Data Quality Reports for Session: 141299 User: bdgantt Completed: 03/20/2012


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D050412.1PYE/MET/M1 - Deployment periodpyemetM1.b1
D081022.10PYE/MET/M1 - Wind Speed data contain offset due to induced voltagepyemetM1.b1
D101020.6PYE/AOS/M1 - Data processing problempyeaosM1.a1, pyenoaaaosM1.b1, pyenoaaaosavgM1.b1


DQRID : D050412.1
Start DateStart TimeEnd DateEnd Time
02/11/2005000003/14/20052359
Subject:
PYE/MET/M1 - Deployment period
DataStreams:pyemetM1.b1
Description:
During the listed time period, the AMF was in the beginning phase of deployment.  Data 
during this time may not be representative and should be used with caution.
Measurements:pyemetM1.b1:
  • Humidity, relative, at 2-m height, 1-min avg(rh_mean)
  • Barometric Pressure(atmos_pressure)
  • alt
  • Wind direction, vector, lower, 1-min avg(wdir_vec_mean)
  • rh_std
  • pwd_pw_code_1hr
  • Mean Air Temperature or Hardware Error(temp_mean)
  • PWD 1 minute mean visibility(pwd_mean_vis_1min)
  • vapor_pressure_std
  • pwd_precip_rate_mean_1min
  • pwd_cumul_rain
  • time
  • pwd_err_code
  • lat
  • logger_temp
  • time_offset
  • temp_std
  • Wind speed, lower, 1-min avg(wspd_arith_mean)
  • base_time
  • PWD 10 minute mean visibility(pwd_mean_vis_10min)
  • lon
  • pwd_pw_code_inst
  • vapor_pressure_mean
  • Wind speed vector mean(wspd_vec_mean)
  • ORG precipitation rate mean(org_precip_rate_mean)
  • pwd_pw_code_15min
  • wdir_vec_std
  • logger_volt


Back To Table of Contents

DQRID : D081022.10
Start DateStart TimeEnd DateEnd Time
02/01/2005000009/01/20052359
Subject:
PYE/MET/M1 - Wind Speed data contain offset due to induced voltage
DataStreams:pyemetM1.b1
Description:
The marine model of the RM Young wind sensor needs twisted pair shielded cable.  The 
manufacturer did not originally supply the pigtail with twisted pair shielded cable.  Use of 
twisted pair shielded cable for the rest of the cable run to the logger was also not 
specified.  This caused an induced voltage in the wind speed line from the wind direction 
excitation.  The induced voltage varied from sensor to sensor and probably changed anytime 
the cable was replaced or parts of the wind sensor were repaired or replaced. The induced 
voltage has been seen to vary from 0.15 m/s to as much as 1.08 m/s.
Measurements:pyemetM1.b1:
  • Wind speed, lower, 1-min avg(wspd_arith_mean)
  • Wind speed vector mean(wspd_vec_mean)


Back To Table of Contents

DQRID : D101020.6
Start DateStart TimeEnd DateEnd Time
03/09/2005000009/15/20050000
Subject:
PYE/AOS/M1 - Data processing problem
DataStreams:pyeaosM1.a1, pyenoaaaosM1.b1, pyenoaaaosavgM1.b1
Description:
The AOS data was reprocessed for the AMF, SGP and NSA sites. There was an error in the 
processing of the corrections for the PSAP absorption signal such that a low pass filter was 
implemented for the scattering data used in these corrections.The difference between the 
corrected and previous data is < 0.05 Mm-1 (below the instrument detection limit) and 
should not have a significant effect on the previous data. Data will be reprocessed at a 
later time when more significant changes to the data processing are implemented.
Measurements:pyeaosM1.a1:
  • Absorption Coefficient Blue, corrected for flow and sample area, 3 wavelength
    PSAP, 10 um size cut(Ba_B_Dry_10um_PSAP3W_1)
  • Absorption Coefficient Blue, corrected for flow and sample area, 3 wavelength
    PSAP, 1 um size cut(Ba_B_Dry_1um_PSAP3W_1)
  • Absorption Coefficient Red, corrected for flow and sample area, 3 wavelength
    PSAP, 10 um size cut(Ba_R_Dry_10um_PSAP3W_1)
  • Absorption Coefficient Green, corrected for flow and sample area, 3 wavelength
    PSAP, 10 um size cut(Ba_G_Dry_10um_PSAP3W_1)
  • Aerosol light absorption coefficient, green channel, 1 um particle diameter(Ba_G_Dry_1um_PSAP1W_1)
  • Aerosol light absorption coefficient, green channel, 10 um particle diameter(Ba_G_Dry_10um_PSAP1W_1)
  • Absorption Coefficient Green, corrected for flow and sample area, 3 wavelength
    PSAP, 1 um size cut(Ba_G_Dry_1um_PSAP3W_1)
  • Absorption Coefficient Red, corrected for flow and sample area, 3 wavelength
    PSAP, 10 um size cut(Ba_R_Dry_1um_PSAP3W_1)

pyenoaaaosavgM1.b1:
  • Aerosol light absorption coefficient, green channel, 1 um particle diameter(Ba_G_Dry_1um_PSAP1W_1)
  • Aerosol light absorption coefficient, green channel, 10 um particle diameter(Ba_G_Dry_10um_PSAP1W_1)

pyenoaaaosM1.b1:
  • Absorption Coefficient Red, corrected for flow and sample area, 3 wavelength
    PSAP, 10 um size cut(Ba_R_Dry_1um_PSAP3W_1)
  • Aerosol light absorption coefficient, green channel, 10 um particle diameter(Ba_G_Dry_10um_PSAP1W_1)
  • Absorption Coefficient Blue, corrected for flow and sample area, 3 wavelength
    PSAP, 10 um size cut(Ba_B_Dry_10um_PSAP3W_1)
  • Absorption Coefficient Red, corrected for flow and sample area, 3 wavelength
    PSAP, 10 um size cut(Ba_R_Dry_10um_PSAP3W_1)
  • Absorption Coefficient Green, corrected for flow and sample area, 3 wavelength
    PSAP, 1 um size cut(Ba_G_Dry_1um_PSAP3W_1)
  • Absorption Coefficient Green, corrected for flow and sample area, 3 wavelength
    PSAP, 10 um size cut(Ba_G_Dry_10um_PSAP3W_1)
  • Aerosol light absorption coefficient, green channel, 1 um particle diameter(Ba_G_Dry_1um_PSAP1W_1)
  • Absorption Coefficient Blue, corrected for flow and sample area, 3 wavelength
    PSAP, 1 um size cut(Ba_B_Dry_1um_PSAP3W_1)


Back To Table of Contents