Data Quality Reports for Session: 122344 User: dmfoper Completed: 09/06/2009


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D071217.6SGP/IRT/E15 - Instrument noise problemsgpirt200msE15.a1, sgpirtE15.00, sgpirtE15.b1
D080904.2SGP/IRT/E15 - Communication problemsgpirt200msE15.a1


DQRID : D071217.6
Start DateStart TimeEnd DateEnd Time
07/12/2005153003/28/20081722
Subject:
SGP/IRT/E15 - Instrument noise problem
DataStreams:sgpirt200msE15.a1, sgpirtE15.00, sgpirtE15.b1
Description:
The data was much noisier than other nearby sites (a factor of 4-5 times more). The 
connections and communication and other settings were checked. The IRT was replaced on 
1/29/2008 at 17:37 but the response time was incorrectly set to 0.01 s. This problem was 
corrected when the setting was changed to 0.10 s.
Measurements:sgpirtE15.b1:
  • Sky Infra Red Temperature Minima(sky_ir_temp_min)
  • Sky Infra-Red Temperature(sky_ir_temp)

sgpirt200msE15.a1:
  • Sky Infra-Red Temperature(sky_ir_temp)

sgpirtE15.00:
  • Raw data stream - documentation not supported(raw)


Back To Table of Contents

DQRID : D080904.2
Start DateStart TimeEnd DateEnd Time
08/26/2008130009/26/20081326
Subject:
SGP/IRT/E15 - Communication problem
DataStreams:sgpirt200msE15.a1
Description:
The raw data were truncated multiple times per hour resulting in occasional missing 
values. The remaining data are good. 

The serial cable connection was re-seated.
Measurements:sgpirt200msE15.a1:
  • Internal Reference Infra Red Temperature(ref_ir_temp)
  • Sky Infra-Red Temperature(sky_ir_temp)


Back To Table of Contents



END OF DATA