Data Quality Reports for Session: 138589 User: bouniol Completed: 11/07/2011


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D040416.1TWP/GNDRAD/C3 - IRT biastwpgndrad60sC3.b1
D050919.4TWP/GNDRAD/C3 - Tower inverted for instrument testing and replacementtwpgndrad60sC3.b1
D051114.3TWP/SKYRAD/C3 - tracker inoperativetwpskyradC3.00, twpskyrad20sC3.a0, twpskyrad60sC3.b1
D051202.3TWP/SKYRAD/C3 - Reprocess: Incorrect datalogger program affecting PIR1 and PIR2
downwelling longwave
twpskyrad60sC3.b1
D051206.5TWP/GNDRAD/C3 - IRT malfunctiontwpgndrad60sC3.b1
D060110.3TWP/SKYRAD/C3 - IRT damagedtwpskyrad20sC3.a0, twpskyrad60sC3.b1
D060111.1TWP/GNDRAD/C3 - IRT powertwpgndrad60sC3.b1
D060630.33TWP/SKYRAD/C3 - Reprocess: Longwave Calibration errortwpskyrad60sC3.b1
D060630.36TWP/GNDRAD/C3 - Reprocess: Longwave Calibration errortwpgndrad60sC3.b1
D061214.1TWP/SKYRAD/C3 - NIP instrument failuretwpskyradC3.00, twpskyrad20sC3.a0, twpskyrad60sC3.b1
D071001.4TWP/GNDRAD/C3 - Maintenance activities affected GNDRADtwpgndrad20sC3.a0, twpgndrad60sC3.b1


DQRID : D040416.1
Start DateStart TimeEnd DateEnd Time
04/07/2004023004/08/20041200
04/12/2004060004/27/20041400
05/08/2004030005/15/20042200
05/27/2004110006/09/20040000
12/29/2004110001/06/20052300
Subject:
TWP/GNDRAD/C3 - IRT bias
DataStreams:twpgndrad60sC3.b1
Description:
The IRT behaved erratically, jumping frequently from reporting about 50 Wm^(-2) higher 
than the PIR for surface irradiance values to about 100Wm^(-2) lower than the PIR.  
Generally, the IRT measured lower surface temperatures than expected from the PIR.
Measurements:twpgndrad60sC3.b1:
  • Effective Surface (Skin) Temperature(sfc_ir_temp)
  • sfc_ir_temp_std
  • Surface Infra-Red Temperature Maxima(sfc_ir_temp_max)
  • Surface Infra-Red Temperature Minima(sfc_ir_temp_min)


Back To Table of Contents

DQRID : D050919.4
Start DateStart TimeEnd DateEnd Time
08/05/2005003008/05/20050200
08/09/2005000008/09/20050200
Subject:
TWP/GNDRAD/C3 - Tower inverted for instrument testing and replacement
DataStreams:twpgndrad60sC3.b1
Description:
The tower at Darwin was brought down to test and replace the Gndrad PIR on 8/5/2005 and 
the PSP on 8/9/2005
Measurements:twpgndrad60sC3.b1:
  • Effective Surface (Skin) Temperature(sfc_ir_temp)
  • sfc_ir_temp_std
  • up_long_hemisp_std
  • Surface Infra-Red Temperature Maxima(sfc_ir_temp_max)
  • Upwelling Longwave Hemispheric Irradiance, Unventilated Pyrgeometer, Maxima(up_long_hemisp_max)
  • up_short_hemisp_std
  • Upwelling Shortwave Hemispheric Irradiance, Unventilated Pyranometer, Maxima(up_short_hemisp_max)
  • Upwelling Shortwave Hemispheric Irradiance, Unventilated Pyranometer, Minima(up_short_hemisp_min)
  • Upwelling Longwave Hemispheric Irradiance, Unventilated Pyrgeometer, Minima(up_long_hemisp_min)
  • Radiation, shortwave, at 10-m height, upwelling hemispheric irrad., 1-min avg(up_short_hemisp)
  • Surface Infra-Red Temperature Minima(sfc_ir_temp_min)
  • Radiation, longwave, at 10-m height, upwelling hemispheric irrad., 1-min avg(up_long_hemisp)


Back To Table of Contents

DQRID : D051114.3
Start DateStart TimeEnd DateEnd Time
11/14/2005045312/02/20050000
Subject:
TWP/SKYRAD/C3 - tracker inoperative
DataStreams:twpskyradC3.00, twpskyrad20sC3.a0, twpskyrad60sC3.b1
Description:
A lightning strike at Darwin removed the tracker from service.  Instruments installed on 
the tracker were rendered either inoperative or were not shaded during this time period.  
The longwave hemispheric instruments PIR1 and PIR2 appear to have been unaffected.  They 
are not suseptable to incorrect tracker operation.  The downwelling hemispheric shortwave 
PSP instrument mounted on Skyrad will not be affected by incorrect tracker operation.
Measurements:twpskyrad60sC3.b1:
  • Shortwave Direct Normal Irradiance, Pyrheliometer, Minima(short_direct_normal_min)
  • Radiation, shortwave, direct normal irradiance, 1-min avg(short_direct_normal)
  • Downwelling Shortwave Diffuse Hemisp. Irrad., Ventilated Pyranometer, Maxima(down_short_diffuse_hemisp_max)
  • down_short_diffuse_hemisp_std
  • Radiation, shortwave, downwelling diffuse hemispheric irradiance, 1-min avg(down_short_diffuse_hemisp)
  • Downwelling Shortwave Diffuse Hemisp. Irrad., Ventilated Pyranometer, Minima(down_short_diffuse_hemisp_min)
  • Shortwave Direct Normal Irradiance, Pyrheliometer, Maxima(short_direct_normal_max)
  • short_direct_normal_std

twpskyrad20sC3.a0:
  • inst_direct_normal
  • inst_diffuse

twpskyradC3.00:
  • Raw data stream - documentation not supported


Back To Table of Contents

DQRID : D051202.3
Start DateStart TimeEnd DateEnd Time
05/21/2004000009/09/20052012
Subject:
TWP/SKYRAD/C3 - Reprocess: Incorrect datalogger program affecting PIR1 and PIR2 
downwelling longwave
DataStreams:twpskyrad60sC3.b1
Description:
The excitation delays using in measuring the PIR1 case thermistor commonly named "Case1" 
and the PIR2 dome thermistor called "Dome2" were discovered to have been incorrectly set 
to 0 in the Skyrad datalogger program when the new CR23X based Skyrad loggers were 
installed.  The result of no delay in these measurements results in an error in the measured 
resistance of the thermistors, Case1 and Dome2, and thus the calculated PIR1 and PIR2 
irradiances.  The errors are inversely proportial to temperature.

Users can apply temperature dependent irradiance corrections as shown in the table below.  
Recall that the reported uncertainty of PIRs is typically +-.25% to +-.5% or 1 to 2 
W/m2. The error corrections become the same order of magnitude or larger than the instrument 
uncertainties at or below 260K or 9F. Irradiance Correction Table for Skyrad (TWP and 
NSA) DIR1 and DIR2 Irradiance

		DIR1	DIR1	DIR2	DIR2
	T[F]	Tcase[K]	Correction[W/m2]	Tdome[K]       
		Correction[W/m2]
	109.4	316.13	0.00	316.13	0.00
	100.5	311.20	-0.15	311.20	0.12
	93.2	307.14	-0.12	307.14	0.10
	87.0	303.70	-0.10	303.70	0.08
	81.6	300.70	-0.17	300.70	0.14
	77.0	298.15	-0.22	298.15	0.18
	59.4	288.35	-0.34	288.35	0.27
	47.4	281.72	-0.23	281.72	0.18
	38.5	276.74	-0.42	276.74	0.34
	31.3	272.77	-0.48	272.77	0.38
	25.4	269.48	-0.61	269.48	0.49
	20.3	266.67	-0.67	266.67	0.54
	16.0	264.25	-0.82	264.25	0.66
	12.1	262.11	-0.98	262.11	0.79
	8.7	260.19	-0.97	260.19	0.78
	5.6	258.46	-1.07	258.46	0.86
	2.7	256.89	-1.17	256.89	0.94
	0.1	255.44	-1.33	255.44	1.07
	-2.3	254.11	-1.41	254.11	1.13
	-4.5	252.87	-1.43	252.87	1.15
	-6.6	251.71	-1.52	251.71	1.22
	-8.5	250.63	-1.63	250.63	1.31
	-10.4	249.62	-1.73	249.62	1.39
	-12.2	248.58	-1.79	248.58	1.44
	-15.4	246.82	-1.93	246.82	1.55
	-18.3	245.23	-2.05	245.23	1.65
	-20.9	243.78	-2.22	243.78	1.78
	-23.3	242.44	-2.36	242.44	1.89
	-25.5	241.21	-2.58	241.21	2.07
	-27.5	240.07	-2.65	240.07	2.13
	-29.5	239.01	-2.75	239.01	2.21
	-31.3	238.00	-3.02	238.00	2.42
Measurements:twpskyrad60sC3.b1:
  • down_long_hemisp_shaded2_max
  • Irradiance, longwave, downwelling, hemispheric, shaded pyrgeometer1(down_long_hemisp_shaded1)
  • Irradiance, longwave, downwelling, hemispheric, shaded pyrgeometer2(down_long_hemisp_shaded2)
  • down_long_hemisp_shaded1_max
  • down_long_hemisp_shaded1_std
  • down_long_hemisp_shaded2_std
  • down_long_hemisp_shaded2_min
  • down_long_hemisp_shaded1_min


Back To Table of Contents

DQRID : D051206.5
Start DateStart TimeEnd DateEnd Time
11/21/2005150012/05/20050100
Subject:
TWP/GNDRAD/C3 - IRT malfunction
DataStreams:twpgndrad60sC3.b1
Description:
The internal IRT power supply failed (probably due to a lightning strike on 11/14) and was 
drawing excessive current on the data logger's DC system. The problem was corrected by 
replacing the instrument with a spare.
Measurements:twpgndrad60sC3.b1:
  • Effective Surface (Skin) Temperature(sfc_ir_temp)
  • sfc_ir_temp_std
  • Surface Infra-Red Temperature Maxima(sfc_ir_temp_max)
  • Surface Infra-Red Temperature Minima(sfc_ir_temp_min)


Back To Table of Contents

DQRID : D060110.3
Start DateStart TimeEnd DateEnd Time
11/13/2005231403/07/20060257
Subject:
TWP/SKYRAD/C3 - IRT damaged
DataStreams:twpskyrad20sC3.a0, twpskyrad60sC3.b1
Description:
The analog output of the IRT was damaged by a lightning strike. The problem was corrected 
by replacing the instrument with a spare. The damaged IRT was returned to the vendor for 
repair.
Measurements:twpskyrad60sC3.b1:
  • sky_ir_temp_min
  • IR Brightness Temperature(sky_ir_temp)
  • sky_ir_temp_std

twpskyrad20sC3.a0:
  • inst_sky_ir_signal


Back To Table of Contents

DQRID : D060111.1
Start DateStart TimeEnd DateEnd Time
12/20/2005233001/02/20062300
Subject:
TWP/GNDRAD/C3 - IRT power
DataStreams:twpgndrad60sC3.b1
Description:
The IRT failed to report after a planned power outage to add lightning protection modules 
to the AC power wiring. The problem was corrected when a loose power connection to the 
logger was found, rectified, and power restored to the logger.
Measurements:twpgndrad60sC3.b1:
  • Effective Surface (Skin) Temperature(sfc_ir_temp)
  • sfc_ir_temp_std
  • Surface Infra-Red Temperature Maxima(sfc_ir_temp_max)
  • Surface Infra-Red Temperature Minima(sfc_ir_temp_min)


Back To Table of Contents

DQRID : D060630.33
Start DateStart TimeEnd DateEnd Time
05/20/2004220012/21/20052136
Subject:
TWP/SKYRAD/C3 - Reprocess: Longwave Calibration error
DataStreams:twpskyrad60sC3.b1
Description:
Modified pyrgeometer calibration procedures were implemented beginning in December 2003.  
These modified procedures introduced a calibration bias in the longwave data.  The 
previous procedures were re-implemented at all sites between December 2005 and February 2006 to 
restore proper calibrations.                                                             
                The data collected while the incorrect procedures were in place are 
being reprocessing to remove the calibration bias.  Note, this also affected standard 
deviation, maximum and minimum data fields.
Measurements:twpskyrad60sC3.b1:
  • Irradiance, longwave, downwelling, hemispheric, shaded pyrgeometer1(down_long_hemisp_shaded1)
  • Irradiance, longwave, downwelling, hemispheric, shaded pyrgeometer2(down_long_hemisp_shaded2)


Back To Table of Contents

DQRID : D060630.36
Start DateStart TimeEnd DateEnd Time
05/20/2004010012/21/20052137
Subject:
TWP/GNDRAD/C3 - Reprocess: Longwave Calibration error
DataStreams:twpgndrad60sC3.b1
Description:
Modified pyrgeometer calibration procedures were implemented beginning in December 2003.  
These modified procedures introduced a calibration bias in the longwave data.  The 
previous procedures were re-implemented at all sites between December 2005 and February 2006 to 
restore proper calibrations.                                                             
                The data collected while the incorrect procedures were in place are 
being reprocessing to remove the calibration bias.  Note, this also affected standard 
deviation, maximum and minimum data fields.
Measurements:twpgndrad60sC3.b1:
  • Radiation, longwave, at 10-m height, upwelling hemispheric irrad., 1-min avg(up_long_hemisp)


Back To Table of Contents

DQRID : D061214.1
Start DateStart TimeEnd DateEnd Time
12/13/2006034201/16/20070440
Subject:
TWP/SKYRAD/C3 - NIP instrument failure
DataStreams:twpskyradC3.00, twpskyrad20sC3.a0, twpskyrad60sC3.b1
Description:
The NIP (Short Direct Normal) was replaced after it displayed symptoms of failure 
associated with either unknown precipitation related issues or a possible lightning strike.

NIP data are not recoverable.  Direct normal data can be reconstructed from global 
hemispheric and global diffuse data using the following relationship:
   global hemispheric = direct*cos(sza) + diffuse hemispheric
Measurements:twpskyrad60sC3.b1:
  • Shortwave Direct Normal Irradiance, Pyrheliometer, Minima(short_direct_normal_min)
  • Radiation, shortwave, direct normal irradiance, 1-min avg(short_direct_normal)
  • Shortwave Direct Normal Irradiance, Pyrheliometer, Maxima(short_direct_normal_max)
  • short_direct_normal_std

twpskyrad20sC3.a0:
  • inst_direct_normal

twpskyradC3.00:
  • Raw data stream - documentation not supported


Back To Table of Contents

DQRID : D071001.4
Start DateStart TimeEnd DateEnd Time
09/17/2007233009/18/20070722
09/18/2007232009/19/20070150
09/19/2007233309/20/20070055
Subject:
TWP/GNDRAD/C3 - Maintenance activities affected GNDRAD
DataStreams:twpgndrad20sC3.a0, twpgndrad60sC3.b1
Description:
While maintenance was performed on the SMET anenometer, the GNDRAD instruments were 
inverted. Disregard GNDRAD data during this time.
Measurements:twpgndrad20sC3.a0:
  • inst_up_long_dome_resist
  • inst_sfc_ir_signal
  • inst_up_short_hemisp
  • lat
  • lon
  • base_time
  • time
  • alt
  • inst_up_long_hemisp_tp
  • inst_up_long_case_resist
  • time_offset

twpgndrad60sC3.b1:
  • net_min
  • net_std
  • up_long_hemisp_std
  • time_offset
  • external_voltage
  • lat
  • Radiation, longwave, at 10-m height, upwelling hemispheric irrad., 1-min avg(up_long_hemisp)
  • lon
  • Effective Surface (Skin) Temperature(sfc_ir_temp)
  • logger_temp
  • base_time
  • net
  • sfc_ir_temp_std
  • net_max
  • alt
  • Surface Infra-Red Temperature Maxima(sfc_ir_temp_max)
  • Upwelling Longwave Hemispheric Irradiance, Unventilated Pyrgeometer, Maxima(up_long_hemisp_max)
  • up_short_hemisp_std
  • Upwelling Shortwave Hemispheric Irradiance, Unventilated Pyranometer, Maxima(up_short_hemisp_max)
  • bit_data
  • Upwelling Shortwave Hemispheric Irradiance, Unventilated Pyranometer, Minima(up_short_hemisp_min)
  • Upwelling Longwave Hemispheric Irradiance, Unventilated Pyrgeometer, Minima(up_long_hemisp_min)
  • internal_voltage
  • Surface Infra-Red Temperature Minima(sfc_ir_temp_min)
  • Radiation, shortwave, at 10-m height, upwelling hemispheric irrad., 1-min avg(up_short_hemisp)


Back To Table of Contents