350 | 02/03/2004 | 1440 |
02/05/2004 | 0925 | 02/05/2004 | 1615 |
Subject: | SGP/EBBR/E22 - Wind Speed Frozen Stopped By Ice |
DataStreams: | sgp30ebbrE22.b1, sgp5ebbrE22.b1, sgp15ebbrE22.b1 |
Description: | An ice storm froze the wind speed sensor to a stopped condition. The wind speed and residual wind speed measurements are incorrect. This sensor failure does not affect the sensible and latent heat fluxes. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/05/2004 | 1120 | 02/05/2004 | 1500 |
Subject: | SGP/EBBR/E22 - Wind Direction Frozen Stopped By Ice |
DataStreams: | sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | The wind direction sensor was frozen stopped by ice from a storm. The wind direction, sigma_wd, and residual wind speed measurements are incorrect. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/02/2004 | 0705 | 02/02/2004 | 1800 |
02/05/2004 | 0205 | 02/05/2004 | 0600 |
Subject: | SGP/EBBR/E27 - Wind Speed Frozen Stopped By Ice |
DataStreams: | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
Description: | An ice storm froze the wind speed sensor to a stopped condition. The wind speed and residual wind speed measurements are incorrect. This sensor failure does not affect the sensible and latent heat fluxes. |
Measurements: | sgp5ebbrE27.b1:
sgp15ebbrE27.b1:
sgp30ebbrE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/01/2004 | 2120 | 02/02/2004 | 1910 |
Subject: | SGP/EBBR/E19 - Wind Speed Frozen Stopped By Ice |
DataStreams: | sgp30ebbrE19.b1, sgp5ebbrE19.b1, sgp15ebbrE19.b1 |
Description: | An ice storm froze the wind speed sensor to a stopped condition. The wind speed and residual wind speed measurements are incorrect. This sensor failure does not affect the sensible and latent heat fluxes. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/02/2004 | 0200 | 02/02/2004 | 1620 |
Subject: | SGP/EBBR/E26 - Wind Speed Frozen Stopped By Ice |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | An ice storm froze the wind speed sensor to a stopped condition. The wind speed and residual wind speed measurements are incorrect. This sensor failure does not affect the sensible and latent heat fluxes. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/02/2004 | 0410 | 02/02/2004 | 1530 |
Subject: | SGP/EBBR/E26 - Wind Direction Frozen Stopped By Ice |
DataStreams: | sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | The wind direction sensor was frozen stopped by ice from a storm. The wind direction, sigma_wd, and residual wind speed measurements are incorrect. |
Measurements: | sgp30ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/26/2004 | 0755 | 01/28/2004 | 1550 |
02/02/2004 | 0840 | 02/03/2004 | 1650 |
02/05/2004 | 2330 | 02/06/2004 | 1210 |
Subject: | SGP/EBBR/E7 - Wind Speed Frozen Stopped By Ice |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | The wind speed sensor was frozen stopped by ice from a storm. The wind speed and residual wind speed measurements are incorrect. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/19/2004 | 2120 | 02/20/2004 | 0305 |
Subject: | SGP/EBBR/E7 - Tref Problem Causes All Temps. to be Incorrect |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | The Tref sensor went crazy for this brief period, making all temperature measurements incorrect. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/25/2004 | 1345 | 01/28/2004 | 1640 |
02/01/2004 | 0105 | 02/03/2004 | 2100 |
Subject: | SGP/EBBR/E2 - Wind Speed Frozen Stopped By Ice |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | An ice storm froze the wind speed sensor to a stopped condition. The wind speed and residual wind speed measurements are incorrect. This sensor failure does not affect the sensible and latent heat fluxes. |
Measurements: | sgp15ebbrE2.b1:
sgp30ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/31/2004 | 1910 | 02/04/2004 | 1750 |
02/05/2004 | 1020 | 02/06/2004 | 0820 |
Subject: | SGP/EBBR/E8 - Wind Speed Frozen Stopped By Ice |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | An ice storm froze the wind speed sensor to a stopped condition. The wind speed and residual wind speed measurements are incorrect. This sensor failure does not affect the sensible and latent heat fluxes. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/25/2004 | 2300 | 01/26/2004 | 2135 |
01/27/2004 | 0425 | 01/27/2004 | 0630 |
01/27/2004 | 1130 | 01/27/2004 | 1740 |
Subject: | SGP/EBBR/E8 - Wind Direction Frozen Stopped By Ice |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The wind direction sensor was frozen stopped by ice from a storm. The wind direction, sigma_wd, and residual wind speed measurements are incorrect. |
Measurements: | sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/27/2004 | 1745 | 01/27/2004 | 1800 |
01/27/2004 | 1935 | 01/28/2004 | 0140 |
01/28/2004 | 0330 | 01/28/2004 | 0355 |
01/31/2004 | 1800 | 02/12/2004 | 1640 |
Subject: | SGP/EBBR/E9 - Wind Speed Frozen Stopped By Ice |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Description: | An ice storm froze the wind speed sensor to a stopped condition. The bearings in the sensor appear to be worn, contributing to the problem. The wind speed and residual wind speed measurements are incorrect. This sensor failure does not affect the sensible and latent heat fluxes. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/02/2004 | 0000 | 02/03/2004 | 1740 |
02/05/2004 | 1120 | 02/05/2004 | 1405 |
02/09/2004 | 0945 | 02/09/2004 | 1430 |
Subject: | SGP/EBBR/E9 - Wind Direction Frozen Stopped By Ice |
DataStreams: | sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Description: | The wind direction sensor was frozen stopped by ice from a storm. The wind direction, sigma_wd, and residual wind speed measurements are incorrect. |
Measurements: | sgp30ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/25/2004 | 0600 | 05/12/2004 | 1930 |
Subject: | SGP/EBBR/E27 - Soil Moisture #5 Insensitive |
DataStreams: | sgp30ebbrE27.b1 |
Description: | Soil moisture sensor #5 was insensitive and put out a value of around 30%. Sensible and latent heat fluxes were not significantly affected by this small error. |
Measurements: | sgp30ebbrE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/26/2004 | 0710 | 02/26/2004 | 1410 |
Subject: | SGP/EBBR/E8 - Wind Speed Sensor Frozen Stopped By Icing |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | An ice storm froze the wind speed sensor to a stopped condition. The wind speed and residual wind speed measurements are incorrect. This sensor failure does not affect the sensible and latent heat fluxes. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/05/2004 | 0000 | 04/06/2004 | 1755 |
Subject: | SGP/EBBR/E8 - AEM Replacement |
DataStreams: | sgp15ebbrE8.b1, sgp30baebbrE8.c1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The AEM at E8 kept hanging in between home positions. This DQR covers only the period during which the AEM was being replaced. DQRs generally are not issued for AEMs that are not working as qc flags indicate this clearly. |
Measurements: | sgp15ebbrE8.b1:
sgp30baebbrE8.c1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/05/2004 | 1700 | 04/15/2004 | 1530 |
Subject: | SGP/EBBR/E19 - Reprocess: Incorrect Net Rad Calibration Coefficient |
DataStreams: | sgp30baebbrE19.c1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | The net radiometer coefficient installed on 2/5/04 for nighttime (negative voltage calibration) was incorrect. Negative net radiations during the subject period can be multiplied by 0.9419 to get corrected net radiation measurements. The sensible (H) and latent heat (LE) fluxes (which are very small at night anyway) can be recalculated using the following equations (where B is Bowen ratio, Rn is corrected net radiation, and G is soil heat flow): LE = -(Rn + G)/(B + 1) H = -(Rn + G + LE) |
Measurements: | sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
sgp30baebbrE19.c1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/11/2004 | 1830 | 06/01/2004 | 1800 |
Subject: | SGP/EBBR/E2 - Soil Heat Flow #1 Spikes Down |
DataStreams: | sgp30ebbrE2.b1 |
Description: | Soil heat flow #1 spiked down at times. Since tightening of connections this is occurring infrequently. In most cases, sensible and latent heat fluxes are not affected. |
Measurements: | sgp30ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/19/2004 | 0500 | 06/13/2004 | 1700 |
Subject: | SGP/EBBR/E18 - Pressure Low |
DataStreams: | sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
Description: | Pressure readings suddenly dropped to unrealistic levels. Sensible Heat Flux was underestimated and Latent Heat Flux was overestimated because of this problem. |
Measurements: | sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/23/2004 | 2130 | 04/27/2004 | 1630 |
Subject: | SGP/EBBR/E15 - Wind Speed Incorrect |
DataStreams: | sgp15ebbrE15.b1, sgp30baebbrE15.c1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | The cups were broken off of the wind speed sensor by hail on 4/23/04. The cups were replaced on 4/27/04. Wind speed measurements were near zero and incorrect during the indicated period. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp30baebbrE15.c1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/01/2004 | 0000 | 03/31/2004 | 2355 |
Subject: | SGP/EBBR/E20 - Left Air Temperature Incorrect |
DataStreams: | sgp15ebbrE20.b1, sgp30baebbrE20.c1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | The left air temperature, and therefore top and bottom air temperatures were offscale or incorrect on occasion. Bowen ratio, sensible heat flux, and latent heat flux were also incorrect during those times. Inspection of the data is required to detect the incorrect periods. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30baebbrE20.c1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/01/2004 | 0000 | 03/31/2004 | 2355 |
Subject: | SGP/EBBR/E27 - Right Air Temperature Incorrect |
DataStreams: | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
Description: | The right air temperature, and therefore top and bottom air temperatures were offscale or incorrect on occasion. Bowen ratio, sensible heat flux, and latent heat flux were also incorrect during those times. Inspection of the data is required to detect the incorrect periods. |
Measurements: | sgp5ebbrE27.b1:
sgp15ebbrE27.b1:
sgp30ebbrE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/30/1997 | 0000 | 05/06/2003 | 2359 |
Subject: | SGP/EBBR/E19 - Metadata errors |
DataStreams: | sgp15ebbrE19.b1, sgp30baebbrE19.c1, sgp30ebbrE19.b1 |
Description: | The latitude, longitude and altitude of the El Reno (E19) site were incorrectly entered into the ARM database. The correct location of the EBBR.E19 instrument is: Lat: 35.557N Lon: 98.017W Alt: 421m |
Measurements: | sgp15ebbrE19.b1:
sgp30ebbrE19.b1:
sgp30baebbrE19.c1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/01/2001 | 0000 | 05/06/2003 | 2359 |
Subject: | SGP/EBBR/E19 - Metadata errors |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | The latitude, longitude and altitude of the El Reno (E19) site were incorrectly entered into the ARM database. The correct location of the EBBR.E19 instrument is: Lat: 35.557N Lon: 98.017W Alt: 421m |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/05/2004 | 0130 | 07/14/2004 | 1515 |
Subject: | SGP/EBBR/E13 - Automatic Exchange Mechanism Failure |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | The AEM was stuck in the home_15 position or reported near zero home signals during much of the indicated period. The Bowen ratio and latent and sensible heat fluxes were incorrect when the AEM was not operating properly; the periods of incorrect data can be determined from the home signal values. On 7/14/04 the AEM was replaced. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/29/2004 | 2215 | 06/10/2004 | 1510 |
Subject: | SGP/EBBR/E19 - AEM Hung Inbetween Home Positions |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | The AEM hung inbetween home positions, producing a home signal of zero. Sensible and latent heat fluxes during this period are incorrect. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/19/2004 | 2212 | 03/10/2004 | 2247 |
Subject: | SGP/EBBR/E12 - Incorrect Times |
DataStreams: | sgp15ebbrE12.b1, sgp30baebbrE12.c1, sgp30ebbrE12.b1, sgp5ebbrE12.b1, sgpebbrE12.00 |
Description: | The ntpd server (time adjustment) stopped working on Feb. 19th at 22:12:16 GMT, and the clock was corrected by -5600.177269 seconds on March 10th at 22:47 GMT. The recorded times for observations are off by amounts ranging from 0 seconds on Feb. 19th to 5600 seconds on March 10th. |
Measurements: | sgp30ebbrE12.b1:
sgp15ebbrE12.b1:
sgp30baebbrE12.c1:
sgpebbrE12.00:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/22/2004 | 0400 | 06/23/2004 | 1650 |
Subject: | SGP/EBBR/E22 - Net Radiation Incorrect |
DataStreams: | sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | Rain filled the bottom of the net radiometer with water, resulting in very high net radiations. Net radiation is incorrect during this period, also causing sensible and latent heat fluxes to be incorrect. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/10/2004 | 0600 | 07/21/2004 | 1725 |
Subject: | SGP/EBBR/E27 - Loose Tair Left Connections |
DataStreams: | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
Description: | Tair left dropped out at times. When this occurred, the sensible and latent heat fluxes are incorrect. Tightening of the electronic connections to the datalogging equipment appears to have solved the problem. |
Measurements: | sgp5ebbrE27.b1:
sgp15ebbrE27.b1:
sgp30ebbrE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/09/2004 | 1745 | 08/04/2004 | 1915 |
Subject: | SGP/EBBR/E20 - Wind Direction Stuck at 100 Degrees |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | The wind direction sensor was stuck at about 100 degrees. This did not affect the primary measurements (net radiation, soil heat flow, sensible heat flux, latent heat flux). |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/28/2004 | 1200 | 06/29/2004 | 1530 |
Subject: | SGP/EBBR/E9 - net radiometer top dome broken, water in bottom dome |
DataStreams: | sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Description: | The Net radiometer top dome was broken. Site Ops did not indicate that there was water in the bottom dome, but the data suggests it. Net radiation values during the period are incorrect. |
Measurements: | sgp30ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/01/2004 | 1330 | 07/07/2004 | 1900 |
Subject: | SGP/EBBR/E22 - Water In Net Radiometer |
DataStreams: | sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | Water entered the Net Radiometer through a cracked upper dome, causing the net radiation values to be too large. As a result, latent and sensible heat flux values are also incorrect. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/28/2004 | 1300 | 07/28/2004 | 1500 |
Subject: | SGP/EBBR/E18 - Soil Temperature and Moisture Probes in Set #1 Spiked Outputs |
DataStreams: | sgp30ebbrE18.b1 |
Description: | Soil moisture #1 and soil temperature #1 occasionally spiked at the same time. Little sensible and latent heat flux data were affected. |
Measurements: | sgp30ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/03/2004 | 0300 | 09/15/2004 | 2000 |
09/24/2004 | 0700 | 09/29/2004 | 1800 |
Subject: | SGP/EBBR/E20 - AEM Failure |
DataStreams: | sgp5ebbrE20.b1, sgp15ebbrE20.b1, sgp30ebbrE20.b1 |
Description: | The AEM hung inbetween positions; home signals were zero. The AEM fuse kept blowing, most likely because of worn rails or sleeves. This condition may reoccur. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/30/2004 | 1600 | 08/05/2004 | 1530 |
Subject: | SGP/EBBR/E26 - Soil Temperature #4 and Soil Heat Flow #4 Incorrect |
DataStreams: | sgp30ebbrE26.b1 |
Description: | The soil temperature and soil heat flow probes for set #4 was pulled partially out of the ground. This caused soil temperature #4 to be too high and soil heat flow #4 to be too low. The error is not so great as to cause a large error in the sensible and latent heat fluxes. |
Measurements: | sgp30ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/06/2004 | 1830 | 07/06/2004 | 2100 |
Subject: | SGP/EBBR/E18 - Soil Temperature #2 Incorrect |
DataStreams: | sgp30ebbrE18.b1 |
Description: | Soil Temperature #2 was incorrect during the indicated period. Average soil heat flow, sensible heat flux, and latent heat flux are incorrect, as a result. |
Measurements: | sgp30ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/10/2004 | 1925 | 08/24/2004 | 1820 |
Subject: | SGP/EBBR/E7 - Pressure Too High |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | When data collection resumed after a missing period, the barometer pressure had become very high, 102 kPa, which was not correct. This condition resulted from an incorrect calibration being used. Sensible heat flux was overestimated by about 4% and latent heat flux underestimated by about 4% because of the high pressure readings. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/11/2004 | 0230 | 08/11/2004 | 1824 |
Subject: | SGP/EBBR/E13 - Net Radiation Incorrect, Fluxes Incorrect |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | Rain entered the net radiometer through a crack in the upper dome and settled in the bottom dome. This caused too large net radiation measurements during the following daylight hours, until the domes were replaced and dried out the same day. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/31/2004 | 1300 | 09/09/2004 | 1500 |
Subject: | SGP/EBBR/E2 - Net Radiometer Dome Broken, Reducing Net Radiation Measurement, Fluxes |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | The upper dome of the net radiometer was broken. Normally this alone does not affect the net radiation measurement. However, in this case the breakage must have been more than just a bird claw hole or a crack, resulting in obstruction of the top sensing surface of the net radiometer and greatly reduced net radiation measurements. Sensible and latent heat fluxes were reduced as a result. |
Measurements: | sgp30ebbrE2.b1:
sgp15ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/30/2004 | 1200 | 09/07/2004 | 1730 |
Subject: | SGP/EBBR/E8 - Net Radiation Incorrect, Fluxes Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The net radiometer dome was broken and water entered the botom dome, causing net radiation measurements, and thus sensible and latent heat flux measurements, to be incorrect. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/08/2004 | 1200 | 09/15/2004 | 1930 |
Subject: | SGP/EBBR/E13 - AEM Failure, Fluxes Incorrect |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | The AEM failed. Sensible and latent heat fluxes were incorrect, as well as gradient measurements. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/08/2004 | 2130 | 09/15/2004 | 1930 |
Subject: | SGP/EBBR/E20 - Soil Heat Flow #3 and Soil Temperature #3 Incorrect |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1 |
Description: | |
Measurements: | sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/11/2001 | 1500 | 09/28/2004 | 1509 |
Subject: | SGP/EBBR/E15 - Soil Moisture Coefficients Incorrect |
DataStreams: | sgp30ebbrE15.b1 |
Description: | The soil Moisture coefficients used for E15 were incorrect from the time of installation of the system in program version 8. On 9 July 02 parameter 4 (an offset value) of program step 136 was changed from 9.5264 to the correct value 0.95264 (this error had been made by the vendor). The error masked the real problem (wrong soil type) and after 9 July 02 resulted in very low soil moisture measurements. The coefficients used were for coarse sand. On 28 Sep 04 I visited the E15 site and found the soil type a fine sandy clay or fine sandy loam. I changed the coefficients at 1509 GMT to those for a very fine sandy loam; this produced reasonable soil moistures for the soil type. The old coefficients in parameters 4 through 9 for step 136 of the program were: 0.95264, -6.7859, 36.224, -58.619, 25.073, 4.2278 The new coefficients are: 10.241, -27.943, 49.523, -12.648, -52.643, 38.054 |
Measurements: | sgp30ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/04/2004 | 1705 | 08/10/2004 | 1925 |
Subject: | SGP/EBBR/E7 - Data Missing |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | A power outage resulted in missing data for the indicated period. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/15/2004 | 1930 | 09/15/2004 | 2030 |
Subject: | SGP/EBBR/E20 - All Soil Measurements Incorrect |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1 |
Description: | All soil measurements were incorrect during the replacement of soil heat flow #3 and soil temperature #3. |
Measurements: | sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/02/2004 | 0630 | 10/21/2004 | 1500 |
Subject: | SGP/EBBR/E2 - Soil Heat Flow Sensor #2 Bad |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1 |
Description: | Soil Heat Flow sensor #2 spiked downwards at times for a while, but eventually became negative continuously. The problem was fixed when the sensor was replaced. |
Measurements: | sgp15ebbrE2.b1:
sgp30ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/21/2003 | 1530 | 10/12/2003 | 2355 |
Subject: | SGP/EBBR/E26 - Spark Gap Bad, Pressure Incorrect |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | In a trip to E26, I found that a bad spark gap through which the barometer was powered was causing pressure measurements to be incorrect. The pressure readings oscillated with the voltage of the EBBR battery. During daytime the voltage and thus the pressure would increase and vice versa during dark hours. Apparently only the positive side of the power supply was properly connected through the spark gap and thus the voltage received by the pressure sensor was lower than required to power it properly. Under this condition the perssure measurement varies with voltage level. The pressure measurements therefore had a diurnal signature and usually were incorrect. The decreased pressure during most times would cause latent heat flux to be slightly overestimated and sensible heat flux to be slightly underestimated. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/05/2004 | 1630 | 10/05/2004 | 1700 |
Subject: | SGP/EBBR/E9 - 6 Month Checks |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/12/2004 | 1630 | 10/12/2004 | 1700 |
Subject: | SGP/EBBR/E12 - 6 Month Checks |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/28/2004 | 1830 | 10/28/2004 | 1900 |
Subject: | SGP/EBBR/E13 - 6 Month Checks |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/26/2004 | 1630 | 10/26/2004 | 1700 |
Subject: | SGP/EBBR/E15 - 6 Month Checks |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/28/2004 | 2230 | 09/28/2004 | 2300 |
Subject: | SGP/EBBR/E18 - 6 Month Checks |
DataStreams: | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/28/2004 | 1430 | 10/28/2004 | 1500 |
Subject: | SGP/EBBR/E19 - 6 Month Checks |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/29/2004 | 1730 | 09/29/2004 | 1800 |
Subject: | SGP/EBBR/E20 - 6 Month Checks |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/27/2004 | 1700 | 10/27/2004 | 1730 |
Subject: | SGP/EBBR/E22 - 6 Month Checks |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/30/2004 | 1530 | 09/30/2004 | 1600 |
Subject: | SGP/EBBR/E26 - 6 Month Checks |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/29/2004 | 2000 | 09/29/2004 | 2030 |
Subject: | SGP/EBBR/E27 - 6 Month Checks |
DataStreams: | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp5ebbrE27.b1:
sgp15ebbrE27.b1:
sgp30ebbrE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/12/2004 | 1545 | 10/27/2004 | 0645 |
Subject: | SGP/EBBR/E12 - AEM home_15 Incorrect, Flux Measurements not affected |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | The home_15 measurement was incorrect during much of the stated period. The AEM was exchanging however, and thus the flux measurements were not affected. |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/02/2004 | 1330 | 03/02/2005 | 1930 |
Subject: | SGP/EBBR/E20 - Soil Moisture #4 and Soil Temperature #4 Incorrect |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1 |
Description: | Failures of a multiplexer channel and a J-panel channel caused soil temperature #4 and soil moisture #4 to spike and drift at times. Replacement of the failed equipment solved the problem. |
Measurements: | sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/27/2004 | 1930 | 11/05/2004 | 0100 |
11/10/2004 | 0600 | 11/24/2004 | 1500 |
Subject: | SGP/EBBR/E26 - Low Battery Voltage - Many Measurements Incorrect |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | During low battery voltage conditions (< 10.5 V especially) many measurements are incorrect. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/29/2004 | 1100 | 10/29/2004 | 1600 |
Subject: | SGP/EBBR/E26 - Most Measurements Incorrect |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | Most of the measurements became either too large or too small. The cause is not known. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/21/2004 | 1500 | 10/21/2004 | 1530 |
Subject: | SGP/EBBR/E2 - 6 Month Checks |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp30ebbrE2.b1:
sgp15ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/05/2004 | 2000 | 10/05/2004 | 2030 |
Subject: | SGP/EBBR/E7 - 6 Month Checks |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/19/2004 | 1730 | 10/19/2004 | 1800 |
Subject: | SGP/EBBR/E8 - 6 Month Checks |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/26/2004 | 2100 | 12/08/2004 | 2100 |
Subject: | SGP/EBBR/E20 - Soil Heat Flow #4 Incorrect |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1 |
Description: | Soil Heat Flow #4 was missing or offscale and therefore incorrect during the stated period. |
Measurements: | sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/01/2004 | 0830 | 01/31/2005 | 1400 |
Subject: | SGP/EBBR/E9 - Wind Speed Threshhold Low |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Description: | When the wind speed is very low, the anemometer cups stop turning. Wind speeds above the threshhold may be underestimated. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/05/2005 | 0510 | 01/08/2005 | 1745 |
01/11/2005 | 0445 | 01/12/2005 | 1310 |
Subject: | SGP/EBBR/E13 - Wind Speed Frozen Stopped By Ice Storm |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | An ice storm froze the wind speed sensor to a stopped condition. The wind speed and residual wind speed measurements are incorrect. This sensor failure does not affect the sensible and latent heat fluxes. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/04/2005 | 1010 | 01/05/2005 | 0200 |
Subject: | SGP/EBBR/E15 - Wind Speed Frozen Stopped By Ice Storm |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | An ice storm froze the wind speed sensor to a stopped condition. The wind speed and residual wind speed measurements are incorrect. This sensor failure does not affect the sensible and latent heat fluxes. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/05/2005 | 1810 | 01/06/2005 | 2045 |
Subject: | SGP/EBBR/E18 - Wind Speed Frozen Stopped By Ice Storm |
DataStreams: | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
Description: | An ice storm froze the wind speed sensor to a stopped condition. The wind speed and residual wind speed measurements are incorrect. This sensor failure does not affect the sensible and latent heat fluxes. |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/05/2005 | 0655 | 01/07/2005 | 1545 |
Subject: | SGP/EBBR/E19 - Wind Speed Frozen Stopped By Ice Storm |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | An ice storm froze the wind speed sensor to a stopped condition. The wind speed and residual wind speed measurements are incorrect. This sensor failure does not affect the sensible and latent heat fluxes. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/05/2005 | 1420 | 01/07/2005 | 1650 |
Subject: | SGP/EBBR/E20 - Wind Speed Frozen Stopped By Ice Storm |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | An ice storm froze the wind speed sensor to a stopped condition. The wind speed and residual wind speed measurements are incorrect. This sensor failure does not affect the sensible and latent heat fluxes. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/05/2005 | 0845 | 01/07/2005 | 1555 |
Subject: | SGP/EBBR/E22 - Wind Speed Frozen Stopped By Ice Storm |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | An ice storm froze the wind speed sensor to a stopped condition. The wind speed and residual wind speed measurements are incorrect. This sensor failure does not affect the sensible and latent heat fluxes. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/05/2005 | 0745 | 01/07/2005 | 1530 |
Subject: | SGP/EBBR/E26 - Wind Speed Frozen Stopped By Ice Storm |
DataStreams: | sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp15ebbrE26.b1 |
Description: | An ice storm froze the wind speed sensor to a stopped condition. The wind speed and residual wind speed measurements are incorrect. This sensor failure does not affect the sensible and latent heat fluxes. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/05/2005 | 1555 | 01/07/2005 | 1605 |
Subject: | SGP/EBBR/E27 - Wind Speed Frozen Stopped By Ice Storm |
DataStreams: | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
Description: | An ice storm froze the wind speed sensor to a stopped condition. The wind speed and residual wind speed measurements are incorrect. This sensor failure does not affect the sensible and latent heat fluxes. |
Measurements: | sgp5ebbrE27.b1:
sgp15ebbrE27.b1:
sgp30ebbrE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/03/2005 | 0530 | 01/09/2005 | 0830 |
Subject: | SGP/EBBR/E2 - Wind Speed Frozen Stopped By Ice Storm |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | An ice storm froze the wind speed sensor to a stopped condition. The wind speed and residual wind speed measurements are incorrect. This sensor failure does not affect the sensible and latent heat fluxes. |
Measurements: | sgp15ebbrE2.b1:
sgp30ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/04/2005 | 1450 | 01/05/2005 | 0100 |
Subject: | SGP/EBBR/E7 - Wind Speed Frozen Stopped By Ice Storm |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | An ice storm froze the wind speed sensor to a stopped condition. The wind speed and residual wind speed measurements are incorrect. This sensor failure does not affect the sensible and latent heat fluxes. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/03/2005 | 0810 | 01/04/2005 | 2100 |
Subject: | SGP/EBBR/E8 - Wind Speed Frozen Stopped By Ice Storm |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | An ice storm froze the wind speed sensor to a stopped condition. The wind speed and residual wind speed measurements are incorrect. This sensor failure does not affect the sensible and latent heat fluxes. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/04/2005 | 1505 | 01/05/2005 | 0800 |
Subject: | SGP/EBBR/E9 - Wind Speed Frozen Stopped By Ice Storm |
DataStreams: | sgp30ebbrE9.b1, sgp5ebbrE9.b1, sgp15ebbrE9.b1 |
Description: | An ice storm froze the wind speed sensor to a stopped condition. The wind speed and residual wind speed measurements are incorrect. This sensor failure does not affect the sensible and latent heat fluxes. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/05/2005 | 0735 | 01/08/2005 | 1700 |
Subject: | SGP/EBBR/E12 - Wind Speed Frozen Stopped By Ice Storm |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | An ice storm froze the wind speed sensor to a stopped condition. The wind speed and residual wind speed measurements are incorrect. This sensor failure does not affect the sensible and latent heat fluxes. |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/03/2005 | 0915 | 01/08/2005 | 2140 |
Subject: | SGP/EBBR/E2 - Wind Direction Frozen Stopped By Ice Storm |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | The wind direction sensor was frozen stopped by ice from a storm. The wind direction, sigma_wd, and residual wind speed measurements are incorrect. |
Measurements: | sgp30ebbrE2.b1:
sgp15ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/07/2005 | 0150 | 01/07/2005 | 1425 |
Subject: | SGP/EBBR/E20 - Wind Direction Frozen Stopped By Ice Storm |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | The wind direction sensor was frozen stopped by ice from a storm. The wind direction, sigma_wd, and residual wind speed measurements are incorrect. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/05/2005 | 0855 | 01/06/2005 | 1940 |
Subject: | SGP/EBBR/E22 - Wind Direction Frozen Stopped By Ice Storm |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | The wind direction sensor was frozen stopped by ice from a storm. The wind direction, sigma_wd, and residual wind speed measurements are incorrect. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/05/2005 | 1020 | 01/06/2005 | 1730 |
Subject: | SGP/EBBR/E26 - Wind Direction Frozen Stopped By Ice Storm |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | The wind direction sensor was frozen stopped by ice from a storm. The wind direction, sigma_wd, and residual wind speed measurements are incorrect. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/06/2005 | 1005 | 01/07/2005 | 1430 |
Subject: | SGP/EBBR/E27 - Wind Direction Frozen Stopped By Ice Storm |
DataStreams: | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
Description: | The wind direction sensor was frozen stopped by ice from a storm. The wind direction, sigma_wd, and residual wind speed measurements are incorrect. |
Measurements: | sgp5ebbrE27.b1:
sgp15ebbrE27.b1:
sgp30ebbrE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/03/2005 | 0850 | 01/04/2005 | 2100 |
Subject: | SGP/EBBR/E8 - Wind Direction Frozen Stopped By Ice Storm |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The wind direction sensor was frozen stopped by ice from a storm. The wind direction, sigma_wd, and residual wind speed measurements are incorrect. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/04/2005 | 1755 | 01/05/2005 | 0800 |
Subject: | SGP/EBBR/E9 - Wind Direction Frozen Stopped By Ice Storm |
DataStreams: | sgp30ebbrE9.b1, sgp5ebbrE9.b1, sgp15ebbrE9.b1 |
Description: | The wind direction sensor was frozen stopped by ice from a storm. The wind direction, sigma_wd, and residual wind speed measurements are incorrect. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/05/2005 | 0935 | 01/08/2005 | 1630 |
Subject: | SGP/EBBR/E12 - Wind Direction Frozen Stopped By Ice Storm |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | The wind direction sensor was frozen stopped by ice from a storm. The wind direction, sigma_wd, and residual wind speed measurements are incorrect. |
Measurements: | sgp30ebbrE12.b1:
sgp15ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/05/2005 | 0625 | 01/07/2005 | 1850 |
01/11/2005 | 1545 | 01/12/2005 | 0930 |
Subject: | SGP/EBBR/E13 - Wind Direction Frozen Stopped By Ice Storm |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | The wind direction sensor was frozen stopped by ice from a storm. The wind direction, sigma_wd, and residual wind speed measurements are incorrect. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/04/2005 | 1605 | 01/05/2005 | 0200 |
Subject: | SGP/EBBR/E15 - Wind Direction Frozen Stopped By Ice Storm |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | The wind direction sensor was frozen stopped by ice from a storm. The wind direction, sigma_wd, and residual wind speed measurements are incorrect. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/07/2005 | 0010 | 01/07/2005 | 1550 |
Subject: | SGP/EBBR/E18 - Wind Direction Frozen Stopped By Ice Storm |
DataStreams: | sgp30ebbrE18.b1, sgp5ebbrE18.b1, sgp15ebbrE18.b1 |
Description: | The wind direction sensor was frozen stopped by ice from a storm. The wind direction, sigma_wd, and residual wind speed measurements are incorrect. |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/05/2005 | 1115 | 01/06/2005 | 2055 |
Subject: | SGP/EBBR/E19 - Wind Direction Frozen Stopped By Ice Storm |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | The wind direction sensor was frozen stopped by ice from a storm. The wind direction, sigma_wd, and residual wind speed measurements are incorrect. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/10/2004 | 1400 | 02/17/2005 | 1630 |
Subject: | SGP/EBBR/E26 - Aging Pressure Sensor Output Changes With Supply Voltage |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | The E26 EBBR pressure sensor experienced the poor internal voltage regulation problem experienced at other sites last year. Aging of the internal voltage regulator caused the pressure output to drop by 1-2 kPa (0.1-0.5 kPa is normal) when the Sun rose and the solar panel goosed the battery voltage up. The sensible and latent heat fluxes were not significantly affected. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/15/2004 | 1600 | 03/07/2005 | 2100 |
Subject: | SGP/EBBR/E27 - Soil Moisture #5 Incorrect |
DataStreams: | sgp30ebbrE27.b1 |
Description: | The wrong CR10 program loaded from the storage module on 9/15/04. During the previous change of program, an attempt was made to store the new program in the storage module; site ops techs did not notice that this attempt failed, leaving the old program in the storage module. When power was cyled on 9/15/04, the old program with flat sm5 coefficients loaded into the CR10. The "flat" coefficients resulted in the sm5 soil moisture hovering around 30%. 15 minute raw sm5 resistances were correct. Average soil heat flow, sensible heat flux, and latent heat flux were not significantly affected and so are not included in the list of incorrect measurements. |
Measurements: | sgp30ebbrE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/11/2005 | 1330 | 01/12/2005 | 0050 |
Subject: | SGP/EBBR/E7 - Wind Speed Frozen Stopped By Ice |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | An ice storm froze the wind speed sensor to a stopped condition. The wind speed and residual wind speed measurements are incorrect. This sensor failure does not affect the sensible and latent heat fluxes. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/12/2005 | 0720 | 01/12/2005 | 1605 |
02/07/2005 | 0200 | 02/08/2005 | 1930 |
Subject: | SGP/EBBR/E8 - Wind Speed Frozen Stopped By Ice Storm |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | An ice storm froze the wind speed sensor to a stopped condition. The wind speed and residual wind speed measurements are incorrect. This sensor failure does not affect the sensible and latent heat fluxes. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/07/2005 | 0500 | 02/08/2005 | 0330 |
Subject: | SGP/EBBR/E8 - Wind Direction Frozen Stopped By Ice |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The wind direction sensor was frozen stopped by ice from a storm. The wind direction, sigma_wd, and residual wind speed measurements are incorrect. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/10/2005 | 2300 | 01/11/2005 | 1645 |
Subject: | SGP/EBBR/E9 - Wind Direction Frozen Stopped By Ice |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Description: | The wind direction sensor was frozen stopped by ice from a storm. The wind direction, sigma_wd, and residual wind speed measurements are incorrect. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/08/2005 | 0530 | 02/09/2005 | 1830 |
Subject: | SGP/EBBR/E9 - Wind Speed Frozen Stopped By Ice |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Description: | An ice storm froze the wind speed sensor to a stopped condition. The wind speed and residual wind speed measurements are incorrect. This sensor failure does not affect the sensible and latent heat fluxes. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/11/2005 | 1040 | 01/11/2005 | 0755 |
Subject: | SGP/EBBR/E12 - Wind Speed Frozen Stopped By Ice |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | An ice storm froze the wind speed sensor to a stopped condition. The wind speed and residual wind speed measurements are incorrect. This sensor failure does not affect the sensible and latent heat fluxes. |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/11/2005 | 1040 | 01/12/2005 | 0410 |
Subject: | SGP/EBBR/E12 - Wind Direction Frozen Stopped By Ice |
DataStreams: | sgp30ebbrE12.b1, sgp5ebbrE12.b1, sgp15ebbrE12.b1 |
Description: | The wind direction sensor was frozen stopped by ice from a storm. The wind direction, sigma_wd, and residual wind speed measurements are incorrect. |
Measurements: | sgp30ebbrE12.b1:
sgp15ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/12/2005 | 0030 | 01/12/2005 | 1545 |
Subject: | SGP/EBBR/E15 - Wind Speed Frozen Stopped By Ice |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | An ice storm froze the wind speed sensor to a stopped condition. The wind speed and residual wind speed measurements are incorrect. This sensor failure does not affect the sensible and latent heat fluxes. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/12/2005 | 0030 | 01/12/2005 | 0930 |
Subject: | SGP/EBBR/E15 - Wind Direction Frozen Stopped By Ice |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | The wind direction sensor was frozen stopped by ice from a storm. The wind direction, sigma_wd, and residual wind speed measurements are incorrect. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/29/2005 | 0930 | 01/29/2005 | 1300 |
Subject: | SGP/EBBR/E19 - Wind Speed Constant |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | Wind Speed was constant at 5 m/s during this period; the reason is unknown. This sensor failure does not affect the sensible and latent heat fluxes. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/11/2005 | 0915 | 01/11/2005 | 2100 |
Subject: | SGP/EBBR/E22 - Wind Speed Frozen Stopped By Ice |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | An ice storm froze the wind speed sensor to a stopped condition. The wind speed and residual wind speed measurements are incorrect. This sensor failure does not affect the sensible and latent heat fluxes. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/11/2005 | 1730 | 01/11/2005 | 1825 |
Subject: | SGP/EBBR/E22 - Wind Direction Frozen Stopped By Ice |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | The wind direction sensor was frozen stopped by ice from a storm. The wind direction, sigma_wd, and residual wind speed measurements are incorrect. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/11/2005 | 1640 | 01/13/2005 | 1630 |
02/07/2005 | 0600 | 02/09/2005 | 1930 |
Subject: | SGP/EBBR/E2 - Wind Speed Frozen Stopped By Ice |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | An ice storm froze the wind speed sensor to a stopped condition. The wind speed and residual wind speed measurements are incorrect. This sensor failure does not affect the sensible and latent heat fluxes. |
Measurements: | sgp15ebbrE2.b1:
sgp30ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/08/2005 | 2330 | 02/09/2005 | 1600 |
Subject: | SGP/EBBR/E2 - Wind Direction Frozen Stopped By Ice |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | The wind direction sensor was frozen stopped by ice from a storm. The wind direction, sigma_wd, and residual wind speed measurements are incorrect. |
Measurements: | sgp30ebbrE2.b1:
sgp15ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/25/2005 | 0430 | 02/25/2005 | 1130 |
Subject: | SGP/EBBR/E12 - Wind Speed and Direction Frozen by Ice Storm |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | The wind speed and direction sensors were frozen by ice from and ice storm. |
Measurements: | sgp30ebbrE12.b1:
sgp15ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/06/2005 | 1400 | 03/16/2005 | 1900 |
Subject: | SGP/EBBR/E22 - Net Radiation Too Low - Top Dome Broken |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | Net radiation values were too low because of a broken top dome. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/08/2004 | 1830 | 03/01/2005 | 1630 |
Subject: | SGP/EBBR/E15 - Soil Heat Flows Undermeasured, and Latent and Sensible Heat Fluxes |
DataStreams: | sgp30ebbrE15.b1 |
Description: | Use of the wrong soil type soil moisture coefficients caused soil moisture to be greatly undermeasured (a typical wrong value was 4% instead of the correct 17%), soil heat flows to be undermeasured, and sensible and latent heat fluxes to be overestimated. Soil heat flow plate outputs were undermeasured by up to 10%, change of energy storage was undermeasured by up to 80% and total soil heat flow was undermeasured by up to 50% as a result. This would result in approximately a 5% overestimation of both sensible and latent heat fluxes. |
Measurements: | sgp30ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/05/2004 | 1700 | 03/05/2005 | 1730 |
Subject: | SGP/EBBR/E19 - Reprocess: Negative Net Rad calibrations |
DataStreams: | sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | The calibration for negative net radiation was incorrectly entered in the datalogger program. |
Measurements: | sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/10/2003 | 1700 | 03/03/2005 | 1600 |
Subject: | SGP/EBBR/E26 - Incorrect soil type coefficients used |
DataStreams: | sgp30ebbrE26.b1 |
Description: | The soil type coefficents for soil moisture were incorrect from the day of the system installation on, making soil moisture values too large. As a result, soil heat flows were overmeasured, and sensible and latent heat fluxes were underestimated. |
Measurements: | sgp30ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/08/2005 | 1945 | 03/22/2005 | 1900 |
Subject: | SGP/EBBR/E7 - Wrong Datalogger Program |
DataStreams: | sgp30ebbrE7.b1 |
Description: | The wrong version 9 datalogger program (unit 12 instead of 13) was installed during the period listed, causing soil heat flows to be 2 to 3 W m-2 too high, pressure to be 0.35 kPa too high, and soil moistures to be 10% too low. Effects on the sensible and latent heat flow are much smaller than the system error. |
Measurements: | sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/30/2005 | 2000 | 04/13/2005 | 1700 |
Subject: | SGP/EBBR/E20 - Wrong Datalogger Program Installed |
DataStreams: | sgp30ebbrE20.b1 |
Description: | ts4 went offscale and sm4 dropped to a constant 16% when a "new" program was installed on 3/30/05. The wrong program was installed. Because ts4 did not change with time, it did not mess up the calculation of e and h. e and h are slightly overestimated, but only by about 1%, far less than the system error. This is a rare type of instrument problem where the concequences are minimal. Therefore, e and h are not listed as incorrect in this DQR and are still usable during the stated period. |
Measurements: | sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/20/2005 | 1230 | 04/27/2005 | 2000 |
Subject: | SGP/EBBR/E20 - Temp. of thum_left Incorrect At Times |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | The temperature portion of the left T/RH probe temperature spiked at times. This was caused by a faulty surge protector in the J-Panel. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/21/2005 | 0230 | 04/26/2005 | 2215 |
Subject: | SGP/EBBR/E8 - Battery Voltage Problem |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | Battery voltage was too low to give correct data. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/04/2004 | 1645 | 03/02/2005 | 1600 |
Subject: | SGP/EBBR/E27 - SM2 Coefficients Incorect |
DataStreams: | sgp15ebbrE27.b1, sgp30ebbrE27.b1 |
Description: | Soil Moisture #2 coefficients in the CR10 datalogger program were incorrect during the time that an old program version was installed. |
Measurements: | sgp15ebbrE27.b1:
sgp30ebbrE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/11/2005 | 1430 | 05/13/2005 | 1900 |
Subject: | SGP/EBBR/E13 - Net Radiometer Tilted Half Degree West |
DataStreams: | sgp30ebbrE13.b1, sgp5ebbrE13.b1, sgp15ebbrE13.b1 |
Description: | The net radiation daily peak on sunny days occured 1/2 hour after the net radiation peak at E2, E9, and the SIRS because the sensor was tipped to the west about 1/2 degree. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/02/2005 | 1215 | 06/11/2005 | 0600 |
Subject: | SGP/EBBR/E9 - Soil moisture too high |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1 |
Description: | The 15 and 30 minute data for sm3 indicated that there was a problem with the sensor that kept the soil moisture artificially high. This corrected itself on 6/11, but it is not clear why. A new sensor was installed on 6/17 and it has behaved well since then. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/08/2005 | 1900 | 07/06/2005 | 1630 |
Subject: | SGP/EBBR/E13 - Sensible and Latent Heat Fluxes Incorrect (AEM failure) |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | The AEM was stuck in the home 30 position most of this period. Therefore, the sensible and latent heat fluxes are incorrect during most of this period. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/13/2005 | 0800 | 05/24/2005 | 1700 |
Subject: | SGP/EBBR/E15 - Net Radiation Too Large - Water in Bottom |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | Water in the bottom dome of the net radiometer made the measurements incorrectly large. Sensible and latent heat fluxes were therefore incorrect. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/07/2005 | 2300 | 05/12/2005 | 1445 |
Subject: | SGP/EBBR/E19 - Net Radiation Incorrect - Cable Chewed by Rodents |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | The net radiation measurement was incorrect after the net radiometer cable was chewed by rodents. Sensible and latent heat fluxes were therefore incorrect. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/01/2001 | 0000 | 03/31/2005 | 2100 |
Subject: | SGP/EBBR/E13 - Improved EBBR CR10 Program |
DataStreams: | sgp30ebbrE13.b1 |
Description: | Effective 20050331.2100, the EBBR.E13 CR10 program was revised to improve the quality of the primary measurements as follows: 1) An RMS procedure is used to determine max and min limits of acceptable soil heat flow. This is applied to the individual soil sets. Measurements outside the limits are rejected and the measurements within the limits are used to calculate the average soil heat flow. 2) Max and min limits are used to determine incorrect values of net radiation, sensible heat flux (h), latent heat flux (e), and automatic exchange mechanism (AEM) signal. If AEM signal is outside the limits, h and e are set to 999s. If net radiation is outside the limits, h and e are set to 999s. If h or e are outside the limits, h and e are set to 999s. Virtually no incorrect soil measurement will affect the primary measurements of h and e. By setting h and e to 999s, it can be easily seen that the primary variables are incorrect; no other interpretation is possible. Prior to 20050331, the improved CR10 program was NOT in effect. DQRs have been submitted for known instances of incorrect soil measurements which affected the quality of the primary measurements of h and e. Note: the DQR begin date is the begin date of the sgp30ebbrE13.b1 data stream. The earlier version of the CR10 program was also used on previous EBBR datastream names (e.g. sgp30ebbrE13.a1). |
Measurements: | sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/01/2001 | 0000 | 03/29/2005 | 1900 |
Subject: | SGP/EBBR/E15 - Improved EBBR CR10 Program |
DataStreams: | sgp30ebbrE15.b1 |
Description: | Effective 20050329.1900, the EBBR.E15 CR10 program was revised to improve the quality of the primary measurements as follows: 1) An RMS procedure is used to determine max and min limits of acceptable soil heat flow. This is applied to the individual soil sets. Measurements outside the limits are rejected and the measurements within the limits are used to calculate the average soil heat flow. 2) Max and min limits are used to determine incorrect values of net radiation, sensible heat flux (h), latent heat flux (e), and automatic exchange mechanism (AEM) signal. If AEM signal is outside the limits, h and e are set to 999s. If net radiation is outside the limits, h and e are set to 999s. If h or e are outside the limits, h and e are set to 999s. Virtually no incorrect soil measurement will affect the primary measurements of h and e. By setting h and e to 999s, it can be easily seen that the primary variables are incorrect; no other interpretation is possible. Prior to 20050329, the improved CR10 program was NOT in effect. DQRs have been submitted for known instances of incorrect soil measurements which affected the quality of the primary measurements of h and e. Note: the DQR begin date is the begin date of the sgp30ebbrE15.b1 data stream. The earlier version of the CR10 program was also used on previous EBBR datastream names (e.g. sgp30ebbrE15.a1). |
Measurements: | sgp30ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/01/2001 | 0000 | 03/30/2005 | 0000 |
Subject: | SGP/EBBR/E18 - Improved EBBR CR10 Program |
DataStreams: | sgp30ebbrE18.b1 |
Description: | Effective 20050330.0000, the EBBR.E18 CR10 program was revised to improve the quality of the primary measurements as follows: 1) An RMS procedure is used to determine max and min limits of acceptable soil heat flow. This is applied to the individual soil sets. Measurements outside the limits are rejected and the measurements within the limits are used to calculate the average soil heat flow. 2) Max and min limits are used to determine incorrect values of net radiation, sensible heat flux (h), latent heat flux (e), and automatic exchange mechanism (AEM) signal. If AEM signal is outside the limits, h and e are set to 999s. If net radiation is outside the limits, h and e are set to 999s. If h or e are outside the limits, h and e are set to 999s. Virtually no incorrect soil measurement will affect the primary measurements of h and e. By setting h and e to 999s, it can be easily seen that the primary variables are incorrect; no other interpretation is possible. Prior to 20050330, the improved CR10 program was NOT in effect. DQRs have been submitted for known instances of incorrect soil measurements which affected the quality of the primary measurements of h and e. Note: the DQR begin date is the begin date of the sgp30ebbrE18.b1 data stream. The earlier version of the CR10 program was also used on previous EBBR datastream names (e.g. sgp30ebbrE18.a1). |
Measurements: | sgp30ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/01/2001 | 0000 | 03/31/2005 | 1600 |
Subject: | SGP/EBBR/E19 - Improved EBBR CR10 Program |
DataStreams: | sgp30ebbrE19.b1 |
Description: | Effective 20050331.1600, the EBBR.E19 CR10 program was revised to improve the quality of the primary measurements as follows: 1) An RMS procedure is used to determine max and min limits of acceptable soil heat flow. This is applied to the individual soil sets. Measurements outside the limits are rejected and the measurements within the limits are used to calculate the average soil heat flow. 2) Max and min limits are used to determine incorrect values of net radiation, sensible heat flux (h), latent heat flux (e), and automatic exchange mechanism (AEM) signal. If AEM signal is outside the limits, h and e are set to 999s. If net radiation is outside the limits, h and e are set to 999s. If h or e are outside the limits, h and e are set to 999s. Virtually no incorrect soil measurement will affect the primary measurements of h and e. By setting h and e to 999s, it can be easily seen that the primary variables are incorrect; no other interpretation is possible. Prior to 20050331, the improved CR10 program was NOT in effect. DQRs have been submitted for known instances of incorrect soil measurements which affected the quality of the primary measurements of h and e. Note: the DQR begin date is the begin date of the sgp30ebbrE19.b1 data stream. The earlier version of the CR10 program was also used on previous EBBR datastream names (e.g. sgp30ebbrE19.a1). |
Measurements: | sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/01/2001 | 0000 | 03/30/2005 | 1900 |
Subject: | SGP/EBBR/E20 - Improved EBBR CR10 Program |
DataStreams: | sgp30ebbrE20.b1 |
Description: | Effective 20050330.1900, the EBBR.E20 CR10 program was revised to improve the quality of the primary measurements as follows: 1) An RMS procedure is used to determine max and min limits of acceptable soil heat flow. This is applied to the individual soil sets. Measurements outside the limits are rejected and the measurements within the limits are used to calculate the average soil heat flow. 2) Max and min limits are used to determine incorrect values of net radiation, sensible heat flux (h), latent heat flux (e), and automatic exchange mechanism (AEM) signal. If AEM signal is outside the limits, h and e are set to 999s. If net radiation is outside the limits, h and e are set to 999s. If h or e are outside the limits, h and e are set to 999s. Virtually no incorrect soil measurement will affect the primary measurements of h and e. By setting h and e to 999s, it can be easily seen that the primary variables are incorrect; no other interpretation is possible. Prior to 20050330, the improved CR10 program was NOT in effect. DQRs have been submitted for known instances of incorrect soil measurements which affected the quality of the primary measurements of h and e. Note: the DQR begin date is the begin date of the sgp30ebbrE20.b1 data stream. The earlier version of the CR10 program was also used on previous EBBR datastream names (e.g. sgp30ebbrE20.a1). |
Measurements: | sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/01/2001 | 0000 | 03/30/2005 | 1930 |
Subject: | SGP/EBBR/E22 - Improved EBBR CR10 Program |
DataStreams: | sgp30ebbrE22.b1 |
Description: | Effective 20050330.1930, the EBBR.E20 CR10 program was revised to improve the quality of the primary measurements as follows: 1) An RMS procedure is used to determine max and min limits of acceptable soil heat flow. This is applied to the individual soil sets. Measurements outside the limits are rejected and the measurements within the limits are used to calculate the average soil heat flow. 2) Max and min limits are used to determine incorrect values of net radiation, sensible heat flux (h), latent heat flux (e), and automatic exchange mechanism (AEM) signal. If AEM signal is outside the limits, h and e are set to 999s. If net radiation is outside the limits, h and e are set to 999s. If h or e are outside the limits, h and e are set to 999s. Virtually no incorrect soil measurement will affect the primary measurements of h and e. By setting h and e to 999s, it can be easily seen that the primary variables are incorrect; no other interpretation is possible. Prior to 20050330, the improved CR10 program was NOT in effect. DQRs have been submitted for known instances of incorrect soil measurements which affected the quality of the primary measurements of h and e. Note: the DQR begin date is the begin date of the sgp30ebbrE20.b1 data stream. The earlier version of the CR10 program was also used on previous EBBR datastream names (e.g. sgp30ebbrE20.a1). |
Measurements: | sgp30ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/01/2001 | 0000 | 03/31/2005 | 1700 |
Subject: | SGP/EBBR/E26 - Improved EBBR CR10 Program |
DataStreams: | sgp30ebbrE26.b1 |
Description: | Effective 20050331.1700, the EBBR.E26 CR10 program was revised to improve the quality of the primary measurements as follows: 1) An RMS procedure is used to determine max and min limits of acceptable soil heat flow. This is applied to the individual soil sets. Measurements outside the limits are rejected and the measurements within the limits are used to calculate the average soil heat flow. 2) Max and min limits are used to determine incorrect values of net radiation, sensible heat flux (h), latent heat flux (e), and automatic exchange mechanism (AEM) signal. If AEM signal is outside the limits, h and e are set to 999s. If net radiation is outside the limits, h and e are set to 999s. If h or e are outside the limits, h and e are set to 999s. Virtually no incorrect soil measurement will affect the primary measurements of h and e. By setting h and e to 999s, it can be easily seen that the primary variables are incorrect; no other interpretation is possible. Prior to 20050331, the improved CR10 program was NOT in effect. DQRs have been submitted for known instances of incorrect soil measurements which affected the quality of the primary measurements of h and e. Note: the DQR begin date is the begin date of the sgp30ebbrE26.b1 data stream. The earlier version of the CR10 program was also used on previous EBBR datastream names (e.g. sgp30ebbrE26.a1). ~ |
Measurements: | sgp30ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/07/2003 | 0000 | 03/30/2005 | 1730 |
Subject: | SGP/EBBR/E27 - Improved EBBR CR10 Program |
DataStreams: | sgp30ebbrE27.b1 |
Description: | Effective 20050330.1730, the EBBR.E27 CR10 program was revised to improve the quality of the primary measurements as follows: 1) An RMS procedure is used to determine max and min limits of acceptable soil heat flow. This is applied to the individual soil sets. Measurements outside the limits are rejected and the measurements within the limits are used to calculate the average soil heat flow. 2) Max and min limits are used to determine incorrect values of net radiation, sensible heat flux (h), latent heat flux (e), and automatic exchange mechanism (AEM) signal. If AEM signal is outside the limits, h and e are set to 999s. If net radiation is outside the limits, h and e are set to 999s. If h or e are outside the limits, h and e are set to 999s. Virtually no incorrect soil measurement will affect the primary measurements of h and e. By setting h and e to 999s, it can be easily seen that the primary variables are incorrect; no other interpretation is possible. Prior to 20050330, the improved CR10 program was NOT in effect. DQRs have been submitted for known instances of incorrect soil measurements which affected the quality of the primary measurements of h and e. Note: the DQR begin date is the begin date of the sgp30ebbrE27.b1 data stream. The earlier version of the CR10 program was also used on previous EBBR datastream names (e.g. sgp30ebbrE27.a1). |
Measurements: | sgp30ebbrE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/01/2001 | 0000 | 03/24/2005 | 1700 |
Subject: | SGP/EBBR/E2 - Improved EBBR CR10 Program |
DataStreams: | sgp30ebbrE2.b1 |
Description: | Effective 20050324.1700, the EBBR.E2 CR10 program was revised to improve the quality of the primary measurements as follows: 1) An RMS procedure is used to determine max and min limits of acceptable soil heat flow. This is applied to the individual soil sets. Measurements outside the limits are rejected and the measurements within the limits are used to calculate the average soil heat flow. 2) Max and min limits are used to determine incorrect values of net radiation, sensible heat flux (h), latent heat flux (e), and automatic exchange mechanism (AEM) signal. If AEM signal is outside the limits, h and e are set to 999s. If net radiation is outside the limits, h and e are set to 999s. If h or e are outside the limits, h and e are set to 999s. Virtually no incorrect soil measurement will affect the primary measurements of h and e. By setting h and e to 999s, it can be easily seen that the primary variables are incorrect; no other interpretation is possible. Prior to 20050324, the improved CR10 program was NOT in effect. DQRs have been submitted for known instances of incorrect soil measurements which affected the quality of the primary measurements of h and e. Note: the DQR begin date is the begin date of the sgp30ebbrE2.b1 data stream. The earlier version of the CR10 program was also used on previous EBBR datastream names (e.g. sgp30ebbrE2.a1). |
Measurements: | sgp30ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/01/2001 | 0000 | 03/22/2005 | 1900 |
Subject: | SGP/EBBR/E7 - Improved EBBR CR10 Program |
DataStreams: | sgp30ebbrE7.b1 |
Description: | Effective 20050322.1900, the EBBR.E7 CR10 program was revised to improve the quality of the primary measurements as follows: 1) An RMS procedure is used to determine max and min limits of acceptable soil heat flow. This is applied to the individual soil sets. Measurements outside the limits are rejected and the measurements within the limits are used to calculate the average soil heat flow. 2) Max and min limits are used to determine incorrect values of net radiation, sensible heat flux (h), latent heat flux (e), and automatic exchange mechanism (AEM) signal. If AEM signal is outside the limits, h and e are set to 999s. If net radiation is outside the limits, h and e are set to 999s. If h or e are outside the limits, h and e are set to 999s. Virtually no incorrect soil measurement will affect the primary measurements of h and e. By setting h and e to 999s, it can be easily seen that the primary variables are incorrect; no other interpretation is possible. Prior to 20050322, the improved CR10 program was NOT in effect. DQRs have been submitted for known instances of incorrect soil measurements which affected the quality of the primary measurements of h and e. Note: the DQR begin date is the begin date of the sgp30ebbrE7.b1 data stream. The earlier version of the CR10 program was also used on previous EBBR datastream names (e.g. sgp30ebbrE7.a1). |
Measurements: | sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/01/2001 | 0000 | 03/22/2005 | 1730 |
Subject: | SGP/EBBR/E8 - Improved EBBR CR10 Program |
DataStreams: | sgp30ebbrE8.b1 |
Description: | Effective 20050322.1730, the EBBR.E8 CR10 program was revised to improve the quality of the primary measurements as follows: 1) An RMS procedure is used to determine max and min limits of acceptable soil heat flow. This is applied to the individual soil sets. Measurements outside the limits are rejected and the measurements within the limits are used to calculate the average soil heat flow. 2) Max and min limits are used to determine incorrect values of net radiation, sensible heat flux (h), latent heat flux (e), and automatic exchange mechanism (AEM) signal. If AEM signal is outside the limits, h and e are set to 999s. If net radiation is outside the limits, h and e are set to 999s. If h or e are outside the limits, h and e are set to 999s. Virtually no incorrect soil measurement will affect the primary measurements of h and e. By setting h and e to 999s, it can be easily seen that the primary variables are incorrect; no other interpretation is possible. Prior to 20050322, the improved CR10 program was NOT in effect. DQRs have been submitted for known instances of incorrect soil measurements which affected the quality of the primary measurements of h and e. Note: the DQR begin date is the begin date of the sgp30ebbrE8.b1 data stream. The earlier version of the CR10 program was also used on previous EBBR datastream names (e.g. sgp30ebbrE8.a1). |
Measurements: | sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/01/2001 | 0000 | 03/22/2005 | 1600 |
Subject: | SGP/EBBR/E9 - Improved EBBR CR10 Program |
DataStreams: | sgp30ebbrE9.b1 |
Description: | Effective 20050322.1600, the EBBR.E9 CR10 program was revised to improve the quality of the primary measurements as follows: 1) An RMS procedure is used to determine max and min limits of acceptable soil heat flow. This is applied to the individual soil sets. Measurements outside the limits are rejected and the measurements within the limits are used to calculate the average soil heat flow. 2) Max and min limits are used to determine incorrect values of net radiation, sensible heat flux (h), latent heat flux (e), and automatic exchange mechanism (AEM) signal. If AEM signal is outside the limits, h and e are set to 999s. If net radiation is outside the limits, h and e are set to 999s. If h or e are outside the limits, h and e are set to 999s. Virtually no incorrect soil measurement will affect the primary measurements of h and e. By setting h and e to 999s, it can be easily seen that the primary variables are incorrect; no other interpretation is possible. Prior to 20050322, the improved CR10 program was NOT in effect. DQRs have been submitted for known instances of incorrect soil measurements which affected the quality of the primary measurements of h and e. Note: the DQR begin date is the begin date of the sgp30ebbrE9.b1 data stream. The earlier version of the CR10 program was also used on previous EBBR datastream names (e.g. sgp30ebbrE9.a1). |
Measurements: | sgp30ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/01/2001 | 0000 | 03/29/2005 | 1830 |
Subject: | SGP/EBBR/E12 - Improved EBBR CR10 Program |
DataStreams: | sgp30ebbrE12.b1 |
Description: | Effective 20050329.1830, the EBBR.E12 CR10 program was revised to improve the quality of the primary measurements as follows: 1) An RMS procedure is used to determine max and min limits of acceptable soil heat flow. This is applied to the individual soil sets. Measurements outside the limits are rejected and the measurements within the limits are used to calculate the average soil heat flow. 2) Max and min limits are used to determine incorrect values of net radiation, sensible heat flux (h), latent heat flux (e), and automatic exchange mechanism (AEM) signal. If AEM signal is outside the limits, h and e are set to 999s. If net radiation is outside the limits, h and e are set to 999s. If h or e are outside the limits, h and e are set to 999s. Virtually no incorrect soil measurement will affect the primary measurements of h and e. By setting h and e to 999s, it can be easily seen that the primary variables are incorrect; no other interpretation is possible. Prior to 20050329, the improved CR10 program was NOT in effect. DQRs have been submitted for known instances of incorrect soil measurements which affected the quality of the primary measurements of h and e. Note: the DQR begin date is the begin date of the sgp30ebbrE12.b1 data stream. The earlier version of the CR10 program was also used on previous EBBR datastream names (e.g. sgp30ebbrE12.a1). |
Measurements: | sgp30ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/05/2005 | 1730 | 07/06/2005 | 1930 |
08/23/2005 | 1900 | 08/31/2005 | 1700 |
09/01/2005 | 1730 | 09/02/2005 | 0300 |
09/02/2005 | 1100 | 09/15/2005 | 2230 |
09/16/2005 | 0100 | 09/28/2005 | 1830 |
Subject: | SGP/EBBR/E20 - Soil Temperature and Soil Moisture #4 Measurements Incorrect |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1 |
Description: | The ts4 sensor reported 99999 values because the sensor leads were loose. Soil moisture #4 reported low values (steady 16%) during this time period. The sensible and latent heat fluxes were not affected by this problem, only soil temperature and soil moisture #4. See soil temperatures and soil moistures from sets #2, #3, and #5 for correct values (note: soil temperature #1 was having problems during the same time). |
Measurements: | sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/20/2005 | 2030 | 10/07/2005 | 1530 |
Subject: | SGP/EBBR/E7 - Data Missing |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | Data was missing because of a communication problem. However, previously missing data from 10/07-18 was recovered from the EBRR storage module and ingested, thereby reducing the data lost period significantly. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/31/2005 | 0900 | 11/01/2005 | 1630 |
Subject: | SGP/EBBR/E9 - Net Radiation Incorrect: Sensible and Latent Heat Fluxes Incorrect |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Description: | Water had entered the top of the net radiometer during the rain on 10/31 early morning. The net radiation, and thus sensible and latent heat fluxes were incorreect during the stated period. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/01/2005 | 2030 | 10/12/2005 | 2130 |
Subject: | SGP/EBBR/E13 - Fluxes Incorrect |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | The AEM has not been reaching the home_30 position fully a few times a day on most days, resulting in zero home_30 measurements. The sensible and latent heat fluxes are incorrect when the home_30 measurement is near zero. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/17/2005 | 2230 | 10/05/2005 | 1600 |
Subject: | SGP/EBBR/E13 - Loose connections |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | The tair_bot spiked around midday, as if it is temperature related, and may have indicated a loose connection. The sensible and latyent heat fluxes and Bowen ratio were incorrect when this condition occurred. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/04/2005 | 2200 | 08/11/2005 | 1530 |
Subject: | SGP/EBBR/E2 - Net Radiation Incorrect; Latent and Sensible Heat Fluxes Incorrect |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | The top dome of the net radiometer was broken and water entered, causing incorrect net radiation measurements. This resulted in incorrect sensible and latent heat fluxes. |
Measurements: | sgp30ebbrE2.b1:
sgp15ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/27/2005 | 0730 | 08/03/2005 | 2100 |
Subject: | SGP/EBBR/E18 - Net Radiation Incorrect; Sensible and Latent Heat Fluxes Incorrect |
DataStreams: | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
Description: | The net radiometer top dome was broken and water got inside, making the net radiation, and sensible and latent heat flux measurements incorrect. |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/27/2005 | 2000 | 08/31/2005 | 1730 |
Subject: | SGP/EBBR/E22 - Net Radiation Incorrect; Sensible and Latent Heat Fluxes Incorrect |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | The net radiometer top dome was broken and water got inside, making the net radiation, and sensible and latent heat flux measurements incorrect. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/05/2005 | 2200 | 08/17/2005 | 1600 |
Subject: | SGP/EBBR/E27 - Net Radiation Incorrect; Sensible and Latent Heat Fluxes Incorrect |
DataStreams: | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
Description: | The net radiometer top dome was broken and water got inside, making the net radiation, and sensible and latent heat flux measurements incorrect. |
Measurements: | sgp5ebbrE27.b1:
sgp15ebbrE27.b1:
sgp30ebbrE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/17/2005 | 1630 | 08/27/2005 | 1930 |
Subject: | SGP/EBBR/E27 - Net Radiation Incorrect; Sensible and Latent Heat Fluxes Incorredt |
DataStreams: | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
Description: | Residual Moisture left in the top dome of the net radiometer after the 8/17/05 servicing caused the net radiation measurements to be biased low, and resulting in sensible and latent heat flux being low also. |
Measurements: | sgp5ebbrE27.b1:
sgp15ebbrE27.b1:
sgp30ebbrE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/10/2005 | 1600 | 03/24/2005 | 1700 |
Subject: | SGP/EBBR/E2 - Average Soil Heat Flux Incorrect |
DataStreams: | sgp30ebbrE2.b1 |
Description: | A typo in the version 9 EBBR program caused average soil heat flow to be zero, resulting in both sensible and latent heat fluxes being approximately 20 watts per meter squared too large during the middle of the day. |
Measurements: | sgp30ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/08/2005 | 2000 | 03/22/2005 | 1900 |
Subject: | SGP/EBBR/E7 - Average Soil Heat Flux Incorrect |
DataStreams: | sgp30ebbrE7.b1 |
Description: | A typo in the version 9 EBBR program caused average soil heat flow to be zero, resulting in both sensible and latent heat fluxes being approximately 20 watts per meter squared too large during the middle of the day. |
Measurements: | sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/08/2005 | 1800 | 03/22/2005 | 1730 |
Subject: | SGP/EBBR/E8 - Average Soil Heat Flux Incorrect |
DataStreams: | sgp30ebbrE8.b1 |
Description: | A typo in the version 9 EBBR program caused average soil heat flow to be zero, resulting in both sensible and latent heat fluxes being approximately 20 watts per meter squared too large during the middle of the day. |
Measurements: | sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/08/2005 | 1630 | 03/22/2005 | 1600 |
Subject: | SGP/EBBR/E9 - Average Soil Heat Flux Incorrect |
DataStreams: | sgp30ebbrE9.b1 |
Description: | A typo in the version 9 EBBR program caused average soil heat flow to be zero, resulting in both sensible and latent heat fluxes being approximately 20 watts per meter squared too large during the middle of the day. |
Measurements: | sgp30ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/01/2005 | 1800 | 03/29/2005 | 1830 |
Subject: | SGP/EBBR/E12 - Average Soil Heat Flux Incorrect |
DataStreams: | sgp30ebbrE12.b1 |
Description: | A typo in the version 9 EBBR program caused average soil heat flow to be zero, resulting in both sensible and latent heat fluxes being approximately 20 watts per meter squared too large during the middle of the day. |
Measurements: | sgp30ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/03/2005 | 2130 | 03/31/2005 | 2100 |
Subject: | SGP/EBBR/E13 - Average Soil Heat Flux Incorrect |
DataStreams: | sgp30ebbrE13.b1 |
Description: | A typo in the version 9 EBBR program caused average soil heat flow to be zero, resulting in both sensible and latent heat fluxes being approximately 20 watts per meter squared too large during the middle of the day. |
Measurements: | sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/01/2005 | 1800 | 03/29/2005 | 1900 |
Subject: | SGP/EBBR/E15 - Average Soil Heat Flux Incorrect |
DataStreams: | sgp30ebbrE15.b1 |
Description: | A typo in the version 9 EBBR program caused average soil heat flow to be zero, resulting in both sensible and latent heat fluxes being approximately 20 watts per meter squared too large during the middle of the day. |
Measurements: | sgp30ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/01/2005 | 2330 | 03/30/2005 | 0000 |
Subject: | SGP/EBBR/E18 - Average Soil Heat Flux Incorrect |
DataStreams: | sgp30ebbrE18.b1 |
Description: | A typo in the version 9 EBBR program caused average soil heat flow to be zero, resulting in both sensible and latent heat fluxes being approximately 20 watts per meter squared too large during the middle of the day. |
Measurements: | sgp30ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/03/2005 | 1730 | 03/31/2005 | 1600 |
Subject: | SGP/EBBR/E19 - Average Soil Heat Flux Incorrect |
DataStreams: | sgp30ebbrE19.b1 |
Description: | A typo in the version 9 EBBR program caused average soil heat flow to be zero, resulting in both sensible and latent heat fluxes being approximately 20 watts per meter squared too large during the middle of the day. |
Measurements: | sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/02/2005 | 1900 | 03/30/2005 | 1900 |
Subject: | SGP/EBBR/E20 - Average Soil Heat Flux Incorrect |
DataStreams: | sgp30ebbrE20.b1 |
Description: | A typo in the version 9 EBBR program caused average soil heat flow to be zero, resulting in both sensible and latent heat fluxes being approximately 20 watts per meter squared too large during the middle of the day. |
Measurements: | sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/16/2005 | 1830 | 03/30/2005 | 1930 |
Subject: | SGP/EBBR/E22 - Average Soil Heat Flux Incorrect |
DataStreams: | sgp30ebbrE22.b1 |
Description: | A typo in the version 9 EBBR program caused average soil heat flow to be zero, resulting in both sensible and latent heat fluxes being approximately 20 watts per meter squared too large during the middle of the day. |
Measurements: | sgp30ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/14/2005 | 0000 | 09/20/2005 | 2000 |
Subject: | SGP/EBBR/E7 - Net Radiation Incorrect; Sensible and Latent Heat Fluxes Incorrect |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | The top dome of the net radiometer was cracked; net radiation, and sensible and latent heat flux measurements were incorrect. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/03/2005 | 1800 | 03/31/2005 | 1700 |
Subject: | SGP/EBBR/E26 - Average Soil Heat Flux Incorrect |
DataStreams: | sgp30ebbrE26.b1 |
Description: | A typo in the version 9 EBBR program caused average soil heat flow to be zero, resulting in both sensible and latent heat fluxes being approximately 20 watts per meter squared too large during the middle of the day. |
Measurements: | sgp30ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/02/2005 | 1600 | 03/30/2005 | 1730 |
Subject: | SGP/EBBR/E27 - Average Soil Heat Flux Incorrect |
DataStreams: | sgp30ebbrE27.b1 |
Description: | A typo in the version 9 EBBR program caused average soil heat flow to be zero, resulting in both sensible and latent heat fluxes being approximately 20 watts per meter squared too large during the middle of the day. |
Measurements: | sgp30ebbrE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/28/2005 | 1930 | 06/28/2005 | 1930 |
Subject: | SGP/EBBR/E8 - Soil Heat Flow #2 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow #2 spiked upwards briefly at 1930 GMT. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/04/2005 | 2000 | 08/09/2005 | 2030 |
Subject: | SGP/EBBR/E8 - Net Radiation Incorrect; Sensible and Latent Heat Fluxes Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The top dome of the net radiometer was cracked and water had gotten inside it, so net radiation, and sensible and latent heat fluxes were incorrect. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/15/2005 | 2130 | 07/19/2005 | 1700 |
Subject: | SGP/EBBR/E15 - Soil Set #5 Temperature, Moisture, and Heat Flow Incorrect |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1 |
Description: | Rodents chewed the cables on all three soil set #5 sensors, making the measurements incorrect. |
Measurements: | sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/13/2003 | 0000 | 07/22/2005 | 2359 |
Subject: | SGP/EBBR/E12 - metadata corrections |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | On 20050723, a series of metadata corrections and additions were completed. These metadata changes apply to all EBBR.E12 data collected by ARM back to the installation of the instrument in August 1993. Please see the current metadata for correct information. These changes do not affect data values or quality. The changes are summarized below: 1) Update of "sensor location" information 2) Addition of installation dates for the systems 3) Correction of soil type 4) Correction of soil moisture units (from "by volume" to "gravimetric") |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/13/2003 | 0000 | 07/22/2005 | 2359 |
Subject: | SGP/EBBR/E13 - metadata corrections |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | On 20050723, a series of metadata corrections and additions were completed. These metadata changes apply to all EBBR.E13 data collected by ARM back to the installation of the instrument in September 1992. Please see the current metadata for correct information. These changes do not affect data values or quality. The changes are summarized below: 1) Update of "sensor location" information 2) Addition of installation dates for systems 3) Correction of soil type 4) Correction of soil moisture units (from "by volume" to "gravimetric") |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/13/2003 | 0000 | 07/22/2005 | 2359 |
Subject: | SGP/EBBR/E15 - metadata corrections |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | On 20050723, a series of metadata corrections and additions were completed. These metadata changes apply to all EBBR.E15 data collected by ARM back to the installation of the instrument in September 1992. Please see the current metadata for correct information. These changes do not affect data values or quality. The changes are summarized below: 1) Update of "sensor location" information 2) Addition of installation dates for the systems 3) Correction of soil type 4) Correction of soil moisture units (from "by volume" to "gravimetric") |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/13/2003 | 0000 | 07/22/2005 | 2359 |
Subject: | SGP/EBBR/E18 - metadata corrections |
DataStreams: | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
Description: | On 20050723, a series of metadata corrections and additions were completed. These metadata changes apply to all EBBR.E18 data collected by ARM back to the installation of the instrument in July 1997. Please see the current metadata for correct information. These changes do not affect data values or quality. The changes are summarized below: 1) Update of "sensor location" information 2) Addition of installation dates for the systems 3) Correction of soil type 4) Correction of soil moisture units (from "by volume" to "gravimetric") |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/13/2003 | 0000 | 07/22/2005 | 2359 |
Subject: | SGP/EBBR/E19 - metadata corrections |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | On 20050723, a series of metadata corrections and additions were completed. These metadata changes apply to all EBBR.E19 data collected by ARM back to the beginning of the data set in May 1997. Please see the current metadata for correct information. These changes do not affect data values or quality. The changes are summarized below: 1) Update of "sensor location" information 2) Addition of installation dates for the systems 3) Correction of soil type 4) Correction of soil moisture units (from "by volume" to "gravimetric") |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/13/2003 | 0000 | 08/16/2005 | 2359 |
Subject: | SGP/EBBR/E27 - metadata corrections |
DataStreams: | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
Description: | On 20050816, a series of metadata corrections and additions were completed. These metadata changes apply to all EBBR.E27 data collected by ARM back to the installation of the instrument in May 2003. Please see the current metadata for correct information. These changes do not affect data values or quality. The changes are summarized below: 1) Update of "sensor location" information 2) Addition of installation dates for the systems 3) Correction of soil moisture units (from "by volume" to "gravimetric") |
Measurements: | sgp5ebbrE27.b1:
sgp15ebbrE27.b1:
sgp30ebbrE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/13/2003 | 0000 | 07/22/2005 | 2359 |
Subject: | SGP/EBBR/E2 - metadata corrections |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | On 20050723, a series of metadata corrections and additions were completed. These metadata changes apply to all EBBR.E2 data collected by ARM back to the installation of the instrument in May 1997. Please see the current metadata for correct information. These changes do not affect data values or quality. The changes are summarized below: 1) Update of "sensor location" information 2) Addition of installation dates for the systems 3) Correction of soil moisture units (from "by volume" to "gravimetric") |
Measurements: | sgp30ebbrE2.b1:
sgp15ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/13/2003 | 0000 | 07/22/2005 | 2359 |
Subject: | SGP/EBBR/E7 - metadata corrections |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | On 20050723, a series of metadata corrections and additions were completed. These metadata changes apply to all EBBR.E7 data collected by ARM back to the installation of the instrument in May 1993. Please see the current metadata for correct information. These changes do not affect data values or quality. The changes are summarized below: 1) Update of "sensor location" information 2) Addition of installation dates for the systems 3) Correction of soil moisture units (from "by volume" to "gravimetric") |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/13/2003 | 0000 | 07/22/2005 | 2359 |
Subject: | SGP/EBBR/E8 - metadata corrections |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | On 20050723, a series of metadata corrections and additions were completed. These metadata changes apply to all EBBR.E8 data collected by ARM back to the installation of the instrument in December 1992. Please see the current metadata for correct information. These changes do not affect data values or quality. The changes are summarized below: 1) Update of "sensor location" information 2) Addition of installation dates for the systems 3) Correction of soil moisture units (from "by volume" to "gravimetric") |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/13/2003 | 0000 | 04/19/2005 | 2359 |
Subject: | SGP/EBBR/E9 - metadata corrections |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Description: | On 20050420, a series of metadata corrections and additions were completed. These metadata changes apply to all EBBR.E9 data collected by ARM back to the installation of the instrument in December 1992. Please see the current metadata for correct information. These changes do not affect data values or quality. The changes are summarized below: 1) Update of "sensor location" information 2) Addition of installation dates for the systems 3) Correction of soil moisture units (from "by volume" to "gravimetric") |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/15/2003 | 0000 | 07/22/2005 | 2359 |
Subject: | SGP/EBBR/E20 - metadata corrections |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | On 20050723, a series of metadata corrections and additions were completed. These metadata changes apply to all EBBR.E20 data collected by ARM back to the installation of the instrument in April 1993. Please see the current metadata for correct information. These changes do not affect data values or quality. The changes are summarized below: 1) Update of "sensor location" information 2) Addition of installation dates for the systems 3) Correction of soil moisture units (from "by volume" to "gravimetric") |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/13/2003 | 0000 | 07/22/2005 | 2359 |
Subject: | SGP/EBBR/E22 - metadata corrections |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | On 20050723, a series of metadata corrections and additions were completed. These metadata changes apply to all EBBR.E22 data collected by ARM back to the installation of the instrument in April 1993. Please see the current metadata for correct information. These changes do not affect data values or quality. The changes are summarized below: 1) Update of "sensor location" information 2) Addition of installation dates for the systems 3) Correction of soil type 4) Correction of soil moisture units (from "by volume" to "gravimetric") |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/13/2003 | 0000 | 07/22/2005 | 2359 |
Subject: | SGP/EBBR/E26 - metadata corrections |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | On 20050723, a series of metadata corrections and additions were completed. These metadata changes apply to all EBBR.E26 data collected by ARM back to the installation of the instrument in June 1992. Please see the current metadata for correct information. These changes do not affect data values or quality. The changes are summarized below: 1) Update of "sensor location" information 2) Addition of installation dates for the systems 3) Correction of soil moisture units (from "by volume" to "gravimetric") |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/01/2001 | 0000 | 04/08/2002 | 2359 |
Subject: | SGP/EBBR/E25 - metadata errors |
DataStreams: | sgp15ebbrE25.b1, sgp30ebbrE25.b1 |
Description: | During the entire span of EBBR.E25 data, the cdf data files contained a few minor global metadata errors. These errors do not affect the data values or quality. The errors include: * incomplete sensor location information * incomplete installation date information * soil moisture documented as "by volume" should have read "gravimetric" |
Measurements: | sgp30ebbrE25.b1:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/24/2005 | 0600 | 10/25/2005 | 0100 |
10/31/2005 | 1030 | 12/20/2005 | 1940 |
Subject: | SGP/EBBR/E19 - Net Radiation, Sensible and Latent Heat Fluxes Incorrect |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | The net radiometer cable was chewed by an animal. Net Radiation, Sensible Heat Flux, and Latent Heat Flux were incorrect during the stated periods. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/13/2005 | 0000 | 01/19/2006 | 1630 |
Subject: | SGP/EBBR/E19 - Incorrect Sensible and Latent Heat Fluxes |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | The gradient measurements and latent and sensible heat fluxes were incorrect because the AEM was not exchanging. The problem was fixed by replacing the AEM. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/30/2005 | 0700 | 11/01/2005 | 1615 |
Subject: | SGP/EBBR/E9 - Net Radiation, Sensible and Latent Heat Fluxes Incorrect |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Description: | The top dome of the net radiometer was broken, resulting in net radiations that were too small, and thus sensible and latent heat fluxes that were too small. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/30/2005 | 0000 | 11/08/2005 | 1815 |
Subject: | SGP/EBBR/E15 - Net Radiation, Sensible and Latent Heat Fluxes Incorrect. |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | The top dome of the net radiometer was broken, resulting in net radiations that were too small, and thus sensible and latent heat fluxes that were too small. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/20/2005 | 0800 | 12/22/2005 | 1500 |
Subject: | SGP/EBBR/E26 - Data Suspect During Low Supply Voltage |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | Low battery supply voltage resulted in high home signals and questionable data at times. Data for battery voltage <10.5 volts should be considered to be incorrect. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/17/2005 | 2130 | 12/21/2005 | 1930 |
12/08/2005 | 0700 | 12/08/2005 | 0700 |
12/08/2005 | 1130 | 12/08/2005 | 1430 |
12/09/2005 | 0700 | 12/09/2005 | 0700 |
Subject: | SGP/EBBR/E2 - Sensible and Latent Heat Fluxes Incorrect |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | The sensible and latent heat fluxes and the gradient measurements were incorrect during this period because the AEM did not switch positions properly. |
Measurements: | sgp30ebbrE2.b1:
sgp15ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/10/2006 | 1330 | 01/17/2006 | 1730 |
Subject: | SGP/EBBR/E15 - Incorrect Sensible and Latent Heat Fluxes |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | Incorrect gradient measurements and sensible and latent heat fluxes resulted from the AEM not exchanging. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/27/2005 | 0000 | 12/18/2005 | 1200 |
Subject: | SGP/EBBR/E15 - Reprocess: Sensible and Latent Heat Fluxes, Soil Set #2 Sometimes Incorrect |
DataStreams: | sgp30ebbrE15.b1 |
Description: | Spikes in soil set #2 sometimes caused sensible and latent heat fluxes to be incorrect. There is no electronic connection between the soil set #2 sensors, so it is suspected that there was a problem in the soil set #2 cable bundle. A power outage ensued, after which the soil set #2 problem did not re-occur. |
Measurements: | sgp30ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/21/2005 | 0930 | 12/21/2005 | 1730 |
Subject: | SGP/EBBR/E20 - Data Incorrect During J-Panel Replacement |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | Much of the data was incorrect during replacement of J-Panel #1. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/04/2005 | 1700 | 02/20/2006 | 2230 |
Subject: | SGP/EBBR/E7 - Barometric Pressure Incorrect |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | The pressure measurement sometimes dropped to zero during daytime because of unstable battery voltage. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/26/2005 | 1900 | 01/10/2006 | 1630 |
Subject: | SGP/EBBR/E8 - AEM Not Exchanging |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The AEM hung in one position frequently and sometimes for many days during this period While the AEM was not switching, the sensible and latent heat fluxes and gradient measurements were incorrect. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/27/2005 | 1730 | 01/02/2006 | 2330 |
Subject: | SGP/EBBR/E12 - Sensible and Latent Heat Fluxes Incorrect at Times |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1 |
Description: | Large spikes in soil temperature #2 are causing the soil heat storage term to blow up, and thus the average soil heat flow, sensible heat flux, and latent heat flux to be incorrect at times. |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/08/2005 | 1100 | 12/08/2005 | 1300 |
Subject: | SGP/EBBR/E12 - Multiplexer failure |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | Sensible and latent heat fluxes are incorrect during a period when multiplexer #2 experienced a failure. |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/18/2005 | 0900 | 12/20/2005 | 1600 |
12/21/2005 | 0500 | 12/21/2005 | 1830 |
Subject: | SGP/EBBR/E13 - Wind Speed Sensor Frozen |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | The wind speed sensor was frozen to a stopped posiiton; wind speed data is incorrect. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/10/2006 | 0730 | 01/18/2006 | 2000 |
Subject: | SGP/EBBR/E22 - Sensible and Latent Heat Fluxes Incorrect |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/03/2006 | 0000 | 05/10/2006 | 1130 |
05/25/2006 | 2100 | 06/20/2006 | 1730 |
Subject: | SGP/EBBR/E27 - Sensible and Latent Heat Fluxes Incorrect |
DataStreams: | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
Description: | The left T/RH probe temperature sensor was spiking some of the time during these periods. Therefore, sensible and latent heat fluxes were incorrect during parts of the periods as well. |
Measurements: | sgp5ebbrE27.b1:
sgp15ebbrE27.b1:
sgp30ebbrE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/12/2006 | 1900 | 05/24/2006 | 1700 |
Subject: | SGP/EBBR/E22 - Fluxes Incorrect When Soil Heat Flow #3 Out of Range |
DataStreams: | sgp30ebbrE22.b1 |
Description: | When soil heat flow #3 exceeded the range limit in the EBBR program, it was replaced with 9999es. This caused the ave_shf and the sensible and latent heat fluxes to be incorrect. |
Measurements: | sgp30ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/15/2006 | 2300 | 04/25/2006 | 1700 |
Subject: | SGP/EBBR/E12 - Fluxes and Gradient Measurements Incorrect |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | The AEM hung in one position; therefore the fluxes and gradient measurements were incorrect. |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/06/2006 | 1730 | 01/09/2006 | 0730 |
01/10/2006 | 1330 | 01/10/2006 | 1930 |
02/17/2006 | 2005 | 02/20/2006 | 1030 |
Subject: | SGP/EBBR/E13 - Missing Data |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | The Data was missing during these three periods for various reasons. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/22/2006 | 2330 | 01/26/2006 | 2040 |
Subject: | SGP/EBBR/E13 - Flux and Gradient Measurements Incorrect |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | The AEM hung in one position; flux and gradient measurements were therefore incorrect. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/02/2006 | 0330 | 04/05/2006 | 1435 |
Subject: | SGP/EBBR/E13 - Net Radiation and Fluxes Suspect |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | The net radiometer had water in it, making the net radiation and flux measurements suspect; the latter do not appear to be incorrect, but should be considered as suspect anyway. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/22/2006 | 1300 | 02/22/2006 | 1530 |
Subject: | SGP/EBBR/E15 - Wind Speed Sensor Frozen Stopped |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | Freezing rain froze the wind speed sensor to a stopped condition during the period. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/22/2006 | 0000 | 01/26/2006 | 0530 |
02/18/2006 | 0400 | 02/18/2006 | 2030 |
02/20/2006 | 0200 | 02/24/2006 | 1630 |
03/31/2006 | 0530 | 03/31/2006 | 0900 |
Subject: | SGP/EBBR/E18 - Wind Speed Zero |
DataStreams: | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
Description: | Wind Speed was zero, presumably because the sensor was frozen stopped during freezing rain. |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/21/2006 | 0400 | 01/21/2006 | 1530 |
Subject: | SGP/EBBR/E2 - Wind Speed Sensor Frozen Stopped |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | Freezing rain froze the wind speed sensor to a stopped condition. |
Measurements: | sgp15ebbrE2.b1:
sgp30ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/22/2006 | 0300 | 02/25/2006 | 1630 |
Subject: | SGP/EBBR/E20 - Suspect Measurements |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | Some measurements may be incorrect because of low supply voltage. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/21/2006 | 1130 | 01/21/2006 | 1700 |
01/22/2006 | 0100 | 01/22/2006 | 0730 |
Subject: | SGP/EBBR/E22 - Wind Speed Sensor Frozen Stopped |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | The wind speed sensor was frozen to a stopped condition. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/21/2006 | 0400 | 02/25/2006 | 1700 |
Subject: | SGP/EBBR/E26 - Measurements Incorrect |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | Most measurements were incorrect during low battery voltage conditions. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/18/2006 | 0330 | 03/01/2006 | 1615 |
Subject: | SGP/EBBR/E27 - Flux and Gradient Measurements Incorrect |
DataStreams: | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
Description: | The sensible and latent heat flux and gradient measurements were incorrect while the AEM was hung in one position. |
Measurements: | sgp5ebbrE27.b1:
sgp15ebbrE27.b1:
sgp30ebbrE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/21/2006 | 1000 | 03/21/2006 | 2030 |
Subject: | SGP/EBBR/E7 - Fluxes and Gradient Measurements Incorrect |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | The left AEM aspirator was not working during this period, resulting in incorrect gradient and flux measurements. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/07/2005 | 0000 | 01/24/2006 | 2330 |
Subject: | SGP/EBBR/E8 - Flux and Gradient Measurements Suspect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The flux and gradient measurements were often incorrect when the AEM did not function properly. The user needs to inspect the data (particularly the home signals) to determine when the AEM was and was not working. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/21/2006 | 0230 | 03/21/2006 | 1600 |
Subject: | SGP/EBBR/E8 - Wind Speed Sensor Frozen Stopped |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | Freezing rain froze the wind speed sensor to a stopped condition. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/22/2006 | 2030 | 04/04/2006 | 1500 |
Subject: | SGP/EBBR/E8 - Fluxes Incorrect; Left Humidity Near Zero |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The left humidity was intermittently incorrect, therefore the sensible and latent heat fluxes were also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/22/2006 | 1030 | 05/31/2006 | 0030 |
Subject: | SGP/EBBR/E8 - Sensible and Latent Heat Flux Suspect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The left humidity sensor produced large spikes frequently. This resulted in incorrect measurements of vapor pressure gradient and thus calculations of sensible and latent heat flux. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/09/2006 | 1900 | 04/16/2006 | 1600 |
Subject: | SGP/EBBR/E27 - Instrument problem, Loose connections |
DataStreams: | sgp15ebbrE27.b1, sgp30ebbrE27.b1 |
Description: | The soil heat flow #1 sensor went offscale low on occasion. This sometimes caused sensible and latent heat fluxes to be 9999ed out by the EBBR programming, but not always. |
Measurements: | sgp15ebbrE27.b1:
sgp30ebbrE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/30/2006 | 2230 | 05/02/2006 | 1550 |
Subject: | SGP/EBBR/E9 - Wind Speed Incorrect |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Description: | The wind speed measurement was incorrect after the cup anemometer lost a cup. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/23/2006 | 2300 | 04/23/2006 | 2300 |
06/03/2006 | 1900 | 06/04/2006 | 2330 |
Subject: | SGP/EBBR/E13 - Left Air Temperature Incorrect - Fluxes Affected |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | The left air temperature was incorrect, thus the Bowen ratio and sensible and latent heat fluxes were incorrect also. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/30/2006 | 2100 | 05/01/2006 | 1530 |
06/05/2006 | 2330 | 06/06/2006 | 1600 |
06/14/2006 | 2030 | 06/15/2006 | 1430 |
Subject: | SGP/EBBR/E13 - Fluxes Incorrect, AEM Problems |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | The AEM did not function properly, thus sensible and latent heat fluxes were incorrect. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/17/2006 | 0900 | 06/20/2006 | 1915 |
Subject: | SGP/EBBR/E20 - Net Radiation, Fluxes Too Large |
DataStreams: | sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | Net Radiation was way too large because of water in the bottom dome. This resulted in the sensible and latent heat fluxes being too large also. |
Measurements: | sgp5ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/11/2006 | 0130 | 07/21/2006 | 1730 |
Subject: | SGP/EBBR/E27 - Field fire burned AEM |
DataStreams: | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
Description: | A fire in the E27 field burned the AEM, making it non-functional. Sensible and latent heat fluxes and gradients were incorrect during this period. |
Measurements: | sgp5ebbrE27.b1:
sgp15ebbrE27.b1:
sgp30ebbrE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/04/2006 | 1930 | 05/04/2006 | 2200 |
05/05/2006 | 0230 | 05/05/2006 | 0800 |
05/06/2006 | 0430 | 06/20/2006 | 1730 |
Subject: | SGP/EBBR/E27 - Sensible, Latent, and Average Soil Heat Fluxes Sometimes Incorrect |
DataStreams: | sgp15ebbrE27.b1, sgp30ebbrE27.b1 |
Description: | The Soil Heat Flow #1 sensor kept going offscale. Sensible and Latent heat fluxes and the ave_shf were affected and sometimes incorrect. |
Measurements: | sgp15ebbrE27.b1:
sgp30ebbrE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/31/2006 | 0600 | 06/07/2006 | 1730 |
06/14/2006 | 0500 | 06/20/2006 | 1730 |
Subject: | SGP/EBBR/E27 - Low Battery Condition |
DataStreams: | sgp30ebbrE27.b1, sgp5ebbrE27.b1, sgp15ebbrE27.b1 |
Description: | A low battery condition caused most of the measurements to be incorrect. |
Measurements: | sgp5ebbrE27.b1:
sgp15ebbrE27.b1:
sgp30ebbrE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/01/2006 | 1700 | 07/06/2006 | 2200 |
Subject: | SGP/EBBR/E13 - Sensible and Latent Heat Fluxes Incorrect When Pressure Offscale |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | There were several half hours during this period when the pressure went offscale or at least to a high value, causing the sensible and latent heat fluxes to be incorrect. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/26/2006 | 1830 | 08/17/2006 | 1730 |
Subject: | SGP/EBBR/E27 - Most Measurements Suspect |
DataStreams: | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
Description: | The wrong EBBR program was used at this site. Therefore, most of the measurements are suspect. Note: Mentor reports reprocessing will not correct this problem. |
Measurements: | sgp5ebbrE27.b1:
sgp15ebbrE27.b1:
sgp30ebbrE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/11/2006 | 1300 | 08/21/2006 | 2330 |
Subject: | SGP/EBBR/E27 - Battery Voltage Problem - Measurements Incorrect |
DataStreams: | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
Description: | Battery voltage was too low much of the time during the period, thus most measurements were incorrect during much of the time period. |
Measurements: | sgp5ebbrE27.b1:
sgp15ebbrE27.b1:
sgp30ebbrE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/31/2006 | 0430 | 06/02/2006 | 2130 |
Subject: | SGP/EBBR/E18 - Sensible and Latent Heat Fluxes and Net Radiation Incorrect |
DataStreams: | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
Description: | The net radiometer dome was cracked and water had gotten in, resulting in incorrect net radiation measurements and therefore incorrect sensible and latent heat flux measurements. |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/23/2006 | 0130 | 06/27/2006 | 1730 |
Subject: | SGP/EBBR/E8 - Sensible and Latent Heat Fluxes and Net Radiation Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The net radiometer top dome was broken and water had gotten into it. Sensible and latent heat fluxes and net radiation were therefore incorrect. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/21/2006 | 1200 | 08/29/2006 | 1630 |
Subject: | SGP/EBBR/E12 - Net Radiation, and Sensible and Latent Heat Fluxes Incorrect |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | Rain entered into the net radiometer through a broken top dome. Most of the net radiation values during the period are much too large, thus sensible and latent fluxes are also too large. |
Measurements: | sgp30ebbrE12.b1:
sgp15ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/01/2006 | 0130 | 09/12/2006 | 1630 |
Subject: | SGP/EBBR/E12 - Soil Temperature #1 Incorrect |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1 |
Description: | During much of this period soil temperature #1 was offscale or erratic. Sensible and latent heat fluxes were not ususally affected. |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/05/2006 | 1730 | 08/10/2006 | 1730 |
08/23/2006 | 1700 | 08/25/2006 | 1700 |
08/31/2006 | 2030 | 09/01/2006 | 1900 |
Subject: | SGP/EBBR/E13 - Soil Heat Flow #3 Large Negative |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
Description: | During this period Soil Heat Flow #3 was sometimes incorrect (large negative); during those times, sensible and latent heat fluxes and average soil heat flux were sometimes (but not always incorrect). |
Measurements: | sgp15ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/22/2006 | 1630 | 08/22/2006 | 1830 |
08/24/2006 | 1530 | 08/24/2006 | 2000 |
08/29/2006 | 0830 | 08/29/2006 | 1200 |
Subject: | SGP/EBBR/E13 - Pressure and Sensible and Latent Heat Fluxes Incorrect |
DataStreams: | sgp30ebbrE13.b1, sgp5ebbrE13.b1, sgp15ebbrE13.b1 |
Description: | During spikes in pressure upwards or downwards the sensible and latent heat fluxes were incorrect. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/28/2006 | 0030 | 08/28/2006 | 0730 |
Subject: | SGP/EBBR/E13 - Most Measurements Incorrect When Battery Voltage Too low |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | Most measurements were incorrect during this period because the battery voltage was below 10.5 volts. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/21/2006 | 0630 | 08/30/2006 | 1800 |
Subject: | SGP/EBBR/E22 - Net Radiation and Sensible and Latent Heat Fluxes Incorrect |
DataStreams: | sgp30ebbrE22.b1, sgp5ebbrE22.b1, sgp15ebbrE22.b1 |
Description: | Rain entered the net radiometer, causing net radiation, and thus sensible and latent heat fluxes, to be too large. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/30/2006 | 2200 | 11/07/2006 | 2130 |
Subject: | SGP/EBBR/E18 - Sensible and Latent Heat Fluxes Incorrect |
DataStreams: | sgp30ebbrE18.b1 |
Description: | Sensible and latent heat fluxes were incorrect during this period because the AEM was not exchanging. |
Measurements: | sgp30ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/16/2006 | 0000 | 11/08/2006 | 1940 |
Subject: | SGP/EBBR/E22 - Soil Heat Flow #4 Incorrect - Sensible and Latent Heat Fluxes Sometimes |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1 |
Description: | Soil heat flow sensor #4 dropped to low values or offscale. Sensible and latent heat fluxes were sometimes affected. |
Measurements: | sgp30ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/17/2006 | 0700 | 10/23/2006 | 1615 |
Subject: | SGP/EBBR/E13 - Battery failure |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | Data was missing entirely on parts of 10/22 and 10/23. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/09/2006 | 1500 | 10/12/2006 | 1630 |
Subject: | SGP/EBBR/E19 - Sensible and Latent Heat Fluxes Incorrect |
DataStreams: | sgp30ebbrE19.b1 |
Description: | The sensible and latent heat fluxes were incorrect when the top and bottom air temperatures were incorrect. |
Measurements: | sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/22/1997 | 2100 | 10/12/2003 | 2330 |
Subject: | SGP/EBBR/E2 - Wind Direction Dependency of Surface Fluxes and General Data Quality |
DataStreams: | sgp30ebbrE2.b1 |
Description: | Beginning in FY2006, Data Quality Reports are not written for missing data or for situations when qc flags clearly show that the data is incorrect (this is true for most of the conditions listed below). DQRs are written for periods when data is incorrect, when the situation is not represented by qc flags in the data, and it's not obvious that the data should have been flagged as incorrect. The EBBR data is only useful for particular wind directions at each Extended Facility. Please use the following resources to help in interpreting the EBBR data. 1) qc flags in the EBBR data. 2) Instrument Mentor Monthly Summaries (IMMS) at http://www.db.arm.gov/IMMS/ 3) SGP EF Surface Conditions Observations (EFSCO) at http://198.124.96.210:591/sfc_cond1/default.htm 4) Surface fluxes are valid only for the following wind directions (in degrees): 71-137, 223-289 Note that there are wind directions at most sites for which the fetch is insufficient and therefore EBBR data is invalid. Appropriate fetch was determined from a 1/40 measurement height to fetch ratio, resulting in a required minimum fetch of 120 m. Particular attention should be paid to the limited valid wind directions for the Central Facility EBBR. The surface vegetation type for the EBBRs is grazed or ungrazed grassland (refer to the EFSCO for state). 5) Some conditions that commonly cause the EBBR primary measurements to be incorrect or missing follow: a) Sensible and latent heat fluxes are not accurate during times when the Automatic Exchange Mechanism (AEM) is not functioning properly. The AEM switches the gradient measuring instrumentation between the top and bottom positions every 15 minutes; this reduces the effects of instrument offsets. Sometimes the AEM does not reach it's full extent of travel, resulting in the home signal being zero. b) Sensible and latent heat fluxes are sometimes incorrect when surface soil heat flow is out of range, as seen in the average soil heat flux (ave_shf). c) Very light winds may be seen at nights for brief periods. Wind direction flops around a lot during low wind conditions (< 1.5 m/s) and is often unreliable during those periods. d) Missing data periods occur at times; this is usually a site data system collection/communication problem; it can also occur during replacement of the system with a newly calibrated unit. Missing data is sometimes filled in later from manual or automatic re-collection of the data. |
Measurements: | sgp30ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/10/1997 | 2000 | 10/12/2003 | 2330 |
Subject: | SGP/EBBR/E18 - Wind Direction Dependency of Surface Fluxes and General Data Quality |
DataStreams: | sgp30ebbrE18.b1 |
Description: | Beginning in FY2006, Data Quality Reports are not written for missing data or for situations when qc flags clearly show that the data is incorrect (this is true for most of the conditions listed below). DQRs are written for periods when data is incorrect, when the situation is not represented by qc flags in the data, and it's not obvious that the data should have been flagged as incorrect. The EBBR data is only useful for particular wind directions at each Extended Facility. Please use the following resources to help in interpreting the EBBR data. 1) qc flags in the EBBR data. 2) Instrument Mentor Monthly Summaries (IMMS) at http://www.db.arm.gov/IMMS/ 3) SGP EF Surface Conditions Observations (EFSCO) at http://198.124.96.210:591/sfc_cond1/default.htm 4) Surface fluxes are valid only for the following wind directions (in degrees): 138-325 Note that there are wind directions at most sites for which the fetch is insufficient and therefore EBBR data is invalid. Appropriate fetch was determined from a 1/40 measurement height to fetch ratio, resulting in a required minimum fetch of 120 m. Particular attention should be paid to the limited valid wind directions for the Central Facility EBBR. The surface vegetation type for the EBBRs is grazed or ungrazed grassland (refer to the EFSCO for state). 5) Some conditions that commonly cause the EBBR primary measurements to be incorrect or missing follow: a) Sensible and latent heat fluxes are not accurate during times when the Automatic Exchange Mechanism (AEM) is not functioning properly. The AEM switches the gradient measuring instrumentation between the top and bottom positions every 15 minutes; this reduces the effects of instrument offsets. Sometimes the AEM does not reach it's full extent of travel, resulting in the home signal being zero. b) Sensible and latent heat fluxes are sometimes incorrect when surface soil heat flow is out of range, as seen in the average soil heat flux (ave_shf). c) Very light winds may be seen at nights for brief periods. Wind direction flops around a lot during low wind conditions (< 1.5 m/s) and is often unreliable during those periods. d) Missing data periods occur at times; this is usually a site data system collection/communication problem; it can also occur during replacement of the system with a newly calibrated unit. Missing data is sometimes filled in later from manual or automatic re-collection of the data. |
Measurements: | sgp30ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/30/1997 | 1000 | 10/12/2003 | 2330 |
Subject: | SGP/EBBR/E19 - Wind Direction Dependency of Surface Fluxes and General Data Quality |
DataStreams: | sgp30ebbrE19.b1 |
Description: | Beginning in FY2006, Data Quality Reports are not written for missing data or for situations when qc flags clearly show that the data is incorrect (this is true for most of the conditions listed below). DQRs are written for periods when data is incorrect, when the situation is not represented by qc flags in the data, and it's not obvious that the data should have been flagged as incorrect. The EBBR data is only useful for particular wind directions at each Extended Facility. Please use the following resources to help in interpreting the EBBR data. 1) qc flags in the EBBR data. 2) Instrument Mentor Monthly Summaries (IMMS) at http://www.db.arm.gov/IMMS/ 3) SGP EF Surface Conditions Observations (EFSCO) at http://198.124.96.210:591/sfc_cond1/default.htm 4) Surface fluxes are valid only for the following wind directions (in degrees): 0-133, 151-360 Note that there are wind directions at most sites for which the fetch is insufficient and therefore EBBR data is invalid. Appropriate fetch was determined from a 1/40 measurement height to fetch ratio, resulting in a required minimum fetch of 120 m. Particular attention should be paid to the limited valid wind directions for the Central Facility EBBR. The surface vegetation type for the EBBRs is grazed or ungrazed grassland (refer to the EFSCO for state). 5) Some conditions that commonly cause the EBBR primary measurements to be incorrect or missing follow: a) Sensible and latent heat fluxes are not accurate during times when the Automatic Exchange Mechanism (AEM) is not functioning properly. The AEM switches the gradient measuring instrumentation between the top and bottom positions every 15 minutes; this reduces the effects of instrument offsets. Sometimes the AEM does not reach it's full extent of travel, resulting in the home signal being zero. b) Sensible and latent heat fluxes are sometimes incorrect when surface soil heat flow is out of range, as seen in the average soil heat flux (ave_shf). c) Very light winds may be seen at nights for brief periods. Wind direction flops around a lot during low wind conditions (< 1.5 m/s) and is often unreliable during those periods. d) Missing data periods occur at times; this is usually a site data system collection/communication problem; it can also occur during replacement of the system with a newly calibrated unit. Missing data is sometimes filled in later from manual or automatic re-collection of the data. |
Measurements: | sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/06/1993 | 0630 | 10/04/2003 | 2330 |
Subject: | SGP/EBBR/E20 - Wind Direction Dependency of Surface Fluxes and General Data Quality |
DataStreams: | sgp30ebbrE20.b1 |
Description: | Beginning in FY2006, Data Quality Reports are not written for missing data or for situations when qc flags clearly show that the data is incorrect (this is true for most of the conditions listed below). DQRs are written for periods when data is incorrect, when the situation is not represented by qc flags in the data, and it's not obvious that the data should have been flagged as incorrect. The EBBR data is only useful for particular wind directions at each Extended Facility. Please use the following resources to help in interpreting the EBBR data. 1) qc flags in the EBBR data. 2) Instrument Mentor Monthly Summaries (IMMS) at http://www.db.arm.gov/IMMS/ 3) SGP EF Surface Conditions Observations (EFSCO) at http://198.124.96.210:591/sfc_cond1/default.htm 4) Surface fluxes are valid only for the following wind directions (in degrees): 0-230, 310-360 Note that there are wind directions at most sites for which the fetch is insufficient and therefore EBBR data is invalid. Appropriate fetch was determined from a 1/40 measurement height to fetch ratio, resulting in a required minimum fetch of 120 m. Particular attention should be paid to the limited valid wind directions for the Central Facility EBBR. The surface vegetation type for the EBBRs is grazed or ungrazed grassland (refer to the EFSCO for state). 5) Some conditions that commonly cause the EBBR primary measurements to be incorrect or missing follow: a) Sensible and latent heat fluxes are not accurate during times when the Automatic Exchange Mechanism (AEM) is not functioning properly. The AEM switches the gradient measuring instrumentation between the top and bottom positions every 15 minutes; this reduces the effects of instrument offsets. Sometimes the AEM does not reach it's full extent of travel, resulting in the home signal being zero. b) Sensible and latent heat fluxes are sometimes incorrect when surface soil heat flow is out of range, as seen in the average soil heat flux (ave_shf). c) Very light winds may be seen at nights for brief periods. Wind direction flops around a lot during low wind conditions (< 1.5 m/s) and is often unreliable during those periods. d) Missing data periods occur at times; this is usually a site data system collection/communication problem; it can also occur during replacement of the system with a newly calibrated unit. Missing data is sometimes filled in later from manual or automatic re-collection of the data. |
Measurements: | sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/04/1993 | 1200 | 10/12/2003 | 2330 |
Subject: | SGP/EBBR/E22 - Wind Direction Dependency of Surface Fluxes and General Data Quality |
DataStreams: | sgp30ebbrE22.b1 |
Description: | Beginning in FY2006, Data Quality Reports are not written for missing data or for situations when qc flags clearly show that the data is incorrect (this is true for most of the conditions listed below). DQRs are written for periods when data is incorrect, when the situation is not represented by qc flags in the data, and it's not obvious that the data should have been flagged as incorrect. The EBBR data is only useful for particular wind directions at each Extended Facility. Please use the following resources to help in interpreting the EBBR data. 1) qc flags in the EBBR data. 2) Instrument Mentor Monthly Summaries (IMMS) at http://www.db.arm.gov/IMMS/ 3) SGP EF Surface Conditions Observations (EFSCO) at http://198.124.96.210:591/sfc_cond1/default.htm 4) Surface fluxes are valid only for the following wind directions (in degrees): 0-49, 139-360 Note that there are wind directions at most sites for which the fetch is insufficient and therefore EBBR data is invalid. Appropriate fetch was determined from a 1/40 measurement height to fetch ratio, resulting in a required minimum fetch of 120 m. Particular attention should be paid to the limited valid wind directions for the Central Facility EBBR. The surface vegetation type for the EBBRs is grazed or ungrazed grassland (refer to the EFSCO for state). 5) Some conditions that commonly cause the EBBR primary measurements to be incorrect or missing follow: a) Sensible and latent heat fluxes are not accurate during times when the Automatic Exchange Mechanism (AEM) is not functioning properly. The AEM switches the gradient measuring instrumentation between the top and bottom positions every 15 minutes; this reduces the effects of instrument offsets. Sometimes the AEM does not reach it's full extent of travel, resulting in the home signal being zero. b) Sensible and latent heat fluxes are sometimes incorrect when surface soil heat flow is out of range, as seen in the average soil heat flux (ave_shf). c) Very light winds may be seen at nights for brief periods. Wind direction flops around a lot during low wind conditions (< 1.5 m/s) and is often unreliable during those periods. d) Missing data periods occur at times; this is usually a site data system collection/communication problem; it can also occur during replacement of the system with a newly calibrated unit. Missing data is sometimes filled in later from manual or automatic re-collection of the data. |
Measurements: | sgp30ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/22/1997 | 2030 | 04/08/2002 | 2330 |
Subject: | SGP/EBBR/E25 - Wind Direction Dependency of Surface Fluxes and General Data Quality |
DataStreams: | sgp30ebbrE25.b1 |
Description: | Beginning in FY2006, Data Quality Reports are not written for missing data or for situations when qc flags clearly show that the data is incorrect (this is true for most of the conditions listed below). DQRs are written for periods when data is incorrect, when the situation is not represented by qc flags in the data, and it's not obvious that the data should have been flagged as incorrect. The EBBR data is only useful for particular wind directions at each Extended Facility. Please use the following resources to help in interpreting the EBBR data. 1) qc flags in the EBBR data. 2) Instrument Mentor Monthly Summaries (IMMS) at http://www.db.arm.gov/IMMS/ 3) SGP EF Surface Conditions Observations (EFSCO) at http://198.124.96.210:591/sfc_cond1/default.htm 4) Surface fluxes are valid only for the following wind directions (in degrees): 30-300 Note that there are wind directions at most sites for which the fetch is insufficient and therefore EBBR data is invalid. Appropriate fetch was determined from a 1/40 measurement height to fetch ratio, resulting in a required minimum fetch of 120 m. Particular attention should be paid to the limited valid wind directions for the Central Facility EBBR. The surface vegetation type for the EBBRs is grazed or ungrazed grassland (refer to the EFSCO for state). 5) Some conditions that commonly cause the EBBR primary measurements to be incorrect or missing follow: a) Sensible and latent heat fluxes are not accurate during times when the Automatic Exchange Mechanism (AEM) is not functioning properly. The AEM switches the gradient measuring instrumentation between the top and bottom positions every 15 minutes; this reduces the effects of instrument offsets. Sometimes the AEM does not reach it's full extent of travel, resulting in the home signal being zero. b) Sensible and latent heat fluxes are sometimes incorrect when surface soil heat flow is out of range, as seen in the average soil heat flux (ave_shf). c) Very light winds may be seen at nights for brief periods. Wind direction flops around a lot during low wind conditions (< 1.5 m/s) and is often unreliable during those periods. d) Missing data periods occur at times; this is usually a site data system collection/communication problem; it can also occur during replacement of the system with a newly calibrated unit. Missing data is sometimes filled in later from manual or automatic re-collection of the data. |
Measurements: | sgp30ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/05/1993 | 0000 | 10/12/2003 | 2330 |
Subject: | SGP/EBBR/E26 - Wind Direction Dependency of Surface Fluxes and General Data Quality |
DataStreams: | sgp30ebbrE26.b1 |
Description: | Beginning in FY2006, Data Quality Reports are not written for missing data or for situations when qc flags clearly show that the data is incorrect (this is true for most of the conditions listed below). DQRs are written for periods when data is incorrect, when the situation is not represented by qc flags in the data, and it's not obvious that the data should have been flagged as incorrect. The EBBR data is only useful for particular wind directions at each Extended Facility. Please use the following resources to help in interpreting the EBBR data. 1) qc flags in the EBBR data. 2) Instrument Mentor Monthly Summaries (IMMS) at http://www.db.arm.gov/IMMS/ 3) SGP EF Surface Conditions Observations (EFSCO) at http://198.124.96.210:591/sfc_cond1/default.htm 4) Surface fluxes are valid only for the following wind directions (in degrees): 0-33, 243-360 Note that there are wind directions at most sites for which the fetch is insufficient and therefore EBBR data is invalid. Appropriate fetch was determined from a 1/40 measurement height to fetch ratio, resulting in a required minimum fetch of 120 m. Particular attention should be paid to the limited valid wind directions for the Central Facility EBBR. The surface vegetation type for the EBBRs is grazed or ungrazed grassland (refer to the EFSCO for state). 5) Some conditions that commonly cause the EBBR primary measurements to be incorrect or missing follow: a) Sensible and latent heat fluxes are not accurate during times when the Automatic Exchange Mechanism (AEM) is not functioning properly. The AEM switches the gradient measuring instrumentation between the top and bottom positions every 15 minutes; this reduces the effects of instrument offsets. Sometimes the AEM does not reach it's full extent of travel, resulting in the home signal being zero. b) Sensible and latent heat fluxes are sometimes incorrect when surface soil heat flow is out of range, as seen in the average soil heat flux (ave_shf). c) Very light winds may be seen at nights for brief periods. Wind direction flops around a lot during low wind conditions (< 1.5 m/s) and is often unreliable during those periods. d) Missing data periods occur at times; this is usually a site data system collection/communication problem; it can also occur during replacement of the system with a newly calibrated unit. Missing data is sometimes filled in later from manual or automatic re-collection of the data. |
Measurements: | sgp30ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/07/2003 | 1030 | 10/12/2003 | 2330 |
Subject: | SGP/EBBR/E27 - Wind Direction Dependency of Surface Fluxes and General Data Quality |
DataStreams: | sgp30ebbrE27.b1 |
Description: | Beginning in FY2006, Data Quality Reports are not written for missing data or for situations when qc flags clearly show that the data is incorrect (this is true for most of the conditions listed below). DQRs are written for periods when data is incorrect, when the situation is not represented by qc flags in the data, and it's not obvious that the data should have been flagged as incorrect. The EBBR data is only useful for particular wind directions at each Extended Facility. Please use the following resources to help in interpreting the EBBR data. 1) qc flags in the EBBR data. 2) Instrument Mentor Monthly Summaries (IMMS) at http://www.db.arm.gov/IMMS/ 3) SGP EF Surface Conditions Observations (EFSCO) at http://198.124.96.210:591/sfc_cond1/default.htm 4) Surface fluxes are valid only for the following wind directions (in degrees): 20-156 Note that there are wind directions at most sites for which the fetch is insufficient and therefore EBBR data is invalid. Appropriate fetch was determined from a 1/40 measurement height to fetch ratio, resulting in a required minimum fetch of 120 m. Particular attention should be paid to the limited valid wind directions for the Central Facility EBBR. The surface vegetation type for the EBBRs is grazed or ungrazed grassland (refer to the EFSCO for state). 5) Some conditions that commonly cause the EBBR primary measurements to be incorrect or missing follow: a) Sensible and latent heat fluxes are not accurate during times when the Automatic Exchange Mechanism (AEM) is not functioning properly. The AEM switches the gradient measuring instrumentation between the top and bottom positions every 15 minutes; this reduces the effects of instrument offsets. Sometimes the AEM does not reach it's full extent of travel, resulting in the home signal being zero. b) Sensible and latent heat fluxes are sometimes incorrect when surface soil heat flow is out of range, as seen in the average soil heat flux (ave_shf). c) Very light winds may be seen at nights for brief periods. Wind direction flops around a lot during low wind conditions (< 1.5 m/s) and is often unreliable during those periods. d) Missing data periods occur at times; this is usually a site data system collection/communication problem; it can also occur during replacement of the system with a newly calibrated unit. Missing data is sometimes filled in later from manual or automatic re-collection of the data. |
Measurements: | sgp30ebbrE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/01/2001 | 0000 | 04/08/2002 | 2330 |
Subject: | SGP/EBBR/E25 - Wind Direction Dependency of Surface Fluxes and General Data Quality |
DataStreams: | sgp30ebbrE25.b1 |
Description: | Beginning in FY2006, Data Quality Reports are not written for missing data or for situations when qc flags clearly show that the data is incorrect (this is true for most of the conditions listed below). DQRs are written for periods when data is incorrect, when the situation is not represented by qc flags in the data, and it's not obvious that the data should have been flagged as incorrect. The EBBR data is only useful for particular wind directions at each Extended Facility. Please use the following resources to help in interpreting the EBBR data. 1) qc flags in the EBBR data. 2) Instrument Mentor Monthly Summaries (IMMS) at http://www.db.arm.gov/IMMS/ 3) SGP EF Surface Conditions Observations (EFSCO) at http://198.124.96.210:591/sfc_cond1/default.htm 4) Surface fluxes are valid only for the following wind directions (in degrees): 30-300 Note that there are wind directions at most sites for which the fetch is insufficient and therefore EBBR data is invalid. Appropriate fetch was determined from a 1/40 measurement height to fetch ratio, resulting in a required minimum fetch of 120 m. Particular attention should be paid to the limited valid wind directions for the Central Facility EBBR. The surface vegetation type for the EBBRs is grazed or ungrazed grassland (refer to the EFSCO for state). 5) Some conditions that commonly cause the EBBR primary measurements to be incorrect or missing follow: a) Sensible and latent heat fluxes are not accurate during times when the Automatic Exchange Mechanism (AEM) is not functioning properly. The AEM switches the gradient measuring instrumentation between the top and bottom positions every 15 minutes; this reduces the effects of instrument offsets. Sometimes the AEM does not reach it's full extent of travel, resulting in the home signal being zero. b) Sensible and latent heat fluxes are sometimes incorrect when surface soil heat flow is out of range, as seen in the average soil heat flux (ave_shf). c) Very light winds may be seen at nights for brief periods. Wind direction flops around a lot during low wind conditions (< 1.5 m/s) and is often unreliable during those periods. d) Missing data periods occur at times; this is usually a site data system collection/communication problem; it can also occur during replacement of the system with a newly calibrated unit. Missing data is sometimes filled in later from manual or automatic re-collection of the data. |
Measurements: | sgp30ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/17/1993 | 1530 | 10/12/2003 | 2330 |
Subject: | SGP/EBBR/E7 - Wind Direction Dependency of Surface Fluxes and General Data Quality |
DataStreams: | sgp30ebbrE7.b1 |
Description: | Beginning in FY2006, Data Quality Reports are not written for missing data or for situations when qc flags clearly show that the data is incorrect (this is true for most of the conditions listed below). DQRs are written for periods when data is incorrect, when the situation is not represented by qc flags in the data, and it's not obvious that the data should have been flagged as incorrect. The EBBR data is only useful for particular wind directions at each Extended Facility. Please use the following resources to help in interpreting the EBBR data. 1) qc flags in the EBBR data. 2) Instrument Mentor Monthly Summaries (IMMS) at http://www.db.arm.gov/IMMS/ 3) SGP EF Surface Conditions Observations (EFSCO) at http://198.124.96.210:591/sfc_cond1/default.htm 4) Surface fluxes are valid only for the following wind directions (in degrees): 0-244, 296-360 Note that there are wind directions at most sites for which the fetch is insufficient and therefore EBBR data is invalid. Appropriate fetch was determined from a 1/40 measurement height to fetch ratio, resulting in a required minimum fetch of 120 m. Particular attention should be paid to the limited valid wind directions for the Central Facility EBBR. The surface vegetation type for the EBBRs is grazed or ungrazed grassland (refer to the EFSCO for state). 5) Some conditions that commonly cause the EBBR primary measurements to be incorrect or missing follow: a) Sensible and latent heat fluxes are not accurate during times when the Automatic Exchange Mechanism (AEM) is not functioning properly. The AEM switches the gradient measuring instrumentation between the top and bottom positions every 15 minutes; this reduces the effects of instrument offsets. Sometimes the AEM does not reach it's full extent of travel, resulting in the home signal being zero. b) Sensible and latent heat fluxes are sometimes incorrect when surface soil heat flow is out of range, as seen in the average soil heat flux (ave_shf). c) Very light winds may be seen at nights for brief periods. Wind direction flops around a lot during low wind conditions (< 1.5 m/s) and is often unreliable during those periods. d) Missing data periods occur at times; this is usually a site data system collection/communication problem; it can also occur during replacement of the system with a newly calibrated unit. Missing data is sometimes filled in later from manual or automatic re-collection of the data. |
Measurements: | sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/12/1993 | 1630 | 10/12/2003 | 2330 |
Subject: | SGP/EBBR/E8 - Wind Direction Dependency of Surface Fluxes and General Data Quality |
DataStreams: | sgp30ebbrE8.b1 |
Description: | Beginning in FY2006, Data Quality Reports are not written for missing data or for situations when qc flags clearly show that the data is incorrect (this is true for most of the conditions listed below). DQRs are written for periods when data is incorrect, when the situation is not represented by qc flags in the data, and it's not obvious that the data should have been flagged as incorrect. The EBBR data is only useful for particular wind directions at each Extended Facility. Please use the following resources to help in interpreting the EBBR data. 1) qc flags in the EBBR data. 2) Instrument Mentor Monthly Summaries (IMMS) at http://www.db.arm.gov/IMMS/ 3) SGP EF Surface Conditions Observations (EFSCO) at http://198.124.96.210:591/sfc_cond1/default.htm 4) Surface fluxes are valid only for the following wind directions (in degrees): 0-224, 314-360 Note that there are wind directions at most sites for which the fetch is insufficient and therefore EBBR data is invalid. Appropriate fetch was determined from a 1/40 measurement height to fetch ratio, resulting in a required minimum fetch of 120 m. Particular attention should be paid to the limited valid wind directions for the Central Facility EBBR. The surface vegetation type for the EBBRs is grazed or ungrazed grassland (refer to the EFSCO for state). 5) Some conditions that commonly cause the EBBR primary measurements to be incorrect or missing follow: a) Sensible and latent heat fluxes are not accurate during times when the Automatic Exchange Mechanism (AEM) is not functioning properly. The AEM switches the gradient measuring instrumentation between the top and bottom positions every 15 minutes; this reduces the effects of instrument offsets. Sometimes the AEM does not reach it's full extent of travel, resulting in the home signal being zero. b) Sensible and latent heat fluxes are sometimes incorrect when surface soil heat flow is out of range, as seen in the average soil heat flux (ave_shf). c) Very light winds may be seen at nights for brief periods. Wind direction flops around a lot during low wind conditions (< 1.5 m/s) and is often unreliable during those periods. d) Missing data periods occur at times; this is usually a site data system collection/communication problem; it can also occur during replacement of the system with a newly calibrated unit. Missing data is sometimes filled in later from manual or automatic re-collection of the data. |
Measurements: | sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/11/1993 | 1500 | 10/12/2003 | 2330 |
Subject: | SGP/EBBR/E9 - Wind Direction Dependency of Surface Fluxes and General Data Quality |
DataStreams: | sgp30ebbrE9.b1 |
Description: | Beginning in FY2006, Data Quality Reports are not written for missing data or for situations when qc flags clearly show that the data is incorrect (this is true for most of the conditions listed below). DQRs are written for periods when data is incorrect, when the situation is not represented by qc flags in the data, and it's not obvious that the data should have been flagged as incorrect. The EBBR data is only useful for particular wind directions at each Extended Facility. Please use the following resources to help in interpreting the EBBR data. 1) qc flags in the EBBR data. 2) Instrument Mentor Monthly Summaries (IMMS) at http://www.db.arm.gov/IMMS/ 3) SGP EF Surface Conditions Observations (EFSCO) at http://198.124.96.210:591/sfc_cond1/default.htm 4) Surface fluxes are valid only for the following wind directions (in degrees): 0-360 Note that there are wind directions at most sites for which the fetch is insufficient and therefore EBBR data is invalid. Appropriate fetch was determined from a 1/40 measurement height to fetch ratio, resulting in a required minimum fetch of 120 m. Particular attention should be paid to the limited valid wind directions for the Central Facility EBBR. The surface vegetation type for the EBBRs is grazed or ungrazed grassland (refer to the EFSCO for state). 5) Some conditions that commonly cause the EBBR primary measurements to be incorrect or missing follow: a) Sensible and latent heat fluxes are not accurate during times when the Automatic Exchange Mechanism (AEM) is not functioning properly. The AEM switches the gradient measuring instrumentation between the top and bottom positions every 15 minutes; this reduces the effects of instrument offsets. Sometimes the AEM does not reach it's full extent of travel, resulting in the home signal being zero. b) Sensible and latent heat fluxes are sometimes incorrect when surface soil heat flow is out of range, as seen in the average soil heat flux (ave_shf). c) Very light winds may be seen at nights for brief periods. Wind direction flops around a lot during low wind conditions (< 1.5 m/s) and is often unreliable during those periods. d) Missing data periods occur at times; this is usually a site data system collection/communication problem; it can also occur during replacement of the system with a newly calibrated unit. Missing data is sometimes filled in later from manual or automatic re-collection of the data. |
Measurements: | sgp30ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/23/1993 | 1500 | 10/12/2003 | 2330 |
Subject: | SGP/EBBR/E12 - Wind Direction Dependency of Surface Fluxes and General Data Quality |
DataStreams: | sgp30ebbrE12.b1 |
Description: | Beginning in FY2006, Data Quality Reports are not written for missing data or for situations when qc flags clearly show that the data is incorrect (this is true for most of the conditions listed below). DQRs are written for periods when data is incorrect, when the situation is not represented by qc flags in the data, and it's not obvious that the data should have been flagged as incorrect. The EBBR data is only useful for particular wind directions at each Extended Facility. Please use the following resources to help in interpreting the EBBR data. 1) qc flags in the EBBR data. 2) Instrument Mentor Monthly Summaries (IMMS) at http://www.db.arm.gov/IMMS/ 3) SGP EF Surface Conditions Observations (EFSCO) at http://198.124.96.210:591/sfc_cond1/default.htm 4) Surface fluxes are valid only for the following wind directions (in degrees): 0-360 Note that there are wind directions at most sites for which the fetch is insufficient and therefore EBBR data is invalid. Appropriate fetch was determined from a 1/40 measurement height to fetch ratio, resulting in a required minimum fetch of 120 m. Particular attention should be paid to the limited valid wind directions for the Central Facility EBBR. The surface vegetation type for the EBBRs is grazed or ungrazed grassland (refer to the EFSCO for state). 5) Some conditions that commonly cause the EBBR primary measurements to be incorrect or missing follow: a) Sensible and latent heat fluxes are not accurate during times when the Automatic Exchange Mechanism (AEM) is not functioning properly. The AEM switches the gradient measuring instrumentation between the top and bottom positions every 15 minutes; this reduces the effects of instrument offsets. Sometimes the AEM does not reach it's full extent of travel, resulting in the home signal being zero. b) Sensible and latent heat fluxes are sometimes incorrect when surface soil heat flow is out of range, as seen in the average soil heat flux (ave_shf). c) Very light winds may be seen at nights for brief periods. Wind direction flops around a lot during low wind conditions (< 1.5 m/s) and is often unreliable during those periods. d) Missing data periods occur at times; this is usually a site data system collection/communication problem; it can also occur during replacement of the system with a newly calibrated unit. Missing data is sometimes filled in later from manual or automatic re-collection of the data. |
Measurements: | sgp30ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/13/1993 | 1200 | 10/12/2003 | 2330 |
Subject: | SGP/EBBR/E13 - Wind Direction Dependency of Surface Fluxes and General Data Quality |
DataStreams: | sgp30ebbrE13.b1 |
Description: | Beginning in FY2006, Data Quality Reports are not written for missing data or for situations when qc flags clearly show that the data is incorrect (this is true for most of the conditions listed below). DQRs are written for periods when data is incorrect, when the situation is not represented by qc flags in the data, and it's not obvious that the data should have been flagged as incorrect. The EBBR data is only useful for particular wind directions at each Extended Facility. Please use the following resources to help in interpreting the EBBR data. 1) qc flags in the EBBR data. 2) Instrument Mentor Monthly Summaries (IMMS) at http://www.db.arm.gov/IMMS/ 3) SGP EF Surface Conditions Observations (EFSCO) at http://198.124.96.210:591/sfc_cond1/default.htm 4) Surface fluxes are valid only for the following wind directions (in degrees): 0-52, 142-194, 328-360 Note that there are wind directions at most sites for which the fetch is insufficient and therefore EBBR data is invalid. Appropriate fetch was determined from a 1/40 measurement height to fetch ratio, resulting in a required minimum fetch of 120 m. Particular attention should be paid to the limited valid wind directions for the Central Facility EBBR. The surface vegetation type for the EBBRs is grazed or ungrazed grassland (refer to the EFSCO for state). 5) Some conditions that commonly cause the EBBR primary measurements to be incorrect or missing follow: a) Sensible and latent heat fluxes are not accurate during times when the Automatic Exchange Mechanism (AEM) is not functioning properly. The AEM switches the gradient measuring instrumentation between the top and bottom positions every 15 minutes; this reduces the effects of instrument offsets. Sometimes the AEM does not reach it's full extent of travel, resulting in the home signal being zero. b) Sensible and latent heat fluxes are sometimes incorrect when surface soil heat flow is out of range, as seen in the average soil heat flux (ave_shf). c) Very light winds may be seen at nights for brief periods. Wind direction flops around a lot during low wind conditions (< 1.5 m/s) and is often unreliable during those periods. d) Missing data periods occur at times; this is usually a site data system collection/communication problem; it can also occur during replacement of the system with a newly calibrated unit. Missing data is sometimes filled in later from manual or automatic re-collection of the data. |
Measurements: | sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/11/1993 | 1100 | 10/12/2003 | 2330 |
Subject: | SGP/EBBR/E15 - Wind Direction Dependency of Surface Fluxes and General Data Quality |
DataStreams: | sgp30ebbrE15.b1 |
Description: | Beginning in FY2006, Data Quality Reports are not written for missing data or for situations when qc flags clearly show that the data is incorrect (this is true for most of the conditions listed below). DQRs are written for periods when data is incorrect, when the situation is not represented by qc flags in the data, and it's not obvious that the data should have been flagged as incorrect. The EBBR data is only useful for particular wind directions at each Extended Facility. Please use the following resources to help in interpreting the EBBR data. 1) qc flags in the EBBR data. 2) Instrument Mentor Monthly Summaries (IMMS) at http://www.db.arm.gov/IMMS/ 3) SGP EF Surface Conditions Observations (EFSCO) at http://198.124.96.210:591/sfc_cond1/default.htm 4) Surface fluxes are valid only for the following wind directions (in degrees): 133-360 Note that there are wind directions at most sites for which the fetch is insufficient and therefore EBBR data is invalid. Appropriate fetch was determined from a 1/40 measurement height to fetch ratio, resulting in a required minimum fetch of 120 m. Particular attention should be paid to the limited valid wind directions for the Central Facility EBBR. The surface vegetation type for the EBBRs is grazed or ungrazed grassland (refer to the EFSCO for state). 5) Some conditions that commonly cause the EBBR primary measurements to be incorrect or missing follow: a) Sensible and latent heat fluxes are not accurate during times when the Automatic Exchange Mechanism (AEM) is not functioning properly. The AEM switches the gradient measuring instrumentation between the top and bottom positions every 15 minutes; this reduces the effects of instrument offsets. Sometimes the AEM does not reach it's full extent of travel, resulting in the home signal being zero. b) Sensible and latent heat fluxes are sometimes incorrect when surface soil heat flow is out of range, as seen in the average soil heat flux (ave_shf). c) Very light winds may be seen at nights for brief periods. Wind direction flops around a lot during low wind conditions (< 1.5 m/s) and is often unreliable during those periods. d) Missing data periods occur at times; this is usually a site data system collection/communication problem; it can also occur during replacement of the system with a newly calibrated unit. Missing data is sometimes filled in later from manual or automatic re-collection of the data. |
Measurements: | sgp30ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/07/2001 | 1505 | 12/11/2001 | 1630 |
01/30/2002 | 1805 | 02/06/2002 | 1225 |
02/08/2002 | 2005 | 02/11/2002 | 1300 |
02/16/2002 | 0405 | 02/18/2002 | 2100 |
02/23/2002 | 0305 | 02/28/2002 | 1300 |
03/09/2002 | 2105 | 03/11/2002 | 2005 |
Subject: | SGP/EBBR/E2 - Missing data |
DataStreams: | sgp15ebbrE2.b1, sgp30baebbrE2.c1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | Data are missing and unrecoverable. |
Measurements: | sgp15ebbrE2.b1:
sgp30ebbrE2.b1:
sgp30baebbrE2.c1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/03/2004 | 2105 | 02/12/2004 | 1555 |
01/23/2007 | 1530 | 01/25/2007 | 1725 |
Subject: | SGP/EBBR/E2 - Missing data |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | Data are missing and unrecoverable |
Measurements: | sgp15ebbrE2.b1:
sgp30ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/03/2007 | 1000 | 02/22/2007 | 1645 |
Subject: | SGP/EBBR/E2 - Windspeed sensor failure |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | The windspeed sensor developed an abnormally high wind speed threshold due to worn bearings. Wind speeds were often measured to be too low. Inspect the data to identify suspect time periods. |
Measurements: | sgp15ebbrE2.b1:
sgp30ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/01/2007 | 0100 | 02/20/2007 | 2100 |
Subject: | SGP/EBBR/E7 - Windspeed threshhold too high |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | The wind speed sensor developed worn bearings and therefore too high a starting threshhold. The data user should compare the EBBR wind speed with the SMOS wind speed to detect periods when the EBBR wind speed threshhold caused poor data quality. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/22/2007 | 1630 | 03/06/2007 | 2100 |
Subject: | SGP/EBBR/E8 - Failed Temp/RH sensor |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The right Temperature-Humidity sensor failed. The measured temperature gradients were incorrect and thus the sensible and latent heat fluxes were incorrect for this extended period. The sensor was replaced March 6. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/16/2007 | 0400 | 04/26/2007 | 1630 |
Subject: | SGP/EBBR/E22 - Wind Direction Incorrect |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | Wind direction was flat-lining at 135 degrees and occasionally changed to 0.4 degrees. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/10/2007 | 1500 | 09/12/2007 | 2045 |
Subject: | SGP/EBBR/E20 - Reprocess: Wind Direction 20 Degrees Low |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | The EBBR wind direction was found to be misaligned 20 degrees low. This bias should have had no effect on the res_ws or sigma_wd measurements. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/07/2007 | 2130 | 08/13/2007 | 1600 |
08/19/2007 | 1300 | 08/29/2007 | 2130 |
Subject: | SGP/EBBR/E27 - Multiplexer #2 failure |
DataStreams: | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
Description: | A problem with the #2 Multiplexer caused a large number of measurements to be incorrect. Sensible and latent heat fluxes were often affected. |
Measurements: | sgp5ebbrE27.b1:
sgp15ebbrE27.b1:
sgp30ebbrE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/20/2007 | 0630 | 08/01/2007 | 2200 |
Subject: | SGP/EBBR/E22 - H and LE Incorrect |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1 |
Description: | Soil Heat Flow 1 bounced around and went offscale at times, resulting in the EBBR program setting sensible and latent heat fluxes to 9999es. The problem ended during a PM visit, possibly because of routine checks and tightening of leads in terminal strips. |
Measurements: | sgp30ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/02/2006 | 0400 | 09/14/2006 | 1900 |
Subject: | SGP/EBBR/E13 - Radiometer dome broken |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | The top net radiometer dome was broken, yielding net radiation, sensible heat flux and latent heat flux that were too low. The BEFLUX (SIRS) net radiation can be used as a replacement for the EBBR net radiation. To correct the sensible and latent heat fluxes, determine the ratio of SIRS net radiation to EBBR net radiation for each daylight half hour and multiply that ratio times H and LE to obtain approximate corrected flux values. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/28/2007 | 0140 | 10/11/2007 | 1730 |
Subject: | SGP/EBBR/E26 - Sensible and Latent Heat Fluxes Affected by bad temp sensor |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | The left Thum (temperature of the humidity sensor chamber) measurement was incorrect at times, affecting several measurements, including the sensible and latent heat fluxes. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/15/2007 | 0000 | 10/24/2007 | 1900 |
Subject: | SGP/EBBR/E20 - Reprocess: Wind Direction 65 Degrees Low |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | The wind direction sensor reported wind directions that were 65 degrees too low because of a mis-alignment. This bias should have no effect on the res_ws or sigma_wd measurements. The wind direction data could be re-processed by adding 65 degrees. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/22/2007 | 1500 | 01/08/2008 | 1800 |
Subject: | SGP/EBBR/E9 - Net Radiation Incorrect |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Description: | There was water in the net radiometer. The net radiation data was incorrect. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/22/2007 | 1800 | 03/26/2008 | 1730 |
Subject: | SGP/EBBR/E27 - Measurements offscale |
DataStreams: | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
Description: | All J-panel 4 measurements (soil heat flows, net radiation, air temperatures, home signals, pressure, vapor pressure) were sometimes offscale or otherwise incorrect. J-Panel was replaced 20080326 correcting the problem. |
Measurements: | sgp5ebbrE27.b1:
sgp15ebbrE27.b1:
sgp30ebbrE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/11/2008 | 1100 | 02/12/2008 | 1730 |
02/21/2008 | 1230 | 02/21/2008 | 2300 |
Subject: | SGP/EBBR/E7 - Wind Speed Frozen Stopped |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | Freezing rain stopped wind speed sensor. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/18/2008 | 1000 | 03/27/2008 | 1930 |
Subject: | SGP/EBBR/E13 - Wind direction failure |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | The EBBR wind direction was stuck at 135 degrees; this reading indicates a failure of the sensor electronics. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/08/2008 | 0600 | 05/07/2008 | 2000 |
Subject: | SGP/EBBR/E20 - Wind Direction 50 Degrees High |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | The wind direction was offset high about 50 degrees. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/15/2008 | 0030 | 05/06/2008 | 2110 |
Subject: | SGP/EBBR/E18 - Soil Heat Flow #4 Offscale Low; Sensible and Latent Heat Flux Sometimes |
DataStreams: | sgp15ebbrE18.b1, sgp30ebbrE18.b1 |
Description: | Soil heat flow #4 was sometimes lower than the lower limit imposed in the EBBR programming; sensible and latent heat flux were therefore deemed to be suspect during those times. The cause of the low values is not clear, but the values are not reasonable for the vegetation type at EF18. |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/06/2008 | 0330 | 05/28/2008 | 1745 |
Subject: | SGP/EBBR/E8 - Wind Direction Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The wind direction flatlined at times at 135 degrees (an indication of failed electronics) and at other times trended correctly with the co-located SMOS but differed from it by 15 to 100 degrees. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/02/2008 | 1630 | 05/15/2008 | 2000 |
Subject: | SGP/EBBR/E2 - Net Radiation and Fluxes Incorrect |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | The net radiometer was tipped 37 degrees eastward, resulting in apparent solar noon being 2.5 hours early. The net radiation error results in incorrect sensible and latent heat fluxes. |
Measurements: | sgp30ebbrE2.b1:
sgp15ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/07/2008 | 0600 | 05/07/2008 | 0900 |
05/07/2008 | 1530 | 05/07/2008 | 2130 |
05/09/2008 | 1330 | 05/09/2008 | 1330 |
Subject: | SGP/EBBR/E19 - Sensible and latent heat flux and net radiation incorrect |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | Net radiation was too large, resulting in sensible and latent heat fluxes being too large. The cause of the problem is unknown. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/07/2008 | 0030 | 06/12/2008 | 1730 |
Subject: | SGP/EBBR/E2 - AEM Ventilator Dangling - Sensible and Latent Heat Fluxes Incorrect |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | The right AEM ventilator unit with temperature and humidity sensors inside was dangling from the AEM. Therefore, the gradients of temperature and relative humidity and thus sensible and latent heat flux are suspect for this period. Often, the fluxes look reasonable. |
Measurements: | sgp30ebbrE2.b1:
sgp15ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/11/2008 | 1500 | 06/20/2008 | 1500 |
Subject: | SGP/EBBR/E15 - Water in dome |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | There was water in the bottom dome of the net radiometer. Net radiation, and sensible and latent heat fluxes were incorrect during the period. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/01/2008 | 0030 | 06/15/2008 | 1500 |
Subject: | SGP/EBBR/E15 - Soil Heat Flow #4 Incorrect |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1 |
Description: | Soil heat flux #4 produced too large a soil heat flow at times, resulting in it exceeding the qc check limit and causing sensible and latent heat fluxes to be incorrect. |
Measurements: | sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/06/2008 | 1215 | 06/17/2008 | 2120 |
Subject: | SGP/EBBR/E18 - SHF #4 Measurements Too Low; Sensible and Latent Heat Fluxes Suspect |
DataStreams: | sgp15ebbrE18.b1, sgp30ebbrE18.b1 |
Description: | Soil heat flow #4 was replaced at the beginning of this period. However, the soil heat flow measurements from that sensor were sometimes still lower than the lower limit imposed in the EBBR programming; sensible and latent heat flux were therefore deemed to be suspect during those times. The cause of the low values could not be determined, but the values are not reasonable for the vegetation type at EF18. To remedy the situation, the soil heat flow sensor calibration was changed so that the sensor provided measurements that resembled the other four soil heat flow sensor measurements. Soil heat flow measurements have been reasonable since the calibration change. |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/15/2008 | 0230 | 09/10/2008 | 2230 |
Subject: | SGP/EBBR/E20 - Wind Direction Incorrect |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | The wind direction differed from the co-located SMOS by 50 deg during part of the stated period and the electronics had failed completely resulting in a constant 135 degrees for the following dates and times: 08/15 0230-0430, 0830-1530 and 09/03 0200-0300, 0500 - 09/10 2230. Sensible and latent heat fluxes were not affected by this problem. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/08/2009 | 1430 | 04/15/2009 | 1800 |
Subject: | SGP/EBBR/E7 - AEM voltage too high |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | The AEM was putting out too high a home 30 voltage. Sensible and latent flux data was affected by this condition, at first just at night, but later (especially after 03/17/09 0730 GMT) fluxes were affected during all hours of the day. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/27/2009 | 0350 | 01/28/2009 | 1910 |
Subject: | SGP/EBBR/E9 - Wind Speed Incorrect |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Description: | The wind speed sensor was frozen stopped by freezing rain. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/27/2009 | 0245 | 01/28/2009 | 2005 |
Subject: | SGP/EBBR/E12 - Wind Speed Incorrect |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | The wind speed sensor was stopped by freezing rain. |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/26/2009 | 2015 | 01/28/2009 | 2000 |
Subject: | SGP/EBBR/E13 - Wind Speed Incorrect |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | The wind speed sensor was stopped by freezing rain. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/26/2009 | 1830 | 01/28/2009 | 1935 |
Subject: | SGP/EBBR/E15 - Wind Speed Incorrect |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | The wind speed sensor was frozen stopped by freezing rain. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/03/2009 | 1230 | 01/03/2009 | 1500 |
Subject: | SGP/EBBR/E8 - Wind Speed Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The wind speed sensor was frozen stopped. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/26/2009 | 1910 | 01/27/2009 | 0800 |
Subject: | SGP/EBBR/E20 - Wind Speed Incorrect |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | The wind speed sensor was frozen stopped by freezing rain. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/27/2009 | 1930 | 02/25/2009 | 1830 |
Subject: | SGP/EBBR/E15 - Soil Moisture and Soil Temp #1 Incorrect |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1 |
Description: | Soil Moisture #1 and Soil Temperature #1 were incorrect during this period. This situation did not affect the sensible and latent heat fluxes (EBBR programming used only the 4 good soil measurements to determine soil surface heat flux). |
Measurements: | sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/26/2009 | 2150 | 01/29/2009 | 1550 |
Subject: | SGP/EBBR/E18 - Wind Speed Incorrect |
DataStreams: | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
Description: | The wind speed sensor was frozen stopped and thus wind speed was incorrect during this period. |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/26/2009 | 1835 | 01/28/2009 | 1825 |
Subject: | SGP/EBBR/E19 - Wind Speed Incorrect |
DataStreams: | sgp30ebbrE19.b1, sgp5ebbrE19.b1, sgp15ebbrE19.b1 |
Description: | Wind Speed was frozen stopped by ice and was incorrect during this period. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/26/2009 | 1845 | 01/28/2009 | 1810 |
Subject: | SGP/EBBR/E22 - Wind Speed Incorrect |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | The wind speed sensor was frozen stopped by freezing rain. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/26/2009 | 1805 | 01/28/2009 | 1955 |
Subject: | SGP/EBBR/E26 - Wind Speed Incorrect |
DataStreams: | sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp15ebbrE26.b1 |
Description: | The wind speed sensor was frozen stopped by freezing rain. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/26/2009 | 1800 | 01/28/2009 | 2145 |
Subject: | SGP/EBBR/E27 - Wind Speed Incorrect |
DataStreams: | sgp15ebbrE27.b1, sgp30ebbrE27.b1, sgp5ebbrE27.b1 |
Description: | The wind speed sensor was frozen stopped by freezing rain. |
Measurements: | sgp5ebbrE27.b1:
sgp15ebbrE27.b1:
sgp30ebbrE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/12/2009 | 0305 | 04/22/2009 | 1400 |
Subject: | SGP/EBBR/E20 - Wind Direction Incorrect |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | The wind direction measurement was stuck at 40 degrees. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/20/2009 | 1830 | 06/22/2009 | 1500 |
Subject: | SGP/EBBR/E7 - Fluxes intermittently incorrect |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | Soil temperature #2 frequently spiked, causing incorrect sensible and latent heat fluxes. The condition was not always obvious from the qc checks. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/16/2009 | 1000 | 06/22/2009 | 1500 |
Subject: | SGP/EBBR/E7 - Wind direction incorrect |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | The wind direction flat-lined at ~10 degrees. System was replaced. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/10/2009 | 0930 | 08/19/2009 | 1800 |
Subject: | SGP/EBBR/E8 - Insufficient voltage |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | Low DC supply voltage caused many measurements to be incorrect during this period. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/29/2009 | 0000 | 10/20/2009 | 1330 |
Subject: | SGP/EBBR/E2 - Low Supply Voltage |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | Low supply voltage caused incorrect and missing data during much of the period. This facility was decommissioned 10/20/2009 and maintenance was discontinued effective 10/01/2009. |
Measurements: | sgp15ebbrE2.b1:
sgp30ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/14/2009 | 0230 | 10/17/2009 | 1530 |
Subject: | SGP/EBBR/E26 - Low supply voltage |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | Low supply voltage during much of the period resulted in missing or incorrect data most of the time. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/03/2010 | 1400 | 01/04/2010 | 0005 |
Subject: | SGP/EBBR/E7 - Wind speed frozen stopped |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | The wind speed sensor was frozen stopped by freezing rain. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/29/2010 | 1130 | 01/31/2010 | 2200 |
Subject: | SGP/EBBR/E19 - Wind direction sensor frozen near north |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | The wind direction sensor was frozen in place near north by freezing rain. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/28/2010 | 2030 | 02/01/2010 | 1930 |
Subject: | SGP/EBBR/E20 - Wind speed sensor frozen stopped |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | The wind speed sensor was frozen stopped by freezing rain. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/28/2010 | 1630 | 02/02/2010 | 1530 |
Subject: | SGP/EBBR/E19 - Wind speed sensor frozen stopped |
DataStreams: | sgp5ebbrE19.b1, sgp15ebbrE19.b1, sgp30ebbrE19.b1 |
Description: | The wind speed sensor was frozen stopped by freezing rain. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/21/2010 | 2300 | 02/22/2010 | 1730 |
Subject: | SGP/EBBR/E12 - Wind Speed Sensor Frozen Stopped by Icing |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | Icing stopped the wind speed sensor cups and resulted in the measurement of wind speed being equivalent to the sensor offset. |
Measurements: | sgp30ebbrE12.b1:
sgp15ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/19/2010 | 1930 | 05/05/2010 | 1745 |
Subject: | SGP/EBBR/E15 - Sensible and Latent Heat Fluxes Low (Net Radiation too low) |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | Sensible and Latent Heat Fluxes were measured as being low because of damage to the net radiometer. The Sensible and Latent Heat fluxes can be approximated by taking the ratio of the daytime SIRS net radiation to EBBR net radiation, and then multiplying the Sensible and Latent Heat Fluxes by that ratio. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/20/2010 | 1730 | 05/10/2010 | 1800 |
Subject: | SGP/EBBR/E9 - Sensible and latent heat flux incorrect |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1 |
Description: | Sensible and latent heat fluxes were sometimes incorrect when soil temperature 4 and soil moisture 4 were acting up. Sometimes the sensible and latent heat fluxes are indicated as incorrect by an offscale reading produced by the data logger error checking routine. This occurs mostly during daytime. However, you will have to examine the soil moisture 4 and soil temperature 4 data to see when they were incorrect or offscale to determine all times for which sensible and latent heat fluxes were incorrect. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/03/2010 | 0000 | 06/08/2010 | 1800 |
Subject: | SGP/EBBR/E12 - Intermittent data loss |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | The datalogger at E12 EBBR reset it's time to Dec 3, 1999 on a periodic basis. This occurred on 5/3, 5/15, 5/17, 5/18, 5/27, 6/5 and 6/6 of 2010. Each time this occurred we lost 1-2 hours of data. The datalogger was apparently at fault. The problem has not re-occurred since the datalogger was replaced on 6/8/2010. |
Measurements: | sgp30ebbrE12.b1:
sgp15ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/10/2010 | 1200 | 06/23/2010 | 1630 |
Subject: | SGP/EBBR/E12 - Sensible and Latent Heat Fluxes and Other Measurements Incorrect |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | The power supply leads for the combined temperature probe shorted somehow, resulting in several measurements from the probe and for the multiplexer that it was connected to, to be incorrect. |
Measurements: | sgp30ebbrE12.b1:
sgp15ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/17/2010 | 2105 | 07/22/2010 | 1600 |
Subject: | SGP/EBBR/E7 - Incorrect Sensible and Latent Heat Fluxes |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | Soil heat flow 4 failed. Latent and sensible heat flux values were incorrect at times, when the EBBR program did not reject the incorrect soil heat flow. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/06/2010 | 1930 | 07/21/2010 | 1400 |
Subject: | SGP/EBBR/E13 - Incorrect Sensible and Latent Heat Fluxes |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | The multiplexer to which the air temperatures were connected experienced a problem, resulting in incorrect sensible and latent heat fluxes at times. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/08/2010 | 0730 | 07/29/2010 | 1100 |
Subject: | SGP/EBBR/E20 - Incorrect Sensible and Latent Heat Fluxes |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | Because of a periodic AEM failure, home signal 30 was often stuck at the home signal 15 position. These occurrences resulted in incorrect latent and sensible heat flux values at times. Some missing data occurred because of low battery voltage (the CR10 program stops running when battery voltage is insufficient. The data outage events were preceded and followed by several hours of bad/suspicious data. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/09/2010 | 1030 | 08/04/2010 | 0200 |
Subject: | SGP/EBBR/E20 - Incorrect Sensible and Latent Heat Fluxes - a result of low battery |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | Supply voltage sometimes dropped too low at night, causing incorrect measurements of sensible and latent heat fluxes, pressure, AEM home signal, and soil heat flow. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/01/2010 | 1200 | 09/29/2010 | 1600 |
Subject: | SGP/EBBR/E12 - Net radiation too high |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | Net radiation was too high, resulting in overestimated sensible and latent heat fluxes. Net rad domes replaced. |
Measurements: | sgp30ebbrE12.b1:
sgp15ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/21/2010 | 1330 | 10/01/2010 | 2000 |
Subject: | SGP/EBBR/E13 - Net Radiation Too Large |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | The EBBR Net Radiation (Rn) values maintained a ~200 Wm^-2 reading because of poor cable lead connections. Sensible and latent heat fluxes were therefore overestimated. Sensible (H) and latent heat (LE) fluxes can be estimated in the following way: H (new) = H/[1 + (EBBR Rn - SIRS net radiation)/SIRS net radiation] LE (new) = LE/[1 + (EBBR Rn - SIRS net radiation)/SIRS net radiation] |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/24/2010 | 1630 | 10/05/2010 | 1600 |
Subject: | SGP/EBBR/E15 - Sensible and Latent Heat Flux Suspect - Soil Heat Flow 5 Problem |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1 |
Description: | Soil Heat Flow sensor 5 was offscale at times, causing sensible and latent heat fluxes to be incorrect during those times. |
Measurements: | sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/09/2010 | 1830 | 10/06/2010 | 1830 |
Subject: | SGP/EBBR/E7 - Sensible and latent heat fluxes incorrect - soil heat flow 1 problem |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | Soil Heat Flow 1 was offscale at times, affecting sensible and latent heat fluxes during those times. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/28/2010 | 1630 | 10/06/2010 | 2130 |
Subject: | SGP/EBBR/E9 - T/RH Probe Temperature Problem - Sensible and Latent Heat Fluxes Affected |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Description: | During mid-day the left T/RH probe temperature was offscale, sometimes resulting in incorrect sensible and latent heat fluxes. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/22/2007 | 1830 | 11/10/2009 | 1430 |
Subject: | SGP/EBBR/E8 - RH Maximum too high |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The absolute RH measurements were too high for RH > 100%. Use of first (instead of second order) coefficients was no improvement. The calculated sensible and latent heat fluxes were unaffected. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/14/2010 | 0800 | 12/07/2010 | 1600 |
Subject: | SGP/EBBR/E9 - Sensible and Latent Heat Fluxes Incorrect |
DataStreams: | sgp5ebbrE9.b1, sgp15ebbrE9.b1, sgp30ebbrE9.b1 |
Description: | The right Air Temperature probe leads were loose. During nighttime the right Air Temperature went offscale, resulting in incorrect sensible and latent heat fluxes. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/30/2011 | 1500 | 07/15/2011 | 1400 |
Subject: | SGP/EBBR/E13 - Soil temperature #3 offscale |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
Description: | Soil Temp Sensor 3 data were missing or offscale due to a bad connection. H and LE were not normally affected, as the EBBR program "kicked" out the incorrect temperature measurements before calculating H and LE. |
Measurements: | sgp15ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/29/2011 | 1530 | 08/03/2011 | 1500 |
Subject: | SGP/EBBR/E12 - Loose connection (SHF #4) |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1 |
Description: | Soil Heat Flow #4 was sometimes offscale due to a bad connection. Sensible and latent heat fluxes also affected at times. |
Measurements: | sgp30ebbrE12.b1:
sgp15ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/09/2011 | 0500 | 09/29/2011 | 1600 |
Subject: | SGP/EBBR/E9 - Sensible and Latent Heat Fluxes Incorrect |
DataStreams: | sgp30ebbrE9.b1 |
Description: | The AEM stuck in the home 15 or home 30 position, resulting in incorrect sensible and latent heat fluxes. |
Measurements: | sgp30ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/21/2011 | 2030 | 08/03/2011 | 1930 |
Subject: | SGP/EBBR/E7 - Loose connection (ST #3) |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | Soil temperature 3 was going offscale at times due to loose connections. Sensible and latent heat fluxes were not affected. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/05/2011 | 1030 | 09/19/2011 | 1700 |
Subject: | SGP/EBBR/E19 - Nighttime data missing |
DataStreams: | sgp30ebbrE19.b1, sgp5ebbrE19.b1, sgp15ebbrE19.b1 |
Description: | Battery supply voltage was insufficient at night to power the system resulting in missing data. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/03/2011 | 1330 | 09/28/2011 | 1900 |
Subject: | SGP/EBBR/E7 - Sensible and Latent Heat Flux Incorrect |
DataStreams: | sgp30ebbrE7.b1 |
Description: | Latent and sensible heat flux were incorrect because the AEM was not working properly. |
Measurements: | sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/25/2011 | 1200 | 08/24/2011 | 1930 |
Subject: | SGP/EBBR/E20 - Sensible and latent heat fluxes overestimated |
DataStreams: | sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | Net Radiation was too large resulting in overestimated sensible heat and latent heat fluxes. |
Measurements: | sgp5ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/28/2011 | 2300 | 01/24/2012 | 1730 |
Subject: | SGP/EBBR/E15 - Soil heat flow #1 failure |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1 |
Description: | Soil heat flow 1 sensor wires chewed by rodents. Values less than the accepted minimum. |
Measurements: | sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/28/2011 | 1400 | 02/06/2012 | 1830 |
Subject: | SGP/EBBR/E15 - Soil heat flow #5 failure |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1 |
Description: | Soil_heat_flow_5 sensor failed at times, resulting in sensible and latent heat fluxes being incorrect at times. Sensor was replaced 2/6/2012. |
Measurements: | sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/19/2012 | 1700 | 02/12/2012 | 1630 |
Subject: | SGP/EBBR/E12 - Soil heat flux #4 intermittently incorrect |
DataStreams: | sgp30ebbrE12.b1 |
Description: | Soil heat flux #4 measurements were sometimes incorrect and sometimes affected the sensible and latent heat flux calculations. |
Measurements: | sgp30ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/06/1993 | 0631 | 07/06/1993 | 0632 |
Subject: | EBBR - soil heat flux plate giving opposite sign |
DataStreams: | sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | EDITOR'S NOTE: This DQR refers to data collected by the EBBR at E20 prior to the begin date of regular ARM data. At that time, the data streams which contain the EBBR.E20 data were named Dsgp15ebbr6.a0 Dsgp30ebbr6.a1 Dsgp5ebbr6.a0 These data are not readily available from the ARM archive, but can be made available by special request. The actual time range of the problem described here is 930405.0000-930429.2359. Soil heat flux sensor #3 of EBBR6, at Meeker, Ok extended site was determined by me to have one heat flux plate whose output was out of phase with the other four. The absolute value of the soil heat flux value produced by the plate was good. Normally, this condition is caused either by improper wiring of the sensor into the data acquisition system (this was checked by site operations personnel on 29 April; wiring convention was correct) or by the plate being installed upside down. The latter seems fairly obvious, although the plate was not dug up to check it's orientation; that is not neccesary and would be detrimental since it takes time for such sensors to acclimate themselves to their soil environment. The consequence of this problem is that 5 soil heat flux plate outputs were being averaged, with one of the outputs being of opposite sign from the other four. Since the outputs are similar, in absolute terms, this essentially resulted in an average soil heat flux plate value that is 40% too low. In terms of output, this a reduction of around 10 W per meters squared, or 1-2% of the RN+G term during maximum solar insolation. This is clearly a small error. Site operations personnel switched the soil heat flux plate leads to give the proper sign of output, on 29 April 1993, time unknown (this can be determined from looking at the ingested data later; it hasn't been ingested yet). The instrument operated properly at that point. Thus, the problem has been corrected. |
Measurements: | sgp5ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/13/1993 | 1200 | 11/02/1993 | 2121 |
Subject: | SGP/EBBR/E13 - pressure calibration |
DataStreams: | sgp30ebbrE13.b1 |
Description: | In the original Campbell CR10 software provided by the vendor (version 1), the EBBR had incorrect calibration pressure sensor slopes and offsets. The slopes and offset was carried through into version 2 of the software and was not discovered by the mentor (me) until sometime later. The calibration and slope did not deleteriously affect calculations in the program since only a four percent reduction in atmospheric pressure measurement resulted (the calculations are not very sensitive to small errors in pressure). The vendor does not recall how the wrong calibrations came to be used. Version 4 of the CR10 program includes the proper calibration and offset. This remedies the apparent offset problem (it was actually a calibration and offset problem). In comparisons with the co-located SMOS pressure sensor, the difference between the SMOS and EBBR pressure values appeared almost constant with time, hence the assumption of an offset problem. However, the EBBR pressure sensor offsets are so large (a consequence of tailoring the calibration curve to a small range of elevations in which the calibration applies) that the very small error in slope was difficult to recognize. The outputted pressure values for the EBBR system can be corrected with the following information and equations. Old Slope Old Offset New Slope New Offset 0.01952 81.24 0.02031 81.27 Correct Pressure = ((Current Pressure Data - 81.24)*0.02031/0.01952) + 81.27 = 1.04047*(Current Pressure Data) - 3.258 |
Measurements: | sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/13/1993 | 1200 | 11/02/1993 | 2121 |
Subject: | SGP/EBBR/E13 - solid moisture correction |
DataStreams: | sgp30ebbrE13.b1 |
Description: | In the original EBBR software (Version 1) for the Campbell CR10 at the Central Facility (E13), adjustment of the soil moisture for soil type was performed differently than for the other EBBR systems. This was supposed to have been changed prior to installation (according to the vendor), but the change was not made. This adjustment was carried into Version 2 and was not discovered by the mentor (me) until later. A polynomial is used to adjust the measured soil moisture for differences between the soil type at the location of installation (clay-loam assumed) and the soil type (sand) used in the calibration of the soil moisture sensor. The adjustment in the E13 EBBR only, was computed on the basis of a sandy-loam soil. Presently, a clay-loam soil type is assumed for each EBBR location. Soil types at the extended facilities have not been characterized yet, although there are hopes of doing so in the future. The effect of using a polynomial for sandy-loam soil instead of clay-loam soil is to decrease the resultant soil moisture values by approximately 14.5% when the clay-loam polynomial would indicate a soil moisture of 28.3%, and by approximately 38.3% when the clay-loam polynomial would indicate a soil moisture of 6.3%. The normal range of soil moisture at the EBBR sites is approximately 3% to 35%. The effect of the soil moisture decreases above on the adjustment to soil heat flux plate measurements is less than 1.0%. The effect on the change in energy storage in the soil (determined from the change in soil temperature over time) is a 10% decrease for a soil moisture of 28.3% and a decrease of 13.0% for a soil moisture of 6.3%. The change in energy storage is typically 5% of the total surface energy budget during mid-day and 25% at night. Therefore, the soil type adjustment error in the Central Facility EBBR resulted in approximately a 0.5% error in the surface energy budget during mid-day and 2.5% at night. These are small and acceptable errors and should not cause one to need to recalculate energy budget values. |
Measurements: | sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/11/1993 | 0300 | 11/01/1993 | 1700 |
Subject: | SGP/EBBR/E9 - pressure calibration |
DataStreams: | sgp30ebbrE9.b1 |
Description: | In the original Campbell CR10 software provided by the vendor (version 1), the EBBR had incorrect calibration pressure sensor slopes and offsets. The slopes and offset was carried through into version 2 of the software and was not discovered by the mentor (me) until sometime later. The calibration and slope did not deleteriously affect calculations in the program since only a four percent reduction in atmospheric pressure measurement resulted (the calculations are not very sensitive to small errors in pressure). The vendor does not recall how the wrong calibrations came to be used. Version 4 of the CR10 program includes the proper calibration and offset. This remedies the apparent offset problem (it was actually a calibration and offset problem). In comparisons with the co-located SMOS pressure sensor, the difference between the SMOS and EBBR pressure values appeared almost constant with time, hence the assumption of an offset problem. However, the EBBR pressure sensor offsets are so large (a consequence of tailoring the calibration curve to a small range of elevations in which the calibration applies) that the very small error in slope was difficult to recognize. The outputted pressure values for the EBBR system can be corrected with the following information and equations. Old Slope Old Offset New Slope New Offset 0.01952 81.24 0.02031 81.27 Correct Pressure = ((Current Pressure Data - 81.24)*0.02031/0.01952) + 81.27 = 1.04047*(Current Pressure Data) - 3.258 |
Measurements: | sgp30ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/22/1993 | 1630 | 11/03/1993 | 1730 |
Subject: | SGP/EBBR/E8 - Drop in atmospheric pressure |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | Beginning September 22, 1993, the atmospheric pressure sensor at the E8 EBBR began showing drops of as much as 5 kPa for about half a day. These could be easily seen in comparisons with the SMOS pressure sensor. The drops were diurnal and occurred during daytime, at the same time as other quantities such as temperature, net radiation, and battery voltage were at the highest levels of the day. It is unclear what exactly caused this situation. Replacement of the Campbell multiplexer to which the pressure sensor was connected caused the problem to go away. The same multiplexer was then used as a replacement for the other multiplexer in the E8 EBBR and has not shown any odd behavior on the channel that previously served the pressure sensor. My conclusion is that there was poor electronic contact of the pressure sensor to the multiplexer. Affected pressure data can only be easily seen by comparison with SMOS pressure data. No EBBR data other than pressure is affected by the pressure problem; calculations of specific heat at constant pressure (which includes the term 1/(p-vp)) times other terms (including p-.378*vp) result in a very small error in the resulting Bowen Ratio. For example; assume a common vapor pressure (vp) of 2.0 kPa, a common atmospheric pressure of 97.0 kPa, and a corrupted atmospheric pressure of 92.0 kPa. Uncorrupted Corrupted (97.0-.378*2.0)/(97.0-2.0) = 1.0131 (92.0-.378*2.0)/(92.0-2.0) = 1.0138 1.0138/1.0131 = 1.00069 or 0.069% increase in the Bowen Ratio as a result of the corruption Therefore, no correction to the quantities calculated using pressure is warranted. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/11/1993 | 1100 | 12/01/1993 | 1830 |
Subject: | Incorrect readings for pressure sensor |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | ARM PROBLEM IDENTIFICATION FORM (PIF) PIF No. P940105.1 Subject: SGP/Ebbr - Incorrect readings for pressure sensor at E15 Submitted By: David R. Cook Organization: Argonne National Laboratory Date Submitted: December 13, 1993 How may we contact you? e-mail: cook@anler.er.anl.gov phone: (708) 252-5840 FAX: (708) 252-9792 Submitter's Priority: 3 [Critical-1 Very Important-2 Important-3 Inconvenient-4 Interesting-5] Where was this Problem Identified: DA [Site Data System-SDS Experiment Center-EC] [Archive-A At Field Instrument-FI ] [During Data Analyis-DA Other- List Location] Does this PIF result in a Software Change Request? No Where: [ SDS, EC, or A ] Type: [Development-1 Problem-2 Enhancement-3] List Programs/Documents Affected: Configuration Identification: Does problem impact data values or cause data loss? Data values. which platform(s): EBBR, E15, Ringwood, OK Specify (or estimate) begin and end dates Begin Date 7/11/93 End Date 12/01/93 Time 1830 GMT Apparent cause of data loss, if known: SUGGESTED CAUSES: Calibration Drift. [human error, component failure, temperature, lightning ] [foreign matter, power loss, software failure, ] [communications failure, modification difficulites, specify other] Problem Description/Change Description. o For Instrument Problems ONLY, Identify MODE of Operation: Normal o Give a brief explanation with details, attach examples and any supporting information. This should include a description of analysis leading to identification of problem and, if known, recommended action. Beginning July 11, 1993, the atmospheric pressure sensor at the E15 EBBR began indicating approximately 0.2 KPa lower than the colocated SMOS. Between July 31 and August 22, 1993 the difference increased to between 0.25 and 0.30 KPa. After August 22, the difference increased to 0.3 to 0.35 KPa and remained that way until December 1, 1993. A comparison of the EBBR pressure sensor with the portable standard on December 1, 1993 indicated that the EBBR sensor was 0.04 KPa higher than the standard; this is excellent agreement considering that the uncertainty of the EBBR sensor calibration is 0.15 KPa. The consistent bias in EBBR pressure between July 11 and December 1 can be easily detected in comparisons with SMOS pressure data. No EBBR data other than pressure was affected by the pressure problem; calculations of specific heat at constant pressure (which includes the term 1/(p-vp)) times other terms (including p-.378*vp) result in no error to the third decimal point at least in the resulting Bowen Ratio. For example; assume a common vapor pressure (vp) of 2.0 kPa, a common atmospheric pressure of 97.0 kPa, and a corrupted atmospheric pressure of 96.65 kPa at the extreme. Uncorrupted Corrupted (97.0-.378*2.0)/(97.0-2.0) = 1.0131 (96.65-.378*2.0)/(96.65-2.0) = 1.0131 1.0131/1.0131 = 1.0000 There is no effect on the Bowen Ratio as a result of the corruption. Therefore, no correction to the quantities calculated using pressure is needed. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/01/1993 | 0000 | 12/21/1993 | 1800 |
Subject: | SGP/EBBR/E22 - Air Temperature Sensor malfunction |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | The right air temperature probe (thermocouple) malfunctioned during the periods stated above at Cordell, OK (E22). It was replaced on December 21, 1993 and good data began again with the 1830 GMT half hour average. For the periods of interest, 5 and 30 minute values of tair_top and tair_bot are incorrect when the right probe was in the specified position; 15 minute values of tair_r are incorrect; and 30 minute values of bowen, e, and h are incorrect. None of these values are recoverable or recalculatable through the use of other quantities since both air temperature probes are needed to determine bowen ratio. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/30/1993 | 1300 | 11/12/1993 | 1600 |
Subject: | SGP/EBBR/E20 - Blown fuse in AEM at E20 |
DataStreams: | sgp30ebbrE20.b1, sgp5ebbrE20.b1, sgp15ebbrE20.b1 |
Description: | On October 30, 1993, before 1300 GMT, the fuse in the AEM (Automatic Exchange Mechanism) of Meeker, Ok (E20) EBBR blew out. The AEM remained nonfunctional until the fuse was replaced just before 1600 GMT on November 12, 1993. This situation resulted in the following invalid data: 5 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, home 15 minute: rr_thum_r, rr_thum_l, mv_hum_r, mv_hum_l, mv_home, tair_r, tair_l 30 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, bowen, e, h, home_15, home_30 If the biases of the sensors were know, bowen, h, and e could be recalculated from the 15 minute and/or 5 minute data. Since we will not have any idea of what those biases are until an annual calibration of the sensors can take place (which may be months away), the recalculations cannot be performed at this time. Furthermore, it may be invalid to use sensor biases, if determined too far removed in time, due to the gradual calibration drift of some of the sensors. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/09/1993 | 1700 | 11/22/1993 | 1700 |
Subject: | SGP/EBBR/E22 - Incorrect averaging interval at E22 |
DataStreams: | sgp30ebbrE22.b1 |
Description: | On November 9, 1993 version 4 of the EBBR program installed at Cordell, OK (E22) was mistakenly modified in one area during the process of modifying a calibration in another part of the program. The mistaken modification was a change in the execution interval from 30 seconds to 10 seconds. This modification remained in the program until November 22, 1993 when it was discovered and changed back to a 30 second execution interval. The modification produced no ill effects as the execution interval only changes sampling and averaging periods. A 10 second execution interval results in three times as many samples being taken, which, when averaged, should actually produce more accurate results. Therefore, data quality was not affected. This PIF was written only to document the fact that the mistaken modification was made, and when it was made. |
Measurements: | sgp30ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/24/1993 | 1215 | 11/24/1993 | 1515 |
Subject: | SGP/EBBR/E20 - Reprocess: Questionable data in soil heat fluxes at E20 |
DataStreams: | sgp30ebbrE20.b1, sgp15ebbrE20.b1 |
Description: | Soil heat flux probes #4 had been indicating soil heat fluxes that were higher than the other four probes. It was decided to replace it. On November 24, 1993 Soil Heat Flux probe #4 was replaced, resulting in outputs more similar to the other four probes. This PIF does not invalidate previous measurements by the old probe #4, but only indicates data that is bad as a result of the replacement procedure. This procedure resulted in some bad data for the adjoining soil heat flux probes #3 and #5 also. Bad data for #3 includes the 1600 GMT half hour, and 5 and 15 minute periods between 1530 and 1600 GMT. Bad data for #4 includes half hour times of 1600 through 1900 GMT, with 5 and 15 minute data inbetween those times being invalid as well. Bad data for #5 includes half hour times of 1630 and 1700 GMT, with 5 and 15 minute periods inbetween those times being invalid as well. Corrupted 30 minute values of shf4, c_shf4, g4, ave_shf, h, and e, and corrupted 15 minute values of mv_hft4 resulted. The quantities shf4, c_shf4, g4, and mv_hft4 are bad data values and are not recoverable. The other quantities can be recalculated using the remaining functioning set(s) of soil sensors. Use the following equations: ave_shf = (gs summed)/n, where n is the number of probes functioning e = -(q + ave_shf)/(1 + bowen) where q is net radiation, and bowen is Bowen Ratio h = e * bowen |
Measurements: | sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/17/1993 | 1100 | 12/29/1993 | 1700 |
Subject: | SGP/EBBR/E8 - Invalid data due to AEM fuse blow out at E8 |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | On December 17, 1993, before 1100 GMT, the fuse in the AEM (Automatic Exchange Mechanism) of Coldwater, KS (E8) EBBR blew out. The AEM remained nonfunctional until the fuse was replaced just before 1700 GMT on December 29, 1993. During the 1000 GMT half hour on December 17 the AEM was hung up; this is clear from the home signal. These situations resulted in the following invalid data: 5 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, home 15 minute: rr_thum_r, rr_thum_l, mv_hum_r, mv_hum_l, mv_home, tair_r, tair_l 30 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, bowen, e, h, home_15, home_30 Bowen, h, and e can not be recalculated from the 15 minute and/or 5 minute data since we cannot determine the position of the sensors on the AEM from the home signal; home signal is near zero when there is no power to the AEM. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/20/1993 | 0730 | 12/22/1993 | 1500 |
Subject: | SGP/EBBR/E12 - Invalid data due to blown AEM fuse at E12 |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | On December 20, 1993, before 0730 GMT, the fuse in the AEM (Automatic Exchange Mechanism) of Pawhuska, OK (E12) EBBR blew out. The AEM remained nonfunctional until the fuse was replaced just before 1500 GMT on December 22, 1993. This situation resulted in the following invalid data: 5 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, home 15 minute: rr_thum_r, rr_thum_l, mv_hum_r, mv_hum_l, mv_home, tair_r, tair_l 30 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, bowen, e, h, home_15, home_30 Bowen, h, and e can not be recalculated from the 15 minute and/or 5 minute data since we cannot determine the position of the sensors on the AEM from the home signal; home signal is near zero when there is no power to the AEM. |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/31/1993 | 1950 | 12/31/1993 | 2100 |
Subject: | SGP/EBBR/E13 - Invalid data during disconnection of AEM at E13 |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | On December 31, 1993 at 1950 GMT the AEM on the Central Facility EBBR was disconnected. Another AEM was then connected, on which some testing was needed before site operations could be confident that it was ready for field installation at an extended facility. At 2041 GMT the original AEM was reconnected. The following 5, 15, and 30 minute data between or for the stated times below is invalid: 5 minute (1950 to 2045 GMT): tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, home 15 minute (2000 to 2045 GMT): rr_thum_r, rr_thum_l, mv_hum_r, mv_hum_l, mv_home, tair_r, tair_l 30 minute (2000 through 2100 GMT): tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, bowen, e, h, home_15, home_30 |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/28/1993 | 1645 | 12/28/1993 | 1900 |
Subject: | SGP/EBBR/E26 - Invalid data during Preventative Maintenance |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | During a preventative maintenance visit the +12V DC supply to the wind direction sensor at E26, Cement, Ok was temporarily disconnected, resulting in invalid wind direction data for December 28, 1993 from 1645 through 1900 GMT. The following fields are therefore invalid for that period: 30 minute: wind_d, sigma_wd, res_ws 15 minute: mv_wind_d 5 minute: wind_d, sigma_wd, res_ws. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/12/1994 | 1100 | 01/12/1994 | 1630 |
Subject: | SGP/EBBR/E8 - Invalid wind data frozen sensor |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | On January 12, 1994 the EBBR wind speed sensor at E8, Coldwater, KS was frozen to a stationary by ice. Invalid data for the time period 1100 to 1630 GMT is: 30 minute: wind_s, res_ws 15 minute: wind_s 5 minute: wind_s, res_ws. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/01/1993 | 0000 | 02/02/1994 | 2359 |
Subject: | SGP/EBBR/E26 - Winds, SM2 Invalid data 10/10/93 - 2/2/94 |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | Beginning October 10, 1993 the EBBR system at E26, Cement, OK began to have problems with outputs from soil moisture sensor number 2 and the wind direction sensor. This situation also occured in early 1993 and was eliminated by replacing the soil moisture sensor (after replacing the wind direction sensor did not help). This time, replacing the soil moisture sensor did not help, but the problem has not re-occured since the wind direction sensor was replaced on February 2, 1994. During periods of bad data the wind direction output is a constant 135 degrees and soil moisture two is offscale. Invalid data for a constant wind direction of 135 degrees is: 30 minute: res_ws, wind_d, sigma_wd 15 minute: mv_wind_d 5 minute: res_ws, wind_d, sigma_wd. Invalid data for soil moisture 2 offscale is: 30 minute: sm2, c_shf2, cs2, ces2, g2, ave_shf, h, e 15 minute: r_sm2 5 minute: no data affected. Apparent cause of data loss, if known: Apparently an electronic problem in interfacing an intermittently malfunctioning wind direction sensor to the data acquisition system multiplexer, thereby affecting the other sensor (soil moisture 2) on that multiplexer pair. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/21/1994 | 2330 | 02/23/1994 | 1800 |
Subject: | SGP/EBBR/E8 - Invalid wind speed frozen sensor |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | On February 21, 1994 the EBBR wind speed sensor at E8, Coldwater, KS was frozen to a stationary condition by ice. Invalid data for the time period indicated is: 30 minute: wind_s, res_ws 15 minute: wind_s 5 minute: wind_s, res_ws. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/23/1994 | 1030 | 02/24/1994 | 1540 |
Subject: | Invalid wind speed frozen sensor |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | ARM PROBLEM IDENTIFICATION FORM (PIF) PIF No. P940301.4 Submitted By: David R. Cook Organization: Argonne National Laboratory Date Submitted: March 1, 1994 Problem Description/Change Description. o For Instrument Problems ONLY, Identify MODE of Operation: Degraded o Give a brief explanation with details, attach examples and any supporting information. This should include a description of analysis leading to identification of problem and, if known, recommended action. This is an informational PIF. On February 23, 1994 the EBBR wind speed sensor at E7, Elk Falls, KS was frozen to a stationary condition by ice. Invalid data for the time period indicated is: 30 minute: wind_s, res_ws 15 minute: wind_s 5 minute: wind_s, res_ws. How may we contact you? e-mail: cook@anler.er.anl.gov phone: (708) 252-5840 FAX: (708) 252-9792 Submitter's Priority: 3 [Critical-1 Very Important-2 Important-3 Inconvenient-4 Interesting-5] Where was this Problem Identified: FI [Site Data System-SDS Experiment Center-EC] [Archive-A At Field Instrument-FI ] [During Data Analyis-DA Other- List Location] Does this PIF result in a Software Change Request? NO Where: [ SDS, EC, or A ] Type: [Development-1 Problem-2 Enhancement-3] List Programs/Documents Affected: Configuration Identification: Does problem impact data values or cause data loss? Data Loss. which platform(s): EBBR, E7, Elk Falls, KS Specify (or estimate) begin and end dates for data loss Begin Date 2/23/94 Time 1030 GMT End Date 2/23/94 Time 1540 GMT Apparent cause of data loss, if known: Ice SUGGESTED CAUSES: [human error, component failure, temperature, lightning ] [foreign matter, power loss, software failure, ] [communications failure, modification difficulites, specify other] |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/23/1994 | 0400 | 02/24/1994 | 1300 |
Subject: | SGP/EBBR/E9 - Invalid wind speed frozen sensor |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Description: | On February 23, 1994 the EBBR wind speed sensor at E9, Ashton, KS was frozen to a stationary condition by ice. It broke free in strong winds on February 24, 1994. Invalid data for the time period indicated is: 30 minute: wind_s, res_ws 15 minute: wind_s 5 minute: wind_s, res_ws. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/07/1994 | 1400 | 03/16/1994 | 1700 |
Subject: | SGP/EBBR/E26 - Wind Direction Sensor Malfunction |
DataStreams: | sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp15ebbrE26.b1 |
Description: | The wind direction sensor malfunctioned, outputting nearly a constant value, invalidating the following data values. 30 minute: wind_d, sigma_wd, res_ws 15 minute: mv_wind_d 5 minute: wind_d, sigma_wd, res_ws The first good data after the affected period was the 1700 GMT, 30 minute data on March 16, 1994. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/17/1994 | 1600 | 03/17/1994 | 1630 |
Subject: | SGP/EBBR/E20 - Pressure Sensor Malfunction |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | Pressure sensor malfunction during attempt to monitor it during Preventative Maintenance visit. 30 minute values of pressure for 1600 GMT and 1630 GMT, 15 minute values for 1600, 1615, and 1630, and 5 minute values for 1600, 1605, 1610, 1615, 1620, 1625, and 1630 GMT are incorrect. However, the range of the incorrect values was such as to not adversely affect flux measurements. Some noise was apparently induced in the pressure output voltage as a result of the monitoring, although it is not known how. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/11/1993 | 1501 | 03/22/1994 | 1643 |
Subject: | SGP/E9/EBBR - AEM Malfunction |
DataStreams: | sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Description: | EDITOR'S NOTE: This DQR refers, in part, to data collected prior to the begin date of regular ARM data. During the first month covered by this DQR, the EBBR.E9 data were not archived with standard ARM names. The data are therefore not readily available, but can be made available by special request. The total period covered by the problem documented here is 930615.0000-940322.1643. The AEM at E9 frequently did not function properly during the time period stated. Please see the notes in the EBBR platform describing home signal conditions for which certain of the data fields are suspect. The newly installed AEM is performing properly so far. The files could be flagged, but it is up to the user to read the platform notes concerning the condition of the home signal and then view the quality of the data accordingly. I would be at my terminal full- time in attempting to list all days and times and data fields that are affected by the numerous AEM malfunctions. Furthermore, I have done so previously in several PIFs that I will reference here for information: P931101.4, P931115.2, P940105.2, P940211.3 |
Measurements: | sgp30ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/05/1994 | 1500 | 04/19/1994 | 1525 |
Subject: | SGP/EBBR/E9 - AEM failure |
DataStreams: | sgp30ebbrE9.b1 |
Description: | The Automatic Exchange Mechanism would not function. It was decided to take the unit back to SGP CART to try to repair it. A spare AEM was not available for a replacement, so the aspirated radiation shields housing temperature and humidity gradient sensors were attached to the EBBR framework, one meter apart vertically, with the right side in the lower position. The housings remained this way for two weeks. During this time period, home signal, sensible heat flux, and latent heat flux values are incorrect. Significant reprocessing would be required to obtain correct fluxes, and that would be possible only if one assumed that offsets in the sensor calibrations are negligible; that is a very debatable assumption considering that the EBBR system is already half a year past due for calibration. |
Measurements: | sgp30ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/06/1994 | 1515 | 05/06/1994 | 1600 |
Subject: | SGP/EBBR/E13 - SM4 Replacement |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
Description: | Soil Moisture Probe #4 has been indicating large soil moistures in the past couple of months. At times the values have exceeded the upper limit of the acceptable range. This data could be considered questionable, although the affect on average soil moisture and quantities affected by it is very small. Therefore, I do not suggest additional flagging of the data other than that imposed by the limits. The probe was changed from S/N 92004 to 93004 on May 6, 1994 at 1530 GMT. During the exchange process a few selected periods of soil moisture #4 data is incorrect, but not so different from the other soil moisture probes as to have a significant affect on the average soil moisture or the fluxes. Additionally, the EBBR output was interrupted for a half hour while the calibration of the new probe was entered into the CR10 data logger. Therefore, all 5 minute data from 1535 to 1600 GMT, 15 minute data for 1545 and 1600 GMT, and 30 minute data for 1600 GMT is incorrect. Furthermore, disturbance of the area near the #4 soil probes (temperature, soil moisture, and heat flux) caused all three to be incorrect for the following periods: 30 minute 1530 GMT 15 minute 1515 and 1530 GMT 5 minute 1515, 1520, 1525, and 1530 GMT. Complete recovery by all three sensors did not occur for one day (which is fairly typical for near surface installations), although the differences from the other four sets of soil sensors did not deleteriously affect the averages of the soil quantities. |
Measurements: | sgp15ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/22/1994 | 2200 | 05/25/1994 | 1645 |
Subject: | SGP/EBBR/E26 - Wind Speed Malfunction |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | For an unknown reason, the wind speed sensor at Cement began to record very low wind speeds on a consistent basis, with most of the speeds recorded reflecting what appeared to be zero wind speed (calibration offset). When the sensor was replaced on 5/25/94, it was damaged, although this damage apparently did not occur until a thunderstorm around midnight (local time) May 24. The wind direction sensor was also damaged by the storm at the same time. At any rate, the wind speed data during the period listed should be considered questionable. The wind speed sensor was replaced on May 25, 1994 at 1639. Old sensor S/N K1032, New sensor S/N K1035. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/25/1994 | 0430 | 05/25/1994 | 1640 |
Subject: | SGP/EBBR/E26 - Wind Direction Sensor Failure |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | Apparently, a fairly severe thunderstorm passed over the Cement, OK EBBR early on May 25, 1994 (around midnight May 24, local time). The storm damaged the wind direction sensor and the wind speed sensor. Wind Direction and Sigma of Wind Direction data for the period listed should be considered incorrect. The sensors were replaced on May 25, 1994 at 1631. Old sensor S/N K3038, New sensor S/N K3040. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/03/1994 | 1200 | 06/08/1994 | 1515 |
Subject: | SGP/EBBR/E12 - Wind Direction |
DataStreams: | sgp30ebbrE12.b1, sgp5ebbrE12.b1, sgp15ebbrE12.b1 |
Description: | For much of the period above, a malfunction in the E12 EBBR wind direction sensor (insect larvae shorting out the circuit board) caused incorrect values for 5 minute res_ws, wind_d, and sigma_wd, 15 minute mv_wind_d, and 30 minute res_ws, wind_d, and sigma_wd. Wind direction normally indicated north (near 0 or 360 degrees); there were some times during the period when wind directions looked good, but the lack of variation in direction and sigma implies that the data is suspect. Therefore, it would be wise to flag all of the values listed above during the entire time period as incorrect. The wind direction sensor (S/N 3081) was replaced with S/N 3042 at 1515 GMT on June 8, 1994. |
Measurements: | sgp30ebbrE12.b1:
sgp15ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/06/1994 | 1600 | 06/09/1994 | 1545 |
Subject: | SGP/EBBR/E20 - Soil Moisture Probe |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1 |
Description: | Beginning on May 6, 1994, soil moisture probe #3 began a long decline in output and sensitivity until it stabilized at a low value of around 7, with virtually no sensitivity at midday (GMT) on May 22, 1994. The output remained around this low value until the probe was replaced on June 9, 1994 at 1545 GMT. Old Probe S/N 92038, New Probe S/N 93003. |
Measurements: | sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/17/1994 | 1900 | 06/17/1994 | 1930 |
06/18/1994 | 0200 | 06/25/1994 | 1930 |
06/25/1994 | 2300 | 06/28/1994 | 1600 |
06/15/1994 | 2100 | 06/15/1994 | 2130 |
06/15/1994 | 2300 | 06/16/1994 | 0030 |
06/16/1994 | 2130 | 06/17/1994 | 1530 |
Subject: | SGP/EBBR/E9 - Reprocess: Soil Heat Flow Sensor Failure |
DataStreams: | sgp30ebbrE9.b1, sgp15ebbrE9.b1 |
Description: | Failure of soil heat flow sensor #5 at E9, Ashton, KS resulted in incorrect data values for 6 fields for much of the period of time listed above. The soil heat flow values from #5 were so large (in the absolute) that it deleteriosly affected the soil heat flux average and the sensible and latent heat fluxes. The fields affected were: 30 minute: shf5, c_shf5, g5, ave_shf, e, h 15 minute: mv_hft5. I list only the 30 minute times affected below; it can be assumed that the 15 minute value contained in the half hour are incorrect. Sensor #5 was replaced at 1545 GMT on June 28. Old S/N 923029, new S/N 923092. Half hour 1600 GMT is missing due to having to reload the CR10 program with a new shf5 calibration. Sensible and latent heat fluxes can be recomputed by averaging soil heat flow sensors 1-4; use the following equations: ave_shf = (g1 + g2 + g3 + g4)/4 e = -(q +ave_shf)/(1 + bowen) h = e * bowen, where g is total soil heat flow, q is net radiation, and bowen is Bowen Ratio. Affected half hours: Date Times (GMT) 6/15 2130, 2300, 2330 6/16 0000, 0030, 2130-2330 6/17 0000-1530, 1900, 1930 6/18 0200-2330 6/19 All 6/20 All 6/21 All 6/22 All 6/23 All 6/24 All 6/25 0000-1930, 2300, 2330 6/26 All 6/27 All 6/28 0000-1600 |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/18/1994 | 2000 | 06/22/1994 | 1600 |
Subject: | SGP/EBBR/E12 - Aspirator Failure |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | On June 18, 1994 the aspirator fan in the EBBR at E12, Pawhuska stopped working. This situation was not discovered until a Preventative Maintenance visit by site operations personnel on June 22, 1994. I determined the time of the beginning of the malfunction from close inspection of the sensible and latent heat flux data. There is no aspiration status output for the EBBRs, nor is one presently possible. Wind speeds during most of the period listed were sufficient to provide some (although inadequate) aspiration to the temperature and relative humidity probes, making it difficult for most users of the data to be able to tell (except in a few circumstances) that the sensible and latent heat flux values produced were incorrect. All of the data fields listed below are incorrect for the period listed above. Latent and Sensible heat fluxes cannot be recomputed, as the measurements that they are based on are what was corrupted. 30 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, bowen, e, h 15 minute: rr_thum_r, rr_thum_l, mv_hum_r, mv_hum_l, tair_r, tair_l 5 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/28/1994 | 1700 | 06/29/1994 | 1700 |
Subject: | SGP/EBBR/E7 - Reprocess: Soil Temperature Probe Failure |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | Failure of soil temperature probe #4 at E7, Elk Falls, KS resulted in incorrect data values for 6 fields for the period of time listed above. The soil temperature value from probe #4 went offscale, resulting in a zero value for the soil energy storage term calculated from that probe and thus no contribution to the total soil heat flux for soils set 4 and the same for average soil heat flux. This led to incorrect sensible and latent heat flux values. The fields affected were: 30 minute: ts4, ces4, g4, ave_shf, e, h 15 minute: rr_ts4. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/22/1994 | 1200 | 08/04/1994 | 1600 |
Subject: | SGP/EBBR/E12 - Stopped Ventilation Fan |
DataStreams: | sgp30ebbrE12.b1, sgp5ebbrE12.b1, sgp15ebbrE12.b1 |
Description: | On August 4, 1994, a site operator discovered the ventilation fan of the EBBR to be stopped; the fan guard was clogged with animal remains. Inspection of the EBBR data from Pawhuska, OK, E12 indicates that this condition occurred from July 22 through August 4. Lack of ventilation of the temperature and relative humidity sensors attached to the AEM results in questionable measurements of air temperature, relative humidity, and all quantities calculated from them, including sensible and latent heat fluxes. The quantities that are questionable for the period indicated are: 5 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, and vp_bot 15 minute: rr_thum_r, rr_thum_l, mv_hum_r, mv_hum_l, tair_r, tair_l 30 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, bowen, e, h |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/31/1994 | 2100 | 08/17/1994 | 1700 |
Subject: | SGP/EBBR/E26 - Soil Moisture Sensor Malfunction |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1 |
Description: | Soil moisture sensor #1 at Cement, Ok, E26 began a slow drift downwards on July 31, 1994. This continued until the sensor was replaced at 1700 GMT on August 17, 1994. The downwards drift occured coincident with a constant decrease in soil moisture at the site. As a result, the average soil moisture (for the set of five sensors) was reduced by only about 2% (for a range of 10 to 22%), and thus had a negligible effect on the calculations of sensible and latent heat flux. The sm1 output dropped below zero at times during August 14-17. Soil Moisture #1 Data Fields: 15 minute: r_sm1 30 minute: sm1 |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/19/1994 | 0100 | 08/23/1994 | 1530 |
Subject: | SGP/EBBR/E9 - Malfunctioning Wind Dir Sensor, Incorrect SM2 Results |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Description: | It has been reported previously that a malfunction of the wind direction sensor resulting in zero or 135 degree outputs can cause the soil moisture #2 sensor to read offscale. The two sensors are attached to a multiplexer pair. The manufacturer of the multiplexer and data acquisition system have not been able to explain how our instrument configuration can cause this effect. Shortly after 0030 GMT on August 20, 1994 significant rainfall occurred, being reflected in more than a doubling of soil moisture within an hour. During the time period listed above, the wind direction sensor mulfunctioned (apparently caused by water in the wind direction sensor electronics), causing the following data fields to be incorrect: 5 minute: res_ws, wind_d, sigma_wd 15 minute: mv_wind_dir 30 minute: res_ws, wind_d, sigma_wd. During part of the listed period, Soil Moisture #2 was also incorrect, being offscale; during those times the following 30 minute data is also incorrect: 30 minute: sm2, c_shf2, ave_shf, cs2, ces2, g2, h, e. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/17/1994 | 1700 | 08/17/1994 | 2300 |
Subject: | SGP/EBBR/E9 - Reprocess: Soil Heat Flux Out of Range |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1 |
Description: | For an unknown reason, soil heat flux sensor #5 indicated too small a value for the period listed above. This resulted in an incorrect average of soil heat flow for the five sets of soil heat flow plates, and therefore incorrect e (latent heat flux) and h (sensible heat flux) in the 30 minute data. The following data values are incorrect: 15 minute: mv_hft5 30 minute: shf5, c_shf5, ave_shf, g5, e, h The latent and sensible heat flux can be recalculated by using only soil heat flows 1 through 4 to calculate the average soil heat flow, ave_shf = (g1 + g2 + g3 + g4)/4 e = -(q + ave_shf)/(1 + bowen) h = -(e + ave_shf + q). |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/19/1994 | 1900 | 08/24/1994 | 1600 |
Subject: | SGP/EBBR/E8 - Battery Voltage Drop |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | Sometime after 0000 GMT on August 19, 1994 the cable from the solar panel to the EBBR battery was broken, presumably the result of severe weather that was in the area. The result was that the battery could not be charged by the solar panel, leading to a gradual reduction in battery voltage. By 1730 GMT on the 19th, the battery voltage had dropped to 10.8 volts and at about 1905 GMT had dropped too low to power the AEM. Some thirty minute data after 1900 GMT is therefore incorrect. By 2000 GMT the voltage had dropped to 10.7 volts and was too low to properly power the barometric pressure sensor (it's output dropped off from then on). At 1530 GMT on August 24, 1994 the solar panel connection to the battery was reestablished and normal data acquisition resumed at 1600 GMT. The last data file to be produced was apparently 0000 GMT (15 minute file) on 20 August. In an apparently related incident, soil temperature #3 skyrocketed in output beginning at about 1815 GMT; the cause for this is unknown, but is probably a result of the inability of the CR10 data logger to correctly measure the sensor resistance. By 1900 GMT this situation had caused elevated average soil heat flux, which in turn affected calculated e and h. Data fields that are incorrect for the time period listed, follow: 5 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot 15 minute: rr_thum_r, rr_thum_l, mv_hum_r, mv_hum_l, tair_r, tair_l 30 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, bowen, e, h |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/20/1994 | 0130 | 08/23/1994 | 1600 |
Subject: | SGP/EBBR/E9 - Hail Damage to Net Radiometer and Wind Speed Sensor |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Description: | Golf ball size hail was reported for the Ashton site for early August 20, 1994. It damaged the wind speed sensor and the net radiometer shortly before 0130 GMT. The following data fields are to be flagged as incorrect for the periods listed above (resultant wind speed has already been indicated as incorrect in a previous DQR that documented the malfunction of the wind direction sensor): 5 minute: wind_s, res_ws, q 15 minute: wind_s, mv_q 30 minute: wind_s, res_ws, q, h, e. Note that h and e are also incorrect as a consequence of q being incorrect. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/09/1994 | 0000 | 08/23/1994 | 1922 |
Subject: | Stopped Ventilation Fan |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | DQR No: Platform: EBBR Subject: Stopped Ventilation Fan Date Submitted: 30 August 1994 Submitted By: _x_ Instrument Mentor ___ EST Member ___ Science Team Member ___ Other _____________________________ For questions or problems, please contact the ARM Experiment Center at 509-375-6898 or via email at dqr@arm.gov. Platform/Measurement: What level data: (raw,a0,a1,b1,c1 etc): a0 (5 minute), a1 (15 and 30 minute) What location was the data collected at: E7, Elk Falls, KS Period of time in question Begin Date 8/9/94 End Date 8/23/94 Time 1922 GMT Data should be labeled: ___ questionable ___ All data fields affected ___ incorrect _x_ Only some data fields affected ___ wrong calibration _x_ others (information only) Discussion of Problem: Sometime between August 9 and 23, 1994 the ventilation fan of the EBBR stopped; the fan guard was clogged with insect remains. The fan was restarted at 1922 GMT on 23 August. Inspection of the EBBR data does not indicate any ill affects on the calculated sensible and latent heat during the period. The fan may have stopped soon before the arrival of the site operator on 23 August. I cannot suggest flagging the data as incorrect, but this message should still be attached to the data so that anyone who might discover something unusual may have some basis for explaining it. The quantities that are normally affected by stoppage of the ventilation to the temperature and relative humidity sensors are: 5 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, and vp_bot 15 minute: rr_thum_r, rr_thum_l, mv_hum_r, mv_hum_l, tair_r, tair_l 30 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, bowen, e, h Other observations/measurements impacted by this problem: see list above Suggested Corrections of the Problem: (e.g. change calibration factor and recompute, flag data with this comment, etc.) Enclose the message above with the data if possible. Data Processing Notes Date ------------------------------------------------------------------- END |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/20/1994 | 0230 | 08/31/1994 | 1542 |
Subject: | SGP/EBBR/E12 - Miscellaneous Problems Affecting Data |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | Three hardware problems occurred at E12, Pawhuska, OK in late August 1994. On August 12 the wind direction sensor malfunctioned. Therefore, all wind direction, sigma, and resultant wind speed values for the period above are incorrect. The wind direction sensor was replaced on August 31. Affected data are: 5 minute: wind_d, sigma_wd, res_ws 15 minute: mv_wind_d 30 minute: wind_d, sigma_wd, res_ws. Water accumulated in the AEM motor box and apparently caused a reduction in the voltage level of the home signal (battery condition was very good and so was not the cause); by late August 27 the hom30 level had fallen below the minimum qc check level. The AEM continued to exchange properly. The minimum qc level check is intended to help detect either a non-operating AEM or low battery condition. The Campbell CR10 software operates properly unless the hom30 signal is thirty-five or greater. So the low hom30 values do not result in any incorrect data!!!. I point this out because this situation demonstrates a condition flagged by the qc checks, even though no effect on the data has occurred. The AEM box was dried out and new dessicant placed in it on August 31. On August 28, 1994 the left AEM housing ventilation fan ceased to operate because of clogging of the ventilation system with debris. Fifteen minute data seems to indicate that the fan stopped operating just before 1245 GMT. This is at least the third such incidence of clogging this summer at different EBBRs. The clog was removed and ventilation restored on August 31. The lack of ventilation caused improper measurements of temperature and relative humidity in the left housing (these are quite obvious in the 15 minute data but difficult to detect in the 30 minute data) and therefore incorrect values of sensible and latent heat. Affected data are: 5 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot 15 minute: rr_thum_l, mv_hum_l, tair_l 30 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, bowen, e, h. |
Measurements: | sgp30ebbrE12.b1:
sgp15ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/01/1994 | 1700 | 09/01/1994 | 2230 |
09/03/1994 | 1530 | 09/07/1994 | 1700 |
Subject: | SGP/EBBR/E8 - Data Corruption and Loss |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | Data loss was experienced during the periods indicated for unknown reasons. My understanding is that ingestion of the data failed for these periods because the record length was shorter than it is supposed to be. Site Operations personnel visited the site on September 7 and powered down and then up, the EBBR, essentially causing a reload of the CR10 program. All output locations had been -9999 or 0000 when they had arrived. The battery condition was normal. Failure of a portion of the storage module may be indicated; data sneakernetted from the storage module shows an alternating sequence of two days, September 2 and 3 and then a stretch of garbage characters. No other days of data are found. The portion of the storage module memory that contained the program was not corrupted, as the CR10 program compiled and ran properly when it was reloaded from the storage module. The data after the period above is reasonable. The last part of the storage module memory may be bad and perhaps in trying to access it remotely multiple times the situation above resulted. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/12/1994 | 1300 | 09/14/1994 | 1847 |
Subject: | SGP/EBBR/E20 - Fan Stopped, T and RH Biases |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | Another fan stoppage situation occurred. This is the fourth one in the last few months and may indicate a weakening of the fans that aspirate the temperature, relative humidity housings. The 15 minute data reveals that the stoppage occurred at approximately 1300 GMT on September 12. The left fan was cleaned out by site operations personnel and began running again at 1847 GMT on September 14. The 15 minute data during this period shows that during daytime, the left temperature was greater than the right temperature and that the left relative humidity was lower than the right relative humidity (a consequence of greater temperature), irregardless of the positions of the housings. The bowen ratio, sensible heat flux and latent heat flux for this period, as well as all left housing temperatures and relative humidities are incorrect. The data values that are incorrect are: 5 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot 15 minute: rr_thum_l, mv_hum_l, tair_l, 30 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, thum_bot, vp_top, vp_bot, bowen, e, h Because of the increased frequency of fan stoppage occurrences, I have submitted a work request to site operations to use the portable vane anemometer to check the flow of the housing fans on each EBBR and visually check (and clean if needed) the fans for debris that would restrict the flow. Included in this work request are instructions for checking the performance of the temperature and relative humidity sensors by manually putting the aspirator housings at the same height for a half hour period (so that two 15 minute and one half hour comparison result). The latter part of the work request is prompted by my close inspection of the Meeker data for the last few weeks. It is clear that significant biases have developed in the pairs of temperature and relative humidity sensors; left thermocouple ~0.5 deg. C greater than right thermocouple, left relative humidity ~5% lower than right relative humidity. Biases were expected in operating the systems over long time periods, and these appear to be the result of simple offsets, not changes in calibrations. At this point, I just want to determine how large the offsets are. Comparison of these offsets between EBBR units, in light of the different lengths of time that EBBR units and individual sensors have been deployed, may provide some interesting information on likely offset changes with time. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/18/1994 | 1200 | 09/21/1994 | 1530 |
Subject: | Intermittent Stopped or Slowed Aspirator Fans |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | DQR No: Platform: EBBR Subject: Intermittent Stopped or Slowed Aspirator Fans Date Submitted: 26 September 1994 Submitted By: _x_ Instrument Mentor ___ EST Member ___ Science Team Member ___ Other _____________________________ For questions or problems, please contact the ARM Experiment Center at 509-375-6898 or via email at dqr@arm.gov. Platform/Measurement: EBBR/T and RH What level data: (raw,a0,a1,b1,c1 etc): a0 (5 minute), a1 (15 and 30 minute) What location was the data collected at: E7, Elk Falls, KS Period of time in question Begin Date 8/18/94 Time 1200 GMT End Date 9/21/94 Time 1530 GMT Data should be labeled: ___ questionable ___ All data fields affected _X_ incorrect _x_ Only some data fields affected ___ wrong calibration ___ others (information only) Discussion of Problem: Around 1200 GMT on September 21, 1994 the aspirator fans of the EBBR intermittently stopped or slowed significantly; the fans were clogged with insect remains and spider webs. Only a few weeks earlier the same situation occurred at this EBBR. The 15 minute data reveals that most of the time the right aspirator fan was the one that failed to work, although there are times that it appears that the left fan was either not working also or was the only one not working. Unlike the previous occurrence, the 30 minute flux data does show some indication of this problem, although, as usual, it is far easier to see in the 15 minute data. The fans were cleaned out and restarted at 1530 GMT on September 21, 1994. The data listed below should be flagged as incorrect for the period shown above. 5 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, and vp_bot 15 minute: rr_thum_r, rr_thum_l, mv_hum_r, mv_hum_l, tair_r, tair_l 30 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, bowen, e, h Other observations/measurements impacted by this problem: see list above Suggested Corrections of the Problem: (e.g. change calibration factor and recompute, flag data with this comment, etc.) Enclose the message above with the data if possible. Data Processing Notes Date ------------------------------------------------------------------- END |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/29/1994 | 1530 | 09/29/1994 | 1600 |
Subject: | SGP/EBBR/E20 - Soil Moisture Probe Failure |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1 |
Description: | Soil moisture probe #4 went offscale for a short period, for unknown reasons. However, it may have resulted from Site Operations Preventative Maintenance activities that were taking place at the time. The times of incorrect values are listed below. 30 minute: 1600 GMT; sm4, c_shf4, cs4, ces4, g4, ave_shf, e, h 15 minute: 1600 GMT; r_sm4 |
Measurements: | sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/13/1994 | 1600 | 09/27/1994 | 1600 |
Subject: | SGP/EBBR/E22 - Net Radiometer Hemisphere Changes |
DataStreams: | sgp5ebbrE22.b1, sgp30ebbrE22.b1, sgp15ebbrE22.b1 |
Description: | The upper hemisphere of the net radiometer at the site was punctured allowing water to collect in the lower (and sometimes upper) section of the sensor. The punctures are most likely the result of migratory birds (a problem that we have experienced almost every year here at ANL also). The punctures (and water in some net radiometers) did not result in obvious invalidation of the data. Whether there is water in the lower hemisphere or not, a small puncture of the upper hemisphere would not have a major effect on the net radiation measurement. Significant amounts of water in the lower hemisphere should have a major effect on nighttime measurements particularly, although that effect is hard to detect in the data. The date when the puncture occurred is approximate - but is certainly no more than two weeks before the hemisphere was replaced. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/14/1994 | 1600 | 09/28/1994 | 1515 |
Subject: | SGP/EBBR/E26 - Net Radiometer Hemisphere Changes |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | The upper hemisphere of the net radiometer at the site was punctured allowing water to collect in the lower (and sometimes upper) section of the sensor. The punctures are most likely the result of migratory birds (a problem that we have experienced almost every year here at ANL also). The punctures (and water in some net radiometers) did not result in obvious invalidation of the data. Whether there is water in the lower hemisphere or not, a small puncture of the upper hemisphere would not have a major effect on the net radiation measurement. Significant amounts of water in the lower hemisphere should have a major effect on nighttime measurements particularly, although that effect is hard to detect in the data. The date when the puncture occurred is approximate - but is certainly no more than two weeks before the hemisphere was replaced. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/15/1994 | 1600 | 09/29/1994 | 1430 |
Subject: | SGP/EBBR/E12 - Net Radiometer Hemisphere Changes |
DataStreams: | sgp5ebbrE12.b1, sgp30ebbrE12.b1, sgp15ebbrE12.b1 |
Description: | The upper hemisphere of the net radiometer at the site was punctured allowing water to collect in the lower (and sometimes upper) section of the sensor. The punctures are most likely the result of migratory birds (a problem that we have experienced almost every year here at ANL also). The punctures (and water in some net radiometers) did not result in obvious invalidation of the data. Whether there is water in the lower hemisphere or not, a small puncture of the upper hemisphere would not have a major effect on the net radiation measurement. Significant amounts of water in the lower hemisphere should have a major effect on nighttime measurements particularly, although that effect is hard to detect in the data. The date when the puncture occurred is approximate - but is certainly no more than two weeks before the hemisphere was replaced. |
Measurements: | sgp30ebbrE12.b1:
sgp15ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/26/1994 | 1530 | 10/11/1994 | 1910 |
Subject: | SGP/EBBR/E15 - Net Radiometer Hemisphere Changes |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | The upper hemisphere of the net radiometer at the site was punctured allowing water to collect in the lower (and sometimes upper) section of the sensor. The punctures are most likely the result of migratory birds (a problem that we have experienced almost every year here at ANL also). The punctures (and water in some net radiometers) did not result in obvious invalidation of the data. Whether there is water in the lower hemisphere or not, a small puncture of the upper hemisphere would not have a major effect on the net radiation measurement. Significant amounts of water in the lower hemisphere should have a major effect on nighttime measurements particularly, although that effect is hard to detect in the data. The date when the puncture occurred is approximate - but is certainly no more than two weeks before the hemisphere was replaced. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/29/1994 | 1530 | 09/29/1994 | 1730 |
Subject: | SGP/EBBR/E20 - Aspirator Fan Stoppage |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1 |
Description: | During Site Operations Preventative Maintenance activities the left AEM fan stopped operating. This affected some of the data being collected. The times of incorrect values are listed below. 15 minute: 1530-1715 GMT; tair_l, mv_hum_l, rr_thum_l 30 minute: 1530-1730 GMT; thum_top, thum_bot, tair_top, tair_bot, hum_top, hum_bot, e, h, bowen |
Measurements: | sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/13/1994 | 2300 | 10/14/1994 | 1900 |
Subject: | SGP/EBBR/E13 - Soil Temperature Aberations |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
Description: | For a period of not quite a day, soil temperature probe #4 indicated values that were either too high or too low, for unknown reasons. After Site Operations personnel checked connectors and the probe itself, it recovered to proper values. The probe has functioned properly since then. The times and incorrect data fields are listed below. 15 minute: rr_ts4, 10/13/94: 2245 GMT; 10/14/94: 0000, 0200, 0230, 0515, 0600, 0630, 1300-1500, 1530-1715, 1815, 1845 GMT 30 minute: ts4, 10/13/94: 2300 GMT; 10/14/94: 0000, 0230, 0530-0630, 1300-1700, 1830-1900 GMT ces4, g4, ave_shf, e, h, 10/14/94: 0030, 0200-0230, 0600-0630, 1300-1900 GMT |
Measurements: | sgp15ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/03/1994 | 1900 | 09/15/1994 | 1330 |
Subject: | SGP/EBBR/E12 - Low home_15 Values |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | The home signal values for the EBBR system installed at E12, Pawhuska, OK have generally been lower than for the other EBBR systems. However, during the period above the home_15 value was often lower than 35, which is then improperly interpreted by the CR10 software. A home_15 value less than 35 is interpreted as meaning that the AEM aspirated housings (which hold the temperature and relative humidity sensors) are in the opposite orientation to what they actually are. The AEM continues to exchange properly, but the resulting calculations of Bowen ratio, latent heat flux, and sensible heat flux are incorrect. On September 15, 1994 Site Operations personnel refurbished the wiring for the left housing and the home_15 signal returned to proper levels. When 30 minute home_15 values are below 35, the data fields listed below are incorrect. 5 minute: home, tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot 15 minute: mv_home 30 minute: home_15, tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, Bowen, e, h It is possible to recalculate the Bowen ratio, e, and h by using the 15 minute data to compute the correct gradients of temperature and relative humidity (two 15 minute periods are averaged to give a half hour value); the orientation of the AEM is obvious in the 15 minute data, based on the home signal (which is largest for the first 15 minutes of the half hour). More information on performing this recalculation can be obtained from the mentor. |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/29/1994 | 1715 | 09/29/1994 | 1800 |
Subject: | SGP/EBBR/E20 - Net Radiometer Hemisphere Changes |
DataStreams: | sgp30ebbrE20.b1, sgp15ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | Incorrect data during hemisphere change. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/11/1994 | 1915 | 10/11/1994 | 1930 |
Subject: | SGP/EBBR/E15 - Net Radiometer Hemisphere Changes |
DataStreams: | sgp30ebbrE15.b1, sgp15ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | Incorrect data during hemisphere change. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/14/1994 | 1600 | 09/29/1994 | 1735 |
Subject: | SGP/EBBR/E20 - Net Radiometer Hemisphere Changes |
DataStreams: | sgp30ebbrE20.b1, sgp15ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | The upper hemisphere of the net radiometer at the site was punctured allowing water to collect in the lower (and sometimes upper) section of the sensor. The punctures are most likely the result of migratory birds (a problem that we have experienced almost every year here at ANL also). The punctures (and water in some net radiometers) did not result in obvious invalidation of the data. Whether there is water in the lower hemisphere or not, a small puncture of the upper hemisphere would not have a major effect on the net radiation measurement. Significant amounts of water in the lower hemisphere should have a major effect on nighttime measurements particularly, although that effect is hard to detect in the data. The date when the puncture occurred is approximate - but is certainly no more than two weeks before the hemisphere was replaced. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/21/1994 | 0300 | 11/02/1994 | 1715 |
Subject: | SGP/EBBR/E8 - Water in Net Radiometer |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | At some time previous to October 21, 1994, the top hemisphere of the net radiometer of the EBBR at E8 was pierced, probably by migrating birds. A DQR was issued by me a couple of weeks ago indicating other extended facilities at which this has happened. Net radiometer data was correct until October 21 at sometime between 0230 and 0300 GMT, when rain began to enter the radiometer; the effect can be easily seen in the data. Collection of 3/4 inch of water in the bottom hemisphere was not the only consequence of the situation. Unlike other net radiometers that have had this experience, this one produced incorrect data during daylight hours as well as during nightime hours until the water was removed, the sensor dried, and the hemispheres were replaced on November 2, 1994. During the period listed above, the following data are incorrect and cannot be recalculated: 5 minute: q 15 minute: mv_q 30 minute: q, h, e Science Team members or others using the data should recognize that the flux data is not correct for this fairly lengthy time period. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/20/1994 | 0000 | 11/01/1994 | 1540 |
Subject: | SGP/EBBR/E7 - Net Radiometer not level |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | Sometime between 0000 and 1200 GMT on October 20, 1994, the net radiometer support of the EBBR at E7 was bent down at about a 30 degree angle, probably by migrating birds. A couple of DQRs have been issued by me indicating other extended facilities at which birds have caused incorrect data from net radiometers. The net radiometer support was probably loose to begin with. The effect can be easily seen in the data, when comparison is made to the EBBR data at Ashton, KS. Since the net radiometer support extends south from the EBBR, the net radiation values are high during the period above, leading to incorrect values of latent and sensible heat fluxes. Until the net radiometer was levelled on November 1, 1994, the net radiation was approximately 26% high at solar noon, although it may have been incorrect by a smaller percentage at night. From October 22 through 28 the half hour of data nearest solar noon averaged 497 watts per meter squared at Elk Falls and only 396 at Ashton. During the period listed above, the following data are incorrect and cannot be recalculated: 5 minute: q 15 minute: mv_q 30 minute: q, h, e Science Team members or others using the data should recognize that the flux data is not correct for this fairly lengthy time period. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/02/1994 | 1937 | 12/02/1994 | 2000 |
Subject: | SGP/EBBR/E13 - Net Radiation - Hemisphere Replacement |
DataStreams: | sgp30ebbrE13.b1, sgp5ebbrE13.b1, sgp15ebbrE13.b1 |
Description: | The upper hemisphere of the net radiometer was punctured (pinhole) and thus was replaced at 1937 GMT on December 2, 1994. This resulted in a spike in the net radiation data. 5 minute q data for 1935, 1940, 1945, and 1955 GMT, 15 minute mv_q data for 1945 and 2000 GMT, and 30 minute q, e, and h data for 2000 GMT are incorrect. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/15/1994 | 2300 | 11/30/1994 | 1600 |
Subject: | SGP/EBBR/E7 - Radiation Shield Problem |
DataStreams: | sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | The right radiation shield was out of position or off during the period of time indicated. This may or may not have occurred as a result of preventative maintenance on November 15. All air and humidity probe temperatures (except the reference temperature), relative humidities, vapor pressures, Bowen ratio, sensible heat flux, and latent heat flux during this period are questionable. During a large part of the period the resultant sensible and latent heat fluxes look reasonable. However, during part of the period, the difference in relative humidity between the two EBBR levels appears to be much too large. Unusual differences in temperatures are not as obvious. |
Measurements: | sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/07/1994 | 0430 | 12/07/1994 | 0800 |
Subject: | SGP/EBBR/E26 - Wind Direction Sensor Affect on Soil Moisture 2 |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | I have written a number of previous PIFs and DQRs about this same problem previously. This occurred most recently at E9, Ashton, KS in August 1994. The wind direction sensor is on the same data acquisition system multiplexer pair as soil moisture sensor 2. Sometimes a malfunctioning wind direction sensor (resulting in a constant wind direction of 135 degrees) will cause the soil moisture 2 measurement to be offscale. Between 0430 and 0800 GMT the wind direction sensor did not function properly at E22, Cordell, KS. As has often been the case, soil moisture measurements were affected for a shorter time period, from 0500 through 0730 GMT. The cause of this problem is unknown and only occurs on the multiplexer pair to which the wind direction sensor is connected, not on any other pair to which a soil moisture sensor is attached. None of the manufacturers of the applicable equipment can explain the problem. As done before, I will monitor the data for further occurrences of this condition; if the frequency of occurrence increases, further action will be taken. The following data and times are incorrect: Winds: 0430-0800 GMT 5 minute: wind_d, sigma_wd, res_ws 15 minute: mv_wind_d 30 minute: wind_d, sigma_wd, res_ws Soil Moisture 2: 0500-0730 GMT 15 minute: r_sm2 30 minute: sm2, c_shf2, cs2, ces2, g2, ave_shf, h, e |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/06/1994 | 1200 | 12/09/1994 | 1530 |
Subject: | SGP/EBBR/E8 - Frozen Wind Instruments |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | On December 6, 1994 a frontal system deposited freezing rain in central Kansas that resulted in the EBBR wind instruments becoming frozen in place at two extended facilities (E4 and E8) for a few days. The direction and speed instruments were frozen for slightly differents lengths of time. The times and fields to be indicated as incorrect are shown below (all 5, 15, and 30 minute data measured during the periods, inclusive, is incorrect): Wind Speed: 12/06/94 1200 GMT - 12/09/94 1530 GMT 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws Wind Direction: 12/06/94 1730 GMT - 12/07/94 1830 GMT 5 minute: wind_d, sigma_wd 15 minute: mv_wind_d 30 minute: wind_d, sigma_wd |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/14/1994 | 1600 | 12/14/1994 | 1630 |
Subject: | SGP/EBBR/E7 - Water in Net Radiometer |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | A small amount of water collected in the bottom dome of the net radiometer. It is not obvious that it had been there for more than several hours before the site operators arrived for a preventative maintenance visit and there is no evidence of any effects on the data (there had been heavy cloudiness for the previous several hours). The only incorrect data occurred during the removal of the water. The dome was not replaced. The following data is incorrect between 1600 and 1630 GMT on December 14, 1994: 5 minute: q 15 minute: mv_q 30 minute: q, e, h |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/09/1994 | 1615 | 11/09/1994 | 1800 |
Subject: | SGP/EBBR/E26 - T/RH Check |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | During the month of November 1994, site operations personnel conducted field comparisons of the temperature and relative humidity sensors in the EBBRs and the HMI31 portable T/RH meter (with the removable probe). As is reflected in the home signal output, the automatic exchange mechanism was disabled in order to conduct the checks. Measurements were taken every 5 minutes, for a half hour each in the aspirated radiation shields (the HMI31 was inserted into the EBBR aspirated radiation shield to provide the same aspiration as the EBBR sensors received). Measurements of aspirator flow were also recorded. This information was provided to the mentor. The mentor then compared the field checks with the data recorded by the EBBRs. The periods of data and data fields that are incorrect because of the checks are listed below. Times are those at which the data are reported. Data Fields 5 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot 15 minute: rr_thum_r, rr_thum_l, mv_hum_r, mv_hum_l, tair_r, tair_l 30 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, bowen, e, h 5 minute: 1615, 1620, 1625, 1630, 1635, 1640, 1645, 1650, 1655, 1700, 1705, 1710, 1715, 1720, 1725, 1730, 1735 15 minute: 1615, 1630, 1645, 1700, 1715, 1730, 1745 30 minute: 1630, 1700, 1730, 1800 (EDITOR'S NOTE: The following analysis refers to all EBBRs installed in Nov 1994) Results of the T/RH comparisons: The combined accuracy of the HMI31 and EBBR temperature sensors (thermocouples or PRTDs) is +/- 0.4 degrees C. The combined accuracy of the HMI31 and EBBR RH sensors is +/- 4% for RH less than 80%, and +/- 6% for RH greater than 80%. The combined accuracy for each pair of EBBR sensors of the same type is essentially the same as the quantities stated above. All of the pairs of EBBR sensors show differences that fall within the stated combined accuracies. Although this sounds wonderful, a few of the differences are close to the combined accuracy, which, because of the small differences in temperature being measured (normally smaller than the magnitude of the combined accuracy) could lead to some incorrect estimation of sensible and latent heat flux. However, unless there would be substantial differences of calibration slope of two sensors in a pair, the switching of height of the sensors every 15 minutes by the automatic exchange mechanism will remove the offset between the sensors. The comparisons between the EBBR sensors and the HMI31 T/RH meter showed much larger differences. The range of differences were: -2.4 to 0.0 degrees for the thermocouples, -2.1 to 0.1 degrees for the RH probe PRTDs, and -6.6 to 6.7 percent for the RH sensors. Six Thermocouples indicated differences from the HMI31 greater than the combined accuracy. Seven PRTDs indicated differences from the HMI31 greater than the combined accuracy. Four RH sensors indicated differences from the HMI31 greater than the combined accuracy. The temperature differences (for both thermocouples and PRTDs) were nearly all of one sign; the HMI31 indicated the greater temperature, which may suggest a bias in the HMI31 temperature measurement, even though a dozen differences of +/- 0.2 degrees were measured. Four differences of zero and only one positive difference (out of 40) were measured. Thirteen of the 20 RH differences were positive (HMI31 indicating the lower RH), with the differences near the level of the combined accuracies being approximately 50% positive and 50% negative. It does not appear that the HMI31 has a RH bias. Plots of EBBR sensor/HMI31 differences as a function of time from the installation of the individual sensors (a range of 10 to 30 months in the field) shows no obvious pattern of sensor drift with time. This is encouraging. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/18/1994 | 1630 | 12/18/1994 | 1930 |
12/19/1994 | 0000 | 12/19/1994 | 1600 |
12/20/1994 | 0200 | 12/20/1994 | 0400 |
12/23/1994 | 0830 | 12/23/1994 | 2030 |
12/24/1994 | 0800 | 12/25/1994 | 0100 |
12/25/1994 | 0200 | 12/25/1994 | 1400 |
12/25/1994 | 1500 | 12/25/1994 | 1600 |
12/25/1994 | 1700 | 12/26/1994 | 1800 |
12/27/1994 | 0200 | 12/27/1994 | 1700 |
12/30/1994 | 1730 | 12/30/1994 | 1800 |
12/13/1994 | 0430 | 12/13/1994 | 0500 |
12/13/1994 | 1800 | 12/13/1994 | 1830 |
12/18/1994 | 0900 | 12/18/1994 | 1030 |
Subject: | SGP/EBBR/E13 - Incorrect Soil Temperature #4 |
DataStreams: | sgp30ebbrE13.b1, sgp15ebbrE13.b1 |
Description: | For periods of a number of days in December 1994, soil temperature probe #4 indicated incorrect values that were either too high or too low. Site Operations personnel replaced the sensor on December 30, 1994. The times and incorrect data fields are listed below. Only half hour times are listed. The incorrect data on December 30 was the result of the process of replacing the sensor during that time. Data Fields: 15 minute: rr_ts4 30 minute: ts4, ces4, g4, ave_shf, e, h Dates and Times (GMT): 12/13/94 0500, 1830 12/18/94 0900-1030, 1630-1930 12/19/94 0000-1600 12/20/94 0200-0400 12/23/94 0830-2030 12/24/94 0800-2330 12/25/94 0000-0100, 0200-1400, 1530, 1600, 1700-2330 12/26/94 0000-1800 12/27/94 0200-1700 12/30/94 1730-1800 Recomputation of the change in energy storage in the soil could be done roughly from the soil temperature sensors that were still working (roughly because the difference in soil temperature with time is determined with real-time (one minute samples) data, not from the 15 or 30 minute data). However, this would require considerable effort. The process for doing so is therefore not stated here. Contact the mentor for more information if neccesary. |
Measurements: | sgp15ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/19/1994 | 1700 | 08/19/1994 | 2300 |
10/13/1994 | 1800 | 10/14/1994 | 1500 |
10/17/1994 | 2030 | 10/17/1994 | 2100 |
10/28/1994 | 2030 | 10/28/1994 | 2230 |
12/26/1994 | 1700 | 12/27/1994 | 0130 |
12/27/1994 | 1600 | 12/27/1994 | 1630 |
01/02/1995 | 0600 | 01/05/1995 | 1700 |
01/05/1995 | 2200 | 01/09/1995 | 0230 |
01/14/1995 | 0800 | 01/15/1995 | 0300 |
01/15/1995 | 1200 | 01/16/1995 | 1600 |
12/20/1994 | 0630 | 12/23/1994 | 0530 |
01/09/1995 | 0900 | 01/14/1995 | 0400 |
Subject: | SGP/EBBR/E9 - Reprocess: Soil Heat Flow Sensor 5 Incorrect Data |
DataStreams: | sgp30ebbrE9.b1, sgp15ebbrE9.b1 |
Description: | Soil heat flow sensor #5 periodically outputted incorrect values from mid-August 1994 until mid-January 1995. The output often indicated large negative heat flow. The sensor was replaced on January 23, 1995, even though the previous week of data was correct. The data affected (and thus incorrect) includes: 15 minute: mv_hft5 30 minute: shf5, c_shf5, g5, ave_shf, e, h. The dates and times (GMT) of incorrect values are listed below. Only half hour times are listed. 08/19/94 1700-2300 10/13/94 1800-2330 10/14/94 0000-1500 10/17/94 2030-2100 10/28/94 2030-2230 12/20/94 0630-2330 12/21/94 0000-2330 12/22/94 0000-2330 12/23/94 0000-0530 12/26/94 1700-2330 12/27/94 0000-0130, 1630 01/02/95 0600-2330 01/03/95 0000-2330 01/04/95 0000-2330 01/05/95 0000-1700, 2200-2330 01/06/95 0000-2330 01/07/95 0000-2330 01/08/95 0000-2330 01/09/95 0000-0230, 0900-2330 01/10/95 0000-2330 01/11/95 0000-2330 01/12/95 0000-2330 01/13/95 0000-2330 01/14/95 0000-0400, 0800-2330 01/15/95 0000-0300, 1200-2330 01/16/95 0000-1600 The latent and sensible heat fluxes can be recalculated by using only soil heat flows 1 through 4 to calculate the average soil heat flow, ave_shf = (g1 + g2 + g3 + g4)/4 e = -(q + ave_shf)/(1 + bowen) h = -(e + ave_shf + q) |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/31/1994 | 1330 | 01/01/1995 | 1745 |
Subject: | SGP/EBBR/E8 - Frozen Wind Speed Instrument |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | On December 31, 1994 EBBR wind speed instrument froze in place from icing. Wind speeds increased sufficiently on January 1, 1995 to break the anemometer free. The wind direction instrument was not affected. The times and fields that are incorrect are shown below. 5 minute: wind_s, res_ws 12/31 1335 GMT - 1/1 1745 GMT 15 minute: wind_s 12/31 1345 GMT - 1/1 1745 GMT 30 minute: wind_s, res_ws 12/31 1400 GMT - 1/1 1800 GMT |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/31/1994 | 2140 | 01/01/1995 | 1730 |
Subject: | SGP/EBBR/E15 - Frozen Wind Speed Instrument |
DataStreams: | sgp30ebbrE15.b1, sgp5ebbrE15.b1, sgp15ebbrE15.b1 |
Description: | On December 31, 1994 EBBR wind speed instrument froze in place from icing. Wind speeds increased sufficiently on January 1, 1995 to break the anemometer free. The wind direction instrument was not affected. The times and fields that are incorrect are shown below. 5 minute: wind_s, res_ws 12/31 2140 GMT - 1/1 1705 GMT 15 minute: wind_s 12/31 2145 GMT - 1/1 1715 GMT 30 minute: wind_s, res_ws 12/31 2200 GMT - 1/1 1730 GMT |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/21/1995 | 0030 | 02/21/1995 | 1600 |
Subject: | SGP/EBBR/E9 - Radiation Shield Problem |
DataStreams: | sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Description: | The left radiation shield was off during the period of time above. Close inspection of the 15 minute data was required to detect when the shield fell off. As is often the case in such a situation, the 30 minute data does not adequately indicate such a failure. A radiation shield has fallen off of an EBBR before, apparently a result of vibration of the AEM during exchanges. All air and humidity probe temperatures (except the reference temperature), relative humidities, vapor pressures, Bowen ratio, sensible heat flux, and latent heat flux during this period are considered to be incorrect. |
Measurements: | sgp30ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/16/1995 | 1500 | 02/17/1995 | 1230 |
Subject: | SGP/EBBR/E26 - Frozen Wind Speed Sensor |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | On February 16, 1994 a frontal system deposited freezing rain in central Oklahoma that resulted in the EBBR wind speed instrument becoming frozen in place at Cement (E26) for part of two days. All 5, 15, and 30 minute data values indicated below are incorrect for all times within the period indicated. 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/14/1995 | 1100 | 02/14/1995 | 1730 |
Subject: | SGP/EBBR/E13 - Frozen Wind Speed Sensor |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | DQR No: Platform: EBBR Subject: Frozen Wind Speed Sensor Date Submitted: JUNE 22, 1995 Submitted By: _x_ Instrument Mentor ___ EST Member ___ Science Team Member ___ Other _____________________________ For questions or problems, please contact the ARM Experiment Center at 509-375-6898 or via email at dqr@arm.gov. Platform/Measurement: EBBR/ Wind Speed What level data: (raw,a0,a1,b1,c1 etc): a1 (15 and 30 minute) a0 (5 minute) What location was the data collected at: E13, Central Facility Period of time in question Begin Date 2/14/95 Time 1100 GMT End Time 1730 GMT Data should be labeled: ___ questionable ___ All data fields affected _X_ incorrect _X_ Only some data fields affected ___ wrong calibration ___ others Discussion of Problem: On February 14, 1995 a frontal system deposited freezing rain in central Oklahoma that resulted in the EBBR wind speed instrument becoming frozen in place at the Central Facillity (E13) for part of a day. All 5, 15, and 30 minute data values indicated below are incorrect for the period listed above. 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws Other observations/measurements impacted by this problem: Suggested Corrections of the Problem: (e.g. change calibration factor and recompute, flag data with this comment, etc.) Flag data with this comment. Data Processing Notes Date ------------------------------------------------------------------- END |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/11/1995 | 1230 | 04/12/1995 | 0615 |
04/12/1995 | 1315 | 04/12/1995 | 1400 |
Subject: | SGP/EBBR/E26 - Stopped Ventilation Fan |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | On April 11, 1995, a site operator discovered the right aspirator ventilation fan of the EBBR to be stopped. Inspection of the data indicates that this condition occurred during the two periods shown above; the fan apparently worked briefly inbetween the periods. Lack of ventilation of the temperature and relative humidity sensors attached to the AEM results in incorrect measurements of air temperature and relative humidity by the right aspirator probes, and therefore all quantities calculated from them, including sensible and latent heat fluxes, are incorrect as well. The quantities that are incorrect for the periods indicated are: 5 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, and vp_bot 15 minute: rr_thum_r, mv_hum_r, tair_r 30 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, bowen, e, h |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/18/1995 | 0130 | 04/18/1995 | 1620 |
Subject: | SGP/EBBR/E9 - Stopped Ventilation Fan |
DataStreams: | sgp30ebbrE9.b1, sgp5ebbrE9.b1, sgp15ebbrE9.b1 |
Description: | On April 18, 1995, a site operator discovered the right aspirator ventilation fan of the EBBR to be stopped. Inspection of the data indicates that this condition began early in the GMT day. Interpretation of the start time of the condition is made more difficult by a slight bias low in absolute temperature in comparison to the left temperature probe. Lack of ventilation of the temperature and relative humidity sensors attached to the AEM results in incorrect measurements of air temperature and relative humidity by the right aspirator probes, and therefore all quantities calculated from them, including sensible and latent heat fluxes, are incorrect as well. The quantities that are incorrect for the period above are: 5 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, and vp_bot 15 minute: rr_thum_r, mv_hum_r, tair_r 30 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, bowen, e, h |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/18/1995 | 1130 | 04/18/1995 | 1530 |
Subject: | SGP/EBBR/E9 - Water in Net Radiometer |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Description: | The top dome of the net radiometer was broken and a small amount of water had collected in the bottom dome of the net radiometer. It is not obvious that it had been there for more than several hours before the site operators arrived for a preventative maintenance visit. There are noticable effects on the data for the period listed above. The domes were replaced. The following data is incorrect for the period above: 5 minute: q 15 minute: mv_q 30 minute: q, e, h |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/26/1995 | 1945 | 04/26/1995 | 2000 |
Subject: | SGP/EBBR/E26 - Reprocess: Soil Heat Flow Sensor 1 Replacement |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1 |
Description: | During preventative maintenance, site operations personnel observed that soil heat flow sensor #1 at Cement, OK (E26) was outputting larger values than the other four sensors. Therefore, they replaced sensor #1. A review of the data indicates that the sensor was not malfunctioning, but that its calibration was just too large. The longer the sensors are in the ground, the greater the likelihood that they will drift from their original calibration. The difference in the average soil heat flux caused by the #1 sensor data values being included in the average will have less than a 5% affect on the calculated latent and sensible heat fluxes. The data should therefore not be labelled as incorrect. However, during the replacement of sensor #1, the #1 data was spiked, causing the following data values to be incorrect for the period listed: 15 minute: mv_hft1 at 1945 GMT. 30 minute: shf5, c_shf5, g5, ave_shf, e, h, bowen at 2000 GMT. The latent and sensible heat fluxes can be recalculated by using only soil heat flows 2 through 5 to calculate the average soil heat flow, ave_shf = (g2 + g3 + g4 + g5)/4 e = -(q + ave_shf)/(1 + bowen) h = -(e + ave_shf + q). |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/23/1995 | 0020 | 04/26/1995 | 1630 |
Subject: | SGP/EBBR/E12 - Wind Direction Malfunction |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | For most of the period above, a malfunction in the E12 EBBR wind direction sensor caused the following data values to be incorrect: 5 minute: res_ws, wind_d, and sigma_wd 15 minute: mv_wind_d 30 minute: res_ws, wind_d, and sigma_wd. Wind direction normally indicated near north or 135 degrees (the latter indicates a malfunction). The wind direction sensor (S/N 3036) was replaced with S/N 3034 at 1620 GMT on April 26, 1995. |
Measurements: | sgp30ebbrE12.b1:
sgp15ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/06/1995 | 0700 | 06/07/1995 | 1830 |
Subject: | SGP/EBBR/E12 - Wind Direction Malfunction |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | For the period indicated, a malfunction in the E12 EBBR wind direction sensor caused the following data values to be incorrect: 5 minute: res_ws, wind_d, and sigma_wd 15 minute: mv_wind_d 30 minute: res_ws, wind_d, and sigma_wd. Wind direction normally indicated smaller or larger than it should have. The affected time period was determined from comparison with data from E7, Elk Falls EBBR, which is located north of E12. The wind direction sensor (S/N 3034) was replaced with S/N 3039 at 1822 GMT on June 7, 1995. |
Measurements: | sgp30ebbrE12.b1:
sgp15ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/09/1995 | 1705 | 05/16/1995 | 1900 |
05/24/1995 | 1930 | 05/28/1995 | 0100 |
05/28/1995 | 0130 | 06/01/1995 | 1630 |
06/05/1995 | 1730 | 06/05/1995 | 1900 |
06/06/1995 | 1530 | 06/06/1995 | 1900 |
Subject: | SGP/EBBR/E26 - Right Air Temperature Malfunction |
DataStreams: | sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp15ebbrE26.b1 |
Description: | The right air temperature probe malfunctioned continuously during the periods of time indicated above as continuous. On May 16, 1995, by 1900 GMT, the probe was replaced (old S/N 92007, new S/N 92013). There was a complete loss of data between 1200 and 1900 GMT as Site Operations repaired a corroded battery terminal. For several days the right air temperature probe worked well. However, between May 24 and 28, the right air temperature probe malfunctioned continuously again. The right air temperature probe malfunctioned intermittently between May 28 and June 1, as well as during short periods on June 5 and 6. The cause of these latter periods is unknown; the time periods do not correspond with any Site Operations activities. During the intermittent periods the data that is not offscale should be viewed as questionable, as it is not always easy to detect what effect the malfunction might have had on that data. The malfunction of the right air temperature probe results in incorrect Bowen ratio, sensible heat flux, and latent heat flux during the periods indicated. The data values that are affected are: 5 minute: tair_top, tair_bot 15 minute: tair_r 30 minute: tair_top, tair_bot, bowen, e, h. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/07/1995 | 0245 | 03/08/1995 | 1640 |
Subject: | SGP/EBBR/E13 - Frozen Wind Speed Sensor |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | On March 7, 1995 freezing rain stopped the cups of the EBBR wind speed instrument at the Central Facility (E13). All 5, 15, and 30 minute data values indicated below are incorrect for the period listed above. 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/06/1995 | 1645 | 06/06/1995 | 1830 |
Subject: | SGP/EBBR/E15 - Battery Voltage Drop, Storage Module Change |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | For a short period on June 6, 1995 at E15, Ringwood, OK, two events resulted in a loss of data: replacement of the storage module and a low battery condition. The low battery condition apparently affected only the 1815 GMT data output. All data during this period is missing (1645-1800 GMT) of incorrect (1815-1830). |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/07/1995 | 0145 | 03/07/1995 | 1725 |
Subject: | SGP/EBBR/E15 - Frozen Wind Speed Sensor |
DataStreams: | sgp30ebbrE15.b1, sgp5ebbrE15.b1, sgp15ebbrE15.b1 |
Description: | On March 7, 1995 freezing rain stopped the cups of the EBBR wind speed instrument at Ringwood, OK (E15). All 5, 15, and 30 minute data values indicated below are incorrect for the period indicated. 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/07/1995 | 0555 | 03/07/1995 | 2145 |
Subject: | SGP/EBBR/E20 - Frozen Wind Speed Sensor |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | On March 7, 1995 freezing rain stopped the cups of the EBBR wind speed instrument at Meeker, OK (E20). All 5, 15, and 30 minute data values indicated below are incorrect for the period listed above. 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/06/1995 | 2205 | 03/07/1995 | 2230 |
Subject: | SGP/EBBR/E26 - Frozen Wind Speed Sensor |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | On March 6, 1995 freezing rain stopped the cups of the EBBR wind speed instrument at Cement, OK (E26). All 5, 15, and 30 minute data values indicated below are incorrect for the period listed. 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/07/1995 | 0245 | 03/08/1995 | 1640 |
Subject: | SGP/EBBR/E9 - Frozen Wind Speed Sensor |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Description: | On March 7, 1995 freezing rain stopped the cups of the EBBR wind speed instrument at Ashton, KS (E9). All 5, 15, and 30 minute data values indicated below are incorrect for the period listed above. 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/17/1995 | 1400 | 02/22/1995 | 1500 |
Subject: | SGP/EBBR/E13 - Reprocess: Soil Temperature Malfunction |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
Description: | For several days, soil temperature probe #4 indicated values that were either too high or too low, for unknown reasons. After Site Operations personnel checked connections, it recovered to proper values. Affected quantities are shown below. 15 minute: rr_ts4 30 minute: ts4, ces4, g4, ave_shf, e, h Sensible, latent, and soil heat fluxes need to be recalculated using the change in energy storage of probes 1-3 and 5, as follows. ave_shf = (g1 + g2 + g3 + g5)/4 e = -(q + ave_shf)/(1 +bowen) h = -(e + ave_shf +q) |
Measurements: | sgp15ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/11/1995 | 0100 | 03/11/1995 | 0600 |
03/11/1995 | 1530 | 03/12/1995 | 1930 |
03/13/1995 | 0430 | 03/13/1995 | 0500 |
03/16/1995 | 1830 | 03/18/1995 | 1100 |
03/19/1995 | 2100 | 03/20/1995 | 0400 |
03/20/1995 | 1530 | 03/20/1995 | 1700 |
03/10/1995 | 2100 | 03/10/1995 | 2130 |
Subject: | SBP/EBBR/E9 - Reprocess: Soil Heat Flow Sensor 5 Incorrect Data |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1 |
Description: | Soil heat flow sensor #5 periodically outputted incorrect values during the period above. The output often indicated large negative heat flow. The sensor connections were tightened on March 20, 1995, which seems to have corrected the problem. The data affected (and thus incorrect) includes: 15 minute: mv_hft5 30 minute: shf5, c_shf5, g5, ave_shf, e, h. The dates and times (GMT) of incorrect values are listed below. Only half hour times are listed. 03/10/95 2130 03/11/95 0100-0600, 1530-2330 03/12/95 0000-1930 03/13/95 0500 03/16/95 1830-2330 03/17/95 0000-2330 03/18/95 0000-1100 03/19/95 2100-2330 03/20/95 0000-0400, 1530-1700 The latent and sensible heat fluxes can be recalculated by using only soil heat flows 1 through 4 to calculate the average soil heat flow, ave_shf = (g1 + g2 + g3 + g4)/4 e = -(q + ave_shf)/(1 + bowen) h = -(e + ave_shf + q) |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/17/1995 | 2000 | 03/19/1995 | 0500 |
03/19/1995 | 1905 | 04/12/1995 | 1430 |
Subject: | SGP/EBBR/E26 - Frozen Wind Speed Sensor |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | A broken lead on the EBBR wind speed instrument cable at Cement, OK (E26) resulted in the loss of wind speed data. It was nearly a month before the cable could be repaired and reinstalled. All 5, 15, and 30 minute data values indicated below are incorrect for the periods listed above. 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/03/1995 | 0600 | 07/07/1995 | 1605 |
Subject: | SGP/EBBR/E12 - Wind Direction Malfunction |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | For the period indicated, a malfunction in the E12 EBBR wind direction sensor caused the following data values to be incorrect: 5 minute: res_ws, wind_d, and sigma_wd 15 minute: mv_wind_d 30 minute: res_ws, wind_d, and sigma_wd. Wind direction normally indicated small values (15 to 25 degrees) from 3 July 1995 at 0600 GMT through 6 July 1995 at 1500 GMT (at which time the sensor was removed). No sensor was in place between 6 July 1995 at 1500 GMT and 7 July 1995 at 1605 GMT (at which time a new sensor was installed). Old sensor S/N K3039, new S/N K3033. |
Measurements: | sgp30ebbrE12.b1:
sgp15ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/18/1995 | 0400 | 07/18/1995 | 1400 |
07/19/1995 | 0430 | 07/19/1995 | 0500 |
07/19/1995 | 0730 | 07/19/1995 | 1500 |
07/29/1995 | 0700 | 07/29/1995 | 1630 |
Subject: | SGP/EBBR/E20 - Reprocess: Soil Temperature Malfunction |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1 |
Description: | During parts of three days, soil temperature probe #4 indicated values that were either too high or too low, for unknown reasons. Site Operations personnel checked connections and repacked the soil around the probe on 21 July at about 1535 GMT, although ts4 had recovered to proper values the previous day. The times of incorrect data and the affected quantities are shown below. Some of the ts4 values during the listed periods appear to be correct, although the change in energy storage values appear to indicate otherwise. 18 July: 0400-1400 GMT 19 July: 0430-0500, 0730-1500 GMT 29 July: 0700-1630 GMT 15 minute: rr_ts4 30 minute: ts4, ces4, g4, ave_shf, e, h Sensible, latent, and soil heat fluxes need to be recalculated using the change in energy storage of probes 1-3 and 5, as follows. ave_shf = (g1 + g2 + g3 + g5)/4 e = -(q + ave_shf)/(1 +bowen) h = -(e + ave_shf +q) |
Measurements: | sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/01/1995 | 1500 | 08/02/1995 | 1615 |
Subject: | SGP/EBBR/E13 - Low Battery Voltage and AEM Left on Manual |
DataStreams: | sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | Low battery voltage caused some data values to be incorrect beginning at 1500 GMT on August 1, 1995. The data should be carefully inspected to determine which values are incorrect. During replacement of the battery at the Central Facility EBBR on August 1, 1995, the AEM exchange toggle switch was left in the manual position, which caused the right aspirator to remain in the home position until the switch was correctly placed in the exchange position at 1616 GMT on August 2, 1995. All 5, 15, and 30 minute values of temperature, relative humidity , vapor pressure, bowen, e, and h are incorrect during the period. 15 and 30 minute values are also incorrect at 1630 GMT. |
Measurements: | sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/09/1995 | 0930 | 08/09/1995 | 2030 |
Subject: | SGP/EBBR/E20 - Annual Calibration Changeout |
DataStreams: | sgp30ebbrE20.b1 |
Description: | On August 9, 1995 between 0930-2030 GMT, the changeout for annual calibration occurred at E20, Meeker, OK. Soil measurements data should be used with caution for the following week at least, as the sensors require time to "settle" into the soil. |
Measurements: | sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/20/1995 | 1700 | 09/14/1995 | 2359 |
Subject: | SGP/EBBR/E12 - AEM Failure |
DataStreams: | sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | As noted in a previous DQR, the home signal values at E12, Pawhuska, OK began decreasing in July. At the beginning of the period listed, the home 30 value dropped below 15; this causes a data quality message to be output in the site operations log as the stated acceptable range for home 30 is 15 to 34.999999. No data is incorrect as a result of this condition. The home 15 value dropped below 35 beginning July 21, 1995, sometime after 2000 GMT; this causes a data quality message to be output in the site operations log as the stated acceptable range for home 15 is 35 to 70. Data is incorrect as a result of this condition, that being 30 minute e, h, and bowen. The home values frequently dropped below their proper levels until 10 August, 1995, at some time after 1100 GMT, when the set screws in the AEM would no longer hold and the aspirator units were stranded inbetween the home positions. Thereafter, the reported home signals were negative. All 30 minute e, h, and bowen data are incorrect for this condition. On August 16, 1995, at about 1600 GMT, site operators determined that they could not repair the AEM on-site and they determined that the replacement AEM that they took with them has an electronic short. Therefore, the AEM was left in manual mode (no exchange). Replacement of the AEM will not take place until a working spare is available and all 30 minute e, h, and bowen data will be incorrect until then. I will submit another DQR or addendum when successful replacement of the AEM is accomplished. |
Measurements: | sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/23/1995 | 0000 | 08/16/1995 | 2359 |
Subject: | SGP/EBBR/E15 - AEM Home 15 Values High |
DataStreams: | sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | Beginning on June 23, 1995, the home 15 value indicated greater than 70 at times. This indicates that the home 15 and 30 outputs are somehow being electronically added in the AEM. Home 15 values above 70 cause a data quality message to be output in the site operations log as the stated acceptable range for home 15 is 35 to 70. However, no data is incorrect as a result of this condition. No replacement AEM is available and site operations was not able to diagnose the problem. At present, data quality is not jeopardized. I will submit another DQR or addendum when successful repair or replacement of the AEM is accomplished. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/09/1995 | 2030 | 08/16/1995 | 2359 |
Subject: | SGP/EBBR/E20 - AEM Home 30 Values Low |
DataStreams: | sgp5ebbrE20.b1, sgp30ebbrE20.b1 |
Description: | The first EBBR to achieve annual calibration changeout was Meeker, Ok, E20. However, soon after changeout on August 9, 1995, the home 30 value dropped below 15, intermittently, to typically 0.3 to 0.5; this causes a data quality message to be output in the site operations log as the stated acceptable range for home 30 is 15 to 34.999999. However, no data is incorrect as a result of this condition. No replacement AEM is available, as the changeout kit AEMs are somewhat different from the initial units. Site operations hopes to have time to diagnose the problem on the next PM visit and correct the problem. At present, data quality is not jeopardized. I will submit another DQR or addendum when successful replacement of the AEM is accomplished. |
Measurements: | sgp5ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/16/1995 | 0600 | 08/16/1995 | 1530 |
08/18/1995 | 1230 | 08/18/1995 | 1500 |
08/19/1995 | 1100 | 08/19/1995 | 1500 |
08/20/1995 | 0900 | 08/20/1995 | 1500 |
08/21/1995 | 0600 | 08/21/1995 | 1500 |
Subject: | SGP/EBBR/E13 - Battery Voltage Drop |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | Sometime before August 16, 1995 the AC/DC converter was switched off and the instrument heaters were switched on. The solar panel was not able to keep the battery voltage up under these conditions. Much of the data for the periods indicated is incorrect, especially data from the sensors that require 12 volts DC power. The data would have to be inspected to determine what data is correct and what is incorrect. It would be safest to exclude all data from those periods from any analysis. Site operations switched the heaters off and turned the converter back on on August 21, 1995. As reported in a recent DQR, the fluxes from the EBBR are incorrect presently because of a malfunctioning T/RH probe; no spares are presently available. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/18/1995 | 0000 | 08/28/1995 | 2359 |
Subject: | AEM Home 30 Values Low |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | DQR No: Platform: EBBR Subject: E8 EBBR AEM Home 30 Values Low Date Submitted: August 28, 1995 Submitted By: _x_ Instrument Mentor ___ EST Member ___ Science Team Member ___ Other _____________________________ For questions or problems, please contact the ARM Experiment Center at 509-375-6898 or via email at dqr@arm.gov. Platform/Measurement: EBBR/AEM What level data: (raw,a0,a1,b1,c1 etc): a1 (5 and 30 minute) a0 (15 minute) What location was the data collected at: E8, Coldwater, KS Period of time in question Begin Date 08/18/95 to present Data should be labeled: ___ questionable ___ All data fields affected ___ incorrect ___ Only some data fields affected ___ wrong calibration _X_ others Discussion of Problem: At least by August 18, 1995, the home 30 value intermittently dropped below 15, to typically a small negative number; this causes a data quality message to be output in the site operations log as the stated acceptable range for home 30 is 15 to 34.999999. However, no data is incorrect as a result of this condition. No replacement AEM is available, as the changeout kit AEMs are somewhat different from the initial units. I have requested site operations to diagnose the cause of the problem on the next PM visit and hopefully correct the problem. I will submit another DQR or addendum when successful replacement of the AEM is accomplished. Other observations/measurements impacted by this problem: Suggested Corrections of the Problem: (e.g. change calibration factor and recompute, flag data with this comment, etc.) Flag data with this comment. Data Processing Notes Date ------------------------------------------------------------------- END |
Measurements: | sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/27/1995 | 1615 | 09/14/1995 | 1600 |
Subject: | SGP/EBBR/E15 - AEM Not Exchanging |
DataStreams: | sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | On August 27, 1995 at approximately 1615 GMT the upper sprocket shaft of the AEM broke in two, causing the AEM to stop exchanging. The home signal values then remained at around 22 for both home_15 and Home_30. On August 28 site operations personnel removed the fuse (to prevent exchange) with the left aspirator in the down position. The home signal values remained around -2.3 after that. The AEM will remain in this condition until a replacement is available. Therefore, the calculations of Bowen ratio, latent heat flux, and sensible heat flux are incorrect for all times after August 27, 1995 at 1615 GMT. The data fields listed below are incorrect after this time: 5 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot 30 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, Bowen, e, h It is not possible to reliably calculate the Bowen ratio, e, and h using the 15 minute data. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/26/1995 | 0345 | 09/12/1995 | 2000 |
Subject: | SGP/EBBR/E26 - AEM not exchanging |
DataStreams: | sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | On August 26, 1995 at approximately 0345 GMT the gears on the lower sprocket shaft of the AEM appear to have been damaged, causing the AEM to stop exchanging. The home signal values then remained at around -0.5 for both home_15 and Home_30. On August 29 site operations personnel removed the fuse (to prevent exchange) with the left aspirator in the down position. The AEM will remain in this condition until a replacement is available. Therefore, the calculations of Bowen ratio, latent heat flux, and sensible heat flux are incorrect for all times after August 26, 1995 at 0345 GMT. The data fields listed below are incorrect after this time: 5 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot 30 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, Bowen, e, h. It is not possible to reliably calculate the Bowen ratio, e, and h using the 15 minute data. |
Measurements: | sgp30ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/16/1995 | 0015 | 08/30/1995 | 1757 |
Subject: | SGP/EBBR/E12 - Wind Direction Malfunction |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | For the period above, a malfunction in the E12 EBBR wind direction sensor caused the wind direction to output a constant 212 degrees. Site operations replaced the sensor (S/N K3033) with K3042. The following data values are incorrect for the period listed: 5 minute: res_ws, wind_d, and sigma_wd 15 minute: mv_wind_d 30 minute: res_ws, wind_d, and sigma_wd. |
Measurements: | sgp30ebbrE12.b1:
sgp15ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/27/1995 | 1615 | 09/14/1995 | 1630 |
Subject: | SGP/EBBR/E15 - AEM Replaced |
DataStreams: | sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | Please see DQR P950908.1, where this problem was first documented. On August 27, 1995 at approximately 1615 GMT the upper sprocket shaft of the AEM broke in two, causing the AEM to stop exchanging. The home signal values then remained at around 22 for both home_15 and Home_30. On August 28 site operations personnel removed the fuse (to prevent exchange) with the left aspirator in the down position. The home signal values remained around -2.3 after that. On 14 September 1995, site operations personnel replaced the AEM (serial numbers unknown) and it has operated correctly since then. The calculations of Bowen ratio, latent heat flux, and sensible heat flux are incorrect for all times during the period listed above. The data fields listed below are incorrect: 5 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot 30 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, Bowen, e, h It is not possible to reliably calculate the Bowen ratio, e, and h using the 15 minute data. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/20/1995 | 1700 | 09/14/1995 | 1830 |
Subject: | SGP/EBBR/E12 - AEM replacement |
DataStreams: | sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | Please see DQR P950816.2, where this problem was first documented. As noted in a previous DQR, the home signal values at E12, Pawhuska, OK began decreasing in July. At the beginning of the period listed, the home 30 value dropped below 15; this causes a data quality message to be output in the site operations log as the stated acceptable range for home 30 is 15 to 34.999999. No data is incorrect as a result of this condition. The home 15 value dropped below 35 beginning July 21, 1995, sometime after 2000 GMT; this causes a data quality message to be output in the site operations log as the stated acceptable range for home 15 is 35 to 70. Data is incorrect as a result of this condition, that being 30 minute e, h, and bowen. The home values frequently dropped below their proper levels until 10 August, 1995, at some time after 1100 GMT, when the set screws in the AEM would no longer hold and the aspirator units were stranded inbetween the home positions. Thereafter, the reported home signals were negative. All 30 minute e, h, and bowen data are incorrect for this condition. On August 16, 1995, at about 1600 GMT, site operators determined that they could not repair the AEM on-site and they determined that the replacement AEM that they took with them has an electronic short. Therefore, the AEM was left in manual mode (no exchange). All 30 minute e, h, and bowen data are incorrect until the end of the period listed above. On 13 September 1995 site operations personnel replaced the AEM (old S/N 92005, new S/N 94008). The new AEM would not exchange when the right sensor was in the home position: this was confirmed on 14 September when site operations personnel again visited the site. However, soon after the site operations personnel left Pawhuska on 14 September (at 1810 GMT), the AEM started to exchange properly (data correct for 1900 GMT and after). The AEM continues to operate properly. |
Measurements: | sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/26/1995 | 0345 | 09/12/1995 | 2000 |
Subject: | SGP/EBBR/E26 - AEM replacement |
DataStreams: | sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | Please see DQR P950908.2, where this problem was first documented. On August 26, 1995 at approximately 0345 GMT the gears on the lower sprocket shaft of the AEM appear to have been damaged, causing the AEM to stop exchanging. The home signal values then remained at around -0.5 for both home_15 and Home_30. On August 29 site operations personnel removed the fuse (to prevent exchange) with the left aspirator in the down position. On 12 September 1995 the AEM was replaced at 1959 GMT (old S/N 92004, new S/N 92008) and is functioning properly. Calculations of Bowen ratio, latent heat flux, and sensible heat flux are incorrect for all times during the period above. The data fields listed below are incorrect during the period: 5 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot 30 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, Bowen, e, h. It is not possible to reliably calculate the Bowen ratio, e, and h using the 15 minute data. |
Measurements: | sgp30ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/18/1995 | 0630 | 09/05/1995 | 1800 |
Subject: | SGP/EBBR/E8 - Home 30 Values Correct |
DataStreams: | sgp30ebbrE8.b1 |
Description: | Beginning 0630 GMT on August 18, 1995, the home 30 value intermittently dropped below 15, to typically a small negative number; this causes a data quality message to be output in the site operations log as the stated acceptable range for home 30 is 15 to 34.999999. However, no data is incorrect as a result of this condition. On September 15 at 1800 GMT the home 30 signal returned to normal and has remained normal since then. This occurred without the intervention of site operations personnel. |
Measurements: | sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/27/1995 | 2100 | 10/13/1995 | 1230 |
Subject: | SGP/EBBR/E26 - Battery Voltage Drop |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | Shortly after the PM visit on September 26, 1995 the EBBR battery voltage at E26, Cement, OK declined to lower than acceptable levels. During the October 13, 1995 PM visit the battery was replaced. I do not know the specifics of why the battery voltage declined so suddenly, but failure of one or more of the battery cells is a possibility. I have not received the site operations log of the PM visit yet. The voltage dropped to 10 volts at night. At various times, soil temperature, soil heat flux plate output, and atmospheric pressure (at least) were not reported correctly by the CR10 datalogger. Amazingly, the AEM continued to work and reported millivolt outputs that did not fall below the acceptable ranges. All data between 2100 GMT on September 26, 1995 and 1230 GMT on October 13, 1995 should be considered suspect, particularly when the battery voltage fell below 11 volts (every nighttime). This general caution is issued because incorrect reporting of data also took place during times when the battery voltage was reported as 12 Volts, a normally acceptable level; the reason for this behavior is unknown. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/12/1995 | 0000 | 10/13/1995 | 1230 |
Subject: | SGP/EBBR/E26 - Net Radiometer Dome Punctured |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | The upper dome of the net radiometer was punctured sometime between the PM visits of September 12 and 26, 1995. After the puncture, water collected in the top hemisphere. The water was removed on September 26, but a replacement dome was not available. The dome was replaced during the PM visit on October 13, 1995. The puncture and water did not result in obviously incorrect data. I cannot tell when the puncture occurred from the data. The data, particularly during the two weeks between September 12 and 26, 1995, should be considered suspect. Suspect net radiation data results in the latent heat flux and sensible heat flux also being suspect. Suspect data then includes: 5 minute: q 15 minute: mv_q 30 minute: q, e, h. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/05/1995 | 1530 | 10/05/1995 | 1830 |
Subject: | SGP/EBBR/E9 - 3-month Check |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during this period, at least, do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/26/1995 | 0000 | 11/03/1995 | 2045 |
Subject: | SGP/EBBR/E13 - Wind Direction Set Pin Slippage |
DataStreams: | sgp30ebbrE13.b1, sgp5ebbrE13.b1, sgp15ebbrE13.b1 |
Description: | The set pin bolt on the wind vane loosened, allowing the vane to turn on the shaft. Wind directions are incorrect as a result, for the period listed above. The bolt was tightened on 3 Nov 95 so that correct wind directions would result. Values that are incorrect for the period are: 5 minute: wind_d, res_ws, sigma_wd 15 minute: mv_wind_d 30 minute: wind_d, res_ws, sigma_wd |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/01/1995 | 1545 | 11/01/1995 | 1845 |
Subject: | SGP/EBBR/E26 - Right Tair Offscale |
DataStreams: | sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp15ebbrE26.b1 |
Description: | On 1 November 1995 at E26, Cement, OK EBBR the right thermocouple output went offscale for a few hours. The cause for this is unknown and the sensor recovered fully. Incorrect data for the period above includes: 5 minute: tair_top, tair_bot 15 minute: tair_r, tair_l 30 minute: tair_top, tair_bot, bowen, h, e |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/04/1995 | 1200 | 11/04/1995 | 1400 |
11/05/1995 | 0330 | 11/05/1995 | 1430 |
11/06/1995 | 0000 | 11/06/1995 | 1530 |
11/07/1995 | 0000 | 11/07/1995 | 1400 |
11/08/1995 | 0000 | 11/08/1995 | 1400 |
Subject: | SGP/EBBR/E13 - Battery voltage drop |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | On Friday 2 November, 1995 the instrument heaters were switched on. By 4 November, the solar panel was not able to maintain the battery voltage for the cold night conditions that existed. On Monday, 5 November the heaters were turned off, but the poor condition of the battery did not allow it, once partially drained, to recharge sufficiently for nighttime conditions. On 8 November, at 2106 GMT, the battery was replaced (resulting in missing data for 2030-2130 GMT). Most of the data for the period above is incorrect, especially data from the sensors that require at least 11 volts DC power. The data that is incorrect is listed below. 5 minute: tref, tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, q, pres, home 15 minute: rr_thum_r, rr_thum_l, rr_ts1, rr_ts2, rr_ts3, rr_ts4, rr_ts5, rr_tref, mv_hum_r, mv_hum_l, mv_pres, mv_home, mv_q, mv_hft1, mv_hft2, mv_hft3, mv_hft4, mv_hft5, tair_r, tair_l 30 minute: tair_top, tair_bot, tref, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, pres, home_15, home_30, bowen, e, h, shf, shf1, shf2, shf3, shf4, shf5, c_shf1, c_shf2, c_shf3, c_shf4, c_shf5, ts1, ts2, ts3, ts4, ts5, q, g1, g2, g3, g4, g5, ces1, ces2, ces3, ces4, ces5 Periods of time affected (these are the 30 minute periods that are incorrect - incorrect 5 and 15 minute data may also occur for all or part of the previous half hour): 4 Nov 95: 1200-1400 GMT 5 Nov 95: 0330-1430 GMT 6 Nov 95: 0000-1530 GMT 7 Nov 95: 0000-1400 GMT 8 Nov 95: 0000-1400 GMT |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/11/1995 | 0700 | 11/11/1995 | 0730 |
Subject: | SGP/EBBR/E13 - AEM Exchange Stopped in Snow and Freezing Rain |
DataStreams: | sgp30ebbrE13.b1 |
Description: | On the night of November 10-11, a north-south line from south central Kansas through south central Oklahoma received snow and freezing rain and experienced below freezing temperatures. This caused the EBBR automatic exchange mechanism (AEM) to stop exchanging temporarily. Note that temperatures, relative humidities, and vapor pressures are not affected; these are correct as measured, but the height at which the measurements occurred must be determined from the home signal value. |
Measurements: | sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/10/1995 | 0030 | 11/10/1995 | 1530 |
Subject: | SGP/EBBR/E20 - Soil Heat Flow Zero Under Unchanging Conditions |
DataStreams: | sgp30ebbrE20.b1 |
Description: | This DQR is written to inform users of the data that the EBBR soil heat flow measurements of zero during much of 11 November 1995 at E20, Meeker, OK are correct values. Nearly unchanging conditions of net radiation, air temperature, and soil moisture, plus high wind speeds and near neutral atmospheric stability, resulted in almost unchanging conditions of soil temperature and ten and a half hours of zero soil heat flow output from all soil heat flow plates (some plates outputted zero for longer than this). The change in energy storage (determined from the change in temperature with time of the soil above the plates) was very nearly zero also during the period of interest. Such an extended period of zero soil heat flow and the accompanying unchanging soil and atmospheric conditions is quite rare and therefore worth noting here in case someone using the data might mistakenly think that the data is incorrect. |
Measurements: | sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/11/1995 | 0240 | 11/11/1995 | 1505 |
Subject: | SGP/EBBR/E26 - Wind Speed Cups Frozen |
DataStreams: | sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp15ebbrE26.b1 |
Description: | Freezing rain, snow, and below freezing conditions during the night of 10-11 November 1995 resulted in the wind speed sensor cups becoming frozen in a stopped state at E26, Cement, OK on early 11 November 1995. When temperatures rose to almost freezing under cloudless skies later in the day, the ice thawed and the cups were free to turn again. Measurements affected by the stalled wind speed sensor include: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/11/1995 | 0115 | 11/15/1995 | 2359 |
Subject: | SGP/EBBR/E15 - AEM Exchange Stopped in Snow and Freezing Rain |
DataStreams: | sgp30ebbrE15.b1 |
Description: | On the night of November 10-11, a north-south line from south central Kansas through south central Oklahoma received snow and freezing rain and experienced below freezing temperatures. This caused the EBBR automatic exchange mechanism (AEM) to stop exchanging temporarily. Note that temperatures, relative humidities, and vapor pressures are not affected; these are correct as measured, but the height at which the measurements occurred must be determined from the home signal value. |
Measurements: | sgp30ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/11/1995 | 0145 | 11/11/1995 | 1600 |
Subject: | SGP/EBBR/E9 - AEM Exchange Stopped in Snow and Freezing Rain |
DataStreams: | sgp30ebbrE9.b1 |
Description: | On the night of November 10-11, a north-south line from south central Kansas through south central Oklahoma received snow and freezing rain and experienced below freezing temperatures. This caused the EBBR automatic exchange mechanism (AEM) to stop exchanging temporarily. Note that temperatures, relative humidities, and vapor pressures are not affected; these are correct as measured, but the height at which the measurements occurred must be determined from the home signal value. |
Measurements: | sgp30ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/11/1995 | 0300 | 11/11/1995 | 1600 |
Subject: | SGP/EBBR/E20 - AEM Exchange Stopped in Snow and Freezing Rain |
DataStreams: | sgp30ebbrE20.b1 |
Description: | On the night of November 10-11, a north-south line from south central Kansas through south central Oklahoma received snow and freezing rain and experienced below freezing temperatures. This caused the EBBR automatic exchange mechanism (AEM) to stop exchanging temporarily. Note that temperatures, relative humidities, and vapor pressures are not affected; these are correct as measured, but the height at which the measurements occurred must be determined from the home signal value. |
Measurements: | sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/15/1995 | 1230 | 11/15/1995 | 2100 |
Subject: | SGP/EBBR/E9 - Annual calibration changeout |
DataStreams: | sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Description: | On November 15, 1995 the changeout for annual calibration occurred at E9, Ashton, KS. SEBS unit 4 was removed and unit 6 was installed. Between 1230 and 2100 GMT no data was collected; data during this period is missing. |
Measurements: | sgp30ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/18/1995 | 0315 | 11/21/1995 | 2130 |
Subject: | SGP/EBBR/E12 - AEM Failure/Right Radiation Shield OFF |
DataStreams: | sgp30ebbrE12.b1 |
Description: | On 18 November 1995 the AEM at E12, Pawhuska, OK stopped exchanging, with the right radiation shield at the bottom position. The 30 minute home signal reflects this condition. During the 21 November PM visit, the site operators also found that the right radiation shield was no longer attached to the AEM, but was dangling by its conduit. It is not clear whether both conditions occurred simultaneously. Comparison of E12 data with that at the Central Facility indicates that E12 data was correct on 17 November. The AEM had stopped exchanging because a transistor became unsoldered from the AEM circuit board. Site operators repaired this and reattached the right radiation shield to bring the system to operational status at 2130 GMT. All 30 minute latent heat (e), sensible heat (h), and Bowen ratio (bowen) data during the period above are incorrect. |
Measurements: | sgp30ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/14/1995 | 1650 | 12/14/1995 | 2100 |
Subject: | SGP/EBBR/E15 - Annual calibration changeout |
DataStreams: | sgp5ebbrE15.b1, sgp30ebbrE15.b1 |
Description: | On December 14, 1995 the changeout for annual calibration occurred at E15, Ringwood, OK. Between 1650 and 2050 GMT no data was remotely collected. Furthermore, 5 minute data for 2050 GMT is incorrect, 15 minute data for 2045 GMT is incorrect, and 30 minute data for 2100 GMT is incorrect. A rather large jump in soil moisture appears in the data after the installation was completed (the bags holding the probes were soaked with water before installation). Soil measurements indicated greater than ambient soil moisture until a rain on late 18 December saturated the soil; after this period the probes reflected actual soil moisture conditions. The larger than normal soil moistures have little impact on the small soil heat flux during this time of year, and thus no significant affect on computed sensible and latent heat fluxes. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/09/1995 | 0530 | 12/09/1995 | 1415 |
12/13/1995 | 0400 | 12/13/1995 | 1800 |
12/13/1995 | 2300 | 12/14/1995 | 1430 |
12/15/1995 | 0745 | 12/15/1995 | 1515 |
Subject: | SGP/EBBR/E13 - battery voltage drop |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | The E13 battery condition was such that the solar panel was not able to keep the battery voltage up (greater than 11 V DC) under cloudy conditions. The data for the periods above is incorrect; many sensors require at least 11 volts DC power. On 15 December 1995 at 1900 GMT the EBBR lost it's connection to the site data system; data after this period will be discussed in a separate DQR. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/02/1995 | 1530 | 11/28/1995 | 2200 |
Subject: | Annual calibration changeout |
DataStreams: | sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | DQR No: Platform: EBBR Subject: Annual Calibration Changeout - E7, Elk Falls, KS Date Submitted: December 21, 1995 Submitted By: _x_ Instrument Mentor ___ EST Member ___ Science Team Member ___ Other _____________________________ For questions or problems, please contact the ARM Experiment Center at 509-375-6898 or via email at dqr@arm.gov. Platform/Measurement: EBBR/Changeout What level data: (raw,a0,a1,b1,c1 etc): a1 (5 and 30 minute) a0 (15 minute) What location was the data collected at: E7, Elk Falls, KS Period of time in question Begin Date 11/02/95 Time 1530 GMT End Date 11/28/95 Time 2200 GMT Data should be labeled: ___ questionable ___ All data fields affected part of the period) _X_ incorrect _X_ Only some data fields affected ___ wrong calibration ___ others Discussion of Problem: On November 2, 1995 the changeout for annual calibration occurred at E7, Elk Falls, KS. SEBS unit 9 was removed and unit 12 was installed. Between 1530 and 1930 GMT no data was collected. The soil moisture probes were shipped dry from the vendor. Site operations did not realize this and installed them without wetting the bags that hold the soil moisture probes in soil. The water did not penetrate into the bags of soil. Therefore, some soil moisture probe outputs were initially offscale. Different soil moisture probes indicated offscale at different times until 28 November 1995. On 28 November, site operations soaked the soil moisture probe bags with water, causing the soil moisture data values to come on scale. No data was collected during the time of the installation on November 2, between 1530 and 1930 GMT. The incorrect soil moisture after SEBS 12 was installed affected numerous data values including: 15 minute: rr_sm1, rr_sm2, rr_sm3, rr_sm4, rr_sm5 30 minute: h, e, ave_shf, shf1, shf2, shf3, shf4, shf5, sm1, sm2, sm3, sm4, sm5, cs1, cs2, cs3, cs4, cs5, ces1, ces2, ces3, ces4, g1, g2, g3, g4, g5. Inspection of the data readily shows which soil moisture probes were offscale. When one or more were offscale, the e, h, and ave_shf data and the respective numbered data variables above were incorrect. h and e can be recalculated using only correct soil moisture probe values and equations shown in many previous DQRs. Other observations/measurements impacted by this problem: see above Suggested Corrections of the Problem: (e.g. change calibration factor and recompute, flag data with this comment, etc.) Flag data with this comment. Data Processing Notes Date ------------------------------------------------------------------- END |
Measurements: | sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/18/1995 | 2110 | 12/19/1995 | 1425 |
Subject: | SGP/EBBR/E26 - Wind Speed Cups Frozen |
DataStreams: | sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp15ebbrE26.b1 |
Description: | Freezing rain, snow, and below freezing conditions during 18-19 January 1995 caused the wind speed sensor cups to freeze in a stopped state at E26, Cement, OK on 18 January 1995. When temperatures rose to above freezing on 19 January, the ice thawed and the cups turned freely again. Measurements affected by the stalled wind speed sensor include: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/18/1995 | 1705 | 12/19/1995 | 1450 |
Subject: | SGP/EBBR/E22 - Wind Speed Cups Frozen |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | Freezing rain, snow, and below freezing conditions during 18-19 January 1995 caused the wind speed sensor cups to freeze in a stopped state at E22, Cordell, OK on 18 January 1995. When temperatures rose to above freezing on 19 January, the ice thawed and the cups turned freely again. Measurements affected by the stalled wind speed sensor include: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/18/1995 | 1845 | 12/19/1995 | 1755 |
Subject: | SGP/EBBR/E15 - Wind Speed Cups Frozen |
DataStreams: | sgp30ebbrE15.b1, sgp5ebbrE15.b1, sgp15ebbrE15.b1 |
Description: | Freezing rain, snow, and below freezing conditions during 18-19 January 1995 caused the wind speed sensor cups to freeze in a stopped state at E15, Ringwood, OK on 18 January 1995. When temperatures rose to above freezing on 19 January, the ice thawed and the cups turned freely again. Measurements affected by the stalled wind speed sensor include: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/19/1995 | 1040 | 12/19/1995 | 1650 |
Subject: | SGP/EBBR/E7 - Wind Speed Cups Frozen |
DataStreams: | sgp30ebbrE7.b1, sgp5ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Freezing rain, snow, and below freezing conditions during 19 January 1995 caused the wind speed sensor cups to freeze in a stopped state at E7, Elk Falls, KS on 19 January 1995. When temperatures rose to above freezing later that day, the ice thawed and the cups turned freely again. Measurements affected by the stalled wind speed sensor include: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/26/1995 | 1630 | 12/31/1995 | 1930 |
Subject: | SGP/EBBR/E26 - Right Tair Offscale |
DataStreams: | sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp15ebbrE26.b1 |
Description: | On 26 and 31 December 1995 at E26, Cement, OK EBBR the right thermocouple output went offscale. The cause for this is unknown and the sensor recovered fully. Five minute data would need to be inspected during the periods above for incorrect five minute values. Site Operations has been asked to check the sensor connections. Incorrect data for the period indicated includes: 5 minute: tair_top, tair_bot 15 minute: tair_r 30 minute: tair_top, tair_bot, bowen, h, e |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/20/1995 | 1600 | 01/16/1996 | 2030 |
Subject: | SGP/EBBR/E15 - Reprocess: Soil Temperature Probe Malfunction |
DataStreams: | sgp30ebbrE15.b1, sgp15ebbrE15.b1 |
Description: | Malfunction of soil temperature probe #3 at 157, Ringwood, OK resulted in incorrect data values for 7 fields for the period of time listed above. The soil temperature value from probe #3 went offscale, resulting in a zero value for the soil energy storage term calculated from that probe and thus no contribution to the total soil heat flux for soils set 4 and the same for average soil heat flux. This led to incorrect sensible and latent heat flux values. The probe may have become disconnected during the 20 December site operations preventative maintenance visit. However, it also had rodent chew damage and so was removed and replaced. The fields that are incorrect are: 30 minute: ts3, ces3, g3, ave_shf, e, h 15 minute: rr_ts3. Soil temperature probe #3 was replaced at 2030 GMT on 16 January 1996. The old S/N was 95043, the new S/N is 93018. Sensible and latent heat fluxes can be recomputed by averaging the soil heat flux from the four sets of functioning soil probes; use the following equations: ave_shf = (g1 + g2 + g4 + g5)/4 e = -(q +ave_shf)/(1 + bowen) h = e * bowen, where g is total soil heat flow, q is net radiation, and bowen is Bowen Ratio. |
Measurements: | sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/21/1995 | 0630 | 12/22/1995 | 1445 |
12/22/1995 | 2230 | 12/23/1995 | 1445 |
12/24/1995 | 0245 | 12/24/1995 | 1445 |
12/25/1995 | 0330 | 12/25/1995 | 1515 |
12/26/1995 | 0230 | 12/26/1995 | 1445 |
12/27/1995 | 0215 | 12/27/1995 | 1445 |
12/28/1995 | 0000 | 12/29/1995 | 1700 |
12/29/1995 | 1800 | 12/29/1995 | 1815 |
12/29/1995 | 1930 | 01/03/1996 | 1430 |
01/04/1996 | 0530 | 01/04/1996 | 1430 |
01/04/1996 | 1500 | 01/04/1996 | 1645 |
01/04/1996 | 1900 | 01/05/1996 | 1730 |
01/05/1996 | 1930 | 01/05/1996 | 2000 |
01/05/1996 | 2045 | 01/06/1996 | 1445 |
01/07/1996 | 0500 | 01/07/1996 | 1500 |
01/08/1996 | 0300 | 01/08/1996 | 1445 |
01/09/1996 | 0745 | 01/09/1996 | 1415 |
01/10/1996 | 0730 | 01/10/1996 | 1615 |
01/11/1996 | 0130 | 01/11/1996 | 1430 |
01/12/1996 | 0230 | 01/12/1996 | 1445 |
01/13/1996 | 0100 | 01/13/1996 | 1445 |
01/14/1996 | 0030 | 01/14/1996 | 1445 |
01/15/1996 | 0045 | 01/15/1996 | 1600 |
01/15/1996 | 2330 | 01/16/1996 | 2015 |
Subject: | SGP/EBBR/E15 - Battery Voltage Drop |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | From the installation of the changeout unit at E15 on 15 December 1995, the battery exhibited a slow decline in voltage. However, after a day of the AEM not operating, site operations restarted the AEM on 20 December 1995 and the battery condition declined more rapidly, such that the solar panel was not able to keep the battery voltage up (greater than 11 V DC) for nighttime conditions, even after a cloudless day. Battery voltage first dropped below 11 volts at 0635 GMT on 21 December. I requested that the solar panel voltage regulator be tested. On 16 January 1996, site operations found that the fuse in the solar panel voltage regulator was loose. After the fuse was repositioned, the battery voltage increased almost immediately to 13.5 volts and has been good since. All data for the periods above should be considered incorrect; many sensors require at least 11 volts DC power and the performance of the CR10 data logger can not be trusted at less than 11 volts. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/04/1996 | 1515 | 01/05/1996 | 0430 |
01/07/1996 | 2015 | 01/07/1996 | 2330 |
01/08/1996 | 1545 | 01/09/1996 | 1330 |
01/09/1996 | 2000 | 01/09/1996 | 2030 |
Subject: | SGP/EBBR/E26 - Right tair Offscale |
DataStreams: | sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp15ebbrE26.b1 |
Description: | On five days in January 1996 at E26, Cement, OK EBBR the right thermocouple output went offscale. The cause for this is unknown, although poor probe connections are common. Site Operation personnel tightened the probe connections in mid January. Five minute data would need to be inspected during the periods indicated for incorrect five minute values. Incorrect data for the period indicated includes: 5 minute: tair_top, tair_bot 15 minute: tair_r 30 minute: tair_top, tair_bot, bowen, h, e The data are not recoverable from other measurements. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/07/1996 | 0530 | 01/09/1996 | 1730 |
Subject: | SGP/EBBR/E7 - Soil Moisture Probes Offscale |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1 |
Description: | DQR No: Platform: EBBR Subject: E7 Soil Moisture Probes Offscale Date Submitted: January 19, 1996 Submitted By: _x_ Instrument Mentor ___ EST Member ___ Science Team Member ___ Other _____________________________ For questions or problems, please contact the ARM Experiment Center at 509-375-6898 or via email at dqr@arm.gov. Platform/Measurement: EBBR/Soil Moisture Probes What level data: (raw,a0,a1,b1,c1 etc): a1 (15 and 30 minute) What location was the data collected at: E7, Elk Falls, KS Period of time in question (Times vary for different days) Date 01/07/96 Time 0530-2400 GMT Probe sm2 0530-2130 GMT sm4 1100-1900 GMT sm5 01/08/96 0000-2400 GMT sm2 0130-2300 GMT sm4 0130-2030 GMT sm5 01/09/96 0000-2200 GMT sm2 1130-1900 GMT sm4 1130-1730 GMT sm5 Data should be labeled: ___ questionable ___ All data fields affected _X_ incorrect _X_ Only some data fields affected ___ wrong calibration ___ others Discussion of Problem: During numerous periods within the days listed above, individual soil moisture probes indicated high or offscale, on the dry side. The cause of the problems is unknown. Data since 9 January 1996 has been correct. Recalculation of ave_shf, e, and h must be made using only the soil moisture sets with correct soil moisture values. Affected data values are (# for probe number): 15 minute: rr_sm# 30 minute: sm#, c_shf#, cs#, ces#, g#, ave_shf, e, h Sensible, latent, and soil heat fluxes can be recalculated using the soil probe sets that provide correct data. For example, if soil moisture 2 is incorrect, ave_shf = (g1 + g3 + g4 + g5)/4 e = -(q + ave_shf)/(1 +bowen) h = -(e + ave_shf +q), If soil moisture 2 and 4 are incorrect, ave_shf = (g1 + g3 + g5)/3 e = -(q + ave_shf)/(1 +bowen) h = -(e + ave_shf +q), If soil moisture 2, 4, and 5 are incorrect, ave_shf = (g1 + g3)/2 e = -(q + ave_shf)/(1 +bowen) h = -(e + ave_shf +q), Other observations/measurements impacted by this problem: see above Suggested Corrections of the Problem: (e.g. change calibration factor and recompute, flag data with this comment, etc.) Flag data with this comment. Data Processing Notes Date ------------------------------------------------------------------- END |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/23/1996 | 0500 | 01/24/1996 | 1830 |
Subject: | SGP/EBBR/E7 - Wind Speed Cups Frozen |
DataStreams: | sgp30ebbrE7.b1, sgp5ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Freezing rain, snow, and below freezing conditions during 23 January 1996 caused the wind speed sensor cups to freeze in a stopped state at E7, Elk Falls, KS on 23 January 1996. When temperatures rose to above freezing on 24 January 1996, the ice thawed and the cups turned freely again. Measurements affected by the stalled wind speed sensor include: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/23/1996 | 0730 | 01/24/1996 | 1615 |
01/18/1996 | 0500 | 01/18/1996 | 2100 |
Subject: | SGP/EBBR/E12 - Wind Speed Cups Frozen |
DataStreams: | sgp30ebbrE12.b1, sgp5ebbrE12.b1, sgp15ebbrE12.b1 |
Description: | Freezing rain, snow, and below freezing conditions during 18 and 23 January 1996 caused the wind speed sensor cups to freeze in a stopped state at E12, Pawhuska, OK. The cups were freed up later on 18 January when the wind speed rose to above 10 meters per second. When temperatures rose to above freezing on 24 January 1996, the ice thawed and the cups turned freely again. Measurements affected by the stalled wind speed sensor include: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/23/1996 | 0450 | 01/23/1996 | 1555 |
Subject: | SGP/EBBR/E9 - Wind Speed Cups Frozen |
DataStreams: | sgp30ebbrE9.b1, sgp5ebbrE9.b1, sgp15ebbrE9.b1 |
Description: | Freezing rain, snow, and below freezing conditions during 23 January 1996 caused the wind speed sensor cups to freeze in a stopped state at E9, Ashton, KS. Site operations personnel removed the ice from the cups later in the day to allow them to turn freely again. Measurements affected by the stalled wind speed sensor include: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/10/1996 | 0230 | 02/10/1996 | 1300 |
02/10/1996 | 1400 | 02/10/1996 | 1600 |
02/11/1996 | 0100 | 02/11/1996 | 0900 |
02/11/1996 | 1030 | 02/11/1996 | 1400 |
02/11/1996 | 1730 | 02/12/1996 | 0100 |
02/12/1996 | 1200 | 02/12/1996 | 1230 |
02/13/1996 | 0130 | 02/13/1996 | 0230 |
02/13/1996 | 1600 | 02/13/1996 | 1700 |
02/14/1996 | 0130 | 02/14/1996 | 0230 |
02/23/1996 | 0330 | 02/23/1996 | 0530 |
02/23/1996 | 1200 | 02/24/1996 | 0200 |
02/28/1996 | 2330 | 02/29/1996 | 0000 |
03/03/1996 | 1500 | 03/03/1996 | 1830 |
03/04/1996 | 0100 | 03/04/1996 | 0230 |
03/04/1996 | 0400 | 03/04/1996 | 1530 |
03/05/1996 | 1200 | 03/05/1996 | 1430 |
03/05/1996 | 1630 | 03/05/1996 | 1830 |
03/06/1996 | 0200 | 03/07/1996 | 0430 |
03/07/1996 | 2300 | 03/07/1996 | 2330 |
03/08/1996 | 0330 | 03/08/1996 | 1730 |
03/10/1996 | 0400 | 03/10/1996 | 1130 |
03/12/1996 | 0400 | 03/12/1996 | 1130 |
03/12/1996 | 1430 | 03/12/1996 | 1600 |
03/13/1996 | 0830 | 03/13/1996 | 1130 |
03/16/1996 | 0400 | 03/16/1996 | 0830 |
03/16/1996 | 1030 | 03/16/1996 | 1200 |
03/16/1996 | 1500 | 03/16/1996 | 1700 |
03/17/1996 | 0600 | 03/17/1996 | 0630 |
03/17/1996 | 0730 | 03/17/1996 | 2030 |
03/18/1996 | 0300 | 03/18/1996 | 1730 |
03/20/1996 | 1700 | 03/20/1996 | 1830 |
03/21/1996 | 0200 | 03/21/1996 | 0230 |
03/22/1996 | 0330 | 03/22/1996 | 0400 |
03/22/1996 | 1600 | 03/22/1996 | 1700 |
03/23/1996 | 0400 | 03/23/1996 | 1930 |
03/27/1996 | 1800 | 03/27/1996 | 2000 |
01/31/1996 | 1430 | 01/31/1996 | 1500 |
02/12/1996 | 1600 | 02/12/1996 | 1830 |
03/03/1996 | 0130 | 03/03/1996 | 0200 |
03/07/1996 | 2030 | 03/07/1996 | 2130 |
03/19/1996 | 0130 | 03/20/1996 | 1430 |
03/24/1996 | 2200 | 03/25/1996 | 1500 |
02/13/1996 | 1400 | 02/13/1996 | 1430 |
03/15/1996 | 0400 | 03/15/1996 | 2200 |
Subject: | SGP/EBBR/E22 - Reprocess: Soil Heat Flow Sensor 4 Incorrect Data |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1 |
Description: | Soil heat flow sensor #4 periodically output incorrect values during the period indicated. The output often indicated large negative heat flow. The sensor was replaced on March 19, 1996, which corrected the problem. The data affected (and thus incorrect) includes: 15 minute: mv_hft4 30 minute: shf4, c_shf4, g4, ave_shf, e, h. The dates and times (GMT) of incorrect values are listed below. Only half hour data times are listed. 01/31/96 1500 (shf4 only) 02/10/96 0230-1300, 1400-1600 02/11/96 0100-0900, 1030-1400, 1730-2330 02/12/96 0000-0100, 1200-1230, 1600-1830 02/13/96 0130-0230, 1430, 1600-1700 02/14/96 0130-0230 02/23/96 0330-0530, 1200-2330 02/24/96 0000-0200 02/29/96 0000 03/03/96 0130-0200, 1500-1830 03/04/96 0100-0230, 0400-1530 03/05/96 1200-1430, 1630-1830 03/06/96 0200-2330 03/07/96 0000-0430, 2030-2130, 2330 03/08/96 0330-1730 03/10/96 0400-1130 03/12/96 0400-1130, 1430-1600 03/13/96 0830-1130 03/15/96 0400-2200 03/16/96 0400-0830, 1030-1200, 1500-1700 03/17/96 0630, 0730-2030 03/18/96 0300-1730 03/19/96 0130-2330 03/20/96 0000-1430, 1700-1830 03/21/96 0230 03/22/96 0400, 1600-1700 03/23/96 0400-1930 03/24/96 2200-2330 03/25/96 0000-1500 03/27/96 1800-2000 The latent and sensible heat fluxes can be recalculated by using only soil heat flows 1, 2, 3, and 5 to calculate the average soil heat flow, ave_shf = (g1 + g2 + g3 + g5)/4 e = -(q + ave_shf)/(1 + bowen) h = -(e + ave_shf + q) |
Measurements: | sgp30ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/01/1996 | 1715 | 04/01/1996 | 1830 |
Subject: | SGP/EBBR/E26 - Right Air Temperature Malfunction |
DataStreams: | sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp15ebbrE26.b1 |
Description: | The right air temperature probe malfunctioned and produced incorrect data during the period of time indicated above. The cause of the problem is unknown. The right air temperature probe appears to be working well presently. The malfunction of the right air temperature probe results in incorrect Bowen ratio, sensible heat flux, and latent heat flux during the period above. The data values that are affected are: 5 minute: tair_top, tair_bot 15 minute: tair_r (1715-1830 GMT values) 30 minute: tair_top, tair_bot, bowen, e, and h (1730-1830 GMT values) |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/23/1996 | 1500 | 04/23/1996 | 2115 |
Subject: | SGP/EBBR/E26 - Right Air Temp Malfunction |
DataStreams: | sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp15ebbrE26.b1 |
Description: | The right air temperature probe malfunctioned and produced incorrect data during the period of time indicated above. The probe (S/N 92013) was replaced by Site Operations on 23 April 1996 at 2115 GMT with S/N 92002. The malfunction of the right air temperature probe results in incorrect Bowen ratio, sensible heat flux, and latent heat flux during the period indicated. The data values that are affected are: 5 minute: tair_top, tair_bot 15 minute: tair_r (1515-2115 GMT values) 30 minute: tair_top, tair_bot, bowen, e, and h (1530-2130 GMT values) |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/01/1995 | 0000 | 05/08/1996 | 1755 |
Subject: | SGP/EBBR/E20 - Temperature/RH Probes Malfunction and Replacement |
DataStreams: | sgp30ebbrE20.b1, sgp5ebbrE20.b1, sgp15ebbrE20.b1, sgp5ebbrE20.00 |
Description: | Beginning on approximately October 1, 1996 the calibration of the temperature sensor in one or both T/RH probes began to drift; temperatures started to drift higher than the thermocouple probes. This drift caused the upper and lower temperatures measured by the T/RH probe to be 2 to 10 degrees C greater than the thermocouple probes. Comparison with adjacent EBBR sites shows that the thermocouples were measuring properly. The problem affected the vapor pressure outputs primarily, making them as much as 1 kPa greater than they should be (vapor pressure is calculated from probe temperature and relative humidity). The effect on sensible and latent heat fluxes was negligible, although latent heat flux may be overestimated by about 5% presently and sensible heat flux underestimated by approximately the same percentage. Site operations replaced the T/RH probes on 8 May 1996; the new probes are performing well. The serial numbers of replaced probes are: Left 632415, Right 632356. New probe serial numbers are: Left 460428, Right 460412. |
Measurements: | sgp5ebbrE20.b1:
sgp5ebbrE20.00:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/09/1996 | 1820 | 04/09/1996 | 2000 |
Subject: | SGP/EBBR/E26 - 3-month check |
DataStreams: | sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp15ebbrE26.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during this period, at least, do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. The period of incorrect data for the EBBR is listed; data times affected do not include the beginning time and do include the ending time (data files are denoted by the end time). |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/02/1996 | 1441 | 05/02/1996 | 1512 |
Subject: | SGP/EBBR/E7 - 3-month check |
DataStreams: | sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | During EBBR 3-month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during this period, at least, do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. The period of incorrect data for the EBBR is listed above; data times affected do not include the beginning time and do include the ending time (data files are denoted by the end time). |
Measurements: | sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/17/1996 | 1502 | 04/17/1996 | 1553 |
Subject: | SGP/EBBR/E8 - 3-month check |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | During EBBR 3-month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during this period, at least, do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. The period of incorrect data for the EBBR is listed above; data times affected do not include the beginning time and do include the ending time (data files are denoted by the end time). |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/30/1996 | 1731 | 04/30/1996 | 1800 |
Subject: | SGP/EBBR/E9 - 3-month check |
DataStreams: | sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during this period, at least, do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. The period of incorrect data for the EBBR is listed above; data times affected do not include the beginning time and do include the ending time (data files are denoted by the end time). |
Measurements: | sgp30ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/11/1996 | 1550 | 04/11/1996 | 1640 |
Subject: | SGP/EBBR/E12 - 3-month check |
DataStreams: | sgp30ebbrE12.b1, sgp5ebbrE12.b1, sgp15ebbrE12.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during this period, at least, do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. The period of incorrect data for the EBBR is listed above; data times affected do not include the beginning time and do include the ending time (data files are denoted by the end time). |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/06/1996 | 1755 | 05/06/1996 | 1835 |
Subject: | SGP/EBBR/E13 - 3-month check |
DataStreams: | sgp30ebbrE13.b1, sgp5ebbrE13.b1, sgp15ebbrE13.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during this period, at least, do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. The period of incorrect data for the EBBR is listed above; data times affected do not include the beginning time and do include the ending time (data files are denoted by the end time). |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/23/1996 | 1610 | 04/23/1996 | 1703 |
Subject: | SGP/EBBR/E15 - 3-month check |
DataStreams: | sgp30ebbrE15.b1, sgp5ebbrE15.b1, sgp15ebbrE15.b1 |
Description: | During EBBR 3-month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during this period, at least, do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. The period of incorrect data for the EBBR is listed above; data times affected do not include the beginning time and do include the ending time (data files are denoted by the end time). |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/10/1996 | 1500 | 04/10/1996 | 1830 |
Subject: | SGP/EBBR/E20 - 3-month check |
DataStreams: | sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | During EBBR 3-month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during this period, at least, do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. The period of incorrect data for the EBBR is listed above; data times affected do not include the beginning time and do include the ending time (data files are denoted by the end time). |
Measurements: | sgp5ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/24/1996 | 1315 | 04/24/1996 | 1402 |
Subject: | SGP/EBBR/E22 - 3-month check |
DataStreams: | sgp30ebbrE22.b1, sgp5ebbrE22.b1, sgp15ebbrE22.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during this period, at least, do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. The period of incorrect data for the EBBR is listed in the times indicated; data times affected do not include the beginning time and do include the ending time (data files are denoted by the end time). |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/24/1996 | 1800 | 05/24/1996 | 2130 |
Subject: | SGP/EBBR/E13 - Annual Calibration changeout |
DataStreams: | sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | On May 24, 1996 the changeout for annual calibration occurred at E13, Central Facility, OK. Between 1800 and 2130 GMT no data was remotely collected. Furthermore, 5 minute data inbetween 1800 and 2130 is incorrect, 15 minute data between 1800 and 2130 is incorrect, and 30 minute data for 2130 GMT is incorrect. A large jump in soil moisture appears in the data after the installation was completed (the soil bags holding the probes had a greater than ambient soil moisture). Soil measurements indicated greater than ambient soil moisture until a rain just after 0600 GMT on 1 June saturated the soil; after this period the probes reflected actual soil moisture conditions. The larger than normal soil moistures have some impact on the soil heat flux during this time of year, and thus soil heat flux from the soil flow plates and soil temperature probes, as well as average soil heat flux, computed sensible and latent heat fluxes; these quantities should be considered to be incorrect from 24 June at 2130 GMT until about 1200 GMT on June 1. Because of communication problems, data is missing from the time of the EBBR installation until 29 May 1996 at about 1815 GMT. This data may be recovered from the sneakernetted data set. |
Measurements: | sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/30/1996 | 1430 | 07/30/1996 | 2100 |
07/31/1996 | 1400 | 07/31/1996 | 1430 |
Subject: | SGP/EBBR/E8 - Annual Calibration Changeout |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | On 30 July 1996 the EBBR changeout for annual calibration occurred at E8, Coldwater, KS. SEBS unit 5 was removed and unit 4 was installed. Between 1430 and 2100 GMT no data or incorrect data was collected; most of the data is missing. Data for half hours 1400-1430 GMT on 31 July is incorrect; 3 month calibration checks were being performed. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/20/1996 | 1430 | 09/20/1996 | 1500 |
Subject: | SGP/EBBR/E9 - Air Temperature Thermocouples Incorrect |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Description: | Both air temperature thermocouples were offscale, for unknown reasons, for 1500 GMT for both 15 and 30 minute air temperature values. The malfunction of the air temperature thermocouples results in incorrect Bowen ratio, sensible heat flux, and latent heat flux during the period above. The data values that are affected are: 5 minute: tair_top, tair_bot 15 minute: tair_r, tair_l 30 minute: tair_top, tair_bot, bowen, e, and h |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/20/1996 | 0700 | 08/30/1996 | 1830 |
Subject: | SGP/EBBR/E13 - Wind Direction Incorrect |
DataStreams: | sgp30ebbrE13.b1, sgp5ebbrE13.b1, sgp15ebbrE13.b1 |
Description: | Wind direction was incorrect during the period indicated due to a sensor electronics failure. On 30 August 1996 site operations replaced S/N K3035 with S/N K3034. Correct data collection resumed. Values that are incorrect for the period are: 5 minute: wind_d, res_ws, sigma_wd 15 minute: mv_wind_d 30 minute: wind_d, res_ws, sigma_wd |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/24/1996 | 1015 | 08/13/1996 | 1800 |
Subject: | SGP/EBBR/E26 - Wind Direction Sensor Failure |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | Wind Direction and resultant wind speed data for the period listed above is incorrect. Wind direction would not record outside 0 to 20 degrees. This is a common malfunction in this style of wind direction sensor. The sensor was replaced on 13 August 1996. The old sensor S/N was K3038 and new sensor is S/N K3033. The affected data values are: 5 minute: wind_d, sigma_wd, res_ws 15 minute: mv_wind_d 30 minute: wind_d, sigma_wd, res_ws |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/12/1996 | 0000 | 07/05/1996 | 2359 |
Subject: | SGP/EBBR/E26 - Right Air Temperature Malfunction |
DataStreams: | sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp15ebbrE26.b1 |
Description: | The right air temperature probe malfunctioned again and produced incorrect data (outside of range) during some of the period indicated above. The cause of the problem is unknown. On 18 June site operations tightened the connection to the J-panel and repaired a broken aspirator fan wire. However, incorrect data resumed shortly after site operations left the site. Data continued to be incorrect during some times through July 5. No incorrect data was recorded from July 6 through July 17. The EBBR unit was changed out on July 17, 1996. The times of incorrect right air temperatures are easy to recognize from the EBBR data, so the precise times of incorrect data are not enumerated here. Typically, the top and bottom air temperature is out of range, resulting in the sensible heat flux being zero. The malfunction of the right air temperature probe results in incorrect Bowen ratio, sensible heat flux, and latent heat flux. The data values that are affected are: 5 minute: tair_top, tair_bot 15 minute: tair_r 30 minute: tair_top, tair_bot, bowen, e, and h |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/05/1996 | 0200 | 08/07/1996 | 1445 |
08/11/1996 | 1200 | 08/11/1996 | 1845 |
08/11/1996 | 2315 | 08/11/1996 | 2330 |
08/12/1996 | 0000 | 08/12/1996 | 0200 |
08/12/1996 | 1130 | 08/12/1996 | 1900 |
08/12/1996 | 2230 | 08/14/1996 | 2200 |
08/15/1996 | 1630 | 08/15/1996 | 1730 |
08/16/1996 | 2200 | 08/20/1996 | 1945 |
08/11/1996 | 1015 | 08/11/1996 | 1045 |
Subject: | SGP/EBBR/E8 - Incorrect Right Humidity Probe Temperature |
DataStreams: | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The temperature portion of the right humidity probe produced incorrect values during the periods indicated. No site operations intervention occurred to correct the problem. The problem has not reoccurred. The following data values are incorrect for the periods listed below: 15 minute: rr_thum_r 30 minute: hum_top, hum_bot, vp_top, vp_bot, e, h, bowen August 5: 0200-2330 6: 0000-2330 7: 0000-1445 11: 1045, 1200-1845, 2315-2330 12: 0000-0200, 1130-1900, 2230-2330 13: 0000-2330 14: 0000-2200 15: 1630-1730 16: 2200-2330 17: 0000-2330 18: 0000-2330 19: 0000-2330 20: 0000-1945 |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/19/1996 | 0000 | 06/19/1996 | 0230 |
06/19/1996 | 0800 | 06/19/1996 | 1000 |
06/30/1996 | 0230 | 06/30/1996 | 1530 |
06/30/1996 | 2130 | 06/30/1996 | 2330 |
07/01/1996 | 0100 | 07/01/1996 | 0830 |
07/02/1996 | 0500 | 07/02/1996 | 1500 |
07/03/1996 | 0130 | 07/03/1996 | 0400 |
07/03/1996 | 1430 | 07/03/1996 | 1530 |
07/04/1996 | 0200 | 07/04/1996 | 0230 |
07/05/1996 | 0000 | 07/05/1996 | 0030 |
07/05/1996 | 0200 | 07/05/1996 | 0230 |
07/02/1996 | 0000 | 07/02/1996 | 0400 |
Subject: | SGP/EBBR/E8 - Reprocess: Soil Heat Flow Sensor 2 Incorrect Data |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil heat flow sensor #2 periodically outputted incorrect values during the period indicated. The output often indicated large negative heat flow. Problems with good connections of the sensor leads have caused this problem in the past. However, no maintenance was performed before the problem ceased on 15 July 1996. No data is available for 20-27 and 29 July, so it is not known if the problem occurred during that time. The data affected (and thus incorrect) includes: 15 minute: mv_hft2 30 minute: shf2, c_shf2, g2, ave_shf, e, h. The dates and times (GMT) of incorrect values are listed below. Only half hour times are listed. 06/19/96 0230, 0800-1000 06/30/96 0230-1530, 2130-2330 07/01/96 0100-0830 07/02/96 0400, 0500-1500 07/03/96 0130-0400, 1430-1530 07/04/96 0200-0230 07/05/96 0030, 0230 The latent and sensible heat fluxes can be recalculated by using only soil heat flows 1 and 3 through 5 to calculate the average soil heat flow, ave_shf = (g1 + g3 + g4 + g5)/4 e = -(q + ave_shf)/(1 + bowen) h = -(e + ave_shf + q) |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/08/1996 | 2145 | 06/19/1996 | 1700 |
Subject: | SGP/EBBR/E20 - Incorrect Data from Left Humidity Probe |
DataStreams: | sgp30ebbrE20.b1, sgp15ebbrE20.b1 |
Description: | The left humidity probe output started drifting upward at 2145 GMT on May 8, 1996 and went offscale (over 105%) at 0200 GMT on May 9, 1996. The output settled at approximately 113% and remained there until the probe was replaced on June 19, 1996. The first correct 15 minute values were at 1630 GMT on June 19 and the first correct 30 minute values were at 1700 GMT. During the period of incorrect left probe RH data, the following values were incorrect: 15 minute: mv_hum_l 30 minute: hum_top, hum_bot, vp_top, vp_bot, h, e, bowen. |
Measurements: | sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/06/1996 | 1300 | 11/06/1996 | 2300 |
Subject: | SGP/EBBR/E22 - Annual Calibration Changeout |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | On 6 November 1996 the EBBR changeout for annual calibration occurred at E22, Cordell, OK. SEBS unit 7 was removed and unit 3 was installed. Between 1300 and 2300 GMT no data or incorrect data was collected; most of the data is missing. Data for half hour 1930 GMT on 7 November is incorrect; 3 month calibration checks were being performed. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/13/1996 | 1516 | 11/13/1996 | 1552 |
Subject: | SGP/EBBR/E7 - EBBR 3-Month Checks - 10 EFs |
DataStreams: | sgp30ebbrE7.b1, sgp5ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | During EBBR 3-month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for periods indicated. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/30/1996 | 1726 | 10/30/1996 | 1801 |
Subject: | SGP/EBBR/E8 - EBBR 3 Month Checks - 10 EFs |
DataStreams: | sgp15ebbrE8.b1, sgp5ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for periods indicated. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/15/1996 | 1557 | 11/15/1996 | 1630 |
Subject: | SGP/EBBR/E9 - EBBR 3 Month Checks - 10 EFs |
DataStreams: | sgp15ebbrE9.b1, sgp5ebbrE9.b1, sgp30ebbrE9.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for periods indicated. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/25/1996 | 1600 | 10/25/1996 | 1640 |
Subject: | SGP/EBBR/E12 - EBBR 3 Month Checks - 10 EFs |
DataStreams: | sgp30ebbrE12.b1, sgp15ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for periods indicated. |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/18/1996 | 1942 | 11/18/1996 | 2022 |
Subject: | SGP/EBBR/E13 - EBBR 3 Month Checks - 10 EFs |
DataStreams: | sgp15ebbrE13.b1, sgp5ebbrE13.b1, sgp30ebbrE13.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for periods indicated. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/05/1996 | 1658 | 11/05/1996 | 1732 |
Subject: | SGP/EBBR/E15 - EBBR 3 Month Checks - 10 EFs |
DataStreams: | sgp30ebbrE15.b1, sgp15ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for periods indicated. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/23/1996 | 1730 | 10/23/1996 | 1755 |
Subject: | SGP/EBBR/E20 - EBBR 3 Month Checks - 10 EFs |
DataStreams: | sgp30ebbrE20.b1, sgp15ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for periods indicated. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/06/1996 | 1500 | 11/06/1996 | 1553 |
Subject: | SGP/EBBR/E22 - EBBR 3 Month Checks - 10 EFs |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for periods indicated. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/24/1996 | 1550 | 10/24/1996 | 1613 |
Subject: | SGP/EBBR/E26 - EBBR 3 Month Checks - 10 EFs |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for periods indicated. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/09/1996 | 1000 | 10/09/1996 | 2000 |
Subject: | SGP/EBBR/E12 - Annual Calibration Changeout |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | On 9 October 1996 the EBBR changeout for annual calibration occurred at E12, Pawhuska, OK. SEBS unit 10 was removed and unit 1 was installed. Between 1000 and 1500 GMT no data was collected. Data for half hours 1530-2000 GMT is incorrect; 3 month calibration checks were being performed. |
Measurements: | sgp30ebbrE12.b1:
sgp15ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/16/1996 | 1500 | 07/16/1996 | 2230 |
Subject: | SGP/EBBR/E26 - Annual Calibration Changeout |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | On 16 November 1996 the EBBR changeout for annual calibration occurred at E26, Cement, OK. SEBS unit 2 was removed and unit 8 was installed. Between 1500 and 2230 GMT no data or incorrect data was collected; most of the data is missing. 3 month calibration checks were being performed. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/06/1996 | 2000 | 11/19/1996 | 1830 |
Subject: | SGP/EBBR/E26 - Wind Direction Sensor Failure |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | Wind Direction and resultant wind speed data for much of the period listed above is incorrect. Wind directions of 135, 0, or 360 would continue for extended periods. This behavior is indicative of a failed sensor. This is a common malfunction in this style of wind direction sensor. A work request has been submitted to site operations for replacement of the sensor. All values of 135, 0, and 360 should be considered to be incorrect during the period above. Furthermore, more incorrect data may occur until the sensor is replaced (by 6 December). The affected data values are: 5 minute: wind_d, sigma_wd, res_ws 15 minute: mv_wind_d 30 minute: wind_d, sigma_wd, res_ws |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/01/1996 | 1500 | 11/01/1996 | 1530 |
11/10/1996 | 0900 | 11/10/1996 | 0930 |
11/09/1996 | 0800 | 11/09/1996 | 0830 |
Subject: | SGP/EBBR/E8 - Incorrect Soil Moisture #1 |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Moisture probe #1 went offscale at these half hour times, for an unknown reason. This problem may reoccur, and at present is extremely intermittent. A work request has been submitted to site operations to check the soil and electrical contacts of the probe. Affected data values are: 15 minute: r_sm1 30 minute: sm1, c_shf1, ces1, cs1, g1, ave_shf, e, h |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/03/1996 | 0300 | 08/03/1996 | 0830 |
08/03/1996 | 2030 | 08/03/1996 | 2100 |
08/05/1996 | 0230 | 08/07/1996 | 1430 |
09/20/1996 | 1030 | 09/20/1996 | 1800 |
09/20/1996 | 2100 | 09/20/1996 | 2130 |
09/20/1996 | 2230 | 09/20/1996 | 2330 |
09/21/1996 | 2330 | 09/22/1996 | 0130 |
09/22/1996 | 0530 | 09/23/1996 | 0130 |
09/23/1996 | 0400 | 09/26/1996 | 0700 |
09/26/1996 | 1900 | 09/26/1996 | 2030 |
09/27/1996 | 0100 | 09/29/1996 | 2330 |
09/30/1996 | 0000 | 09/30/1996 | 1930 |
10/07/1996 | 1400 | 10/07/1996 | 1830 |
10/07/1996 | 2000 | 10/15/1996 | 1900 |
08/03/1996 | 1400 | 08/03/1996 | 1700 |
09/21/1996 | 1500 | 09/21/1996 | 1530 |
09/30/1996 | 2130 | 10/07/1996 | 0530 |
Subject: | SGP/EBBR/E8 - Incorrect Soil Moisture #2, Soil Temperature #2 |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Moisture probe #2 and soil temperature probe #2 went offscale at times soon after installation of a new EBBR unit at E8. The problem may have been poor soil contact to the probes or poor electrical contact at the j- panel. No site operation intervention occurred to correct the problem; the problem has not reoccurred since 15 October 1996. Affected data values were: 15 minute: r_sm2, rr_ts2 30 minute: sm2, ts2, c_shf2, ces2, cs2, g2, ave_shf, e, h The affected times (GMT) and dates are listed below: Aug 3: 0300-0830, 1400-1700, 2030-2100 5: 0230-2330 6: 0000-2330 7: 0000-1430 Sep 20: 1030-1800, 2130, 2230-2330 21: 1530, 2330 22: 0000-0130, 0530-2330 23: 0000-0130, 0400-2330 24: 0000-2330 25: 0000-2330 26: 0000-0700, 1900-2030 27: 0100-2330 28: 0000-2330 29: 0000-2330 30: 0000-1930, 2130-2330 Oct 1: 0000-2330 2: 0000-2330 3: 0000-2330 4: 0000-2330 5: 0000-2330 6: 0000-2330 7: 0000-0530, 1400-1830, 2000-2330 8: 0000-2330 9: 0000-2330 10: 0000-2330 11: 0000-2330 12: 0000-2330 13: 0000-2330 14: 0000-2330 15: 0000-1900 |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/24/1996 | 1715 | 09/24/1996 | 1730 |
Subject: | SGP/EBBR/E12 - Left Air Temperature Thermocouple Incorrect |
DataStreams: | sgp30ebbrE12.b1, sgp5ebbrE12.b1, sgp15ebbrE12.b1 |
Description: | The left air temperature thermocouple was offscale, for unknown reasons, for 1715 GMT for 15 minute, and 1730 GMT for 30 minute air temperature values. The malfunction of the air temperature thermocouple results in incorrect Bowen ratio, sensible heat flux, and latent heat flux during the period indicated. The data values that are affected are: 5 minute: tair_top, tair_bot 15 minute: tair_l 30 minute: tair_top, tair_bot, bowen, e, and h |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/13/1996 | 0000 | 10/09/1996 | 2359 |
Subject: | SGP/EBBR/E12 - Battery Voltage Low - Fluxes Affected |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | The battery condition deteriorated over time so that the solar panel was not always able to keep the battery voltage high enough to produce correct home signal values. The home 15 signal was too low much of the period above, with home 30 being low also occasionally. The home signal qc checks should be consulted to determine which half hour values of sensible (h) and latent (e) heat fluxes are incorrect; no other data values were affected. The first occurrence of low home 15 signal appears to be 0630 GMT on 13 July 1996. Battery voltage did not drop low enough to cause any sensors to stop functioning. The battery was replaced on 9 October 1996 at 2000 GMT, during annual calibration changeout of the entire EBBR unit (the mentor assumes that the battery was replaced - no instrument maintenance reports are available for that time). |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/07/1997 | 0530 | 02/07/1997 | 1340 |
Subject: | SGP/EBBR/E13 - Frozen Wind Speed Sensor |
DataStreams: | sgp30ebbrE13.b1, sgp5ebbrE13.b1, sgp15ebbrE13.b1 |
Description: | On February 7, 1997 freezing rain stopped the cups of the EBBR wind speed instrument at the Central Facility (E13). All 5, 15, and 30 minute data values indicated below are incorrect for the period listed. 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/18/1995 | 1845 | 12/19/1995 | 1755 |
Subject: | SGP/EBBR/E15 - Wind Speed Cups frozen |
DataStreams: | sgp30ebbrE15.b1, sgp5ebbrE15.b1, sgp15ebbrE15.b1 |
Description: | Freezing rain, snow, and below freezing conditions during 18-19 January 1995 caused the wind speed sensor cups to freeze in a stopped state at E15, Ringwood, OK on 18 January 1995. When temperatures rose to above freezing on 19 January, the ice thawed and the cups turned freely again. Measurements affected by the stalled wind speed sensor include: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/23/1996 | 0500 | 01/24/1996 | 1830 |
Subject: | SGP/EBBR/E7- Wind Speed Cups Frozen |
DataStreams: | sgp30ebbrE7.b1, sgp5ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | Freezing rain, snow, and below freezing conditions during 23 January 1996 caused the wind speed sensor cups to freeze in a stopped state at E7, Elk Falls, KS on 23 January 1996. When temperatures rose to above freezing on 24 January 1996, the ice thawed and the cups turned freely again. Measurements affected by the stalled wind speed sensor include: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/24/1996 | 1010 | 11/25/1996 | 1805 |
Subject: | SGP/EBBR/E20 - Wind Speed Cups Frozen |
DataStreams: | sgp30ebbrE20.b1, sgp5ebbrE20.b1, sgp15ebbrE20.b1 |
Description: | Freezing rain, snow, and below freezing conditions during 24 November 1996 caused the wind speed sensor cups to freeze in a stopped state at E20, Meeker, OK. On 25 November, the cups broke free of the ice and turned freely again. Measurements affected by the stalled wind speed sensor include: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/06/1996 | 2000 | 11/19/1996 | 1825 |
Subject: | SGP/EBBR/E26 - Wind Direction Sensor Failure |
DataStreams: | sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp15ebbrE26.b1 |
Description: | Wind Direction and Sigma of Wind Direction data for the period listed is incorrect. The wind direction sensor began to malfunction on 6 November, with the characteristic constant 135 degree output. At other times the output remained a nearly constant 230 to 240 degrees. On 16 November at 1819 GMT site operations replaced sensor S/N 3033 with S/N 3036. Apparently they left it aligned at north when they left the EF; output for the next few days was 0 or 360 degrees, until they returned on 19 November at 1825 GMT and gave the vane it's freedom. Affected fields are: 5 minute: wind_d, sigma_wd 15 minute: mv_wind_d 30 minute: wind_d, sigma_wd |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/20/1997 | 2255 | 03/31/1997 | 2359 |
Subject: | Controlled burn at E12 on 3/20/97 and affected data |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1, sgpsirosE12.a1 |
Description: | DQR No: Platforms: sgpsirosE12 sgpebbrE12 Subject: Controlled burn at E12 on 3/20/97 and affected data Date Submitted: 3/31/97 Submitted By: Mike Splitt ___ Instrument Mentor ___ EST Member ___ Science Team Member _x_ Other _Site_Scientist_Team_ For questions or problems, please contact the ARM Experiment Center at 509-375-6898 or via email at problems@arm.gov. Platform/Measurement: What level data: (raw,a0,a1,b1,c1 etc): all levels What location was the data collected at: sgp E12 Period of time in question Begin Date 3/20/97 Time 22:55 (GMT) End Date * Time : (GMT) * The "end date" is not necessarily in need of definition. The EBBR data appears to be directly affected for a period of about one hour. The effect on the local shortwave albedo will be longer lasting and could be said to "end" when the vegetative state recovers to being representative of similar areas. Data should be labeled: ___ questionable ___ All data fields affected ___ incorrect ___ Only some data fields affected ___ wrong calibration _x_ others Discussion of Problem: This report is reflective of a change in environmental conditions thought to be worth noting at E12, Pawhuska, and does not concern any issue of substandard sensor performance. A contolled burn was conducted at the the Pawhuska Extened Facillity (E12) by the Tallgrass Prairie Preserve. The Preserve director, Bob Hamilton, contacted the ARM SGP Site prior to the activity which allowed Site Operations to take precautionary steps. Danny Nelson reported the following activities by Site Operations: 1) Site Ops cut and removed all dried grass within the EF fence and also to a 10' perimeter around the outside of the fence. Final grass height was 1-3". This was done on Monday the 17th of March. 2) Tallgrass Prairie Preserve personnel put down a 10' wet line around our EF prior to the burn on the 20th, so the EF area was not burned. 3) Site Ops had originally planned to deploy a field crew to inspect the site on the day following the burn. TPP director Bob Hamilton had agreed to call us the morning of the burn, but didn't actually call us until this morning; the day following the burn. Due to the short notification and prior commitment of personnel, Site Ops will not be able to travel to the site until Monday or Tuesday of next week. John Shatz was given the following information about the duration of the burn: >Bob Hamilton called and indicated that they did a control burn of the >Tall Grass Prairie yesterday(3-20-97). He indicated that the fire in >our area lasted between 4:55 PM to 5:40 PM. He also said that they >plan to put this control burning on a two to three year cycle. Affect on ARM Data: Review of the EBBR temperature data (the sensor in the lower position) shows a dramatic increase of temperature from 26 to 37 Celsius in a 10 minute period. The temperatures appear to have been affected for approximately 1 hour. Analysis of the broadband shortwave albedo during this period reveals changes that appear to be related to the burn. The following is a listing of the daily average shortwave albedos (caluclated by running a linear regression fit on the "curve" created by plotting the upwelling shortwave vs. the downwelling shortwave values from the E12 SIROS broadband instrumentation): Date Broadband Shortwave Albedo ________ ____________________________ 3-14-97 .171 3-15-97 .171 3-16-97 .169 3-17-97 .166 3-18-97 .135 3-19-97 .171 3-20-97 .174 3-21-97 .141 3-22-97 .143 3-23-97 .148 3-24-97 .147 The grass cutting activity at the EF does not appear to be responsible for the abledo change on 3-18; this is likely related to the drizzle and overcast sky conditions in the area on that day. Note that the albedo recovers to previous values by the next day. The burn occured late on the 20th, and it appears the albedo after the burn sustained a more long-lasting decrease in value. Examination of both the upwelling and downwelling solar irradiance indicates that the albedo change was a resultant of the decrease in upwelling irradiance, as the downwelling values were comparable to other sites. |
Measurements: | sgp30ebbrE12.b1:
sgp15ebbrE12.b1:
sgpsirosE12.a1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/03/1996 | 0000 | 06/10/1997 | 2200 |
Subject: | SGP/EBBR/E8 - Incorrect Soil Temperature #s 1 and 2 |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil temperature probes #1 and #2 went offscale at times for many months. Replacement probes were not available until recently. The probes were replaced on 16 June, 1997 at 2200 GMT. Affected data values were: 15 minute: r_sm1, r_sm2, rr_ts1, rr_ts2 30 minute: sm1, sm2, ts1, ts2, c_shf1, c_shf2, ces1, ces2, cs1, cs2, g1, g2, ave_shf, e, h |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/15/1996 | 1715 | 06/18/1997 | 1645 |
Subject: | SGP/EBBR/E9 - Temperature Portion of T/RH Probe |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1 |
Description: | The Temperature portion of the right T/RH probe indicated too high a temperature for nearly 11 months. I twice placed work requests to have the probe replaced, but no spares were available. Spares recently became available and I placed a new work request to have the probe changed. The replacement occurred on 18 June 1997; this fixed the problem. The main results of the high temperatures were probe top and bottom temperatures (used only for the calculation of vapor pressure) and vapor pressures that were too large. Furthermore, this caused latent heat flux to be reported to be too large, and therefore sensible heat flux was reported to be too small. Recalculation of the fluxes using thermocouple temperatures to calculate vapor pressure would be quite time consuming and would have to assume that the thermocouple temperature and probe temperature were nearly the same. The EBBR procedures have been changed to include comparison of the thermocouple and probe temperatures to make it easier for the site operators to detect this condition. The data affected (and thus incorrect) includes: 15 minute: rr_thum_r 30 minute: thum_top, thum_bot, vp_top, vp_bot, e, h, bowen. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/03/1996 | 1930 | 06/17/1997 | 2030 |
Subject: | SGP/EBBR/E26 - Reprocess: E26 Soil Heat Flow Sensor 1 Replacement |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1 |
Description: | Apparently, during installation of the present system on 17 July 1996, soil heat flow sensor #4 was intentionally placed upside down in the soil to produce the correct sign of flux. The leads from the sensor must have been connected with the wrong polarity. On 3 December 1996 site operations corrected some wiring problems and reconnected soil heat flow sensor #4 according to the proper color code; this caused the sign of the soil heat flow from that sensor to then be opposite of the other four sensors. All data from sensor #4 is the wrong sign during the period listed above. On 17 June 1997 site operations removed the sensor and installed it with the correct side up, fixing the problem. The result of this problem was a decrease of reported average soil heat flow of approximately 40%. This translates into a 0 to 20% error in latent and sensible heat fluxes (soil heat flow is rarely more than 50% of net radiation, upon which the Bowen ratio calculation depends). Therefore, sensible and latent heat fluxes were typically overestimated by about 10% during the daytime. This is of the same magnitude as the EBBR system/technique error. The following data values were affected for the period listed above: 15 minute: mv_hft4 30 minute: shf4, c_shf4, g4, ave_shf, e, h The latent and sensible heat fluxes can be recalculated by using only soil heat flows 1, 2, 3, and 5 to calculate the average soil heat flow, ave_shf = (g1 + g2 + g3 + g5)/4 e = -(q + ave_shf)/(1 + bowen) h = -(e + ave_shf + q). |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/15/1997 | 1500 | 06/24/1997 | 2000 |
Subject: | SGP/EBBR/E8 - Incorrect Soil Temperature #s 1 and 2 |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil temperature probes #1 and #2 were replaced on June 10, 1997. However, the leads were apparently not connected tightly enough, resulting in some large or offscale resistance measurements and therefore temperature values for those probes. The problem occurred only occasionally until June 20, 1997. On June 20 - 24, 1997 probes 1 and 2 outputs were offscale for several hours per day, both not always at the same time. On June 24, 1997, site operations checked and tightened the connections. No offscale values have occurred since then. Affected data values were: 15 minute: r_sm1, r_sm2, rr_ts1, rr_ts2 30 minute: sm1, sm2, ts1, ts2, c_shf1, c_shf2, ces1, ces2, cs1, cs2, g1, g2, ave_shf, e, h |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/15/1997 | 2245 | 04/22/1997 | 1830 |
Subject: | SGP/EBBR/E15 - Connections Affecting Home Signals |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | For most of the period above at E15, Ringwood, OK, home signals were very large, resulting in incorrect data. On April 22, 1997, site operations found some loose connections on the J-panels that were causing the problem. This problem affected a wide range of measurements and calculated values, rendering much of the data, including the sensible and latent heat fluxes as incorrect. Only careful examination of the data would allow some to be used as is and other quantities to be recalculated. I would request that anyone needing this data contact me for information on it's quality. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/14/1997 | 0000 | 04/09/1997 | 1730 |
Subject: | SGP/EBBR/E22 - Reprocess: E22 Soil Heat Flow Sensor # 5 Incorrect Data |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1 |
Description: | Soil heat flow sensor #5 frequently produced large negative (and incorrect) values during the period above. The problem ended during a PM visit by site operations. However, there is no documentation by Site Operations in the instrumentation report or corrective maintenance report that Site Operations personnel took action to correct the problem. Site Operations also told me that they have no record of any changes or corrective maintenance. Apparently, something fortuitous happened during the PM to correct the problem. The start date and time of the problem cannot be determined exactly, but it did begin during a period of missing data from Feb. 14-25, 1997. Values affected (and thus incorrect) include: 15 minute: mv_hft5 30 minute: shf5, c_shf5, g5, ave_shf, e, h. The latent and sensible heat fluxes can be recalculated by using only soil heat flows 1, 2, 3, and 4 to calculate the average soil heat flow, ave_shf = (g1 + g2 + g3 + g4)/4 e = -(q + ave_shf)/(1 + bowen) h = -(e + ave_shf + q) |
Measurements: | sgp30ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/19/1997 | 0130 | 07/08/1997 | 1915 |
Subject: | SGP/EBBR/E8 - AEM Inoperative |
DataStreams: | sgp30ebbrE8.b1 |
Description: | Beginning on 19 June 1997, the set screw holding one of the AEM sprockets to the drive motor became stripped, preventing the aspirator units from exchanging every time. The problem was intermittent until 24 June when site operations personnel determined that a new sprocket (therefore a replacement AEM) was needed. The AEM was therefore left in the position such that the right aspirator was in the bottom position continuously, with home signals indicating values near -1. The AEM was replaced on 8 July 1997. For most of the time period above, home signals, bowen, e, h, and all thermocouple and probe tmperatures, probe relative humidities, and vapor pressures do not reflect a switching AEM. However, temperatures, RH and vapor pressures are accurate for their height. Obviously, bowen, e, and h are incorrect. |
Measurements: | sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/11/1997 | 0000 | 07/16/1997 | 2359 |
Subject: | SGP/EBBR/E20 - Reprocess: E20 Soil Probes #3 Cables Cut |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1 |
Description: | The cables going to #3 soil temperature, moisture, and heat flow probes were cut by a mower on 11 June 1997 and were repaired on 16 July 1997. Data for the temperature probe was offscale, for the soil moisture probe was a constant 10.1 (close to the actual), and for the soil heat flow was large negative during the period. The following are the data quantities that were affected: 15 minute: rr_ts3, mv_hft3, r_sm3 30 minute: ts3, ces3, shf3, c_shf3, cs3, g3, ave_shf, e, h Sensible, latent, and soil heat fluxes can be recalculated using the values from probes 1, 2, 4, and 5, as follows. ave_shf = (g1 + g2 + g4 + g5)/4 e = -(q + ave_shf)/(1 +bowen) h = -(e + ave_shf +q) |
Measurements: | sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/13/1997 | 0305 | 07/16/1997 | 1500 |
Subject: | SGP/EBBR/E22 - Right Aspirator Unit Stolen |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | On 13 June 1997, vandals stole the right AEM aspirator unit, including the thermocouple and T/RH probe inside. These were replaced on 16 July 1997. Incorrect data fields during the period above are: 5 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot 15 minute: rr_thum_r, mv_hum_r, tair_r 30 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, bowen, e, h |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/03/1997 | 1200 | 06/08/1997 | 1515 |
Subject: | SGP/EBBR/E12 - Wind Direction Sensor Malfunctions |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | For much of the period above, a malfunction in the E12 EBBR wind direction sensor (insect larvae shorting out the circuit board) caused incorrect values for 5 minute res_ws, wind_d, and sigma_wd, 15 minute mv_wind_d, and 30 minute res_ws, wind_d, and sigma_wd. Wind direction normally indicated north (near 0 or 360 degrees); there were some times during the period when wind directions looked good, but the lack of variation in direction and sigma implies that the data is suspect. Therefore, it would be wise to flag all of the values listed above during the entire time period as incorrect. The wind direction sensor (S/N 3081) was replaced with S/N 3042 at 1515 GMT on June 8, 1997. |
Measurements: | sgp30ebbrE12.b1:
sgp15ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/17/1997 | 0930 | 07/25/1997 | 1600 |
Subject: | SGP/EBBR/E13 - CR10 Malfunction |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | The CR10 apparently malfunctioned on 17 July 1997. This caused a variety of effects, including incorrect datalogger clock times, missing data periods, and some corrupted data fields. The CR10 was replaced on 25 July 1997; all data fields are now correct. Data users may contact the mentor for more detailed information on data quality during this period. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/10/1997 | 2030 | 07/14/1997 | 0915 |
Subject: | SGP/EBBR/E8 - Reprocess: E8 EBBR Program Compile Error, Effect On Net Radiation |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | An error in CR10X program compilation occurred when new software was used to perform the compilation. The new software created an erroneous data table listing, which the program works off of. An incorrect data location was used in the adjustment of net radiation for wind speed, causing all net radiation values to be divided by 2. The following data was incorrect during the period above: 5 minute: q 15 minute: mv_q 30 minute: q, e, h The latent and sensible heat fluxes can be corrected by recomputing as follows: e = -(2q + ave_shf)/(1 + bowen) h = -(e + ave_shf + 2q) |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/14/1997 | 0915 | 08/05/1997 | 2300 |
Subject: | E8 EBBR Lightning Strike |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | DQR No: Platform: EBBR Subject: E8 EBBR Lightning Strike Date Submitted: October 16, 1997 Submitted By: _x_ Instrument Mentor ___ EST Member ___ Science Team Member ___ Other Platform/Measurement: EBBR/Numerous What level data: (raw,a0,a1,b1,c1 etc): a1 (15 and 30 minute) a0 (5 minute) What location was the data collected at: E8, Coldwater, KS Period of time in question Begin Date 07/14/97 Time 0915 GMT End Date 08/05/97 Time 2300 GMT Discussion of Problem: A lightning "strike" to the EF 8 Facility shortly after 0900 GMT on 14 July 1997 caused the failure of several EBBR sensors, the solar panel voltage regulator, and the CR10X datalogger. All data is missing for the period above, with degraded operations thereafter for a couple of months while replacement parts slowly became available (a separate DQR will cover the degraded data period). |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/30/1997 | 1000 | 08/14/1997 | 1500 |
Subject: | SGP/EBBR/E19 - Reprocess: E19 EBBR Soil heat Flow Probe #5 Upsidedown |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1 |
Description: | Soil heat flow probe #5 was installed upsidedown at the initial installation of the EBBR system. The data from it had the wrong sign. Site operations personnel reversed its orientation in the soil on August 14, 1997, thereby correcting the problem. Values affected (and thus incorrect) include: 15 minute: mv_hft5 30 minute: shf5, c_shf5, g5, ave_shf, e, h. The latent and sensible heat fluxes can be recalculated by reversing the sign of cshf5, adding it to ces5 and recomputing the average soil heat flow, ave_shf = (g1 + g2 + g3 + g4 + g5)/5 e = -(q + ave_shf)/(1 + bowen) h = -(e + ave_shf + q). |
Measurements: | sgp15ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/30/1997 | 1700 | 07/30/1997 | 1830 |
Subject: | SGP/EBBR/E22 - Reprocess: E22 EBBR Soil Heat Flow Probe #5 Incorrect Data |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1 |
Description: | Soil heat flow probe #5 produced large negative (and incorrect) values during the period above. The probe was replaced. Somehow, two half hour records were produced at 1900 GMT as a result, the first record having 9s for the data fields; the second record is correct. Values affected (and thus incorrect) include: 15 minute: mv_hft5 30 minute: shf5, c_shf5, g5, ave_shf, e, h. The latent and sensible heat fluxes can be recalculated by using only soil heat flows 1, 2, 3, and 4 to calculate the average soil heat flow, ave_shf = (g1 + g2 + g3 + g4)/4 e = -(q + ave_shf)/(1 + bowen) h = -(e + ave_shf + q) |
Measurements: | sgp30ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/05/1997 | 2330 | 10/15/1997 | 1830 |
Subject: | SGP/EBBR/E8 - Reprocess: E8 EBBR Lightning Strike / sensor Replacements |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | DQR No: Platform: EBBR Subject: E8 EBBR Lightning Strike Date Submitted: October 17, 1997 Submitted By: _x_ Instrument Mentor ___ EST Member ___ Science Team Member ___ Other Platform/Measurement: EBBR/Numerous What level data: (raw,a0,a1,b1,c1 etc): a1 (15 and 30 minute) a0 (5 minute) What location was the data collected at: E8, Coldwater, KS Period of time in question Begin Date 08/05/97 Time 2330 GMT End Date 10/15/97 Time 1830 GMT Discussion of Problem: A lightning "strike" to the EF 8 Facility shortly after 0900 GMT on 14 July 1997 caused the failure of several EBBR sensors, the solar panel voltage regulator, and the CR10 datalogger. A previous DQR indicated this. During the period listed above, all but one of the damaged parts were replaced. Replacements occurred on 22 July, 5 August, 19 August, and 16 September. Some sensors began producing correct data immediately after their replacement; however the temperature measurements were incorrect until 15 October, when the reference temperature probe, with which they are corrected, was reconnected. Only the soil heat flow probe #2 remains to be replaced. Average soil, sensible, and latent heat fluxes are therefore now ,still, incorrect, but can be recomputed using only four sets of soil heat flow probes, as below. ave_shf = (g1 +g3 + g4 + g5)/4 e = -(q + ave_shf)/(1 + bowen) h = -(e + ave_shf + q) |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/16/1997 | 1533 | 04/16/1997 | 1625 |
Subject: | SGP/EBBR/E7 - EBBR Biannual Checks |
DataStreams: | sgp30ebbrE7.b1, sgp5ebbrE7.b1, sgp15ebbrE7.b1 |
Description: | During EBBR 6-month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for periods indicated. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/29/1997 | 1447 | 05/29/1997 | 1534 |
Subject: | SGP/EBBR/E2 - EBBR Biannual Checks |
DataStreams: | sgp15ebbrE2.b1, sgp5ebbrE2.b1, sgp30ebbrE2.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for periods indicated. |
Measurements: | sgp30ebbrE2.b1:
sgp15ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/29/1997 | 2003 | 04/29/1997 | 2030 |
Subject: | SGP/EBBR/E8 - EBBR Biannual Checks |
DataStreams: | sgp15ebbrE8.b1, sgp5ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for periods indicated. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/15/1997 | 1845 | 04/15/1997 | 1915 |
Subject: | SGP/EBBR/E9 - EBBR Biannual Checks |
DataStreams: | sgp15ebbrE9.b1, sgp5ebbrE9.b1, sgp30ebbrE9.b1 |
Description: | During EBBR 6 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for periods indicated. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/10/1997 | 1652 | 04/10/1997 | 1758 |
Subject: | SGP/EBBR/E12 - EBBR Biannual Checks |
DataStreams: | sgp30ebbrE12.b1, sgp15ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for periods indicated. |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/28/1997 | 1419 | 04/28/1997 | 1445 |
Subject: | SGP/EBBR/E13 - EBBR Biannual Checks |
DataStreams: | sgp15ebbrE13.b1, sgp5ebbrE13.b1, sgp30ebbrE13.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for periods indicated. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/22/1997 | 1525 | 04/22/1997 | 1546 |
Subject: | SGP/EBBR/E15 - EBBR Biannual Checks |
DataStreams: | sgp30ebbrE15.b1, sgp5ebbrE15.b1, sgp15ebbrE15.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for periods indicated. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/09/1997 | 2120 | 04/09/1997 | 2144 |
Subject: | SGP/EBBR/E20 - EBBR Biannual Checks |
DataStreams: | sgp30ebbrE20.b1, sgp15ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for periods indicated. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/23/1997 | 1525 | 04/23/1997 | 1605 |
Subject: | SGP/EBBR/E22 - EBBR Biannual Checks |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | During EBBR 6 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for periods indicated. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/23/1997 | 2008 | 04/23/1997 | 2050 |
Subject: | SGP/EBBR/E26 - EBBR Biannual Checks |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for periods indicated. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/10/1997 | 2000 | 10/23/1997 | 1700 |
Subject: | SGP/EBBR/E18 - Left RH Probe Temperature Incorrect |
DataStreams: | sgp15ebbrE18.b1, sgp30ebbrE18.b1 |
Description: | The left humidity probe temperature produced incorrect values from the time of installation of the E18 EBBR until 23 October at 1700 GMT because of a poor connection. |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/08/1997 | 1431 | 07/08/1997 | 1500 |
Subject: | SGP/EBBR/E2 - EBBR E2 Half Hour Incorrect |
DataStreams: | sgp30ebbrE2.b1 |
Description: | During one half hour period all data is incorrect. The problem appears to be just a bad record as the two 15 minute periods during the half hour were not affected. The cause of the problem is unknown. |
Measurements: | sgp30ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/10/1997 | 0000 | 09/30/1997 | 1550 |
Subject: | SGP/EBBR/E9 - Wind Direction Sensor Malfunction |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Description: | During numerous intermittent times during the period listed, the wind direction sensor at E9, Ashton, KS malfunctioned, putting out a value of 135. The sensor was replaced on 30 September 1997, at 1550 GMT. All wind direction data with values of 135 during the period listed should be considered incorrect. The affected data are: 5 minute: res_ws, wind_d, sigma_d 15 minute: mv_wind_d 30 minute: res_ws, wind_d, sigma_d |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/10/1997 | 0000 | 09/10/1997 | 2359 |
Subject: | SGP/EBBR/E26 - Bad Dates |
DataStreams: | sgp30ebbrE26.b1 |
Description: | At the end of the netcdf 30 minute file for 10 September 1997 there are two half hours with date 9 September 1997, and time 2200 GMT. It is unknown how this happens, but it does occasionally. Those two half hour records should be ignored. |
Measurements: | sgp30ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/14/1997 | 0900 | 10/29/1997 | 2000 |
Subject: | SGP/EBBR/E8 - Reprocess: SHF2 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil heat flow probe #2 outputted incorrect, large positive values during the period indicated. Replacement of the probe on 29 October 1997 corrected the problem. The serial number of the probe replaced is 923025 and of the replacement probe is 923110. The data affected (and thus incorrect) includes: 15 minute: mv_hft2 30 minute: shf2, c_shf2, g2, ave_shf, e, h. The latent and sensible heat fluxes can be recalculated by using only soil heat flows 1, and 3 through 5 to calculate the average soil heat flow, ave_shf = (g1 + g3 + g4 + g5)/4 e = -(q + ave_shf)/(1 + bowen) h = -(e + ave_shf + q) |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/30/1997 | 1500 | 12/31/1997 | 1520 |
Subject: | SGP/EBBR/E19 - Battery Voltage Drop |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | EBBR battery voltage at E19, El Reno, OK declined to lower than acceptable levels after long periods of clouds after initial installation. The low voltage condition worsened as Autumn progressed. In November 1997 the battery tested okay, but would not hold sufficient capacity at night or during extended periods of clouds to keep the voltage high enough for the EBBR to work properly. At my request, the battery was replaced with a brand new battery on 31 December 1997. The new battery is presently maintaining voltage level at night. The voltage dropped to 10 volts or less many days before the battery change. Voltages less than 10.3 volts affected a wide range of measured values, but at different times. When the voltage was much below 10 volts the home signal became very large (several volts), a quirk of the AEM circuitry. At various times, latent and sensible heat fluxes, soil temperature, air temperature, relative humidity, vapor pressure, wind direction, soil heat flux plate output, home signal, and atmospheric pressure (at least) were not reported correctly by the CR10 datalogger. At least the sensible and latent heat fluxes should be considered to be incorrect when the battery voltage is less than 10.3 volts. The battery voltage is found in the 15 minute data. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/30/1997 | 1500 | 07/20/1999 | 1530 |
Subject: | SGP/EBBR/E19 - Noisy Sensible and Latent Heat Fluxes |
DataStreams: | sgp30ebbrE19.b1, sgp5ebbrE19.b1, sgp15ebbrE19.b1 |
Description: | "Noisy" sensible and latent heat fluxes often occur during daytime periods when the net radiation is smaller (less than 150 watts/m2), but this is not the only condition under which it occurs. During this condition especially, the temperature gradient, RH gradient, and thus the vapor pressure gradient are near zero; this causes the sensible and latent heat flux gradients to be inaccurately calculated. Good examples of these conditions occurred on 28 December 1997 and 6 January 1998. However, even when there is substantial net radiation the problem does sometimes occur; good examples of this are found during daytime or nighttime variously on 1, 2, 3, and 4 January 1998. The conditions occur more fequently than at other EBBR sites; they do occur at other EBBR sites, but less frequently. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/01/1997 | 0000 | 04/21/2000 | 1640 |
Subject: | SGP/EBBR/E22 - Reprocess: E22 Soil Moisture Values Off End of Calibration at High Soil |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1 |
Description: | Intermittently, since November 1997 soil moisture probes 1, 2, and 3 indicate small positive or else negative values when soil moisture is high (greater than 35%). This situation is similar to that at the Cement EBBR, although the soil at Cordell is not capable of holding as much water as that at Cement. The cause of the problem is inadequate range in the calibration curve equation. Very high soil moisture cannot be obtained with the calibration soil used by REBS. The soil moisture equations could be manipulated to allow larger soil moistures to be calculated. However, the infrequency of occurence does not justify doing so. The soil moisture data must be inspected by the data user to see when soil moisture values for one or more probes is inconsistent with the others. Recalculation of ave_shf, e, and h must be made using only the soil moisture sets with correct soil moisture values. Affected data values are (# for probe number): 15 minute: rr_sm# 30 minute: sm#, c_shf#, cs#, ces#, g#, ave_shf, e, h Sensible, latent, and soil heat fluxes can be recalculated using the soil probe sets that provide correct data. For example, if soil moisture 2 is incorrect, ave_shf = (g1 + g3 + g4 + g5)/4 e = -(q + ave_shf)/(1 +bowen) h = -(e + ave_shf +q), If soil moistures 1 and 2 are incorrect, ave_shf = (g3 + g4 + g5)/3 e = -(q + ave_shf)/(1 +bowen) h = -(e + ave_shf +q), If soil moistures 1, 2, and 3 are incorrect, ave_shf = (g4 + g5)/2 e = -(q + ave_shf)/(1 +bowen) h = -(e + ave_shf +q), etc. |
Measurements: | sgp30ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/01/1997 | 0000 | 01/08/1998 | 1711 |
Subject: | SGP/EBBR/E2 - Reprocess: E2 Soil Heat Flow #4 Incorrect - Replaced |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1 |
Description: | Soil heat flow probe #4 produced intermittently large positive and negative values from October 1997 until 3 January 1998, after which the values remained high until the probe was replaced on 8 January 1998 at 1711 GMT. The following data values were affected for the period listed above (when soil heat flow #4 was large): 15 minute: mv_hft4 30 minute: shf4, c_shf4, g4, ave_shf, e, h |
Measurements: | sgp15ebbrE2.b1:
sgp30ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/09/1998 | 0250 | 01/10/1998 | 1710 |
12/08/1997 | 1000 | 12/08/1997 | 1600 |
12/21/1997 | 1210 | 12/22/1997 | 1840 |
01/04/1998 | 1045 | 01/05/1998 | 2010 |
Subject: | SGP/EBBR/E2 - EBBR Frozen Wind Speed |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | Wind speed cups were frozen during the indicated time periods. All 5, 15, and 30 minute data values indicated are incorrect for the periods listed. |
Measurements: | sgp15ebbrE2.b1:
sgp30ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/21/1997 | 0715 | 12/21/1997 | 1115 |
01/04/1998 | 0745 | 01/05/1998 | 1710 |
01/09/1998 | 0945 | 01/09/1998 | 1610 |
Subject: | SGP/EBBR/E9 - EBBR Frozen Wind Speed |
DataStreams: | sgp30ebbrE9.b1, sgp5ebbrE9.b1, sgp15ebbrE9.b1 |
Description: | Wind speed cups were frozen during the indicated time periods. All 5, 15, and 30 minute data values indicated are incorrect for the periods listed. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/21/1997 | 0440 | 12/21/1997 | 1820 |
01/04/1998 | 0525 | 01/05/1998 | 1830 |
01/09/1998 | 2150 | 01/12/1998 | 0330 |
Subject: | SGP/EBBR/E15 - EBBR Frozen Wind Speed |
DataStreams: | sgp5ebbrE15.b1, sgp30ebbrE15.b1, sgp15ebbrE15.b1 |
Description: | Wind speed cups were frozen during the indicated time periods. All 5, 15, and 30 minute data values indicated are incorrect for the periods listed. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/04/1998 | 0950 | 01/04/1998 | 1950 |
01/08/1998 | 0850 | 01/08/1998 | 1645 |
Subject: | SGP/EBBR/E7 - EBBR Frozen Wind Speed |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | Wind speed cups were frozen during the indicated time periods. All 5, 15, and 30 minute data values indicated are incorrect for the periods listed. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/21/1997 | 0830 | 12/22/1997 | 1700 |
01/09/1998 | 0000 | 01/09/1998 | 1300 |
Subject: | SGP/EBBR/E8- EBBR Frozen Wind Speed |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | Wind speed cups were frozen during the indicated time periods. All 5, 15, and 30 minute data values indicated are incorrect for the periods listed. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/04/1998 | 0750 | 01/04/1998 | 1900 |
01/08/1998 | 0705 | 01/08/1998 | 1925 |
Subject: | SGP/EBBR/E12- EBBR Frozen Wind Speed |
DataStreams: | sgp5ebbrE12.b1, sgp15ebbrE12.b1, sgp30ebbrE12.b1 |
Description: | Wind speed cups were frozen during the indicated time periods. All 5, 15, and 30 minute data values indicated are incorrect for the periods listed. |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/04/1998 | 0920 | 01/04/1998 | 1710 |
01/07/1998 | 0745 | 01/09/1998 | 2359 |
Subject: | SGP/EBBR/E19- EBBR Frozen Wind Speed |
DataStreams: | sgp5ebbrE19.b1, sgp30ebbrE19.b1, sgp15ebbrE19.b1 |
Description: | Wind speed cups were frozen during the indicated time periods. All 5, 15, and 30 minute data values indicated are incorrect for the periods listed. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/04/1998 | 0625 | 01/05/1998 | 0115 |
Subject: | SGP/EBBR/E13- EBBR Frozen Wind Speed |
DataStreams: | sgp5ebbrE13.b1, sgp30ebbrE13.b1, sgp15ebbrE13.b1 |
Description: | Wind speed cups were frozen during the indicated time periods. All 5, 15, and 30 minute data values indicated are incorrect for the periods listed. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/08/1998 | 0330 | 01/08/1998 | 1940 |
Subject: | SGP/EBBR/E20- EBBR Frozen Wind Speed |
DataStreams: | sgp5ebbrE20.b1, sgp30ebbrE20.b1, sgp15ebbrE20.b1 |
Description: | Wind speed cups were frozen during the indicated time periods. All 5, 15, and 30 minute data values indicated are incorrect for the periods listed. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/08/1998 | 0815 | 01/08/1998 | 1420 |
Subject: | SGP/EBBR/E25- EBBR Frozen Wind Speed |
DataStreams: | sgp15ebbrE25.b1, sgp30ebbrE25.b1, sgp5ebbrE25.b1 |
Description: | Wind speed cups were frozen during the indicated time periods. All 5, 15, and 30 minute data values indicated are incorrect for the periods listed. |
Measurements: | sgp30ebbrE25.b1:
sgp5ebbrE25.b1:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/07/1998 | 1950 | 01/08/1998 | 2230 |
Subject: | SGP/EBBR/E26- EBBR Frozen Wind Speed |
DataStreams: | sgp5ebbrE26.b1, sgp15ebbrE26.b1, sgp30ebbrE26.b1 |
Description: | Wind speed cups were frozen during the indicated time periods. All 5, 15, and 30 minute data values indicated are incorrect for the periods listed. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/08/1997 | 0200 | 12/08/1997 | 1615 |
12/21/1997 | 2130 | 12/21/1997 | 2131 |
12/22/1997 | 0030 | 12/22/1997 | 1830 |
Subject: | SGP/EBBR/E2 - EBBR AEMs Not Exchanging (Iced Or Fuse Blown) |
DataStreams: | sgp15ebbrE2.b1, sgp5ebbrE2.b1, sgp30ebbrE2.b1 |
Description: | The EBBR AEM stopped exchanging because of icing at during this time period. The calculations of Bowen ratio, latent heat flux, and sensible heat flux are incorrect for the time period indicated. The other measurements listed are also incorrect for those times: |
Measurements: | sgp30ebbrE2.b1:
sgp15ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/04/1998 | 0630 | 01/05/1998 | 1830 |
12/21/1997 | 0745 | 12/21/1997 | 1715 |
Subject: | SGP/EBBR/E15 - EBBR AEMs Not Exchanging (Iced Or Fuse Blown) |
DataStreams: | sgp30ebbrE15.b1, sgp5ebbrE15.b1, sgp15ebbrE15.b1 |
Description: | The EBBR AEM stopped exchanging because of icing at during this time period. The calculations of Bowen ratio, latent heat flux, and sensible heat flux are incorrect for the time period indicated. The other measurements listed are also incorrect for those times: |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/04/1998 | 1845 | 01/08/1998 | 1715 |
Subject: | SGP/EBBR/E9 - EBBR AEMs Not Exchanging (Iced Or Fuse Blown) |
DataStreams: | sgp15ebbrE9.b1, sgp5ebbrE9.b1, sgp30ebbrE9.b1 |
Description: | The EBBR AEM stopped exchanging because of icing at during this time period. The calculations of Bowen ratio, latent heat flux, and sensible heat flux are incorrect for the time period indicated. The other measurements listed are also incorrect for those times: |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/04/1998 | 1300 | 01/15/1998 | 1715 |
Subject: | SGP/EBBR/E12 - EBBR AEMs Not Exchanging (Iced Or Fuse Blown) |
DataStreams: | sgp30ebbrE12.b1, sgp15ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | The EBBR AEM stopped exchanging because of icing at during this time period. The calculations of Bowen ratio, latent heat flux, and sensible heat flux are incorrect for the time period indicated. The other measurements listed are also incorrect for those times: Note: Mentor supplied the end date/time for this DQR after submittal |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/04/1998 | 0745 | 01/16/1998 | 1600 |
Subject: | SGP/EBBR/E13 - EBBR AEMs Not Exchanging (Iced Or Fuse Blown) |
DataStreams: | sgp15ebbrE13.b1, sgp5ebbrE13.b1, sgp30ebbrE13.b1 |
Description: | The EBBR AEM stopped exchanging because of icing at during this time period. The calculations of Bowen ratio, latent heat flux, and sensible heat flux are incorrect for the time period indicated. The other measurements listed are also incorrect for those times: Note: Mentor supplied the end date/time for this DQR after submittal |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/08/1998 | 1100 | 01/14/1998 | 1545 |
Subject: | SGP/EBBR/E20 - EBBR AEMs Not Exchanging (Iced Or Fuse Blown) |
DataStreams: | sgp30ebbrE20.b1, sgp5ebbrE20.b1, sgp15ebbrE20.b1 |
Description: | The EBBR AEM stopped exchanging because of icing at during this time period. The calculations of Bowen ratio, latent heat flux, and sensible heat flux are incorrect for the time period indicated. The other measurements listed are also incorrect for those times: Note: Mentor supplied the end date/time to this DQR after submission. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/16/1997 | 1600 | 12/29/1997 | 1730 |
Subject: | SGP/EBBR/E15 - EBBR Net Radiometer Domes Pierced, Water Inside - Data Affected |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | The upper dome of the EBBR net radiometers was punctured. After the puncture, water collected in the top and/or bottom dome. The water was removed and the dome was replaced during PM visits. In some cases, the puncture and water resulted in incorrect data. Comparison with net radiation values from surrounding EBBR sites and the SIRS is neccesary to determine if the EBBR net radiation data are correct or incorrect. It is often difficult to tell from the data when the punctures occurred, but they would have existed for no more than 2 weeks, since the domes are checked during each two-weekly PM visit. Incorrect net radiation data results in the latent heat flux and sensible heat flux also being incorrect. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/27/1997 | 0000 | 05/06/1998 | 1645 |
Subject: | SGP/EBBR/E25 - Bowen Ratio Usually Negative - Fluxes Incorrect |
DataStreams: | sgp15ebbrE25.b1, sgp30ebbrE25.b1, sgp5ebbrE25.b1 |
Description: | Submitted By: _x_ Instrument Mentor ___ EST Member ___ Science Team Member ___ Other Platform/Measurement: EBBR/Bowen ratio, sensible and latent heat fluxes What level data: (raw,a0,a1,b1,c1 etc): a1 (15 and 30 minute) What location was the data collected at: E25, Seminole, OK Period of time in question (Times vary for different days) Begin Date 10/22/97 and on Discussion of Problem: Bowen ratio is normally negative, as if it is somehow being biased. Latent and sensible heat fluxes are apparently incorrect much of the time, especially during the daytime. All supporting measurements seem to be incorrect; therefore the error must be occurring in the measured T and RH gradients. This situation has occured since the EBBR unit was installed on 22 October 1997. A work request has been submitted for a full six month check to be performed on the system. Hopefully the checks will point to the source of the problem. The right relative humidity was biased. The right T/RH probe was replaced on 4/22/98 at 1700 GMT, but it had a more defective RH sensor in it, causing right RH to sit at 112-114 percent. This probe was itself replaced on 5/6/98 at 1645 GMT, thereby taking care of the negative Bowen ratio problem. |
Measurements: | sgp30ebbrE25.b1:
sgp5ebbrE25.b1:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/24/1997 | 1630 | 01/08/1998 | 1800 |
Subject: | SGP/EBBR/E9 - EBBR Net Radiometer Domes Pierced, Water Inside - Data Affected |
DataStreams: | sgp30ebbrE9.b1, sgp5ebbrE9.b1, sgp15ebbrE9.b1 |
Description: | The upper dome of the EBBR net radiometers was punctured. After the puncture, water collected in the top and/or bottom dome. The water was removed and the dome was replaced during PM visits. In some cases, the puncture and water resulted in incorrect data. Comparison with net radiation values from surrounding EBBR sites and the SIRS is neccesary to determine if the EBBR net radiation data are correct or incorrect. It is often difficult to tell from the data when the punctures occurred, but they would have existed for no more than 2 weeks, since the domes are checked during each two-weekly PM visit. Incorrect net radiation data results in the latent heat flux and sensible heat flux also being incorrect. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/18/1997 | 1700 | 12/31/1997 | 1615 |
Subject: | SGP/EBBR/E12 - EBBR Net Radiometer Domes Pierced, Water Inside - Data Affected |
DataStreams: | sgp5ebbrE12.b1, sgp30ebbrE12.b1, sgp15ebbrE12.b1 |
Description: | The upper dome of the EBBR net radiometers was punctured. After the puncture, water collected in the top and/or bottom dome. The water was removed and the dome was replaced during PM visits. In some cases, the puncture and water resulted in incorrect data. Comparison with net radiation values from surrounding EBBR sites and the SIRS is neccesary to determine if the EBBR net radiation data are correct or incorrect. It is often difficult to tell from the data when the punctures occurred, but they would have existed for no more than 2 weeks, since the domes are checked during each two-weekly PM visit. Incorrect net radiation data results in the latent heat flux and sensible heat flux also being incorrect. |
Measurements: | sgp30ebbrE12.b1:
sgp15ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/16/1997 | 2000 | 12/29/1997 | 1840 |
Subject: | SGP/EBBR/E26 - EBBR Net Radiometer Domes Pierced, Water Inside - Data Affected |
DataStreams: | sgp5ebbrE26.b1, sgp30ebbrE26.b1, sgp15ebbrE26.b1 |
Description: | The upper dome of the EBBR net radiometers was punctured. After the puncture, water collected in the top and/or bottom dome. The water was removed and the dome was replaced during PM visits. In some cases, the puncture and water resulted in incorrect data. Comparison with net radiation values from surrounding EBBR sites and the SIRS is neccesary to determine if the EBBR net radiation data are correct or incorrect. It is often difficult to tell from the data when the punctures occurred, but they would have existed for no more than 2 weeks, since the domes are checked during each two-weekly PM visit. Incorrect net radiation data results in the latent heat flux and sensible heat flux also being incorrect. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/08/1998 | 0100 | 01/08/1998 | 2030 |
Subject: | SGP/EBBR/E26 - EBBR AEMs Not Exchanging (Iced Or Fuse Blown) |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | The EBBR AEM stopped exchanging because of icing at during this time period. The calculations of Bowen ratio, latent heat flux, and sensible heat flux are incorrect for the time period indicated. The other measurements listed are also incorrect for those times: |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/27/1997 | 0530 | 12/18/1997 | 1805 |
Subject: | SGP/EBBR/E12 - AEM Failure and Replacement |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | The E12 AEM did not exchange for most of the period above, yielding small negative home signals. The AEM was replaced on 18 December by site operations personnel. All 30 minute e, h, and bowen data are incorrect for the times when the home signal is incorrect, and the temperature, relative humidity, and vapor pressures do not represent properly switched values. |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/04/1997 | 0000 | 01/06/1998 | 1840 |
Subject: | SGP/EBBR/E8 - Battery Low - Data Affected |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The battery condition deteriorated over time so that the solar panel and AC charger were not always able to keep the battery voltage high enough to produce correct data after 3 December 1997, especially at night. The condition became particularly bad after 8 December 1997. Data for conditions of battery voltage below 10.4 volts should be considered incorrect; inspection of the battery voltage data is required to determine this. The battery was replaced on 6 January 1998 at 1840 GMT, after which data was again correct. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/24/1997 | 1530 | 01/08/1998 | 1515 |
Subject: | SGP/EBBR/E2 - EBBR Net Radiometer Domes Pierced, Water Inside - Data Affected |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | The upper dome of the EBBR net radiometers was punctured. After the puncture, water collected in the top and/or bottom dome. The water was removed and the dome was replaced during PM visits. In some cases, the puncture and water resulted in incorrect data. Comparison with net radiation values from surrounding EBBR sites and the SIRS is neccesary to determine if the EBBR net radiation data are correct or incorrect. It is often difficult to tell from the data when the punctures occurred, but they would have existed for no more than 2 weeks, since the domes are checked during each two-weekly PM visit. Incorrect net radiation data results in the latent heat flux and sensible heat flux also being incorrect. |
Measurements: | sgp30ebbrE2.b1:
sgp15ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/11/1997 | 1900 | 10/29/1997 | 2010 |
Subject: | SGP/EBBR/E8 - Reprocess: E8 Soil Heat Flow #2 Failure - Replaced |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil heat flow probe #2 values were positive and in the thousands after an apparently catastrophic failure at 1900 GMT on 11 October 1997. The probe was replaced on 29 October 1997 at 2010 GMT. The data affected (and thus incorrect) includes: 15 minute: mv_hft2 30 minute: shf2, c_shf2, g2, ave_shf, e, h. The latent and sensible heat fluxes can be recalculated by using only soil heat flows 1 and 3 through 5 to calculate the average soil heat flow, ave_shf = (g1 + g3 + g4 + g5)/4 e = -(q + ave_shf)/(1 + bowen) h = -(e + ave_shf + q). |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/28/1997 | 1600 | 02/04/1998 | 2230 |
Subject: | SGP/EBBR/E7 - Right T/RH Probe Temperature Incorrect |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1, sgp5ebbrE7.00 |
Description: | The temperature portion of the right T/RH probe was reporting too high for the period. The following fields are incorrect: 5 minute: thum_top, thum_bot, vp_top, vp_bot 15 minute: rr_thum-r 30 minute: thum_top, thum_bot, vp_top, vp_bot, e, h, bowen |
Measurements: | sgp5ebbrE7.00:
sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/13/1998 | 0330 | 01/15/1998 | 1630 |
Subject: | SGP/EBBR/E20 - Wind Speed Sensor Frozen |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1, sgp5ebbrE20.00 |
Description: | The wind speed sensor was frozen during the period. Data values that were incorrect are: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp5ebbrE20.b1:
sgp5ebbrE20.00:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/14/1998 | 0630 | 01/14/1998 | 1100 |
Subject: | SGP/EBBR/E2 - EBBR E2 Wind Direction Sensors Frozen |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | Wind direction sensors were frozen stalled during this period. Wind direction from 0630 through 1100 GMT. Data values that were affected were: 5 minute: sigma_wd, wind_d 15 minute: mv_wind_d 30 minute: sigma_wd, wind_d |
Measurements: | sgp30ebbrE2.b1:
sgp15ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/14/1998 | 0430 | 01/14/1998 | 1830 |
Subject: | SGP/EBBR/E7 - Wind Speed Sensor Frozen |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | The wind speed sensor was frozen stalled during this period. Data values that were affected are: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/14/1998 | 0745 | 01/14/1998 | 1600 |
Subject: | SPG/EBBR/E8 - Wind Speed Sensor and AEM Frozen |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The Automatic Exchange Mechanism was frozen during the entire period and did not exchange. Data values that were affected by the inoperative AEM are: 5 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot 15 minute: rr_thum_r, rr_thum_l, mv_hum_r, mv_hum_l, tair_r, tair_l 30 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, e, h, bowen The wind speed sensor was frozen stalled during part of this period (1100-1600 GMT). Data values that were affected are: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/14/1998 | 0500 | 01/15/1998 | 0030 |
Subject: | SGP/EBBR/E9 - Wind Speed Sensor Frozen |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1, sgp5ebbrE9.00 |
Description: | The wind speed sensor was frozen stalled during this period. Data values that were affected are: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp30ebbrE9.b1:
sgp5ebbrE9.00:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/15/1998 | 1015 | 01/15/1998 | 1445 |
Subject: | SGP/EBBR/E13 - Wind Speed Sensor Frozen |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | The wind speed sensor was frozen during the period. Data values that were incorrect are: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/14/1998 | 0915 | 01/14/1998 | 1920 |
Subject: | SGP/EBBR/E15 - Wind Speed Sensor Frozen |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1, sgp5ebbrE15.00 |
Description: | The wind speed sensor was frozen during the period. Data values that were incorrect are: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp5ebbrE15.00:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/15/1998 | 0300 | 01/15/1998 | 1415 |
Subject: | SGP/EBBR/E22 - Low Battery - Resulting In Incorrect Data |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | A low battery condition (less than 10.5 volts) caused a number of data fields to be incorrect at some times and all data fields to be incorrect at a more limited number of times. All data fields are suspect when the battery voltage drops below 10.5 volts because the CR10 cannot operate properly. Some sensors are affected slightly (i.e. pressure) at slightly larger voltages. The low battery condition resulted from an initially weak battery and a week of extremely low solar radiation (from heavy cloudiness) on the solar panel that provides most of the charging of the battery. Comments added by Marv Wesely: David Cook has been working with Dave Breedlove to try to assess the age of the batteries. The dates on the batteries needs to be read during site visits and provided to David C. It's quite possible that they need to be replaced. It's work in progress, though slow at times. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/19/1998 | 2200 | 01/21/1998 | 0230 |
Subject: | SGP/EBBR/E2 - EBBR E2 Incorrect Data |
DataStreams: | sgp30ebbrE2.b1, sgp5ebbrE2.b1, sgp15ebbrE2.b1 |
Description: | During the indicated period, when either home signal is in the hundreds, all soil heat fluxes, soil moistures, net radiation, air temperatures, bowen ratio, sensible heat flux, and latent heat flux are incorrect. This condition usually results from an AEM circuit problem that allows the home signals to go high and corrupts the data. The cause has never been determined. Comments added by Marv Wesely: David Cook said that this is a very rare occurence; it's happened about three times in total for all EBBRs. The problem might be on the circuit boards (moisture?), but that is only a guess. The problem has always disappeared by itself after a while. |
Measurements: | sgp30ebbrE2.b1:
sgp15ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/31/1997 | 1602 | 10/31/1997 | 1640 |
Subject: | SGP/EBBR/E2 - EBBR E2 3 Month Check |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for the period indicated. |
Measurements: | sgp30ebbrE2.b1:
sgp15ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/16/1997 | 1928 | 10/16/1997 | 2002 |
Subject: | SGP/EBBR/E9 - 3-Month Check |
DataStreams: | sgp30ebbrE9.b1, sgp5ebbrE9.b1, sgp15ebbrE9.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for the period above. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/23/1997 | 1955 | 10/23/1997 | 2015 |
Subject: | SGP/EBBR/E12 - 3-Month Check |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for the period above. |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/10/1997 | 1850 | 10/10/1997 | 1932 |
Subject: | SGP/EBBR/E13 - 3-Month Check |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for the period above. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/07/1997 | 1750 | 10/07/1997 | 1832 |
Subject: | SGP/EBBR/E15 - 3-Month Check |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for the period above. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/23/1997 | 1355 | 10/23/1997 | 1420 |
Subject: | SGP/EBBR/E18 - 3 Month Check |
DataStreams: | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for the period above. |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/09/1997 | 1528 | 10/09/1997 | 1602 |
Subject: | SGP/EBBR/E19 - 3-Month Check |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for the period indicated. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/22/1997 | 1645 | 10/22/1997 | 1705 |
Subject: | SGP/EBBR/E20 - 3 Month Check |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for the period above. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/08/1997 | 1830 | 10/08/1997 | 1902 |
Subject: | SGP/EBBR/E22 - 3-Month Check |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for the period above. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/21/1997 | 2125 | 10/21/1997 | 2148 |
Subject: | SGP/EBBR/E26 - 3-Month Check |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for the period indicated. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/14/1998 | 1820 | 03/15/1998 | 1625 |
Subject: | SGP/EBBR/E8 - Wind Speed Frozen |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | During the period above the wind speed sensor cups were frozen stalled by freezing rain. The data fields affected are: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/16/1998 | 1530 | 03/16/1998 | 1830 |
Subject: | SGP/EBBR/E8 - Wind Speed Frozen |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | During the period above the wind speed sensor cups were frozen stalled by freezing rain. The data fields affected are: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/17/1998 | 1510 | 03/17/1998 | 1555 |
Subject: | SGP/EBBR/E8 - Wind Speed Frozen |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | During the period above the wind speed sensor cups were frozen stalled by freezing rain. The data fields affected are: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/16/1997 | 1528 | 10/16/1997 | 1602 |
Subject: | SGP/EBBR/E7 - 3-Month Check |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for the period above. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/29/1997 | 1930 | 10/29/1997 | 2125 |
Subject: | SGP/EBBR/E8 - 3-Month Check |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | During EBBR 3-month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for the period above. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/18/1997 | 0130 | 06/18/1997 | 0200 |
Subject: | SGP/EBBR/E26 - Character In EBBR E26, 30 Minute Data |
DataStreams: | sgp30ebbrE26.b1 |
Description: | A non-numeric character appears in the field for shf1 in the 30 minute data file at the time shown. Data processing programs that assume that a numeric character is there will display an error condition. I had to create a short routine to put in my analysis program to avoid using that data point. |
Measurements: | sgp30ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/03/1998 | 0000 | 04/21/1998 | 1515 |
Subject: | SGP/EBBR/E15 - Pressure Diurnal Fluctuations |
DataStreams: | sgp30ebbrE15.b1 |
Description: | During this period the barometric pressure exhibited increasingly larger, and unatural, diurnal fluctuations. This has been observed at other sites previously. The sensor was replaced on April 21. The diurnal fluctuations had only a minor effect on the calculated fluxes of latent and sensible heat; the fluxes are not incorrect as a result of the problem. |
Measurements: | sgp30ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/21/1998 | 1530 | 04/28/1998 | 1530 |
Subject: | SGP/EBBR/E15 - Pressure Too High |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | During this period the barometric pressure was consistently much greater than the true pressure, hanging around 113 kPa all of the time. The defective sensor was replaced on April 28. The calculated fluxes of latent and sensible heat are incorrect during the period. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/12/1998 | 0930 | 05/01/1998 | 1530 |
Subject: | SGP/EBBR/E15 - Left T/RH Probe Temperature Too Large |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | Beginning in February, the temperature of the left T/RH probe reported too high a value at some times. The frequency of the problem increased with time until 28 April at 1520 GMT, when a defective probe was installed; the latter probe temperature indicated very high values. This continued until 1 May at 1530 GMT when a working probe was installed. Times of incorrect humidity probe temperatures can be determined by comparing 5 or 30 minute values of air temperature and humidity probe temperature; at these times the calculated bowen ratio, and the fluxes of latent and sensible heat are incorrect. The incorrect variables were: 5 minute: thum_top, thum_bot, vp_top, vp_bot 15 minute: rr_thum_l 30 minute: thum_top, thum_bot, vp_top, vp_bot, bowen, e, h |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/24/1998 | 1800 | 05/06/1998 | 1740 |
Subject: | SGP/EBBR/E25 - Right Humidity Incorrect |
DataStreams: | sgp15ebbrE25.b1, sgp30ebbrE25.b1, sgp5ebbrE25.b1 |
Description: | On 24 April 1998 the humidity reported by the right T/RH probe soared to 112% to 114% and remained there until replacement of the probe on 6 May 1998. Sensible and latent heat fluxes and some humidity and vapor pressure data for the period are incorrect. The data fields that are incorrect are: 5 minute: hum_top, hum_bot, vp_top, vp_bot 15 minute: mv_hum_r 30 minute: hum_top, hum_bot, vp_top, vp_bot, bowen, e, h |
Measurements: | sgp30ebbrE25.b1:
sgp5ebbrE25.b1:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/14/1997 | 2130 | 05/12/1998 | 1850 |
Subject: | EBBR E8 Net Radiation Data Quality |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | On 14 July 1997 lightning struck the EBBR at E8. A number of sensors and components needed to be replaced. One component that was not replaced, a J-panel, suffered some damage that was not obvious at the time. Also, changes in sensors eventually required a change in the CR10 program to accomodate new calibration factors. A new program was installed on 14 October. Unfortunately, an incorrect nighttime calibration for the net radiation was contained in the new program provided by the EBBR vendor REBS. The calibration problem will be covered in a separate DQR. The J-panel problem began to surface on 14 October as well, with most net radiation values being smaller than correct values during the day, and large negative at night; the latter was partially from the J-panel problem and partially from the incorrect nighttime calibration. All net radiation values, nighttime and daytime from 14 October 1997 through 12 May 1998 need to be examined carefully (and compared with other nearby sites) to determine if they are correct, and nighttime values need to be corrected. Much of the latent and sensible heat flux data is corrupted by both problems. The correction to nightime values will be covered in two subsequent DQRs, as two net radiometers, with different calibrations, have been used at E8 since 14 October 1997. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/07/1998 | 0130 | 05/21/1998 | 1830 |
Subject: | SGP/EBBR/E12 - Wind Direction Sensor Failure |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | The wind direction sensor failed, producing zero values. The cause is unknown. Data values affected are below: 5 minute: res_ws, wind_d, sigma_wd 15 minute: mv_wind_d 30 Minute: res_ws, wind_d, sigma_wd |
Measurements: | sgp30ebbrE12.b1:
sgp15ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/07/1998 | 2000 | 05/07/1998 | 2030 |
Subject: | SGP/EBBR/E13 - Six Month Checks |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | During EBBR 6 month checks (formally called 3 month checks) the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed just before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for the period indicated. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/05/1998 | 1459 | 05/05/1998 | 1530 |
Subject: | SGP/EBBR/E15 - Six Month Checks |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | During EBBR 6 month checks (formally called 3 month checks) the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed just before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for the period above. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/21/1998 | 1402 | 05/21/1998 | 1430 |
Subject: | SGP/EBBR/E18 - Six Month Checks |
DataStreams: | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
Description: | During EBBR 6 month checks (formally called 3 month checks) the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed just before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for the period above. |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/07/1998 | 1430 | 05/07/1998 | 1500 |
Subject: | SGP/EBBR/E19 - Six-Month Checks |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | During EBBR 6 month checks (formally called 3 month checks) the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed just before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for the period indicated. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/20/1998 | 1450 | 05/20/1998 | 1530 |
Subject: | SGP/EBBR/E20 - Six Month Checks |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | During EBBR 6 month checks (formally called 3 month checks) the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed just before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for the period above. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/06/1998 | 1800 | 05/06/1998 | 1840 |
Subject: | SGP/EBBR/E22 - Six Month Checks |
DataStreams: | sgp30ebbrE22.b1, sgp5ebbrE22.b1, sgp15ebbrE22.b1 |
Description: | During EBBR 6 month checks (formally called 3 month checks) the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed just before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for the period indicated. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/21/1998 | 1948 | 05/21/1998 | 2030 |
Subject: | SGP/EBBR/E25 - Six-Month Checks |
DataStreams: | sgp15ebbrE25.b1, sgp30ebbrE25.b1, sgp5ebbrE25.b1 |
Description: | During EBBR 6-month checks (formally called 3 month checks) the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed just before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for the period indicated. |
Measurements: | sgp30ebbrE25.b1:
sgp5ebbrE25.b1:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/19/1998 | 1646 | 05/19/1998 | 1717 |
Subject: | SGP/EBBR/E26 - Six Month Checks |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | During EBBR 6 month checks (formally called 3 month checks) the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed just before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for the period indicated. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/30/1998 | 1928 | 04/30/1998 | 2000 |
Subject: | SGP/EBBR/E2 - EBBR E2 Six Month Checks |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | During EBBR 6 month checks (formally called 3 month checks) the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed just before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for the period indicated. |
Measurements: | sgp30ebbrE2.b1:
sgp15ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/12/1998 | 1940 | 05/12/1998 | 2037 |
Subject: | SGP/EBBR/E7 - Six-Month Checks |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | During EBBR 6 month checks (formally called 3 month checks) the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed just before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for the period above. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/28/1998 | 1958 | 04/28/1998 | 2031 |
Subject: | SGP/EBBR/E8 - Six Month Checks |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | During EBBR 6 month checks (formally called 3 month checks) the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed just before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for the period above. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/14/1998 | 1740 | 05/14/1998 | 1759 |
Subject: | SGP/EBBR/E9 - Six Month Checks |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Description: | During EBBR 6 month checks (formally called 3 month checks) the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed just before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for the period above. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/21/1998 | 1855 | 05/21/1998 | 1918 |
Subject: | SGP/EBBR/E12 - Six Month Checks |
DataStreams: | sgp30ebbrE12.b1, sgp5ebbrE12.b1, sgp15ebbrE12.b1 |
Description: | During EBBR 6 month checks (formally called 3 month checks) the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed just before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for the period indicated. |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/14/1997 | 2200 | 04/20/1998 | 1720 |
Subject: | EBBR E8 Adjustment of Negative Net Radiation Values |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | Please see DQR D980526.2 for some background information on the cause of the problem described here. On 14 October 1997 a new CR10 program was installed at E8 EBBR. The program contained the wrong calibration coefficient for negative net radiations. A multiplier had been left out of the computation of the coefficient by the vendor. All negative net radiation values during the stated period need to be DIVIDED by 13.88. Only the 5 minute and 30 minute q values are affected. The 15 minute mv_q values are correct. Other values affected are 30 minute h and e, which would need to be recalculated from: e = -(q + ave_shf)/(1 + bowen) h = -(e + ave_shf + q) |
Measurements: | sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/20/1998 | 1725 | 05/26/1998 | 1920 |
Subject: | EBBR E8 Adjustment of Negative Net Radiation Values |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | Please see DQR D980526.2 for some background information on the cause of the problem described here, and D980603.1 for correction to the data obtained with the previous net radiometer. On 14 October 1997 a new CR10 program was installed at E8 EBBR. The program contained the wrong calibration coefficient for negative net radiations. A multiplier had been left out of the computation of the coefficient by the vendor. On 20 April 1998 a new net radiometer was installed. It was not noticed that the mutiplier was missing. A slightly different calibration was entered for the net radiometer, again without the multiplier applied. A few weeks later the mentor, in trying to diagnose the obvious incorrect negative net radiation values, discovered the missing multiplier. A new program with the multiplier included was installed on 26 May 1998 at 1917 GMT. All negative net radiation values during the stated period above need to be DIVIDED by 13.775. Only the 5 minute and 30 minute q values are affected. The 15 minute mv_q values are correct. Other values affected are 30 minute h and e, which would need to be recalculated from: e = -(q + ave_shf)/(1 + bowen) h = -(e + ave_shf + q) |
Measurements: | sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/24/1998 | 0300 | 05/26/1998 | 1700 |
Subject: | SGP/EBBR/E8 - Wind Direction Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The wind direction sensor showed the usual failure signature much of the time, an output of 135 degrees, beginning on 24 May 1998. Not all times showed 135 degrees, but all wind direction data during the stated period should be considered incorrect. The values affected are: 5 minute: res_ws, wind_d, sigma_d 15 minute: mv_wind_d 30 minute: res_ws, wind_d, sigma_d |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/10/1997 | 2000 | 06/18/1998 | 1420 |
Subject: | SGP/EBBR/E18 - Reprocess: E18 EBBR Net Radiometer Calibration |
DataStreams: | sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
Description: | When the CR10 program was loaded in the storage module of the newly installed EBBR unit at E18 on 10 September 1997, an incorrect calibration for the net radiometer was present. The incorrect calibration had apparently been entered into the storage module by the vendor before the unit was shipped. The .DOC file showed the correct calibration; therefore, we did not suspect a problem until comparisons with radiometers at other sites nearby were performed. The program with the correct calibration was entered on 18 June 1998. The 15 minute mv_q values were not affected. The 5 and 30 minute q values during the period above need to be corrected; multiply positive q values by 0.684, multiply negative values by 0.811; Sensible (h) and Latent Heat (e) fluxes for the period need to be recalculated with the equations below. e = -(q + ave_shf)/(1 + bowen) h = -(e + ave_shf + q) |
Measurements: | sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/03/1998 | 1115 | 07/15/1998 | 1930 |
Subject: | SGP/EBBR/E20 - Wind Speed Incorrect |
DataStreams: | sgp5ebbrE20.b1, sgp15ebbrE20.b1, sgp30ebbrE20.b1 |
Description: | On 3 July 1998 the cup anemometer lost a cup. Until the cup assembly was replaced by site ops personnel on 15 July 1998, wind speed and vector averaged wind speed data were incorrect. The values affected are listed below. 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/15/1998 | 1445 | 07/15/1998 | 1500 |
Subject: | SGP/EBBR/E25 - Data Incorrect |
DataStreams: | sgp15ebbrE25.b1, sgp30ebbrE25.b1, sgp5ebbrE25.b1 |
Description: | On 15 July 1998 the EBBR AEM was moved to a higher position. The complexity of the task required that the AEM be disabled for what amounted to two half hour average periods, during which the temperatures, relative humidities, vapor pressures, Bowen ratio, and latent and sensible heat fluxes were all incorrect. |
Measurements: | sgp30ebbrE25.b1:
sgp5ebbrE25.b1:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/15/1998 | 1630 | 07/15/1998 | 1700 |
Subject: | SGP/EBBR/E25 - Data Incorrect |
DataStreams: | sgp5ebbrE25.b1, sgp15ebbrE25.b1, sgp30ebbrE25.b1 |
Description: | On 15 July 1998 the EBBR CR10 program was reloaded with changed calibration coefficients for a new soil temperature probe #1. Therefore the 5, 15, and 30 minute data for 1700 GMT are incorrect. |
Measurements: | sgp30ebbrE25.b1:
sgp5ebbrE25.b1:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/30/1998 | 0345 | 10/01/1998 | 1500 |
Subject: | SGP/EBBR/E2 - EBBR E2 Net Radiation Incorrect - Water in Bottom Dome |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | Water entered the net radiometer top dome through a hole made, probably, by migrating birds (this happens every Spring and Autumn) and accumulated in the bottom dome. This causes nighttime net radiations to be too large and daytime net radiations to be too small. The data are therefore incorrect. |
Measurements: | sgp30ebbrE2.b1:
sgp15ebbrE2.b1:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/02/1998 | 0345 | 10/08/1998 | 1930 |
Subject: | SGP/EBBR/E13 - Net Radiation Incorrect - Water in Bottom Dome |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | Water entered the net radiometer top dome through a hole made, probably, by migrating birds (this happens every Spring and Autumn) and accumulated in the bottom dome. This causes nighttime net radiations to be too large and daytime net radiations to be too small. The data is therefore incorrect. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/19/1998 | 2215 | 10/21/1998 | 1600 |
Subject: | SGP/EBBR/E20 - Incorrect Data - Cattle Vandalism |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | Cattle penetrated the electric fence and pulled the aspirator housings from the AEM. Thus T/RH/VP and sensible and latent heat fluxes are incorrect for the period (until 1540 GMT; LE and H through the 30 minute period ending at 1600 GMT) |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/28/1998 | 1559 | 10/28/1998 | 1600 |
Subject: | SGP/EBBR/E7 - Data Incorrect Because of Recompile |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | All data for the 5, 15, and 30 minute periods ending at 1600 GMT are incorrect or missing because the CR10 program was recompiled. |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/05/1998 | 1200 | 11/05/1998 | 1800 |
Subject: | SGP/EBBR/E15 - Aspirator loosened from AEM mount |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | The left aspirator was loosened from the AEM mount. Thus all gradients of T/RH/VP and sensible and latent heat fluxes for the period are incorrect. T/RH/VP measurements from the left probe are incorrect only through 1745 GMT. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/15/1998 | 0100 | 11/17/1998 | 2100 |
Subject: | SGP/EBBR/E18 - Water In Net Radiometer, Flux Data Incorrect |
DataStreams: | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1 |
Description: | Incorrect Net Radiation values resulted from water in the bottom dome. Thus incorrect sensible and latent heat fluxes resulted. Affected values are: 5 minute: q 15 minute: mv_q 30 minute: q, e, h |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/30/1998 | 0600 | 11/30/1998 | 1400 |
Subject: | SGP/EBBR/E25 - All Data Incorrect (zeros) |
DataStreams: | sgp30ebbrE25.b1, sgp15ebbrE25.b1, sgp5ebbrE25.b1, sgp5ebbrE25.00 |
Description: | All data values are zero and incorrect; reason unknown. |
Measurements: | sgp30ebbrE25.b1:
sgp5ebbrE25.b1:
sgp5ebbrE25.00:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/12/1998 | 2100 | 11/25/1998 | 1830 |
Subject: | SGP/EBBR/E9 - Wind Direction Data Incorrect |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1, sgp5ebbrE9.00 |
Description: | The wind direction sensor stopped working correctly, resulting in the following values being affected: 5 minute: res_ws, wind_d, sigma_wd 15 minute: mv_wind_d 30 minute: res_ws, wind_d, sigma_wd |
Measurements: | sgp30ebbrE9.b1:
sgp5ebbrE9.00:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/20/1998 | 1505 | 12/22/1998 | 1800 |
Subject: | SGP/EBBR/E8 - Wind Direction and Speed Data Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The wind direction and speed sensors were iced stopped by freezing rain, resulting in the following values being incorrect: 5 minute: wind_s, res_ws, wind_d, sigma_wd 15 minute: wind_s, mv_wind_d 30 minute: wind_s, res_ws, wind_d, sigma_wd |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/21/1998 | 1100 | 12/24/1998 | 1830 |
Subject: | SGP/EBBR/E25 - Wind Direction and Speed Data Incorrect |
DataStreams: | sgp15ebbrE25.b1, sgp30ebbrE25.b1, sgp5ebbrE25.b1, sgp5ebbrE25.00 |
Description: | The wind direction and speed sensors were iced stopped by freezing rain, resulting in the following values being incorrect: 5 minute: wind_s, res_ws, wind_d, sigma_wd 15 minute: wind_s, mv_wind_d 30 minute: wind_s, res_ws, wind_d, sigma_wd |
Measurements: | sgp30ebbrE25.b1:
sgp5ebbrE25.b1:
sgp5ebbrE25.00:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/03/1998 | 0615 | 11/10/1998 | 1430 |
Subject: | SGP/EBBR/E22 - Battery Voltage Low, Data Incorrect |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1, sgp5ebbrE22.00 |
Description: | Battery voltages less than 10.4 volts occurred much of the time. All data should be considered incorrect when the voltage is less than 10.4 volts. |
Measurements: | sgp5ebbrE22.00:
sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/21/1998 | 0245 | 12/22/1998 | 1630 |
Subject: | SGP/EBBR/E12 - Wind Direction Data Incorrect (revised) |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | The wind direction sensor was iced stopped by freezing rain, resulting in the following values being incorrect: 5 minute: res_ws, wind_d, sigma_wd 15 minute: mv_wind_d 30 minute: res_ws, wind_d, sigma_wd |
Measurements: | sgp30ebbrE12.b1:
sgp15ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/22/1998 | 1005 | 12/22/1998 | 1600 |
Subject: | SGP/EBBR/E7 - Wind Direction Data Incorrect (revised) |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1, sgp5ebbrE7.00 |
Description: | The wind direction sensor was iced stopped by freezing rain, resulting in the following values being incorrect: 5 minute: res_ws, wind_d, sigma_wd 15 minute: mv_wind_d 30 minute: res_ws, wind_d, sigma_wd |
Measurements: | sgp5ebbrE7.00:
sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/21/1998 | 1635 | 12/23/1998 | 2000 |
Subject: | SGP/EBBR/E18 - Wind Direction Data Incorrect (revised) |
DataStreams: | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1, sgp5ebbrE18.00 |
Description: | The wind direction sensor was iced stopped by freezing rain, resulting in the following values being incorrect: 5 minute: res_ws, wind_d, sigma_wd 15 minute: mv_wind_d 30 minute: res_ws, wind_d, sigma_wd |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
sgp5ebbrE18.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/22/1998 | 1105 | 12/23/1998 | 0400 |
Subject: | SGP/EBBR/E19 - Wind Direction Data Incorrect (revised) |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1, sgp5ebbrE19.00 |
Description: | The wind direction sensor was iced stopped by freezing rain, resulting in the following values being incorrect: 5 minute: res_ws, wind_d, sigma_wd 15 minute: mv_wind_d 30 minute: res_ws, wind_d, sigma_wd |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp5ebbrE19.00:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/21/1998 | 1435 | 12/23/1998 | 0200 |
Subject: | SGP/EBBR/E20 - Wind Direction Data Incorrect (revised) |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | The wind direction sensor was iced stopped by freezing rain, resulting in the following values being incorrect: 5 minute: res_ws, wind_d, sigma_wd 15 minute: mv_wind_d 30 minute: res_ws, wind_d, sigma_wd |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/08/1999 | 0005 | 01/08/1999 | 0900 |
Subject: | SGP/EBBR/E15 - Wind Direction Data Incorrect (revised) |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | The wind direction sensor was iced stopped by freezing rain, resulting in the following values being incorrect: 5 minute: res_ws, wind_d, sigma_wd 15 minute: mv_wind_d 30 minute: res_ws, wind_d, sigma_wd |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/08/1999 | 0105 | 01/09/1999 | 0030 |
Subject: | SGP/EBBR/E25 - Wind Direction Data Incorrect |
DataStreams: | sgp15ebbrE25.b1, sgp30ebbrE25.b1, sgp5ebbrE25.b1 |
Description: | The wind direction sensor was iced stopped by freezing rain, resulting in the following values being incorrect: 5 minute: res_ws, wind_d, sigma_wd 15 minute: mv_wind_d 30 minute: res_ws, wind_d, sigma_wd |
Measurements: | sgp30ebbrE25.b1:
sgp5ebbrE25.b1:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/08/1999 | 1005 | 01/08/1999 | 1730 |
Subject: | SGP/EBBR/E20 - Wind Direction Data Incorrect |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | The wind direction sensor was iced stopped by freezing rain, resulting in the following values being incorrect: 5 minute: res_ws, wind_d, sigma_wd 15 minute: mv_wind_d 30 minute: res_ws, wind_d, sigma_wd |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/07/1999 | 2355 | 01/08/1999 | 1100 |
Subject: | SGP/EBBR/E22 - Wind Direction Data Incorrect |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1, sgp5ebbrE22.00 |
Description: | The wind direction sensor was iced stopped by freezing rain, resulting in the following values being incorrect: 5 minute: res_ws, wind_d, sigma_wd 15 minute: mv_wind_d 30 minute: res_ws, wind_d, sigma_wd |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.00:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/08/1999 | 0835 | 01/08/1999 | 1600 |
Subject: | SGP/EBBR/E26 - Wind Direction Data Incorrect |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | The wind direction sensor was iced stopped by freezing rain, resulting in the following values being incorrect: 5 minute: res_ws, wind_d, sigma_wd 15 minute: mv_wind_d 30 minute: res_ws, wind_d, sigma_wd |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/21/1998 | 0015 | 12/24/1998 | 1730 |
01/01/1999 | 1515 | 01/01/1999 | 2230 |
01/08/1999 | 0205 | 01/08/1999 | 1600 |
Subject: | SGP/EBBR/E13 - Wind Speed Data Incorrect |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | The wind speed sensor was iced stopped by freezing rain, resulting in the following values being incorrect: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws, |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/01/1999 | 1515 | 01/01/1999 | 2230 |
Subject: | SGP/EBBR - EBBR E13 Wind Speed Data Incorrect |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1, sgp5ebbrE13.00 |
Description: | The wind speed sensor was iced stopped by freezing rain, resulting in the following values being incorrect: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws, |
Measurements: | sgp5ebbrE13.00:
sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/08/1999 | 0205 | 01/08/1999 | 1600 |
Subject: | SGP/EBBR - EBBR E13 Wind Speed Data Incorrect |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1, sgp5ebbrE13.00 |
Description: | The wind speed sensor was iced stopped by freezing rain, resulting in the following values being incorrect: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws, |
Measurements: | sgp5ebbrE13.00:
sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/07/1999 | 2240 | 01/10/1999 | 1530 |
Subject: | SGP/EBBR/E19 - Wind Speed Data Incorrect |
DataStreams: | sgp30ebbrE19.b1, sgp5ebbrE19.b1, sgp5ebbrE19.00, sgp15ebbrE19.b1 |
Description: | The wind speed sensor was iced stopped by freezing rain, resulting in the following values being incorrect: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws, |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp5ebbrE19.00:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/21/1998 | 0505 | 12/24/1998 | 1830 |
Subject: | SGP/EBBR/E19 - Wind Speed Data Incorrect |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1, sgp5ebbrE19.00 |
Description: | The wind speed sensor was iced stopped by freezing rain, resulting in the following values being incorrect: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws, |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp5ebbrE19.00:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/21/1998 | 1005 | 12/22/1998 | 2000 |
Subject: | SGP/EBBR - EBBR E18 Wind Speed Data Incorrect |
DataStreams: | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1, sgp5ebbrE18.00 |
Description: | The wind speed sensor was iced stopped by freezing rain, resulting in the following values being incorrect: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws, |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
sgp5ebbrE18.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/02/1999 | 0805 | 01/02/1999 | 1800 |
Subject: | SGP/EBBR/E18 - Wind Speed Data Incorrect |
DataStreams: | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1, sgp5ebbrE18.00 |
Description: | The wind speed sensor was iced stopped by freezing rain, resulting in the following values being incorrect: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws, |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
sgp5ebbrE18.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/08/1999 | 1015 | 01/10/1999 | 1300 |
Subject: | SGP/EBBR/E18 - Wind Speed Data Incorrect |
DataStreams: | sgp15ebbrE18.b1, sgp30ebbrE18.b1, sgp5ebbrE18.b1, sgp5ebbrE18.00 |
Description: | The wind speed sensor was iced stopped by freezing rain, resulting in the following values being incorrect: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws, |
Measurements: | sgp15ebbrE18.b1:
sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
sgp5ebbrE18.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/08/1999 | 0545 | 01/09/1999 | 2200 |
Subject: | SGP/EBBR/E20 - Wind Speed Data Incorrect |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | The wind speed sensor was iced stopped by freezing rain, resulting in the following values being incorrect: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws, |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/21/1998 | 0735 | 12/21/1998 | 2000 |
Subject: | SGP/EBBR/E20 - Wind Speed Data Incorrect |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | The wind speed sensor was iced stopped by freezing rain, resulting in the following values being incorrect: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws, |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/21/1998 | 0705 | 12/24/1998 | 1630 |
01/08/1999 | 0315 | 01/09/1999 | 0600 |
Subject: | SGP/EBBR/E26 - Wind Speed Data Incorrect |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | The wind speed sensor was iced stopped by freezing rain, resulting in the following values being incorrect: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws, |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/08/1999 | 0315 | 01/09/1999 | 0600 |
Subject: | SGP/EBBR - EBBR E26 Wind Speed Data Incorrect |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp5ebbrE26.00 |
Description: | The wind speed sensor was iced stopped by freezing rain, resulting in the following values being incorrect: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws, |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
sgp5ebbrE26.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/20/1998 | 2135 | 12/25/1998 | 1730 |
01/01/1999 | 1620 | 01/02/1999 | 2000 |
01/07/1999 | 2215 | 01/09/1999 | 0130 |
Subject: | SGP/EBBR/E15 - Wind Speed Data Incorrect |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1, sgp5ebbrE15.00 |
Description: | The wind speed sensor was iced stopped by freezing rain, resulting in the following values being incorrect: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws, |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp5ebbrE15.00:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/01/1999 | 1620 | 01/02/1999 | 2000 |
Subject: | SGP/EBBR - EBBR E15 Wind Speed Data Incorrect |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1, sgp5ebbrE15.00 |
Description: | The wind speed sensor was iced stopped by freezing rain, resulting in the following values being incorrect: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws, |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp5ebbrE15.00:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/07/1999 | 2215 | 01/09/1999 | 0130 |
Subject: | SGP/EBBR - EBBR E15 Wind Speed Data Incorrect |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1, sgp5ebbrE15.00 |
Description: | The wind speed sensor was iced stopped by freezing rain, resulting in the following values being incorrect: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws, |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp5ebbrE15.00:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/09/1999 | 0200 | 01/14/1999 | 1830 |
Subject: | SGP/EBBR/E26 - Battery Voltage Low, Data Incorrect |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp5ebbrE26.00 |
Description: | Battery voltages less than 10.4 volts occurred some of the time. All data should be considered incorrect when the voltage is less than 10.4 volts. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
sgp5ebbrE26.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/12/1998 | 1230 | 11/12/1998 | 2000 |
Subject: | SGP/EBBR/E9 - EBBR E9 Net Radiometer Dome Broken, Flux Data Incorrect |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1, sgp5ebbrE9.00 |
Description: | Incorrect Net Radiation values resulted in incorrect sensible and latent heat fluxes. Affected values are: 5 minute: q 15 minute: mv_q 30 minute: q, e, h |
Measurements: | sgp30ebbrE9.b1:
sgp5ebbrE9.00:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/20/1998 | 2335 | 12/22/1998 | 1630 |
01/07/1999 | 1805 | 01/09/1999 | 0330 |
01/01/1999 | 1700 | 01/03/1999 | 1130 |
Subject: | SGP/EBBR/E12 - Wind Speed Data Incorrect |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | The wind speed sensor was iced stopped by freezing rain, resulting in the following values being incorrect: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws, |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/02/1999 | 0735 | 01/02/1999 | 1700 |
Subject: | SGP/EBBR/E25 - Wind Speed Data Incorrect |
DataStreams: | sgp15ebbrE25.b1, sgp30ebbrE25.b1, sgp5ebbrE25.b1 |
Description: | The wind speed sensor was iced stopped by freezing rain, resulting in the following values being incorrect: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws, |
Measurements: | sgp30ebbrE25.b1:
sgp5ebbrE25.b1:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/08/1999 | 1145 | 01/09/1999 | 0300 |
Subject: | SGP/EBBR/E25 - Wind Speed Data Incorrect |
DataStreams: | sgp15ebbrE25.b1, sgp30ebbrE25.b1, sgp5ebbrE25.b1 |
Description: | The wind speed sensor was iced stopped by freezing rain, resulting in the following values being incorrect: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws, |
Measurements: | sgp30ebbrE25.b1:
sgp5ebbrE25.b1:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/07/1999 | 1805 | 01/09/1999 | 0330 |
Subject: | SGP/EBBR - EBBR E12 Wind Speed Data Incorrect |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1, sgp5ebbrE12.00 |
Description: | The wind speed sensor was iced stopped by freezing rain, resulting in the following values being incorrect: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws, |
Measurements: | sgp30ebbrE12.b1:
sgp15ebbrE12.b1:
sgp5ebbrE12.00:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/01/1999 | 1505 | 01/03/1999 | 2100 |
Subject: | SGP/EBBR/E7 - EBBR E7 Wind Speed Data Incorrect |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | The wind speed sensor was iced stopped by freezing rain, resulting in the following values being incorrect: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws, |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/08/1999 | 1035 | 01/08/1999 | 2300 |
Subject: | SGP/EBBR/E7 - Wind Speed Data Incorrect |
DataStreams: | sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.b1 |
Description: | The wind speed sensor was iced stopped by freezing rain, resulting in the following values being incorrect: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws, |
Measurements: | sgp15ebbrE7.b1:
sgp30ebbrE7.b1:
sgp5ebbrE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/01/1999 | 1445 | 01/05/1999 | 0400 |
Subject: | SGP/EBBR/E2 - EBBR E2 Wind Speed Data Incorrect |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1, sgp5ebbrE2.00 |
Description: | The wind speed sensor was iced stopped by freezing rain, resulting in the following values being incorrect: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws, |
Measurements: | sgp15ebbrE2.b1:
sgp30ebbrE2.b1:
sgp5ebbrE2.00:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/01/1999 | 1515 | 01/03/1999 | 1830 |
Subject: | SGP/EBBR/E9 - Wind Speed Data Incorrect |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1, sgp5ebbrE9.00 |
Description: | The wind speed sensor was iced stopped by freezing rain, resulting in the following values being incorrect: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws, |
Measurements: | sgp30ebbrE9.b1:
sgp5ebbrE9.00:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/08/1999 | 0605 | 01/08/1999 | 1900 |
Subject: | SGP/EBBR/E9 - Wind Speed Data Incorrect |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1, sgp5ebbrE9.00 |
Description: | The wind speed sensor was iced stopped by freezing rain, resulting in the following values being incorrect: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws, |
Measurements: | sgp30ebbrE9.b1:
sgp5ebbrE9.00:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/08/1999 | 0835 | 01/08/1999 | 1500 |
Subject: | SGP/EBBR/E8 - Wind Speed Data Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The wind speed sensor was iced stopped by freezing rain, resulting in the following values being incorrect: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws, |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/01/1999 | 1700 | 01/03/1999 | 1130 |
Subject: | SGP/EBBR - EBBR E12 Wind Speed Data Incorrect |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1, sgp5ebbrE12.00 |
Description: | The wind speed sensor was iced stopped by freezing rain, resulting in the following values being incorrect: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws, |
Measurements: | sgp30ebbrE12.b1:
sgp15ebbrE12.b1:
sgp5ebbrE12.00:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/28/1998 | 1600 | 12/28/1998 | 1630 |
Subject: | SGP/EBBR/E12 - Data Incorrect |
DataStreams: | sgp30ebbrE12.b1, sgp5ebbrE12.b1, sgp5ebbrE12.00, sgp15ebbrE12.b1 |
Description: | Most of the data values for just this time (1630 GMT) for the 5, 15, and 30 minute periods are incorrect; this apparently was the result of PM activities. |
Measurements: | sgp30ebbrE12.b1:
sgp15ebbrE12.b1:
sgp5ebbrE12.00:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/28/1998 | 1345 | 12/28/1998 | 1730 |
Subject: | SGP/EBBR/E12 - Temperature, RH, and Vapor Pressure Data Incorrect |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.b1 |
Description: | The right aspirator housing fell off of the AEM around 1345 GMT, by looking at the data. The housing was put back into place later in the day during a PM visit. The following values were affected: 5 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot 15 minute: rr_thum_r, mv_hum_r, tair_r 30 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, bowen, e, h |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/24/1998 | 1315 | 11/24/1998 | 1530 |
Subject: | SGP/EBBR/E12 - All Temperature Measurements Incorrect |
DataStreams: | sgp30ebbrE12.b1, sgp5ebbrE12.b1, sgp15ebbrE12.b1 |
Description: | Failure of the multiplexer connection for the reference temperature thermocouple in the control box or a problem with the Campbell CR10 bridge circuitry caused all air and soil temperature measurements to be incorrect. Therefore, the computed 30 minute bowen, e, and h were also incorrect. |
Measurements: | sgp15ebbrE12.b1:
sgp30ebbrE12.b1:
sgp5ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/30/1998 | 0300 | 12/30/1998 | 1300 |
Subject: | SGP/EBBR/E26 - Battery Voltage Low |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp5ebbrE26.00 |
Description: | Battery voltage dropped below 10.4 volts. All data should be considered to be incorrect. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
sgp5ebbrE26.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/25/1998 | 1735 | 12/29/1998 | 1630 |
Subject: | SGP/EBBR/E20 - Data Incorrect |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1, sgp5ebbrE20.00 |
Description: | Version 6 of the CR10 program was installed during this period instead of version 7 (the initial version 7 program did not work). The calibrations and coefficients in version 6 were incorrect; therefore all data during this period is of questionable quality and e and h are clearly incorrect. The correct version 7 program was installed on 29 Dec 1998. |
Measurements: | sgp5ebbrE20.b1:
sgp5ebbrE20.00:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/29/1999 | 0945 | 01/29/1999 | 2130 |
Subject: | SGP/EBBR/E2 - Wind Speed Iced, Data Incorrect |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1, sgp5ebbrE2.00 |
Description: | The wind speed sensor was stopped by icing. Wind speed data are incorrect. The values are: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp15ebbrE2.b1:
sgp30ebbrE2.b1:
sgp5ebbrE2.00:
sgp5ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/28/1999 | 1605 | 01/29/1999 | 0830 |
Subject: | SGP/EBBR - EBBR E8 Wind Speed Iced, Data Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp5ebbrE8.00 |
Description: | The wind speed sensor was stopped by icing. Wind speed data is incorrect. The values are: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
sgp5ebbrE8.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/28/1999 | 1530 | 02/02/1999 | 1800 |
Subject: | SGP/EBBR/E8 - Wind Direction Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The wind direction sensor output read 135 degrees for much of the period, an indication of sensor failure. Energy Balance fluxes are not affected by the problem. The values below were incorrect when wind direction was 135: 5 minute: res_ws, wind_d, sigma_wd 15 minute: mv_wind_d 30 minute: res_ws, wind_d, sigma_wd |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/30/1999 | 0220 | 01/31/1999 | 2030 |
Subject: | SGP/EBBR/E8 - Wind Speed Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp5ebbrE8.00 |
Description: | The wind speed sensor was iced stopped for the period. Energy Balance fluxes are not affected by the problem. The values below were incorrect for the period: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
sgp5ebbrE8.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/13/1999 | 2150 | 02/10/1999 | 1640 |
Subject: | SGP/EBBR/E20 - Pressure Low |
DataStreams: | sgp5ebbrE20.b1, sgp15ebbrE20.b1, sgp30ebbrE20.b1 |
Description: | The pressure sensor values dropped several kPa on 13 January 1999 because of a sensor malfunction and remained that way until the sensor was replaced on 10 February 1999. Comparison of the pressure before and after the problem period shows that Bowen ratio was underestimated by approximately 6% during the period, resulting in a 6% overestimation of latent heat flux and a 6% underestimation of sensible heat flux. Adjustment of the flux data during the period by these amounts should be made by the data user. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/02/1999 | 0400 | 02/23/1999 | 1500 |
Subject: | SGP/EBBR/E15 - Temperatures Incorrect |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | Poor contact of the reference temperature probe leads to the J-panel or multiplexer caused frequent offscale and incorrect values of the reference temperature. All other temperature measurements (air, soil, and humidity probe) depend on the reference temperature for a correction. Therefore all temperature measurements are frequently offscale and incorrect during the specified period. The data clearly shows the offscale times. |
Measurements: | sgp5ebbrE15.b1:
sgp30ebbrE15.b1:
sgp15ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/10/1999 | 2045 | 03/10/1999 | 2100 |
Subject: | SGP/EBBR/E22 - Right Aspirator Fan Replaced |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1 |
Description: | Replacement of the right aspirator fan on the EBBR AEM caused temperature, relative humidity, and vapor pressure data to be incorrect for the 2045 and 2100 GMT 15 minute data and Bowen ratio, sensible heat flux, and latent heat flux to be incorrect for the 2100 GMT data. |
Measurements: | sgp30ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/01/1999 | 1415 | 04/02/1999 | 1645 |
Subject: | SGP/EBBR/E26 - Calibration Changeout |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp5ebbrE26.00 |
Description: | The data from this period was either missing or incorrect during the calibration changeout of the EBBR unit. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
sgp5ebbrE26.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/25/1999 | 1545 | 03/25/1999 | 2355 |
Subject: | SGP/EBBR/E25 - CR10 Program Revision (Vs. 7) |
DataStreams: | sgp15ebbrE25.b1, sgp30ebbrE25.b1, sgp5ebbrE25.b1, sgp5ebbrE25.00 |
Description: | The data from this period was either missing during and after the installation of the revised version 7 of the CR10 program. The new program version has made a colossal improvement in the flux data. Please refer back to DQR D980117.3 and PIF P971218.1, which describe the incorrect way in which calculations were being performed by the previous CR10 program version. |
Measurements: | sgp30ebbrE25.b1:
sgp5ebbrE25.b1:
sgp5ebbrE25.00:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/08/1999 | 0500 | 03/11/1999 | 1540 |
Subject: | SGP/EBBR/E19 - Net Radiation Incorrect |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | Water in the net radiometer dome made the net radiation data incorrect. Comparison with surrounding EBBR sites should allow for an approximate correction of the net radiation data. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/13/1999 | 2330 | 03/14/1999 | 0500 |
Subject: | SGP/EBBR/E13 - Wind Speed Frozen |
DataStreams: | sgp15ebbrE13.b1, sgp30ebbrE13.b1, sgp5ebbrE13.b1 |
Description: | Wind speed and residual wind speed during this period are incorrect caused by an ice storm. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/12/1999 | 0955 | 03/14/1999 | 1800 |
Subject: | SGP/EBBR/E8 - Wind Speed Frozen |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | Wind speed and residual wind speed during this period are incorrect caused by an ice storm. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/25/1999 | 0040 | 04/07/1999 | 1745 |
Subject: | SGP/EBBR/E20 - Air Temperature Incorrect |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | Right air temperature (15 minute) and top and bottom air temperatures (5 and 30 minute) during this period are incorrect because the right air temperature sensor was not working. |
Measurements: | sgp5ebbrE20.b1:
sgp15ebbrE20.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/02/1999 | 1645 | 04/02/1999 | 1800 |
Subject: | SGP/EBBR/E26 - Incorrect Data |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgpebbrE26.00, sgp5ebbrE26.00 |
Description: | All data is incorrect for this time. This problem was related to changeout activities. |
Measurements: | sgpebbrE26.00:
sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
sgp5ebbrE26.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/08/1999 | 1430 | 04/16/1999 | 1730 |
Subject: | SGP/EBBR/E19 - Incorrect Tref, Tair Data, Bowen Ratio, latent and sensible heat fluxes |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | All Tref and Tair temperatures, plus bowen ratio, latent heat flux, and sensible heat flux values are incorrect for this period. This problem was caused by mistakenly removing the Tref sensor. |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/08/1997 | 2030 | 06/17/1999 | 1600 |
Subject: | SGP/EBBR/E18 - Pressure Incorrect |
DataStreams: | sgp30ebbrE18.b1, sgp5ebbrE18.b1, sgp5ebbrE18.00 |
Description: | The pressure sensor was replaced at E18 EBBR without the replacement being noted by Site Ops and therefore without an indication of serial number and thus correct calibration. The serial number of the sensor was observed during a PM (after months of confusion concerning the thing) and a several month search for the calibration sheet at ANL, the CF, and the manufacturer ensued. The sheet was finally found and I confirmed that an incorrect offset was being used (the same as used for the previous sensor). This is an easy error to make as there are two models of pressure sensors (with different calibration ranges) being used in the EBBR systems. On June 17, 1999 the correct offset was entered into the CR10 and storage module programming. The pressure for the period should be increased by 6.757 kPa. The result of the error is that Bowen ratio was underestimated by 6.8% on average, Latent Heat Flux was overestimated by 3.6% on average, and Sensible Heat Flux was underestimated by 3.6% on average. The errors in the fluxes are smaller than the estimated 10% error in the Bowen ratio flux measurement technique and therefore the data does not need to be reprocessed. Anyone performing area averages of pressure or looking at the variation in pressure over the CART Site should adjust the E18 EBBR pressure values for this period using the offset error indicated above. |
Measurements: | sgp30ebbrE18.b1:
sgp5ebbrE18.b1:
sgp5ebbrE18.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/25/1999 | 1630 | 04/22/1999 | 1500 |
Subject: | SGP/EBBR/E19 - Temperature of Left Humidity Probe Incorrect |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | A poorly seated lead caused the temperature reading from the left humidity probe to be incorrect. This also affects the calculated vapor pressures,the Bowen ratio, and the sensible and latent heat fluxes. The affected values are: 5 minute: thum_top, thum_bot, vp_top, vp_bot 15 minute: rr_thum_l 30 minute: thum_top, thum_bot, vp_top, vp_bot, bowen, e, h |
Measurements: | sgp15ebbrE19.b1:
sgp5ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/04/1999 | 1630 | 06/04/1999 | 1700 |
Subject: | SGP/EBBR/E12 - RH top Data Zero |
DataStreams: | sgp30ebbrE12.b1 |
Description: | The one data value hum_top was incorrect in the 30 minute data. 5 and 15 minute data were not affected, and the 30 minute flux data was not affected. The data value may have been lost in remote data retrieval, resulting in a zero value instead. |
Measurements: | sgp30ebbrE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/22/1999 | 0750 | 04/29/1999 | 1850 |
Subject: | SGP/EBBR/E9 - Wind Speed Incorrect |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1, sgp5ebbrE9.00 |
Description: | The wind speed sensor lost two of the three cups. Flux data is not affected. The data values affected are: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp30ebbrE9.b1:
sgp5ebbrE9.00:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/15/1999 | 1830 | 04/15/1999 | 1900 |
Subject: | SGP/EBBR/E26 - Battery Wire Problem Causes Incorrect Data |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp5ebbrE26.00 |
Description: | The battery voltage dropped suddenly to 8 volts at 1900 GMT, making all data incorrect for that time. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
sgp5ebbrE26.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/28/1999 | 2245 | 04/29/1999 | 0600 |
Subject: | SGP/EBBR/E26 - Battery Wire Problem Causes Questionable Data |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1, sgp5ebbrE26.00 |
Description: | The battery voltage dropped to less than 10.4 volts, causing all data to be suspect for the time period. |
Measurements: | sgp30ebbrE26.b1:
sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
sgp5ebbrE26.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/13/1999 | 1230 | 06/11/1999 | 1645 |
Subject: | SGP/EBBR/E9 - Right Aspirator Not Working |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Description: | The result of the right aspirator not working was a reduction in measured temperature gradient and thus a reduction in estimated sensible heat flux and an increase in latent heat flux. The fluxes could be in error by as much as 10%. Reported right air and humidity probe temperatures, top and bottom air and humidity probe temperatures, right relative humidity, and top and bottom vapor pressure are affected. |
Measurements: | sgp30ebbrE9.b1:
sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/16/1999 | 1830 | 06/16/1999 | 1900 |
Subject: | SGP/EBBR/E25 - Data Incorrect During PM |
DataStreams: | sgp15ebbrE25.b1, sgp30ebbrE25.b1, sgp5ebbrE25.b1, sgp5ebbrE25.00 |
Description: | All data was incorrect during some PM work. |
Measurements: | sgp30ebbrE25.b1:
sgp5ebbrE25.b1:
sgp5ebbrE25.00:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/16/1999 | 1905 | 06/16/1999 | 2010 |
Subject: | SGP/EBBR/E25 - Data Incorrect During PM |
DataStreams: | sgp15ebbrE25.b1, sgp30ebbrE25.b1, sgp5ebbrE25.b1 |
Description: | Air and humidity probe temperatures and the reference temperature were incorrect during this period because of the PM work taking place. |
Measurements: | sgp30ebbrE25.b1:
sgp5ebbrE25.b1:
sgp15ebbrE25.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/16/1999 | 2015 | 06/17/1999 | 0500 |
Subject: | SGP/EBBR/E25 - Data Incorrect During PM |
DataStreams: | sgp15ebbrE25.b1, sgp30ebbrE25.b1, sgp5ebbrE25.b1 |
Description: | Right and left air temperatures and top and bottom air temperatures were incorrect during this period because of the PM work taking place. Sensible and latent heat fluxes and Bowen ratio were therefore also incorrect. |
Measurements: | sgp30ebbrE25.b1:
sgp5ebbrE25.b1:
sgp15ebbrE25.b1:
|