Data Quality Reports for Session: 128407 User: anuskitade Completed: 06/14/2010


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D080307.1SGP/ORG/C1 - Frost on lenssgporgC1.b1
D080321.1SGP/ORG/C1 - Lens heater failuresgporgC1.b1
D080905.3SGP/ORG/C1 - Erroneous precipitation observationssgporgC1.b1
D090223.6SGP/ORG/C1 - False precip measurementssgporgC1.b1


DQRID : D080307.1
Start DateStart TimeEnd DateEnd Time
02/06/2008000002/06/20082359
Subject:
SGP/ORG/C1 - Frost on lens
DataStreams:sgporgC1.b1
Description:
Frost build up on the instrument lenses caused a faulty precipitation measurement.
Measurements:sgporgC1.b1:
  • Precipitation maximum(precip_max)
  • Precipitation standard deviation(precip_sd)
  • Precipitation mean(precip_mean)
  • precipitation minimum(precip_min)


Back To Table of Contents

DQRID : D080321.1
Start DateStart TimeEnd DateEnd Time
01/01/2008000003/10/20081600
Subject:
SGP/ORG/C1 - Lens heater failure
DataStreams:sgporgC1.b1
Description:
Any measurements made in freezing conditions are suspect due to failure of the lens 
heaters.  Users of the data can use the SMOS E13 data to determine if temperatures below 0 
degrees C occurred.
Measurements:sgporgC1.b1:
  • Precipitation maximum(precip_max)
  • Precipitation standard deviation(precip_sd)
  • Precipitation mean(precip_mean)
  • precipitation minimum(precip_min)


Back To Table of Contents

DQRID : D080905.3
Start DateStart TimeEnd DateEnd Time
07/28/2008000007/28/20081200
Subject:
SGP/ORG/C1 - Erroneous precipitation observations
DataStreams:sgporgC1.b1
Description:
Precipitation observed by the optical rain gauge was not observed by the SMOS tipping 
bucket or by the disdrometer nearby.  The optical rain gauge results should be considered 
erroneous.
Measurements:sgporgC1.b1:
  • Precipitation maximum(precip_max)
  • Precipitation standard deviation(precip_sd)
  • Precipitation mean(precip_mean)
  • precipitation minimum(precip_min)


Back To Table of Contents

DQRID : D090223.6
Start DateStart TimeEnd DateEnd Time
01/15/2009000001/15/20092359
01/23/2009000001/23/20092359
01/26/2009000001/26/20092359
01/27/2009000001/27/20092359
Subject:
SGP/ORG/C1 - False precip measurements
DataStreams:sgporgC1.b1
Description:
False measurements of precipitation occurred due to icing or conditions where the dew 
point temperature became close to the air temperature.
Measurements:sgporgC1.b1:
  • Precipitation maximum(precip_max)
  • Precipitation standard deviation(precip_sd)
  • Precipitation mean(precip_mean)
  • precipitation minimum(precip_min)


Back To Table of Contents



END OF DATA