DQR ID | Subject | Data Streams Affected |
---|---|---|
D070807.1 | FKB/RWP/M1 - Amplifier failure | fkb1290rwpwindconM1.a1, fkb1290rwpwindmomM1.a0, fkb1290rwpwindspecM1.a0 |
D070809.3 | FKB/RWP/M1 - Chaff released in Murg Valley | fkb1290rwpM1.00, fkb1290rwprawM1.00, fkb1290rwpwindconM1.a1, fkb1290rwpwindmomM1.a0, fkb1290rwpwindspecM1.a0 |
D071106.1 | FKB/GNDRAD/M1 - Reprocessed: Incorrect calibrations | fkbgndrad20sM1.a0, fkbgndrad60sM1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/26/2007 | 0912 | 08/03/2007 | 1539 |
Subject: | FKB/RWP/M1 - Amplifier failure |
DataStreams: | fkb1290rwpwindconM1.a1, fkb1290rwpwindmomM1.a0, fkb1290rwpwindspecM1.a0 |
Description: | The RWP stopped getting signals, presumably due to failure of the final amplifier. System was turned off on Aug 3, 2007 and no further data were collected. |
Measurements: | fkb1290rwpwindmomM1.a0:
fkb1290rwpwindconM1.a1:
fkb1290rwpwindspecM1.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/26/2007 | 1307 | 07/26/2007 | 1400 |
07/30/2007 | 1307 | 07/30/2007 | 1400 |
Subject: | FKB/RWP/M1 - Chaff released in Murg Valley |
DataStreams: | fkb1290rwpM1.00, fkb1290rwprawM1.00, fkb1290rwpwindconM1.a1, fkb1290rwpwindmomM1.a0, fkb1290rwpwindspecM1.a0 |
Description: | As part of the research activities associated with the COPS field campaign, the research center at Karlsruhe released aluminum chaff (2.5 cm in length) in the Murg Valley in order to study atmospheric diffusion. It is possible that some of this chaff drifted over the AMF site and thus may have influenced the radar observations (as spurious echos). |
Measurements: | fkb1290rwpM1.00:
fkb1290rwpwindmomM1.a0:
fkb1290rwpwindconM1.a1:
fkb1290rwprawM1.00:
fkb1290rwpwindspecM1.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/16/2007 | 0000 | 11/05/2007 | 1344 |
Subject: | FKB/GNDRAD/M1 - Reprocessed: Incorrect calibrations |
DataStreams: | fkbgndrad20sM1.a0, fkbgndrad60sM1.b1 |
Description: | An incorrect calibration coefficient was applied to the up_short_hemispheric instrument (PSP) for the specified time period. Changes in the ARM PSP calibration procedure now generates two different calibration coefficients for the same instrument, Cf(non "IR corrected")and Cfc("IR corrected"). CFc should have been applied to this instrument but CF was erroneously used. The affected PSP is S/N 34292F3 Cf=117.61 (incorrect applied), Cfc(correct)=116.07W/m2/mV These data were reprocessed and archived in March 2008. |
Measurements: | fkbgndrad60sM1.b1:
fkbgndrad20sM1.a0:
|