Data Quality Reports for Session: 137642 User: obbe1tuine Completed: 09/26/2011


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D060928.3SGP/SONDE/C1 - Increased precision for alt, dp, pres, rh, tdrysgpsondewnpnC1.b1
D070613.2SGP/RAIN/C1 - Funnel cloggedsgprainC1.b1
D080307.1SGP/ORG/C1 - Frost on lenssgporgC1.b1
D080321.1SGP/ORG/C1 - Lens heater failuresgporgC1.b1
D080905.3SGP/ORG/C1 - Erroneous precipitation observationssgporgC1.b1


DQRID : D060928.3
Start DateStart TimeEnd DateEnd Time
09/28/2006172812/31/99992359
Subject:
SGP/SONDE/C1 - Increased precision for alt, dp, pres, rh, tdry
DataStreams:sgpsondewnpnC1.b1
Description:
The output precision of the following variables was increased 20060928

alt  changed from 1m to 0.1m
dp   changed from 0.1degC to 0.01degC
pres changed from 0.1 hPa to 0.01 hPa
RH   changed from 1%RH to 0.01%RH
tdry changed from .1degC to 0.01degC

Note that the increase in precision does not necessarily imply an increase in accuracy.  
Users are cautioned to continue to use their best scientific judgement when using these or 
any other data.
Measurements:sgpsondewnpnC1.b1:
  • alt
  • Dry Bulb Temperature(tdry)
  • Dewpoint Temperature(dp)
  • Humidity, relative(rh)
  • Pressure, atmospheric, constant value of model level, RUC model output(pres)


Back To Table of Contents

DQRID : D070613.2
Start DateStart TimeEnd DateEnd Time
06/01/2007000006/02/20072359
Subject:
SGP/RAIN/C1 - Funnel clogged
DataStreams:sgprainC1.b1
Description:
The funnel for the instrument was clogged and the results for this time period are not 
reliable.
Measurements:sgprainC1.b1:
  • lon
  • time
  • base_time
  • lat
  • Rainfall rate(rain_rate)
  • Precipitation(precip_tbrg)
  • time_offset
  • alt


Back To Table of Contents

DQRID : D080307.1
Start DateStart TimeEnd DateEnd Time
02/06/2008000002/06/20082359
Subject:
SGP/ORG/C1 - Frost on lens
DataStreams:sgporgC1.b1
Description:
Frost build up on the instrument lenses caused a faulty precipitation measurement.
Measurements:sgporgC1.b1:
  • Precipitation maximum(precip_max)
  • precip_sd
  • Precipitation mean(precip_mean)
  • Precipitation minimum(precip_min)


Back To Table of Contents

DQRID : D080321.1
Start DateStart TimeEnd DateEnd Time
01/01/2008000003/10/20081600
Subject:
SGP/ORG/C1 - Lens heater failure
DataStreams:sgporgC1.b1
Description:
Any measurements made in freezing conditions are suspect due to failure of the lens 
heaters.  Users of the data can use the SMOS E13 data to determine if temperatures below 0 
degrees C occurred.
Measurements:sgporgC1.b1:
  • Precipitation maximum(precip_max)
  • precip_sd
  • Precipitation mean(precip_mean)
  • Precipitation minimum(precip_min)


Back To Table of Contents

DQRID : D080905.3
Start DateStart TimeEnd DateEnd Time
07/28/2008000007/28/20081200
Subject:
SGP/ORG/C1 - Erroneous precipitation observations
DataStreams:sgporgC1.b1
Description:
Precipitation observed by the optical rain gauge was not observed by the SMOS tipping 
bucket or by the disdrometer nearby.  The optical rain gauge results should be considered 
erroneous.
Measurements:sgporgC1.b1:
  • Precipitation maximum(precip_max)
  • precip_sd
  • Precipitation mean(precip_mean)
  • Precipitation minimum(precip_min)


Back To Table of Contents



END OF DATA