Data Quality Reports for Session: 130637 User: pschmid Completed: 09/30/2010


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D080604.2SGP/ECOR/E14 - Sensor failuresgp30ecorE14.b1


DQRID : D080604.2
Start DateStart TimeEnd DateEnd Time
04/02/2008133006/05/20081930
Subject:
SGP/ECOR/E14 - Sensor failure
DataStreams:sgp30ecorE14.b1
Description:
The water vapor output of the LI-7500 was too low, indicating that water vapor leaked into 
the sensor and that the chemicals are spent or there was an electronic failure.

Latent heat flux was not normally affected by this problem.
Measurements:sgp30ecorE14.b1:
  • Water vapor density (q)(mean_q)


Back To Table of Contents



END OF DATA