Data Quality Reports for Session: 128352 User: wlj Completed: 06/10/2010


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D040421.3NSA/SKYRAD/C1 - itermittant IRT biasnsaskyrad60sC1.b1
D050219.2TWP/SKYRAD/C2 - IRT biastwpskyrad60sC2.b1
D050504.2TWP/GNDRAD/C2 - IRT connectortwpgndrad60sC2.b1
D060630.34TWP/GNDRAD/C1 - Reprocess: Longwave Calibration errortwpgndrad60sC1.b1
D060630.35TWP/GNDRAD/C2 - Reprocess: Longwave Calibration errortwpgndrad60sC2.b1
D060630.36TWP/GNDRAD/C3 - Reprocess: Longwave Calibration errortwpgndrad60sC3.b1
D060630.39NSA/GNDRAD/C1 - Reprocess: Longwave Calibration errornsagndrad60sC1.b1
D060630.40NSA/GNDRAD/C2 - Reprocess: Longwave Calibration errornsagndrad60sC2.b1
D060630.44PYE/GNDRAD/M1 - Reprocessed: Longwave calibration errorpyegndrad60sM1.b1


DQRID : D040421.3
Start DateStart TimeEnd DateEnd Time
03/28/2003020006/21/20060215
Subject:
NSA/SKYRAD/C1 - itermittant IRT bias
DataStreams:nsaskyrad60sC1.b1
Description:
A problem with the SKYRAD IRT that itermittantly produced a gradual increase in the sky 
temperature measurement then a sudden correction, as compared to AERI, appears to have been 
caused by a combination of a failing IR detector, a poor signal cable, a contaminated 
(with condensation or frost) mirror or lens, and low ambient temperatures. The positive 
bias was apparently not consistant in either frequency or duration but on days when the 
unusual behavior is observed, it usually occurred from about 00:00 to 06:00 local time. 

The instrument was repaired and recalibrated and the signal cable replaced.
Measurements:nsaskyrad60sC1.b1:
  • Sky Infra Red Temperature Minima(sky_ir_temp_min)
  • Standard Deviation of Sky Infra Red Temperature(sky_ir_temp_std)
  • Sky Infra Red Temperature Maxima(sky_ir_temp_max)
  • Sky Infra-Red Temperature(sky_ir_temp)


Back To Table of Contents

DQRID : D050219.2
Start DateStart TimeEnd DateEnd Time
02/07/2005045404/17/20052023
Subject:
TWP/SKYRAD/C2 - IRT bias
DataStreams:twpskyrad60sC2.b1
Description:
The IRT was measuring higher sky temperatures than AERI. The problem was corrected by 
changing the mirror.
Measurements:twpskyrad60sC2.b1:
  • Sky Infra-Red Temperature(sky_ir_temp)
  • Standard Deviation of Sky Infra Red Temperature(sky_ir_temp_std)
  • Sky Infra Red Temperature Maxima(sky_ir_temp_max)
  • Sky Infra Red Temperature Minima(sky_ir_temp_min)


Back To Table of Contents

DQRID : D050504.2
Start DateStart TimeEnd DateEnd Time
03/27/2005233004/03/20050240
Subject:
TWP/GNDRAD/C2 - IRT connector
DataStreams:twpgndrad60sC2.b1
Description:
The IRT had a problem with the connector of either the signal or power cable which 
produced an output of 0 volts equivalent to a measurement of 223K. The problem was corrected 
when the connectors were cleaned and tightened.
Measurements:twpgndrad60sC2.b1:
  • Surface Infra-Red Temperature, Standard Deviation(sfc_ir_temp_std)
  • Effective Surface (Skin) Temperature(sfc_ir_temp)
  • Surface Infra-Red Temperature Minima(sfc_ir_temp_min)
  • Surface Infra-Red Temperature Maxima(sfc_ir_temp_max)


Back To Table of Contents

DQRID : D060630.34
Start DateStart TimeEnd DateEnd Time
08/02/2004070012/21/20052204
Subject:
TWP/GNDRAD/C1 - Reprocess: Longwave Calibration error
DataStreams:twpgndrad60sC1.b1
Description:
Modified pyrgeometer calibration procedures were implemented beginning in December 2003.  
These modified procedures introduced a calibration bias in the longwave data.  The 
previous procedures were re-implemented at all sites between December 2005 and February 2006 to 
restore proper calibrations.                                                             
                The data collected while the incorrect procedures were in place are 
being reprocessing to remove the calibration bias.  Note, this also affected standard 
deviation, maximum and minimum data fields.
Measurements:twpgndrad60sC1.b1:
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Pyrgeometer(up_long_hemisp)


Back To Table of Contents

DQRID : D060630.35
Start DateStart TimeEnd DateEnd Time
07/05/2004220212/21/20052149
Subject:
TWP/GNDRAD/C2 - Reprocess: Longwave Calibration error
DataStreams:twpgndrad60sC2.b1
Description:
Modified pyrgeometer calibration procedures were implemented beginning in December 2003.  
These modified procedures introduced a calibration bias in the longwave data.  The 
previous procedures were re-implemented at all sites between December 2005 and February 2006 to 
restore proper calibrations.                                                             
                The data collected while the incorrect procedures were in place are 
being reprocessing to remove the calibration bias.  Note, this also affected standard 
deviation, maximum and minimum data fields.
Measurements:twpgndrad60sC2.b1:
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Pyrgeometer(up_long_hemisp)


Back To Table of Contents

DQRID : D060630.36
Start DateStart TimeEnd DateEnd Time
05/20/2004010012/21/20052137
Subject:
TWP/GNDRAD/C3 - Reprocess: Longwave Calibration error
DataStreams:twpgndrad60sC3.b1
Description:
Modified pyrgeometer calibration procedures were implemented beginning in December 2003.  
These modified procedures introduced a calibration bias in the longwave data.  The 
previous procedures were re-implemented at all sites between December 2005 and February 2006 to 
restore proper calibrations.                                                             
                The data collected while the incorrect procedures were in place are 
being reprocessing to remove the calibration bias.  Note, this also affected standard 
deviation, maximum and minimum data fields.
Measurements:twpgndrad60sC3.b1:
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Pyrgeometer(up_long_hemisp)


Back To Table of Contents

DQRID : D060630.39
Start DateStart TimeEnd DateEnd Time
11/04/2004160012/21/20051855
Subject:
NSA/GNDRAD/C1 - Reprocess: Longwave Calibration error
DataStreams:nsagndrad60sC1.b1
Description:
Modified pyrgeometer calibration procedures were implemented beginning in December 2003.  
These modified procedures introduced a calibration bias in the longwave data.  The 
previous procedures were re-implemented at all sites between December 2005 and February 2006 to 
restore proper calibrations.                                                             
                The data collected while the incorrect procedures were in place are 
being reprocessing to remove the calibration bias.  Note, this also affected standard 
deviation, maximum and minimum data fields.
Measurements:nsagndrad60sC1.b1:
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Pyrgeometer(up_long_hemisp)


Back To Table of Contents

DQRID : D060630.40
Start DateStart TimeEnd DateEnd Time
11/17/2004003712/22/20052152
Subject:
NSA/GNDRAD/C2 - Reprocess: Longwave Calibration error
DataStreams:nsagndrad60sC2.b1
Description:
Modified pyrgeometer calibration procedures were implemented beginning in December 2003.  
These modified procedures introduced a calibration bias in the longwave data.  The 
previous procedures were re-implemented at all sites between December 2005 and February 2006 to 
restore proper calibrations.                                                             
                The data collected while the incorrect procedures were in place are 
being reprocessing to remove the calibration bias.  Note, this also affected standard 
deviation, maximum and minimum data fields.
Measurements:nsagndrad60sC2.b1:
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Pyrgeometer(up_long_hemisp)


Back To Table of Contents

DQRID : D060630.44
Start DateStart TimeEnd DateEnd Time
02/01/2005000009/14/20050000
Subject:
PYE/GNDRAD/M1 - Reprocessed: Longwave calibration error
DataStreams:pyegndrad60sM1.b1
Description:
Modified pyrgeometer calibration procedures were implemented beginning in December 2003. 
These modified procedures introduced a calibration bias in the longwave data.  The 
previous procedures were re-implemented at all sites between December 2005 and February 2006 to 
restore proper calibrations.

The data collected while the incorrect procedures were in place have been reprocessed to 
remove the calibration bias.  The reprocessed 60 second averaged data are based on 3 
instantaneous 20 second data records rather than on 60 1 second instantaneous data records. 
Still, these data are considered far superior to the originally processed data. The 
reprocessed data were archived in February 2007.
Measurements:pyegndrad60sM1.b1:
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Pyrgeometer(up_long_hemisp)


Back To Table of Contents



END OF DATA