Data Quality Reports for Session: 140628 User: lijun1ljr Completed: 02/16/2012


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D110628.2SGP/SIRS/C1 - NIP failuresgpsirs20sC1.a0, sgpsirsC1.b1
D111202.1SGP/SIRS/C1 - NIP Failuresgpsirs20sC1.a0, sgpsirsC1.b1


DQRID : D110628.2
Start DateStart TimeEnd DateEnd Time
06/12/2011000006/24/20111740
Subject:
SGP/SIRS/C1 - NIP failure
DataStreams:sgpsirs20sC1.a0, sgpsirsC1.b1
Description:
The short_direct_normal (NIP instrument) cable failed during this time period.	Data are 
not recoverable.

Users can derive an estimate of direct normal values using the following formula:
short_direct_normal = (down_short_hemisp - down_short_diffuse_hemisp)/cos(zenith angle)
Measurements:sgpsirs20sC1.a0:
  • inst_direct_normal

sgpsirsC1.b1:
  • Radiation, shortwave, direct normal irradiance, 1-min avg(short_direct_normal)
  • short_direct_normal_std
  • Shortwave Direct Normal Irradiance, Pyrheliometer, Maxima(short_direct_normal_max)
  • Shortwave Direct Normal Irradiance, Pyrheliometer, Minima(short_direct_normal_min)


Back To Table of Contents

DQRID : D111202.1
Start DateStart TimeEnd DateEnd Time
08/16/2011214408/25/20111435
Subject:
SGP/SIRS/C1 - NIP Failure
DataStreams:sgpsirs20sC1.a0, sgpsirsC1.b1
Description:
The short_direct_normal (NIP instrument) cable failed (broken wire in NIP connector) 
during this time period. Data are not recoverable.


An estimate of direct normal values can be derived using the following formula:

short_direct_normal = (down_short_hemisp - down_short_diffuse_hemisp)/cos(zenith angle)
Measurements:sgpsirs20sC1.a0:
  • inst_direct_normal

sgpsirsC1.b1:
  • Radiation, shortwave, direct normal irradiance, 1-min avg(short_direct_normal)
  • short_direct_normal_std
  • Shortwave Direct Normal Irradiance, Pyrheliometer, Maxima(short_direct_normal_max)
  • Shortwave Direct Normal Irradiance, Pyrheliometer, Minima(short_direct_normal_min)


Back To Table of Contents



END OF DATA