Data Quality Reports for Session: 119016 User: lindsay1m1 Completed: 04/18/2009


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D060508.1SGP/SIRS/E16 - NIP Instrument Failuresgpsirs20sE16.a0, sgpsirsE16.b1


DQRID : D060508.1
Start DateStart TimeEnd DateEnd Time
04/28/2006154305/12/20061608
Subject:
SGP/SIRS/E16 - NIP Instrument Failure
DataStreams:sgpsirs20sE16.a0, sgpsirsE16.b1
Description:
The short_direct_normal dropped below MIN value on 4/28/06 resulting in failure of the 
up_long_hemisp started 3-component test.  No data are available from 20060428.1900 through 
20060503.21:39.  When the data returned, the short_direct normal fluctuated wildly 
throughout the day, and down_long_hemisp & up_long_hemisp failed the 3-component test.  
Technicians determined the NIP had failed and replaced it 5/12.

Use component method ((global - diffuse)/cos(ZA)) to calculate direct normal. For more 
information on this summation method please see section 7.2 (Theory of Operation) of the 
SIRS Handbook:
http://www.arm.gov/publications/tech_reports/handbooks/sirs_handbook.pdf
Measurements:sgpsirsE16.b1:
  • Shortwave Direct Normal Irradiance, Pyrgeometer(short_direct_normal)
  • Shortwave Direct Normal Irradiance, Pyrgeometer, Standard Deviation(short_direct_normal_std)
  • Shortwave Direct Normal Irradiance, Pyrgeometer, Maxima(short_direct_normal_max)
  • Shortwave Direct Normal Irradiance, Pyrgeometer, Minima(short_direct_normal_min)

sgpsirs20sE16.a0:
  • Instantaneous Direct Normal Shortwave Irradiance, Pyheliometer Thermopile
    Voltage(inst_direct_normal)


Back To Table of Contents