Data Quality Reports for Session: 136419 User: ql_data Completed: 07/26/2011


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D090209.3NSA/GNDRAD/C2 - Reprocess: IRT insufficiently heated nsagndrad60sC2.b1


DQRID : D090209.3
Start DateStart TimeEnd DateEnd Time
08/01/2000000006/19/20011830
Subject:
NSA/GNDRAD/C2 - Reprocess: IRT insufficiently heated
DataStreams:nsagndrad60sC2.b1
Description:
IRT had insufficient heating to maintain an internal reference temperature above 0 degrees 
C. Measurements of surface temperature were over-estimated when instrument was below 
freezing.

Data will be reprocessed, but users can correct data using the following correction 
factor:

If T_reference < -5?C, then

  IRT_corrected = (IRT_original - 32.993K)/0.87238

where T_reference can be estimated with the ambient temperature (e.g. 
inst_up_long_hemisp_dome_temp or inst_up_long_dome_temp)
Measurements:nsagndrad60sC2.b1:
  • Surface Infra-Red Temperature Maxima(sfc_ir_temp_max)
  • Effective Surface (Skin) Temperature(sfc_ir_temp)
  • Surface Infra-Red Temperature Minima(sfc_ir_temp_min)
  • sfc_ir_temp_std


Back To Table of Contents



END OF DATA