Data Quality Reports for Session: 128041 User: pmshaw Completed: 05/24/2010


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D080418.1NSA/SKYRAD/C1 - Direct and Diffuse incorrectnsaskyrad20sC1.a0, nsaskyrad60sC1.b1


DQRID : D080418.1
Start DateStart TimeEnd DateEnd Time
04/17/2008063204/23/20081815
Subject:
NSA/SKYRAD/C1 - Direct and Diffuse incorrect
DataStreams:nsaskyrad20sC1.a0, nsaskyrad60sC1.b1
Description:
The short_direct_normal (NIP) instrument and cable were damaged by tracker rewind 
malfunction. Tracker and NIP replaced on 4/22 2200GMT but tracker not operational until 4/23 
1815GMT.  Thus, down_short_diffuse_hemisp (8-48) was not shaded 4/22 to 4/23.

Do not use 8-48 data from 4/22 to 4/23. NIP data is not recoverable during entire time but 
can be estimated using component sum for each 1 minute data: Direct = (global - 
diffuse)/cos(zenith angle)
Measurements:nsaskyrad60sC1.b1:
  • Shortwave Direct Normal Irradiance, Pyrgeometer(short_direct_normal)
  • Downwelling Shortwave Diffuse Hemispheric Irradiance, Ventilated Pyranometer(down_short_diffuse_hemisp)
  • Downwelling Shortwave Diffuse Hemispheric Irradiance, Ventilated Pyranometer,
    Maxima(down_short_diffuse_hemisp_max)
  • Shortwave Direct Normal Irradiance, Pyrgeometer, Standard Deviation(short_direct_normal_std)
  • Downwelling Shortwave Hemispheric Irradiance, Pyranometer, Standard Deviation(down_short_diffuse_hemisp_std)
  • Shortwave Direct Normal Irradiance, Pyrgeometer, Minima(short_direct_normal_min)
  • Shortwave Direct Normal Irradiance, Pyrgeometer, Maxima(short_direct_normal_max)
  • Downwelling Shortwave Diffuse Hemispheric Irradiance, Ventilated Pyranometer,
    Minima(down_short_diffuse_hemisp_min)

nsaskyrad20sC1.a0:
  • Instantaneous Uncorrected Downwelling Shortwave Diffuse, Shaded Pyranometer
    Thermopile Voltage(inst_diffuse)
  • Instantaneous Direct Normal Shortwave Irradiance, Pyheliometer Thermopile
    Voltage(inst_direct_normal)


Back To Table of Contents



END OF DATA