Data Quality Reports for Session: 132145 User: reh10d Completed: 12/13/2010


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D050105.2SGP/IRT/C1 - Reprocess: Some min and max values swappedsgpirtC1.b1
D060630.6SGP/IRT/C1 - Reprocess: Longwave Calibration errorsgpirt25mC1.b1
D060828.2SGP/IRT/C1 - UPS failingsgpirt10mC1.b1
D080930.2SGP/IRT/C1 - Contamination of 25M lenssgpirt25mC1.b1


DQRID : D050105.2
Start DateStart TimeEnd DateEnd Time
12/18/2003160302/02/20052359
Subject:
SGP/IRT/C1 - Reprocess: Some min and max values swapped
DataStreams:sgpirtC1.b1
Description:
During the time period specified the minimum and maximum sky_ir_temp
values for the first record of each hour are swapped with the minimum
being reported as sky_ir_temp_max and the maximum being reported as
sky_ir_temp_min.
Measurements:sgpirtC1.b1:
  • sky_ir_temp_max
  • sky_ir_temp_min


Back To Table of Contents

DQRID : D060630.6
Start DateStart TimeEnd DateEnd Time
01/15/2004205202/14/20062050
Subject:
SGP/IRT/C1 - Reprocess: Longwave Calibration error
DataStreams:sgpirt25mC1.b1
Description:
Modified pyrgeometer calibration procedures were implemented beginning in December 2003.  
These modified procedures introduced a calibration bias in the longwave data.  The 
previous procedures were re-implemented at all sites between December 2005 and February 2006 to 
restore proper calibrations.                                                             
                The data collected while the incorrect procedures were in place are 
being reprocessing to remove the calibration bias.  Note, this also affected standard 
deviation, maximum and minimum data fields.
Measurements:sgpirt25mC1.b1:
  • Radiation, longwave, at 10-m height, upwelling hemispheric irrad., 1-min avg(up_long_hemisp)
  • Instantaneous Upwelling Longwave Hemispheric Irradiance, Pyrgeometer(inst_up_long_hemisp)


Back To Table of Contents

DQRID : D060828.2
Start DateStart TimeEnd DateEnd Time
08/26/2006085309/18/20061530
Subject:
SGP/IRT/C1 - UPS failing
DataStreams:sgpirt10mC1.b1
Description:
The IRT data were intermittently bad during this period due to a failing UPS. The problem 
was corrected by replacing the UPS.
Measurements:sgpirt10mC1.b1:
  • internal_voltage
  • Surface Infra-Red Temperature Maxima(sfc_ir_temp_max)
  • sfc_ir_temp_std
  • Surface Infra-Red Temperature Minima(sfc_ir_temp_min)
  • Effective Surface (Skin) Temperature(sfc_ir_temp)


Back To Table of Contents

DQRID : D080930.2
Start DateStart TimeEnd DateEnd Time
05/25/2008000010/01/20081905
Subject:
SGP/IRT/C1 - Contamination of 25M lens
DataStreams:sgpirt25mC1.b1
Description:
The IRT exhibited a positive bias compared to PIR. The problem was corrected by cleaning 
the lens.
Measurements:sgpirt25mC1.b1:
  • Effective Surface (Skin) Temperature(sfc_ir_temp)
  • Instantaneous Surface Infra-red Temperature(inst_sfc_ir_temp)
  • Surface Infra-Red Temperature Minima(sfc_ir_temp_min)
  • sfc_ir_temp_std
  • Surface Infra-Red Temperature Maxima(sfc_ir_temp_max)


Back To Table of Contents



END OF DATA