Data Quality Reports for Session: 123920 User: koontz Completed: 10/23/2009


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D030324.1SGP/IRT25m/C1 - Reprocess: up_short_hemisp needs -1 multipliersgpirt25m20sC1.a0, sgpirt25mC1.00, sgpirt25mC1.a1, sgpirt25mC1.a2


DQRID : D030324.1
Start DateStart TimeEnd DateEnd Time
10/13/2001200203/21/20032024
Subject:
SGP/IRT25m/C1 - Reprocess: up_short_hemisp needs -1 multiplier
DataStreams:sgpirt25m20sC1.a0, sgpirt25mC1.00, sgpirt25mC1.a1, sgpirt25mC1.a2
Description:
The upwelling hemispheric shortwave measurement
   in the SGP.C1 IRT25m data streams for the time
   period noted requires a (-1) multiplier to correct
   for swapped connections at installation.
Measurements:sgpirt25mC1.a2:
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Pyranometer(up_short_hemisp)
  • Instantaneous Upwelling Shortwave Hemispheric Irradiance, Pyranometer(inst_up_short_hemisp)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Pyranometer, Maxima(up_short_hemisp_max)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Pyranometer, Minima(up_short_hemisp_min)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Pyranometer, Standard
    Deviation(up_short_hemisp_std)

sgpirt25mC1.a1:
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Pyranometer, Minima(up_short_hemisp_min)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Pyranometer, Standard
    Deviation(up_short_hemisp_std)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Pyranometer(up_short_hemisp)
  • Instantaneous Upwelling Shortwave Hemispheric Irradiance, Pyranometer(inst_up_short_hemisp)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Pyranometer, Maxima(up_short_hemisp_max)

sgpirt25mC1.00:
  • null(Raw data stream - documentation not supported)

sgpirt25m20sC1.a0:
  • Instantaneous Upwelling Pyrgeometer Thermopile Voltage, Pyrgeometer(inst_up_short_hemisp_tp)


Back To Table of Contents



END OF DATA