Data Quality Reports for Session: 128092 User: pmshaw Completed: 05/26/2010


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D041027.1NSA/MWRP/C1 - Missing datansamwrpC1.b1
D080409.2NSA/MWRP/C1 - Spare radiometer installednsamwrpC1.00, nsamwrpC1.b1, nsamwrprawC1.00
D080421.1NSA/MWRP/C1 - Radiometer in mixed scanning-zenith modensamwrpC1.b1
D080523.3NSA/MWRP/C1 - Intermittent datansamwrpC1.00, nsamwrpC1.b1, nsamwrprawC1.00
D080523.4NSA/MWRP/C1 - Reprocess: LN2 calibrationnsamwrpC1.b1
D080523.5NSA/MWRP/C1 - Increased RF intereferencensamwrpC1.b1


DQRID : D041027.1
Start DateStart TimeEnd DateEnd Time
02/21/2004092502/23/20042238
10/07/2004085910/27/20041237
11/20/2004003911/22/20041851
04/10/2005000004/11/20051622
09/04/2005053109/06/20051857
09/10/2005005209/12/20051830
09/21/2005180510/12/20052311
06/23/2006180008/08/20062250
08/11/2006235908/14/20061454
10/09/2006210110/11/20061733
01/05/2007235801/08/20072019
02/28/2007230903/05/20070027
07/13/2007183812/08/20070125
01/04/2008235701/06/20081903
02/16/2008181802/21/20080006
02/23/2008100502/26/20082320
03/12/2008014304/09/20081540
04/10/2008184004/14/20081657
04/18/2008225804/21/20081422
04/27/2008184604/29/20080525
05/01/2008032805/28/20082241
10/14/2008165902/13/20092234
Subject:
NSA/MWRP/C1 - Missing data
DataStreams:nsamwrpC1.b1
Description:
Data are missing and unrecoverable.
Measurements:nsamwrpC1.b1:
  • Level of free convection pressure(levelFreeConvectionPres)
  • Derived virtual temperature(virtualTemperature)
  • Lifting condensation level pressure(liftingCondensationLevelPres)
  • Equilibrium level pressure(equilibriumLevelPres)
  • Data quality flags(dataQualityFlags)
  • Expected root-mean-square error in precipitable water retrieval using only
    23.835 and 30.0 GHz(totalPrecipitableWater2RmsError)
  • Surface water vapor density at instrument(surfaceWaterVaporDensity)
  • Derived relative humidity(relativeHumidity)
  • Flag indicating moisture sensor status(wetWindowFlag)
  • Retrieved cloud liquid water content(liquidWaterContent)
  • Retrieved liquid water path using only 23.835 and 30.0 GHz(liquidWaterPath2)
  • Equilibrium level(equilibriumLevel)
  • Zenith-pointing infrared temperature at 10um(infraredTemperature)
  • Convective Available Potential Energy(cape)
  • time(time)
  • Time offset of tweaks from base_time(time_offset)
  • Expected root-mean-square error in liquid water path retrieval(liquidWaterPathRmsError)
  • Expected root-mean-square error in water vapor density retrieval(waterVaporDensityRmsError)
  • atmospheric pressure at mean sea level and at tropopause (NGM250 predicted)(pressure)
  • Internal blackbody reference temperature(blackbodyTemperature)
  • Microwave brightness temperature(brightnessTemperature)
  • Retrieved liquid water path(liquidWaterPath)
  • Interpolated water vapor mixing ratio(waterVaporMixingRatio)
  • Dummy altitude for Zeb(alt)
  • Ambient surface pressure(surfacePressure)
  • Lifting condensation level(liftingCondensationLevel)
  • Array of heights for the range gates(height)
  • Level of free convection(levelFreeConvection)
  • elevation(elevation)
  • Expected root-mean-square error in liquid water content retrieval(liquidWaterContentRmsError)
  • Expected root-mean-square error in liquid water path retrieval using only 23.835
    and 30.0 GHz(liquidWaterPath2RmsError)
  • Ambient surface relative humidity(surfaceRelativeHumidity)
  • Interpolated dewpoint temperature(dewpointTemperature)
  • Ambient surface absolute temperature(surfaceTemperature)
  • Total precipitable water vapor, from microwave radiometer(totalPrecipitableWater)
  • Expected root-mean-square error in temperature retrieval(temperatureRmsError)
  • Frequency(frequency)
  • air temperature (NGM250 predicted)(temperature)
  • base time(base_time)
  • Retrieved water vapor density(waterVaporDensity)
  • Expected root-mean-square error in precipitable water retrieval(totalPrecipitableWaterRmsError)
  • lon(lon)
  • Cloud base height(cloudBaseHeight)
  • lat(lat)
  • Retrieved total precipitable water vapor using only 23.835 and 30.0 GHz(totalPrecipitableWater2)
  • azimuth(azimuth)


Back To Table of Contents

DQRID : D080409.2
Start DateStart TimeEnd DateEnd Time
04/09/2008153905/01/20080300
Subject:
NSA/MWRP/C1 - Spare radiometer installed
DataStreams:nsamwrpC1.00, nsamwrpC1.b1, nsamwrprawC1.00
Description:
A spare radiometer was provided by Radiometrics while they try to fix our MWRP. The spare 
radiometer is of the same family as the one we have in Barrow and it will stay at the 
site until our radiometer is repaired.
A LN2 calibration will be performed next week.
Measurements:nsamwrprawC1.00:
  • Raw data stream - documentation not supported(raw)

nsamwrpC1.b1:
  • Level of free convection pressure(levelFreeConvectionPres)
  • Derived virtual temperature(virtualTemperature)
  • Lifting condensation level pressure(liftingCondensationLevelPres)
  • Equilibrium level pressure(equilibriumLevelPres)
  • Data quality flags(dataQualityFlags)
  • Expected root-mean-square error in precipitable water retrieval using only
    23.835 and 30.0 GHz(totalPrecipitableWater2RmsError)
  • Surface water vapor density at instrument(surfaceWaterVaporDensity)
  • Derived relative humidity(relativeHumidity)
  • Flag indicating moisture sensor status(wetWindowFlag)
  • Retrieved cloud liquid water content(liquidWaterContent)
  • Retrieved liquid water path using only 23.835 and 30.0 GHz(liquidWaterPath2)
  • Equilibrium level(equilibriumLevel)
  • Zenith-pointing infrared temperature at 10um(infraredTemperature)
  • Convective Available Potential Energy(cape)
  • time(time)
  • Time offset of tweaks from base_time(time_offset)
  • Expected root-mean-square error in liquid water path retrieval(liquidWaterPathRmsError)
  • Expected root-mean-square error in water vapor density retrieval(waterVaporDensityRmsError)
  • atmospheric pressure at mean sea level and at tropopause (NGM250 predicted)(pressure)
  • Internal blackbody reference temperature(blackbodyTemperature)
  • Microwave brightness temperature(brightnessTemperature)
  • Retrieved liquid water path(liquidWaterPath)
  • Interpolated water vapor mixing ratio(waterVaporMixingRatio)
  • Dummy altitude for Zeb(alt)
  • Ambient surface pressure(surfacePressure)
  • Lifting condensation level(liftingCondensationLevel)
  • Array of heights for the range gates(height)
  • Level of free convection(levelFreeConvection)
  • elevation(elevation)
  • Expected root-mean-square error in liquid water content retrieval(liquidWaterContentRmsError)
  • Expected root-mean-square error in liquid water path retrieval using only 23.835
    and 30.0 GHz(liquidWaterPath2RmsError)
  • Ambient surface relative humidity(surfaceRelativeHumidity)
  • Interpolated dewpoint temperature(dewpointTemperature)
  • Ambient surface absolute temperature(surfaceTemperature)
  • Total precipitable water vapor, from microwave radiometer(totalPrecipitableWater)
  • Expected root-mean-square error in temperature retrieval(temperatureRmsError)
  • Frequency(frequency)
  • air temperature (NGM250 predicted)(temperature)
  • base time(base_time)
  • Retrieved water vapor density(waterVaporDensity)
  • Expected root-mean-square error in precipitable water retrieval(totalPrecipitableWaterRmsError)
  • lon(lon)
  • Cloud base height(cloudBaseHeight)
  • lat(lat)
  • Retrieved total precipitable water vapor using only 23.835 and 30.0 GHz(totalPrecipitableWater2)
  • azimuth(azimuth)

nsamwrpC1.00:
  • Raw data stream - documentation not supported(raw)


Back To Table of Contents

DQRID : D080421.1
Start DateStart TimeEnd DateEnd Time
04/08/2008000004/21/20081420
Subject:
NSA/MWRP/C1 - Radiometer in mixed scanning-zenith mode
DataStreams:nsamwrpC1.b1
Description:
After installing the spare radiometer (D080409.2), the data started to be intermittent and 
the software froze several times. To test if the problem was due to the scanning 
procedure used with the other radiometer, the instrument was run with a zenith-only procedure at 
times and with the usual scanning procedure at restart. When restarted with the scanning 
procedure the software had problems, so it was decided to run the spare instrument only 
with the zenith-viewing mode until our instrument comes back.

The quality of the data should be unaffected. The only difference is that only data 
collected at 90 degree angle will be available until our instrument comes back.
Measurements:nsamwrpC1.b1:
  • Level of free convection pressure(levelFreeConvectionPres)
  • Derived virtual temperature(virtualTemperature)
  • Lifting condensation level pressure(liftingCondensationLevelPres)
  • Equilibrium level pressure(equilibriumLevelPres)
  • Data quality flags(dataQualityFlags)
  • Expected root-mean-square error in precipitable water retrieval using only
    23.835 and 30.0 GHz(totalPrecipitableWater2RmsError)
  • Surface water vapor density at instrument(surfaceWaterVaporDensity)
  • Derived relative humidity(relativeHumidity)
  • Flag indicating moisture sensor status(wetWindowFlag)
  • Retrieved cloud liquid water content(liquidWaterContent)
  • Retrieved liquid water path using only 23.835 and 30.0 GHz(liquidWaterPath2)
  • Equilibrium level(equilibriumLevel)
  • Zenith-pointing infrared temperature at 10um(infraredTemperature)
  • Convective Available Potential Energy(cape)
  • time(time)
  • Time offset of tweaks from base_time(time_offset)
  • Expected root-mean-square error in liquid water path retrieval(liquidWaterPathRmsError)
  • Expected root-mean-square error in water vapor density retrieval(waterVaporDensityRmsError)
  • atmospheric pressure at mean sea level and at tropopause (NGM250 predicted)(pressure)
  • Internal blackbody reference temperature(blackbodyTemperature)
  • Microwave brightness temperature(brightnessTemperature)
  • Retrieved liquid water path(liquidWaterPath)
  • Interpolated water vapor mixing ratio(waterVaporMixingRatio)
  • Dummy altitude for Zeb(alt)
  • Ambient surface pressure(surfacePressure)
  • Lifting condensation level(liftingCondensationLevel)
  • Array of heights for the range gates(height)
  • Level of free convection(levelFreeConvection)
  • elevation(elevation)
  • Expected root-mean-square error in liquid water content retrieval(liquidWaterContentRmsError)
  • Expected root-mean-square error in liquid water path retrieval using only 23.835
    and 30.0 GHz(liquidWaterPath2RmsError)
  • Ambient surface relative humidity(surfaceRelativeHumidity)
  • Interpolated dewpoint temperature(dewpointTemperature)
  • Ambient surface absolute temperature(surfaceTemperature)
  • Total precipitable water vapor, from microwave radiometer(totalPrecipitableWater)
  • Expected root-mean-square error in temperature retrieval(temperatureRmsError)
  • Frequency(frequency)
  • air temperature (NGM250 predicted)(temperature)
  • base time(base_time)
  • Retrieved water vapor density(waterVaporDensity)
  • Expected root-mean-square error in precipitable water retrieval(totalPrecipitableWaterRmsError)
  • lon(lon)
  • Cloud base height(cloudBaseHeight)
  • lat(lat)
  • Retrieved total precipitable water vapor using only 23.835 and 30.0 GHz(totalPrecipitableWater2)
  • azimuth(azimuth)


Back To Table of Contents

DQRID : D080523.3
Start DateStart TimeEnd DateEnd Time
04/09/2008153905/01/20080300
Subject:
NSA/MWRP/C1 - Intermittent data
DataStreams:nsamwrpC1.00, nsamwrpC1.b1, nsamwrprawC1.00
Description:
A spare radiometer was installed a the site to substitute the malfunctioning radiometer 
(see DQRs D080409.2 and D080421.1). During this time the spare radiometer had problems 
starting and was stopping collection suddenly. Data during this time are very intermittent 
with large gaps.
Measurements:nsamwrprawC1.00:
  • Raw data stream - documentation not supported(raw)

nsamwrpC1.b1:
  • Level of free convection pressure(levelFreeConvectionPres)
  • Derived virtual temperature(virtualTemperature)
  • Lifting condensation level pressure(liftingCondensationLevelPres)
  • Equilibrium level pressure(equilibriumLevelPres)
  • Data quality flags(dataQualityFlags)
  • Expected root-mean-square error in precipitable water retrieval using only
    23.835 and 30.0 GHz(totalPrecipitableWater2RmsError)
  • Surface water vapor density at instrument(surfaceWaterVaporDensity)
  • Derived relative humidity(relativeHumidity)
  • Flag indicating moisture sensor status(wetWindowFlag)
  • Retrieved cloud liquid water content(liquidWaterContent)
  • Retrieved liquid water path using only 23.835 and 30.0 GHz(liquidWaterPath2)
  • Equilibrium level(equilibriumLevel)
  • Zenith-pointing infrared temperature at 10um(infraredTemperature)
  • Convective Available Potential Energy(cape)
  • time(time)
  • Time offset of tweaks from base_time(time_offset)
  • Expected root-mean-square error in liquid water path retrieval(liquidWaterPathRmsError)
  • Expected root-mean-square error in water vapor density retrieval(waterVaporDensityRmsError)
  • atmospheric pressure at mean sea level and at tropopause (NGM250 predicted)(pressure)
  • Internal blackbody reference temperature(blackbodyTemperature)
  • Microwave brightness temperature(brightnessTemperature)
  • Retrieved liquid water path(liquidWaterPath)
  • Interpolated water vapor mixing ratio(waterVaporMixingRatio)
  • Dummy altitude for Zeb(alt)
  • Ambient surface pressure(surfacePressure)
  • Lifting condensation level(liftingCondensationLevel)
  • Array of heights for the range gates(height)
  • Level of free convection(levelFreeConvection)
  • elevation(elevation)
  • Expected root-mean-square error in liquid water content retrieval(liquidWaterContentRmsError)
  • Expected root-mean-square error in liquid water path retrieval using only 23.835
    and 30.0 GHz(liquidWaterPath2RmsError)
  • Ambient surface relative humidity(surfaceRelativeHumidity)
  • Interpolated dewpoint temperature(dewpointTemperature)
  • Ambient surface absolute temperature(surfaceTemperature)
  • Total precipitable water vapor, from microwave radiometer(totalPrecipitableWater)
  • Expected root-mean-square error in temperature retrieval(temperatureRmsError)
  • Frequency(frequency)
  • air temperature (NGM250 predicted)(temperature)
  • base time(base_time)
  • Retrieved water vapor density(waterVaporDensity)
  • Expected root-mean-square error in precipitable water retrieval(totalPrecipitableWaterRmsError)
  • lon(lon)
  • Cloud base height(cloudBaseHeight)
  • lat(lat)
  • Retrieved total precipitable water vapor using only 23.835 and 30.0 GHz(totalPrecipitableWater2)
  • azimuth(azimuth)

nsamwrpC1.00:
  • Raw data stream - documentation not supported(raw)


Back To Table of Contents

DQRID : D080523.4
Start DateStart TimeEnd DateEnd Time
04/09/2008153905/01/20080300
Subject:
NSA/MWRP/C1 - Reprocess: LN2 calibration
DataStreams:nsamwrpC1.b1
Description:
On 4/9 a spare radiometer was installed at the site. On 4/18 a LN2 calibration was 
conducted. Upon examination of the data it appeared that the LN2 calibration was generally 
successful, however the 30-GHz channel was not calibrated properly.
New calibration coefficients for all K-band channels were derived and data for the whole 
month should be reprocessed with these newly derived coefficients. I will provide the new 
configuration file.

An explanation of the effects of the poor calibration is given in the Instrument Mentor 
Monthly Summary report. The poor calibration affected the LWP in particular. Re-calibration 
will correct the problem.
Measurements:nsamwrpC1.b1:
  • Total precipitable water vapor, from microwave radiometer(totalPrecipitableWater)
  • Derived relative humidity(relativeHumidity)
  • Retrieved cloud liquid water content(liquidWaterContent)
  • Microwave brightness temperature(brightnessTemperature)
  • Retrieved liquid water path(liquidWaterPath)
  • Interpolated water vapor mixing ratio(waterVaporMixingRatio)
  • Retrieved liquid water path using only 23.835 and 30.0 GHz(liquidWaterPath2)
  • Retrieved total precipitable water vapor using only 23.835 and 30.0 GHz(totalPrecipitableWater2)


Back To Table of Contents

DQRID : D080523.5
Start DateStart TimeEnd DateEnd Time
04/09/2008153905/01/20080300
Subject:
NSA/MWRP/C1 - Increased RF intereference
DataStreams:nsamwrpC1.b1
Description:
The spare radiometer experienced increased levels of RF interference. The interference is 
prominent on the 30-GHz channel and LWP retrievals that appear particularly noisy.
Measurements:nsamwrpC1.b1:
  • Microwave brightness temperature(brightnessTemperature)
  • Retrieved liquid water path(liquidWaterPath)
  • Retrieved liquid water path using only 23.835 and 30.0 GHz(liquidWaterPath2)


Back To Table of Contents



END OF DATA