Data Quality Reports for Session: 120071 User: r1b1skeie Completed: 06/08/2009


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D000203.3SGP/AOS/C1 - CN counter removedsgpaosC1.a0
D000801.4SGP/AOS/C1 - Nephelometer failuresgpaosC1.a0, sgpaosauxC1.a0
D000801.5SGP/AOS/C1 - Nephelometer failuresgpaosC1.a0, sgpaosauxC1.a0
D020709.3SGP/AOS/C1 - AOS Computer FailuresgpaosC1.a0, sgpaosauxC1.a0
D021007.1SGP/AOS/C1 - Ozone instrument out of servicesgpaosC1.a0, sgpaosauxC1.a0
D030210.1SGP/AOS/C1 - Data missingsgpaosC1.00, sgpaosC1.a0, sgpaosauxC1.a0
D040130.1SGP/AOS/C1 - Ozone Instrument removed for repairsgpaosC1.a0, sgpaosauxC1.a0
D041026.3SGP/AOS/C1 - PCASP instrument problemsgpaosC1.a0, sgpaosauxC1.a0
D041026.4SGP/AOS/C1 - Ozone instrument removed for repairsgpaosC1.a0, sgpaosauxC1.a0
D041026.5SGP/AOS/C1 - Condensation Particle Counter failuresgpaosC1.a0, sgpaosauxC1.a0
D041026.6SGP/AOS/C1 - Wet nephelometer problemssgpaosC1.a0, sgpaosauxC1.a0
D050120.2SGP/AOS/C1 - Failed RH sensorsgpaosC1.a0
D050401.1SGP/AOS/C1 - wet nephelometer out of servicesgpaosC1.a0
D051024.4SGP/AOS/C1 - Nephelometer calibration driftsgpaosC1.a0, sgpaosauxC1.a0
D060906.2SGP/AOS/C1 - Missing Condensation Nuclei Counter datasgpaosC1.a0
D070420.1SGP/AOS/C1 - Humidifier controller failuresgpaosC1.00, sgpaosC1.a0, sgpaosauxC1.a0
D070905.2SGP/AOS/C1 - CPC failuresgpaosC1.a1
D090120.5SGP/AOS/C1 - Pump failure and subsequent failure of humidified nephelometersgpaosC1.a1
D980505.1SGP/AOS/C1 - Annual MaintenancesgpaosC1.a0
D980528.1SGP/AOS/C1 - Optical Particle Counter (PCASP-X) downsgpaosC1.a0, sgpaosauxC1.a0


DQRID : D000203.3
Start DateStart TimeEnd DateEnd Time
11/16/1999044801/15/20000000
07/08/2002113108/05/20021424
Subject:
SGP/AOS/C1 - CN counter removed
DataStreams:sgpaosC1.a0
Description:
The AOS condensation particle counter performance degraded to unacceptable
levels and was removed for repair.  

For the first time period: upon inspection of the internal condition of the instrument, it 
was found that oil was present in the intake line.  It is unknown how this oil could 
have arrived in this position.  The instrument was thoroughly cleaned and back-flushed, and 
returned to service during the maintenance visit.

For the second time period: The degradation of began July 8th.  Data collected during this 
time period are either incorrect or missing and should not be used.
Measurements:sgpaosC1.a0:
  • CPC Particle concentration(CPCPartConc)


Back To Table of Contents

DQRID : D000801.4
Start DateStart TimeEnd DateEnd Time
06/20/2000134007/03/20001833
Subject:
SGP/AOS/C1 - Nephelometer failure
DataStreams:sgpaosC1.a0, sgpaosauxC1.a0
Description:
An inexperienced AOS technician inadvertently closed off the flow to the dual-nephelometer 
humidograph system during a Friday Preventative Maintenance.  This resulted in liquid 
water getting into the humidified nephelometer.  Flow was restored to the system on Monday, 
but the nephelometers were producing strange data.  A number of tests were performed 
over the next few days to remotely troubleshoot the situation, to no avail.  Eventually, the 
decision was made to pull both nephelometers and return them to Boulder for repair.  
This DQR refers only to the reference nephelometer, which was out of service for 2 periods 
due to 2 separate problems.  This DQR is for the second period, from 000620 (1340 GMT) to 
000703 (1833 GMT).  Light absorption measurements are invalid whenever the reference 
nephelometer is not operational, because those data are required for correction of the 
absorption data.
Measurements:sgpaosauxC1.a0:
  • TSI Low RH Neph. 550 nm back scattering coefficient for zero gas(GrnBScatZero_LRH)
  • TSI Low RH Neph. 550 nm rayleigh scattering coefficient for zero gas(GrnRScatZero_LRH)
  • TSI Low RH Neph. 700 nm total scattering coefficient for zero gas(RedTScatZero_LRH)
  • TSI Low RH Neph. 700 nm rayleigh scattering coefficient for zero gas(RedRScatZero_LRH)
  • TSI Low RH Neph. 450 nm total scattering coefficient for cal. gas(BluTScatSpn_LRH)
  • TSI Low RH Neph. 450 nm back scattering coefficient for zero gas(BluBScatZero_LRH)
  • TSI Low RH Neph. 700 nm back scattering coefficient for zero gas(RedBScatZero_LRH)
  • TSI Low RH Neph. 450 nm rayleigh scattering coefficient for zero gas(BluRScatZero_LRH)
  • TSI Low RH Neph. 550 nm back scattering coefficient for cal. gas(GrnBScatSpn_LRH)
  • TSI Low RH Neph. 450 nm back scattering coefficient for cal. gas(BluBScatSpn_LRH)
  • TSI Low RH Neph. 450 nm total scattering coefficient for zero gas(BluTScatZero_LRH)
  • TSI Low RH Neph. 550 nm total scattering coefficient for cal. gas(GrnTScatSpn_LRH)
  • TSI Low RH Neph. 700 nm back scattering coefficient for cal. gas(RedBScatSpn_LRH)
  • TSI Low RH Neph. 700 nm total scattering coefficient for cal. gas(RedTScatSpn_LRH)
  • CLAP Signal(BapSignal)
  • CLAP reference signal(BapRef)
  • TSI Low RH Neph. 550 nm total scattering coefficient for zero gas(GrnTScatZero_LRH)

sgpaosC1.a0:
  • TSI Low RH Neph. 550 nm total scat. coef. at 1 um(GrnTScatCoef_1um_LRH)
  • TSI Low RH Neph. 700 nm total scat. coef. at 10 um(RedTScatCoef_10um_LRH)
  • TSI Low RH Neph. 450 nm total scat. coef. at 10 um(BluTScatCoef_10um_LRH)
  • TSI Low RH Neph. 450 nm backscat. coef. at 1 um(BluBScatCoef_1um_LRH)
  • TSI Low RH Neph. 450 nm total scat. coef. at 1 um(BluTScatCoef_1um_LRH)
  • 1 um Absorption coefficient(Bap_I_1um)
  • 10 um Absorption coefficient(Bap_I_10um)
  • TSI Low RH Neph. 700 nm backscat. coef. at 10 um(RedBScatCoef_10um_LRH)
  • TSI Low RH Neph. 700 nm total scat. coef. at 1 um(RedTScatCoef_1um_LRH)
  • TSI Low RH Neph. 550 nm backscat. coef. at 1 um(GrnBScatCoef_1um_LRH)
  • TSI Low RH Neph. 550 nm backscat. coef. at 10 um(GrnBScatCoef_10um_LRH)
  • TSI Low RH Neph. 700 nm backscat. coef. at 1 um(RedBScatCoef_1um_LRH)
  • TSI Low RH Neph. 550 nm total scat. coef. at 10 um(GrnTScatCoef_10um_LRH)
  • TSI Low RH Neph. 450 nm backscat. coef. at 10 um(BluBScatCoef_10um_LRH)


Back To Table of Contents

DQRID : D000801.5
Start DateStart TimeEnd DateEnd Time
06/02/2000184207/03/20001833
Subject:
SGP/AOS/C1 - Nephelometer failure
DataStreams:sgpaosC1.a0, sgpaosauxC1.a0
Description:
An inexperienced AOS technician inadvertently closed off the flow to the dual-nephelometer 
humidograph system during a Friday Preventative Maintenance.  This resulted in liquid 
water getting into the humidified nephelometer.  Flow was restored to the system on Monday, 
but the nephelometers were producing strange data.  A number of tests were performed 
over the next few days to remotely troubleshoot the situation, to no avail.  Eventually, the 
decision was made to pull both nephelometers and return them to Boulder for repair.  
This DQR refers only to the humidified nephelometer, which was out of service for the entire 
period from 000602 through 000703.
Measurements:sgpaosauxC1.a0:
  • TSI High RH Neph. 550 nm back scattering coefficient for zero gas(GrnBScatZero_HRH)
  • TSI High RH Neph. 550 nm total scattering coefficient for zero gas(GrnTScatZero_HRH)
  • TSI High RH Neph. 700 nm total scattering coefficient for cal. gas(RedTScatSpn_HRH)
  • TSI High RH Neph. 550 nm total scattering coefficient for cal. gas(GrnTScatSpn_HRH)
  • TSI High RH Neph. 450 nm total scattering coefficient for cal. gas(BluTScatSpn_HRH)
  • TSI High RH Neph. 700 nm total scattering coefficient for zero gas(RedTScatZero_HRH)
  • TSI High RH Neph. 550 nm rayleigh scattering coefficient for zero gas(GrnRScatZero_HRH)
  • TSI High RH Neph. 450 nm back scattering coefficient for zero gas(BluBScatZero_HRH)
  • TSI High RH Neph. 550 nm back scattering coefficient for cal. gas(GrnBScatSpn_HRH)
  • TSI High RH Neph. 450 nm total scattering coefficient for zero gas(BluTScatZero_HRH)
  • TSI High RH Neph. 450 nm rayleigh scattering coefficient for zero gas(BluRScatZero_HRH)
  • TSI High RH Neph. 700 nm back scattering coefficient for zero gas(RedBScatZero_HRH)
  • TSI High RH Neph. 700 nm back scattering coefficient for cal. gas(RedBScatSpn_HRH)
  • TSI High RH Neph. 700 nm rayleigh scattering coefficient for zero gas(RedRScatZero_HRH)
  • TSI High RH Neph. 450 nm back scattering coefficient for cal. gas(BluBScatSpn_HRH)
  • CLAP Signal(BapSignal)
  • CLAP reference signal(BapRef)

sgpaosC1.a0:
  • TSI High RH Neph. 700 nm total scat. coef. at 1 um(RedTScatCoef_1um_HRH)
  • TSI High RH Neph. 700 nm backscat. coef. at 10 um(RedBScatCoef_10um_HRH)
  • 1 um Absorption coefficient(Bap_I_1um)
  • 10 um Absorption coefficient(Bap_I_10um)
  • TSI High RH Neph. 700 nm backscat. coef. at 1 um(RedBScatCoef_1um_HRH)
  • TSI High RH Neph. 550 nm backscat. coef. at 1 um(GrnBScatCoef_1um_HRH)
  • TSI High RH Neph. 550 nm backscat. coef. at 10 um(GrnBScatCoef_10um_HRH)
  • TSI High RH Neph. 550 nm total scat. coef. at 1 um(GrnTScatCoef_1um_HRH)
  • TSI High RH Neph. 700 nm total scat. coef. at 10 um(RedTScatCoef_10um_HRH)
  • TSI High RH Neph. 450 nm backscat. coef. at 1 um(BluBScatCoef_1um_HRH)
  • TSI High RH Neph. 550 nm total scat. coef. at 10 um(GrnTScatCoef_10um_HRH)
  • TSI High RH Neph. 450 nm total scat. coef. at 10 um(BluTScatCoef_10um_HRH)
  • TSI High RH Neph. 450 nm total scat. coef. at 1 um(BluTScatCoef_1um_HRH)
  • TSI High RH Neph. 450 nm backscat. coef. at 10 um(BluBScatCoef_10um_HRH)


Back To Table of Contents

DQRID : D020709.3
Start DateStart TimeEnd DateEnd Time
05/23/2002201105/26/20022033
05/27/2002023205/28/20021439
05/29/2002025805/29/20021516
05/29/2002215905/30/20021113
05/31/2002231006/04/20021447
Subject:
SGP/AOS/C1 - AOS Computer Failure
DataStreams:sgpaosC1.a0, sgpaosauxC1.a0
Description:
The AOS computer started to experience periodic failures caused by an aging and failing 
system disk.  Replacement of the AOS computer's system disk occurred between 2-4 June 2002.
Measurements:sgpaosauxC1.a0:
  • TSI Low RH Neph. 550 nm back scattering coefficient for zero gas(GrnBScatZero_LRH)
  • TSI High RH Neph. 550 nm total scattering coefficient for zero gas(GrnTScatZero_HRH)
  • Radiance neph zero scattering coefficient(Zero530nm)
  • TSI Low RH Neph. 550 nm rayleigh scattering coefficient for zero gas(GrnRScatZero_LRH)
  • Pressure difference across TSI neph.(Line8DeltaP)
  • TSI High RH Neph. 700 nm total scattering coefficient for zero gas(RedTScatZero_HRH)
  • TSI High RH Neph. 550 nm back scattering coefficient for cal. gas(GrnBScatSpn_HRH)
  • TSI Low RH Neph. 450 nm total scattering coefficient for cal. gas(BluTScatSpn_LRH)
  • unknown(O3_Cont)
  • TSI Low RH Neph. 450 nm rayleigh scattering coefficient for zero gas(BluRScatZero_LRH)
  • TSI Low RH Neph. 450 nm back scattering coefficient for cal. gas(BluBScatSpn_LRH)
  • Relative humidity in main sample manifold heater(Line11RH)
  • Dummy altitude for Zeb(alt)
  • Ozone offset value(O3_Offset)
  • TSI Low RH Neph. 550nm reference intensity(TSISensGrn_LRH)
  • TSI High RH Neph. 450 nm rayleigh scattering coefficient for zero gas(BluRScatZero_HRH)
  • Ozone sample pressure(O3_Press)
  • CLAP flow(BapFlow)
  • Main Sample manifold temperature(Line11Temp)
  • Room temperature(RoomTemp)
  • Ozone span concentration(O3_Spn)
  • Flowrate of zero air line 2(ZeroAirFlow)
  • TSI Low RH Neph. 700 nm total scattering coefficient for cal. gas(RedTScatSpn_LRH)
  • TSI neph. rack temperature(CabinetTemp)
  • Pressure transducer V for line 3 (extra line)(Line3Vac)
  • TSI Low RH Neph. 550 nm total scattering coefficient for zero gas(GrnTScatZero_LRH)
  • TSI High RH Neph. 550 nm back scattering coefficient for zero gas(GrnBScatZero_HRH)
  • TSI High RH Neph. lamp voltage(TSINephLV_HRH)
  • TSI Low RH Neph. 700 nm total scattering coefficient for zero gas(RedTScatZero_LRH)
  • TSI High RH Neph. 450 nm total scattering coefficient for cal. gas(BluTScatSpn_HRH)
  • Pressure transducer V for line 4 (extra line)(Line4Vac)
  • TSI Low RH Neph. 700 nm rayleigh scattering coefficient for zero gas(RedRScatZero_LRH)
  • OPC sample flow(OPCFlow)
  • TSI High RH Neph. 450 nm total scattering coefficient for zero gas(BluTScatZero_HRH)
  • TSI Low RH Neph. 450 nm back scattering coefficient for zero gas(BluBScatZero_LRH)
  • TSI CPC saturator temperature(CPCSaturT)
  • TSI Low RH Neph. 700 nm back scattering coefficient for zero gas(RedBScatZero_LRH)
  • TSI Low RH Neph. 550 nm back scattering coefficient for cal. gas(GrnBScatSpn_LRH)
  • Analog signal from PMS PCASP(OPCLaserVolt)
  • Pressure difference across radiance neph.(Line6DeltaP)
  • TSI Low RH Neph. 700 nm back scattering coefficient for cal. gas(RedBScatSpn_LRH)
  • TSI Low RH Neph. lamp current(TSINephLC_LRH)
  • TSI High RH Neph. 700 nm rayleigh scattering coefficient for zero gas(RedRScatZero_HRH)
  • Ozone line sample temperature (omega temp. sensor)(O3_LinTemp)
  • Temperature in main sample manifold heater(Line11Heater)
  • CLAP reference signal(BapRef)
  • Mass flow output from line 8 (neph. 2 line)(Line8Flow)
  • Transmittance through CLAP filter(BapFiltr)
  • Temperature in back of Whatlow(WatAmbT)
  • OPC sheath(OPCSheath)
  • TSI High RH Neph. 450 nm back scattering coefficient for zero gas(BluBScatZero_HRH)
  • Ozone zero concentration(O3_Zro)
  • Reference object brightness(Calib530nm)
  • lat(lat)
  • Ozone sample frequency(O3_Samp)
  • Time offset of tweaks from base_time(time_offset)
  • TSI Low RH Neph. 550 nm total scattering coefficient for cal. gas(GrnTScatSpn_LRH)
  • Pressure transducer V for line 6 (neph. 1 line)(Line6Vac)
  • TSI High RH Neph. 700 nm back scattering coefficient for zero gas(RedBScatZero_HRH)
  • time(time)
  • TSI Low RH Neph. lamp voltage(TSINephLV_LRH)
  • TSI High RH Neph. 700 nm back scattering coefficient for cal. gas(RedBScatSpn_HRH)
  • Ozone line temp. from thermocouple attached to heater(O3_LinHeater)
  • Main stack air flow(MainAirFlow)
  • TSI CPC condensor temperature(CPCCondT)
  • Temperature between racks 0 and 1(Cab_4_5_Temp)
  • TSI High RH Neph. 700 nm total scattering coefficient for cal. gas(RedTScatSpn_HRH)
  • TSI High RH Neph. 550 nm total scattering coefficient for cal. gas(GrnTScatSpn_HRH)
  • Ozone span value(O3_Span)
  • Pressure transducer V for line 5 (extra line)(Line5Vac)
  • TSI CPC condensor and saturator temperature difference(CPCDeltT)
  • TSI High RH Neph. 550 nm rayleigh scattering coefficient for zero gas(GrnRScatZero_HRH)
  • TSI High RH Neph. lamp current(TSINephLC_HRH)
  • Ozone line temperature set point(O3_SetPt)
  • Temperature of line 8 air (neph. 2 line)(Line8Temp)
  • base time(base_time)
  • TSI High RH Neph. 550nm reference intensity(TSISensGrn_HRH)
  • TSI Low RH Neph. 450 nm total scattering coefficient for zero gas(BluTScatZero_LRH)
  • Ozone sample temperature(O3_Temp)
  • TSI High RH Neph. 450 nm back scattering coefficient for cal. gas(BluBScatSpn_HRH)
  • CLAP Signal(BapSignal)
  • lon(lon)
  • Ozone mode(O3_Mode)

sgpaosC1.a0:
  • CPC Particle concentration(CPCPartConc)
  • lat(lat)
  • PMS PCASP Chan. 7 (0.23 micrometers < Dp < 0.26)(pa23_p26conc)
  • PMS PCASP Chan. 26 (3.00 micrometers < Dp < 3.50)(pa300_p350conc)
  • Temperature in sample line down of heater(SampTemp)
  • PMS PCASP Chan. 0 (Dp > 10 micrometers)(Pa1000Conc)
  • TSI Low RH Neph. 700 nm backscat. coef. at 10 um(RedBScatCoef_10um_LRH)
  • lon(lon)
  • base time(base_time)
  • PMS PCASP Chan. 23 1/cm^3 (2.00 micrometers < Dp < 2.30)(pa200_p230conc)
  • TSI Low RH Neph. relative humidity(TSINephRHSamp_LRH)
  • PMS PCASP Chan. 21 (1.60 micrometers < Dp < 1.80)(pa160_p180conc)
  • PMS PCASP Chan. 9 (0.30 micrometers < Dp < 0.35)(pa30_p35conc)
  • TSI Low RH Neph. 550 nm backscat. coef. at 1 um(GrnBScatCoef_1um_LRH)
  • PMS PCASP Chan. 27 (3.50 micrometers < Dp < 4.00)(pa350_p400conc)
  • PMS PCASP Chan. 13 (0.50 micrometers < Dp < 0.60)(pa50_p60conc)
  • TSI High RH Neph. relative humidity(TSINephRHSamp_HRH)
  • TSI Low RH Neph. 550 nm total scat. coef. at 10 um(GrnTScatCoef_10um_LRH)
  • TSI Low RH Neph. Pressure(TSINephPres_LRH)
  • Time offset of tweaks from base_time(time_offset)
  • PMS PCASP Chan. 5 (0.18 micrometers < Dp < 0.20)(pa18_p20conc)
  • PMS PCASP Chan. 10 (0.35 micrometers < Dp < 0.40)(pa35_p40conc)
  • TSI High RH Neph. 450 nm backscat. coef. at 10 um(BluBScatCoef_10um_HRH)
  • Ozone concentration(Ozone)
  • PMS PCASP Chan. 16 (0.80 micrometers < Dp < 0.90)(pa80_p90conc)
  • TSI High RH Neph. 700 nm total scat. coef. at 1 um(RedTScatCoef_1um_HRH)
  • TSI Low RH Neph. 550 nm total scat. coef. at 1 um(GrnTScatCoef_1um_LRH)
  • TSI Low RH Neph. 700 nm total scat. coef. at 10 um(RedTScatCoef_10um_LRH)
  • TSI Low RH Neph. 450 nm total scat. coef. at 10 um(BluTScatCoef_10um_LRH)
  • TSI Low RH Neph. 450 nm total scat. coef. at 1 um(BluTScatCoef_1um_LRH)
  • Single wave Neph. pressure(Pres530nm)
  • RH upstream of humidograph(HGupstream_RH)
  • PMS PCASP Chan. 30 (6.50 micrometers < Dp < 8.00)(pa650_p800conc)
  • PMS PCASP Chan. 17 (0.90 micrometers < Dp < 1.00)(pa90_p100conc)
  • PMS PCASP Chan. 3 (0.14 micrometers < Dp < 0.16)(pa14_p16conc)
  • TSI High RH Neph. 550 nm total scat. coef. at 1 um(GrnTScatCoef_1um_HRH)
  • TSI High RH Neph. inlet temperature(TSINephTin_HRH)
  • PMS PCASP Chan. 31 (8.00 micrometers < Dp < 10.0)(pa800_p1000conc)
  • TSI High RH Neph. 450 nm total scat. coef. at 1 um(BluTScatCoef_1um_HRH)
  • PMS PCASP Chan. 22 (1.80 micrometers < Dp < 2.00)(pa180_p200conc)
  • PMS PCASP Chan. 24 (2.30 micrometers < Dp < 2.60)(pa230_p260conc)
  • TSI Low RH Neph. 550 nm backscat. coef. at 10 um(GrnBScatCoef_10um_LRH)
  • TSI Low RH Neph. 700 nm backscat. coef. at 1 um(RedBScatCoef_1um_LRH)
  • PMS PCASP Chan. 20 (1.40 micrometers < Dp < 1.60)(pa140_p160conc)
  • PMS PCASP Chan. 14 (0.60 micrometers < Dp < 0.70)(pa60_p70conc)
  • TSI Low RH Neph. 450 nm backscat. coef. at 10 um(BluBScatCoef_10um_LRH)
  • Outdoor air temperature(AmbTemp)
  • Radiance neph. temperature(Neph530Temp)
  • TSI High RH Neph. Pressure(TSINephPres_HRH)
  • PMS PCASP Chan. 15 (0.70 micrometers < Dp < 0.80)(pa70_p80conc)
  • TSI High RH Neph. Sample Temperature(TSINephTSamp_HRH)
  • TSI Low RH Neph. 450 nm backscat. coef. at 1 um(BluBScatCoef_1um_LRH)
  • TSI High RH Neph. 700 nm backscat. coef. at 10 um(RedBScatCoef_10um_HRH)
  • PMS PCASP Chan. 2 (0.12 micrometers < Dp < 0.14)(pa12_p14conc)
  • 1 um Absorption coefficient(Bap_I_1um)
  • PMS PCASP Chan. 18 (1.00 micrometers < Dp < 1.30)(pa100_p130conc)
  • TSI High RH Neph. 550 nm backscat. coef. at 10 um(GrnBScatCoef_10um_HRH)
  • TSI High RH Neph. 700 nm total scat. coef. at 10 um(RedTScatCoef_10um_HRH)
  • Dummy altitude for Zeb(alt)
  • TSI High RH Neph. 550 nm total scat. coef. at 10 um(GrnTScatCoef_10um_HRH)
  • TSI High RH Neph. 450 nm total scat. coef. at 10 um(BluTScatCoef_10um_HRH)
  • TSI Low RH Neph. inlet temperature(TSINephTin_LRH)
  • PMS PCASP Chan. 11 (0.35 micrometers < Dp < 0.40)(pa40_p45conc)
  • PMS PCASP Chan. 29 (5.00 micrometers < Dp < 6.50)(pa500_p650conc)
  • Temp. downstream of humidograph(HGdownstream_T)
  • TSI Low RH Neph. Sample Temperature(TSINephTSamp_LRH)
  • time(time)
  • PMS PCASP Chan. 28 (4.00 micrometers < Dp < 5.00)(pa400_p500conc)
  • PMS PCASP Chan. 25 (2.60 micrometers < Dp < 3.00)(pa260_p300conc)
  • Relative humidity in sample line down. of heater(SampRH)
  • B_sp, at 530 nm from single wave. neph.(Bscat530nm)
  • TSI High RH Neph. 700 nm backscat. coef. at 1 um(RedBScatCoef_1um_HRH)
  • 10 um Absorption coefficient(Bap_I_10um)
  • TSI High RH Neph. 550 nm backscat. coef. at 1 um(GrnBScatCoef_1um_HRH)
  • Temp. upstream of humidograph(HGupstream_T)
  • PMS PCASP Chan. 6 (0.20 micrometers < Dp < 0.23)(pa20_p23conc)
  • PMS PCASP Chan. 19 (1.30 micrometers < Dp < 1.40)(pa130_p140conc)
  • TSI Low RH Neph. 700 nm total scat. coef. at 1 um(RedTScatCoef_1um_LRH)
  • PMS PCASP Chan. 1 1/cm^3 (0.10 micrometers < Dp < 0.12)(pa10_p12conc)
  • TSI High RH Neph. 450 nm backscat. coef. at 1 um(BluBScatCoef_1um_HRH)
  • PMS PCASP Chan. 4 (0.16 micrometers < Dp < 0.18)(pa16_p18conc)
  • PMS PCASP Chan. 8 (0.26 micrometers < Dp < 0.30)(pa26_p30conc)
  • RH downstream of humidograph(HGdownstream_RH)
  • Single wave Neph. RH(RH530nm)
  • PMS PCASP Chan. 12 (0.45 micrometers < Dp < 0.50)(pa45_p50conc)


Back To Table of Contents

DQRID : D021007.1
Start DateStart TimeEnd DateEnd Time
01/30/2002000010/01/20021934
Subject:
SGP/AOS/C1 - Ozone instrument out of service
DataStreams:sgpaosC1.a0, sgpaosauxC1.a0
Description:
The ice storm of 30-31 January, 2002 shut down the AOS for ~2 weeks.  A controlled 
power-up of the system found a problem with the ozone monitor.  Smoke was observed to come from 
the ozone instrument, at which time it was shut down.

The ozone monitor was removed from the AOS for remote troubleshooting. The cause of the 
smoke was identified as being from underneath a printed circuit board.  At that point, Site 
Ops staff were requested to send the unit to Boulder for additional troubleshooting.  
NOAA staff received the unit on 20 February 2002.

After a brief inspection, the cause of the smoking was discovered.  A one-half inch hole 
was burned completely through the motherboard.  The determination was made that this could 
not be repaired at the NOAA facility, and a purchase order was initiated to return the 
unit to Dasibi, Inc. (the manufacturer), for repair/replacement of the motherboard.

On 4 March, 2002, NOAA personnel received the OK from ARM to return the instrument to 
Dasibi, Inc. for service.  Dasibi, Inc., received the unit on or around 8 March, 2002.  The 
manufacturer was to test the unit, repair or replace the damaged motherboard, and 
calibrate the unit according to the NIST standard protocol.

On 8 May, 2002, NOAA personnel emailed SGP Site Ops personnel and asked them to check on 
the status of the repair, which was taking longer then expected.   Upon trying to contact 
the manufacturer, it was discovered that they went out of business.  The return of 
instruments from the manufacturer to customers was being handled by a contract company.  The 
ozone monitor arrived back in Boulder, in an unrepaired state, on or about 1 July 2002.

Since there were few other options, NOAA personnel undertook the difficult task of 
repairing the motherboard.  This required the re-routing of several electrical conduits from the 
burned area in the multilayer board to other acceptable places on the board.  This took 
a considerable amount of electronics technician time and effort.  After the motherboard 
was repaired, extensive testing was required to verify that the repair was done 
successfully and that other problems were not caused in the unit by the board burning..

The repair was deemed successful on 24 September, 2002.  The unit was shipped back to the 
SGP site immediately thereafter.  It was installed during a NOAA AOS maintenance visit on 
1 October, 2002.
Measurements:sgpaosauxC1.a0:
  • Ozone sample frequency(O3_Samp)
  • Ozone offset value(O3_Offset)
  • Ozone sample pressure(O3_Press)
  • Ozone span value(O3_Span)
  • Ozone line temp. from thermocouple attached to heater(O3_LinHeater)
  • Ozone sample temperature(O3_Temp)
  • Ozone line temperature set point(O3_SetPt)
  • Ozone span concentration(O3_Spn)
  • Ozone zero concentration(O3_Zro)
  • Ozone line sample temperature (omega temp. sensor)(O3_LinTemp)
  • unknown(O3_Cont)
  • Ozone mode(O3_Mode)

sgpaosC1.a0:
  • Ozone concentration(Ozone)


Back To Table of Contents

DQRID : D030210.1
Start DateStart TimeEnd DateEnd Time
12/27/2002180512/31/20021543
Subject:
SGP/AOS/C1  - Data missing
DataStreams:sgpaosC1.00, sgpaosC1.a0, sgpaosauxC1.a0
Description:
The UPS supplying power to the AOS computer failed on Friday, 12/27/2002.  The UPS was 
repaired on Monday 12/30/2002, but the AOS computer was locked up and would not reboot.  
Telephone communication was established with Richard Eagan, ANL, who was able to do a remote 
reboot of the computer using his account.
Measurements:sgpaosauxC1.a0:
  • TSI Low RH Neph. 550 nm back scattering coefficient for zero gas(GrnBScatZero_LRH)
  • TSI High RH Neph. 550 nm total scattering coefficient for zero gas(GrnTScatZero_HRH)
  • Radiance neph zero scattering coefficient(Zero530nm)
  • TSI Low RH Neph. 550 nm rayleigh scattering coefficient for zero gas(GrnRScatZero_LRH)
  • Pressure difference across TSI neph.(Line8DeltaP)
  • TSI High RH Neph. 700 nm total scattering coefficient for zero gas(RedTScatZero_HRH)
  • TSI High RH Neph. 550 nm back scattering coefficient for cal. gas(GrnBScatSpn_HRH)
  • TSI Low RH Neph. 450 nm total scattering coefficient for cal. gas(BluTScatSpn_LRH)
  • unknown(O3_Cont)
  • TSI Low RH Neph. 450 nm rayleigh scattering coefficient for zero gas(BluRScatZero_LRH)
  • TSI Low RH Neph. 450 nm back scattering coefficient for cal. gas(BluBScatSpn_LRH)
  • Relative humidity in main sample manifold heater(Line11RH)
  • Dummy altitude for Zeb(alt)
  • Ozone offset value(O3_Offset)
  • TSI Low RH Neph. 550nm reference intensity(TSISensGrn_LRH)
  • TSI High RH Neph. 450 nm rayleigh scattering coefficient for zero gas(BluRScatZero_HRH)
  • Ozone sample pressure(O3_Press)
  • CLAP flow(BapFlow)
  • Main Sample manifold temperature(Line11Temp)
  • Room temperature(RoomTemp)
  • Ozone span concentration(O3_Spn)
  • Flowrate of zero air line 2(ZeroAirFlow)
  • TSI Low RH Neph. 700 nm total scattering coefficient for cal. gas(RedTScatSpn_LRH)
  • TSI neph. rack temperature(CabinetTemp)
  • Pressure transducer V for line 3 (extra line)(Line3Vac)
  • TSI Low RH Neph. 550 nm total scattering coefficient for zero gas(GrnTScatZero_LRH)
  • TSI High RH Neph. 550 nm back scattering coefficient for zero gas(GrnBScatZero_HRH)
  • TSI High RH Neph. lamp voltage(TSINephLV_HRH)
  • TSI Low RH Neph. 700 nm total scattering coefficient for zero gas(RedTScatZero_LRH)
  • TSI High RH Neph. 450 nm total scattering coefficient for cal. gas(BluTScatSpn_HRH)
  • Pressure transducer V for line 4 (extra line)(Line4Vac)
  • TSI Low RH Neph. 700 nm rayleigh scattering coefficient for zero gas(RedRScatZero_LRH)
  • OPC sample flow(OPCFlow)
  • TSI High RH Neph. 450 nm total scattering coefficient for zero gas(BluTScatZero_HRH)
  • TSI Low RH Neph. 450 nm back scattering coefficient for zero gas(BluBScatZero_LRH)
  • TSI CPC saturator temperature(CPCSaturT)
  • TSI Low RH Neph. 700 nm back scattering coefficient for zero gas(RedBScatZero_LRH)
  • TSI Low RH Neph. 550 nm back scattering coefficient for cal. gas(GrnBScatSpn_LRH)
  • Analog signal from PMS PCASP(OPCLaserVolt)
  • Pressure difference across radiance neph.(Line6DeltaP)
  • TSI Low RH Neph. 700 nm back scattering coefficient for cal. gas(RedBScatSpn_LRH)
  • TSI Low RH Neph. lamp current(TSINephLC_LRH)
  • TSI High RH Neph. 700 nm rayleigh scattering coefficient for zero gas(RedRScatZero_HRH)
  • Ozone line sample temperature (omega temp. sensor)(O3_LinTemp)
  • Temperature in main sample manifold heater(Line11Heater)
  • CLAP reference signal(BapRef)
  • Mass flow output from line 8 (neph. 2 line)(Line8Flow)
  • Transmittance through CLAP filter(BapFiltr)
  • Temperature in back of Whatlow(WatAmbT)
  • OPC sheath(OPCSheath)
  • TSI High RH Neph. 450 nm back scattering coefficient for zero gas(BluBScatZero_HRH)
  • Ozone zero concentration(O3_Zro)
  • Reference object brightness(Calib530nm)
  • lat(lat)
  • Ozone sample frequency(O3_Samp)
  • Time offset of tweaks from base_time(time_offset)
  • TSI Low RH Neph. 550 nm total scattering coefficient for cal. gas(GrnTScatSpn_LRH)
  • Pressure transducer V for line 6 (neph. 1 line)(Line6Vac)
  • TSI High RH Neph. 700 nm back scattering coefficient for zero gas(RedBScatZero_HRH)
  • time(time)
  • TSI Low RH Neph. lamp voltage(TSINephLV_LRH)
  • TSI High RH Neph. 700 nm back scattering coefficient for cal. gas(RedBScatSpn_HRH)
  • Ozone line temp. from thermocouple attached to heater(O3_LinHeater)
  • Main stack air flow(MainAirFlow)
  • TSI CPC condensor temperature(CPCCondT)
  • Temperature between racks 0 and 1(Cab_4_5_Temp)
  • TSI High RH Neph. 700 nm total scattering coefficient for cal. gas(RedTScatSpn_HRH)
  • TSI High RH Neph. 550 nm total scattering coefficient for cal. gas(GrnTScatSpn_HRH)
  • Ozone span value(O3_Span)
  • Pressure transducer V for line 5 (extra line)(Line5Vac)
  • TSI CPC condensor and saturator temperature difference(CPCDeltT)
  • TSI High RH Neph. 550 nm rayleigh scattering coefficient for zero gas(GrnRScatZero_HRH)
  • TSI High RH Neph. lamp current(TSINephLC_HRH)
  • Ozone line temperature set point(O3_SetPt)
  • Temperature of line 8 air (neph. 2 line)(Line8Temp)
  • base time(base_time)
  • TSI High RH Neph. 550nm reference intensity(TSISensGrn_HRH)
  • TSI Low RH Neph. 450 nm total scattering coefficient for zero gas(BluTScatZero_LRH)
  • Ozone sample temperature(O3_Temp)
  • TSI High RH Neph. 450 nm back scattering coefficient for cal. gas(BluBScatSpn_HRH)
  • CLAP Signal(BapSignal)
  • lon(lon)
  • Ozone mode(O3_Mode)

sgpaosC1.00:
  • null(Raw data stream - documentation not supported)

sgpaosC1.a0:
  • CPC Particle concentration(CPCPartConc)
  • lat(lat)
  • PMS PCASP Chan. 7 (0.23 micrometers < Dp < 0.26)(pa23_p26conc)
  • PMS PCASP Chan. 26 (3.00 micrometers < Dp < 3.50)(pa300_p350conc)
  • Temperature in sample line down of heater(SampTemp)
  • PMS PCASP Chan. 0 (Dp > 10 micrometers)(Pa1000Conc)
  • TSI Low RH Neph. 700 nm backscat. coef. at 10 um(RedBScatCoef_10um_LRH)
  • lon(lon)
  • base time(base_time)
  • PMS PCASP Chan. 23 1/cm^3 (2.00 micrometers < Dp < 2.30)(pa200_p230conc)
  • TSI Low RH Neph. relative humidity(TSINephRHSamp_LRH)
  • PMS PCASP Chan. 21 (1.60 micrometers < Dp < 1.80)(pa160_p180conc)
  • PMS PCASP Chan. 9 (0.30 micrometers < Dp < 0.35)(pa30_p35conc)
  • TSI Low RH Neph. 550 nm backscat. coef. at 1 um(GrnBScatCoef_1um_LRH)
  • PMS PCASP Chan. 27 (3.50 micrometers < Dp < 4.00)(pa350_p400conc)
  • PMS PCASP Chan. 13 (0.50 micrometers < Dp < 0.60)(pa50_p60conc)
  • TSI High RH Neph. relative humidity(TSINephRHSamp_HRH)
  • TSI Low RH Neph. 550 nm total scat. coef. at 10 um(GrnTScatCoef_10um_LRH)
  • TSI Low RH Neph. Pressure(TSINephPres_LRH)
  • Time offset of tweaks from base_time(time_offset)
  • PMS PCASP Chan. 5 (0.18 micrometers < Dp < 0.20)(pa18_p20conc)
  • PMS PCASP Chan. 10 (0.35 micrometers < Dp < 0.40)(pa35_p40conc)
  • TSI High RH Neph. 450 nm backscat. coef. at 10 um(BluBScatCoef_10um_HRH)
  • Ozone concentration(Ozone)
  • PMS PCASP Chan. 16 (0.80 micrometers < Dp < 0.90)(pa80_p90conc)
  • TSI High RH Neph. 700 nm total scat. coef. at 1 um(RedTScatCoef_1um_HRH)
  • TSI Low RH Neph. 550 nm total scat. coef. at 1 um(GrnTScatCoef_1um_LRH)
  • TSI Low RH Neph. 700 nm total scat. coef. at 10 um(RedTScatCoef_10um_LRH)
  • TSI Low RH Neph. 450 nm total scat. coef. at 10 um(BluTScatCoef_10um_LRH)
  • TSI Low RH Neph. 450 nm total scat. coef. at 1 um(BluTScatCoef_1um_LRH)
  • Single wave Neph. pressure(Pres530nm)
  • RH upstream of humidograph(HGupstream_RH)
  • PMS PCASP Chan. 30 (6.50 micrometers < Dp < 8.00)(pa650_p800conc)
  • PMS PCASP Chan. 17 (0.90 micrometers < Dp < 1.00)(pa90_p100conc)
  • PMS PCASP Chan. 3 (0.14 micrometers < Dp < 0.16)(pa14_p16conc)
  • TSI High RH Neph. 550 nm total scat. coef. at 1 um(GrnTScatCoef_1um_HRH)
  • TSI High RH Neph. inlet temperature(TSINephTin_HRH)
  • PMS PCASP Chan. 31 (8.00 micrometers < Dp < 10.0)(pa800_p1000conc)
  • TSI High RH Neph. 450 nm total scat. coef. at 1 um(BluTScatCoef_1um_HRH)
  • PMS PCASP Chan. 22 (1.80 micrometers < Dp < 2.00)(pa180_p200conc)
  • PMS PCASP Chan. 24 (2.30 micrometers < Dp < 2.60)(pa230_p260conc)
  • TSI Low RH Neph. 550 nm backscat. coef. at 10 um(GrnBScatCoef_10um_LRH)
  • TSI Low RH Neph. 700 nm backscat. coef. at 1 um(RedBScatCoef_1um_LRH)
  • PMS PCASP Chan. 20 (1.40 micrometers < Dp < 1.60)(pa140_p160conc)
  • PMS PCASP Chan. 14 (0.60 micrometers < Dp < 0.70)(pa60_p70conc)
  • TSI Low RH Neph. 450 nm backscat. coef. at 10 um(BluBScatCoef_10um_LRH)
  • Outdoor air temperature(AmbTemp)
  • Radiance neph. temperature(Neph530Temp)
  • TSI High RH Neph. Pressure(TSINephPres_HRH)
  • PMS PCASP Chan. 15 (0.70 micrometers < Dp < 0.80)(pa70_p80conc)
  • TSI High RH Neph. Sample Temperature(TSINephTSamp_HRH)
  • TSI Low RH Neph. 450 nm backscat. coef. at 1 um(BluBScatCoef_1um_LRH)
  • TSI High RH Neph. 700 nm backscat. coef. at 10 um(RedBScatCoef_10um_HRH)
  • PMS PCASP Chan. 2 (0.12 micrometers < Dp < 0.14)(pa12_p14conc)
  • 1 um Absorption coefficient(Bap_I_1um)
  • PMS PCASP Chan. 18 (1.00 micrometers < Dp < 1.30)(pa100_p130conc)
  • TSI High RH Neph. 550 nm backscat. coef. at 10 um(GrnBScatCoef_10um_HRH)
  • TSI High RH Neph. 700 nm total scat. coef. at 10 um(RedTScatCoef_10um_HRH)
  • Dummy altitude for Zeb(alt)
  • TSI High RH Neph. 550 nm total scat. coef. at 10 um(GrnTScatCoef_10um_HRH)
  • TSI High RH Neph. 450 nm total scat. coef. at 10 um(BluTScatCoef_10um_HRH)
  • TSI Low RH Neph. inlet temperature(TSINephTin_LRH)
  • PMS PCASP Chan. 11 (0.35 micrometers < Dp < 0.40)(pa40_p45conc)
  • PMS PCASP Chan. 29 (5.00 micrometers < Dp < 6.50)(pa500_p650conc)
  • Temp. downstream of humidograph(HGdownstream_T)
  • TSI Low RH Neph. Sample Temperature(TSINephTSamp_LRH)
  • time(time)
  • PMS PCASP Chan. 28 (4.00 micrometers < Dp < 5.00)(pa400_p500conc)
  • PMS PCASP Chan. 25 (2.60 micrometers < Dp < 3.00)(pa260_p300conc)
  • Relative humidity in sample line down. of heater(SampRH)
  • B_sp, at 530 nm from single wave. neph.(Bscat530nm)
  • TSI High RH Neph. 700 nm backscat. coef. at 1 um(RedBScatCoef_1um_HRH)
  • 10 um Absorption coefficient(Bap_I_10um)
  • TSI High RH Neph. 550 nm backscat. coef. at 1 um(GrnBScatCoef_1um_HRH)
  • Temp. upstream of humidograph(HGupstream_T)
  • PMS PCASP Chan. 6 (0.20 micrometers < Dp < 0.23)(pa20_p23conc)
  • PMS PCASP Chan. 19 (1.30 micrometers < Dp < 1.40)(pa130_p140conc)
  • TSI Low RH Neph. 700 nm total scat. coef. at 1 um(RedTScatCoef_1um_LRH)
  • PMS PCASP Chan. 1 1/cm^3 (0.10 micrometers < Dp < 0.12)(pa10_p12conc)
  • TSI High RH Neph. 450 nm backscat. coef. at 1 um(BluBScatCoef_1um_HRH)
  • PMS PCASP Chan. 4 (0.16 micrometers < Dp < 0.18)(pa16_p18conc)
  • PMS PCASP Chan. 8 (0.26 micrometers < Dp < 0.30)(pa26_p30conc)
  • RH downstream of humidograph(HGdownstream_RH)
  • Single wave Neph. RH(RH530nm)
  • PMS PCASP Chan. 12 (0.45 micrometers < Dp < 0.50)(pa45_p50conc)


Back To Table of Contents

DQRID : D040130.1
Start DateStart TimeEnd DateEnd Time
11/21/2003163601/27/20042208
Subject:
SGP/AOS/C1 - Ozone Instrument removed for repair
DataStreams:sgpaosC1.a0, sgpaosauxC1.a0
Description:
The ozone instrument was removed for calibration on 2003 11/21.  It was shipped to Boulder 
and was cleaned and calibrated.  It was sent back to SGP on 2003 12/02, but was damaged 
in transit by FedEx.  The instrument was reinstalled on 2003 12/03 but was found to be 
defective.  Large spikes were observed in the data.  Afetr a few days the instrument was 
removed and shipped to a vendor for repair.  The ozone monitor was returned to service 
after repair at 2004 01/27 (2208 GMT).
Measurements:sgpaosauxC1.a0:
  • Ozone sample frequency(O3_Samp)
  • Ozone offset value(O3_Offset)
  • Ozone sample pressure(O3_Press)
  • Ozone span value(O3_Span)
  • Ozone line temp. from thermocouple attached to heater(O3_LinHeater)
  • Ozone sample temperature(O3_Temp)
  • Ozone line temperature set point(O3_SetPt)
  • Ozone span concentration(O3_Spn)
  • Ozone zero concentration(O3_Zro)
  • Ozone line sample temperature (omega temp. sensor)(O3_LinTemp)
  • unknown(O3_Cont)
  • Ozone mode(O3_Mode)

sgpaosC1.a0:
  • Ozone concentration(Ozone)


Back To Table of Contents

DQRID : D041026.3
Start DateStart TimeEnd DateEnd Time
01/10/2004000010/26/20041200
Subject:
SGP/AOS/C1 - PCASP instrument problem
DataStreams:sgpaosC1.a0, sgpaosauxC1.a0
Description:
PCASP (optical particle counter) gave unusually low aerosol signal counts despite numerous 
attempts to fix the instrument by the mentor and manufacturer.
Measurements:sgpaosauxC1.a0:
  • OPC sheath(OPCSheath)
  • OPC sample flow(OPCFlow)
  • Analog signal from PMS PCASP(OPCLaserVolt)

sgpaosC1.a0:
  • PMS PCASP Chan. 7 (0.23 micrometers < Dp < 0.26)(pa23_p26conc)
  • PMS PCASP Chan. 26 (3.00 micrometers < Dp < 3.50)(pa300_p350conc)
  • PMS PCASP Chan. 28 (4.00 micrometers < Dp < 5.00)(pa400_p500conc)
  • PMS PCASP Chan. 15 (0.70 micrometers < Dp < 0.80)(pa70_p80conc)
  • PMS PCASP Chan. 25 (2.60 micrometers < Dp < 3.00)(pa260_p300conc)
  • PMS PCASP Chan. 0 (Dp > 10 micrometers)(Pa1000Conc)
  • PMS PCASP Chan. 2 (0.12 micrometers < Dp < 0.14)(pa12_p14conc)
  • PMS PCASP Chan. 30 (6.50 micrometers < Dp < 8.00)(pa650_p800conc)
  • PMS PCASP Chan. 6 (0.20 micrometers < Dp < 0.23)(pa20_p23conc)
  • PMS PCASP Chan. 19 (1.30 micrometers < Dp < 1.40)(pa130_p140conc)
  • PMS PCASP Chan. 17 (0.90 micrometers < Dp < 1.00)(pa90_p100conc)
  • PMS PCASP Chan. 18 (1.00 micrometers < Dp < 1.30)(pa100_p130conc)
  • PMS PCASP Chan. 3 (0.14 micrometers < Dp < 0.16)(pa14_p16conc)
  • PMS PCASP Chan. 31 (8.00 micrometers < Dp < 10.0)(pa800_p1000conc)
  • PMS PCASP Chan. 1 1/cm^3 (0.10 micrometers < Dp < 0.12)(pa10_p12conc)
  • PMS PCASP Chan. 23 1/cm^3 (2.00 micrometers < Dp < 2.30)(pa200_p230conc)
  • PMS PCASP Chan. 24 (2.30 micrometers < Dp < 2.60)(pa230_p260conc)
  • PMS PCASP Chan. 21 (1.60 micrometers < Dp < 1.80)(pa160_p180conc)
  • PMS PCASP Chan. 22 (1.80 micrometers < Dp < 2.00)(pa180_p200conc)
  • PMS PCASP Chan. 9 (0.30 micrometers < Dp < 0.35)(pa30_p35conc)
  • PMS PCASP Chan. 27 (3.50 micrometers < Dp < 4.00)(pa350_p400conc)
  • PMS PCASP Chan. 13 (0.50 micrometers < Dp < 0.60)(pa50_p60conc)
  • PMS PCASP Chan. 4 (0.16 micrometers < Dp < 0.18)(pa16_p18conc)
  • PMS PCASP Chan. 20 (1.40 micrometers < Dp < 1.60)(pa140_p160conc)
  • PMS PCASP Chan. 8 (0.26 micrometers < Dp < 0.30)(pa26_p30conc)
  • PMS PCASP Chan. 5 (0.18 micrometers < Dp < 0.20)(pa18_p20conc)
  • PMS PCASP Chan. 14 (0.60 micrometers < Dp < 0.70)(pa60_p70conc)
  • PMS PCASP Chan. 10 (0.35 micrometers < Dp < 0.40)(pa35_p40conc)
  • PMS PCASP Chan. 11 (0.35 micrometers < Dp < 0.40)(pa40_p45conc)
  • PMS PCASP Chan. 16 (0.80 micrometers < Dp < 0.90)(pa80_p90conc)
  • PMS PCASP Chan. 12 (0.45 micrometers < Dp < 0.50)(pa45_p50conc)
  • PMS PCASP Chan. 29 (5.00 micrometers < Dp < 6.50)(pa500_p650conc)


Back To Table of Contents

DQRID : D041026.4
Start DateStart TimeEnd DateEnd Time
11/21/2003160001/28/20040000
Subject:
SGP/AOS/C1 - Ozone instrument removed for repair
DataStreams:sgpaosC1.a0, sgpaosauxC1.a0
Description:
The ozone monitor was removed and sent back to the manufacturer for repair.
Measurements:sgpaosauxC1.a0:
  • Ozone sample frequency(O3_Samp)
  • Ozone offset value(O3_Offset)
  • Ozone sample pressure(O3_Press)
  • Ozone span value(O3_Span)
  • Ozone line temp. from thermocouple attached to heater(O3_LinHeater)
  • Ozone sample temperature(O3_Temp)
  • Ozone line temperature set point(O3_SetPt)
  • Ozone span concentration(O3_Spn)
  • Ozone zero concentration(O3_Zro)
  • Ozone line sample temperature (omega temp. sensor)(O3_LinTemp)
  • unknown(O3_Cont)
  • Ozone mode(O3_Mode)

sgpaosC1.a0:
  • Ozone concentration(Ozone)


Back To Table of Contents

DQRID : D041026.5
Start DateStart TimeEnd DateEnd Time
07/28/2004140008/27/20041540
Subject:
SGP/AOS/C1 - Condensation Particle Counter failure
DataStreams:sgpaosC1.a0, sgpaosauxC1.a0
Description:
The CPC signal declined to a low value. The instrument was shipped back to the mentor, 
cleaned and repaired. The instrument inlet was clogged.
Measurements:sgpaosauxC1.a0:
  • TSI CPC saturator temperature(CPCSaturT)
  • TSI CPC condensor and saturator temperature difference(CPCDeltT)
  • TSI CPC condensor temperature(CPCCondT)

sgpaosC1.a0:
  • CPC Particle concentration(CPCPartConc)


Back To Table of Contents

DQRID : D041026.6
Start DateStart TimeEnd DateEnd Time
01/24/2004000002/18/20041800
Subject:
SGP/AOS/C1 - Wet nephelometer problems
DataStreams:sgpaosC1.a0, sgpaosauxC1.a0
Description:
The humidifier tube was coated with mold and had to be replaced. This affects the wet 
nepheleometer measurements and the calculations of the aerosol hygroscopic growth.

The humidifer was cleaned and replaced. A filter was added to the water circulation to 
minimize mold.
Measurements:sgpaosauxC1.a0:
  • TSI High RH Neph. 550 nm back scattering coefficient for zero gas(GrnBScatZero_HRH)
  • TSI High RH Neph. 550 nm total scattering coefficient for zero gas(GrnTScatZero_HRH)
  • TSI High RH Neph. 700 nm total scattering coefficient for cal. gas(RedTScatSpn_HRH)
  • TSI High RH Neph. lamp voltage(TSINephLV_HRH)
  • TSI High RH Neph. 550 nm total scattering coefficient for cal. gas(GrnTScatSpn_HRH)
  • TSI High RH Neph. 450 nm total scattering coefficient for cal. gas(BluTScatSpn_HRH)
  • TSI High RH Neph. 700 nm total scattering coefficient for zero gas(RedTScatZero_HRH)
  • TSI High RH Neph. 550 nm rayleigh scattering coefficient for zero gas(GrnRScatZero_HRH)
  • TSI High RH Neph. lamp current(TSINephLC_HRH)
  • TSI High RH Neph. 550 nm back scattering coefficient for cal. gas(GrnBScatSpn_HRH)
  • TSI High RH Neph. 450 nm back scattering coefficient for zero gas(BluBScatZero_HRH)
  • TSI High RH Neph. 450 nm total scattering coefficient for zero gas(BluTScatZero_HRH)
  • TSI High RH Neph. 550nm reference intensity(TSISensGrn_HRH)
  • TSI High RH Neph. 450 nm rayleigh scattering coefficient for zero gas(BluRScatZero_HRH)
  • TSI High RH Neph. 700 nm back scattering coefficient for zero gas(RedBScatZero_HRH)
  • TSI High RH Neph. 700 nm back scattering coefficient for cal. gas(RedBScatSpn_HRH)
  • TSI High RH Neph. 700 nm rayleigh scattering coefficient for zero gas(RedRScatZero_HRH)
  • TSI High RH Neph. 450 nm back scattering coefficient for cal. gas(BluBScatSpn_HRH)

sgpaosC1.a0:
  • TSI High RH Neph. 700 nm total scat. coef. at 1 um(RedTScatCoef_1um_HRH)
  • Temp. downstream of humidograph(HGdownstream_T)
  • TSI High RH Neph. Pressure(TSINephPres_HRH)
  • TSI High RH Neph. Sample Temperature(TSINephTSamp_HRH)
  • TSI High RH Neph. 700 nm backscat. coef. at 10 um(RedBScatCoef_10um_HRH)
  • RH upstream of humidograph(HGupstream_RH)
  • TSI High RH Neph. 700 nm backscat. coef. at 1 um(RedBScatCoef_1um_HRH)
  • TSI High RH Neph. 550 nm backscat. coef. at 1 um(GrnBScatCoef_1um_HRH)
  • Temp. upstream of humidograph(HGupstream_T)
  • TSI High RH Neph. 550 nm backscat. coef. at 10 um(GrnBScatCoef_10um_HRH)
  • TSI High RH Neph. 550 nm total scat. coef. at 1 um(GrnTScatCoef_1um_HRH)
  • TSI High RH Neph. inlet temperature(TSINephTin_HRH)
  • TSI High RH Neph. 700 nm total scat. coef. at 10 um(RedTScatCoef_10um_HRH)
  • TSI High RH Neph. 450 nm backscat. coef. at 1 um(BluBScatCoef_1um_HRH)
  • TSI High RH Neph. 550 nm total scat. coef. at 10 um(GrnTScatCoef_10um_HRH)
  • TSI High RH Neph. 450 nm total scat. coef. at 10 um(BluTScatCoef_10um_HRH)
  • TSI High RH Neph. 450 nm total scat. coef. at 1 um(BluTScatCoef_1um_HRH)
  • TSI High RH Neph. relative humidity(TSINephRHSamp_HRH)
  • RH downstream of humidograph(HGdownstream_RH)
  • TSI High RH Neph. 450 nm backscat. coef. at 10 um(BluBScatCoef_10um_HRH)


Back To Table of Contents

DQRID : D050120.2
Start DateStart TimeEnd DateEnd Time
01/01/2004000005/01/20052359
Subject:
SGP/AOS/C1 - Failed RH sensor
DataStreams:sgpaosC1.a0
Description:
The RH sensor inside the humidified nephelometer began to slowly decline in value in the 
upper RH range beginning in early 2004 and becoming more noticable poor on 20041117. The 
sensor currently reaches a high RH value of about 60% when the actual RH is closer to 80%.

The sensor was replaced in the Spring of 2005, but the problem remained.  Effective 
1/1/2004, this sensor is not used to calculate scientific data and is therefore essentially 
abandoned in the field.  An end-date of 20050501 is being used as a placeholder.
Measurements:sgpaosC1.a0:
  • TSI High RH Neph. relative humidity(TSINephRHSamp_HRH)


Back To Table of Contents

DQRID : D050401.1
Start DateStart TimeEnd DateEnd Time
03/04/2005000003/22/20051955
Subject:
SGP/AOS/C1 - wet nephelometer out of service
DataStreams:sgpaosC1.a0
Description:
The humidified nephelometer was removed and shipped back to CMDL to test and work on the 
relative humidity sensor inside the instrument.
Measurements:sgpaosC1.a0:
  • TSI High RH Neph. 700 nm total scat. coef. at 1 um(RedTScatCoef_1um_HRH)
  • TSI High RH Neph. 450 nm backscat. coef. at 1 um(BluBScatCoef_1um_HRH)
  • TSI High RH Neph. 550 nm total scat. coef. at 10 um(GrnTScatCoef_10um_HRH)
  • TSI High RH Neph. 450 nm total scat. coef. at 10 um(BluTScatCoef_10um_HRH)
  • Temp. downstream of humidograph(HGdownstream_T)
  • TSI High RH Neph. Pressure(TSINephPres_HRH)
  • TSI High RH Neph. 450 nm total scat. coef. at 1 um(BluTScatCoef_1um_HRH)
  • TSI High RH Neph. Sample Temperature(TSINephTSamp_HRH)
  • TSI High RH Neph. 700 nm backscat. coef. at 10 um(RedBScatCoef_10um_HRH)
  • RH upstream of humidograph(HGupstream_RH)
  • TSI High RH Neph. relative humidity(TSINephRHSamp_HRH)
  • TSI High RH Neph. 700 nm backscat. coef. at 1 um(RedBScatCoef_1um_HRH)
  • Temp. upstream of humidograph(HGupstream_T)
  • TSI High RH Neph. 550 nm backscat. coef. at 1 um(GrnBScatCoef_1um_HRH)
  • RH downstream of humidograph(HGdownstream_RH)
  • TSI High RH Neph. 550 nm backscat. coef. at 10 um(GrnBScatCoef_10um_HRH)
  • TSI High RH Neph. inlet temperature(TSINephTin_HRH)
  • TSI High RH Neph. 550 nm total scat. coef. at 1 um(GrnTScatCoef_1um_HRH)
  • TSI High RH Neph. 700 nm total scat. coef. at 10 um(RedTScatCoef_10um_HRH)
  • TSI High RH Neph. 450 nm backscat. coef. at 10 um(BluBScatCoef_10um_HRH)


Back To Table of Contents

DQRID : D051024.4
Start DateStart TimeEnd DateEnd Time
10/14/2005153610/28/20051940
Subject:
SGP/AOS/C1 - Nephelometer calibration drift
DataStreams:sgpaosC1.a0, sgpaosauxC1.a0
Description:
Bad span check on the humidified nepehelometer.  Instrument was recalibrated and returned 
to service on 10/28/2005 at 1940 UTC.
Measurements:sgpaosauxC1.a0:
  • TSI High RH Neph. 450 nm rayleigh scattering coefficient for zero gas(BluRScatZero_HRH)
  • TSI High RH Neph. 450 nm total scattering coefficient for cal. gas(BluTScatSpn_HRH)
  • TSI High RH Neph. 450 nm back scattering coefficient for zero gas(BluBScatZero_HRH)
  • TSI High RH Neph. 450 nm total scattering coefficient for zero gas(BluTScatZero_HRH)
  • TSI High RH Neph. 450 nm back scattering coefficient for cal. gas(BluBScatSpn_HRH)

sgpaosC1.a0:
  • TSI High RH Neph. 700 nm total scat. coef. at 1 um(RedTScatCoef_1um_HRH)
  • TSI High RH Neph. 450 nm backscat. coef. at 1 um(BluBScatCoef_1um_HRH)
  • TSI High RH Neph. 550 nm total scat. coef. at 10 um(GrnTScatCoef_10um_HRH)
  • TSI High RH Neph. 450 nm total scat. coef. at 10 um(BluTScatCoef_10um_HRH)
  • TSI High RH Neph. Pressure(TSINephPres_HRH)
  • TSI High RH Neph. 450 nm total scat. coef. at 1 um(BluTScatCoef_1um_HRH)
  • TSI High RH Neph. Sample Temperature(TSINephTSamp_HRH)
  • TSI High RH Neph. 700 nm backscat. coef. at 10 um(RedBScatCoef_10um_HRH)
  • RH upstream of humidograph(HGupstream_RH)
  • TSI High RH Neph. relative humidity(TSINephRHSamp_HRH)
  • TSI High RH Neph. 700 nm backscat. coef. at 1 um(RedBScatCoef_1um_HRH)
  • TSI High RH Neph. 550 nm backscat. coef. at 1 um(GrnBScatCoef_1um_HRH)
  • RH downstream of humidograph(HGdownstream_RH)
  • TSI High RH Neph. 550 nm backscat. coef. at 10 um(GrnBScatCoef_10um_HRH)
  • TSI High RH Neph. 550 nm total scat. coef. at 1 um(GrnTScatCoef_1um_HRH)
  • TSI High RH Neph. inlet temperature(TSINephTin_HRH)
  • TSI High RH Neph. 700 nm total scat. coef. at 10 um(RedTScatCoef_10um_HRH)
  • TSI High RH Neph. 450 nm backscat. coef. at 10 um(BluBScatCoef_10um_HRH)


Back To Table of Contents

DQRID : D060906.2
Start DateStart TimeEnd DateEnd Time
08/31/2006135509/11/20061900
Subject:
SGP/AOS/C1 - Missing Condensation Nuclei Counter data
DataStreams:sgpaosC1.a0
Description:
AOS Condensation Nuclei Counter stopped detecting particle counts and was returned to 
mentor for cleaning.  The instrument has small orifices and has a tendency to clog.
Measurements:sgpaosC1.a0:
  • CPC Particle concentration(CPCPartConc)


Back To Table of Contents

DQRID : D070420.1
Start DateStart TimeEnd DateEnd Time
02/27/2007191204/12/20072107
Subject:
SGP/AOS/C1 - Humidifier controller failure
DataStreams:sgpaosC1.00, sgpaosC1.a0, sgpaosauxC1.a0
Description:
Watlow PID controller which controls the humidifier stopped working. This mostly affects 
calculated fits of fRH
Measurements:sgpaosauxC1.a0:
  • TSI Low RH Neph. 550 nm back scattering coefficient for zero gas(GrnBScatZero_LRH)
  • TSI High RH Neph. 550 nm total scattering coefficient for zero gas(GrnTScatZero_HRH)
  • TSI Low RH Neph. 550 nm rayleigh scattering coefficient for zero gas(GrnRScatZero_LRH)
  • TSI High RH Neph. 700 nm total scattering coefficient for zero gas(RedTScatZero_HRH)
  • TSI High RH Neph. 550 nm back scattering coefficient for cal. gas(GrnBScatSpn_HRH)
  • TSI High RH Neph. 450 nm back scattering coefficient for zero gas(BluBScatZero_HRH)
  • TSI Low RH Neph. 450 nm total scattering coefficient for cal. gas(BluTScatSpn_LRH)
  • TSI Low RH Neph. 450 nm back scattering coefficient for cal. gas(BluBScatSpn_LRH)
  • TSI Low RH Neph. 450 nm rayleigh scattering coefficient for zero gas(BluRScatZero_LRH)
  • Relative humidity in main sample manifold heater(Line11RH)
  • TSI Low RH Neph. 550nm reference intensity(TSISensGrn_LRH)
  • TSI Low RH Neph. 550 nm total scattering coefficient for cal. gas(GrnTScatSpn_LRH)
  • TSI High RH Neph. 450 nm rayleigh scattering coefficient for zero gas(BluRScatZero_HRH)
  • TSI High RH Neph. 700 nm back scattering coefficient for zero gas(RedBScatZero_HRH)
  • TSI Low RH Neph. lamp voltage(TSINephLV_LRH)
  • TSI High RH Neph. 700 nm back scattering coefficient for cal. gas(RedBScatSpn_HRH)
  • TSI Low RH Neph. 700 nm total scattering coefficient for cal. gas(RedTScatSpn_LRH)
  • TSI Low RH Neph. 550 nm total scattering coefficient for zero gas(GrnTScatZero_LRH)
  • TSI High RH Neph. 550 nm back scattering coefficient for zero gas(GrnBScatZero_HRH)
  • TSI High RH Neph. 700 nm total scattering coefficient for cal. gas(RedTScatSpn_HRH)
  • TSI High RH Neph. lamp voltage(TSINephLV_HRH)
  • TSI Low RH Neph. 700 nm total scattering coefficient for zero gas(RedTScatZero_LRH)
  • TSI High RH Neph. 550 nm total scattering coefficient for cal. gas(GrnTScatSpn_HRH)
  • TSI High RH Neph. 450 nm total scattering coefficient for cal. gas(BluTScatSpn_HRH)
  • TSI High RH Neph. 550 nm rayleigh scattering coefficient for zero gas(GrnRScatZero_HRH)
  • TSI High RH Neph. lamp current(TSINephLC_HRH)
  • TSI Low RH Neph. 700 nm rayleigh scattering coefficient for zero gas(RedRScatZero_LRH)
  • TSI High RH Neph. 450 nm total scattering coefficient for zero gas(BluTScatZero_HRH)
  • TSI Low RH Neph. 450 nm back scattering coefficient for zero gas(BluBScatZero_LRH)
  • TSI Low RH Neph. 700 nm back scattering coefficient for zero gas(RedBScatZero_LRH)
  • TSI Low RH Neph. 550 nm back scattering coefficient for cal. gas(GrnBScatSpn_LRH)
  • TSI High RH Neph. 550nm reference intensity(TSISensGrn_HRH)
  • TSI Low RH Neph. 450 nm total scattering coefficient for zero gas(BluTScatZero_LRH)
  • TSI Low RH Neph. lamp current(TSINephLC_LRH)
  • TSI Low RH Neph. 700 nm back scattering coefficient for cal. gas(RedBScatSpn_LRH)
  • TSI High RH Neph. 700 nm rayleigh scattering coefficient for zero gas(RedRScatZero_HRH)
  • TSI High RH Neph. 450 nm back scattering coefficient for cal. gas(BluBScatSpn_HRH)

sgpaosC1.00:
  • null(Raw data stream - documentation not supported)

sgpaosC1.a0:
  • TSI High RH Neph. Pressure(TSINephPres_HRH)
  • TSI High RH Neph. Sample Temperature(TSINephTSamp_HRH)
  • TSI Low RH Neph. 450 nm backscat. coef. at 1 um(BluBScatCoef_1um_LRH)
  • TSI High RH Neph. 700 nm backscat. coef. at 10 um(RedBScatCoef_10um_HRH)
  • TSI Low RH Neph. 700 nm backscat. coef. at 10 um(RedBScatCoef_10um_LRH)
  • TSI High RH Neph. 550 nm backscat. coef. at 10 um(GrnBScatCoef_10um_HRH)
  • TSI High RH Neph. 700 nm total scat. coef. at 10 um(RedTScatCoef_10um_HRH)
  • TSI High RH Neph. 550 nm total scat. coef. at 10 um(GrnTScatCoef_10um_HRH)
  • TSI High RH Neph. 450 nm total scat. coef. at 10 um(BluTScatCoef_10um_HRH)
  • TSI Low RH Neph. relative humidity(TSINephRHSamp_LRH)
  • TSI Low RH Neph. inlet temperature(TSINephTin_LRH)
  • TSI Low RH Neph. 550 nm backscat. coef. at 1 um(GrnBScatCoef_1um_LRH)
  • TSI High RH Neph. relative humidity(TSINephRHSamp_HRH)
  • TSI Low RH Neph. 550 nm total scat. coef. at 10 um(GrnTScatCoef_10um_LRH)
  • TSI Low RH Neph. Pressure(TSINephPres_LRH)
  • TSI High RH Neph. 450 nm backscat. coef. at 10 um(BluBScatCoef_10um_HRH)
  • TSI High RH Neph. 700 nm total scat. coef. at 1 um(RedTScatCoef_1um_HRH)
  • TSI Low RH Neph. 550 nm total scat. coef. at 1 um(GrnTScatCoef_1um_LRH)
  • TSI Low RH Neph. Sample Temperature(TSINephTSamp_LRH)
  • TSI Low RH Neph. 700 nm total scat. coef. at 10 um(RedTScatCoef_10um_LRH)
  • Relative humidity in sample line down. of heater(SampRH)
  • TSI Low RH Neph. 450 nm total scat. coef. at 10 um(BluTScatCoef_10um_LRH)
  • TSI Low RH Neph. 450 nm total scat. coef. at 1 um(BluTScatCoef_1um_LRH)
  • RH upstream of humidograph(HGupstream_RH)
  • TSI High RH Neph. 700 nm backscat. coef. at 1 um(RedBScatCoef_1um_HRH)
  • TSI High RH Neph. 550 nm backscat. coef. at 1 um(GrnBScatCoef_1um_HRH)
  • TSI Low RH Neph. 700 nm total scat. coef. at 1 um(RedTScatCoef_1um_LRH)
  • TSI High RH Neph. 550 nm total scat. coef. at 1 um(GrnTScatCoef_1um_HRH)
  • TSI High RH Neph. inlet temperature(TSINephTin_HRH)
  • TSI High RH Neph. 450 nm backscat. coef. at 1 um(BluBScatCoef_1um_HRH)
  • TSI High RH Neph. 450 nm total scat. coef. at 1 um(BluTScatCoef_1um_HRH)
  • TSI Low RH Neph. 550 nm backscat. coef. at 10 um(GrnBScatCoef_10um_LRH)
  • TSI Low RH Neph. 700 nm backscat. coef. at 1 um(RedBScatCoef_1um_LRH)
  • RH downstream of humidograph(HGdownstream_RH)
  • TSI Low RH Neph. 450 nm backscat. coef. at 10 um(BluBScatCoef_10um_LRH)
  • Single wave Neph. RH(RH530nm)


Back To Table of Contents

DQRID : D070905.2
Start DateStart TimeEnd DateEnd Time
07/11/2007094007/24/20072205
08/17/2007214008/30/20071900
Subject:
SGP/AOS/C1 - CPC failure
DataStreams:sgpaosC1.a1
Description:
The Condensation Particle Counter stopped recording counts. Upon inspection the tech found 
that water had condensed in the reservoir. The CPC was shipped back to the mentor for 
repair and a replacement CPC from NOAA was sent to the site.

The replacement Condensation Particle Counter stopped measuring particle counts and was 
removed and returned to NOAA. The thermoelectric cooler in the device had stopped working. 
The ARM CPC that had previously been sent back to NOAA for repair was fixed and returned 
to SGP.
Measurements:sgpaosC1.a1:
  • Condensation Nuclei Concentration Number(N_CPC_1)


Back To Table of Contents

DQRID : D090120.5
Start DateStart TimeEnd DateEnd Time
07/19/2008011201/06/20091912
Subject:
SGP/AOS/C1 - Pump failure and subsequent failure of humidified nephelometer
DataStreams:sgpaosC1.a1
Description:
The pump for the aerosol sampling system failed over a weekend. The humidifier continued 
to operate and soaked the inside of both the reference and wet nephelometers. The 
nephelometers were returned for repair. A replacement neph was installed on 2008-August 5 @1858 
UTC. The nephs were repaired and returned to SGP in December 2008. The humidification 
system was rebuilt and installed on 2009 January 6 at 1912 UTC.
Measurements:sgpaosC1.a1:
  • Aerosol light absorption coefficient, green channel, 10 um size cut(Ba_G_Dry_10um_PSAP1W_1)
  • Absorption Coefficient Red, corrected for flow and sample area, 3 wavelength
    PSAP, 10 um size cut(Ba_R_Dry_10um_PSAP3W_1)
  • Absorption Coefficient Green, corrected for flow and sample area, 3 wavelength
    PSAP, 10 um size cut(Ba_G_Dry_10um_PSAP3W_1)
  • Aerosol backwards-hemispheric light scattering coefficient, green channel, 10 um
    size cut, reference three wavelength nephelometer(Bbs_G_Dry_10um_Neph3W_1)
  • Aerosol backwards-hemispheric light scattering coefficient, red channel, 1 um
    size cut, wet three wavelength nephelometer(Bbs_R_Wet_1um_Neph3W_2)
  • Absorption Coefficient Red, corrected for flow and sample area, 3 wavelength
    PSAP, 1 um size cut(Ba_R_Dry_1um_PSAP3W_1)
  • Aerosol backwards-hemispheric light scattering coefficient, green channel, 1 um
    size cut, wet three wavelength nephelometer(Bbs_G_Wet_1um_Neph3W_2)
  • Aerosol total light scattering coefficient, blue channel, 1 um size cut, wet
    three wavelength nephelometer(Bs_B_Wet_1um_Neph3W_2)
  • Aerosol total light scattering coefficient, green channel, 10 um size cut, wet
    three wavelength nephelometer(Bs_G_Wet_10um_Neph3W_2)
  • Aerosol backwards-hemispheric light scattering coefficient, red channel, 10 um
    size cut, wet three wavelength nephelometer(Bbs_R_Wet_10um_Neph3W_2)
  • Aerosol total light scattering coefficient, green channel, 1 um size cut, wet
    three wavelength nephelometer(Bs_G_Wet_1um_Neph3W_2)
  • Aerosol total light scattering coefficient, red channel, 1 um size cut, wet
    three wavelength nephelometer(Bs_R_Wet_1um_Neph3W_2)
  • Relative humidity of humidograph preheater(RH_preHG)
  • Aerosol total light scattering coefficient, red channel, 1 um size cut,
    reference three wavelength nephelometer(Bs_R_Dry_1um_Neph3W_1)
  • Temperature of humidograph controller(T_postHG)
  • Aerosol total light scattering coefficient, red channel, 10 um size cut, wet
    three wavelength nephelometer(Bs_R_Wet_10um_Neph3W_2)
  • Temperature at wet nephelometer inlet(T_NephInlet_Wet)
  • Relative humidity of humidograph controller(RH_postHG)
  • Aerosol backwards-hemispheric light scattering coefficient, green channel, 1 um
    size cut, reference three wavelength nephelometer(Bbs_G_Dry_1um_Neph3W_1)
  • Aerosol total light scattering coefficient, green channel, 1 um size cut,
    reference three wavelength nephelometer(Bs_G_Dry_1um_Neph3W_1)
  • Relative humidity at wet nephelometer inlet(RH_NephInlet_Wet)
  • Temperature at reference nephelometer inlet(T_NephInlet_Dry)
  • Aerosol backwards-hemispheric light scattering coefficient, blue channel, 10 um
    size cut, wet three wavelength nephelometer(Bbs_B_Wet_10um_Neph3W_2)
  • Absorption Coefficient Green, corrected for flow and sample area, 3 wavelength
    PSAP, 1 um size cut(Ba_G_Dry_1um_PSAP3W_1)
  • Aerosol backwards-hemispheric light scattering coefficient, blue channel, 10 um
    size cut, reference three wavelength nephelometer(Bbs_B_Dry_10um_Neph3W_1)
  • Aerosol backwards-hemispheric light scattering coefficient, blue channel, 1 um
    size cut, reference three wavelength nephelometer(Bbs_B_Dry_1um_Neph3W_1)
  • Relative humidity at reference nephelometer inlet(RH_NephInlet_Dry)
  • Absorption Coefficient Blue, corrected for flow and sample area, 3 wavelength
    PSAP, 10 um size cut(Ba_B_Dry_10um_PSAP3W_1)
  • Aerosol total light scattering coefficient, green channel, 10 um size cut,
    reference three wavelength nephelometer(Bs_G_Dry_10um_Neph3W_1)
  • Temperature of humidograph preheater(T_preHG)
  • Aerosol total light scattering coefficient, blue channel, 10 um size cut, wet
    three wavelength nephelometer(Bs_B_Wet_10um_Neph3W_2)
  • Aerosol backwards-hemispheric light scattering coefficient, blue channel, 1 um
    size cut, wet three wavelength nephelometer(Bbs_B_Wet_1um_Neph3W_2)
  • Aerosol light absorption coefficient, green channel, 1 um size cut(Ba_G_Dry_1um_PSAP1W_1)
  • Aerosol total light scattering coefficient, blue channel, 1 um size cut,
    reference three wavelength nephelometer(Bs_B_Dry_1um_Neph3W_1)
  • Aerosol backwards-hemispheric light scattering coefficient, red channel, 10 um
    size cut, reference three wavelength nephelometer(Bbs_R_Dry_10um_Neph3W_1)
  • Aerosol backwards-hemispheric light scattering coefficient, green channel, 10 um
    size cut, wet three wavelength nephelometer(Bbs_G_Wet_10um_Neph3W_2)
  • Aerosol total light scattering coefficient, red channel, 10 um size cut,
    reference three wavelength nephelometer(Bs_R_Dry_10um_Neph3W_1)
  • Aerosol total light scattering coefficient, blue channel, 10 um size cut,
    reference three wavelength nephelometer(Bs_B_Dry_10um_Neph3W_1)
  • Aerosol backwards-hemispheric light scattering coefficient, red channel, 1 um
    size cut, reference three wavelength nephelometer(Bbs_R_Dry_1um_Neph3W_1)
  • Absorption Coefficient Blue, corrected for flow and sample area, 3 wavelength
    PSAP, 1 um size cut(Ba_B_Dry_1um_PSAP3W_1)


Back To Table of Contents

DQRID : D980505.1
Start DateStart TimeEnd DateEnd Time
08/04/1997145908/08/19971348
04/20/1998144904/23/19982136
07/19/1999145207/30/19991829
01/10/2000140001/15/20000000
09/30/2002130010/04/20021900
03/17/2003154703/21/20032248
Subject:
SGP/AOS/C1  - Annual Maintenance
DataStreams:sgpaosC1.a0
Description:
The AOS Annual maintenance/calibration visit occurred during these periods.  During these 
times, instruments were cleaned, calibrated, and when necessary, repaired.  Instruments 
were online collecting data for some of the time; however, periods of contamination are 
probable because aerosol sampling lines were opened for inspection and cleaning and pumps 
were switched off for periods of time. Any data fields during this period that contain 
data should be considered incorrect.
Measurements:sgpaosC1.a0:
  • CPC Particle concentration(CPCPartConc)
  • PMS PCASP Chan. 7 (0.23 micrometers < Dp < 0.26)(pa23_p26conc)
  • PMS PCASP Chan. 26 (3.00 micrometers < Dp < 3.50)(pa300_p350conc)
  • Temperature in sample line down of heater(SampTemp)
  • PMS PCASP Chan. 15 (0.70 micrometers < Dp < 0.80)(pa70_p80conc)
  • TSI High RH Neph. Pressure(TSINephPres_HRH)
  • TSI High RH Neph. Sample Temperature(TSINephTSamp_HRH)
  • PMS PCASP Chan. 0 (Dp > 10 micrometers)(Pa1000Conc)
  • TSI Low RH Neph. 450 nm backscat. coef. at 1 um(BluBScatCoef_1um_LRH)
  • TSI High RH Neph. 700 nm backscat. coef. at 10 um(RedBScatCoef_10um_HRH)
  • PMS PCASP Chan. 2 (0.12 micrometers < Dp < 0.14)(pa12_p14conc)
  • 1 um Absorption coefficient(Bap_I_1um)
  • PMS PCASP Chan. 18 (1.00 micrometers < Dp < 1.30)(pa100_p130conc)
  • TSI Low RH Neph. 700 nm backscat. coef. at 10 um(RedBScatCoef_10um_LRH)
  • TSI High RH Neph. 550 nm backscat. coef. at 10 um(GrnBScatCoef_10um_HRH)
  • TSI High RH Neph. 700 nm total scat. coef. at 10 um(RedTScatCoef_10um_HRH)
  • TSI High RH Neph. 550 nm total scat. coef. at 10 um(GrnTScatCoef_10um_HRH)
  • PMS PCASP Chan. 23 1/cm^3 (2.00 micrometers < Dp < 2.30)(pa200_p230conc)
  • TSI High RH Neph. 450 nm total scat. coef. at 10 um(BluTScatCoef_10um_HRH)
  • TSI Low RH Neph. relative humidity(TSINephRHSamp_LRH)
  • TSI Low RH Neph. inlet temperature(TSINephTin_LRH)
  • PMS PCASP Chan. 21 (1.60 micrometers < Dp < 1.80)(pa160_p180conc)
  • PMS PCASP Chan. 9 (0.30 micrometers < Dp < 0.35)(pa30_p35conc)
  • TSI Low RH Neph. 550 nm backscat. coef. at 1 um(GrnBScatCoef_1um_LRH)
  • PMS PCASP Chan. 27 (3.50 micrometers < Dp < 4.00)(pa350_p400conc)
  • PMS PCASP Chan. 13 (0.50 micrometers < Dp < 0.60)(pa50_p60conc)
  • TSI High RH Neph. relative humidity(TSINephRHSamp_HRH)
  • TSI Low RH Neph. 550 nm total scat. coef. at 10 um(GrnTScatCoef_10um_LRH)
  • TSI Low RH Neph. Pressure(TSINephPres_LRH)
  • PMS PCASP Chan. 5 (0.18 micrometers < Dp < 0.20)(pa18_p20conc)
  • PMS PCASP Chan. 10 (0.35 micrometers < Dp < 0.40)(pa35_p40conc)
  • PMS PCASP Chan. 11 (0.35 micrometers < Dp < 0.40)(pa40_p45conc)
  • TSI High RH Neph. 450 nm backscat. coef. at 10 um(BluBScatCoef_10um_HRH)
  • Ozone concentration(Ozone)
  • PMS PCASP Chan. 16 (0.80 micrometers < Dp < 0.90)(pa80_p90conc)
  • PMS PCASP Chan. 29 (5.00 micrometers < Dp < 6.50)(pa500_p650conc)
  • TSI High RH Neph. 700 nm total scat. coef. at 1 um(RedTScatCoef_1um_HRH)
  • TSI Low RH Neph. 550 nm total scat. coef. at 1 um(GrnTScatCoef_1um_LRH)
  • Temp. downstream of humidograph(HGdownstream_T)
  • TSI Low RH Neph. Sample Temperature(TSINephTSamp_LRH)
  • PMS PCASP Chan. 28 (4.00 micrometers < Dp < 5.00)(pa400_p500conc)
  • PMS PCASP Chan. 25 (2.60 micrometers < Dp < 3.00)(pa260_p300conc)
  • TSI Low RH Neph. 700 nm total scat. coef. at 10 um(RedTScatCoef_10um_LRH)
  • Relative humidity in sample line down. of heater(SampRH)
  • B_sp, at 530 nm from single wave. neph.(Bscat530nm)
  • TSI Low RH Neph. 450 nm total scat. coef. at 10 um(BluTScatCoef_10um_LRH)
  • TSI Low RH Neph. 450 nm total scat. coef. at 1 um(BluTScatCoef_1um_LRH)
  • Single wave Neph. pressure(Pres530nm)
  • RH upstream of humidograph(HGupstream_RH)
  • TSI High RH Neph. 700 nm backscat. coef. at 1 um(RedBScatCoef_1um_HRH)
  • 10 um Absorption coefficient(Bap_I_10um)
  • TSI High RH Neph. 550 nm backscat. coef. at 1 um(GrnBScatCoef_1um_HRH)
  • PMS PCASP Chan. 6 (0.20 micrometers < Dp < 0.23)(pa20_p23conc)
  • Temp. upstream of humidograph(HGupstream_T)
  • PMS PCASP Chan. 19 (1.30 micrometers < Dp < 1.40)(pa130_p140conc)
  • PMS PCASP Chan. 30 (6.50 micrometers < Dp < 8.00)(pa650_p800conc)
  • PMS PCASP Chan. 17 (0.90 micrometers < Dp < 1.00)(pa90_p100conc)
  • PMS PCASP Chan. 3 (0.14 micrometers < Dp < 0.16)(pa14_p16conc)
  • TSI Low RH Neph. 700 nm total scat. coef. at 1 um(RedTScatCoef_1um_LRH)
  • TSI High RH Neph. 550 nm total scat. coef. at 1 um(GrnTScatCoef_1um_HRH)
  • TSI High RH Neph. inlet temperature(TSINephTin_HRH)
  • PMS PCASP Chan. 31 (8.00 micrometers < Dp < 10.0)(pa800_p1000conc)
  • PMS PCASP Chan. 1 1/cm^3 (0.10 micrometers < Dp < 0.12)(pa10_p12conc)
  • TSI High RH Neph. 450 nm backscat. coef. at 1 um(BluBScatCoef_1um_HRH)
  • TSI High RH Neph. 450 nm total scat. coef. at 1 um(BluTScatCoef_1um_HRH)
  • PMS PCASP Chan. 22 (1.80 micrometers < Dp < 2.00)(pa180_p200conc)
  • PMS PCASP Chan. 24 (2.30 micrometers < Dp < 2.60)(pa230_p260conc)
  • TSI Low RH Neph. 550 nm backscat. coef. at 10 um(GrnBScatCoef_10um_LRH)
  • TSI Low RH Neph. 700 nm backscat. coef. at 1 um(RedBScatCoef_1um_LRH)
  • PMS PCASP Chan. 4 (0.16 micrometers < Dp < 0.18)(pa16_p18conc)
  • PMS PCASP Chan. 20 (1.40 micrometers < Dp < 1.60)(pa140_p160conc)
  • PMS PCASP Chan. 8 (0.26 micrometers < Dp < 0.30)(pa26_p30conc)
  • RH downstream of humidograph(HGdownstream_RH)
  • PMS PCASP Chan. 14 (0.60 micrometers < Dp < 0.70)(pa60_p70conc)
  • TSI Low RH Neph. 450 nm backscat. coef. at 10 um(BluBScatCoef_10um_LRH)
  • Outdoor air temperature(AmbTemp)
  • Single wave Neph. RH(RH530nm)
  • Radiance neph. temperature(Neph530Temp)
  • PMS PCASP Chan. 12 (0.45 micrometers < Dp < 0.50)(pa45_p50conc)


Back To Table of Contents

DQRID : D980528.1
Start DateStart TimeEnd DateEnd Time
07/01/1997093607/03/19971556
08/01/1997132908/08/19971348
05/05/1998092606/16/19981505
06/09/1999082407/26/19992205
09/08/2000042610/13/20001630
05/17/2002060205/20/20051847
05/23/2002201105/31/20021258
09/09/2002000010/07/20021905
Subject:
SGP/AOS/C1 - Optical Particle Counter (PCASP-X) down
DataStreams:sgpaosC1.a0, sgpaosauxC1.a0
Description:
The AOS Optical Particle Spectrometer (PCASP-X sizing probe) was down during these 
periods.  This lowers the laser reference voltage and results in inaccurate sizing.
Measurements:sgpaosauxC1.a0:
  • OPC sheath(OPCSheath)
  • OPC sample flow(OPCFlow)
  • Analog signal from PMS PCASP(OPCLaserVolt)

sgpaosC1.a0:
  • PMS PCASP Chan. 7 (0.23 micrometers < Dp < 0.26)(pa23_p26conc)
  • PMS PCASP Chan. 26 (3.00 micrometers < Dp < 3.50)(pa300_p350conc)
  • PMS PCASP Chan. 28 (4.00 micrometers < Dp < 5.00)(pa400_p500conc)
  • PMS PCASP Chan. 15 (0.70 micrometers < Dp < 0.80)(pa70_p80conc)
  • PMS PCASP Chan. 25 (2.60 micrometers < Dp < 3.00)(pa260_p300conc)
  • PMS PCASP Chan. 0 (Dp > 10 micrometers)(Pa1000Conc)
  • PMS PCASP Chan. 2 (0.12 micrometers < Dp < 0.14)(pa12_p14conc)
  • PMS PCASP Chan. 30 (6.50 micrometers < Dp < 8.00)(pa650_p800conc)
  • PMS PCASP Chan. 6 (0.20 micrometers < Dp < 0.23)(pa20_p23conc)
  • PMS PCASP Chan. 19 (1.30 micrometers < Dp < 1.40)(pa130_p140conc)
  • PMS PCASP Chan. 17 (0.90 micrometers < Dp < 1.00)(pa90_p100conc)
  • PMS PCASP Chan. 18 (1.00 micrometers < Dp < 1.30)(pa100_p130conc)
  • PMS PCASP Chan. 3 (0.14 micrometers < Dp < 0.16)(pa14_p16conc)
  • PMS PCASP Chan. 31 (8.00 micrometers < Dp < 10.0)(pa800_p1000conc)
  • PMS PCASP Chan. 1 1/cm^3 (0.10 micrometers < Dp < 0.12)(pa10_p12conc)
  • PMS PCASP Chan. 23 1/cm^3 (2.00 micrometers < Dp < 2.30)(pa200_p230conc)
  • PMS PCASP Chan. 24 (2.30 micrometers < Dp < 2.60)(pa230_p260conc)
  • PMS PCASP Chan. 21 (1.60 micrometers < Dp < 1.80)(pa160_p180conc)
  • PMS PCASP Chan. 22 (1.80 micrometers < Dp < 2.00)(pa180_p200conc)
  • PMS PCASP Chan. 9 (0.30 micrometers < Dp < 0.35)(pa30_p35conc)
  • PMS PCASP Chan. 27 (3.50 micrometers < Dp < 4.00)(pa350_p400conc)
  • PMS PCASP Chan. 13 (0.50 micrometers < Dp < 0.60)(pa50_p60conc)
  • PMS PCASP Chan. 4 (0.16 micrometers < Dp < 0.18)(pa16_p18conc)
  • PMS PCASP Chan. 20 (1.40 micrometers < Dp < 1.60)(pa140_p160conc)
  • PMS PCASP Chan. 8 (0.26 micrometers < Dp < 0.30)(pa26_p30conc)
  • PMS PCASP Chan. 5 (0.18 micrometers < Dp < 0.20)(pa18_p20conc)
  • PMS PCASP Chan. 14 (0.60 micrometers < Dp < 0.70)(pa60_p70conc)
  • PMS PCASP Chan. 10 (0.35 micrometers < Dp < 0.40)(pa35_p40conc)
  • PMS PCASP Chan. 11 (0.35 micrometers < Dp < 0.40)(pa40_p45conc)
  • PMS PCASP Chan. 16 (0.80 micrometers < Dp < 0.90)(pa80_p90conc)
  • PMS PCASP Chan. 12 (0.45 micrometers < Dp < 0.50)(pa45_p50conc)
  • PMS PCASP Chan. 29 (5.00 micrometers < Dp < 6.50)(pa500_p650conc)


Back To Table of Contents