DQRID : D060112.4
Start DateStart TimeEnd DateEnd Time
11/26/2005000006/07/20061900
Subject:
NIM/SKYRAD/M1 - IRTs do not agree with AERI
DataStreams:nimskyrad60sM1.b1
Description:
Since deployment at PYE, and then at NIM, the AMF SKYRAD IRT measured about 10K higher sky 
temperatures than the AERI and the MWRP IRT measureed about 20K higher than the AERI.  
Several actions were taken to diagnose the problem including confirming the correct 
configuration of the IRTs and data logger, cleaning the mirror and lens, and replacing the mirror.

After several days of rain beginning 6/2/2006, the three instruments came into agreement. 
It is unknown whether this was a problem with the AERI, MWRP-IRT or SKYRAD-IRT.
Measurements:nimskyrad60sM1.b1:
  • sky_ir_temp_min
  • Temperature, Sky Infra-Red(sky_ir_temp)
  • sky_ir_temp_max


Back To Table of Contents

DQRID : D060222.1
Start DateStart TimeEnd DateEnd Time
12/23/2005000002/23/20060800
Subject:
NIM/SKYRAD/M1 - Instrument ventilator problem
DataStreams:nimskyradM1.00, nimskyrad20sM1.a0, nimskyrad60sM1.b1
Description:
Data quality office first reported that PIR1 and PIR2 appear to be deviating more than 
typical during the daylight hours.  It was found that the Skyrad PIR ventilator fans were 
building up debris on the intake air filter.  Obstructed air flow results in solar heating 
of the PIR ventilator housing in the later part of the day and biasing the incomming IR 
irradiance values. The slow buildup was insidious and wasn't discovered and corrected 
(i.e. ventilator fans cleaned on a regular and frequent basis) until Feb 23rd. On or around 
December 23rd was when this problem may have first started.  We are unable to quantify the 
extent of the problem in W/m2.  Normal deviation between PIRs is typically 3-5 W/m2.
Measurements:nimskyrad20sM1.a0:
  • Irradiance, longwave, downwelling, hemispheric, thermopile (Shaded Pyrgeometer1)(inst_down_long_hemisp_shaded1_tp)
  • Irradiance, longwave, downwelling, hemispheric, thermopile (Shaded Pyrgeometer2)(inst_down_long_hemisp_shaded2_tp)
  • inst_down_long_shaded2_case_resist
  • inst_down_long_shaded2_dome_resist
  • inst_down_long_shaded1_case_resist
  • inst_down_long_shaded1_dome_resist

nimskyrad60sM1.b1:
  • inst_down_long_shaded2_dome_temp
  • down_long_hemisp_shaded1_std
  • down_long_hemisp_shaded1_min
  • down_long_hemisp_shaded1_max
  • Irradiance, longwave, downwelling, hemispheric, thermopile (Shaded Pyrgeometer2)(inst_down_long_hemisp_shaded2_tp)
  • Irradiance, longwave, downwelling, hemispheric, shaded pyrgeometer1(down_long_hemisp_shaded1)
  • down_long_hemisp_shaded2_min
  • Irradiance, longwave, downwelling, hemispheric, shaded pyrgeometer2(down_long_hemisp_shaded2)
  • down_long_hemisp_shaded2_max
  • inst_down_long_shaded1_dome_temp
  • down_long_hemisp_shaded2_std
  • inst_down_long_shaded2_case_temp
  • Irradiance, longwave, downwelling, hemispheric, thermopile (Shaded Pyrgeometer1)(inst_down_long_hemisp_shaded1_tp)
  • inst_down_long_shaded1_case_temp

nimskyradM1.00:
  • Raw data stream - documentation not supported


Back To Table of Contents

DQRID : D061006.1
Start DateStart TimeEnd DateEnd Time
09/12/2006120010/05/20061330
Subject:
NIM/SKYRAD/M1 - Tracker inoperative
DataStreams:nimskyradM1.00, nimskyrad20sM1.a0, nimskyrad60sM1.b1
Description:
Starting on 9/12 the NIM radiometer tracker drifted enough due to sinking supports and a 
drifting clock that the direct normal measurments (NIP) and the downwelling hemispheric 
diffuse shortwave measurments (8-48) were adversely affected in the afternoon hours.  The 
shortwave hemispheric global instrument (PSP) and downelling hemispheric longwave shaded1 
and shaded2 (PIR) are unaffected by this problem. 

The tracker problem greatly affected direct measurements and affected the diffuse 
measurement to a lesser extent. There are no redundant measurements for these instruments.  
However, during this time range the data collected before local noon (aprox 1140GMT) for all 
shortwave Skyrad instruments appear to be good.
Measurements:nimskyrad20sM1.a0:
  • inst_diffuse
  • inst_direct_normal

nimskyrad60sM1.b1:
  • down_short_diffuse_hemisp_std
  • Radiation, shortwave, downwelling diffuse hemispheric irradiance, 1-min avg(down_short_diffuse_hemisp)
  • Shortwave Direct Normal Irradiance, Pyrheliometer, Minima(short_direct_normal_min)
  • short_direct_normal_std
  • Radiation, shortwave, direct normal irradiance, 1-min avg(short_direct_normal)
  • Downwelling Shortwave Diffuse Hemisp. Irrad., Ventilated Pyranometer, Minima(down_short_diffuse_hemisp_min)
  • Shortwave Direct Normal Irradiance, Pyrheliometer, Maxima(short_direct_normal_max)
  • Downwelling Shortwave Diffuse Hemisp. Irrad., Ventilated Pyranometer, Maxima(down_short_diffuse_hemisp_max)

nimskyradM1.00:
  • Raw data stream - documentation not supported


Back To Table of Contents



END OF DATA