Data Quality Reports for Session: 140340 User: ak10j Completed: 02/06/2012


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D021004.20NSA/GNDRAD/C1 - PIR thermopile instantaneous values intermittently read zeronsagndrad20sC1.a1
D080728.1NSA/GNDRAD/C1 - IRT incorrectnsagndrad60sC1.a1
D080729.2NSA/GNDRAD/C1 - IRT scale factornsagndrad60sC1.a1
D990924.1NSA/GNDRAD/C1 - Reprocess: IRT insufficiently heatednsagndrad60sC1.a1


DQRID : D021004.20
Start DateStart TimeEnd DateEnd Time
04/06/1998215703/31/20012359
Subject:
NSA/GNDRAD/C1 - PIR thermopile instantaneous values intermittently read zero
DataStreams:nsagndrad20sC1.a1
Description:
GNDRAD intermittently reported false zero values for
the third instantaneous value (50 sec after the
minute) of the PIR thermopile voltage.
Measurements:nsagndrad20sC1.a1:
  • pir1_voltage


Back To Table of Contents

DQRID : D080728.1
Start DateStart TimeEnd DateEnd Time
04/06/1998000005/13/19982152
Subject:
NSA/GNDRAD/C1 - IRT incorrect
DataStreams:nsagndrad60sC1.a1
Description:
The IRT data were very noisy and unrepresentative.
Measurements:nsagndrad60sC1.a1:
  • Surface Infra-Red Temperature Maxima(sfc_ir_temp_max)
  • sfc_ir_temp_std
  • Effective Surface (Skin) Temperature(sfc_ir_temp)
  • Surface Infra-Red Temperature Minima(sfc_ir_temp_min)


Back To Table of Contents

DQRID : D080729.2
Start DateStart TimeEnd DateEnd Time
05/13/1998215206/04/19982208
Subject:
NSA/GNDRAD/C1 - IRT scale factor
DataStreams:nsagndrad60sC1.a1
Description:
The IRT scale factors in the data logger were incorrectly set to m = 100 degC/V, b = -60 
degC. They were corrected to 25 degC/V and -50 degC.
Measurements:nsagndrad60sC1.a1:
  • Surface Infra-Red Temperature Maxima(sfc_ir_temp_max)
  • sfc_ir_temp_std
  • Effective Surface (Skin) Temperature(sfc_ir_temp)
  • Surface Infra-Red Temperature Minima(sfc_ir_temp_min)


Back To Table of Contents

DQRID : D990924.1
Start DateStart TimeEnd DateEnd Time
04/06/1998215612/22/20002020
Subject:
NSA/GNDRAD/C1 - Reprocess: IRT insufficiently heated
DataStreams:nsagndrad60sC1.a1
Description:
IRT had insufficient heating to maintain an internal reference temperature above 0 degrees 
C.  Measurements of surface temperature were over-estimated when instrument was below 
freezing.

Data will be reprocessed, but users can correct data using the following correction 
factor:

If T_reference < -5?C, then

  IRT_corrected = (IRT_original - 32.993K)/0.87238

where T_reference can be estimated with the ambient temperature (e.g. 
inst_up_long_hemisp_dome_temp or inst_up_long_dome_temp)
Measurements:nsagndrad60sC1.a1:
  • irt1_max
  • irt1_min
  • irt1_sd
  • irt1_mean


Back To Table of Contents