Data Quality Reports for Session: 119357 User: lovejoon82 Completed: 05/04/2009


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D050803.4PYE/NFOV/M1 - Intermittent missing datapyenfov2chM1.00, pyenfov2chM1.b1
D050803.5PYE/NFOV/M1 - Intermittent data gapspyenfov2chM1.00, pyenfov2chM1.b1
D050811.6PYE/SONDE/M1 - Broken RH sensorpyesondewnpnM1.b1
D050823.1PYE/SONDE/M1 - Bad temperature and RHpyesondewnpnM1.b1
D050912.3PYE/SONDE/M1 - Bad RH sensorpyesondewnpnM1.b1
D050912.4PYE/SONDE/M1 - Bad RH sensorpyesondewnpnM1.b1


DQRID : D050803.4
Start DateStart TimeEnd DateEnd Time
07/18/2005190009/10/20050055
Subject:
PYE/NFOV/M1 - Intermittent missing data
DataStreams:pyenfov2chM1.00, pyenfov2chM1.b1
Description:
During the Pt Reyes deployment of the NFOV, the instrument experienced intermittent data 
gaps due to communications problems.  Data gaps occurred for the following date ranges:
20050718-20050723
20050727-20050730
20050803-20050804
20050813-20050816
20050820-20050822
20050826-20050827
20050830-20050902
20050908-20050910
Measurements:pyenfov2chM1.00:
  • null(Raw data stream - documentation not supported)

pyenfov2chM1.b1:
  • Calibrated Zenith Radiance, 870 nm(radiance_870nm)
  • Tube Temperature(tube_temp)
  • Raw counts from instrument before calibration, 673 nm(raw_counts_673nm)
  • time(time)
  • Cosine of Solar Zenith Angle(cosine_solar_zenith_angle)
  • Raw counts from instrument before calibration, 870 nm(raw_counts_870nm)
  • base time(base_time)
  • Calibrated Zenith Radiance, 673 nm(radiance_673nm)
  • Detector Temperature(head_temp)
  • Time offset of tweaks from base_time(time_offset)


Back To Table of Contents

DQRID : D050803.5
Start DateStart TimeEnd DateEnd Time
07/18/2005190007/30/20051200
Subject:
PYE/NFOV/M1 - Intermittent data gaps
DataStreams:pyenfov2chM1.00, pyenfov2chM1.b1
Description:
Intermittent data collection of the NFOV.  Data collection not successful regularly during 
day light hours but generally caught up during the nighttime hours.  Occasionally the 
system stopped collecting altogether and a reboot of the network components was required to 
restart collection. Data gaps are largest 18 through 23 of July and 27 through 30 July.
Measurements:pyenfov2chM1.00:
  • null(Raw data stream - documentation not supported)

pyenfov2chM1.b1:
  • Tube Temperature(tube_temp)
  • Calibrated Zenith Radiance, 870 nm(radiance_870nm)
  • Dummy altitude for Zeb(alt)
  • lat(lat)
  • Cosine of Solar Zenith Angle(cosine_solar_zenith_angle)
  • base time(base_time)
  • lon(lon)
  • Time offset of tweaks from base_time(time_offset)
  • Raw counts from instrument before calibration, 673 nm(raw_counts_673nm)
  • time(time)
  • Raw counts from instrument before calibration, 870 nm(raw_counts_870nm)
  • Calibrated Zenith Radiance, 673 nm(radiance_673nm)
  • Detector Temperature(head_temp)


Back To Table of Contents

DQRID : D050811.6
Start DateStart TimeEnd DateEnd Time
08/02/2005112908/02/20051225
Subject:
PYE/SONDE/M1 - Broken RH sensor
DataStreams:pyesondewnpnM1.b1
Description:
One of the RH sensors on this radiosonde was bad.  The RH oscillates rapidly between near 
0 and ambient up to approximately 11 km.  The near-zero values are incorrect.
Measurements:pyesondewnpnM1.b1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)


Back To Table of Contents

DQRID : D050823.1
Start DateStart TimeEnd DateEnd Time
06/30/2005052806/30/20050711
Subject:
PYE/SONDE/M1 - Bad temperature and RH
DataStreams:pyesondewnpnM1.b1
Description:
The temperature appears to be approximately 10 degC too high during the entire sounding.  
This also causes the RH to be too low.  Most likely the problem was a bad temperature 
sensor.
Measurements:pyesondewnpnM1.b1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)
  • Dry bulb temperature(tdry)


Back To Table of Contents

DQRID : D050912.3
Start DateStart TimeEnd DateEnd Time
08/31/2005113008/31/20051140
Subject:
PYE/SONDE/M1 - Bad RH sensor
DataStreams:pyesondewnpnM1.b1
Description:
The RH sensor heating circuit failed on this sounding.  RH values during the first part of 
the sounding oscillate between a reasonable value and a very low value.  It appears that 
(a) the heating was out of phase, and (b) incorrect values were being reported during 
the first 10 minutes or so of flight.
Measurements:pyesondewnpnM1.b1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)


Back To Table of Contents

DQRID : D050912.4
Start DateStart TimeEnd DateEnd Time
08/29/2005112808/29/20051340
Subject:
PYE/SONDE/M1 - Bad RH sensor
DataStreams:pyesondewnpnM1.b1
Description:
Two problems here - failure of the RH sensor heating circuit and a bad RH sensor.
Measurements:pyesondewnpnM1.b1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)


Back To Table of Contents



END OF DATA