Data Quality Reports for Session: 126386 User: hallsd Completed: 02/25/2010


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D060928.4NSA/SONDE/C1 - Increased precision for alt, dp, pres, rh, tdrynsasondewnpnC1.b1
D070302.1NSA/SONDE/C1 - Temperature sensor broke at launchnsasondewnpnC1.b1


DQRID : D060928.4
Start DateStart TimeEnd DateEnd Time
09/29/2006050012/31/99992359
Subject:
NSA/SONDE/C1 - Increased precision for alt, dp, pres, rh, tdry
DataStreams:nsasondewnpnC1.b1
Description:
The output precision of the following variables was increased 20060929:

   alt	changed from 1m to 0.1m
   dp	changed from 0.1degC to 0.01degC
   pres changed from 0.1 hPa to 0.01 hPa
   RH	changed from 1%RH to 0.01%RH
   tdry changed from .1degC to 0.01degC

Note that the increase in precision does not necessarily imply an increase
in accuracy.  Users are cautioned to continue to use their best scientific
judgement when using these or any other data.
Measurements:nsasondewnpnC1.b1:
  • Dummy altitude for Zeb(alt)
  • Dry bulb temperature(tdry)
  • Surface dew point temperature(dp)
  • Retrieved pressure profile(pres)
  • Relative humidity inside the instrument enclosure(rh)


Back To Table of Contents

DQRID : D070302.1
Start DateStart TimeEnd DateEnd Time
02/27/2007195502/27/20072131
Subject:
NSA/SONDE/C1 - Temperature sensor broke at launch
DataStreams:nsasondewnpnC1.b1
Description:
The temperature sensor on this sonde broke at launch.  As a result all the thermodynamic 
data are incorrect.  The wind data are OK, though there will be some uncertainty in the 
altitude values.  This sounding was done at a non-standard time and it isn't clear whether 
it was delayed for some reason or a RHUBC sounding.  In either case, there is no fix.
Measurements:nsasondewnpnC1.b1:
  • Dummy altitude for Zeb(alt)
  • Dry bulb temperature(tdry)
  • Surface dew point temperature(dp)
  • Relative humidity inside the instrument enclosure(rh)
  • Ascent Rate(asc)


Back To Table of Contents



END OF DATA