Data Quality Reports for Session: 135038 User: goulven1mo Completed: 05/23/2011


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D080428.3TWP/SMET/C3 - Reprocess: ORG Data has very small rainrate valuestwpsmet60sC3.b1
D090129.4TWP/SMET/C3 - Reprocess: Station Pressure values reported in hPa instead of kPa.twpsmet60sC3.b1
D090515.1TWP/SMET/C3 - Reprocess: Temperature Biased low by 1.06 Degreestwpsmet60sC3.b1
D091023.1TWP/SMET/C3 - Tower Lowered for Maintenancetwpsmet60sC3.b1


DQRID : D080428.3
Start DateStart TimeEnd DateEnd Time
05/01/2004000006/30/20102359
Subject:
TWP/SMET/C3 - Reprocess: ORG Data has very small rainrate values
DataStreams:twpsmet60sC3.b1
Description:
The ORG data is collected via analog signal and not digital.  Therefore, we
continuously collect data even when no rain is occurring.  We also capture small events 
such as leaves, large bugs, dust, etc blowing through the measurement path.  Because this 
discrete event occurs quickly the value is usually captured for 1 or 2 seconds.  This 
value is then averaged to the minute producing rainrate on the order of .009 mm/hr or less.  

These values are not indicative of rain and should be removed from the record or summing 
over long periods can bias the climate record.  We will reprocess the data to remove these 
small values.
Measurements:twpsmet60sC3.b1:
  • precip_sd
  • Precipitation maximum(precip_max)
  • Precipitation minimum(precip_min)
  • Precipitation mean(precip_mean)


Back To Table of Contents

DQRID : D090129.4
Start DateStart TimeEnd DateEnd Time
01/27/2009185001/30/20091535
Subject:
TWP/SMET/C3 - Reprocess: Station Pressure values reported in hPa instead of kPa.
DataStreams:twpsmet60sC3.b1
Description:
An incorrect program was entered into the datalogger.  This program did not convert the 
station pressure from hPa to kPa.  This made the values larger by a factor of 10 and caused 
the QC bit to be tripped.
Measurements:twpsmet60sC3.b1:
  • Atmospheric pressure(atmos_pressure)


Back To Table of Contents

DQRID : D090515.1
Start DateStart TimeEnd DateEnd Time
04/19/2005042004/04/20090115
Subject:
TWP/SMET/C3 - Reprocess: Temperature Biased low by 1.06 Degrees
DataStreams:twpsmet60sC3.b1
Description:
The temperature of the SMET system was biased low by 1.06 Deg C due to a non-precision 
resistor being used int he measurement circuit.  The data will be reprocessed.  Saturation 
Vapor pressure and Vapor pressure will have to be reprocessed also.

Add 1.06 degrees to the temperature data.  

Correct Sat Vap Press using the Buck approximation:
   Es = 6.1121exp((17.502*T)/(240.97 + T) 
where T is the temperature in C.  

Correct Vapor pressure using: E = (RH*Es)/100
Measurements:twpsmet60sC3.b1:
  • temp_sd
  • Vapor pressure mean(vappress_mean)
  • Temperature mean(temp_mean)
  • vappress_sd


Back To Table of Contents

DQRID : D091023.1
Start DateStart TimeEnd DateEnd Time
09/30/2009004509/30/20090500
Subject:
TWP/SMET/C3 - Tower Lowered for Maintenance
DataStreams:twpsmet60sC3.b1
Description:
The tower was lowered for annual anemometer changes.  The wind data is suspect.
Measurements:twpsmet60sC3.b1:
  • lo_spd_sd
  • lo_spd_vec_avg
  • lo_spd_min
  • lo_dir_sd
  • up_spd_arith_avg
  • lo_spd_max
  • up_spd_vec_avg
  • up_dir_sd
  • up_spd_max
  • up_spd_sd
  • up_spd_min
  • lo_dir_vec_avg
  • lo_spd_arith_avg
  • up_dir_vec_avg