Data Quality Reports for Session: 123326 User: dmfoper Completed: 10/02/2009


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D001004.1TWP/MWR/C2 - damaged teflon windowtwpmwrtipC2.a1
D010227.4TWP/MWR/C2 - Excessive wet_window flagstwpmwrlosC2.b1, twpmwrtipC2.a1
D050928.2TWP/MWR/C2 - New software version (4.15) installedtwpmwrlosC2.b1, twpmwrtipC2.a1


DQRID : D001004.1
Start DateStart TimeEnd DateEnd Time
04/10/2000030006/27/20002200
Subject:
TWP/MWR/C2 - damaged teflon window
DataStreams:twpmwrtipC2.a1
Description:
From the RESET11N daily report for 11Apr00:
 At around 15:00 (4/10/00) it was noticed that the 50hz outside power breaker was
 tripped. When it was reset to get the WSI running the MWR heater smoked.
 The heater and cable burned out. The Teflon window sustained damage and
 needs replacing. 

From the RESET11N daily report for 27Jun00:
 22:00GMT(26JUN) - Replaced the MWR window. DM
Measurements:twpmwrtipC2.a1:
  • 31.4 GHz sky signal(tipsky31)
  • 23.8 GHz sky signal(tipsky23)


Back To Table of Contents

DQRID : D010227.4
Start DateStart TimeEnd DateEnd Time
01/16/1999090003/08/19992100
11/16/1999090002/09/20002100
12/07/2000000012/17/20002359
Subject:
TWP/MWR/C2 - Excessive wet_window flags
DataStreams:twpmwrlosC2.b1, twpmwrtipC2.a1
Description:
The wet_window flag is set high more than expected from the measured value of liquid water 
path. This may be caused by dew or, if the heater sensivity is slightly misadjusted, by 
the diurnal variation in ambient temperature.

The "wet_window" flag is an indicator of moisture on the MWR window which invalidates the 
measurements of precipitable water vapor and liquid water path. Sometimes this flag can 
be incorrectly set high or low if the heater sensivity is slightly misadjusted. The heater 
adjustment is a resistive device that is dependent on ambient temperature. Another 
indicator of moisture on the window is unreasonably high ( > 3*liquid_retrieval_rms_accuracy ) 
values of "liq".
Measurements:twpmwrlosC2.b1:
  • Water on Teflon window (1=WET, 0=DRY)(wet_window)

twpmwrtipC2.a1:
  • Water on Teflon window (1=WET, 0=DRY)(wet_window)


Back To Table of Contents

DQRID : D050928.2
Start DateStart TimeEnd DateEnd Time
11/01/2002210109/15/20051424
Subject:
TWP/MWR/C2 - New software version (4.15) installed
DataStreams:twpmwrlosC2.b1, twpmwrtipC2.a1
Description:
A problem began with the installation of MWR.EXE version 4.12 in November 2002. The 
software had been upgraded from a "DOS" to a "Windows"-compiled program to address an earlier 
problem.  The software upgrade corrected the earlier problem but introduced a new one that 
caused line-of-sight observing cycles to be skipped, a 15% reduction in the number of 
tip curves, and saturation of CPU usage.  Software versions 4.13 and 4.14 also produced 
these problems.

The new MWR software version (4.15) was installed on 9/15/2005. As a consequence of this 
upgrade, the tip curve frequency increased. The tip cycle time decreased from ~60s to ~50s.
Measurements:twpmwrlosC2.b1:
  • 31.4 GHz sky signal(sky31)
  • Noise injection temp at nominal temperature at 31.4 GHz(tnd_nom31)
  • Noise injection temp at 31.4 GHz adjusted to tkbb(tnd31)
  • Mixer kinetic (physical) temperature(tkxc)
  • 23.8 GHz blackbody+noise injection signal(bbn23)
  • Blackbody kinetic temperature(tkbb)
  • 23.8 GHz Blackbody signal(bb23)
  • Temperature correction coefficient at 23.8 GHz(tc23)
  • Sky brightness temperature at 23.8 GHz(tbsky23)
  • 31.4 GHz blackbody(bb31)
  • MWR column precipitable water vapor(vap)
  • 23.8 GHz sky signal(sky23)
  • (tknd)
  • Ambient temperature(tkair)
  • IR Brightness Temperature(ir_temp)
  • Noise injection temp at nominal temperature at 23.8 GHz(tnd_nom23)
  • 31.4 GHz blac2body+noise injection signal(bbn31)
  • Sky Infra-Red Temperature(sky_ir_temp)
  • Noise injection temp at 23.8 GHz adjusted to tkbb(tnd23)
  • Temperature correction coefficient at 31.4 GHz(tc31)
  • Sky brightness temperature at 31.4 GHz(tbsky31)
  • Averaged total liquid water along LOS path(liq)

twpmwrtipC2.a1:
  • 31.4 GHz sky signal(tipsky31)
  • (tknd)
  • Blackbody kinetic temperature(tkbb)
  • Noise injection temp at 31.4 GHz derived from this tip(tnd31I)
  • Noise injection temp at 31.4 GHz adjusted to tkbb(tnd31)
  • Noise injection temp at 23.8 GHz adjusted to tkbb(tnd23)
  • 31.4 GHz goodness-of-fit coefficient(r31)
  • 23.8 GHz goodness-of-fit coefficient(r23)
  • Temperature correction coefficient at 23.8 GHz(tc23)
  • Noise injection temp at nominal temperature at 31.4 GHz(tnd_nom31)
  • 23.8 GHz sky signal(tipsky23)
  • 31.4 GHz blac2body+noise injection signal(bbn31)
  • 31.4 GHz blackbody(bb31)
  • Noise injection temp at 23.8 GHz derived from this tip(tnd23I)
  • Noise injection temp at nominal temperature at 23.8 GHz(tnd_nom23)
  • 23.8 GHz blackbody+noise injection signal(bbn23)
  • Temperature correction coefficient at 31.4 GHz(tc31)
  • Mixer kinetic (physical) temperature(tkxc)
  • Ambient temperature(tkair)


Back To Table of Contents