Data Quality Reports for Session: 121256 User: dmfoper Completed: 08/03/2009


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D071113.3NSA/AERI/C1 - Reprocessed: Failed temp sensor affected calibrationnsaaerich1C1.b1, nsaaerich2C1.b1, nsaaerisummaryC1.b1
D080322.1NSA/AERI/C1 - Reprocessed: Failed temp sensor affected calibrationnsaaerich1C1.b1, nsaaerich2C1.b1, nsaaerisummaryC1.b1


DQRID : D071113.3
Start DateStart TimeEnd DateEnd Time
11/01/2007224611/28/20071930
Subject:
NSA/AERI/C1 - Reprocessed: Failed temp sensor affected calibration
DataStreams:nsaaerich1C1.b1, nsaaerich2C1.b1, nsaaerisummaryC1.b1
Description:
On 11 Nov 2007, a thermistor that measures the temperature of the blackbody support 
structure failed, resulting in erroneously high values.  This measurement, which is captured by 
the field "BBsupportStructureTemp", provides the temperature of the radiance that is 
reflected into the blackbody (i.e., captured by the field "calibrationAmbientTemp"), and 
thus the calibrated radiance (the field "mean_rad") was incorrect and outside the ARM 
specification that the AERI data should be absolutely calibrated to within 1% of the ambient 
radiance.  The broken thermistor was fixed on 28 Nov 2007.

Fortunately, since the emissivity of the blackbodies is so high, absolute accuracy of the 
calibrationAmbientTemp is not required but should be within a few degrees.  The radiance 
data have been recalibrated using the ambient blackbody temperature (the field 
"calibrationCBBtemp") as a surrogate of the calibrationAmbientTemp. Historically, the 
calibrationAmbientTemp and calibrationCBBtemp measurements have been within 1-2 deg. As a result of 
this recalibration, the radiance data are now absolutely calibrated to better than 1%.

The reprocessed data were archived 21 Jan 2008.
Measurements:nsaaerich2C1.b1:
  • Mean of radiance spectra ensemble(mean_rad)

nsaaerisummaryC1.b1:
  • Shortwave window radiance average (2510_2515 cm-1)(shortwaveWindowRadiance2510_2515)
  • Longwave elevated air brightness temperature from radiance average (700_705
    cm-1)(elevatedLayerAirTemp700_705)
  • Shortwave window brightness temperature from radiance average (2510_2515 cm-1)(shortwaveWindowAirTemp2510_2515)
  • Shortwave surface air brightness temperature from radiance average (2295_2300
    cm-1)(surfaceLayerAirTemp2295_2300)
  • Longwave radiance average (675-680 cm-1) Surface Air(surfaceLayerRadiance675_680)
  • Shortwave elevated air brightness temperature from radiance average (2282_2287
    cm-1)(elevatedLayerAirTemp2282_2287)
  • Shortwave radiance average (2282_2287 cm-1) Elevated Air(elevatedLayerRadiance2282_2287)
  • Longwave surface air brightness temperature from radiance average (675_680 cm-1)(surfaceLayerAirTemp675_680)
  • Longwave radiance average (700_705 cm-1) Elevated Air(elevatedLayerRadiance700_705)
  • Longwave window radiance average (985_990 cm-1)(longwaveWindowRadiance985_990)
  • Shortwave radiance average (2295_2300 cm-1) Surface Air(surfaceLayerRadiance2295_2300)
  • Longwave window brightness temperature from radiance average (985_990 cm-1)(longwaveWindowAirTemp985_990)

nsaaerich1C1.b1:
  • Mean of radiance spectra ensemble(mean_rad)


Back To Table of Contents

DQRID : D080322.1
Start DateStart TimeEnd DateEnd Time
12/17/2007000002/13/20082330
Subject:
NSA/AERI/C1 - Reprocessed: Failed temp sensor affected calibration
DataStreams:nsaaerich1C1.b1, nsaaerich2C1.b1, nsaaerisummaryC1.b1
Description:
The blackbody support structure temperature sensor on the NSA.C1 AERI were biased high by 
approximately 25K beginning 20071217 and became progressively worse until the sensor was 
replaced 20080213.  The temperature from this sensor is used to calibrate the radiance 
data for the AERI.  The AERI data have been reprocessed by replacing the data from the 
failed sensor with the cold blackbody temperature sensor data and producing 
corrected/recalibrated radiance data.

The reprocessed data were archived in March 2008.
Measurements:nsaaerich2C1.b1:
  • Mean of radiance spectra ensemble(mean_rad)

nsaaerisummaryC1.b1:
  • Shortwave window radiance average (2510_2515 cm-1)(shortwaveWindowRadiance2510_2515)
  • Longwave elevated air brightness temperature from radiance average (700_705
    cm-1)(elevatedLayerAirTemp700_705)
  • Shortwave window brightness temperature from radiance average (2510_2515 cm-1)(shortwaveWindowAirTemp2510_2515)
  • Shortwave surface air brightness temperature from radiance average (2295_2300
    cm-1)(surfaceLayerAirTemp2295_2300)
  • Longwave radiance average (675-680 cm-1) Surface Air(surfaceLayerRadiance675_680)
  • Shortwave elevated air brightness temperature from radiance average (2282_2287
    cm-1)(elevatedLayerAirTemp2282_2287)
  • Shortwave radiance average (2282_2287 cm-1) Elevated Air(elevatedLayerRadiance2282_2287)
  • Longwave surface air brightness temperature from radiance average (675_680 cm-1)(surfaceLayerAirTemp675_680)
  • Longwave radiance average (700_705 cm-1) Elevated Air(elevatedLayerRadiance700_705)
  • Longwave window radiance average (985_990 cm-1)(longwaveWindowRadiance985_990)
  • Shortwave radiance average (2295_2300 cm-1) Surface Air(surfaceLayerRadiance2295_2300)
  • Longwave window brightness temperature from radiance average (985_990 cm-1)(longwaveWindowAirTemp985_990)

nsaaerich1C1.b1:
  • Mean of radiance spectra ensemble(mean_rad)


Back To Table of Contents



END OF DATA