Data Quality Reports for Session: 136479 User: lwymarie Completed: 07/28/2011


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D090507.2GRW/VCEIL/M1 - Receiver failuregrwvceil25kM1.b1
D100216.6GRW/VCEIL/M1 - Reprocess: Incorrect serial_number and sum_backscatter fieldsgrwvceil25kM1.b1


DQRID : D090507.2
Start DateStart TimeEnd DateEnd Time
04/22/2009092505/16/20091010
Subject:
GRW/VCEIL/M1 - Receiver failure
DataStreams:grwvceil25kM1.b1
Description:
The AMF VCEIL had lower-than-normal laser transmitter power, ~84%, so the instrument was 
replaced with a spare on 4/22. The spare VCEIL had just been repaired but began to display 
receiver failure warnings.

Diagnostic information was sent to Vaisala and we discussed either replacing the receiver 
with the one from the original VCEIL (with the degraded transmitter) or replacing the 
transmitter in the original VCEIL with the one from the spare. Vaisala recommended the 
latter because the original VCEIL is a much newer unit.

On 5/16, the site operator replaced the transmitter in the original VCEIL with the one 
from the spare, realigned the optics, recalibrated the laser, and put the instrument back 
into service, successfully correcting the problem.

The data during this period were intermittently incorrect, containing warning flags and 
alarms and unreasonable cloud base heights.
Measurements:grwvceil25kM1.b1:
  • Vertical visibility(vertical_visibility)
  • Lowest cloud base height detected.(first_cbh)
  • background_light
  • Backscatter(backscatter)
  • Third cloud base height(third_cbh)
  • Second lowest cloud base height(second_cbh)
  • alt_highest_signal


Back To Table of Contents

DQRID : D100216.6
Start DateStart TimeEnd DateEnd Time
04/13/2009000010/31/20092359
Subject:
GRW/VCEIL/M1 - Reprocess: Incorrect serial_number and sum_backscatter fields
DataStreams:grwvceil25kM1.b1
Description:
Attribute serial_number: In most cases, this field is not representative of the actual 
serial number of the instrument because the value did not change when an instrument changed 
locations or was replaced with a spare.
   
Variable sum_backscatter: An out-of-range condition for the "sum of detected and 
normalized backscatter" was reported as 0 instead of missing (-9999). The valid range is 0 - 
0.0999 /srad.
Measurements:grwvceil25kM1.b1:
  • sum_backscatter


Back To Table of Contents



END OF DATA