Data Quality Reports for Session: 139611 User: c_long Completed: 01/03/2012


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D110831.1TWP/SKYRAD/C2 - Tracker failuretwpskyrad20sC2.a0, twpskyrad60sC2.b1
D111026.1TWP/SKYRAD/C2 - IRT power failuretwpskyrad20sC2.a0, twpskyrad60sC2.b1


DQRID : D110831.1
Start DateStart TimeEnd DateEnd Time
06/16/2011233508/15/20110130
Subject:
TWP/SKYRAD/C2 - Tracker failure
DataStreams:twpskyrad20sC2.a0, twpskyrad60sC2.b1
Description:
Tracker failed on 6/16/2011. Ops replaced with a spare tracker. The affected instruments 
are the 8-48 (down_short_diffuse) and NIP (direct_normal) shortwave instruments.

The down_long_hemisp1 and 2 (PIR1 and PIR2) performance may have been affected by exposure 
to direct sun.
Measurements:twpskyrad60sC2.b1:
  • inst_down_long_hemisp2_case_temp
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer2(down_long_hemisp2)
  • Radiation, shortwave, downwelling diffuse hemispheric irradiance, 1-min avg(down_short_diffuse_hemisp)
  • down_long_hemisp1_min
  • Radiation, shortwave, direct normal irradiance, 1-min avg(short_direct_normal)
  • inst_down_long_hemisp2_tp
  • down_long_hemisp1_max
  • inst_down_long_hemisp1_tp
  • down_long_hemisp2_std
  • down_short_diffuse_hemisp_std
  • Downwelling Shortwave Diffuse Hemisp. Irrad., Ventilated Pyranometer, Minima(down_short_diffuse_hemisp_min)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer1(down_long_hemisp1)
  • Shortwave Direct Normal Irradiance, Pyrheliometer, Maxima(short_direct_normal_max)
  • Shortwave Direct Normal Irradiance, Pyrheliometer, Minima(short_direct_normal_min)
  • down_long_hemisp1_std
  • inst_down_long_hemisp1_dome_temp
  • short_direct_normal_std
  • down_long_hemisp2_max
  • down_long_hemisp2_min
  • inst_down_long_hemisp2_dome_temp
  • Downwelling Shortwave Diffuse Hemisp. Irrad., Ventilated Pyranometer, Maxima(down_short_diffuse_hemisp_max)
  • inst_down_long_hemisp1_case_temp

twpskyrad20sC2.a0:
  • inst_diffuse
  • inst_down_long_hemisp2_tp
  • inst_direct_normal
  • inst_down_long_hemisp1_tp
  • inst_down_long_hemisp1_case_resist
  • inst_down_long_hemisp2_case_resist
  • inst_down_long_hemisp2_dome_resist
  • inst_down_long_hemisp1_dome_resist


Back To Table of Contents

DQRID : D111026.1
Start DateStart TimeEnd DateEnd Time
10/24/2011085410/25/20110115
Subject:
TWP/SKYRAD/C2 - IRT power failure
DataStreams:twpskyrad20sC2.a0, twpskyrad60sC2.b1
Description:
The IRT measurements were incorrect due to a drop in the logger battery voltage.
Measurements:twpskyrad60sC2.b1:
  • IR Brightness Temperature(sky_ir_temp)
  • sky_ir_temp_std
  • sky_ir_temp_max
  • sky_ir_temp_min

twpskyrad20sC2.a0:
  • inst_sky_ir_signal


Back To Table of Contents