D080409.1NSA/NIMFR/C1 - Tube heater malfunctionnsanimfrC1.a0, nsanimfrC1.b1
D090129.4TWP/SMET/C3 - Reprocess: Station Pressure values reported in hPa instead of kPa.twpsmet60sC3.b1


DQRID : D080409.1
Start DateStart TimeEnd DateEnd Time
03/06/2008150003/29/20080000
Subject:
NSA/NIMFR/C1 - Tube heater malfunction
DataStreams:nsanimfrC1.a0, nsanimfrC1.b1
Description:
The instrument was installed after winter calibration and the tube heater was found to be 
operating erratically. The purpose of the heating the tube is to keep condensation from 
forming on the window.

Though erratic, the tube temperature was generally warm enough that no condensation should 
have formed.  There was never any evidence of this in the irradiance data.  Since it is 
not believed to have affected the adversely, this DQR is being assigned as transparent.
Measurements:nsanimfrC1.a0:
  • Direct Normal Narrowband Irradiance, Filter 5, raw data as read(direct_normal_narrowband_filter5_raw)
  • Direct Normal Narrowband Irradiance, Filter 2(direct_normal_narrowband_filter2)
  • Direct Normal Narrowband Irradiance, Filter 5(direct_normal_narrowband_filter5)
  • Direct Normal Narrowband Irradiance, Filter 1, raw data as read(direct_normal_narrowband_filter1_raw)
  • Direct Normal Narrowband Irradiance, Filter 3, raw data as read(direct_normal_narrowband_filter3_raw)
  • Direct Normal Narrowband Irradiance, Filter 3(direct_normal_narrowband_filter3)
  • Direct Normal Narrowband Irradiance, Filter 4(direct_normal_narrowband_filter4)
  • Direct Normal Narrowband Irradiance, Filter 1(direct_normal_narrowband_filter1)
  • Direct Normal Narrowband Irradiance, Filter 6, raw data as read(direct_normal_narrowband_filter6_raw)
  • Direct Normal Narrowband Irradiance, Filter 2, raw data as read(direct_normal_narrowband_filter2_raw)
  • Direct Normal Broadband Irradiance, raw data as read(direct_normal_broadband_raw)
  • Direct Normal Narrowband Irradiance, Filter 6(direct_normal_narrowband_filter6)
  • Direct Normal Narrowband Irradiance, Filter 4, raw data as read(direct_normal_narrowband_filter4_raw)
  • Direct Normal Broadband Irradiance broadband scale applied(direct_normal_broadband)

nsanimfrC1.b1:
  • Narrowband Direct Horizontal Irradiance, Filter 6(direct_horizontal_narrowband_filter6)
  • Direct Normal Broadband Irradiance broadband scale applied(direct_normal_broadband)
  • Narrowband Direct Horizontal Irradiance, Filter 5(direct_horizontal_narrowband_filter5)
  • Narrowband Direct Horizontal Irradiance, Filter 1(direct_horizontal_narrowband_filter1)
  • Direct Normal Narrowband Irradiance, Filter 4(direct_normal_narrowband_filter4)
  • Direct Normal Narrowband Irradiance, Filter 2(direct_normal_narrowband_filter2)
  • Direct Normal Narrowband Irradiance, Filter 5(direct_normal_narrowband_filter5)
  • Narrowband Direct Horizontal Irradiance, Filter 3(direct_horizontal_narrowband_filter3)
  • Direct Normal Narrowband Irradiance, Filter 6(direct_normal_narrowband_filter6)
  • Direct Normal Narrowband Irradiance, Filter 1(direct_normal_narrowband_filter1)
  • Direct Normal Narrowband Irradiance, Filter 3(direct_normal_narrowband_filter3)
  • Narrowband Direct Horizontal Irradiance, Filter 2(direct_horizontal_narrowband_filter2)
  • Narrowband Direct Horizontal Irradiance, Filter 4(direct_horizontal_narrowband_filter4)
  • Direct Horizontal Broadband Irradiance, broadband scale applied(direct_horizontal_broadband)


Back To Table of Contents

DQRID : D090129.4
Start DateStart TimeEnd DateEnd Time
01/27/2009185001/30/20091535
Subject:
TWP/SMET/C3 - Reprocess: Station Pressure values reported in hPa instead of kPa.
DataStreams:twpsmet60sC3.b1
Description:
An incorrect program was entered into the datalogger.  This program did not convert the 
station pressure from hPa to kPa.  This made the values larger by a factor of 10 and caused 
the QC bit to be tripped.