Data Quality Reports for Session: 125978 User: perez Completed: 02/08/2010


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D040904.2TWP/SKYRAD/C3 - Incorrect NIP cal coefficienttwpskyrad60sC3.b1
D050104.1TWP/SKYRAD/C3 - Reprocess: NIP calibrations incorrecttwpskyrad60sC3.b1, twpskyradC3.00
D051114.3TWP/SKYRAD/C3 - tracker inoperativetwpskyrad20sC3.a0, twpskyrad60sC3.b1, twpskyradC3.00
D060110.3TWP/SKYRAD/C3 - IRT damagedtwpskyrad20sC3.a0, twpskyrad60sC3.b1


DQRID : D040904.2
Start DateStart TimeEnd DateEnd Time
05/24/2004000009/02/20042350
Subject:
TWP/SKYRAD/C3 - Incorrect NIP cal coefficient
DataStreams:twpskyrad60sC3.b1
Description:
During our site visit comparison at Darwin we observed that the skyrad NIP was reading 
about 50 w/m2 lower than the comparison cavity radfiometer and the replacement NIP on the 
comparison stand. We found that the calibration coefficient for the NIP (responsivity at 
45o)was put in as 8.64X10-6. The NIP was new and had not been calibrated at SGP. The 
factory number was 8.04X10-6. We corrected the calibration coefficient and the instruments 
agreed to better than 5 w/m2.
Measurements:twpskyrad60sC3.b1:
  • Shortwave Direct Normal Irradiance, Pyrgeometer, Minima(short_direct_normal_min)
  • Shortwave Direct Normal Irradiance, Pyrgeometer(short_direct_normal)
  • Shortwave Direct Normal Irradiance, Pyrgeometer, Maxima(short_direct_normal_max)
  • Shortwave Direct Normal Irradiance, Pyrgeometer, Standard Deviation(short_direct_normal_std)


Back To Table of Contents

DQRID : D050104.1
Start DateStart TimeEnd DateEnd Time
05/20/2004000009/02/20040400
Subject:
TWP/SKYRAD/C3 - Reprocess: NIP calibrations incorrect
DataStreams:twpskyrad60sC3.b1, twpskyradC3.00
Description:
On 05/20 during installtion of the new Campbell datalogger system and associated program 
the NIP (S/N 33551E6) was assigned an incorrect calibration coefficient.  A coefficient of 
116.01 W/m2/mV was applied when it should have been 124.378 W/m2/mV.
TWP reports SV0405D and SV0409D detail the activities which lead to this DQR.
Measurements:twpskyradC3.00:
  • null(Raw data stream - documentation not supported)

twpskyrad60sC3.b1:
  • Shortwave Direct Normal Irradiance, Pyrgeometer, Minima(short_direct_normal_min)
  • Shortwave Direct Normal Irradiance, Pyrgeometer(short_direct_normal)
  • Shortwave Direct Normal Irradiance, Pyrgeometer, Maxima(short_direct_normal_max)
  • Shortwave Direct Normal Irradiance, Pyrgeometer, Standard Deviation(short_direct_normal_std)


Back To Table of Contents

DQRID : D051114.3
Start DateStart TimeEnd DateEnd Time
11/14/2005045312/02/20050000
Subject:
TWP/SKYRAD/C3 - tracker inoperative
DataStreams:twpskyrad20sC3.a0, twpskyrad60sC3.b1, twpskyradC3.00
Description:
A lightning strike at Darwin removed the tracker from service.  Instruments installed on 
the tracker were rendered either inoperative or were not shaded during this time period.  
The longwave hemispheric instruments PIR1 and PIR2 appear to have been unaffected.  They 
are not suseptable to incorrect tracker operation.  The downwelling hemispheric shortwave 
PSP instrument mounted on Skyrad will not be affected by incorrect tracker operation.
Measurements:twpskyradC3.00:
  • null(Raw data stream - documentation not supported)

twpskyrad60sC3.b1:
  • Shortwave Direct Normal Irradiance, Pyrgeometer, Minima(short_direct_normal_min)
  • Downwelling Shortwave Diffuse Hemispheric Irradiance, Ventilated Pyranometer,
    Maxima(down_short_diffuse_hemisp_max)
  • Shortwave Direct Normal Irradiance, Pyrgeometer(short_direct_normal)
  • Downwelling Shortwave Diffuse Hemispheric Irradiance, Ventilated Pyranometer,
    Minima(down_short_diffuse_hemisp_min)
  • Downwelling Shortwave Hemispheric Irradiance, Pyranometer, Standard Deviation(down_short_diffuse_hemisp_std)
  • Shortwave Direct Normal Irradiance, Pyrgeometer, Maxima(short_direct_normal_max)
  • Shortwave Direct Normal Irradiance, Pyrgeometer, Standard Deviation(short_direct_normal_std)
  • Downwelling Shortwave Diffuse Hemispheric Irradiance, Ventilated Pyranometer(down_short_diffuse_hemisp)

twpskyrad20sC3.a0:
  • Instantaneous Direct Normal Shortwave Irradiance, Pyheliometer Thermopile
    Voltage(inst_direct_normal)
  • Instantaneous Uncorrected Downwelling Shortwave Diffuse, Shaded Pyranometer
    Thermopile Voltage(inst_diffuse)


Back To Table of Contents

DQRID : D060110.3
Start DateStart TimeEnd DateEnd Time
11/13/2005231403/07/20060257
Subject:
TWP/SKYRAD/C3 - IRT damaged
DataStreams:twpskyrad20sC3.a0, twpskyrad60sC3.b1
Description:
The analog output of the IRT was damaged by a lightning strike. The problem was corrected 
by replacing the instrument with a spare. The damaged IRT was returned to the vendor for 
repair.
Measurements:twpskyrad60sC3.b1:
  • Sky Infra Red Temperature Minima(sky_ir_temp_min)
  • Sky Infra-Red Temperature(sky_ir_temp)
  • Standard Deviation of Sky Infra Red Temperature(sky_ir_temp_std)

twpskyrad20sC3.a0:
  • Instantaneous Sky/Cloud Infrared(inst_sky_ir_signal)


Back To Table of Contents



END OF DATA