Data Quality Reports for Session: 124225 User: wlj Completed: 11/10/2009


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D050104.1TWP/SKYRAD/C3 - Reprocess: NIP calibrations incorrecttwpskyrad60sC3.b1, twpskyradC3.00
D060120.1TWP/SKYRAD/C1 - IRT serial number incorrecttwpskyradC1.00


DQRID : D050104.1
Start DateStart TimeEnd DateEnd Time
05/20/2004000009/02/20040400
Subject:
TWP/SKYRAD/C3 - Reprocess: NIP calibrations incorrect
DataStreams:twpskyrad60sC3.b1, twpskyradC3.00
Description:
On 05/20 during installtion of the new Campbell datalogger system and associated program 
the NIP (S/N 33551E6) was assigned an incorrect calibration coefficient.  A coefficient of 
116.01 W/m2/mV was applied when it should have been 124.378 W/m2/mV.
TWP reports SV0405D and SV0409D detail the activities which lead to this DQR.
Measurements:twpskyradC3.00:
  • null(Raw data stream - documentation not supported)

twpskyrad60sC3.b1:
  • Shortwave Direct Normal Irradiance, Pyrgeometer, Minima(short_direct_normal_min)
  • Shortwave Direct Normal Irradiance, Pyrgeometer(short_direct_normal)
  • Shortwave Direct Normal Irradiance, Pyrgeometer, Maxima(short_direct_normal_max)
  • Shortwave Direct Normal Irradiance, Pyrgeometer, Standard Deviation(short_direct_normal_std)


Back To Table of Contents

DQRID : D060120.1
Start DateStart TimeEnd DateEnd Time
08/02/2004234001/18/20062300
Subject:
TWP/SKYRAD/C1 - IRT serial number incorrect
DataStreams:twpskyradC1.00
Description:
The serial number was incorrect in the csi program for the skyrad. It was entered as 868 
when the correct number was 866 until 10 jan 2006 at 0145. It then should have been 1845. 
This does not affect data other than the once per day serial number and calibration field.
Measurements:twpskyradC1.00:
  • null(Raw data stream - documentation not supported)


Back To Table of Contents



END OF DATA