Data Quality Reports for Session: 123957 User: kristi Completed: 10/26/2009


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D030114.1SGP/SONDE/C1 - Bad RH DatasgpsondewnpnC1.a1, sgpsondewnpnC1.b1
D030122.1SGP/SONDE/C1 - Broken temperature sensor SGP/CF soundingsgpsondewnpnC1.a1, sgpsondewnpnC1.b1
D030206.5SGP/SONDE/C1 - Bad temperature/RHsgpsondewnpnC1.a1, sgpsondewnpnC1.b1
D060928.3SGP/SONDE/C1 - Increased precision for alt, dp, pres, rh, tdrysgpsondewnpnC1.b1


DQRID : D030114.1
Start DateStart TimeEnd DateEnd Time
01/11/2003053601/11/20030551
Subject:
SGP/SONDE/C1 - Bad RH Data
DataStreams:sgpsondewnpnC1.a1, sgpsondewnpnC1.b1
Description:
The RH values during the first 15 minutes of this sounding oscillate between reasonable 
values and near-zero values.  The oscillation is irregular and seems to be related to 
failure of the RH sensor heating process.  I have informed Vaisala of the problem and provided 
raw data files to them for analysis.
Measurements:sgpsondewnpnC1.a1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)

sgpsondewnpnC1.b1:
  • Surface dew point temperature(dp)
  • Relative humidity inside the instrument enclosure(rh)


Back To Table of Contents

DQRID : D030122.1
Start DateStart TimeEnd DateEnd Time
01/19/2003113401/19/20031259
Subject:
SGP/SONDE/C1 - Broken temperature sensor SGP/CF sounding
DataStreams:sgpsondewnpnC1.a1, sgpsondewnpnC1.b1
Description:
The radiosonde temperature sensor apparently broke during launch.  This results in bad 
temperature, dew point, and rh data.
Measurements:sgpsondewnpnC1.a1:
  • Dry bulb temperature(tdry)
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)

sgpsondewnpnC1.b1:
  • Surface dew point temperature(dp)
  • Dry bulb temperature(tdry)
  • Relative humidity inside the instrument enclosure(rh)


Back To Table of Contents

DQRID : D030206.5
Start DateStart TimeEnd DateEnd Time
02/02/2003173202/02/20031832
Subject:
SGP/SONDE/C1 - Bad temperature/RH
DataStreams:sgpsondewnpnC1.a1, sgpsondewnpnC1.b1
Description:
The temperature sensor broke shortly after launch.  All temperature and RH-related data 
are incorrect.
Measurements:sgpsondewnpnC1.a1:
  • Dry bulb temperature(tdry)
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)

sgpsondewnpnC1.b1:
  • Surface dew point temperature(dp)
  • Dry bulb temperature(tdry)
  • Relative humidity inside the instrument enclosure(rh)


Back To Table of Contents

DQRID : D060928.3
Start DateStart TimeEnd DateEnd Time
09/28/2006172812/31/99992359
Subject:
SGP/SONDE/C1 - Increased precision for alt, dp, pres, rh, tdry
DataStreams:sgpsondewnpnC1.b1
Description:
The output precision of the following variables was increased 20060928

alt  changed from 1m to 0.1m
dp   changed from 0.1degC to 0.01degC
pres changed from 0.1 hPa to 0.01 hPa
RH   changed from 1%RH to 0.01%RH
tdry changed from .1degC to 0.01degC

Note that the increase in precision does not necessarily imply an increase in accuracy.  
Users are cautioned to continue to use their best scientific judgement when using these or 
any other data.
Measurements:sgpsondewnpnC1.b1:
  • Surface dew point temperature(dp)
  • Dry bulb temperature(tdry)
  • Relative humidity inside the instrument enclosure(rh)
  • Retrieved pressure profile(pres)
  • Dummy altitude for Zeb(alt)


Back To Table of Contents



END OF DATA