Data Quality Reports for Session: 123962 User: kboyd Completed: 10/27/2009


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D040318.10SGP/SMOS/E13 - 6-month Calibration Checkssgp1440smosE13.b1, sgp1smosE13.b1, sgp30smosE13.b1
D070314.2SGP/MWR/C1/E14 - Freezing rain-Incorrect rain flagsgpmwrlosC1.b1, sgpmwrlosE14.b1, sgpmwrtipC1.a1, sgpmwrtipE14.a1
D070406.2SGP/MWR/E14 - Power supply failuresgpmwrlosE14.b1, sgpmwrtipE14.a1
D070420.1SGP/AOS/C1 - Humidifier controller failuresgpaosC1.00, sgpaosC1.a0, sgpaosauxC1.a0
D070613.2SGP/RAIN/C1 - Funnel cloggedsgprainC1.b1
D070618.2SGP/SMOS/E13 - RH sensor failuresgp1440smosE13.b1, sgp1smosE13.b1, sgp30smosE13.b1
D070905.2SGP/AOS/C1 - CPC failuresgpaosC1.a1
D080116.13SGP/SMOS/E13 - Wind Data Incorrect Due to Icingsgp1440smosE13.b1, sgp1smosE13.b1, sgp30smosE13.b1
D080116.24SGP/THWAPS/C1 - RH Data IncorrectsgpthwapsC1.b1
D080128.3SGP/AOS/C1/ - Pump failuresgpaosccnC1.a1
D080222.4SGP/SMOS - Snowfall caused delayed count of precipsgp1smosE1.b1, sgp1smosE11.b1, sgp1smosE13.b1, sgp1smosE15.b1, sgp1smosE20.b1,
sgp1smosE21.b1, sgp1smosE24.b1, sgp1smosE27.b1, sgp1smosE3.b1, sgp1smosE4.b1, sgp1smosE5.b1,
sgp1smosE6.b1, sgp1smosE7.b1, sgp1smosE8.b1, sgp1smosE9.b1, sgp30smosE1.b1,
sgp30smosE11.b1, sgp30smosE13.b1, sgp30smosE15.b1, sgp30smosE20.b1, sgp30smosE21.b1, sgp30smosE24.b1,
sgp30smosE27.b1, sgp30smosE3.b1, sgp30smosE4.b1, sgp30smosE5.b1, sgp30smosE6.b1,
sgp30smosE7.b1, sgp30smosE8.b1, sgp30smosE9.b1
D080403.2SGP/WACR/C1 - Contamination with AMF WACR datasgpwacrC1.00, sgpwacrC1.b1, sgpwacrspecmomC1.00
D080820.9SGP/SMOS/E13 - Missing Datasgp1440smosE13.b1, sgp1smosE13.b1, sgp30smosE13.b1
D081015.1SGP/SMOS/E13 - Rain Gauge Inoperativesgp1smosE13.b1, sgp30smosE13.b1
D081212.2SGP/SMOS/E13 - SMOS T/RH probe solar shield missingsgp1440smosE13.b1, sgp1smosE13.b1, sgp30smosE13.b1
D090120.5SGP/AOS/C1 - Pump failure and subsequent failure of humidified nephelometersgpaosC1.a1
D090306.1SGP/THWAPS/C1 - Time stamp of data record incorrectsgpthwapsC1.b1


DQRID : D040318.10
Start DateStart TimeEnd DateEnd Time
10/16/2003182710/16/20031855
04/15/2004195804/15/20042020
10/28/2004183510/28/20041855
04/15/2005143504/15/20051455
11/23/2005194011/23/20051953
05/01/2006143805/01/20061545
10/13/2006160510/13/20061635
04/16/2007141404/16/20071441
10/09/2007195510/09/20072025
04/11/2008180004/11/20081840
10/30/2008160010/30/20081625
Subject:
SGP/SMOS/E13 - 6-month Calibration Checks
DataStreams:sgp1440smosE13.b1, sgp1smosE13.b1, sgp30smosE13.b1
Description:
During the time periods specified, the SMOS.E13 tower was lowered for 6-month Calibration 
Checks.  During these times wind data are incorrect and temperature and humidity data may 
be incorrect.

Note, exact times when tower was down or maintenance was being performed is sometimes 
unknown.  Users should use the remaining data from these days with caution.
Measurements:sgp1440smosE13.b1:
  • maximum wind speed (gusts)(max_wspd)
  • Time of Maximum Vapor Pressure(time_max_vap_pres)
  • Time of Maximum Barometric Pressure(time_max_bar_pres)
  • Minimum of Wind speed(min_wspd)
  • Time of Minimum Vapor Pressure(time_min_vap_pres)
  • Maximum Relative Humidity(max_rh)
  • Minimum Temperature(min_temp)
  • Time of Minimum Temperature(time_min_temp)
  • Time of Minimum Relative Humidity(time_min_rh)
  • Minimum Barometric Pressure(min_bar_pres)
  • Maximum Vapor Pressure(max_vap_pres)
  • Time of Maximum Temperature(time_max_temp)
  • Time of Maximum Relative Humidity(time_max_rh)
  • Time of Minimum Wind Speed(time_min_wspd)
  • Minimum Relative Humidity(min_rh)
  • Time of Maximum Wind Speed(time_max_wspd)
  • Maximum Temperature(max_temp)
  • Minimum Vapor Pressure(min_vap_pres)
  • Maximum Barometric Pressure(max_bar_pres)
  • Time of Minimum Barometric Pressure(time_min_bar_pres)

sgp30smosE13.b1:
  • Wind Speed (vector averaged)(wspd_va)
  • Standard Deviation of Barometric Pressure(sd_bar_pres)
  • Vapor Pressure (kiloPascals)(vap_pres)
  • Relative humidity inside the instrument enclosure(rh)
  • Standard Deviation of Wind Speed(sd_wspd)
  • Mean Wind Speed(wspd)
  • Unit Vector Wind Direction(wdir)
  • Standard Deviation of Air Temperature(sd_temp)
  • Standard Deviation of Relative Humidity(sd_rh)
  • Standard Deviation of Vapor Pressure(sd_vap_pres)
  • barometric pressure(bar_pres)
  • Standard Deviation of wind direction(sd_deg)
  • Beam 0 Temperature(temp)

sgp1smosE13.b1:
  • barometric pressure(bar_pres)
  • Relative humidity inside the instrument enclosure(rh)
  • Standard Deviation of wind direction(sd_deg)
  • Wind Speed (vector averaged)(wspd_va)
  • Unit Vector Wind Direction(wdir)
  • Mean Wind Speed(wspd)
  • Vapor Pressure (kiloPascals)(vap_pres)
  • Beam 0 Temperature(temp)


Back To Table of Contents

DQRID : D070314.2
Start DateStart TimeEnd DateEnd Time
01/12/2007000001/15/20070000
Subject:
SGP/MWR/C1/E14 - Freezing rain-Incorrect rain flag
DataStreams:sgpmwrlosC1.b1, sgpmwrlosE14.b1, sgpmwrtipC1.a1, sgpmwrtipE14.a1
Description:
Between 01/12 and 01/15 brightness temperatures show high values (Tb > 100 K) indicative 
of rain, however the wet_window flag is 0 (indicative of no rain). During that time there 
where freezing conditions with ice pellets that were not detected by the sensor. The high 
brightness temperatures may be due to melting ice on the window.
Measurements:sgpmwrtipC1.a1:
  • Water on Teflon window (1=WET, 0=DRY)(wet_window)

sgpmwrlosC1.b1:
  • Water on Teflon window (1=WET, 0=DRY)(wet_window)

sgpmwrtipE14.a1:
  • Water on Teflon window (1=WET, 0=DRY)(wet_window)

sgpmwrlosE14.b1:
  • Water on Teflon window (1=WET, 0=DRY)(wet_window)


Back To Table of Contents

DQRID : D070406.2
Start DateStart TimeEnd DateEnd Time
03/20/2007040006/05/20072000
Subject:
SGP/MWR/E14 - Power supply failure
DataStreams:sgpmwrlosE14.b1, sgpmwrtipE14.a1
Description:
On 3/20 the instrument suffered a power supply failure and was removed for repairs
Measurements:sgpmwrtipE14.a1:
  • (tknd)
  • 23.8 GHz goodness-of-fit coefficient(r23)
  • Temperature correction coefficient at 23.8 GHz(tc23)
  • base time(base_time)
  • Water on Teflon window (1=WET, 0=DRY)(wet_window)
  • 31.4 GHz blackbody(bb31)
  • Noise injection temp at 31.4 GHz derived from this tip(tnd31I)
  • 31.4 GHz sky signal(tipsky31)
  • Time offset of tweaks from base_time(time_offset)
  • Noise injection temp at nominal temperature at 31.4 GHz(tnd_nom31)
  • Noise injection temp at 23.8 GHz adjusted to tkbb(tnd23)
  • lat(lat)
  • 31.4 GHz goodness-of-fit coefficient(r31)
  • 31.8 GHz sky brightness temperature derived from tip curve(tbskytip31)
  • Dummy altitude for Zeb(alt)
  • lon(lon)
  • Actual elevation angle(actel)
  • Actual Azimuth(actaz)
  • 31.4 GHz blac2body+noise injection signal(bbn31)
  • Noise injection temp at 31.4 GHz adjusted to tkbb(tnd31)
  • Mixer kinetic (physical) temperature(tkxc)
  • Total water vapor along zenith path using tip-derived brightness temperatures(vaptip)
  • Total liquid water along zenith path using tip-derived brightness temperatures(liqtip)
  • Noise injection temp at 23.8 GHz derived from this tip(tnd23I)
  • 23.8 GHz Blackbody signal(bb23)
  • Ambient temperature(tkair)
  • Temperature correction coefficient at 31.4 GHz(tc31)
  • 23.8 GHz sky brightness temperature derived from tip curve(tbskytip23)
  • Blackbody kinetic temperature(tkbb)
  • 23.8 GHz sky signal(tipsky23)
  • Noise injection temp at nominal temperature at 23.8 GHz(tnd_nom23)
  • 23.8 GHz blackbody+noise injection signal(bbn23)

sgpmwrlosE14.b1:
  • Dummy altitude for Zeb(alt)
  • Sky brightness temperature at 23.8 GHz(tbsky23)
  • MWR column precipitable water vapor(vap)
  • 31.4 GHz blac2body+noise injection signal(bbn31)
  • 23.8 GHz Blackbody signal(bb23)
  • Sky brightness temperature at 31.4 GHz(tbsky31)
  • Noise injection temp at 23.8 GHz adjusted to tkbb(tnd23)
  • Sky Infra-Red Temperature(sky_ir_temp)
  • (tknd)
  • lat(lat)
  • lon(lon)
  • 23.8 GHz sky signal(sky23)
  • Blackbody kinetic temperature(tkbb)
  • Temperature correction coefficient at 31.4 GHz(tc31)
  • Ambient temperature(tkair)
  • Time offset of tweaks from base_time(time_offset)
  • 31.4 GHz sky signal(sky31)
  • Averaged total liquid water along LOS path(liq)
  • Noise injection temp at nominal temperature at 31.4 GHz(tnd_nom31)
  • 23.8 GHz blackbody+noise injection signal(bbn23)
  • base time(base_time)
  • Water on Teflon window (1=WET, 0=DRY)(wet_window)
  • Noise injection temp at 31.4 GHz adjusted to tkbb(tnd31)
  • Mixer kinetic (physical) temperature(tkxc)
  • 31.4 GHz blackbody(bb31)
  • Temperature correction coefficient at 23.8 GHz(tc23)
  • Noise injection temp at nominal temperature at 23.8 GHz(tnd_nom23)


Back To Table of Contents

DQRID : D070420.1
Start DateStart TimeEnd DateEnd Time
02/27/2007191204/12/20072107
Subject:
SGP/AOS/C1 - Humidifier controller failure
DataStreams:sgpaosC1.00, sgpaosC1.a0, sgpaosauxC1.a0
Description:
Watlow PID controller which controls the humidifier stopped working. This mostly affects 
calculated fits of fRH
Measurements:sgpaosauxC1.a0:
  • TSI Low RH Neph. 550 nm back scattering coefficient for zero gas(GrnBScatZero_LRH)
  • TSI High RH Neph. 550 nm total scattering coefficient for zero gas(GrnTScatZero_HRH)
  • TSI Low RH Neph. 550 nm rayleigh scattering coefficient for zero gas(GrnRScatZero_LRH)
  • TSI High RH Neph. 700 nm total scattering coefficient for zero gas(RedTScatZero_HRH)
  • TSI High RH Neph. 550 nm back scattering coefficient for cal. gas(GrnBScatSpn_HRH)
  • TSI High RH Neph. 450 nm back scattering coefficient for zero gas(BluBScatZero_HRH)
  • TSI Low RH Neph. 450 nm total scattering coefficient for cal. gas(BluTScatSpn_LRH)
  • TSI Low RH Neph. 450 nm back scattering coefficient for cal. gas(BluBScatSpn_LRH)
  • TSI Low RH Neph. 450 nm rayleigh scattering coefficient for zero gas(BluRScatZero_LRH)
  • Relative humidity in main sample manifold heater(Line11RH)
  • TSI Low RH Neph. 550nm reference intensity(TSISensGrn_LRH)
  • TSI Low RH Neph. 550 nm total scattering coefficient for cal. gas(GrnTScatSpn_LRH)
  • TSI High RH Neph. 450 nm rayleigh scattering coefficient for zero gas(BluRScatZero_HRH)
  • TSI High RH Neph. 700 nm back scattering coefficient for zero gas(RedBScatZero_HRH)
  • TSI Low RH Neph. lamp voltage(TSINephLV_LRH)
  • TSI High RH Neph. 700 nm back scattering coefficient for cal. gas(RedBScatSpn_HRH)
  • TSI Low RH Neph. 700 nm total scattering coefficient for cal. gas(RedTScatSpn_LRH)
  • TSI Low RH Neph. 550 nm total scattering coefficient for zero gas(GrnTScatZero_LRH)
  • TSI High RH Neph. 550 nm back scattering coefficient for zero gas(GrnBScatZero_HRH)
  • TSI High RH Neph. 700 nm total scattering coefficient for cal. gas(RedTScatSpn_HRH)
  • TSI High RH Neph. lamp voltage(TSINephLV_HRH)
  • TSI Low RH Neph. 700 nm total scattering coefficient for zero gas(RedTScatZero_LRH)
  • TSI High RH Neph. 550 nm total scattering coefficient for cal. gas(GrnTScatSpn_HRH)
  • TSI High RH Neph. 450 nm total scattering coefficient for cal. gas(BluTScatSpn_HRH)
  • TSI High RH Neph. 550 nm rayleigh scattering coefficient for zero gas(GrnRScatZero_HRH)
  • TSI High RH Neph. lamp current(TSINephLC_HRH)
  • TSI Low RH Neph. 700 nm rayleigh scattering coefficient for zero gas(RedRScatZero_LRH)
  • TSI High RH Neph. 450 nm total scattering coefficient for zero gas(BluTScatZero_HRH)
  • TSI Low RH Neph. 450 nm back scattering coefficient for zero gas(BluBScatZero_LRH)
  • TSI Low RH Neph. 700 nm back scattering coefficient for zero gas(RedBScatZero_LRH)
  • TSI Low RH Neph. 550 nm back scattering coefficient for cal. gas(GrnBScatSpn_LRH)
  • TSI High RH Neph. 550nm reference intensity(TSISensGrn_HRH)
  • TSI Low RH Neph. 450 nm total scattering coefficient for zero gas(BluTScatZero_LRH)
  • TSI Low RH Neph. lamp current(TSINephLC_LRH)
  • TSI Low RH Neph. 700 nm back scattering coefficient for cal. gas(RedBScatSpn_LRH)
  • TSI High RH Neph. 700 nm rayleigh scattering coefficient for zero gas(RedRScatZero_HRH)
  • TSI High RH Neph. 450 nm back scattering coefficient for cal. gas(BluBScatSpn_HRH)

sgpaosC1.00:
  • null(Raw data stream - documentation not supported)

sgpaosC1.a0:
  • TSI High RH Neph. Pressure(TSINephPres_HRH)
  • TSI High RH Neph. Sample Temperature(TSINephTSamp_HRH)
  • TSI Low RH Neph. 450 nm backscat. coef. at 1 um(BluBScatCoef_1um_LRH)
  • TSI High RH Neph. 700 nm backscat. coef. at 10 um(RedBScatCoef_10um_HRH)
  • TSI Low RH Neph. 700 nm backscat. coef. at 10 um(RedBScatCoef_10um_LRH)
  • TSI High RH Neph. 550 nm backscat. coef. at 10 um(GrnBScatCoef_10um_HRH)
  • TSI High RH Neph. 700 nm total scat. coef. at 10 um(RedTScatCoef_10um_HRH)
  • TSI High RH Neph. 550 nm total scat. coef. at 10 um(GrnTScatCoef_10um_HRH)
  • TSI High RH Neph. 450 nm total scat. coef. at 10 um(BluTScatCoef_10um_HRH)
  • TSI Low RH Neph. relative humidity(TSINephRHSamp_LRH)
  • TSI Low RH Neph. inlet temperature(TSINephTin_LRH)
  • TSI Low RH Neph. 550 nm backscat. coef. at 1 um(GrnBScatCoef_1um_LRH)
  • TSI High RH Neph. relative humidity(TSINephRHSamp_HRH)
  • TSI Low RH Neph. 550 nm total scat. coef. at 10 um(GrnTScatCoef_10um_LRH)
  • TSI Low RH Neph. Pressure(TSINephPres_LRH)
  • TSI High RH Neph. 450 nm backscat. coef. at 10 um(BluBScatCoef_10um_HRH)
  • TSI High RH Neph. 700 nm total scat. coef. at 1 um(RedTScatCoef_1um_HRH)
  • TSI Low RH Neph. 550 nm total scat. coef. at 1 um(GrnTScatCoef_1um_LRH)
  • TSI Low RH Neph. Sample Temperature(TSINephTSamp_LRH)
  • TSI Low RH Neph. 700 nm total scat. coef. at 10 um(RedTScatCoef_10um_LRH)
  • Relative humidity in sample line down. of heater(SampRH)
  • TSI Low RH Neph. 450 nm total scat. coef. at 10 um(BluTScatCoef_10um_LRH)
  • TSI Low RH Neph. 450 nm total scat. coef. at 1 um(BluTScatCoef_1um_LRH)
  • RH upstream of humidograph(HGupstream_RH)
  • TSI High RH Neph. 700 nm backscat. coef. at 1 um(RedBScatCoef_1um_HRH)
  • TSI High RH Neph. 550 nm backscat. coef. at 1 um(GrnBScatCoef_1um_HRH)
  • TSI Low RH Neph. 700 nm total scat. coef. at 1 um(RedTScatCoef_1um_LRH)
  • TSI High RH Neph. 550 nm total scat. coef. at 1 um(GrnTScatCoef_1um_HRH)
  • TSI High RH Neph. inlet temperature(TSINephTin_HRH)
  • TSI High RH Neph. 450 nm backscat. coef. at 1 um(BluBScatCoef_1um_HRH)
  • TSI High RH Neph. 450 nm total scat. coef. at 1 um(BluTScatCoef_1um_HRH)
  • TSI Low RH Neph. 550 nm backscat. coef. at 10 um(GrnBScatCoef_10um_LRH)
  • TSI Low RH Neph. 700 nm backscat. coef. at 1 um(RedBScatCoef_1um_LRH)
  • RH downstream of humidograph(HGdownstream_RH)
  • TSI Low RH Neph. 450 nm backscat. coef. at 10 um(BluBScatCoef_10um_LRH)
  • Single wave Neph. RH(RH530nm)


Back To Table of Contents

DQRID : D070613.2
Start DateStart TimeEnd DateEnd Time
06/01/2007000006/02/20072359
Subject:
SGP/RAIN/C1 - Funnel clogged
DataStreams:sgprainC1.b1
Description:
The funnel for the instrument was clogged and the results for this time period are not 
reliable.
Measurements:sgprainC1.b1:
  • Time offset of tweaks from base_time(time_offset)
  • Dummy altitude for Zeb(alt)
  • Rainfall rate(rain_rate)
  • lon(lon)
  • lat(lat)
  • Precipitation(precip_tbrg)
  • base time(base_time)
  • time(time)


Back To Table of Contents

DQRID : D070618.2
Start DateStart TimeEnd DateEnd Time
04/16/2007000006/18/20072000
Subject:
SGP/SMOS/E13 - RH sensor failure
DataStreams:sgp1440smosE13.b1, sgp1smosE13.b1, sgp30smosE13.b1
Description:
The T/RH sensor relative humidity data is suspect during this time.  Values from this 
sensor failed the 6-month check and subsequent bi-weekly checks.  The RH values were lower by 
larger than the manufacturer suggested uncertainty of 3%.
Measurements:sgp1440smosE13.b1:
  • Time of Maximum Relative Humidity(time_max_rh)
  • Time of Maximum Vapor Pressure(time_max_vap_pres)
  • Minimum Relative Humidity(min_rh)
  • Time of Minimum Vapor Pressure(time_min_vap_pres)
  • Maximum Relative Humidity(max_rh)
  • Time of Minimum Relative Humidity(time_min_rh)
  • Maximum Vapor Pressure(max_vap_pres)
  • Minimum Vapor Pressure(min_vap_pres)

sgp30smosE13.b1:
  • Standard Deviation of Relative Humidity(sd_rh)
  • Standard Deviation of Vapor Pressure(sd_vap_pres)
  • Vapor Pressure (kiloPascals)(vap_pres)
  • Relative humidity inside the instrument enclosure(rh)

sgp1smosE13.b1:
  • Relative humidity inside the instrument enclosure(rh)
  • Vapor Pressure (kiloPascals)(vap_pres)


Back To Table of Contents

DQRID : D070905.2
Start DateStart TimeEnd DateEnd Time
07/11/2007094007/24/20072205
08/17/2007214008/30/20071900
Subject:
SGP/AOS/C1 - CPC failure
DataStreams:sgpaosC1.a1
Description:
The Condensation Particle Counter stopped recording counts. Upon inspection the tech found 
that water had condensed in the reservoir. The CPC was shipped back to the mentor for 
repair and a replacement CPC from NOAA was sent to the site.

The replacement Condensation Particle Counter stopped measuring particle counts and was 
removed and returned to NOAA. The thermoelectric cooler in the device had stopped working. 
The ARM CPC that had previously been sent back to NOAA for repair was fixed and returned 
to SGP.
Measurements:sgpaosC1.a1:
  • Condensation Nuclei Concentration Number(N_CPC_1)


Back To Table of Contents

DQRID : D080116.13
Start DateStart TimeEnd DateEnd Time
12/10/2007221212/11/20071945
Subject:
SGP/SMOS/E13 - Wind Data Incorrect Due to Icing
DataStreams:sgp1440smosE13.b1, sgp1smosE13.b1, sgp30smosE13.b1
Description:
An ice storm caused the wind data from the RM Young wind sensor to be
incorrect.  In many instances it was frozen and unmoving, at times it
measured values but those are extremely suspect.
Measurements:sgp1440smosE13.b1:
  • maximum wind speed (gusts)(max_wspd)
  • Time of Minimum Wind Speed(time_min_wspd)
  • Minimum of Wind speed(min_wspd)
  • Time of Maximum Wind Speed(time_max_wspd)

sgp30smosE13.b1:
  • Unit Vector Wind Direction(wdir)
  • Wind Speed (vector averaged)(wspd_va)
  • Standard Deviation of Wind Speed(sd_wspd)
  • Standard Deviation of wind direction(sd_deg)
  • Mean Wind Speed(wspd)

sgp1smosE13.b1:
  • Mean Wind Speed(wspd)
  • Standard Deviation of wind direction(sd_deg)
  • Wind Speed (vector averaged)(wspd_va)
  • Unit Vector Wind Direction(wdir)


Back To Table of Contents

DQRID : D080116.24
Start DateStart TimeEnd DateEnd Time
12/22/2007175012/22/20072130
Subject:
SGP/THWAPS/C1 - RH Data Incorrect
DataStreams:sgpthwapsC1.b1
Description:
The RH data was incorrect during the times listed.  Reason unknown.  The sensor recovered 
on its own.
Measurements:sgpthwapsC1.b1:
  • Relative humidity inside the instrument enclosure(rh)
  • Standard Deviation of Vapor Pressure(sd_vap_pres)
  • Standard Deviation of Relative Humidity(sd_rh)
  • Vapor Pressure (kiloPascals)(vap_pres)


Back To Table of Contents

DQRID : D080128.3
Start DateStart TimeEnd DateEnd Time
09/01/2007000011/13/20072000
Subject:
SGP/AOS/C1/ - Pump failure
DataStreams:sgpaosccnC1.a1
Description:
The ratio of CCN to CN started to noticeably decline around 20070901. After a site visit 
on 20071023, the mentor found a problem with the instrument flows and diaphragm pump.  The 
instrument was return to Boulder, CO for repair. The diaphragm pump was repaired and the 
flows re-calibrated. The instrument was returned to service 20071113. 

The CCN data show qualitative changes in the number of droplets formed with changes in 
supersaturation. Because the pump was not working properly and the flows were likely 
incorrect, the numbers are not quantitative.
Measurements:sgpaosccnC1.a1:
  • AOS CCN sample supersaturation calculated by model)(CCN_ss_calc)
  • CCN temperature gradient(CCN_Temperature_Gradient)
  • AOS bin number of lowest channel of OPC include in summation of N_CCN(N_CCN_bin_number)
  • AOS CCN sample pressure(CCN_P_sample)
  • AOS CCN sample saturation setpoint value reported by instrument(CCN_ss_set)
  • AOS CCN volumetric flowrate of sheath air(CCN_Q_sheath)
  • AOS number concentration of CCN(N_CCN)
  • AOS CCN volumetric flowrate of sample air(CCN_Q_sample)
  • AOS CCN number concentration by bin(N_CCN_dN)


Back To Table of Contents

DQRID : D080222.4
Start DateStart TimeEnd DateEnd Time
01/31/2008000002/02/20082000
Subject:
SGP/SMOS - Snowfall caused delayed count of precip
DataStreams:sgp1smosE1.b1, sgp1smosE11.b1, sgp1smosE13.b1, sgp1smosE15.b1, sgp1smosE20.b1,
sgp1smosE21.b1, sgp1smosE24.b1, sgp1smosE27.b1, sgp1smosE3.b1, sgp1smosE4.b1, sgp1smosE5.b1,
sgp1smosE6.b1, sgp1smosE7.b1, sgp1smosE8.b1, sgp1smosE9.b1, sgp30smosE1.b1,
sgp30smosE11.b1, sgp30smosE13.b1, sgp30smosE15.b1, sgp30smosE20.b1, sgp30smosE21.b1, sgp30smosE24.b1,
sgp30smosE27.b1, sgp30smosE3.b1, sgp30smosE4.b1, sgp30smosE5.b1, sgp30smosE6.b1,
sgp30smosE7.b1, sgp30smosE8.b1, sgp30smosE9.b1
Description:
A snow/ice storm moved through the SGP region on 01/31/08 bringing below-freezing 
temperatures, affecting the precip field.  Data is suspect during the date/time range specified 
due to melting ice after temperatures warmed above freezing.
Measurements:sgp1smosE24.b1:
  • Hourly precipitation(precip)

sgp1smosE21.b1:
  • Hourly precipitation(precip)

sgp30smosE7.b1:
  • Hourly precipitation(precip)

sgp1smosE27.b1:
  • Hourly precipitation(precip)

sgp1smosE7.b1:
  • Hourly precipitation(precip)

sgp30smosE15.b1:
  • Hourly precipitation(precip)

sgp30smosE20.b1:
  • Hourly precipitation(precip)

sgp1smosE9.b1:
  • Hourly precipitation(precip)

sgp1smosE4.b1:
  • Hourly precipitation(precip)

sgp30smosE3.b1:
  • Hourly precipitation(precip)

sgp1smosE11.b1:
  • Hourly precipitation(precip)

sgp30smosE27.b1:
  • Hourly precipitation(precip)

sgp1smosE3.b1:
  • Hourly precipitation(precip)

sgp30smosE4.b1:
  • Hourly precipitation(precip)

sgp30smosE8.b1:
  • Hourly precipitation(precip)

sgp1smosE8.b1:
  • Hourly precipitation(precip)

sgp30smosE9.b1:
  • Hourly precipitation(precip)

sgp1smosE13.b1:
  • Hourly precipitation(precip)

sgp30smosE11.b1:
  • Hourly precipitation(precip)

sgp1smosE1.b1:
  • Hourly precipitation(precip)

sgp30smosE5.b1:
  • Hourly precipitation(precip)

sgp1smosE20.b1:
  • Hourly precipitation(precip)

sgp30smosE24.b1:
  • Hourly precipitation(precip)

sgp30smosE13.b1:
  • Hourly precipitation(precip)

sgp30smosE1.b1:
  • Hourly precipitation(precip)

sgp1smosE6.b1:
  • Hourly precipitation(precip)

sgp30smosE6.b1:
  • Hourly precipitation(precip)

sgp1smosE15.b1:
  • Hourly precipitation(precip)

sgp30smosE21.b1:
  • Hourly precipitation(precip)

sgp1smosE5.b1:
  • Hourly precipitation(precip)


Back To Table of Contents

DQRID : D080403.2
Start DateStart TimeEnd DateEnd Time
03/04/2008160003/12/20081800
Subject:
SGP/WACR/C1 - Contamination with AMF WACR data
DataStreams:sgpwacrC1.00, sgpwacrC1.b1, sgpwacrspecmomC1.00
Description:
During this period we were doing an inter-comparison with the AMF WACR.  As a result, the 
AMF WACR contaminated the SGP WACR data.
Measurements:sgpwacrC1.b1:
  • Chiller supply temperature(Temp_Chiller_Supply)
  • Computer enclosure temperature(Temp_Computer_Enclosure)
  • Detected Pulse(txpower)
  • Transmit driver power sampled by receiver(PowerTransmitDriver)
  • Dummy altitude for Zeb(alt)
  • Extended Interaction Klystron Amplifier temperature(Temp_EIKA)
  • Chiller return temperature(Temp_Chiller_Return)
  • Array of heights for each power(heights)
  • Signal to Noise Ratio(SignalToNoiseRatio)
  • Modulator temperature(Temp_modulator)
  • Ambient temperature of radar front end components(Temp_ambient)
  • Antenna bottom temperature(Temp_Antenna_Bottom)
  • Chiller temperature(Temp_Chiller)
  • MMCR Mean Doppler Velocity(MeanDopplerVelocity)
  • time(time)
  • lat(lat)
  • lon(lon)
  • status flag(wacr_status)
  • Hot load power(PowerHotLoad)
  • MMCR Spectral Width(SpectralWidth)
  • Hot load physical temperature(Temp_Hot)
  • Time offset of tweaks from base_time(time_offset)
  • Modulator control temperature(Temp_Modulator_Control)
  • Polarization mode. 0 = copol, 1 = xpol(Polarization)
  • 1000 * log10 (noise signal level)(Noise)
  • Low noise amplifier temperature(Temp_LNA)
  • Outside temperature(Temp_Outside)
  • Antenna top temperature(Temp_Antenna_Top)
  • MMCR Reflectivity(Reflectivity)
  • base time(base_time)
  • Future temperature(Temp_Future)
  • Ambient load power(PowerAmbientLoad)

sgpwacrC1.00:
  • Raw data stream - documentation not supported(raw)

sgpwacrspecmomC1.00:
  • Raw data stream - documentation not supported(raw)


Back To Table of Contents

DQRID : D080820.9
Start DateStart TimeEnd DateEnd Time
01/06/2006170001/09/20060930
02/17/2006200002/20/20061230
05/14/2006210005/16/20060530
08/26/2006080008/29/20060030
12/22/2007180012/24/20070230
04/26/2008000004/28/20081300
Subject:
SGP/SMOS/E13 - Missing Data
DataStreams:sgp1440smosE13.b1, sgp1smosE13.b1, sgp30smosE13.b1
Description:
Data are missing and unrecoverable
Measurements:sgp1440smosE13.b1:
  • Time of Maximum Vapor Pressure(time_max_vap_pres)
  • base time(base_time)
  • Time of Maximum Barometric Pressure(time_max_bar_pres)
  • lat(lat)
  • time(time)
  • Time offset of tweaks from base_time(time_offset)
  • Minimum of Wind speed(min_wspd)
  • Time of Minimum Vapor Pressure(time_min_vap_pres)
  • Minimum Temperature(min_temp)
  • Time of Minimum Relative Humidity(time_min_rh)
  • Minimum Barometric Pressure(min_bar_pres)
  • Maximum Vapor Pressure(max_vap_pres)
  • Time of Minimum Snow Depth(time_min_snow)
  • Time of Maximum Temperature(time_max_temp)
  • Time of Maximum Relative Humidity(time_max_rh)
  • lon(lon)
  • Time of Maximum Wind Speed(time_max_wspd)
  • Minimum Snow Depth(min_snow_depth)
  • Maximum Snow Depth(max_snow)
  • maximum wind speed (gusts)(max_wspd)
  • Maximum Relative Humidity(max_rh)
  • Time of Minimum Temperature(time_min_temp)
  • Time of Maximum Snow Depth(time_max_snow)
  • Time of Minimum Wind Speed(time_min_wspd)
  • Minimum Relative Humidity(min_rh)
  • Dummy altitude for Zeb(alt)
  • Maximum Temperature(max_temp)
  • Minimum Vapor Pressure(min_vap_pres)
  • Maximum Barometric Pressure(max_bar_pres)
  • Time of Minimum Barometric Pressure(time_min_bar_pres)

sgp30smosE13.b1:
  • Vapor Pressure (kiloPascals)(vap_pres)
  • Relative humidity inside the instrument enclosure(rh)
  • base time(base_time)
  • lon(lon)
  • Mean Wind Speed(wspd)
  • Standard Deviation of Air Temperature(sd_temp)
  • Batter Voltage(vbat)
  • Standard Deviation of Vapor Pressure(sd_vap_pres)
  • Time offset of tweaks from base_time(time_offset)
  • lat(lat)
  • Standard Deviation of wind direction(sd_deg)
  • Snow Depth Sensor(snow_sen)
  • Wind Speed (vector averaged)(wspd_va)
  • Standard Deviation of Barometric Pressure(sd_bar_pres)
  • Snow Depth(snow)
  • Standard Deviation of Wind Speed(sd_wspd)
  • Hourly precipitation(precip)
  • Unit Vector Wind Direction(wdir)
  • Standard Deviation of Relative Humidity(sd_rh)
  • barometric pressure(bar_pres)
  • Dummy altitude for Zeb(alt)
  • Beam 0 Temperature(temp)

sgp1smosE13.b1:
  • lon(lon)
  • Standard Deviation of wind direction(sd_deg)
  • Wind Speed (vector averaged)(wspd_va)
  • base time(base_time)
  • Mean Wind Speed(wspd)
  • barometric pressure(bar_pres)
  • Relative humidity inside the instrument enclosure(rh)
  • Unit Vector Wind Direction(wdir)
  • Dummy altitude for Zeb(alt)
  • Time offset of tweaks from base_time(time_offset)
  • lat(lat)
  • Hourly precipitation(precip)
  • Vapor Pressure (kiloPascals)(vap_pres)
  • Beam 0 Temperature(temp)


Back To Table of Contents

DQRID : D081015.1
Start DateStart TimeEnd DateEnd Time
10/13/2008000010/17/20081755
Subject:
SGP/SMOS/E13 - Rain Gauge Inoperative
DataStreams:sgp1smosE13.b1, sgp30smosE13.b1
Description:
The rain gauge measured values did not compare well with the disdrometer.  Subsequent 
testing found the gauge to be incorrect.
Measurements:sgp30smosE13.b1:
  • Hourly precipitation(precip)

sgp1smosE13.b1:
  • Hourly precipitation(precip)


Back To Table of Contents

DQRID : D081212.2
Start DateStart TimeEnd DateEnd Time
12/08/2008154712/10/20082059
Subject:
SGP/SMOS/E13 - SMOS T/RH probe solar shield missing
DataStreams:sgp1440smosE13.b1, sgp1smosE13.b1, sgp30smosE13.b1
Description:
The solar shield protecting the T/RH probe from direct sunlight and the elements became 
unscrewed and fell off.  The probe was affected by sunlight and did not have a consistent 
aspiration during this time.
Measurements:sgp1440smosE13.b1:
  • Time of Maximum Temperature(time_max_temp)
  • Time of Maximum Relative Humidity(time_max_rh)
  • Time of Maximum Vapor Pressure(time_max_vap_pres)
  • Minimum Relative Humidity(min_rh)
  • Time of Minimum Vapor Pressure(time_min_vap_pres)
  • Minimum Temperature(min_temp)
  • Maximum Relative Humidity(max_rh)
  • Time of Minimum Temperature(time_min_temp)
  • Time of Minimum Relative Humidity(time_min_rh)
  • Maximum Temperature(max_temp)
  • Maximum Vapor Pressure(max_vap_pres)
  • Minimum Vapor Pressure(min_vap_pres)

sgp30smosE13.b1:
  • Standard Deviation of Air Temperature(sd_temp)
  • Standard Deviation of Relative Humidity(sd_rh)
  • Vapor Pressure (kiloPascals)(vap_pres)
  • Relative humidity inside the instrument enclosure(rh)
  • Beam 0 Temperature(temp)
  • Standard Deviation of wind direction(sd_deg)

sgp1smosE13.b1:
  • Relative humidity inside the instrument enclosure(rh)
  • Standard Deviation of wind direction(sd_deg)
  • Vapor Pressure (kiloPascals)(vap_pres)
  • Beam 0 Temperature(temp)


Back To Table of Contents

DQRID : D090120.5
Start DateStart TimeEnd DateEnd Time
07/19/2008011201/06/20091912
Subject:
SGP/AOS/C1 - Pump failure and subsequent failure of humidified nephelometer
DataStreams:sgpaosC1.a1
Description:
The pump for the aerosol sampling system failed over a weekend. The humidifier continued 
to operate and soaked the inside of both the reference and wet nephelometers. The 
nephelometers were returned for repair. A replacement neph was installed on 2008-August 5 @1858 
UTC. The nephs were repaired and returned to SGP in December 2008. The humidification 
system was rebuilt and installed on 2009 January 6 at 1912 UTC.
Measurements:sgpaosC1.a1:
  • Aerosol light absorption coefficient, green channel, 10 um size cut(Ba_G_Dry_10um_PSAP1W_1)
  • Absorption Coefficient Red, corrected for flow and sample area, 3 wavelength
    PSAP, 10 um size cut(Ba_R_Dry_10um_PSAP3W_1)
  • Absorption Coefficient Green, corrected for flow and sample area, 3 wavelength
    PSAP, 10 um size cut(Ba_G_Dry_10um_PSAP3W_1)
  • Aerosol backwards-hemispheric light scattering coefficient, green channel, 10 um
    size cut, reference three wavelength nephelometer(Bbs_G_Dry_10um_Neph3W_1)
  • Aerosol backwards-hemispheric light scattering coefficient, red channel, 1 um
    size cut, wet three wavelength nephelometer(Bbs_R_Wet_1um_Neph3W_2)
  • Absorption Coefficient Red, corrected for flow and sample area, 3 wavelength
    PSAP, 1 um size cut(Ba_R_Dry_1um_PSAP3W_1)
  • Aerosol backwards-hemispheric light scattering coefficient, green channel, 1 um
    size cut, wet three wavelength nephelometer(Bbs_G_Wet_1um_Neph3W_2)
  • Aerosol total light scattering coefficient, blue channel, 1 um size cut, wet
    three wavelength nephelometer(Bs_B_Wet_1um_Neph3W_2)
  • Aerosol total light scattering coefficient, green channel, 10 um size cut, wet
    three wavelength nephelometer(Bs_G_Wet_10um_Neph3W_2)
  • Aerosol backwards-hemispheric light scattering coefficient, red channel, 10 um
    size cut, wet three wavelength nephelometer(Bbs_R_Wet_10um_Neph3W_2)
  • Aerosol total light scattering coefficient, green channel, 1 um size cut, wet
    three wavelength nephelometer(Bs_G_Wet_1um_Neph3W_2)
  • Aerosol total light scattering coefficient, red channel, 1 um size cut, wet
    three wavelength nephelometer(Bs_R_Wet_1um_Neph3W_2)
  • Relative humidity of humidograph preheater(RH_preHG)
  • Aerosol total light scattering coefficient, red channel, 1 um size cut,
    reference three wavelength nephelometer(Bs_R_Dry_1um_Neph3W_1)
  • Temperature of humidograph controller(T_postHG)
  • Aerosol total light scattering coefficient, red channel, 10 um size cut, wet
    three wavelength nephelometer(Bs_R_Wet_10um_Neph3W_2)
  • Temperature at wet nephelometer inlet(T_NephInlet_Wet)
  • Relative humidity of humidograph controller(RH_postHG)
  • Aerosol backwards-hemispheric light scattering coefficient, green channel, 1 um
    size cut, reference three wavelength nephelometer(Bbs_G_Dry_1um_Neph3W_1)
  • Aerosol total light scattering coefficient, green channel, 1 um size cut,
    reference three wavelength nephelometer(Bs_G_Dry_1um_Neph3W_1)
  • Relative humidity at wet nephelometer inlet(RH_NephInlet_Wet)
  • Temperature at reference nephelometer inlet(T_NephInlet_Dry)
  • Aerosol backwards-hemispheric light scattering coefficient, blue channel, 10 um
    size cut, wet three wavelength nephelometer(Bbs_B_Wet_10um_Neph3W_2)
  • Absorption Coefficient Green, corrected for flow and sample area, 3 wavelength
    PSAP, 1 um size cut(Ba_G_Dry_1um_PSAP3W_1)
  • Aerosol backwards-hemispheric light scattering coefficient, blue channel, 10 um
    size cut, reference three wavelength nephelometer(Bbs_B_Dry_10um_Neph3W_1)
  • Aerosol backwards-hemispheric light scattering coefficient, blue channel, 1 um
    size cut, reference three wavelength nephelometer(Bbs_B_Dry_1um_Neph3W_1)
  • Relative humidity at reference nephelometer inlet(RH_NephInlet_Dry)
  • Absorption Coefficient Blue, corrected for flow and sample area, 3 wavelength
    PSAP, 10 um size cut(Ba_B_Dry_10um_PSAP3W_1)
  • Aerosol total light scattering coefficient, green channel, 10 um size cut,
    reference three wavelength nephelometer(Bs_G_Dry_10um_Neph3W_1)
  • Temperature of humidograph preheater(T_preHG)
  • Aerosol total light scattering coefficient, blue channel, 10 um size cut, wet
    three wavelength nephelometer(Bs_B_Wet_10um_Neph3W_2)
  • Aerosol backwards-hemispheric light scattering coefficient, blue channel, 1 um
    size cut, wet three wavelength nephelometer(Bbs_B_Wet_1um_Neph3W_2)
  • Aerosol light absorption coefficient, green channel, 1 um size cut(Ba_G_Dry_1um_PSAP1W_1)
  • Aerosol total light scattering coefficient, blue channel, 1 um size cut,
    reference three wavelength nephelometer(Bs_B_Dry_1um_Neph3W_1)
  • Aerosol backwards-hemispheric light scattering coefficient, red channel, 10 um
    size cut, reference three wavelength nephelometer(Bbs_R_Dry_10um_Neph3W_1)
  • Aerosol backwards-hemispheric light scattering coefficient, green channel, 10 um
    size cut, wet three wavelength nephelometer(Bbs_G_Wet_10um_Neph3W_2)
  • Aerosol total light scattering coefficient, red channel, 10 um size cut,
    reference three wavelength nephelometer(Bs_R_Dry_10um_Neph3W_1)
  • Aerosol total light scattering coefficient, blue channel, 10 um size cut,
    reference three wavelength nephelometer(Bs_B_Dry_10um_Neph3W_1)
  • Aerosol backwards-hemispheric light scattering coefficient, red channel, 1 um
    size cut, reference three wavelength nephelometer(Bbs_R_Dry_1um_Neph3W_1)
  • Absorption Coefficient Blue, corrected for flow and sample area, 3 wavelength
    PSAP, 1 um size cut(Ba_B_Dry_1um_PSAP3W_1)


Back To Table of Contents

DQRID : D090306.1
Start DateStart TimeEnd DateEnd Time
12/09/2008000001/23/20092359
Subject:
SGP/THWAPS/C1 - Time stamp of data record incorrect
DataStreams:sgpthwapsC1.b1
Description:
Due to problems with the collector system, the datalogger was being time synched 
inappropriately.  This problem was intermittent and did not affect every time stamp.  The worst 
case time adjustments were approximately +/- 7 minutes.  These were infrequent and most of 
the time adjustments were less than 7 minutes.
Measurements:sgpthwapsC1.b1:
  • Time offset of tweaks from base_time(time_offset)
  • base time(base_time)


Back To Table of Contents



END OF DATA