DQRID : D000801.3
Start DateStart TimeEnd DateEnd Time
06/02/2000184206/16/20002234
Subject:
SGP/AOS/C1 - Nephelometer failure
DataStreams:sgpaosC1.a0, sgpaosauxC1.a0
Description:
An inexperienced AOS technician inadvertently closed off the flow to the dual-nephelometer 
humidograph system during a Friday Preventative Maintenance.  This resulted in liquid 
water getting into the humidified nephelometer.  Flow was restored to the system on Monday, 
but the nephelometers were producing strange data.  A number of tests were performed 
over the next few days to remotely troubleshoot the situation, to no avail.  Eventually, the 
decision was made to pull both nephelometers and return them to Boulder for repair.  
This DQR refers only to the reference nephelometer, which was out of service for 2 periods 
due to 2 separate problems.  This DQR is for the first period, from 000602 (1842 GMT) to 
000616 (2234 GMT).  Light absorption are invalid whenever the reference nephelometer is 
not operational, because those data are required for correction of the absorption data.
Measurements:sgpaosC1.a0:
  • Total scattering coefficient at 1 um from 700 nm TSI Low RH Nephelometer(RedTScatCoef_1um_LRH)
  • Backscattering coefficient at 10 um, from 450 nm TSI Low RH Nephelometer(BluBScatCoef_10um_LRH)
  • Backscattering coefficient at 1 um, from 450 nm TSI Low RH Nephelometer(BluBScatCoef_1um_LRH)
  • Backscattering coefficient at 1 um from 550 nm TSI Low RH Nephelometer(GrnBScatCoef_1um_LRH)
  • Absorption coefficient at 10 um(Bap_I_10um)
  • Total scattering coefficient at 1 um from 550 nm TSI Low RH Nephelometer(GrnTScatCoef_1um_LRH)
  • Total scattering coefficient at 10 um from 700 nm TSI Low RH Nephelometer(RedTScatCoef_10um_LRH)
  • Total scattering coefficient at 10 um from 550 nm TSI Low RH Nephelometer(GrnTScatCoef_10um_LRH)
  • Backscattering coefficient at 1 um from 700 nm TSI Low RH Nephelometer(RedBScatCoef_1um_LRH)
  • Total scattering coefficient at 10 um from 450 nm TSI Low RH Nephelometer(BluTScatCoef_10um_LRH)
  • Absorption coefficient at 1 um(Bap_I_1um)
  • Backscattering coefficient at 10 um from 550 nm TSI Low RH Nephelometer(GrnBScatCoef_10um_LRH)
  • Total scattering coefficient at 1 um from 450 nm TSI Low RH Nephelometer(BluTScatCoef_1um_LRH)
  • Backscattering coefficient at 10 um from 700 nm TSI Low RH Nephelometer(RedBScatCoef_10um_LRH)

sgpaosauxC1.a0:
  • BluBScatZero_LRH
  • BapRef
  • GrnTScatZero_LRH
  • GrnRScatZero_LRH
  • GrnTScatSpn_LRH
  • RedTScatSpn_LRH
  • RedTScatZero_LRH
  • BluRScatZero_LRH
  • RedRScatZero_LRH
  • BapSignal
  • RedBScatSpn_LRH
  • RedBScatZero_LRH
  • GrnBScatSpn_LRH
  • BluBScatSpn_LRH
  • BluTScatZero_LRH
  • GrnBScatZero_LRH
  • BluTScatSpn_LRH


Back To Table of Contents

DQRID : D000801.4
Start DateStart TimeEnd DateEnd Time
06/20/2000134007/03/20001833
Subject:
SGP/AOS/C1 - Nephelometer failure
DataStreams:sgpaosC1.a0, sgpaosauxC1.a0
Description:
An inexperienced AOS technician inadvertently closed off the flow to the dual-nephelometer 
humidograph system during a Friday Preventative Maintenance.  This resulted in liquid 
water getting into the humidified nephelometer.  Flow was restored to the system on Monday, 
but the nephelometers were producing strange data.  A number of tests were performed 
over the next few days to remotely troubleshoot the situation, to no avail.  Eventually, the 
decision was made to pull both nephelometers and return them to Boulder for repair.  
This DQR refers only to the reference nephelometer, which was out of service for 2 periods 
due to 2 separate problems.  This DQR is for the second period, from 000620 (1340 GMT) to 
000703 (1833 GMT).  Light absorption measurements are invalid whenever the reference 
nephelometer is not operational, because those data are required for correction of the 
absorption data.
Measurements:sgpaosC1.a0:
  • Total scattering coefficient at 1 um from 700 nm TSI Low RH Nephelometer(RedTScatCoef_1um_LRH)
  • Backscattering coefficient at 10 um, from 450 nm TSI Low RH Nephelometer(BluBScatCoef_10um_LRH)
  • Backscattering coefficient at 1 um, from 450 nm TSI Low RH Nephelometer(BluBScatCoef_1um_LRH)
  • Backscattering coefficient at 1 um from 550 nm TSI Low RH Nephelometer(GrnBScatCoef_1um_LRH)
  • Absorption coefficient at 10 um(Bap_I_10um)
  • Total scattering coefficient at 1 um from 550 nm TSI Low RH Nephelometer(GrnTScatCoef_1um_LRH)
  • Total scattering coefficient at 10 um from 700 nm TSI Low RH Nephelometer(RedTScatCoef_10um_LRH)
  • Total scattering coefficient at 10 um from 550 nm TSI Low RH Nephelometer(GrnTScatCoef_10um_LRH)
  • Backscattering coefficient at 1 um from 700 nm TSI Low RH Nephelometer(RedBScatCoef_1um_LRH)
  • Total scattering coefficient at 10 um from 450 nm TSI Low RH Nephelometer(BluTScatCoef_10um_LRH)
  • Absorption coefficient at 1 um(Bap_I_1um)
  • Backscattering coefficient at 10 um from 550 nm TSI Low RH Nephelometer(GrnBScatCoef_10um_LRH)
  • Total scattering coefficient at 1 um from 450 nm TSI Low RH Nephelometer(BluTScatCoef_1um_LRH)
  • Backscattering coefficient at 10 um from 700 nm TSI Low RH Nephelometer(RedBScatCoef_10um_LRH)

sgpaosauxC1.a0:
  • BluBScatZero_LRH
  • BapRef
  • GrnTScatZero_LRH
  • GrnRScatZero_LRH
  • GrnTScatSpn_LRH
  • RedTScatSpn_LRH
  • RedTScatZero_LRH
  • BluRScatZero_LRH
  • RedRScatZero_LRH
  • BapSignal
  • RedBScatSpn_LRH
  • RedBScatZero_LRH
  • GrnBScatSpn_LRH
  • BluBScatSpn_LRH
  • BluTScatZero_LRH
  • GrnBScatZero_LRH
  • BluTScatSpn_LRH


Back To Table of Contents

DQRID : D000801.5
Start DateStart TimeEnd DateEnd Time
06/02/2000184207/03/20001833
Subject:
SGP/AOS/C1 - Nephelometer failure
DataStreams:sgpaosC1.a0, sgpaosauxC1.a0
Description:
An inexperienced AOS technician inadvertently closed off the flow to the dual-nephelometer 
humidograph system during a Friday Preventative Maintenance.  This resulted in liquid 
water getting into the humidified nephelometer.  Flow was restored to the system on Monday, 
but the nephelometers were producing strange data.  A number of tests were performed 
over the next few days to remotely troubleshoot the situation, to no avail.  Eventually, the 
decision was made to pull both nephelometers and return them to Boulder for repair.  
This DQR refers only to the humidified nephelometer, which was out of service for the entire 
period from 000602 through 000703.
Measurements:sgpaosC1.a0:
  • Backscattering coefficient at 10 um from 550 nm TSI High RH Nephelometer(GrnBScatCoef_10um_HRH)
  • Total scattering coefficient at 1 um from 550 nm TSI High RH Nephelometer(GrnTScatCoef_1um_HRH)
  • Absorption coefficient at 10 um(Bap_I_10um)
  • Total scattering coefficient at 1 um from 450 nm TSI High RH Nephelometer(BluTScatCoef_1um_HRH)
  • Backscattering coefficient at 1 um from 700 nm TSI High RH Nephelometer(RedBScatCoef_1um_HRH)
  • Backscattering coefficient at 1 um from 550 nm TSI High RH Nephelometer(GrnBScatCoef_1um_HRH)
  • Total scattering coefficient at 10 um from 550 nm TSI High RH Nephelometer(GrnTScatCoef_10um_HRH)
  • Total scattering coefficient at 10 um from 450 nm TSI High RH Nephelometer(BluTScatCoef_10um_HRH)
  • Total scattering coefficient at 1 um from 700 nm TSI High RH Nephelometer(RedTScatCoef_1um_HRH)
  • Absorption coefficient at 1 um(Bap_I_1um)
  • Backscattering coefficient at 10 um from 700 nm TSI High RH Nephelometer(RedBScatCoef_10um_HRH)
  • Backscattering coefficient at 10 um, from 450 nm TSI High RH Nephelometer(BluBScatCoef_10um_HRH)
  • Total scattering coefficient at 10 um from 700 nm TSI High RH Nephelometer(RedTScatCoef_10um_HRH)
  • Backscattering coefficient at 1 um, from 450 nm TSI High RH Nephelometer(BluBScatCoef_1um_HRH)

sgpaosauxC1.a0:
  • RedTScatZero_HRH
  • GrnRScatZero_HRH
  • BluTScatZero_HRH
  • BluRScatZero_HRH
  • GrnBScatSpn_HRH
  • BapRef
  • RedRScatZero_HRH
  • BluBScatZero_HRH
  • GrnTScatSpn_HRH
  • RedBScatZero_HRH
  • BluBScatSpn_HRH
  • BapSignal
  • GrnTScatZero_HRH
  • BluTScatSpn_HRH
  • RedTScatSpn_HRH
  • GrnBScatZero_HRH
  • RedBScatSpn_HRH


Back To Table of Contents



END OF DATA