DQR ID | Subject | Data Streams Affected |
---|---|---|
D110118.1 | SBS/GNDRAD/M1 - Reprocess: IRT calibration error | sbsgndrad60sM1.b1 |
D110131.1 | SBS/MET/M1 - Atmospheric pressure spikes | sbsmetM1.b1 |
D110131.2 | SBS/MET/M1 - Site setup | sbsmetM1.b1 |
D110513.2 | SBS/MET/M1 - Incorrect atmospheric pressure | sbsmetM1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/23/2010 | 1700 | 01/18/2011 | 1556 |
Subject: | SBS/GNDRAD/M1 - Reprocess: IRT calibration error |
DataStreams: | sbsgndrad60sM1.b1 |
Description: | The IRT surface temperature reported was incorrect because the analog output configuration did not agree with the data logger scale factors. The IRT configuration was corrected 1/18. |
Measurements: | sbsgndrad60sM1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/15/2010 | 0000 | 01/28/2011 | 2000 |
Subject: | SBS/MET/M1 - Atmospheric pressure spikes |
DataStreams: | sbsmetM1.b1 |
Description: | A nearby AC unit was causing +/-.05 (approx.) changes in pressure values, most frequently during the daylight hours but occasionally over night. Corrective action was taken to shield the sensor 1/28. |
Measurements: | sbsmetM1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/27/2010 | 0000 | 11/15/2010 | 0000 |
Subject: | SBS/MET/M1 - Site setup |
DataStreams: | sbsmetM1.b1 |
Description: | The instruments were being setup and installed during this period. Some data may be questionable and minimal quality reporting will be done. |
Measurements: | sbsmetM1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/07/2011 | 1945 | 04/12/2011 | 1400 |
Subject: | SBS/MET/M1 - Incorrect atmospheric pressure |
DataStreams: | sbsmetM1.b1 |
Description: | The atmospheric pressure readings showed a ~.3 - 1 mb oscillatory behavior. The sensor tube was found to be obstructed. It was fixed and proper function resumed. |
Measurements: | sbsmetM1.b1:
|