D040127.3TWP/SONDE/C2 - Broken temperature sensorstwpsondewnpnC2.b1
D040815.1TWP/SONDE/C2 - Operator used wrong calibration tapetwpsondewnpnC2.b1


DQRID : D040127.3
Start DateStart TimeEnd DateEnd Time
10/13/2003004010/13/20030203
11/08/2003233511/09/20030215
11/18/2003120511/18/20031340
11/19/2003115611/19/20031346
11/21/2003004411/21/20030113
12/20/2003122512/20/20031408
12/21/2003005612/21/20030126
12/21/2003115512/21/20031344
12/23/2003004212/23/20030212
12/23/2003113712/23/20031259
12/23/2003234212/24/20030132
01/08/2004113501/08/20041309
01/09/2004115701/09/20041342
01/09/2004135901/09/20041537
01/12/2004234201/13/20040214
01/16/2004140901/16/20041548
01/16/2004233701/17/20040109
01/18/2004113501/18/20041256
01/19/2004022301/19/20040346
01/19/2004232901/20/20040157
01/21/2004125201/21/20041400
01/22/2004005201/22/20040227
01/23/2004113701/23/20041319
01/26/2004120201/26/20040146
02/17/2004140602/17/20041536
02/26/2004122102/26/20041350
03/09/2004114503/09/20041315
03/14/2004113103/14/20041300
03/22/2004113403/22/20041300
03/29/2004233203/30/20040047
03/30/2004233403/31/20040059
04/11/2004234804/12/20040120
05/01/2004233505/02/20040110
05/02/2004233605/03/20040109
05/13/2004233405/14/20040112
05/14/2004140605/14/20041411
06/07/2004022406/07/20040351
06/25/2004114606/25/20041328
07/02/2004005907/02/20040217
07/20/2004022007/20/20040358
07/24/2004123507/24/20041355
08/22/2004114408/22/20041207
09/08/2004142509/08/20041605
09/08/2004233809/08/20042345
10/18/2004234610/19/20040128
11/29/2004232111/30/20040105
12/08/2004114312/08/20041306
12/30/2004113012/30/20041239
Subject:
TWP/SONDE/C2 - Broken temperature sensors
DataStreams:twpsondewnpnC2.b1
Description:
The temperature sensors on these radiosondes broke after launch.  All
temperature-related data are incorrect.
Measurements:twpsondewnpnC2.b1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)
  • Dry bulb temperature(tdry)


Back To Table of Contents

DQRID : D040815.1
Start DateStart TimeEnd DateEnd Time
08/12/2004004408/12/20040233
Subject:
TWP/SONDE/C2 - Operator used wrong calibration tape
DataStreams:twpsondewnpnC2.b1
Description:
It appears as though the opperator used the wrong calibration tape for this sounding.  The 
primary symptom is a jump in temperature from a surface value of 30.1 degC to the first 
value aloft of 61.0 degC.  Secondly, the recorded serial number for this sounding is 
Z1430179 which is the same serial number recorded for the previous sounding.

This problem points to three failures of procedure (I have notified TWP site ops by 
e-mail).  The first failure is using the wrong tape. The second failure is not checking the 
radiosonde values before launch which relates to the third failure of not entering the 
radiosonde values as the surface values of the sounding.
Measurements:twpsondewnpnC2.b1:
  • lon(lon)
  • U-component(u_wind)
  • V-component(v_wind)
  • Ascent Rate(asc)
  • Relative humidity inside the instrument enclosure(rh)
  • Time offset of tweaks from base_time(time_offset)
  • base time(base_time)
  • Surface dew point temperature(dp)
  • Wind Status(wstat)
  • Wind Direction(deg)
  • Dummy altitude for Zeb(alt)
  • Mean Wind Speed(wspd)
  • Retrieved pressure profile(pres)
  • lat(lat)
  • Dry bulb temperature(tdry)


Back To Table of Contents