Subject:
SGP/IRT/E19 - Reprocess: Calibration error
DataStreams:sgpirtE19.b1, sgpirt200msE19.a1
Description:
The IRT was measuring ~20K higher sky temperatures than at other nearby extended facilies. 
The calibration of the instrument was incorrect. The problem was corrected by replacing 
the IRT.

The offset in the calibration will be determined and the data will be corrected.
Measurements:sgpirtE19.b1:
  • Temperature, Sky Infra-Red(sky_ir_temp)
  • sky_ir_temp_max
  • sky_ir_temp_min

sgpirt200msE19.a1:
  • Temperature, Sky Infra-Red(sky_ir_temp)


Back To Table of Contents

DQRID : D091215.3
Start DateStart TimeEnd DateEnd Time
12/01/2005000003/30/20062359
Subject:
SGP/IRT/E19 - Response time
DataStreams:sgpirtE19.b1, sgpirt200msE19.a1
Description:
The time response of the temperature measurement was changed from 1.0 to 0.1 sec to 
accommodate the 5Hz sample rate. This caused a 370% decrease in the resolution (to 1.85C at a 
sky temperature of -50C) and a significant decrease in the signal-to-noise ratio.
Measurements:sgpirtE19.b1:
  • sky_ir_temp_std
  • ref_ir_temp_std

sgpirt200msE19.a1:
  • ref_ir_temp
  • Temperature, Sky Infra-Red(sky_ir_temp)