Data Quality Reports for Session: 125335 User: kboyd Completed: 01/20/2010


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D050614.1SGP/VCEIL/C1 - poor signalsgpvceil25kC1.b1
D051011.6SGP/MWR/C1 - New software version (4.15) installedsgpmwrlosC1.b1, sgpmwrtipC1.a1


DQRID : D050614.1
Start DateStart TimeEnd DateEnd Time
05/05/2005184204/19/20061713
Subject:
SGP/VCEIL/C1 - poor signal
DataStreams:sgpvceil25kC1.b1
Description:
When parameters were changed (set noise_h2_compensation to "ON" and units to "METERS") to 
correct a low amplitude problem, the backscatter plots began displaying horizontal 
stripes indicating detector "ringing". However, the cloud base heights are not affected. The 
problem was corrected when the instrument was replaced.
Measurements:sgpvceil25kC1.b1:
  • Detection status. See details(detection_status)
  • Aerosol backscatter coefficient at 355 nm(backscatter)


Back To Table of Contents

DQRID : D051011.6
Start DateStart TimeEnd DateEnd Time
07/31/2002202708/04/20051959
Subject:
SGP/MWR/C1 - New software version (4.15) installed
DataStreams:sgpmwrlosC1.b1, sgpmwrtipC1.a1
Description:
A problem began with the installation of MWR.EXE version 4.12 in July 2002. The software 
had been upgraded from a "DOS" to a "Windows"-compiled program to address an earlier 
problem.  The software upgrade corrected the earlier problem but introduced a new one that 
caused line-of-sight observing cycles to be skipped, a 15% reduction in the number of tip 
curves, and saturation of CPU usage. Software versions 4.13 and 4.14 also produced these 
problems.

The new MWR software, version 4.15, was installed on 08/04/2005. As a consequence of this 
upgrade, the tip curve frequency increased. The tip cycle time decreased from ~60s to 
~50s.
Measurements:sgpmwrtipC1.a1:
  • 31.4 GHz blac2body+noise injection signal(bbn31)
  • Noise injection temp at 23.8 GHz adjusted to tkbb(tnd23)
  • 31.4 GHz blackbody(bb31)
  • Noise injection temp at 23.8 GHz derived from this tip(tnd23I)
  • (tknd)
  • 31.4 GHz sky signal(tipsky31)
  • Noise injection temp at 31.4 GHz derived from this tip(tnd31I)
  • Total liquid water along zenith path using tip-derived brightness temperatures(liqtip)
  • Temperature correction coefficient at 31.4 GHz(tc31)
  • Noise injection temp at 31.4 GHz adjusted to tkbb(tnd31)
  • 23.8 GHz Blackbody signal(bb23)
  • Blackbody kinetic temperature(tkbb)
  • Noise injection temp at nominal temperature at 31.4 GHz(tnd_nom31)
  • 23.8 GHz sky signal(tipsky23)
  • Ambient temperature(tkair)
  • Temperature correction coefficient at 23.8 GHz(tc23)
  • Total water vapor along zenith path using tip-derived brightness temperatures(vaptip)
  • Mixer kinetic (physical) temperature(tkxc)
  • Noise injection temp at nominal temperature at 23.8 GHz(tnd_nom23)
  • 23.8 GHz blackbody+noise injection signal(bbn23)
  • 31.4 GHz sky brightness temperature derived from tip curve(tbsky31tip)
  • 31.4 GHz goodness-of-fit coefficient(r31)
  • 23.8 GHz sky brightness temperature derived from tip curve(tbsky23tip)
  • 23.8 GHz goodness-of-fit coefficient(r23)

sgpmwrlosC1.b1:
  • 31.4 GHz sky signal(sky31)
  • (tknd)
  • 31.4 GHz blac2body+noise injection signal(bbn31)
  • Noise injection temp at nominal temperature at 31.4 GHz(tnd_nom31)
  • Noise injection temp at 23.8 GHz adjusted to tkbb(tnd23)
  • 23.8 GHz sky signal(sky23)
  • Sky brightness temperature at 31.4 GHz(tbsky31)
  • 23.8 GHz blackbody+noise injection signal(bbn23)
  • Noise injection temp at 31.4 GHz adjusted to tkbb(tnd31)
  • Averaged total liquid water along LOS path(liq)
  • Noise injection temp at nominal temperature at 23.8 GHz(tnd_nom23)
  • 31.4 GHz blackbody(bb31)
  • Temperature correction coefficient at 31.4 GHz(tc31)
  • Blackbody kinetic temperature(tkbb)
  • Sky Infra-Red Temperature(sky_ir_temp)
  • MWR column precipitable water vapor(vap)
  • Ambient temperature(tkair)
  • 23.8 GHz Blackbody signal(bb23)
  • Sky brightness temperature at 23.8 GHz(tbsky23)
  • Mixer kinetic (physical) temperature(tkxc)
  • Temperature correction coefficient at 23.8 GHz(tc23)