Data Quality Reports for Session: 118938 User: dmfoper Completed: 04/16/2009


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D080523.2SGP/EBBR/E2 - Net Radiation and Fluxes Incorrectsgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1
D080616.3SGP/EBBR/E2 - AEM Ventilator Dangling - Sensible and Latent Heat Fluxes Incorrectsgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1


DQRID : D080523.2
Start DateStart TimeEnd DateEnd Time
05/02/2008163005/15/20082000
Subject:
SGP/EBBR/E2 - Net Radiation and Fluxes Incorrect
DataStreams:sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1
Description:
The net radiometer was tipped 37 degrees eastward, resulting in apparent
solar noon being 2.5 hours early.  The net radiation error results in incorrect sensible 
and latent heat fluxes.
Measurements:sgp5ebbrE2.b1:
  • net_radiation

sgp15ebbrE2.b1:
  • net_radiation_mv

sgp30ebbrE2.b1:
  • net_radiation
  • latent_heat_flux
  • sensible_heat_flux


Back To Table of Contents

DQRID : D080616.3
Start DateStart TimeEnd DateEnd Time
06/07/2008003006/12/20081730
Subject:
SGP/EBBR/E2 - AEM Ventilator Dangling - Sensible and Latent Heat Fluxes Incorrect
DataStreams:sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1
Description:
The right AEM ventilator unit with temperature and humidity sensors inside was dangling 
from the AEM.  Therefore, the gradients of temperature and relative humidity and thus 
sensible and latent heat flux are suspect for this period.  Often, the fluxes look reasonable.
Measurements:sgp15ebbrE2.b1:
  • temp_trh_right_rr
  • rh_right_mv
  • temp_air_right

sgp5ebbrE2.b1:
  • vapor_pressure_bottom
  • rh_bottom_fraction
  • rh_top_fraction
  • temp_trh_top
  • vapor_pressure_top
  • temp_air_bottom
  • temp_air_top
  • temp_trh_bottom

sgp30ebbrE2.b1:
  • bowen_ratio
  • rh_top_fraction
  • rh_bottom_fraction
  • sensible_heat_flux
  • temp_air_bottom
  • temp_trh_top
  • temp_trh_bottom
  • temp_air_top
  • latent_heat_flux
  • vapor_pressure_bottom
  • vapor_pressure_top


Back To Table of Contents



END OF DATA