Data Quality Reports for Session: 123782 User: mkirn Completed: 10/19/2009


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D070628.1FKB/MWRP/M1 - Reprocess: Incorrect calibrationfkbmwrpM1.b1
D071106.1FKB/GNDRAD/M1 - Reprocessed: Incorrect calibrationsfkbgndrad20sM1.a0, fkbgndrad60sM1.b1


DQRID : D070628.1
Start DateStart TimeEnd DateEnd Time
06/20/2007052006/21/20072000
Subject:
FKB/MWRP/M1 - Reprocess: Incorrect calibration
DataStreams:fkbmwrpM1.b1
Description:
The software was started with the wrong calibration data.
Measurements:fkbmwrpM1.b1:
  • Derived relative humidity(relativeHumidity)
  • Expected root-mean-square error in liquid water content retrieval(liquidWaterContentRmsError)
  • Retrieved water vapor density(waterVaporDensity)
  • Total precipitable water vapor, from microwave radiometer(totalPrecipitableWater)
  • Level of free convection pressure(levelFreeConvectionPres)
  • Lifting condensation level(liftingCondensationLevel)
  • air temperature (NGM250 predicted)(temperature)
  • Derived virtual temperature(virtualTemperature)
  • Retrieved liquid water path using only 23.835 and 30.0 GHz(liquidWaterPath2)
  • Retrieved total precipitable water vapor using only 23.835 and 30.0 GHz(totalPrecipitableWater2)
  • Interpolated water vapor mixing ratio(waterVaporMixingRatio)
  • Expected root-mean-square error in liquid water path retrieval(liquidWaterPathRmsError)
  • Level of free convection(levelFreeConvection)
  • Expected root-mean-square error in liquid water path retrieval using only 23.835
    and 30.0 GHz(liquidWaterPath2RmsError)
  • Microwave brightness temperature(brightnessTemperature)
  • Retrieved cloud liquid water content(liquidWaterContent)
  • Retrieved liquid water path(liquidWaterPath)
  • Lifting condensation level pressure(liftingCondensationLevelPres)


Back To Table of Contents

DQRID : D071106.1
Start DateStart TimeEnd DateEnd Time
03/16/2007000011/05/20071344
Subject:
FKB/GNDRAD/M1 - Reprocessed: Incorrect calibrations
DataStreams:fkbgndrad20sM1.a0, fkbgndrad60sM1.b1
Description:
An incorrect calibration coefficient was applied to the up_short_hemispheric instrument 
(PSP) for the specified time period.  Changes in the ARM PSP calibration procedure now 
generates two different calibration coefficients for the same instrument, Cf(non "IR 
corrected")and Cfc("IR corrected").  CFc should have been applied to this instrument but CF was 
erroneously used.

The affected PSP is S/N 34292F3 Cf=117.61 (incorrect applied), Cfc(correct)=116.07W/m2/mV

These data were reprocessed and archived in March 2008.
Measurements:fkbgndrad60sM1.b1:
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Pyranometer, Minima(up_short_hemisp_min)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Pyranometer, Standard
    Deviation(up_short_hemisp_std)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Pyranometer, Maxima(up_short_hemisp_max)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Pyranometer(up_short_hemisp)

fkbgndrad20sM1.a0:
  • Instantaneous Upwelling Shortwave Hemispheric Irradiance, Pyranometer(inst_up_short_hemisp)


Back To Table of Contents