Data Quality Reports for Session: 118960 User: mark1pleyd Completed: 04/17/2009


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D070416.1FKB/MFRSR/M1 - Reprocessed: Calibration errorfkbmfrsrM1.a0, fkbmfrsrM1.b1
D080703.9FKB/MFRSR/M1 - Incorrect alltime unitsfkbmfrsrM1.b1


DQRID : D070416.1
Start DateStart TimeEnd DateEnd Time
03/21/2007133004/18/20070700
Subject:
FKB/MFRSR/M1 - Reprocessed: Calibration error
DataStreams:fkbmfrsrM1.a0, fkbmfrsrM1.b1
Description:
Incorrect calibrations were originally applied to these data.  The data were reprocessed 
and archived in December 2007.

There is a problem with channel 673.  Numerous drop outs occur throughout the period 
covered by this DQR.
Measurements:fkbmfrsrM1.b1:
  • Direct Normal Narrowband Irradiance, Filter 4(direct_normal_narrowband_filter4)
  • Narrowband Direct Horizontal Irradiance, Filter 4(direct_horizontal_narrowband_filter4)
  • Narrowband Hemispheric Irradiance, Filter 4(hemisp_narrowband_filter4)
  • Cosine correction applied to direct_horizontal_narrowband_filter4(computed_cosine_correction_filter4)
  • Ratio of direct_normal_narrowband_filter4 to diffuse_hemisp_narrowband_filter4(direct_diffuse_ratio_filter4)
  • Narrowband Diffuse Hemispheric Irradiance, Filter 4(diffuse_hemisp_narrowband_filter4)
  • Alltime Narrowband Hemispheric Irradiance, Filter 4(alltime_hemisp_narrowband_filter4)

fkbmfrsrM1.a0:
  • Alltime Narrowband Hemispheric Irradiance, Filter 4(alltime_hemisp_narrowband_filter4)
  • Narrowband Diffuse Hemispheric Irradiance, Filter 4(diffuse_hemisp_narrowband_filter4)
  • Narrowband Direct Horizontal Irradiance, Filter 4(direct_horizontal_narrowband_filter4)
  • Narrowband Hemispheric Irradiance, Filter 4(hemisp_narrowband_filter4)
  • Diffuse Narrowband Hemispheric Irradiance, Filter 4(diffuse_hemisp_narrowband_filter4_raw)
  • Narrowband Hemispheric Irradiance, Filter 4(hemisp_narrowband_filter4_raw)


Back To Table of Contents

DQRID : D080703.9
Start DateStart TimeEnd DateEnd Time
03/21/2007000001/01/20082359
Subject:
FKB/MFRSR/M1 - Incorrect alltime units
DataStreams:fkbmfrsrM1.b1
Description:
Units of alltime_hemisp_broadband and alltime_hemisp_narrowband_filter(N) are incorrect.  
They should be counts.  A low priority reprocessing task has been submitted to correct 
this metadata error.
Measurements:fkbmfrsrM1.b1:
  • Alltime Narrowband Hemispheric Irradiance, Filter 6(alltime_hemisp_narrowband_filter6)
  • Alltime Narrowband Hemispheric Irradiance, Filter 1(alltime_hemisp_narrowband_filter1)
  • Alltime Narrowband Hemispheric Irradiance, Filter 3(alltime_hemisp_narrowband_filter3)
  • Alltime Narrowband Hemispheric Irradiance, Filter 5(alltime_hemisp_narrowband_filter5)
  • Alltime Narrowband Hemispheric Irradiance, Filter 4(alltime_hemisp_narrowband_filter4)
  • Alltime Hemispheric Broadband Irradiance(alltime_hemisp_broadband)
  • Alltime Narrowband Hemispheric Irradiance, Filter 2(alltime_hemisp_narrowband_filter2)


Back To Table of Contents



END OF DATA