Data Quality Reports for Session: 133207 User: roj Completed: 02/12/2011


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D040127.3TWP/SONDE/C2 - Broken temperature sensorstwpsondewnpnC2.b1
D050913.3TWP/SONDE/C2 - Bad RH sensortwpsondewnpnC2.b1
D051122.2TWP/SONDE/C2 - Broken temperature sensortwpsondewnpnC2.b1
D051219.2TWP/SONDE/C2 - Broken temperature sensortwpsondewnpnC2.b1
D060308.1TWP/SONDE/C2 - Broken temperature sensortwpsondewnpnC2.b1
D060403.2TWP/SONDE/C2 - Broken temperature sensortwpsondewnpnC2.b1
D060418.1TWP/SONDE/C2 - Bad RH in lower 8kmtwpsondewnpnC2.b1
D070308.3TWP/SONDE/C2 - Surface Temperature IncorrecttwpsondewnpnC2.b1
D070308.4TWP/SONDE/C2 - Surface RH value incorrecttwpsondewnpnC2.b1
D070313.2TWP/SONDE/C2 - Surface RH value incorrecttwpsondewnpnC2.b1
D070329.2TWP/SONDE/C2 - Surface RH Value incorrecttwpsondewnpnC2.b1
D070404.1TWP/SONDE/C2 - Surface RH too hightwpsondewnpnC2.b1
D070428.1TWP/SONDE/C2 - Broken temperature sensor at launchtwpsondewnpnC2.b1
D070501.2TWP/SONDE/C2 - Broken temperature sensor at launchtwpsondewnpnC2.b1
D080925.1TWP/SONDE/C2 - Bad RH values throughout soundingtwpsondewnpnC2.b1
D081114.1TWP/SONDE/C2 - Defective Temperature SensortwpsondewnpnC2.b1
D090224.4TWP/SONDE/C2 - Suspect RH valuestwpsondewnpnC2.b1
D090629.4TWP/SONDE/C2 - Wind data missingtwpsondewnpnC2.b1


DQRID : D040127.3
Start DateStart TimeEnd DateEnd Time
10/13/2003004010/13/20030203
11/08/2003233511/09/20030215
11/18/2003120511/18/20031340
11/19/2003115611/19/20031346
11/21/2003004411/21/20030113
12/20/2003122512/20/20031408
12/21/2003005612/21/20030126
12/21/2003115512/21/20031344
12/23/2003004212/23/20030212
12/23/2003113712/23/20031259
12/23/2003234212/24/20030132
01/08/2004113501/08/20041309
01/09/2004115701/09/20041342
01/09/2004135901/09/20041537
01/12/2004234201/13/20040214
01/16/2004140901/16/20041548
01/16/2004233701/17/20040109
01/18/2004113501/18/20041256
01/19/2004022301/19/20040346
01/19/2004232901/20/20040157
01/21/2004125201/21/20041400
01/22/2004005201/22/20040227
01/23/2004113701/23/20041319
01/26/2004120201/26/20040146
02/17/2004140602/17/20041536
02/26/2004122102/26/20041350
03/09/2004114503/09/20041315
03/14/2004113103/14/20041300
03/22/2004113403/22/20041300
03/29/2004233203/30/20040047
03/30/2004233403/31/20040059
04/11/2004234804/12/20040120
05/01/2004233505/02/20040110
05/02/2004233605/03/20040109
05/13/2004233405/14/20040112
05/14/2004140605/14/20041411
06/07/2004022406/07/20040351
06/25/2004114606/25/20041328
07/02/2004005907/02/20040217
07/20/2004022007/20/20040358
07/24/2004123507/24/20041355
08/22/2004114408/22/20041207
09/08/2004142509/08/20041605
09/08/2004233809/08/20042345
10/18/2004234610/19/20040128
11/29/2004232111/30/20040105
12/08/2004114312/08/20041306
12/30/2004113012/30/20041239
Subject:
TWP/SONDE/C2 - Broken temperature sensors
DataStreams:twpsondewnpnC2.b1
Description:
The temperature sensors on these radiosondes broke after launch.  All
temperature-related data are incorrect.
Measurements:twpsondewnpnC2.b1:
  • Humidity, relative(rh)
  • Dewpoint Temperature(dp)
  • Dry Bulb Temperature(tdry)


Back To Table of Contents

DQRID : D050913.3
Start DateStart TimeEnd DateEnd Time
07/03/2005235007/04/20050024
07/10/2005234007/11/20050021
07/19/2005233207/20/20050016
Subject:
TWP/SONDE/C2 - Bad RH sensor
DataStreams:twpsondewnpnC2.b1
Description:
It appears that the RH sensor heating circuit failed during this sounding.  The RH 
oscillates rapidly between low (~0 %RH) and not so low (ambient %RH) and the oscillation seems 
to stop when the temperature is below -50 degC.  According to Vaisala these are symptoms 
of failure in the ASIC (application specific integrated circuit) that controls the heating.
Measurements:twpsondewnpnC2.b1:
  • Humidity, relative(rh)
  • Dewpoint Temperature(dp)


Back To Table of Contents

DQRID : D051122.2
Start DateStart TimeEnd DateEnd Time
11/19/2005233411/20/20050153
11/21/2005233211/22/20050140
Subject:
TWP/SONDE/C2 - Broken temperature sensor
DataStreams:twpsondewnpnC2.b1
Description:
The temperature sensor on this radiosonde broke after launch.  All the temperature-related 
data are incorrect.
Measurements:twpsondewnpnC2.b1:
  • Humidity, relative(rh)
  • Dewpoint Temperature(dp)
  • Dry Bulb Temperature(tdry)


Back To Table of Contents

DQRID : D051219.2
Start DateStart TimeEnd DateEnd Time
12/07/2005115312/07/20051323
Subject:
TWP/SONDE/C2  - Broken temperature sensor
DataStreams:twpsondewnpnC2.b1
Description:
The temperature sensor on this radiosonde broke after launch.  All the temperature-related 
data are incorrect.
Measurements:twpsondewnpnC2.b1:
  • Humidity, relative(rh)
  • Dewpoint Temperature(dp)
  • Dry Bulb Temperature(tdry)


Back To Table of Contents

DQRID : D060308.1
Start DateStart TimeEnd DateEnd Time
01/15/2006114401/15/20061328
01/19/2006131801/19/20061320
01/22/2006114901/22/20061315
02/14/2006005502/14/20060207
02/21/2006023202/21/20060359
Subject:
TWP/SONDE/C2  - Broken temperature sensor
DataStreams:twpsondewnpnC2.b1
Description:
The temperature sensor on this radiosonde broke after launch.  All the temperature-related 
data are incorrect.
Measurements:twpsondewnpnC2.b1:
  • Humidity, relative(rh)
  • Dewpoint Temperature(dp)
  • Dry Bulb Temperature(tdry)


Back To Table of Contents

DQRID : D060403.2
Start DateStart TimeEnd DateEnd Time
03/27/2006004303/27/20060208
Subject:
TWP/SONDE/C2  - Broken temperature sensor
DataStreams:twpsondewnpnC2.b1
Description:
The temperature sensor on this radiosonde broke after launch.  All the temperature-related 
data are incorrect.
Measurements:twpsondewnpnC2.b1:
  • Humidity, relative(rh)
  • Dewpoint Temperature(dp)
  • Dry Bulb Temperature(tdry)


Back To Table of Contents

DQRID : D060418.1
Start DateStart TimeEnd DateEnd Time
04/10/2006024704/10/20060314
Subject:
TWP/SONDE/C2  - Bad RH in lower 8km
DataStreams:twpsondewnpnC2.b1
Description:
It appears that the RH sensor heating circuit failed during this sounding.  The RH 
oscillates rapidly between low (~0 %RH) and not so low (ambient %RH) and the oscillation seems 
to stop when the temperature is below ~35 degC.  According to Vaisala these are symptoms 
of failure in the ASIC (application specific integrated circuit) that controls the heating.
Measurements:twpsondewnpnC2.b1:
  • Humidity, relative(rh)
  • Dewpoint Temperature(dp)


Back To Table of Contents

DQRID : D070308.3
Start DateStart TimeEnd DateEnd Time
01/02/2007112801/02/20071129
Subject:
TWP/SONDE/C2 - Surface Temperature Incorrect
DataStreams:twpsondewnpnC2.b1
Description:
The first two temperature values in the datafile are incorrect.  It appears that the 
observer mis-typed the surface temperature as -27 instead of +27.  The next value was 8 which 
is also incorrect and was caused by the system interpolating from the incorrect surface 
value and what the sonde temperature sensor was reading.
Measurements:twpsondewnpnC2.b1:
  • Dry Bulb Temperature(tdry)


Back To Table of Contents

DQRID : D070308.4
Start DateStart TimeEnd DateEnd Time
01/23/2007113001/23/20071131
Subject:
TWP/SONDE/C2 - Surface RH value incorrect
DataStreams:twpsondewnpnC2.b1
Description:
The surface RH value was incorrect.  It appears the observer mis-typed 70 instead of 80.  
This error also caused the dew point temperature to be incorrect.
Measurements:twpsondewnpnC2.b1:
  • Humidity, relative(rh)
  • Dewpoint Temperature(dp)


Back To Table of Contents

DQRID : D070313.2
Start DateStart TimeEnd DateEnd Time
03/08/2007235003/08/20072353
Subject:
TWP/SONDE/C2 - Surface RH value incorrect
DataStreams:twpsondewnpnC2.b1
Description:
The surface RH was reproted as 78%.  The value should have been in the low 50's.  It 
appears the observer did not allow the sonde to equilibrate to ambient conditions before 
taking the reading.  This error also caused an incorrect dew point.
Measurements:twpsondewnpnC2.b1:
  • Humidity, relative(rh)
  • Dewpoint Temperature(dp)


Back To Table of Contents

DQRID : D070329.2
Start DateStart TimeEnd DateEnd Time
03/28/2007231903/28/20072320
Subject:
TWP/SONDE/C2 - Surface RH Value incorrect
DataStreams:twpsondewnpnC2.b1
Description:
The surface value entered by the observers for RH (96%) is incorrect.  At the ascent time 
listed in the .ptu file the readings by the Sonde were around 65%.  This low RH value is 
supported by the nearby SMET tower RH reading of 71.8%.  This also caused the surface dew 
point to be incorrect.
Measurements:twpsondewnpnC2.b1:
  • Humidity, relative(rh)
  • Dewpoint Temperature(dp)


Back To Table of Contents

DQRID : D070404.1
Start DateStart TimeEnd DateEnd Time
04/03/2007233004/03/20072331
Subject:
TWP/SONDE/C2 - Surface RH too high
DataStreams:twpsondewnpnC2.b1
Description:
The surface Rh was entered as 94%.  The ascent time has 72% and this value is supported by 
RH values from the nearby SMET system.  This error also will affect the dew point.
Measurements:twpsondewnpnC2.b1:
  • Humidity, relative(rh)
  • Dewpoint Temperature(dp)


Back To Table of Contents

DQRID : D070428.1
Start DateStart TimeEnd DateEnd Time
04/27/2007114004/27/20071420
Subject:
TWP/SONDE/C2 - Broken temperature sensor at launch
DataStreams:twpsondewnpnC2.b1
Description:
Temperature sensor broken at launch.  All thermodynamic values are incorrect.
Measurements:twpsondewnpnC2.b1:
  • Wind Direction(deg)
  • Humidity, relative(rh)
  • Dewpoint Temperature(dp)


Back To Table of Contents

DQRID : D070501.2
Start DateStart TimeEnd DateEnd Time
04/30/2007235905/01/20070200
Subject:
TWP/SONDE/C2 - Broken temperature sensor at launch
DataStreams:twpsondewnpnC2.b1
Description:
The temperature sensor on this radiosonde broke after launch.  All the temperature-related 
data are incorrect.
Measurements:twpsondewnpnC2.b1:
  • Humidity, relative(rh)
  • Dewpoint Temperature(dp)
  • Dry Bulb Temperature(tdry)


Back To Table of Contents

DQRID : D080925.1
Start DateStart TimeEnd DateEnd Time
09/08/2008114509/08/20081145
Subject:
TWP/SONDE/C2 - Bad RH values throughout sounding
DataStreams:twpsondewnpnC2.b1
Description:
RH values fluctuate largely in a stepping function throughout the sounding.  Suspect bad 
RH sensors.
Measurements:twpsondewnpnC2.b1:
  • Humidity, relative(rh)


Back To Table of Contents

DQRID : D081114.1
Start DateStart TimeEnd DateEnd Time
11/04/2008114511/04/20081145
Subject:
TWP/SONDE/C2 - Defective Temperature Sensor
DataStreams:twpsondewnpnC2.b1
Description:
Temperature sensor shows a +30C offset for duration of flight.
Measurements:twpsondewnpnC2.b1:
  • Dry Bulb Temperature(tdry)


Back To Table of Contents

DQRID : D090224.4
Start DateStart TimeEnd DateEnd Time
01/08/2009233101/08/20092331
Subject:
TWP/SONDE/C2 - Suspect RH values
DataStreams:twpsondewnpnC2.b1
Description:
RH values only are suspect for the duration of this radiosonde flight, exhibiting a 
"sawtooth" pattern.
Measurements:twpsondewnpnC2.b1:
  • Humidity, relative(rh)


Back To Table of Contents

DQRID : D090629.4
Start DateStart TimeEnd DateEnd Time
06/21/2009110006/26/20091100
06/27/2009120006/27/20091200
Subject:
TWP/SONDE/C2 - Wind data missing
DataStreams:twpsondewnpnC2.b1
Description:
A GPS problem associated with Vaisala BBSS ground station software issues caused the 
ground stations to enter into an error state when using GPS wind finding. In some instances, 
the flight was aborted due to the error state and not data was collected.  A workaround 
was developed to allow sondes to be launched without collecting wind data.
Measurements:twpsondewnpnC2.b1:
  • wstat
  • Eastward Wind Component(u_wind)
  • Northward Wind Component(v_wind)
  • Mean Wind Speed(wspd)


Back To Table of Contents



END OF DATA