Data Quality Reports for Session: 126636 User: margaret1l Completed: 03/07/2010


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D021211.1TWP/SMET/C3 - Lower wind direction sensor misaligned twpsmet60sC3.a1, twpsmet60sC3.b1
D030212.1TWP/SMET/C3 - Intermittent Pressure Sensor Failuretwpsmet60sC3.a1, twpsmet60sC3.b1
D040726.3TWP/SMET/C3 - Negative rainfall ratestwpsmet60sC3.b1
D061011.4TWP/SMET/C3 - Reprocess: Tipping bucket rain gauge addedtwpsmet60sC3.b1
D081022.8TWP/SMET/C3 - Wind Speed data contain offset due to induced voltagetwpsmet60sC3.b1


DQRID : D021211.1
Start DateStart TimeEnd DateEnd Time
08/21/2002000012/09/20020000
Subject:
TWP/SMET/C3 - Lower wind direction sensor misaligned
DataStreams:twpsmet60sC3.a1, twpsmet60sC3.b1
Description:
The lower wind direction sensor (sensor #2) was misaligned and off by 15 degrees.
Measurements:twpsmet60sC3.a1:
  • Wind #2 direction standard deviation(wind2_dir_sd)
  • Wind #2 direction vector average(wind2_dir_vec_avg)

twpsmet60sC3.b1:
  • Lower sensor, wind direction standard deviation(lo_dir_sd)
  • Lower sensor, wind direction vector average(lo_dir_vec_avg)


Back To Table of Contents

DQRID : D030212.1
Start DateStart TimeEnd DateEnd Time
10/23/2002000002/06/20032203
Subject:
TWP/SMET/C3 - Intermittent Pressure Sensor Failure
DataStreams:twpsmet60sC3.a1, twpsmet60sC3.b1
Description:
The Barometric Pressure sensor was intermittently failing at TWP site C3 Darwin.  At times 
there were large pressure fluctuations occurring on the order of 24 mb in a few minutes. 
 Times of large pressure fluctuations are readily apparent in the data.  At these times 
the data should be carefully analyzed before determining it's usefulness.  In between the 
periods of large fluctuations the readings appear to be stable.  It is still suggested 
that independent pressure data be obtained for comparison before use.
Measurements:twpsmet60sC3.a1:
  • Barometric Pressure(atmos_pressure)

twpsmet60sC3.b1:
  • Barometric Pressure(atmos_pressure)


Back To Table of Contents

DQRID : D040726.3
Start DateStart TimeEnd DateEnd Time
03/12/2002062805/17/20042300
Subject:
TWP/SMET/C3 - Negative rainfall rates
DataStreams:twpsmet60sC3.b1
Description:
Due to limitations of the Zeno dataloggers all voltage measurements
from the optical rain gauge (ORG) were converted to rainfall rates. 
This resulted in negative values for rain rates.  Any value below
0.10 mm/hr should be considered to be 0 mm/hr.  New dataloggers 
alleviated this problem.
Measurements:twpsmet60sC3.b1:
  • Precipitation standard deviation(precip_sd)
  • Precipitation maximum(precip_max)
  • precipitation minimum(precip_min)
  • Precipitation mean(precip_mean)


Back To Table of Contents

DQRID : D061011.4
Start DateStart TimeEnd DateEnd Time
03/12/2002000009/26/20061435
Subject:
TWP/SMET/C3 - Reprocess: Tipping bucket rain gauge added
DataStreams:twpsmet60sC3.b1
Description:
A tipping bucket rain gauge was added to the SMET suite of instruments effective 20060926. 
 Two new variables (count and rain amount) were added at the end of the twpsmet60sC1.b1 
datastream.  When reprocessing of this datastream occurs these variables will be added 
with -9999 (missing) as the value for dates prior to gauge installation.
Measurements:twpsmet60sC3.b1:
  • base time(base_time)


Back To Table of Contents

DQRID : D081022.8
Start DateStart TimeEnd DateEnd Time
03/12/2002000010/17/20080627
Subject:
TWP/SMET/C3 - Wind Speed data contain offset due to induced voltage
DataStreams:twpsmet60sC3.b1
Description:
The marine model of the RM Young wind sensor needs twisted pair shielded cable.  The 
manufacturer did not originally supply the pigtail with twisted pair shielded cable.  Use of 
twisted pair shielded cable for the rest of the cable run to the logger was also not 
specified.  This caused an induced voltage in the wind speed line from the wind direction 
excitation.  The induced voltage varied from sensor to sensor and probably changed anytime 
the cable was replaced or parts of the wind sensor were repaired or replaced. The induced 
voltage has been seen to vary from 0.15 m/s to as much as 1.08 m/s.
Measurements:twpsmet60sC3.b1:
  • Upper sensor, wind speed vector average(up_spd_vec_avg)
  • Lower sensor, wind speed vector average(lo_spd_vec_avg)
  • Lower sensor, wind speed standard deviation(lo_spd_sd)
  • Lower sensor, wind speed minimum(lo_spd_min)
  • Upper sensor, wind speed maximum(up_spd_max)
  • Upper sensor, wind speed minimum(up_spd_min)
  • Upper sensor, wind speed standard deviation(up_spd_sd)
  • Lower sensor, wind speed arithmetic average(lo_spd_arith_avg)
  • Upper sensor, wind speed arithmetic average(up_spd_arith_avg)
  • Lower sensor, wind speed maximum(lo_spd_max)


Back To Table of Contents