Data Quality Reports for Session: 133140 User: yding Completed: 02/10/2011


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D090507.4sgpvceil25kC1.b1
D100215.1SGP/VCEIL/C1/B4/B5/B6 - Reprocess: Incorrect serial_number and sum_backscatter fieldssgpvceil25kB4.b1, sgpvceil25kB5.b1, sgpvceil25kB6.b1, sgpvceil25kC1.b1


DQRID : D090507.4
Start DateStart TimeEnd DateEnd Time
05/06/2009180005/07/20091910
Subject:
DataStreams:sgpvceil25kC1.b1
Description:
Measurements:sgpvceil25kC1.b1:
  • receiver_sensitivity
  • lon
  • Lowest cloud base height detected.(first_cbh)
  • base_time
  • window_contamination
  • tilt_angle
  • lat
  • alt_highest_signal
  • alt
  • measurement_parameters
  • laser_temperature
  • Second lowest cloud base height(second_cbh)
  • laser_pulse_energy
  • range
  • Vertical visibility(vertical_visibility)
  • sum_backscatter
  • background_light
  • status_string
  • detection_status
  • Backscatter(backscatter)
  • status_flag
  • Third cloud base height(third_cbh)
  • time_offset


Back To Table of Contents

DQRID : D100215.1
Start DateStart TimeEnd DateEnd Time
04/01/2001000010/31/20092359
Subject:
SGP/VCEIL/C1/B4/B5/B6 - Reprocess: Incorrect serial_number and sum_backscatter fields
DataStreams:sgpvceil25kB4.b1, sgpvceil25kB5.b1, sgpvceil25kB6.b1, sgpvceil25kC1.b1
Description:
Attribute serial_number: In most cases, this field is not representative of the actual 
serial number of the instrument because the value did not change when an instrument changed 
locations or was replaced with a spare.

Variable sum_backscatter: An out-of-range condition for the "sum of detected and 
normalized backscatter" was reported as 0 instead of missing (-9999). The valid range is 0 - 
0.0999 /srad.

This DQR also applies to a1 level data collected prior to b1 level.
Measurements:sgpvceil25kC1.b1:
  • sum_backscatter

sgpvceil25kB4.b1:
  • sum_backscatter

sgpvceil25kB6.b1:
  • sum_backscatter

sgpvceil25kB5.b1:
  • sum_backscatter


Back To Table of Contents



END OF DATA