Data Quality Reports for Session: 127592 User: martin1ste Completed: 05/07/2010


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D081216.1TWP/MWR/C2 - Incorrect ambient temperature during raintwpmwrlosC2.b1, twpmwrtipC2.a1


DQRID : D081216.1
Start DateStart TimeEnd DateEnd Time
10/09/2008010001/09/20090300
Subject:
TWP/MWR/C2 - Incorrect ambient temperature during rain
DataStreams:twpmwrlosC2.b1, twpmwrtipC2.a1
Description:
The ambient temperature was approximately 15 degree higher than what it should be when it 
was raining and the heater came. On Jan 9 2009 the dewblower was replaced which solved 
the problem.

The increased ambient temperature (tkair) was happening during times of rain and users 
should always disregard MWR data during rain events.  Therefore this DQR is colored 
transparent.
Measurements:twpmwrtipC2.a1:
  • Ambient temperature(tkair)

twpmwrlosC2.b1:
  • Ambient temperature(tkair)


Back To Table of Contents



END OF DATA