Data Quality Reports for Session: 138352 User: lijun1ljr Completed: 10/27/2011


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D110628.2SGP/SIRS/C1 - NIP failuresgpsirsC1.b1, sgpsirs20sC1.a0


DQRID : D110628.2
Start DateStart TimeEnd DateEnd Time
06/12/2011000006/24/20111740
Subject:
SGP/SIRS/C1 - NIP failure
DataStreams:sgpsirsC1.b1, sgpsirs20sC1.a0
Description:
The short_direct_normal (NIP instrument) cable failed during this time period.	Data are 
not recoverable.

Users can derive an estimate of direct normal values using the following formula:
short_direct_normal = (down_short_hemisp - down_short_diffuse_hemisp)/cos(zenith angle)
Measurements:sgpsirs20sC1.a0:
  • inst_direct_normal

sgpsirsC1.b1:
  • Radiation, shortwave, direct normal irradiance, 1-min avg(short_direct_normal)
  • short_direct_normal_std
  • Shortwave Direct Normal Irradiance, Pyrheliometer, Maxima(short_direct_normal_max)
  • Shortwave Direct Normal Irradiance, Pyrheliometer, Minima(short_direct_normal_min)


Back To Table of Contents