Data Quality Reports for Session: 118839 User: lindsay1m1 Completed: 04/10/2009


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D040907.1SGP/SIRS/C1/E13 - Incorrect calibration coefficientssgpsirs20sC1.a0, sgpsirs20sE13.a0, sgpsirsC1.00, sgpsirsC1.b1, sgpsirsE13.00,
sgpsirsE13.b1
D060630.21SGP/SIRS/E13 - Reprocessed: Longwave Calibration errorsgpsirsE13.b1


DQRID : D040907.1
Start DateStart TimeEnd DateEnd Time
08/27/2004183009/07/20042130
Subject:
SGP/SIRS/C1/E13 - Incorrect calibration coefficients
DataStreams:sgpsirs20sC1.a0, sgpsirs20sE13.a0, sgpsirsC1.00, sgpsirsC1.b1, sgpsirsE13.00,
sgpsirsE13.b1
Description:
After the DIR changeout on 8/27, the DLH measurements at the central facility vary 
significantly between pyrgeometers by as much as 65 Wm^(-2).  It was discovered that the two 
downwelling pyrgeometers destined for C1 and E13 may have been swapped during installation.  
The instrument mentor is unsure of applied calibration coefficients during this DQR's 
time period.  Backing out new downwelling longwave irradiances would be difficult in this 
case. Fortunately, the problem period is less than a fortnight.
Measurements:sgpsirsE13.b1:
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer, Standard
    Deviation(down_long_hemisp_shaded_std)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer, Maxima(down_long_hemisp_shaded_max)
  • Instantaneous Downwelling Pyrgeometer Case Thermistor Temperature, Shaded
    Pyrgeometer(inst_down_long_shaded_case_temp)
  • Downwelling Longwave Hemispheric Net Infrared(down_long_netir)
  • Instantaneous Downwelling Pyrgeometer Dome Thermistor Temperature, Shaded
    Pyrgeometer(inst_down_long_shaded_dome_temp)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer(down_long_hemisp_shaded)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer, Minima(down_long_hemisp_shaded_min)

sgpsirsC1.b1:
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer, Standard
    Deviation(down_long_hemisp_shaded_std)
  • Downwelling Longwave Hemispheric Net Infrared(down_long_netir)
  • Instantaneous Downwelling Pyrgeometer Dome Thermistor Temperature, Shaded
    Pyrgeometer(inst_down_long_shaded_dome_temp)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer, Maxima(down_long_hemisp_shaded_max)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer(down_long_hemisp_shaded)
  • Instantaneous Downwelling Pyrgeometer Case Thermistor Temperature, Shaded
    Pyrgeometer(inst_down_long_shaded_case_temp)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer, Minima(down_long_hemisp_shaded_min)

sgpsirsC1.00:
  • null(Raw data stream - documentation not supported)

sgpsirs20sE13.a0:
  • Instantaneous Downwelling Pyrgeometer Thermopile Voltage, Shaded Pyrgeometer(inst_down_long_hemisp_shaded_tp)
  • Instantaneous Downwelling Pyrgeometer Case Thermistor Resistance, Shaded
    Pyrgeometer(inst_down_long_shaded_case_resist)
  • Instantaneous Downwelling Pyrgeometer Dome Thermistor Resistance, Shaded
    Pyrgeometer(inst_down_long_shaded_dome_resist)

sgpsirs20sC1.a0:
  • Instantaneous Downwelling Pyrgeometer Case Thermistor Resistance, Shaded
    Pyrgeometer(inst_down_long_shaded_case_resist)
  • Instantaneous Downwelling Pyrgeometer Dome Thermistor Resistance, Shaded
    Pyrgeometer(inst_down_long_shaded_dome_resist)
  • Instantaneous Downwelling Pyrgeometer Thermopile Voltage, Shaded Pyrgeometer(inst_down_long_hemisp_shaded_tp)

sgpsirsE13.00:
  • null(Raw data stream - documentation not supported)


Back To Table of Contents

DQRID : D060630.21
Start DateStart TimeEnd DateEnd Time
12/11/2003193502/16/20061905
Subject:
SGP/SIRS/E13 - Reprocessed: Longwave Calibration error
DataStreams:sgpsirsE13.b1
Description:
Modified pyrgeometer calibration procedures were implemented beginning in December 2003. 
These modified procedures introduced a calibration bias in the longwave data. The previous 
procedures were re-implemented at all sites between December 2005 and February 2006 to 
restore proper calibrations. 

The data collected while the incorrect procedures were in place have been reprocessed to 
remove the calibration bias.  The reprocessed 60 second averaged data are based on 3 
instantaneous 20 second data records rather than on 60 1 second instantaneous data records. 
Still, these data are considered far superior to the originally processed data. The 
reprocessed data were archived in October 2006.
Measurements:sgpsirsE13.b1:
  • Upwelling Longwave Hemispheric Net Infrared(up_long_netir)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer(down_long_hemisp_shaded)
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Pyrgeometer(up_long_hemisp)
  • Downwelling Longwave Hemispheric Net Infrared(down_long_netir)


Back To Table of Contents



END OF DATA