Data Quality Reports for Session: 136255 User: wlj Completed: 07/19/2011


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D010831.3SGP/MFR10M/C1 - Head temperature problemssgpmfr10mC1.a0, sgpmfr10mC1.b1
D030916.7SGP/MFR10M/C1 - Artifact in datasgpmfr10mC1.a0, sgpmfr10mC1.b1
D050218.3SGP/MFR10M/C1 - Data discontinuitiessgpmfr10mC1.a0, sgpmfr10mC1.b1
D050218.4SGP/MFR10M/C1 - Dip in data at solar noonsgpmfr10mC1.a0, sgpmfr10mC1.b1
D060411.7SGP/MFR10M/C1 - Missing datasgpmfr10mC1.a0, sgpmfr10mC1.b1
D090520.3SGP/MFR10M/C1 - Heater failuresgpmfr10mC1.a0, sgpmfr10mC1.b1
D091112.3SGP/MFR10M/C1 - Bad connectionsgpmfr10mC1.a0, sgpmfr10mC1.b1
D100722.1SGP/MFR10M/C1 - Poor connectionsgpmfr10mC1.a0, sgpmfr10mC1.b1
D990906.1SGP/MFR10M/C1 - Head heater not workingsgpmfr10mC1.a0, sgpmfr10mC1.b1


DQRID : D010831.3
Start DateStart TimeEnd DateEnd Time
06/11/2001000009/30/20012359
Subject:
SGP/MFR10M/C1 - Head temperature problems
DataStreams:sgpmfr10mC1.a0, sgpmfr10mC1.b1
Description:
Head temperature control was not functioning during the specified period.
Measurements:sgpmfr10mC1.b1:
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • callang_flags
  • cosine_solar_zenith_angle
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • head_temp
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)

sgpmfr10mC1.a0:
  • callang_flags
  • head_temp
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)


Back To Table of Contents

DQRID : D030916.7
Start DateStart TimeEnd DateEnd Time
04/14/2003000009/16/20032359
Subject:
SGP/MFR10M/C1 - Artifact in data
DataStreams:sgpmfr10mC1.a0, sgpmfr10mC1.b1
Description:
A cable was hanging under the instrument and was in the field of view of the instrument. 
This resulted in some erroneous increases in the data, frequently around 22GMT. These 
increases are likely from specular reflections off the cable.
Measurements:sgpmfr10mC1.b1:
  • logger_volt
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • base_time
  • cosine_solar_zenith_angle
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)
  • time_offset
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • alt
  • callang_flags
  • lon
  • head_temp
  • lat

sgpmfr10mC1.a0:
  • callang_flags
  • time_offset
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)
  • logger_volt
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)
  • lon
  • base_time
  • head_temp
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • lat
  • alt


Back To Table of Contents

DQRID : D050218.3
Start DateStart TimeEnd DateEnd Time
06/27/2003143006/27/20031530
01/20/2005190001/20/20052000
Subject:
SGP/MFR10M/C1 - Data discontinuities
DataStreams:sgpmfr10mC1.a0, sgpmfr10mC1.b1
Description:
There is a discontinuity in the data (all channels) just after 1500 UTC on June 27, 2003 
and again, just before 1930 UTC on Jan 20, 2005.  There is nothing unusual in the temp or 
volt data.

At these times the tower was lowered for cleaning so the initial assumption was the 
instrument was dirty, and a cleaning resulted in the discontinuity.  This theory doesn't hold 
up well though as the tower is lowered for instrument cleaning regularly and these are the 
only two times in the past few years that a discontinuity coinciding with a lowering of 
the tower has resulted.  

Also, down pointing instruments are also not prone to collecting dust and dirt.  This 
instrument is located on the E13 SIRS tower and the down pointing PSP does not show a 
discontinuity at either time, which further discounts the dirty instrument theory.  Site 
operations were asked to check the instrument mount to see if it was loose possibly causing the 
instrument to change its position, and they reported that the mount seemed secure and 
stable.  So as of now, the cause of this issue is a mystery.

Compare with other down pointing instruments to help ascertain if data on one side of the 
discontinuity or the other are "better."  Perform an error analysis to determine if this 
would introduce a significant error in the role the data stream plays in a user's research.
Measurements:sgpmfr10mC1.b1:
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)

sgpmfr10mC1.a0:
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)


Back To Table of Contents

DQRID : D050218.4
Start DateStart TimeEnd DateEnd Time
09/01/2004000004/30/20050000
Subject:
SGP/MFR10M/C1 - Dip in data at solar noon
DataStreams:sgpmfr10mC1.a0, sgpmfr10mC1.b1
Description:
On clear days there is a dip in the data (all channels) at solar noon that is more 
pronounced during the winter months.  While we currently don't know the cause of this dip, we 
believe it is related to reflections off the tower pole.  The working theory says at solar 
noon the north-south crossbar shades the tower pole cutting off reflections into the 
detector.  If true, this means data on either side of the dip are being enhanced by 
reflections and the data within the dip are actually "correct."  An experiment is being planned 
for next fall that will wrap the upper portion of the tower in a non-reflecting cloth.  
This should determine if the dip is related to reflections from the tower pole.

Since at this point we don't know if the dip consists of bad or good data, users should be 
aware there is a problem and treat all the data as suspect.  It should noted that the 
magnitude of the problem is relatively small, likely no greater than 2.5%, so the dataset 
as whole is probably still useful as is for most users.
Measurements:sgpmfr10mC1.b1:
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)

sgpmfr10mC1.a0:
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)


Back To Table of Contents

DQRID : D060411.7
Start DateStart TimeEnd DateEnd Time
07/16/1994023007/19/19941831
08/25/1994203009/11/19941631
12/20/1994223012/29/19941248
05/23/1995163005/26/19951327
07/17/1995083001/07/19960533
01/19/1996153003/01/19961731
04/02/1997173004/04/19970019
05/23/1997140805/27/19970104
05/30/1997140806/02/19970314
10/31/1998111311/03/19981935
12/25/1998201312/28/19982014
03/14/1999002203/16/19990135
04/14/1999201305/04/19991914
06/12/1999061606/14/19991414
07/15/1999181307/19/19991813
11/24/2000151311/27/20001814
12/11/2000131312/13/20002115
08/18/2001000208/20/20011602
01/30/2002180202/11/20021502
03/09/2002200203/12/20020702
11/14/2002200211/18/20021602
07/23/2004210207/26/20041458
Subject:
SGP/MFR10M/C1 - Missing data
DataStreams:sgpmfr10mC1.a0, sgpmfr10mC1.b1
Description:
Data are missing and unrecoverable.
Measurements:sgpmfr10mC1.b1:
  • logger_volt
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • base_time
  • cosine_solar_zenith_angle
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)
  • time_offset
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • alt
  • callang_flags
  • lon
  • head_temp
  • lat

sgpmfr10mC1.a0:
  • callang_flags
  • time_offset
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)
  • logger_volt
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)
  • lon
  • base_time
  • head_temp
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • lat
  • alt


Back To Table of Contents

DQRID : D090520.3
Start DateStart TimeEnd DateEnd Time
05/17/2009052805/20/20092300
05/24/2009030006/10/20091600
Subject:
SGP/MFR10M/C1 - Heater failure
DataStreams:sgpmfr10mC1.a0, sgpmfr10mC1.b1
Description:
The heater stopped working on 20090517.  Maintenance on 20090520 seemed to resolve the 
problem, but the heater failed again on 5/24.  Additional maintenance on 20090610 seems to 
have resolved the issue.

The instrument is designed to operate with a head temp of around 40 deg C.  As the head 
was at ambient temp during the period covered by this DQR, the mentor suggests looking at 
the 25m MFR if one is interested in these data.
Measurements:sgpmfr10mC1.b1:
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • head_temp
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)

sgpmfr10mC1.a0:
  • head_temp
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)


Back To Table of Contents

DQRID : D091112.3
Start DateStart TimeEnd DateEnd Time
11/05/2009180911/25/20092000
Subject:
SGP/MFR10M/C1 - Bad connection
DataStreams:sgpmfr10mC1.a0, sgpmfr10mC1.b1
Description:
The instrument was pulled for quarterly calibration on 20091105.  When the instrument was 
reinstalled, the head cable was not making a good connection to the head.  Cleaning the 
connector and reinstalling resolved the issue.
Measurements:sgpmfr10mC1.b1:
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)

sgpmfr10mC1.a0:
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)


Back To Table of Contents

DQRID : D100722.1
Start DateStart TimeEnd DateEnd Time
05/11/2010201405/13/20101505
Subject:
SGP/MFR10M/C1 - Poor connection
DataStreams:sgpmfr10mC1.a0, sgpmfr10mC1.b1
Description:
The instrument was pulled for calibration on 5/11.  Upon re-installation, a good 
connection was not made.  Problem resolved on 5/13.
Measurements:sgpmfr10mC1.b1:
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • head_temp
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)

sgpmfr10mC1.a0:
  • head_temp
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)


Back To Table of Contents

DQRID : D990906.1
Start DateStart TimeEnd DateEnd Time
04/15/1999000009/02/19992221
Subject:
SGP/MFR10M/C1 - Head heater not working
DataStreams:sgpmfr10mC1.a0, sgpmfr10mC1.b1
Description:
On 19990415, the tower upon which the 10-m MFR was mounted was dismantled and replaced 
with a so-called "tippy tower".  After the 10-m MFR was reinstalled on the new tower, the 
mfr head heater did not work, or worked intermittently (at best).  A new mfr head was 
installed on 99.09.02 and that solved the problem.  

Data was collected from the old head in the morning up to the time it was replaced.  Once 
the new head was installed, data collection resumed.  There's a gap of about an hour 
between the time the old head was removed and the new head was installed.  If we try to gauge 
how well the old head was calibrated by comparing the two segments of data (old head, 
new head) it appears as if the old head remained in a good state of calibration even though 
the head heater was not working.
Measurements:sgpmfr10mC1.b1:
  • head_temp

sgpmfr10mC1.a0:
  • head_temp


Back To Table of Contents



END OF DATA